LaMotte ColorQ DW User Manual

8/11 62059-01
Color
ColorQ Photometer for Drinking Water
To insure proper testing results.
Insert tubes into ColorQ with line and arrow facing forward.
Brush and rinse all tubes promptly after use.
Rinse tablet crusher between test factors.
Allow tubes to dry before returning to case.
Replace reagent caps securely.
Use only LaMotte replacement reagents.
If unit should auto-off during testing rinse and fill one tube to Blank and return to testing.
To skip test in the sequence press button to advance past that test factor.
To turn meter off at any time, press and hold the button until “off” is displayed.
bAt = Battery is low.
Q
DW
For software Version 5.73 or higher
Order Code 2059-01
WARNING! This set contains chemicals that may be harmful if misused. Read cautions on individual containers carefully. Not to be used by children except under adult supervision.
PO Box 329 • Chestertown • Maryland • 21620 • USA 800-344-3100 • 410-778-3100 • 410-778-6394 www.lamotte.com
BLANK (CALIBRATION)
1
BLA
A
L
Color
Q
1. Fill sample bottle with water sample. Replace cap.
2. Fill clean tube to the 5mL line with water sample.
3. Insert tube into ColorQ as shown.
4. Press button to turn meter on. When “bLA” appears press button to “Blank” the meter, and go to pH. Remove tube.
PH
2
pH
ColorQ Range: 4.0-9.0
pH - Code 7059-E
p h
Color
5
1. Add 5 drops Wide Range pH Reagent to the same tube.
2. Cap. Invert 3 times to mix. Insert tube into ColorQ.
3. Press button to read “pH” pH. Remove tube.
* Reagent is a potential health hazard. Read MSDS at www.lamotte.com.
2
Q
HARDNESS
3
hb
ColorQ Range: 0-41 GPG
*
Hardness 1 Buffer: Code 7045G Hardness 2 Indicator: Code 7046-G
Color
Q
5 5
+
1. Press button to go to “Hd” Hardness.
2. Fill a clean tube to the 5 mL line with water sample.
3. Add 5 drops Hardness 1 Buffer and Hardness 2 Indicator
to same tube.
4. Cap. Invert 3 times to mix. Insert tube into ColorQ.
5. Press button to read “Hd” Hardness in gpg. Remove tube.
DILUTION
Samples with hardness concentrations above 41 gpg must be diluted. The meter will display “Hi”.
A. Use the 1 mL pipet to add 2 mL water sample to a clean test tube. B. Dilute to the 5 mL line with distilled or deionized water. Cap and mix. C. Blank with the diluted sample from step B. D. Follow the test procedure with this diluted sample starting with step 3. E. Multiply displayed result by 2.5.
Color
Q
* Reagent is a potential health hazard. Read MSDS at www.lamotte.com.
3
Loading...
+ 5 hidden pages