LaMotte DC1100 User Manual

Page 1
MODEL DC1100
Instruction Manual
Helping People Solve Analytical Challenges
800-344-3100 • 410-778-3100 (Outside U.S.A.) • Fax: 410-778-6394
LaMOTTE COMPANY
PO Box 329 • Chestertown • Maryland • 21620 • USA
Visit us on the web at www.lamotte.com
®
Page 2
Most test substances in water are colorless and undetectable to the human eye. In order to test for their presence, we must find a way to “see” them. The LaMotte colorimeter can be used to measure any test substance that is itself colored or can be reacted to produce a color. In fact, a simple definition of colorimetry is “the measurement of color” and a colorimetric method is “any technique used to evaluate an unknown color in reference to known colors.” In a colorimetric chemical test, the intensity of the color from the reaction must be proportional to the concentration of the substance being tested.
A colorimeter can be used to photoelectrically measure the amount of colored light absorbed by a colored sample in reference to a colorless sample (blank).
White light is made up of many different colors or wavelengths of light. A colored sample typically absorbs only one color or one band of wavelengths from the white light. Little difference could be measured between white light before it passes through a colored sample versus after it passes through. The reason for this is that the one color absorbed by the sample is only a small portion of the total amount of light passing through the sample. However, if we could select only that one color or band of wavelengths of light to which the test sample is most sensitive, we would see a large difference between the light before it passes through the sample and after it passes through. A colorimeter passes a white light beam through an optical filter which transmits only one particular color or band of wavelengths of light to the photodetector where it is measured. The difference in the amount of colored light transmitted by a colorless sample (blank) and the amount of colored light transmitted by a colored sample is a measurement of the amount of colored light absorbed by the sample. In most colorimetric tests the amount of colored light absorbed is directly proportional to the concentration of the test factor producing the color and the path length through the sample. However, for some tests the amount of colored light absorbed is inversely proportional to the concentration.
The choice of the correct optical filter and therefore the correct color or wavelength of light is important. It is interesting to note that the filter that gives the most sensitive calibration for your test factor is the complementary color of the test sample. For example, the Nitrate-Nitrogen test produces a pink color proportional to the nitrate concentration in the sample (the greater the nitrate concentration, the darker the pink color). A green filter is used since a pinkish-red solution absorbs mostly green light.
2
Page 3
THE MODEL DC1100 COLORIMETER
The DC1100 colorimeter has been specially calibrated to read the test result directly in parts per million (ppm) of a desired substance (i.e. ppm chlorine, ppm phosphate). The sensitivity of the DC1100’s response has been maximized by selecting a specific, complimentary light source for each test. The light chosen is absorbed by the color of the test solution. For example, a green light is used for reactions that will produce a red sample. Since green light is absorbed in a red solution, the detector can respond to minute variations in the intensity of the transmitted light.
MODEL DC1100 SINGLE-PARAMETER TEST KITS
The DC1100 can be supplied with reagents for a variety of tests, including wastewater monitoring, industrial process control, boiler and cooling water testing, pool and spa water testing, public health inspection, environmental studies, aquaculture, hydroponics, and other general water analysis applications. The meter, reagents, glassware and accessories are packaged in foam in a compact carrying case for easy transportation and field use.
SPECIFICATIONS
MODEL:
INSTRUMENT TYPE:
DC1100 Single wavelength, direct reading colorimeter Readout: 3
1
digit LCD
2
Reading Stability: Within 1% of final value within 5 seconds
PHOTOMETRIC DATA:
Source: Light emitting diode Detector: Filtered photo-diode
Configuration: Round 21 mm diameter sample
tubes (flat bottom) indexed in chamber
CONTROLS:
ON/OFF SET BLANK
POWER REQUIREMENTS:
Battery Operated: Field replaceable NEDA #1604A, 9 volt
Line Operation: 120/220V, 50/60Hz, 2VA with optional adapter
METER HOUSING:
Dimensions: (HxWxD),150mmx80mm
3
3
“x1
x 45 mm, 6" x
4
4
3
Page 4
GENERAL OPERATING INSTRUCTIONS
K
1. Fill the water sample bottle (0688) with the sample water.
2. Fill colorimeter tube (0967) to 10 mL line with water from sample bottle.
The bottom of the curved surface, the meniscus, should rest on the top of the line.
3. Cap the colorimeter tube and insert into the DC1100 colorimeter chamber.
NOTE: Dry all colorimeter tubes with a tissue before inserting into the
chamber. Also be sure that there are no scratches on the tubes. Keep the chamber clean and dry at all times.
4. Press the “ON/OFF” button and adjust the SET BLANK knob until the
display reads 0. Once this adjustment has been made, do not move the SET BLANK knob for the remainder off the test.
Low Battery Indicator
ON/OFF Button
NOTE:
The location of the decimal point may vary for different meters.
~
BA T
SET BLAN Knob
Sample Chamber
4
Page 5
5. For best results, use the same tube that was used to adjust the blank for all
measurements. If the test you are performing requires a wait of more than thirty seconds, turning the meter off will extend battery life.
6. Follow individual test instructions to add reagents to the tube.
7. Cap the treated sample tube, making sure the tube is clean and dry, then
insert the tube into the chamber.
8. Press the “ON/OFF” button if the meter has been turned off. Record
reading directly from the display.
TEST SAMPLE DILUTIONS AND VOLUMETRIC MEASUREMENTS
If the indicated reading is higher than the maximum concentration specified on the reagent specifications, a dilution should be done to ensure accuracy. In rare cases of extremely high concentrations, the display may indicate an over-range condition: the left digit will be a “1,” the other digits will remain blank. The sample must then be diluted with demineralized water and the test repeated to bring the concentration within a range that can be accurately measured by the meter. The reading is then multiplied by the appropriate dilution factor to obtain the test result.
EXAMPLE:
Measure 5 mL of the original sample into a graduated cylinder. Dilute to 10 mL with distilled or demineralized water. The sample has now been diluted by half, and the test results are multiplied by 2.
The following dilution table shows the amount of original sample to be diluted, the amount of distilled or demineralized water required, and the dilution factor. Since the tests are based on a 10 mL sample, the dilutions must be measured accurately to assure reliable results.
Distilled Water to Bring
Size of Sample
Volume to 10 mL Multiplication Factor
10 mL 0 mL 1
5mL 5mL 2
2.5 mL 7.5 mL 4
1 mL 9 mL 10
0.5 mL 9.5 mL 20
NOTE: When measuring small sample sizes, use graduated cylinders or pipets,
and measure the final dilution in a graduated cylinder. Pipet the sample water into the graduated cylinder, then fill the cylinder to the 10 mL line with distilled or demineralized water.
5
Page 6
REAGENT BLANK
For greater accuracy, the test procedure should be run on a 10 mL sample of deionized water, known as the reagent blank. This procedure will compensate for color or turbidity that the reagents may produce. Subtract this result from the test result to receive final reading.
REPLACING THE BATTERY
When the battery indicator appears on the display, the battery should be replaced.
1. Turn meter over and remove four screws from the meter base.
2. Remove old battery and insert new battery. The DC1100 uses a Type
1604A 9 volt Alkaline battery.
CONNECTING THE AC ADAPTER
To connect meter to a power source, plug the connector of the AC adapter into the small hole at the back of the meter. Plug large end into an electrical outlet (120V, 60Hz).
WARRANTY INFORMATION
REPAIRS
If it is necessary to return the instrument for repair, contact LaMotte Company at 1-800-344-3100 for a return authorization number.
INSTRUMENT GUARANTEE
This instrument is guaranteed to be free of defects in material and workmanship for one year from date of original purchase. If, in that time, it is found to be defective, it will be repaired without charge, except for transportation expenses. This guarantee does not cover the batteries.
This guarantee is void under the following circumstances:
operator’s negligence
improper application
unauthorized servicing
LIMITS OF LIABILITY
Under no circumstances shall LaMotte Company be liable for loss of life, profits, or other damages incurred through the use of misuse of their products.
PACKAGING AND DELIVERY
Experienced packaging personnel at LaMotte Company assure adequate protection against normal hazards encountered during shipping. After the product leaves the manufacturer, all responsibility for its safe delivery is assured by the transporter. Damage claims must be filed immediately with the transporter to receive compensation for damaged goods.
6
Page 7
TROUBLE SHOOTING GUIDE
PROBLEM CHECK ACTION
Meter won’t come on
^ AC Adapter? See pg. 6 of manual.
^ AC Wall Outlet? Verify power source.
^ Contact LaMotte for
^
Comes on, but readings drift
^ Are there suspended solids
^ Do reading drift up scale? Bring samples to room temperature.
^ Readings drift down scale
Battery? Replace, see pg. 6 of manual.
return authorization.
Are instrument grade reagents being used?
in sample?
with cold sample?
Return to LaMotte for repair.
Obtain instrument grade reagents.
Filter both blank and sample (after reaction). NOTE: Do not use filtered blank for sample reaction.
Check for sample tube fogging, let sample equilibrate to room temperature.
Suspect Calibration
^ Are instrument grade
^ How old are the reagents? Contact LaMotte to verify age.
^ Check calibration with
^ Is class “A” volumetric
Will meter blank? (zero) If no, see next page.
reagents being used?
standards?
glassware being used to make up the standards?
Obtain instrument grade reagents.
How old is the standard?
If not, obtain and see Standard Methods.
7
Page 8
TROUBLE SHOOTING GUIDE (continued)
PROBLEM CHECK ACTION
Suspect Calibration
^ Verify with another
^ Contact LaMotte for
^
Meter won’t zero Sample tubes free of dirt
^ Sample tubes clean? Clean tubes often with soap and a test
^ Possible internal meter
^ LED contaminated with
^ High suspended solids in
Verify standard with alternate test method?
meter?
return authorization.
& scratches?
components subjected to moisture?
dirt and deposits?
sample blank?
Run alternate test on standard.
Check other meter calibrations.
Return for Calibration check.
Check & replace if necessary.
tube brush. Do not leave reacted samples in tubes, staining may occur.
Always dry tubes before inserting. Examine tube chamber for visible moisture or chemical stains. If yes, remove 4 screws from back, split case & dry with hair dryer under low heat (10 ­15 min).
Clean square LED and Photodetector w/cotton swab & small amount of alcohol.
Filter with coffee filter, or similar, to remove most solids. (Be sure to filter reacted samples as well) NOTE: Do not use filtered blank for sample reaction.
Meter read 1 ­w/o samples in chamber
Meter read 1 ­only when samples in chamber
Meter over range? Contact LaMotte for return
authorization & return.
Meter over range? Extremely high concentration, dilute
sample to see if 1 - condition returns to normal condition.
12/95
Loading...