CHLORINE BLEACH KIT
DIRECT READING TITRATOR, 0-10%
CODE 7105-03
QUANTITY CONTENTS CODE
2 x 30 mL *Hypochlorite Reagent A *7939PS-G
10 g Potassium Iodide Crystals 6809-D
60 mL Hypochlorite Reagent D 2790-H
1 Test Tube, 5-10-15 mL, w/cap 0778
1 Direct Reading Titrator, 0-10 Range 0377
1 Pipet, 0.5 mL, plastic 0353
1 Pipet, 1.0 mL, plastic 0354
1 Spoon, 0.1 g, plastic 0699
*WARNING: Reagents marked with an * are considered to be potential health hazards.
To view or print a Safety Data Sheet (SDS) for these reagents go to www.lamotte.com.
Search for the four digit reagent code number listed on the reagent label, in the contents
list or in the test procedures. Omit any letter that follows or precedes the four digit code
number. For example, if the code is 4450WT-H, search 4450. To obtain a printed copy,
contact LaMotte by email, phone or fax.
Emergency information for all LaMotte reagents is available from Chem-Tel:
(US, 1-800-255-3924) (International, call collect, 813-248-0585).
To order individual reagents or test kit components, use the specifi ed code number.
Carefully read the LaMotte Direct Reading Titrator Instruction Manual (1649) before
proceeding. The Titrator is calibrated in percent available chlorine; each minor
division equals 0.2 percent.
NOTE: If endpoint is not achieved after one full Titrator of reagent has been added in
the titration, refi ll and zero the titrator to continue titration beyond 10 percent.
PROCEDURE
1. Fill test tube (0778) to 10 mL line with tap water.
2. Use the 0.5 mL pipet (0353) to add 0.5 mL of sample to the test tube containing
tap water from Step 1.
3. Use the 1.0 mL pipet (0354) to add 1.0 mL of *Hypochlorite Reagent A (7939).
Swirl to mix.
4. Use 0.1g spoon (0699) to add 0.1g Potassium Iodide Crystals (6809).
Swirl to mix.
5. Fill the Direct Reading Titrator (0377) with Hypochlorite Reagent D (2790). See
Direct Reading Titrator Manual (1649). Insert the Titrator tip into the center hole
of the test tube cap.
6. While gently swirling the tube, slowly depress the plunger to titrate the sample
until the solution changes from brown to colorless and remains colorless for one
minute. Read the test result directly from the scale where the large ring on the
Titrator meets the Titrator barrel. Record as Percent Chlorine.
NOTE: The solution will go through an intermediate yellow stage.
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
800-344-3100 • 410-778-3100 (Outside U.S.A.) • Fax 410-778-6394
PO Box 329 • Chestertown • Maryland • 21620 • USA
Visit us on the web at www.lamotte.com
67105-03 11/18