LaMotte AT Visual Series Water Softener Demo Kit User Manual

Model S/DuoSoft
with Octa-Slide 2
water quality
DEMO OUTFIT
AT 38/40
o
Table of Contents
Safety Information .................................................................... 4
Testing Hints ............................................................................. 4
Water Softeners ......................................................................... 6
Introduction .................................................................................... 6
Instructions for Use
Procedure A - Use of the Softeners ................................................. 7
Procedure B - Replacing the Media in the Softeners ....................... 8
Total Hardness Test .................................................................. 10
Soap Demonstration ................................................................. 11
Precipitation Demonstration ..................................................... 12
pH Test ...................................................................................... 13
Iron Test .................................................................................... 14
Optional Test Kit Modules
Nitrate-Nitrogen ............................................................................ 15
Free Available Chlorine; Total Residual Chlorine
& Combined Chlorine ................................................................. 16
TDS ............................................................................................... 17
Replacement Parts .................................................................... 19
Safety Information
Read the instruction manual thoroughly to familiarize yourself with the test procedures before you begin. Make note of any precautions in the instructions.
Read the labels on all LaMotte reagent containers prior to use. Some containers include precautionary notices and fi rst aid information. Certain reagents are considered hazardous substances and are designated with a * in the instruction manual. Material Safety Data Sheets (MSDS) are available at www.lamote.com. Read the MSDS before using these reagents. Additional emergency information for all LaMotte reagents is available 24 hours a day from the Poison Control Center listed in the front of the phone book. Be prepared to supply the name and four-digit LaMotte code number found on the container label or at the top of the MSDS. LaMotte reagents are registered with a computerized poison control information system available to all local poison control centers.
Keep equipment and reagent chemicals out of the reach of young children.
Protect Yourself and Equipment: Use Proper Analytical Techniques
Testing Hints
1. 2. 3.
Tightly close all reagent containers immediately after use. Be sure not to interchange caps and pipets from containers.
different
Avoid prolonged exposure of equipment and reagents to direct sunlight.
Protect reagents and components from extreme heat and cold.
Wipe up any reagent chemical
4. 5.
spills, liquid or powder, as soon as they occur. Refer to label and accompanying MSDS for proper reagent disposal.
4
Use care when dispensing or handling all reagents due to safety reasons. Some chemicals also may cause permanent stains if spilled.
ANALYTICAL TECHNIQUE
1. Clean glassware is a must for accurate results. Thoroughly rinse test tubes before and after each test. Caps and stoppers should also be cleaned after each use.
2. Use test tube caps or stoppers, not your fi ngers, to cover test tubes and fl asks during shaking or mixing.
3. When adding sample to calibrated test tube, be sure vial is fi lled to the appropriate mark. The bottom of the liquid (meniscus) should be level with the desired mark. (Figure 1)
4. When dispensing reagents from bottles fi lled with dropper plug and cap, be sure to hold bottle vertically and gently squeeze to dispense the appropriate number of uniform drops. (Figure 2)
5. For those reagents to be added with the screwcap pipet assemblies enclosed, remove polyseal cap on bottle and replace with the screwcap pipet. NOTE: Place the polyseal caps back on the reagent bottles for longer periods of storage. Be sure that both pipet assemblies and polyseal caps are thoroughly cleaned before placing on bottles to avoid contamination.
6. When dispensing reagents from pipets, hold pipet vertically to assure uniform drop size. This is extremely important when performing drop count titrations. (Figure 3)
7. To fi ll pipets, squeeze rubber bulb and immerse into reagent. Release bulb to fi ll. (Figure 4)
8. To accurately dispense powdered reagents with spoon, tap spoon on edge of reagent container to remove excess reagent. (Figure 5)
9. When performing tests that include Octa-Slide 2 Comparators, the comparator should be positioned between the operator and non-direct sunlight. This allows the light to enter through the light-diffusing screen at the back of the comparator for optimum color comparison.
5
Model S and DuoSoft Softeners
INTRODUCTION
The Model S and the DuoSoft Water Softeners are designed to produce high quality softened
water. As water passes through the Model S chamber, the resin column causes scale-forming calcium and magnesium ions to be exchanged for non-scale-forming sodium ions. When the resin is exhausted it must be replaced or regenerated. Inexpensive resin refi ll packages are available, or the original resin can be regenerated by chemical treatment.
In the DuoSoft, water passes through both chambers and will be treated by both types of media. The two chambers of the DuoSoft may be easily fi lled with the media of choice for specifi c problem water.
NOTE: These softeners DO NOT yield water suitable for drinking .
READ INSTRUCTIONS BEFORE USE.
OUTLET
HOSE
INLET HOSE
CLAMP
CLAMP
PLUG
CHAMBER
FAUCET
ADAPTER
OUTLET
HOSE
INLET HOSE
CLAMP
CLAMP
PLUG
TOP
CHAMBER
BOTTOM
CHAMBER
FAUCET
ADAPTER
6
Instructions for Use
PROCEDURE A — Use of the Softeners
1. 2.
Allow water to run from faucet briefl y to visually gauge pressure.
3. 4.
SLOWLY turn on faucet and adjust the fl ow of water into softener until the stream of water is about 1/8” in diameter. Maintain a gentle fl ow to prevent excessive pressure in the adapter.
5. 6.
Unclamp outlet hose at bottom of softener.
Attach adapter on inlet hose to faucet. Fully open clamp on inlet hose.
Fill chamber until a 1” layer of water is visible over the media.
Let effl uent run to waste for about 3 minutes to wash out residue from the previous sample.
3
minutes
7.
Adjust fl ow to maintain the 1” layer of water over the media.
9. 10.
When softener is disconnected from faucet, clamp inlet hose.
8.
After enough softened water has been collected, turn off faucet and clamp outlet hose
.
If softener will be stored unused for any length of time, keep a 1” layer of water over resin column to prevent drying and cracking.
7
PROCEDURE B — Replacing the Media in the Softeners
1. 2.
With the aid of a funnel, add the fresh media to the column.
3. 4.
Continue adding media to the column until the resin is about 2” from the top of the column.
2”
ADDITIONREMOVAL
5. 6.
Clean threaded area completely and replace the black plug in the top of column.
Tap water may be added to the media in the funnel to facilitate lling of the column.
Allow excess water to drain from outlet hose.
DuoSoft Unit
To add the media in the bottom chamber, turn unit upside down and repeat Addition Steps 1-5. When media has been added to both chambers, proceed with Procedure A. When using carbon media, it is suggested that it be used in the top chamber.
1.
To replace media, remove black plug from the top of dispenser.
3. 4.
By adding more water to the column and shaking, the remaining media can be easily removed.
2.
To remove the media in the bottom chamber, turn unit upside down and repeat Removal Steps 1-3. To add new media, follow Addition Steps 1-5 for both chambers.
Unclamp outlet hose and turn softener upside down.
Allow exhausted media and water to run to waste.
DuoSoft Unit
8
PROCEDURE C — Regeneration of Media
Follow manufacturers’ instructions for regeneration of media. Cation exchange resin may be regenerated in the following manner. NOTE: Due to build up of air pressure, it is not possible to regenerate both chambers of the DuoSoft unit at the same time. Follow steps 1-8 to regenerate top chamber, then see step 9 to regenerate bottom chamber.
1. 2.
Prepare salt solution by dissolving about ½ pound of salt (sodium chloride) in 1 quart of water.
3. 4.
Clamp outlet tube.
Add salt solution until top chamber is completely full.
5. 6.
Unclamp outlet tube and allow salt solution to run to waste.
Remove black plug from top of softener.
Allow to stand for about 30 minutes.
30
MINUTES
Clean threaded area completely and replace black plug in top of softener.
7.
Attach adapter to faucet.
For the DupSoft unit; turn unit upside down and repeat steps 1-8.
9.
8.
Wash resin with approximately 1 gallon of tap water.
9
Total Hardness
Total Hardness of a water supply generally represents the total concentration of Calcium and Magnesium ions expressed as Calcium Carbonate (CaCO3). Other ions may contribute, however, they are usually present in insignifi cant quantities. Hard waters may form scale on plumbing fi xtures, consume excessive quantities of soap, and leave deposits of fi lm on glassware, fabrics, etc. Excessive hardness may be removed by various treatment methods.
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
TEST PROCEDURE
1. 2. 3.
Thoroughly rinse the sample tube (4488) with the water to be tested.
10 1
20
Fill the sample tube (4488) to the desired line** with the sample water.
4. 5. 6.
Add either 1 Hardness Reagent #6 Tablet (4484) or 5 drops of *Hardness Reagent #6 (4485).
1
0
1
2
0
Swirl to mix.
10 1
20
10 1
20
7. 8.
Counting the number of drops and swirling between drops, add Hardness Reagent #7 (4487WT) one drop at a time until the red color changes to clear
1
0
1 2
0
blue.
Multiply the number of drops used in Step 7 as follows: Tube fi lled to upper line: each drop equals 10 ppm Hardness as CaCO Tube fi lled to middle line: each drop equals 1 gpg Hardness as CaCO Tube fi lled to lower line: each drop equals 20 ppm Hardness as CaCO
10 1
20
1
0
1
2
0
Add 5 drops of *Hardness Reagent #5 (4483).
Swirl to mix.
3
3
3
** When the tube is fi lled to upper line. Each drop of Hardness Reagent #7 is equal to 10 ppm. When the tube is fi lled to middle line. Each drop of Hardness Reagent #7 is equal to 1gpg.
When the hardness level is over 200 ppm, fi ll to lower line. Each drop of Hardness Reagent #7 is equal to 20 ppm.
10
Soap Demonstration
Calcium and Magnesium ions present in a water supply are the principle contributors to the total hardness. Hard water tends to consume excessive quantities of soap and forms curds and deposits on glassware, fabrics, etc.
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
TEST PROCEDURE
1. 2. 3.
Thoroughly rinse the “SOFT” water fl ask (0453) with softened water.
Thoroughly rinse the “HARD” water fl ask (0452) with untreated water.
4. 5. 6.
With the pipet
Fill the “SOFT” fl ask (0453) with softened water to the same level.
7. 9.8.
A thick lather will form in the softened water.
(0392), add 4 drops of *Soap Reagent #4 (4767) to each ask.
Continue to add *Soap Reagent #4 (4767), one drop at a time, to the untreated “HARD” water, shake periodically until a lather forms. Count the number of drops added.
Fill the “HARD” ask (0452) with untreated water until the bottom surface is covered with a layer of water about ½” deep.
Cap and shake the fl ask.
Extremely hard water may require 30, 40, or even 60 drops of *Soap Reagent #4 to produce a lasting lather.
11
Precipitation Demonstration
Again, Calcium and Magnesium ions are the major contributors to water hardness. The chemical reagents in this demonstration pull the Calcium and Magnesium ions out of solution to form a cloudy precipitate in hard water. The water that has been run through the ion exchange column has had these ions removed, therefore, the sample should remain clear.
NOTE: This portion of the AT-38/40 Water Quality Demo Kit is ONLY a visual demonstration illustrating the removal of Calcium and Magnesium ions from tap water after treatment by the ion exchange process. The results should not be interpreted beyond the intent of the demonstration.
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Mate­rial Safety Data Sheet (MSDS) for these reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
TEST PROCEDURE
1. 2. 3.
Thoroughly rinse the “SOFT” water Demo Tube (0298) with softened water.
Thoroughly rinse the “HARD” water Demo Tube (0297) with untreated water.
4. 5. 6.
Fill the “HARD” Demo tube (0297) to the line with untreated water.
Add 7 drops of *Precipitation Reagent A (4542WT) to each tube.
7. 9.8.
Cap and mix.
Add 7 drops of *Precipitation Reagent B (4543WT) to each tube.
Fill the “SOFT” Demo tube (0298) to the line with softened water.
Cap and mix.
Place tubes in the Precipitation Rack (0879) and allow the tubes to stand for 5 minutes.
Heavy Precipitate
Remains clear
12
pH Test
Simply, the term pH can be considered to be an “index” of the amount of hydrogen ions present in a substance. This “index” can be used to quickly identify the acid, neutral, or alkaline (basic) nature of water. On the scale of 0.0 to 14.0, 7.0 is considered to be neutral, acidic water is less than 7.0, and basic or alkaline water is greater than 7.0. Water that is acidic in nature may cause corrosion of plumbing and equipment while alkaline water may contribute to scale buildup. Neutralization by various methods is used to correct pH for proper operation of equipment.
pH measurement may be made electronically with a pH meter or as below with a colorimetric method. The pH indicator used is a mixture of dyes that produces a specifi c color at various pH levels.
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Mate­rial Safety Data Sheet (MSDS) for these reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
Use of the Octa-Slide 2 Viewer
The Octa-Slide Viewer should be held so non-direct light enters through the back of the Viewer. Slide the Octa­Slide 2 Bar into the Viewer. Insert the reacted sample into the top of the Viewer. Match the color of the reaction to the color standards.
TEST PROCEDURE
1. 2. 3.
Insert Wide Range pH Octa­Slide 2 Bar (3483-01) into the Octa­Slide 2 Viewer (1101).
4. 5. 6.
Cap and mix.
Fill a test tube (0106) to the 5 mL line with the water sample.
Insert test tube into
Octa-Slide 2
Viewer.
Add 8 drops of *
Wide Range pH
Indicator
(2218).
Match sample color to a color standard. Record as pH.
13
Iron Test
Most natural waters contain iron, varying from trace to very large amounts in various forms. In the dissolved state (ferrous), iron water is often colorless. Upon exposure to air, or an oxidized state, ferrous iron will undergo a chemical reaction to the suspended (ferric) state causing discoloration, staining, and possibly an objection­able taste. Several methods are available for iron removal, the selection of which is dependent upon the state in which it exists.
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Mate­rial Safety Data Sheet (MSDS) for these reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
TEST PROCEDURE
1. 2. 3.
Insert Iron Octa­Slide 2 Bar (4448-01) into the
Octa-Slide 2
Viewer (1101).
Fill a test tube (0106) to the 5 mL line with the water sample.
4. 5. 6.
Cap and mix.
7. 9.8.
Wait 3 minutes.
Use the 0.05g spoon (0696) to add 0.05g of *Iron Reagent #2 (4451).
Insert test tube into
Octa-Slide 2
Viewer.
Add 5 drops of *Iron Reagent #1 (4450).
Cap and gently shake to dissolve powder.
Match sample color to a color standard. Record as ppm Iron.
14
Optional AT Kit Test Modules
Nitrate-Nitrogen • CODE 4-3004-01
Nitrogen is essential for plant growth, but the presence of excessive amounts in water supplies presents a major pollution problem. Nitrogen compounds may enter water as nitrates or be converted to nitrates from agricultural fertilizers, sewage, industrial and packing house wastes, drainage from livestock feeding areas, farm manures, and legumes.
*WARNING: Reagents marked with an * are considered to be potential health hazards.
To view or print a Material Safety Data Sheet (MSDS) for these reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
TEST PROCEDURE
1. 2. 3.
Insert Nitrate­Nitrogen Octa-Slide 2 Bar (3494-01) into the Octa­Slide 2 Viewer (1101).
Fill a test tube (0106) to the 5 mL line with the water sample.
4. 5. 6.
Cap and mix until tablet disintegrates.
Add one * CTA TesTab (NN-3703). Immediately slide the test tube into the Protective Sleeve (0106-FP).
Nitrate
#2
Add one *
Nitrate
#1 Tablet (2799A).
Cap and mix for two minutes disintegrate the tablet.
to
Wait 5 minutes. Remove
7. 9.8.
the tube from the protective sleeve. NOTE: Nitrate #2 CTA Tablets (NN-3703) are sensitive to UV light. If testing indoors, there is no to use the Protective in this procedure.
need
Sleeve
Insert test tube into
Octa-Slide 2
Viewer.
Match sample color to a color standard. Record as ppm Nitrate-Nitrogen.
To convert to Nitrate, multiply the results by 4.4.
15
Chlorine • CODE 4-3006-01
Water for cities and communities is usually sanitized. Even waters that come from clean sources, protected watersheds, reservoirs, and deep wells are commonly sanitized to assure safety. Chlorine is most commonly used because it is effective against a wide range of microorganisms, its cost is low, and the methods of applying it have been well developed.
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Mate­rial Safety Data Sheet (MSDS) for these reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
TEST PROCEDURE
Free Available Chlorine
1. 2. 3.
Insert DPD Octa-Slide 2 Bar (3401-01) into the Octa­Viewer
Slide 2
(1101).
Fill a test tube (0106) to the 10 mL line with the water sample.
Add one *
Chlorine DPD #1R Tablet (69999A).
4. 5.
Cap and mix until tablet dissolves.
Insert test tube into Octa-Slide 2 Viewer.
6.
Total Residual Chlorine & Combined Chlorine
8.
Insert DPD Octa-Slide 2 Bar (3401-01) into the Octa­Viewer
Cap and mix until tablet disintegrates.
Slide 2
(1101).
9.
11.10. 12.
Add one *
Chlorine DPD #3 Tablet (6905) to the sample from Step 6.
Insert test tube into Octa-Slide 2 Viewer.
Match sample color to a color Record Free Available Chlorine.
Match sample color to a color Record Total Residual Chlorine.
standard.
ppm
standard.
ppm
7.
Retain this sample if Total Residual and Combined Chlorine are to be determined. (Steps 8-12)
16
TDS • CODE 5-0080 • Dual Range TDS Pocketester
BEFORE FIRST USE: Soak electrodes for a few minutes in alcohol to remove oils.
CALIBRATION
(To Be Done Once a Month)
Use a calibration standard between 200 ppm and 1,900 ppm for low range testing.
Use a calibration standard between 1,900 ppm and 7,000 pm for high range testing.
See page 20 for suggested standards.
Pour ½” to 1”
1. 2. 3.
of a Calibration Standard into 2 containers, and tap or deionized water into a third container.
4. 5. 6.
Turn the meter on, and wait several minutes to allow the display to stabilize.
Open battery compartment lid (end with lanyard loop). The two white buttons are Increment (INC) and Decrement (DEC) calibration keys.
Press in INC or DEC keys to adjust reading to match the calibration standard value. After 3 seconds without a key press, the display fl ashes 3 times, then shows “ENT”. The tester accepts calibration value; returns to measurements mode.
Rinse electrode in water. Then rinse in fi rst container of standard, then dip into the second.
Replace battery cap. Rinse the electrode in the water and proceed with test procedure.
SETTING THE TDS FACTOR
This tester lets you select a TDS factor of 0.4 to 1.0. The preset TDS factor of 0.7 is recommended.
1. Open battery compartment. With meter on, press HOLD key, then press the INC key. (INC key is inside battery compartment).
2. Press the INC or DEC keys to adjust the TDS factor.
3. After 3 seconds without a key press, the display fl ashes 3 times, then shows “ENT”. Tester accepts TDS factor and returns to measurement mode.
4. Replace battery cap.
17
PROCEDURE
1. 2. 3.
Remove the cap.
Press ON/OFF button to turn on.
Dip the electrode end into the test solution. Make sure the sensor is fully covered. Stir once. Let the reading stabilize.
4. 5. 6.
7.
Allow time for the Automatic Temperature Compensation to correct the readings for solution temperature changes.
Replace the cap.
Note the reading once the display stabilizes. If the solution is above the range of the TDSTestr, the display will show:
You can replace the electrode module at the fraction of the cost of a new Testr. When the to calibrate or gives fl uctuating readings in calibration standards, the electrode.
1. With dry hands, grip the ribbed Testr collar with electrode facing you. Twist the collar counter clockwise. Save the ribbed Testr collar and O-ring for later use.
2. Pull the old electrode module away from the Testr.
3. Align the four tabs on the new module so they match the four slots on the Testr. Caution: Take care to align pins carefully. Bent or broken pins will cause the meter to malfunction.
4. Gently push the module onto the slots to seat in position. Push the smaller O-ring fully onto the new electrode module. Push the collar
over the module and thread it onto place by fi rmly twisting it clockwise.
Press the ON/OFF button to turn off Testr. Note: Testr will automatically shut off after 8.5 minutes of non-use.
Testr fails
you need to change
HOLD FUNCTION: Press HOLD key to freeze display. Press HOLD again to release. MAINTENANCE: To improve performance, clean the stainless steel electrodes by periodically rinsing them
in alcohol for 10-15 minutes. Replace all 4 batteries if the display becomes faint or disappears, or if the readings are unstable. If drift is detected while electrodes are continuously exposed to solution for longer than one hour, allow electrode to fully dry off periodically.
CHANGING BATTERIES: Open the battery compartment lid. Remove old batteries and replace with fresh
ones noting polarity as shown in the battery compartment. Recalibrate Testr after battery change.
18
Replacement Parts
Model AT-38 & AT-40
To order individual reagents or test components, use the specifi ed code number.
CODE DESCRIPTION CODE DESCRIPTION
1002 Model S Softener (AT-38)
*2218-G *pH, Wide Range Reagent, 25mL 1022 DuoSoft Softener (AT-40)
*4767-H *Soap Reagent #4, 60 mL 1028 DirectFLO Duo-Soft Softener
*4767-L *Soap Reagent #4, 500mL 0879 Precipitation Rack, acrylic
*4483WT-H *Hardness Reagent #5,60mL 0392 Pipet, plain, plastic, w/cap
*4483-L *Hardness Reagent #5, 500 mL 0670 Stopper, rubber, #6, for fl asks
4484-J Hardness Reagent #6 Tablets. 100 0655 Stopper, rubber, #3, for Hardness tube
4487WT-H Hardness Reagent #7, 60 mL 0651
*4542WT-H *Precipitation Reagent A, 60 mL 0106 Test Tube, plastic, w/cap
*4542-L *Precipitation Reagent A, 500 mL 1101 Octa-Slide 2 Viewer
*4543WT-H *Precipitation Reagent B, 60 mL 3483-01 Wide Range, pH Octa-Slide 2 Bar, 5-10
*4543-L *Precipitation Reagent B, 500 mL *4450-G *Iron Reagent #1, 25 mL
0452 Flask, 250 mL, “HARD”, w/cap *4451-S *Iron Reagent #2 Powder, 4.5g
0453 Flask, 250 mL, “SOFT”, w/cap 0696 Spoon, 0.05g
0297 Test Tube, “HARD”, 15 x 120 mm, w/cap 4448-01 Iron Octa-Slide 2 Bar, 0.5-10 ppm
0298 Test Tube, “SOFT”, 15 x 120 mm, w/cap 2-2011 150 mL beaker
4488 Test Tube, Hardness, w/cap
*4485WT-H *Hardness Reagent #6, 60 mL
*4485-L *Hardness Reagent #6, 500 mL
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
Stopper, rubber 00, for Precipitation tubes
19
Replacement Parts
for Optional Test Modules
To order individual reagents or test components, use the specifi ed code number.
CODE DESCRIPTION
*2799A-NN3703ABX *Nitrate #1 Tablet (50) and *Nitrate #2 Tablet (50)
0106 Test tubes, plastic, 2.5, 5 & 10 mL, w/caps (2)
0106-FP Protective Sleeve
3494-01 Nitrate-Nitrogen Octa-Slide 2 Bar, 0–15 ppm
*6905A-6999ABOX *Chlorine DPD #1R Tablet (50) and *Chlorine DPD #3R Tablet (50)
0106 Test Tube, Plastic, 2.5, 5 & 10 mL, w/caps (2)
3401-01 Chlorine Octa-Slide 2 Bar, 0.2–3.0 ppm
6354-L Conductivity Standard, 1413 μmhos/cm (990 ppm TDS), 500 mL
6418-L Conductivity Standard, 6,668 μmhos/cm (4,668 ppm TDS), 500 mL
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
20
21
blank
blank
PO Box 329 • Chestertown • MD • 21620
800-344-3100 • 410-778-3100 • f 410-778-6394
www.lamotte.com
64-3015-WT-01 7.14
Loading...