LaMotte DA-1-1929 User Manual

Page 1
Conductivity
Meter
CODE 1704
MODEL DA-LR
EPA
COMPLIANTCOMPLIANT
LaMOTTE COM PANY
Helping Peo ple Solve An a lyt i cal Challenges
PO Box 329 • Chestertown • Mary land • 21620 • USA
800-344-3100 • 410-778-3100 (Out side U.S.A.) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
SM
4/97
Page 2
INTRODUCTION
The LaMotte Multi-Range Conductivity Meter is a portable, battery-operated instrument with solid state circuitry. The meter accurately measures a wide range of conductivity with a single probe. A conductivity reading measures the capacity of a solution to conduct an electric current and is directly related to the total ionic concentration of dissolved substances in the solution.
The instrument’s extensive capabilities range from checking the purity of distilled water to use as a salinometer for monitoring the conductivity of sea water. Other applications include checking air and water pollution, evaluating electroplating baths, measuring soluble salts in soils, and monitoring industrial and municipal water treatment operations.
This easy-to-use instrument has a rugged 3a” taut-band meter movement with a mirrored scale that eliminates reading errors due to parallax. Other features include a battery check circuit and manual temperature compensation from 0°C to 100°C.
The conductivity meter is equipped with a range selector which monitors conductivities from 0-10,000 micromhos/cm in four different ranges: 0-10; 0-100; 0-1000; and 0-10,000 micromhos/cm. A single conductivity probe on a twelve foot cable is used to measure all four ranges. An optional interchangeable probe is available with fifty feet of cable for measuring conductivity profiles in deep water.
Should it be necessary to return the instrument for repair or servicing, pack the instrument carefully in suitable container with adequate packing material. A return authorization number must be obtained from LaMotte Company by calling 1-800-344-3100. Attach a letter with the authorization number to the shipping carton which describes the reason for the return. This valuable information will enable the service department to make the required repairs more efficiently.
11
Page 3
STANDARD CONDUCTIVITY SOLUTIONS
SIZE DESCRIPTION CODE
500 mL 74 micromhos/cm, 0.0005M KCl 6416-L 500 mL 718 micromhos/cm, 0.005M KCl 6417-L 500 mL 1413 micromhos/cm, 0.01M KCl 6354-L 500 mL 6668 micromhos/cm, 0.05M KCl 6418-L 500 mL 58640 micromhos/cm, 0.5M KCl 6419-L
REPLACING THE BATTERY
When the meter does not indicate greater than 80 micromhos in the “BAT. CK.” position, replace the battery. The battery is located under the meter panel of the instrument. Carefully remove the panel by unscrewing the four screws on the front of the unit and lift the unit out. Make sure battery terminal contacts are aligned with the correct terminals. Replace meter panel to instrument housing.
INSTRUMENT GUARANTEE
This instrument and probe, except for the battery, are guaranteed to be free from defects in material and workmanship for six months from original purchase date. If within that time the instrument is found to be defective, it will be repaired without charge (except for transportation expense to LaMotte Company, PO Box 329, Chestertown, Maryland 21620).
This guarantee is void under the following circumstances: operator’s negligence, improper application, and unauthorized servicing.
GENERAL PRECAUTIONS
Read the instruction manual before attempting to set up or operate this instrument. Failure to do so could result in personal injury or damage to the instrument.
LIMITS OF LIABILITY
Under no circumstances shall LaMotte Company be liable for loss of life, property, profits or other damages incurred through the use or misuse of their products.
PACKAGING AND DELIVERY
Experienced packaging personnel at LaMotte Company assure the adequate protection against normal hazards encountered in transportation of shipment. After the product leaves the manufacturer, all responsiblity for safe delivery is assured by the transportation company. Damage claims must be filed immediately with the transportation company to receive compensation for damaged goods.
TABLE OF CONTENTS
SPECIFICATIONS
Conductivity Meter ....................................................................................... 3
Conductivity Probe........................................................................................ 4
General Operating Procedure .................................................................. 5 - 6
SPECIAL PROCEDURES
Boiler Water................................................................................................... 6
Total Dissolved Salts - Soil ...................................................................... 6 - 7
Soluble Salts - Greenhouse Soil.................................................................... 7
Salinity........................................................................................................... 8
Calibration Procedure.................................................................................... 9
Battery Replacement.................................................................................... 10
Instrument Guarantee.................................................................................. 10
Limits of Liability......................................................................................... 10
Packaging and Delivery ............................................................................... 10
SPECIFICATIONS FOR CONDUCTIVITY METER
METER IN FOR MA TION
Range: Multi-Range
Low: 0-10 in 0.2 micromhos/cm increments Medium: 0-100 in 2 micromhos/cm increments Standard: 0-1,000 in 20 micromhos/cm increments High: 0-10,000 in 200 micromhos/cm increments
Accuracy: Precision:
± 1% full scale reading (all ranges) ± 0.5% full scale reading (all ranges)
FUNC TION CON TROLS
Operation Switch:
BAT.CK.- To check battery reserve OFF - Cuts off energy supply CAL (Calibration) - To calibrate instrument ON - Turns instrument on for reading
Calibration: Calibrates instrument by regulating voltage potential Temperature Compensator:
0° to 100°C in 5°C increments - adjusts instrument to temperature of sample being tested
Page
10
3
Page 4
Range Selector:
X.1 - conductivity reading multiplied by 0.1 X1 - conductivity reading multiplied by 1 X10 - conductivity reading multiplied by 10 X100 - conductivity reading multiplied by 100
Other Specifications
Connector: Standard BNC for conductivity probe Power: Battery - single 9 volt transistor
(Replacement types - Eveready 216, Burgess 2U6, RCA V5323, RAY-O-VAC 1604).
Dimensions: 3 n” width x 2 r” depth x 6 c” height Weight: 1.7 lbs.
CARRYING CASE
Dimensions: 10” width x 6” depth x 6 c” height Weight: With meter, 3.5 lbs.
SPECIFICATIONS FOR CONDUCTIVITY PROBE
Range: 0-10,000 micromhos/cm Size: 3” x t Weight: 2 oz. Cable Length: 3 ft. Probe Constant: 2.0/cm Elements: Carbon
NOTES ON THE USE OF CON DUC TIV ITY PROBE
The sensing elements of the conductivity probe consists of two carbon elements rigidly supported in a rugged plastic material. These elements have a very precise parallel spacing which determines the probe cell constant. To keep this spacing and to assure accurate results, it is essential that the probe is properly maintained and kept clean.
1. When conductivity measurements are made, the probe must be immersed in sufficient solution to cover the bottom groove. The probe may be completely immersed within the sample without adversely affecting the results.
2. The conductivity probe must be kept from contacting the bottom or sides of the container which holds the sample. Contact with surfaces other than the solution might result in inaccurate readings since the current flow may be altered between the probe elements.
3. After the completion of the conductivity measurement, rinse inside and outside of probe thoroughly with water, preferably deionized water. Allow probe to air dry completely before storing. The probe may be stored dry indefinitely without any damage.
CALIBRATION PROCEDURE
Although each conductivity meter is precalibrated before it leaves the manufacturer, it may become necessary to check the calibration, especially if the conductivity probe is different from the one originally supplied with the unit. This procedure should only be attempted by qualified personnel. Please read the instructions thoroughly before starting.
Two accurately-prepared standard conductivity solutions are required for the calibration procedure*. These solutions may be prepared with reagent grade potassium chloride (KCl) in the following manner:
Solution #1: 0.005M KCl: In a 1.0 liter volumetric flask, dissolve 0.3725g anhydrous KCl in deionized or distilled water and dilute to 1.0 liter mark. At 25°C this solution has a conductivity of 718 micromhos/cm.
Solution #2: 0.05 M KCl: In a 1.0 liter volumetric flask dissolve 3.725g anhydrous KCl in deionized or distilled water and dilute to 1.0 liter mark. At 25°C this solution has a conductivity of 6,668 micromhos/cm.
*Prepared standard conductivity solutions may be ordered from LaMotte Company (see page 10).
PRO CE DURE
1. Follow steps 1-3 in General Operating Procedure, page 4.
2. Set the range selector at “X10” and switch the control knob to the “ON”
position. Measure the temperature of the 718 micromhos/cm solution (Solution #1) and set the “TEMP” knob accordingly. Sufficiently immerse the probe in the solution to at least cover the end slots. Adjust the “STD” knob until the meter reading corresponds to the solution value.
3. Remove the 4 screws from the meter panel and lift unit from the base thereby exposing the internal circuitry.
4. Switch the control knob to the “CAL.” position and adjust to a full scale reading of 100 micromhos/cm by using the internal trim pot labeled “full scale” located on the PC board where the switches are mounted.
5. Switch the control knob to the “ON” position and the range selector to “X100". The ”TEMP", “STD”, and internal calibration pot should not be changed for the X100 range. Sufficiently immerse the probe in the 6,668 micromhos/cm solution (Solution #2) to cover the end slots. Adjust the trim pot labeled “X100" located on the meter PC board until the meter reading corresponds to the solution value.
6. The instrument is now calibrated for all ranges.
4
9
Page 5
FOR CHANGING CONDUCTIVITY* INTO SALINITY
CONVERSION TABLE
CONDUCTIVITY SALINITY
0°C 5°C 10°C 15°C 20°C 25°C 30°C
1,200 1,400 1,500 1,700 2,000 2,200 2,400 1 2,220 2,500 2,900 3,300 3,700 4,100 4,500 2 3,200 3,700 4,200 4,700 5,300 5,900 6,500 3 4,100 4,700 5,400 6,100 6,900 7,600 8,400 4 5,000 5,800 6,600 7,500 8,400 9,300 10,300 5 5,900 6,800 7,900 8,800 9,900 11,000 12,100 6 6,700 7,800 8,900 10,100 11,300 12,600 13,900 7 7,600 8,800 10,100 11,400 12,800 14,200 15,700 8 8,500 9,800 11,200 12,700 14,200 15,800 17,400 9
9,300 10,800 12,300 13,900 15,600 17,300 19,100 10 10,200 11,800 13,400 15,200 17,000 18,900 20,800 11 11,000 12,800 14,500 16,400 18,900 20,400 22,500 12 11,900 13,700 15,600 17,600 19,700 21,900 24,100 13 12,600 14,600 16,700 18,900 21,100 23,400 25,800 14 13,400 15,600 17,800 20,100 22,400 24,900 27,400 15 14,200 16,400 18,800 21,200 23,800 26,400 29,100 16 15,000 17,400 19,800 22,400 25,100 27,800 30,700 17 15,800 18,300 20,900 23,600 26,400 29,300 32,300 18 16,600 19,200 21,900 24,800 27,700 30,700 33,900 19 17,400 20,100 23,000 25,900 29,000 32,200 35,500 20 18,200 21,100 24,000 27,100 30,300 33,600 37,000 21 19,000 22,000 25,100 28,300 31,600 35,000 38,600 22 19,800 22,900 26,100 29,400 32,900 36,500 40,100 23 20,600 23,800 27,100 30,600 34,200 37,900 41,700 24 21,400 24,700 28,100 31,700 35,400 39,300 43,200 25 22,100 25,500 29,100 32,800 36,700 40,700 44,800 26 22,800 26,400 30,100 33,900 37,900 42,100 46,300 27 23,600 27,300 31,100 35,100 39,200 43,500 47,800 28 24,400 28,100 32,100 36,200 40,400 44,800 49,400 29 25,200 29,000 33,100 37,300 41,700 46,200 50,900 30 26,000 30,000 34,100 38,500 43,000 47,600 52,400 31 26,800 30,900 35,100 39,600 44,200 49,000 53,900 32 27,500 31,700 36,100 40,700 45,400 50,300 55,400 33 28,300 32,600 37,100 41,800 46,700 51,700 56,800 34 29,100 33,500 38,100 42,900 47,900 53,000 58,300 35 29,700 34,200 39,000 44,000 49,100 54,400 59,800 36 30,500 35,100 40,000 45,100 50,300 55,700 61,300 37 31,200 36,000 41,000 46,200 51,500 57,100 62,800 38 32,000 36,800 41,900 47,200 52,700 58,400 64,200 39 32,700 37,700 42,900 48,300 53,900 59,700 65,700 40
*Conductivity values are given in micromhos/cm. Data derived from the equation of P.K. Weyl, Limnology and Oceanography; 9,75 (1964).
GENERAL OPERATING PROCEDURE
1. Plug the probe into the BNC receptacle (top right corner of panel).
2. With the control knob in the “OFF” position, the meter pointer must
balance at zero. The mirror-faced scale insures accurate readings by
0/
00
eliminating errors due to parallax problems which would otherwise occur when the indicating needle and the scale are not properly aligned by the reader’s eye.
3. Switch the control knob to the “BAT. CK.” position. The meter should indicate greater than 80 micromhos/cm. If not, replace the battery (see instructions for battery replacement, page 8).
4. Set the “TEMP” knob at 25°C, the Range Selector at “X10”, and switch the control knob to the “CAL” position. The meter should indicate exactly 100 micromhos/cm. If not, carefully adjust by using the “STD” knob.
SAM PLE MEA SURE MENT - METHOD A NOR MAL AC CU RACY WITH TEM PER A TURE COM PEN SA TION
5. Use the thermometer (1066) to measure the sample water temperature in degrees centigrade. Set the “TEMP” knob at this figure.
6. Immerse the probe in at least a” of solution. Make sure all bubbles are removed from the probe.
7. Switch the control knob to the “ON” position and the range selector to either “X1”, “X10”, “X100”, or “X1000”. Read the conductivity directly from the meter face. For the best accuracy, select the range which gives readings as close to mid-range of the scale as possible. Multiply the reading from the meter face by the appropriate factor. The probe should not be in contact with any surface other than the solution under test.
8. After completing the conductivity measurement, switch the control knob to the “OFF” position. Rinse the probe thoroughly with water, preferably deionized water, and air dry.
8
5
Page 6
SAM PLE MEA SURE MENT - METHOD B MAX I MUM AC CU RACY WITH TEM PER A TURE COR REC TION
5. Use the thermometer (1066) to measure the sample water temperature. Record this temperature. The “TEMP” knob should be still set at 25°C.
6. Immerse the probe into sufficient solution to at least cover the end slots of the probe. Make sure all bubbles are removed from the probe.
7. Switch the control knob to the “ON” position and the range selector to either “X.1", ”X1", “X10", or ”X100". Select the range which gives readings as close to the middle of the scale as possible. Record the reading. During this test the probe should not touch any surface of the sample container.
8. Calculate result:
Conductivity, micromhos/cm at 25°C =
Meter Reading x Range Factor x [1.0 + 0.02 (T-25)]
T = Temperature of Sample in °C
9. After the measurement, switch the meter to “OFF”. Rinse probe thoroughly with water, preferably deionized water, and air dry.
NOTE: Reference solutions which have known conductivities may be prepared by consulting the latest edition of Standard Methods for the Examination of Water and Wastewater, American Public Health Association, 1015 East Eighteenth Street, NW, Washington, DC 20036. A series of accurately prepared reference solutions in both the high range and low range allow the operator to check on the manipulative procedure as well as the verification of results.
DETERMINATION OF CONDUCTIVITY & TDS IN BOILER WATER
In boiler water where high concentrations of hydroxides may be evident, it is necessary to neutralize the sample solution with gallic acid or lactic acid. Otherwise false high conductivity readings will be noted. For a 100 mL sample add 0.4 grams of gallic acid or lactic acid. This will neutralize approximately 1300 ppm hydroxide alkalinity. If a small amount of gallic acid or lactic acid does not dissolve within the sample solution, the conductivity reading will not be affected. On some highly alkaline waters higher concentrations of gallic acid or lactic acid may be required.
The determination of conductivity for boiler water is made by following the General Operating Procedure given on Page 4. To convert the conductivity reading to Parts Per Million Total Dissolved Solids (ppm TDS), multiply the reading by 0.7.
EXAMPLE: Conductivity reading is 3500 micromhos/cm.
3500 x 0.7 = 2450 ppm TDS
DETERMINATION OF SOIL CONDUCTIVITY & TOTAL DISSOLVED SALTS
An approximation of the Total Dissolved Salts in soil is obtained by making a soil extraction with distilled water and measuring the conductivity of the soil extract. For purposes of comparison, all conductivity values are recorded at 25°C.
NOTE: For determination of Soluble Salts in greenhouse soils, see below.
PRO CE DURE
1. Fill a 50 mL beaker full with the soil to be tested. Tap it lightly to eliminate any air trapped. Strike the excess soil off the surface.
2. Empty the contents of the beaker into a 250 mL widemouth flask and add 100 mL of distilled water.
3. Stopper and shake vigorously. Allow to stand for thirty minutes. During the thirty minutes waiting period the sample should be shaken vigorously three or four times.
4. Filter the contents of the flask and collect filtrate in suitable conductivity chamber.
5. Take conductivity reading according to method given for General Operating Procedure on page 4.
6. To convert to Total Dissolved Salts use the following formula:
micromhos/cm @ 25°C x 0.7 = ppm Total Dissolved Salts
If the Soluble Salts as determined above are greater than 1000 ppm, the chlorides and sulfates should be determined in order to learn whether the Soluble Salts are chlorides or sulfates.
Usually in calcareous soils, the sulfates represent gypsum and cause little injury to plants, while high concentrations of chlorides might result in damage to the plants. Greenhouse and sensitive plants may be damaged if the Souble Salts are over 500 ppm of chlorides.
DETERMINATION OF SOLUBLE SALTS FOR GREENHOUSE SOILS
The LaMotte Multi-Range Conductivity Meter may be used to read conductivity in the limited volume of soil extract obtained through the soil extraction procedure employed in the LaMotte Model GS-01 Greenhouse Outfit. Follow the extraction procedure in the GS-01 manual. Take conductivity reading according to method given for General Operating Procedure on Page 5. Use the following formula to convert the conductivity reading to parts per million Soluble Salts:
Conductivity Reading (micromhos/cm) x 0.7 = ppm Soluble Salts
6
7
Loading...