Read the instruction manual thoroughly to familiarize yourself with the test procedures before you begin.
Make note of any precautions in the instructions.
Read the labels on all LaMotte reagent containers prior to use. Some containers include precautionary notices
and fi rst aid information. Certain reagents are considered hazardous substances and are designated with a * in
the instruction manual. Material Safety Data Sheets (MSDS) are available at www.lamote.com. Read the MSDS
before using these reagents. Additional emergency information for all LaMotte reagents is available 24 hours
a day from the Poison Control Center listed in the front of the phone book. Be prepared to supply the name and
four-digit LaMotte code number found on the container label or at the top of the MSDS. LaMotte reagents are
registered with a computerized poison control information system available to all local poison control centers.
Keep equipment and reagent chemicals out of the reach of young children.
Protect Yourself and Equipment: Use Proper Analytical Techniques
Testing Hints
1.2.3.
Tightly close all
reagent containers
immediately after
use. Be sure not to
interchange caps
and pipets from
different containers.
Avoid
prolonged
exposure of
equipment
and reagents
to direct
sunlight.
Protect reagents
and components
from extreme
heat and cold.
4.5.
Wipe up any reagent chemical
spills, liquid or powder, as
soon as they occur. Refer to
label and accompanying
MSDS for proper
reagent disposal.
4
Use care when dispensing or
handling all reagents due to
safety reasons. Some
chemicals also may cause
permanent stains if spilled.
Page 5
ANALYTICAL TECHNIQUE
1. Clean glassware is a must for accurate results. Thoroughly rinse
test tubes before and after each test. Caps and stoppers should
also be cleaned after each use.
2. Use test tube caps or stoppers, not your fi ngers, to cover test
tubes and fl asks during shaking or mixing.
3. When adding sample to calibrated test tube, be sure vial is fi lled
to the appropriate mark. The bottom of the liquid (meniscus)
should be level with the desired mark. (Figure 1)
4. When dispensing reagents from bottles fi lled with dropper plug
and cap, be sure to hold bottle vertically and gently squeeze to
dispense the appropriate number of uniform drops. (Figure 2)
5. For those reagents to be added with the screwcap pipet assemblies
enclosed, remove polyseal cap on bottle and replace with the
screwcap pipet.
NOTE: Place the polyseal caps back on the reagent bottles for
longer periods of storage. Be sure that both pipet assemblies
and polyseal caps are thoroughly cleaned before placing on
bottles to avoid contamination.
6. When dispensing reagents from pipets, hold pipet vertically
to assure uniform drop size. This is extremely important when
performing drop count titrations. (Figure 3)
7. To fi ll pipets, squeeze rubber bulb and immerse into reagent.
Release bulb to fi ll. (Figure 4)
8. To add tablets, place over test vial with silver plastic side facing
up. Press tablet through the foil into the tube. (Figure 5)
5
Page 6
Model S and DuoSoft Softeners
INTRODUCTION
The Model S and the DuoSoft Water Softeners are designed to produce high quality softened water. As water
passes through the Model S chamber, the resin column causes scale-forming calcium and magnesium ions to be
exchanged for non-scale-forming sodium ions. When the resin is exhausted it must be replaced or regenerated.
Inexpensive
In the DuoSoft, water passes through both chambers and will be treated by both types of media. The two chambers of the DuoSoft may be easily fi lled with the media of choice for specifi c problem water.
NOTE: These softeners DO NOT yield water suitable for drinking .
READ INSTRUCTIONS BEFORE USE.
resin refi ll packages are available, or the original resin can be regenerated by chemical treatment.
OUTLET
HOSE
INLET
HOSE
CLAMP
CLAMP
PLUG
CHAMBER
FAUCET
ADAPTER
OUTLET
HOSE
INLET
HOSE
CLAMP
CLAMP
PLUG
TOP
CHAMBER
BOTTOM
CHAMBER
FAUCET
ADAPTER
6
Page 7
Instructions for Use
PROCEDURE A — Use of the Softeners
1.2.
Allow water
to run from
faucet briefl y to
visually gauge
pressure.
Attach adapter
on inlet hose
to faucet.
Fully open
clamp on inlet
hose.
3.4.
5.6.
7.
SLOWLY turn on
faucet and adjust the fl ow
of water into softener
until
the stream of water
is about 1/8” in diameter.
Maintain
prevent excessive pressure
in the adapter
Unclamp outlet
hose at bottom
of softener.
a gentle fl ow t
Adjust fl ow to
maintain the 1”
layer of water
over the media.
o
.
Let effl uent run
to waste for about
3 minutes to wash
out residue from
the previous
sample.
8.
After enough
softened water has
been collected, turn
off faucet and clamp
outlet hose.
Fill chamber
until a 1” layer
of water is
visible over
the media.
3
minutes
9.10.
When
softener is
disconnected
from faucet,
clamp inlet
hose.
If softener will be stored
unused for any
length of time,
keep a 1” layer
of water over
resin column to
prevent drying
and cracking.
7
Page 8
PROCEDURE B — Replacing the Media in the Softeners
1.2.
With the aid
of a funnel,
add the fresh
media to the
column.
3.4.
Continue adding media to
the column until the resin is about
2” from the top of the column.
2”
ADDITIONREMOVAL
5.6.
Clean threaded
area completely
and replace the
black plug in the
top of column.
Tap water may be
added to the media in the
funnel to facilitate fi lling
of the column.
Allow excess water
to drain from outlet
hose.
DuoSoft Unit
To add the media in the
bottom chamber, turn unit upside
down and repeat Addition Steps
1-5. When media has been added
to both chambers, proceed with
Procedure A. When using carbon
media, it is suggested that it be
used in the top chamber.
1.
T o replace media,
remove black
plug from the top
of dispenser.
3.4.
By adding more
water to the column
and shaking, the
remaining media can
be easily removed.
2.
T o remove the media in the
bottom chamber, turn unit up-
side down and repeat Removal
Steps 1-3. To add new media,
follow Addition Steps 1-5 for
Unclamp outlet hose
and turn softener
upside down.
Allow exhausted
media and water to
run to waste.
DuoSoft Unit
both chambers.
8
Page 9
PROCEDURE C — Regeneration of Media
Follow manufacturers’ instructions for regeneration of media. Cation exchange resin may be regenerated in the
following manner. NOTE: Due to build up of air pressure, it is not possible to regenerate both chambers of the
DuoSoft unit at the same time. Follow steps 1-8 to regenerate top chamber, then see step 9 to regenerate bottom
chamber.
1.2.
Prepare salt
solution by
dissolving about
½ pound of
salt (sodium
Chloride) in 1
quart of water.
3.4.
Clamp
outlet tube.
Add salt
solution
top chamber is
completely full.
until
5.6.
Unclamp outlet
tube and allow salt
solution to run to
waste.
Remove black
plug from top
of softener.
Allow to stand for about 30
minutes.
30
MINUTES
Clean threaded
area completely
and replace
black plug in
top of softener.
7.
Attach adapter
to faucet.
9.
For the DupSoft unit; turn unit upside down and repeat steps 1-8.
8.
Wash resin with
approximately
1 gallon of tap
water.
9
Page 10
Soap Demonstration
Calcium and Magnesium ions present in a water supply are the principle contributors to the total hardness.
Hard water tends to consume excessive quantities of soap and forms curds and deposits on glassware, fabrics, etc.
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a
Material Safety Data Sheet (MSDS) for these reagents go to www.lamotte.com. To obtain a printed copy,
contact LaMotte by e-mail, phone or fax.
TEST PROCEDURE
1.2.3.
Thoroughly rinse
the “SOFT”
water fl ask
(0453) with
softened water.
Thoroughly rinse
the “HARD”
water fl ask
(0452) with
untreated water.
4.5.6.
With the pipet
Fill the “SOFT”
fl ask (0453) with
softened water to
the same level.
7.9.8.
A thick lather will form
in the softened water.
(0392), add 4
drops of *Soap
Reagent #4
(4767) to each
fl ask.
Continue to add
*Soap Reagent #4
(4767), one drop at a
time, to the untreated
“HARD” water,
shake periodically
until a lather forms.
Count the number
of drops added.
Fill the “HARD”
fl ask (0452) with
untreated water
until the bottom
surface is covered
with a layer of
water about ½”
deep.
Cap and shake
the fl ask.
Extremely hard
water may require
30, 40, or even 60
drops of *Soap
Reagent #4 to
produce a lasting
lather.
10
Page 11
Precipitation Demonstration
Again, Calcium and Magnesium ions are the major contributors to water hardness. The chemical reagents in this
demonstration pull the Calcium and Magnesium ions out of solution to form a cloudy precipitate in hard water.
The water that has been run through the ion exchange column has had these ions removed, therefore, the sample
should remain clear.
NOTE: This portion of the AT-38/40 Water Quality Demo Kit is ONLY a visual demonstration illustrating the
removal of Calcium and Magnesium ions from tap water after treatment by the ion exchange process. The
results should not be interpreted beyond the intent of the demonstration.
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents go to www.lamotte.com. To obtain a printed copy, contact
LaMotte by e-mail, phone or fax..
TEST PROCEDURE
1.2.3.
Thoroughly rinse
the “SOFT”
water Demo Tube
(0298) with
softened water.
Thoroughly rinse
the “HARD”
water Demo Tube
(0297) with
softened water.
4.5.6.
Fill the “HARD”
Demo tube
(0297) to the line
with untreated
water.
Add 7 drops of
*Precipitation
Reagent A
(4542WT)
to each tube.
7.9.8.
Cap and mix.
Add 7 drops of
*Precipitation
Reagent B
(4543WT)
to each tube.
Fill the “SOFT”
Demo tube
(0298) to the line
with softened
water.
Cap and mix.
Place tubes in the
Precipitation Rack (0879)
and allow the tubes to
stand for
5 minutes.
Heavy
Precipitate
Remains
clear
11
Page 12
Color
Q
DW
ColorQ Photometer for Drinking Water
For software Version 5.73 or higher
To insure proper testing results.
• Insert tubes into ColorQ with line
and arrow facing forward.
• Brush and rinse all tubes
promptly after use.
• Rinse tablet crusher between test
factors.
• Allow tubes to dry before
returning to case.
• Replace reagent caps securely.
• Use only LaMotte replacement
reagents and ColorQ test tubes
Code 0201.
• If unit should auto-off during
testing rinse and fi ll one tube to
Blank and return to testing.
12
• To skip test in the sequence press
button to advance past that test
factor.
• To turn meter off at any time,
press and hold the button until
“off” is displayed.
• bAt = battery is low.
WARNING! This set contains chemicals that may be harmful
if misused. Read cautions on individual containers carefully.
Not to be used by children except under adult supervision.
Page 13
BLANK (CALIBRATION)
1
BLA
A
L
Color
Q
1. Fill sample bottle with water sample. Replace cap.
2. Fill clean tube (#0201) to the 5mL line with water sample.
3. Insert tube into ColorQ as shown.
4. Press button to turn meter on. When “bLA” appears press button to
“Blank” the meter, and go to pH. Remove tube.
pHColorQ Range: 4.0-9.0
2
pH
pH - Code 7059-E
5
1. Add 5 drops Wide Range pH Reagent (7059) to the same tube.
2. Cap. Invert 3 times to mix. Insert tube into ColorQ.
3. Press button to read “pH” pH. Remove tube.
* Reagent is a potential health hazard. Read MSDS at www.lamotte.com.
h
p
Color
Q
13
Page 14
HARDNESSColorQ Range: 1-41 GPG
*Hardness 1 Buffer: Code 7045-G
3
hb
*Hardness 2 Indicator: Code 7046-G
Color
Q
55
+
1. Press button to go to “Hd” Hardness.
2. Fill a clean tube (0201) to the 5 mL line with water sample.
3. Add 5 drops Hardness 1 Buffer (7045) and Hardness 2 Indicator (7046)
to same tube.
4. Cap. Invert 3 times to mix. Insert tube into ColorQ.
5. Press button to read “Hd” Hardness in gpg. Remove tube.
DILUTION
Samples with hardness concentrations above 41 gpg must be diluted. The
meter will display “Hi”.
A. Use the 1 mL pipet (0354) to add 2 measures (2mL) water
sample to a clean test tube (0201).
B. Dilute to the 5 mL line with distilled or deionized water. Cap and mix.
C. Blank with the diluted sample from Step B. Follow test procedure Step
3-5 with dluted sample. Multiply the displayed result by 2.5
Color
Q
14
NOTE: To convert from GPG to ppm; Multiply GPG by 17.1
* Reagent is a potential health hazard. Read MSDS at www.lamotte.com.
Page 15
TOTAL IRONColorQ Range: 0.0-3.0 ppm
4
i
Color
i
Q
*Ir - Code 3725A-H
1
1. Press button to go to “Ir” Iron.
2. Fill a clean tube (0201) to the 5 mL line with water sample.
3. Add 1 IRON (3725) Tablet. Crush with tablet crusher (0175).
4. Cap. Invert 3 times to mix. Tablet will not disslve completely.
Insert tube into Color Q.
5. Press button to read “Ir” Total Iron in ppm. Remove tube.
DILUTION
Samples with iron concentrations above 3 ppm must be diluted. The meter
will display “Hi”.
A. Use the 1 mL pipet (0354) to add 1 mL water sample to a test tube.
B. Dilute to the 5 mL line with distilled or deionized water. Cap and mix.
C. Blank with the diluted sample from Step B. Follow Steps 3-5 of the test
procedure with the diluted sample.. Multiply the displayed result by 5.
i
Color
Q
* Reagent is a potential health hazard. Read MSDS at www.lamotte.com.
15
Page 16
FREE / TOTAL CHLORINEColorQ Range: 0.0-10.0 ppm Chlorine
5
+
Color
Fcl
L
C
Q
/
+cl
/
*DPD 1 - Code 6903A-H
*DPD 3 - Code 6197A-H
1
Free Chlorine and Total Chlorine
1. Press button to go to “FCL” Free Chlorine.
2. Fill a clean tube (0201) to the 5 mL line with water sample.
3. Add 1 DPD 1(6903A) Tablet. Cap. Shake for TEN seconds. Invert
slowly 5 times.
4. Press button to read “FCL” Free Chlorine in ppm. Remove tube.
1. Press button to go to “tCL” Total Chlorine.
2. Remove cap from reacted FCL tube.
3. Add 1 DPD 3(6197A) Tablet. Cap. Shake for TEN seconds. Invert
slowly 5 times.
4. Insert tube into ColorQ.
5. Press button to read “tCL” Total Chlorine in ppm. Remove tube.
C
+
Color
L
Q
16
Total Chlorine Only
1. Press button past “FCL” Free Chlorine to go to “tCL” Total Chlorine.
2. Fill a clean tube (0201)to the 5 mL line with water sample.
3. Add 1 DPD 1(6903A) Tablet. Cap. Shake for TEN seconds. Invert slowly
5 times.
4. Remove cap. Add 1 DPD 3(6197A) Tablet. Cap. Shake for TEN seconds.
Invert slowly 5 times.
5. Insert tube into ColorQ.
6. Press button to read “tCL” Total Chlorine in ppm. Remove tube.
*Reagent is a potential health hazard. Read MSDS at www.lamotte.com.
Page 17
FREE / TOTAL CHLORINEColorQ Range: 0.0-10.0 ppm Chlorine
*Sulfi de Reagent A: Code V-4458LWT-G
6
Color
Q
*Sulfi de Reagent A: Code V-4458LWT-G
*Sulfi de Reagent A: Code V-4458LWT-G
+
5
3
Color
Q
5
1. Press button to go to “SuL” Sulfi de.
2. Fill a clean tube (0201) to the 5 mL line with water sample.
3. Add 5 drops of SULFIDE REAGENT A (4458).
4. Add 3 drops of SULFIDE REAGENT B (4459).
5. Cap and invert 3 times to mix. Allow tube to sit for ONE minute.
6. Add 5 drops of SULFIDE REAGENT C (4460).
7. Cap. Invert 3 times to mix. Insert tube into ColorQ.
8. Press button to read “SuL” Total Sulfi de in ppm. Remove tube.
DILUTION
Samples with sulfi de concentrations above 3 ppm must be diluted. The meter
will display “Hi”.
A. Use the 1 mL pipet (0354) to add 1 mL water sample to a test tube.
B. Dilute to the 5 mL line with distilled or deionized water. Cap and mix.
C. Blank with the diluted sample from Step B. Follow the test procedure,
steps 3 - 8, with the diluted sample. Multiply the displayed result by 5.
* Reagent is a potential health hazard. Read MSDS at www.lamotte.com.
17
Page 18
NITRATE-NITROGENColorQ Range: 0-15 ppm
7
Color
Q
*Nitrate - Code 3884A-H
1
1. Press button to go to “nit” Nitrate.
2. Fill a clean tube (0201) to the 5 mL line with water sample.
3. Add 1 NITRATE (3884A) Tablet. Crush with tablet crusher.
4. Cap. Invert 30 times per minute for TWO minutes.
5. Wait FIVE minutes. Insert tube into Color Q.
6. Press button to read “nit” Nitrate-Nitrogen in ppm. Remove tube.
Color
Q
18
*Reagent is a potential health hazard. Read MSDS at www.lamotte.com.
Page 19
Color
Q
User Tips
Testing Tips
• Read all instructions before testing.
• If not testing all tests in the sequence, follow
blanking procedure and then press button
until desired test factor is shown. Follow test
procedure.
• In one complete inversion, the tube will be
turned cap down and then right side up. The air
bubble will move slowly to the bottom of the
tube and back again to the cap end.
• If small bubbles form after adding reagents,
tap bottom of tube sharply once or twice to
dislodge bubbles. Bubbles will interfere with
the test results.
• Be sure the outside of the tubes are dry before
putting them into the ColorQ. This is best
accomplished by dispensing the water sample
from the sample bottle.
Low pH Low pH levels can signifi cantly
lower the hardness results.
Error Messages
Er 3 = Insuffi cient light is reaching the detector.
Sample may be too dark, or tube may be in
sideways. Turn off meter and retest.
Er 7 = The blank was not properly set. Turn
off meter and retest blank making sure the black
arrow on the tube is facing forward, toward the
user (not left or right).
SUN = Extremely bright sunlight is penetrating
the chamber. Turn your back to the sun to shade
the meter from bright sunlight.
Lo = Result is below the effective measuring
range for the system (e.g. < 4.0 pH for Code
7059 reagent.)
Hi = Result is above the effective measuring
range for the system (e.g. >9.0 pH for Code 7059
reagent.)
bAt = Battery is low.
Maintenance
• Clean ColorQ optics with a damp cotton swab.
Avoid abrasive cleaners that can damage
plastic.
• Replace stained or scratched tubes (0201).
Battery Replacement
Use a small Phillips head screwdriver to remove
all 5 screws in the base. Gently pull the battery
holder from the inside and replace TWO AA
style batteries. Replace screws securely but do
not over-tighten screws which can tear the rubber
base.
8/11 62059-01
19
Page 20
TDS • CODE 5-0080 • Dual Range TDS Pocketester
BEFORE FIRST USE: Soak electrodes for a few minutes in alcohol to remove oils. Rinse well with distilled
water and shake dry.
CALIBRATION
(To Be Done Once a Month)
Use a calibration standard between 900 ppm and 5,000 ppm for low range testing. LaMotte conductivity standard #6354 (989 ppm TDS) is most commonly used.
Pour ½” to 1”
1.2.3.
of a Calibration
Standard into 2
containers, and
tap or deionized
water into a
third container.
4.5.6.
Rinse electrode
in water. Then
rinse in fi rst container
of standard, then dip
into the second.
Open battery compartment lid
(end with lanyard loop). The
two white buttons are Increment
(INC) and Decrement (DEC)
calibration keys.
Turn the meter on, and wait
several minutes
to allow the
display to stabilize.
DEC Key
INC Key
Press in INC or DEC keys to
adjust reading to match the
calibration standard value.
After 3 seconds without a key
positive
+
negative
batteries
Figure #1
plastic
ribbon
press, the display fl ashes 3 times,
then shows “ENT”. The tester
accepts calibration value; returns
to measurements mode.
7.
Replace battery cap.
Rinse the electrode in
the water and proceed
with test procedure.
20
Page 21
PROCEDURE
1.2.3.
Remove the cap.
Press ON/OFF
button to turn on.
Dip the electrode end
into the test solution.
Make sure the sensor
is fully covered. Stir
once. Let the reading
stabilize.
4.5.6.
7.
Allow time for
the Automatic
T emperature
Compensation to
correct the
readings for
solution
temperature
changes.
Replace the cap.
You can replace the electrode module at the
fraction of the cost of a new Testr. When the
Testr fails to calibrate or gives fl uctuating readings in calibration standards,
you need to change the electrode.
1. With dry hands, grip the ribbed Testr collar with electrode facing you.
Twist the collar counter clockwise. Save the ribbed T estr collar and
O-ring for later use.
2. Pull the old electrode module away from the Testr.
3. Align the four tabs on the new module so they match the four slots
on the Testr.
4. Gently push the module onto the slots to seat in position. Push the
smaller O-ring fully onto the new electrode module. Push the collar
over the module and thread it onto place by fi rmly twisting it clockwise.
Note the reading
once the display
stabilizes. If the
solution is above the
range of the
TDST estr, the
display will show:
Press the ON/OFF
button to turn off Testr.
Note: Testr will
automatically shut
off after 8.5
minutes of
nonuse.
HOLD FUNCTION: Press HOLD key to freeze display. Press HOLD again to release.
MAINTENANCE: To improve performance, clean the stainless steel electrodes by periodically rinsing them
in alcohol for 10-15 minutes. Replace all 4 batteries if the display becomes faint or disappears, or if the readings
are unstable or constant. If drift is detected while electrodes are continuously exposed to solution for longer than
one hour, allow electrode to fully dry off periodically.
CHANGING BATTERIES: Open the battery compartment lid. Remove old batteries and replace with fresh
ones noting polarity as shown in the battery compartment. Recalibrate Testr after battery change.
21
Page 22
Replacement Parts
CODEDESCRIPTIONCODEDESCRIPTION
1002 Model “S” softener (AT-38Q) 0688 Sample bottle, 60 mL with fl ip top
1022 DuoSoft softener (AT-40Q) S-0080 Dual Range TDS PockeTester
2050-DW6 ColorQ DW photometer S-0084 Replacement electrode for Dual
Range TDS PockeTester
0452 Flask, 250 mL, “HARD” with cap 6354-J Optional Calibration standard 989
ppm; 100 mL
0453 Flask, 250 mL, “SOFT” with cap 7059-E WR pH ColorQ reagent
0672 Stopper for 250 mL fl ask*7045-G* Hardness 1 Buffer
*4767-H*Soap reagent #4 60 mL*7046-G* Hardness 2 Indicator
*4767-L*Soap reagent #4 500 mL 0354 Pipet, 1 mL plastic
0392 Pipet, plain plastic with cap 3725 A-H* Iron tablet box of 50
0297 Test tube, “HARD” with cap*3725 A-J*Iron tablet box of 100
0298 Test tube, “SOFT” with cap*6903 A-H*DPD #1, instrument grade box of 50
0651 Stopper for test tubes above*6903 A-J*DPD#1, instrument grade box of 100
0879 Precipitation rack, acrylic*6197 A-H*DPD #3, instrument grade box of 50
*4542WT-H*Precipitation reagent A, 60 mL*6197 A-J*DPD #3, instrument grade box of 100
*4542-L*Precipitation reagent A, 500 mL*V-4458 LWT-G*Sulfi de Reagent A, 60 mL
*4543WT-H*Precipitation reagent B, 60 mL*V-4459-E*Sulfi de Reagent B, 15 mL
*4543-L*Precipitation reagent B, 500 mL 4460 LWT-G4 Sulfi de Reagent C, 60 mL
0392 Pipet, plain plastic with cap* 3384A-H*Nitrate ColorQ TesTab, box of 50
0201 ColorQ test tube with cap 0175 Tablet crusher
4-3051CS Case & foam insert
22
WARNING: Reagents marked with an * are considered to be potential
health hazards. To view or print a Material Safety Data Sheet (MSDS)
for these reagents go to www.lamotte.com. To obtain a printed copy,
contact LaMotte by e-mail, phone or fax.
Page 23
AT-38 & AT-40 Diagram Layout
SOFT
SOAP
4767-H
Iron
3725A-H
0201
-NC
0201
-NC
DPD #3
6197A-H
0201
-NC
WT-G
4460L
Sulfide C
6903A-H
0201
-NC
Sulfide B
V-4459-E
DPD #1
3884A-H
0201
-NC
WT-G
Sufide A
V-4458L
Nitrate
0201
-NC
CH2
7046-G
0809-C c addy
& 0354 pipette (1mL)
CH 1
WR pH
7059-E
Single Softener 1002
0688
0879
0298
0297
4543WT-H
7045-G
26905
0688
0453
0452
4542WT-H
HARD
PRECIPITATION
23
Page 24
PO Box 329 • Chestertown • MD • 21620
800-344-3100 • 410-778-3100 • f 410-778-6394
www.lamotte.com
64-3051-WT-01 8.11
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.