Maintaining the Meter......................................................................... 13
Warranty Information .......................................................................... 15
3
CONDUCTIVITY
Conductivity is defined as the ability of a solution to conduct an electrical
current, or the reciprocal of the solution’s ability to resist the current. This
current is conducted by electrically charged particles called ions, which
are present in almost all solutions. Different solutions have different kinds
and amounts of ions: distilled water has very few ions, and therefore a low
conductivity, while sea water has a large number of ions, and a high
conductivity.
Although a conductivity reading provides an overall measurement of the
ionic content of a solution, it is not possible to distinguish the specific
amounts of individual ions. For this reason, conductivity is often used to
measure the total dissolved solids (TDS) of a solution. TDS is defined as
the amount of solids which will pass through a 45 micron filter. Rather
than filtering a solution, the TDS can be estimated by multiplying the
conductivity measurement by a predetermined factor. This factor, which is
determined gravimetrically, will fall between 0.55 and 0.9; 0.7 is a
commonly used factor.
The conductivity measurement can also be used to estimate the salinity of
water, or the total amount of all salts dissolved in the water. Typically, the
conductivity reading is converted to salinity using charts, such as the one
found on page 10. These charts are based on water containing the same
amount and proportion of ions as standard seawater, so this form of
measurement is most effective for low concentrations and dilutions of
seawater.
Conductivity is measured in microsiemens per centimeter (msiemens/cm).
Siemens are also called mhos. In waters of higher conductivity,
msiemens/cm may be mul million. Therefore, using the information
discussed above:
msiemens/cm x 0.7 = ppm TDS
Salinity is usually measured in parts per thousand (ppt). The chart on page
10can be used to convert conductivity readings to salinity.
Because it is a quick, reliable, and inexpensive way of monitoring the
ionic content of a solution, conductivity measurements are widely used in
many areas of water testing, from environmental monitoring to municipal
water supplies to many industrial applications.
4
METER BASICS
Conductivity is measured using a probe which contains two electrodes,
separated by a fixed distance. When a voltage is applied from the meter
across the electrodes, the ions in solution conduct a current, which flows
between the electrodes. The greater the concentration of ions in the
solution, the larger the current generated and the higher the conductivity.
Likewise, the smaller the concentration of ions, the lower the
conductivity. The meter converts the current measured to a conductivity
reading.
Over time the electrodes may become dirty or fouled with contaminants
from the sample. For specific probe cleaning instructions for the CDS
5000 see Maintaining The Meter on page 13.
Conductivity measurements are very dependent on temperature. The
ability of the ions to move through the solution, and conduct the current,
is related to the temperature of the solution. As the temperature rises, the
ions move more quickly through the solution, increasing the conductivity;
likewise as the temperature decreases the ions move more slowly and the
conductivity decreases. Since the conductivity of the same solution can
change by as much as 2%/°C, accurate temperature measurements must
be made simultaneously to the conductivity reading. The CDS 5000
includes a temperature probe to measure the temperature.
To make conductivity readings taken at different times and places
comparable, measurements are often converted to what the conductivity
of the solution would be at 25°C. The CDS 5000 automatically makes this
conversion before providing a final reading.
The CDS 5000 is precalibrated at the factory; set both the TDS knob and
the SLOPE knob to the 12 o’clock position and proceed. This procedure
will give results within ±10% of the actual reading. For more accurate
results, follow the procedure below.
1. Press “°C” button to turn the meter on.
2. Set TDS knob to the 12 o’clock position.
1
3. Insert temperature and conductivity probes at least
conductivity standard.
“into
2
4. Gently stir with conductivity probe until reading stabilizes. Press
“mS/ppm” or “mS/ppt” button.
5. Adjust SLOPE knob until display reads conductivity of chosen
standard. The CDS 5000 is now calibrated and ready for use.
LaMotte offers several ranges of conductivity standards. Choose the
standard most appropriate for your testing needs and order using the four
digit code number listed.
2. Set TDS and SLOPE knobs to the 12 o’clock position.
NOTE: If meter was calibrated according to procedure on page 7,
leave SLOPE knob set in the same position.
3. Insert temperature and conductivity probes at least 1/2" into sample.
4. Gently stir with conductivity probe until reading stabilizes. Press
“mS/ppm” button. Record reading as msiemens/cm.
5. If a 1 appears on the far left side of the display, the reading is out of
range. Repeat procedure using “mS/ppt” button. Record as
msiemens/cm. To convert to msiemens/cm, multiply reading by 1000.
Record as msiemens/cm.
6. Press “OFF” button when finished testing. Rinse probe with distilled
water and dry thoroughly before storing.
TOTAL DISSOLVED SOLIDS
1. Set TDS knob to desired multiplication factor.
NOTE: 0.7 is a commonly used multiplication factor.
2. Press “°C” button to turn the meter on.
3. Set SLOPE knob to the 12 o’clock position.
NOTE: If meter was calibrated according to procedure on page 7,
leave SLOPE knob set in the same position.
4. Insert temperature and conductivity probes at least 1/2" into sample.
5. Gently stir with conductivity probe until reading stabilizes. Press
“mS/ppm” button. Record reading as msiemens/cm.
6. If a 1 appears on the far left side of the display, the reading is out of
range. Repeat procedure using “mS/ppt” button. Record as
msiemens/cm. To convert to msiemens/cm, multiply reading by 1000.
Record as msiemens/cm.
7. Press “OFF” button when finished testing. Rinse probe with distilled
water and dry thoroughly before storing.
8
TESTING SALINITY
1. Press “°C” button to turn the meter on.
2. Set TDS and SLOPE knobs to the 12 o’clock position.
NOTE: If meter was calibrated according to procedure on page 7,
leave SLOPE knob set in the same position.
3. Insert temperature and conductivity probes at least 1/2" into sample.
4. Gently stir with conductivity probe until reading stabilizes. Press
“mS/ppt” button. Record reading as msiemens/cm.
5. Using chart on the following page, convert conductivity reading to
salinity . Record as ppt Salinity.
6. Press “OFF” button when finished testing. Rinse probe several times
with distilled water and dry thoroughly before storing.
9
FOR CHANGING CONDUCTIVITY INTO SALINITY
CONVERSION TABLE
ConductivitySalinity
0°C5°C10°C15°C20°C25°C30°C
0
1.2001.4001.5001.7002.0002.2002.4001
2.2202.5002.9003.3003.7004.1004.5002
3.2003.7004.2004.7005.3005.9006.5003
4.1004.7005.4006.100
6.900
7.6008.4004
5.0005.8006.6007.5008.4009.30010.3005
5.9006.8007.9008.8009.90011.000 12.1006
6.7007.8008.90010.10011.30012.600 13.9007
7.6008.80010.10011.40012.80014.200 15.7008
8.5009.80011.20012.70014.20015.800 17.4009
9.30010.80012.30013.90015.60017.300 19.10010
10.20011.80013.40015.20017.00018.900 20.80011
11.00012.80014.50017.60018.90020.400 22.50012
11.90013.70015.60018.90019.70021.900 24.10013
12.60014.60016.70020.10021.10023.400 25.80014
00
13.40015.60017.80020.10022.40024.900 27.40015
14.20016.40018.80021.20023.80026.400 29.10016
15.00017.40019.80022.40025.10027.800 30.70017
15.80018.30020.90023.60026.40029.300 32.30018
16.60014.20021.90024.80027.70030.700 33.90019
17.40020.10023.00025.90029.00032.200 35.50020
18.20021.10024.00027.10030.30033.600 37.00021
ConductivitySalinity
0°C5°C10°C15°C20°C25°C30°C
0
19.00022.00025.10028.30031.60035.000 38.60022
19.80022.90026.10029.40032.90036.500 40.10023
20.60023.80027.10030.60034.20037.900 41.70024
21.40024.70028.10031.70035.40039.300 43.20025
22.10025.50029.10032.80036.70040.700 44.80026
22.80026.40030.10033.90037.90042.100 46.30027
23.60027.30031.10035.10039.20043.500 47.80028
24.40028.10032.10036.20040.40044.800 49.40029
25.20029.00033.10037.30041.70046.200 50.90030
00
10
FOR CHANGING CONDUCTIVITY INTO SALINITY
CONVERSION TABLE
ConductivitySalinity
0°C5°C10°C15°C20°C25°C30°C
0
00
26.80030.90035.10039.60044.20049.000 53.90032
27.50031.70036.10040.70045.40050.300 55.40033
28.30032.60037.10041.80046.70051.700 56.80034
29.10033.50038.10042.90047.90053.000 58.30035
29.70034.20039.00044.00049.10054.400 59.80036
30.50035.10040.00045.10050.30055.700 61.30037
31.20036.00041.00046.20051.50057.100 62.80038
32.00036.80041.90047.20052.70058.400 64.20039
32.70037.70042.90048.30053.90059.700 65.70040
Data derived from the equation of P.K. Weyl, Limnology and Oceanography;9,75
(1964).
11
TESTING SOIL
The Total Dissolved Solids (TDS) level of soil samples can be determined
using the CDS 5000. A soil extraction is made using distilled water, and
the TDS level measured.
1. Fill a 50 mL beaker with sample soil. Tap lightly to eliminate trapped
air. Remove excess soil from the surface.
2. Empty beaker into a 250 mL widemouth flask. Add 100 mL of
distilled water. Stopper and shake vigorously. Wait 30 minutes.
NOTE: During the waiting period, vigorously shake the sample three
or four times.
3. Filter contents of flask, collecting filtrate in a beaker or other suitable
container.
NOTE: LaMotte Company offers a funnel (order code 0459) and
filter paper (order code 0465) which can be used for this filtration.
4. Set TDS knob to desired multiplication factor.
NOTE: 0.7 is a commonly used multiplication factor.
5. Press “°C” button to turn the meter on.
6. Set SLOPE knob to the 12 o’clock position.
NOTE: If meter was calibrated according to procedure on page 7,
leave SLOPE knob set in the same position.
7. Insert temperature and conductivity probes at least 1/2" into sample.
8. Gently stir with conductivity probe until reading stabilizes. Press
“mS/ppm” button. Record reading as msiemens/cm.
9. If a 1 appears on the far left side of the display, the reading is out of
range. Repeat procedure using “mS/ppt” button. Record as
msiemens/cm. To convert to msiemens/cm, multiply reading by 1000.
Record as msiemens/cm.
10. Press “OFF” button when finished testing. Rinse probe with distilled
water and dry thoroughly before storing.
12
MAINTAINING THE METER
Adjust This
Potentiometer
REPLACING THE BATTERY
When “BAT” appears on the display, the battery should be replaced. The
temperature reading will be the first function to be affected by a low
battery.
1. Use a #1 Phillips head screwdriver to remove four screws on the back
of the meter case.
2. Gently lift back panel from meter .
3. Lift battery from bottom of meter. Remove from connector.
4. Snap new battery onto connector.
NOTE: The CDS 5000 uses a type 1604A (9 volt) battery .
5. Lower battery into compartment. Replace back panel and screws.
AC ADAPTER
An AC adapter is available for use with the CDS 5000. Order as code
#1708. Insert connector into small hole next to the probe connector.
CLEANING THE PROBE
The graphite probe may occasionally become dirty and need to be
cleaned. After each use the probe should be thoroughly rinsed with
distilled water. If further cleaning is necessary, the probe can be washed
with a mild detergent, and then thoroughly rinsed with distilled water.
Always thoroughly dry the probe before storing.
13
REPLACING THE PROBE
If the probe cannot be adequately cleaned, or becomes damaged, it must
be replaced. When a new probe is attached to the meter, the CDS 5000
must be recalibrated using the following procedure. Do not allow the CDS
5000 to contact any conductive surfaces while performing this procedure.
1. Set SLOPE and TDS knobs to the 12 o’clock position.
2. Use a Phillips head screwdriver to remove four screws from back of
meter. Remove back from meter.
3. Follow the Conductivity testing procedure described on page 8 to
measure the conductivity of a chosen standard. Choose the standard
that is closest to the value of the solutions commonly measured.
4. Using a small, flat edge screwdriver, adjust the top potentiometer
(indicated in the diagram) until value of standard solution is
displayed.
NOTE: Only adjust the top potentiometer; adjusting any other
potentiometer voids the meter’s warranty. If these other
potentiometers are adjusted, the meter must be returned to LaMotte
for repair.
5. Replace back on meter and replace the four screws. The CDS 5000 is
now calibrated and ready for use.
14
WARRANTY INFORMATION
REPAIRS
If it is necessary to return the instrument for repair, contact LaMotte
Company at 1-800-344-3100 for a return authorization number.
INSTRUMENT GUARANTEE
This instrument, excluding the probe, is guaranteed to be free of defects in
material and workmanship for one year from date of original purchase. If,
in that time, it is found to be defective, it will be repaired without charge,
except for transportation expenses. This guarantee does not cover the
batteries.
This guarantee is void under the following circumstances:
operator’s negligence
•
improper application
•
unauthorized servicing
•
LIMITS OF LIABILITY
Under no circumstances shall LaMotte Company be liable for loss of life,
profits, or other damages incurred through the use of misuse of their
products.
PA CKAGING AND DELIVERY
Experienced packaging personnel at LaMotte Company assure adequate
protection against normal hazards encountered during shipping. After the
product leaves the manufacturer, all responsibility for its safe delivery is
assured by the transporter. Damage claims must be filed immediately with
the transporter to receive compensation for damaged goods.