LaMotte ATQ Digital Series Water Softener Demo Kit User Manual

Page 1
Model S/DuoSoft
with ColorQ
water quality
DEMO OUTFIT
AT-Q
38/40
Page 2
o
Page 3
Table of Contents
Analytical T echnique ...................................................................... 5
Introduction .................................................................................... 6
Instructions for Use
Procedure A - Use of the Softeners ................................................. 7
Procedure B - Replacing the Media in the Softeners ....................... 8
Procedure C - Regenerations of the Media .................................... 9
Soap Demonstration ................................................................. 10
Precipitation Demonstration ..................................................... 11
ColorQ DW Photometer ........................................................... 12
Introduction ................................................................................... 12
Blank ............................................................................................. 13
pH .................................................................................................. 13
Hardness ........................................................................................ 14
Iron ................................................................................................ 15
Chlorine ......................................................................................... 16
Sulfi de ........................................................................................... 17
Nitrate ........................................................................................... 18
ColorQ User Tips ...................................................................... 19
TDS Meter (Optional) ............................................................. 20-21
Replacement Parts .................................................................... 22
AT-38 & AT-40 Diagram Layout .............................................. 23
Page 4
Safety Information
Read the instruction manual thoroughly to familiarize yourself with the test procedures before you begin. Make note of any precautions in the instructions.
Read the labels on all LaMotte reagent containers prior to use. Some containers include precautionary notices and fi rst aid information. Certain reagents are considered hazardous substances and are designated with a * in the instruction manual. Material Safety Data Sheets (MSDS) are available at www.lamote.com. Read the MSDS before using these reagents. Additional emergency information for all LaMotte reagents is available 24 hours a day from the Poison Control Center listed in the front of the phone book. Be prepared to supply the name and four-digit LaMotte code number found on the container label or at the top of the MSDS. LaMotte reagents are registered with a computerized poison control information system available to all local poison control centers.
Keep equipment and reagent chemicals out of the reach of young children. Protect Yourself and Equipment: Use Proper Analytical Techniques
Testing Hints
1. 2. 3.
Tightly close all reagent containers immediately after use. Be sure not to interchange caps and pipets from different containers.
Avoid prolonged exposure of equipment and reagents to direct sunlight.
Protect reagents and components from extreme heat and cold.
4. 5.
Wipe up any reagent chemical spills, liquid or powder, as soon as they occur. Refer to label and accompanying MSDS for proper reagent disposal.
4
Use care when dispensing or handling all reagents due to safety reasons. Some chemicals also may cause permanent stains if spilled.
Page 5
ANALYTICAL TECHNIQUE
1. Clean glassware is a must for accurate results. Thoroughly rinse test tubes before and after each test. Caps and stoppers should also be cleaned after each use.
2. Use test tube caps or stoppers, not your fi ngers, to cover test tubes and fl asks during shaking or mixing.
3. When adding sample to calibrated test tube, be sure vial is fi lled to the appropriate mark. The bottom of the liquid (meniscus) should be level with the desired mark. (Figure 1)
4. When dispensing reagents from bottles fi lled with dropper plug and cap, be sure to hold bottle vertically and gently squeeze to dispense the appropriate number of uniform drops. (Figure 2)
5. For those reagents to be added with the screwcap pipet assemblies enclosed, remove polyseal cap on bottle and replace with the screwcap pipet. NOTE: Place the polyseal caps back on the reagent bottles for longer periods of storage. Be sure that both pipet assemblies and polyseal caps are thoroughly cleaned before placing on bottles to avoid contamination.
6. When dispensing reagents from pipets, hold pipet vertically to assure uniform drop size. This is extremely important when performing drop count titrations. (Figure 3)
7. To fi ll pipets, squeeze rubber bulb and immerse into reagent. Release bulb to fi ll. (Figure 4)
8. To add tablets, place over test vial with silver plastic side facing up. Press tablet through the foil into the tube. (Figure 5)
5
Page 6
Model S and DuoSoft Softeners
INTRODUCTION
The Model S and the DuoSoft Water Softeners are designed to produce high quality softened water. As water passes through the Model S chamber, the resin column causes scale-forming calcium and magnesium ions to be exchanged for non-scale-forming sodium ions. When the resin is exhausted it must be replaced or regenerated. Inexpensive
In the DuoSoft, water passes through both chambers and will be treated by both types of media. The two cham­bers of the DuoSoft may be easily fi lled with the media of choice for specifi c problem water.
NOTE: These softeners DO NOT yield water suitable for drinking . READ INSTRUCTIONS BEFORE USE.
resin refi ll packages are available, or the original resin can be regenerated by chemical treatment.
OUTLET
HOSE
INLET HOSE
CLAMP
CLAMP
PLUG
CHAMBER
FAUCET
ADAPTER
OUTLET
HOSE
INLET HOSE
CLAMP
CLAMP
PLUG
TOP
CHAMBER
BOTTOM
CHAMBER
FAUCET
ADAPTER
6
Page 7
Instructions for Use
PROCEDURE A — Use of the Softeners
1. 2.
Allow water to run from faucet briefl y to visually gauge pressure.
Attach adapter on inlet hose to faucet. Fully open clamp on inlet hose.
3. 4.
5. 6.
7.
SLOWLY turn on faucet and adjust the fl ow of water into softener until
the stream of water is about 1/8” in diameter. Maintain prevent excessive pressure in the adapter
Unclamp outlet hose at bottom of softener.
a gentle fl ow t
Adjust fl ow to maintain the 1” layer of water over the media.
o
.
Let effl uent run to waste for about 3 minutes to wash out residue from the previous sample.
8.
After enough softened water has been collected, turn off faucet and clamp outlet hose.
Fill chamber until a 1” layer of water is visible over the media.
3
minutes
9. 10.
When softener is disconnected from faucet, clamp inlet hose.
If softener will be stored unused for any length of time, keep a 1” layer of water over resin column to prevent drying and cracking.
7
Page 8
PROCEDURE B — Replacing the Media in the Softeners
1. 2.
With the aid of a funnel, add the fresh media to the column.
3. 4.
Continue adding media to the column until the resin is about 2” from the top of the column.
2”
ADDITIONREMOVAL
5. 6.
Clean threaded area completely and replace the black plug in the top of column.
Tap water may be added to the media in the funnel to facilitate fi lling of the column.
Allow excess water to drain from outlet hose.
DuoSoft Unit
To add the media in the bottom chamber, turn unit upside down and repeat Addition Steps 1-5. When media has been added to both chambers, proceed with Procedure A. When using carbon media, it is suggested that it be used in the top chamber.
1.
T o replace media, remove black plug from the top of dispenser.
3. 4.
By adding more water to the column and shaking, the remaining media can be easily removed.
2.
T o remove the media in the
bottom chamber, turn unit up-
side down and repeat Removal
Steps 1-3. To add new media, follow Addition Steps 1-5 for
Unclamp outlet hose and turn softener upside down.
Allow exhausted media and water to run to waste.
DuoSoft Unit
both chambers.
8
Page 9
PROCEDURE C — Regeneration of Media
Follow manufacturers’ instructions for regeneration of media. Cation exchange resin may be regenerated in the following manner. NOTE: Due to build up of air pressure, it is not possible to regenerate both chambers of the DuoSoft unit at the same time. Follow steps 1-8 to regenerate top chamber, then see step 9 to regenerate bottom chamber.
1. 2.
Prepare salt solution by dissolving about ½ pound of salt (sodium Chloride) in 1 quart of water.
3. 4.
Clamp outlet tube.
Add salt solution top chamber is completely full.
until
5. 6.
Unclamp outlet tube and allow salt solution to run to waste.
Remove black plug from top of softener.
Allow to stand for about 30 minutes.
30
MINUTES
Clean threaded area completely and replace black plug in top of softener.
7.
Attach adapter to faucet.
9.
For the DupSoft unit; turn unit upside down and repeat steps 1-8.
8.
Wash resin with approximately 1 gallon of tap water.
9
Page 10
Soap Demonstration
Calcium and Magnesium ions present in a water supply are the principle contributors to the total hardness. Hard water tends to consume excessive quantities of soap and forms curds and deposits on glassware, fabrics, etc.
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
TEST PROCEDURE
1. 2. 3.
Thoroughly rinse the “SOFT” water fl ask (0453) with softened water.
Thoroughly rinse the “HARD” water fl ask (0452) with untreated water.
4. 5. 6.
With the pipet
Fill the “SOFT” ask (0453) with softened water to the same level.
7. 9.8.
A thick lather will form in the softened water.
(0392), add 4 drops of *Soap Reagent #4 (4767) to each ask.
Continue to add *Soap Reagent #4 (4767), one drop at a time, to the untreated “HARD” water, shake periodically until a lather forms. Count the number of drops added.
Fill the “HARD” ask (0452) with untreated water until the bottom surface is covered with a layer of water about ½” deep.
Cap and shake the fl ask.
Extremely hard water may require 30, 40, or even 60 drops of *Soap Reagent #4 to produce a lasting lather.
10
Page 11
Precipitation Demonstration
Again, Calcium and Magnesium ions are the major contributors to water hardness. The chemical reagents in this demonstration pull the Calcium and Magnesium ions out of solution to form a cloudy precipitate in hard water. The water that has been run through the ion exchange column has had these ions removed, therefore, the sample should remain clear.
NOTE: This portion of the AT-38/40 Water Quality Demo Kit is ONLY a visual demonstration illustrating the removal of Calcium and Magnesium ions from tap water after treatment by the ion exchange process. The results should not be interpreted beyond the intent of the demonstration.
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Mate­rial Safety Data Sheet (MSDS) for these reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax..
TEST PROCEDURE
1. 2. 3.
Thoroughly rinse the “SOFT” water Demo Tube (0298) with softened water.
Thoroughly rinse the “HARD” water Demo Tube (0297) with softened water.
4. 5. 6.
Fill the “HARD” Demo tube (0297) to the line with untreated water.
Add 7 drops of *Precipitation Reagent A (4542WT) to each tube.
7. 9.8.
Cap and mix.
Add 7 drops of *Precipitation Reagent B (4543WT) to each tube.
Fill the “SOFT” Demo tube (0298) to the line with softened water.
Cap and mix.
Place tubes in the Precipitation Rack (0879) and allow the tubes to stand for 5 minutes.
Heavy Precipitate
Remains clear
11
Page 12
Color
Q
DW
ColorQ Photometer for Drinking Water
For software Version 5.73 or higher
To insure proper testing results.
• Insert tubes into ColorQ with line and arrow facing forward.
• Brush and rinse all tubes promptly after use.
• Rinse tablet crusher between test factors.
• Allow tubes to dry before returning to case.
• Replace reagent caps securely.
• Use only LaMotte replacement reagents and ColorQ test tubes Code 0201.
• If unit should auto-off during testing rinse and fi ll one tube to Blank and return to testing.
12
• To skip test in the sequence press button to advance past that test factor.
• To turn meter off at any time, press and hold the button until “off” is displayed.
• bAt = battery is low.
WARNING! This set contains chemicals that may be harmful if misused. Read cautions on individual containers carefully. Not to be used by children except under adult supervision.
Page 13
BLANK (CALIBRATION)
1
BLA
A
L
Color
Q
1. Fill sample bottle with water sample. Replace cap.
2. Fill clean tube (#0201) to the 5mL line with water sample.
3. Insert tube into ColorQ as shown.
4. Press button to turn meter on. When “bLA” appears press button to “Blank” the meter, and go to pH. Remove tube.
pH ColorQ Range: 4.0-9.0
2
pH
pH - Code 7059-E
5
1. Add 5 drops Wide Range pH Reagent (7059) to the same tube.
2. Cap. Invert 3 times to mix. Insert tube into ColorQ.
3. Press button to read “pH” pH. Remove tube.
* Reagent is a potential health hazard. Read MSDS at www.lamotte.com.
h
p
Color
Q
13
Page 14
HARDNESS ColorQ Range: 1-41 GPG
*Hardness 1 Buffer: Code 7045-G
3
hb
*Hardness 2 Indicator: Code 7046-G
Color
Q
5 5
+
1. Press button to go to “Hd” Hardness.
2. Fill a clean tube (0201) to the 5 mL line with water sample.
3. Add 5 drops Hardness 1 Buffer (7045) and Hardness 2 Indicator (7046) to same tube.
4. Cap. Invert 3 times to mix. Insert tube into ColorQ.
5. Press button to read “Hd” Hardness in gpg. Remove tube.
DILUTION
Samples with hardness concentrations above 41 gpg must be diluted. The meter will display “Hi”.
A. Use the 1 mL pipet (0354) to add 2 measures (2mL) water sample to a clean test tube (0201).
B. Dilute to the 5 mL line with distilled or deionized water. Cap and mix. C. Blank with the diluted sample from Step B. Follow test procedure Step
3-5 with dluted sample. Multiply the displayed result by 2.5
Color
Q
14
NOTE: To convert from GPG to ppm; Multiply GPG by 17.1
* Reagent is a potential health hazard. Read MSDS at www.lamotte.com.
Page 15
TOTAL IRON ColorQ Range: 0.0-3.0 ppm
4
i
Color
i
Q
*Ir - Code 3725A-H
1
1. Press button to go to “Ir” Iron.
2. Fill a clean tube (0201) to the 5 mL line with water sample.
3. Add 1 IRON (3725) Tablet. Crush with tablet crusher (0175).
4. Cap. Invert 3 times to mix. Tablet will not disslve completely. Insert tube into Color Q.
5. Press button to read “Ir” Total Iron in ppm. Remove tube.
DILUTION
Samples with iron concentrations above 3 ppm must be diluted. The meter will display “Hi”.
A. Use the 1 mL pipet (0354) to add 1 mL water sample to a test tube. B. Dilute to the 5 mL line with distilled or deionized water. Cap and mix. C. Blank with the diluted sample from Step B. Follow Steps 3-5 of the test
procedure with the diluted sample.. Multiply the displayed result by 5.
i
Color
Q
* Reagent is a potential health hazard. Read MSDS at www.lamotte.com.
15
Page 16
FREE / TOTAL CHLORINE ColorQ Range: 0.0-10.0 ppm Chlorine
5
+
Color
Fcl
L
C
Q
/
+cl
/
*DPD 1 - Code 6903A-H *DPD 3 - Code 6197A-H
1
Free Chlorine and Total Chlorine
1. Press button to go to “FCL” Free Chlorine.
2. Fill a clean tube (0201) to the 5 mL line with water sample.
3. Add 1 DPD 1(6903A) Tablet. Cap. Shake for TEN seconds. Invert slowly 5 times.
4. Press button to read “FCL” Free Chlorine in ppm. Remove tube.
1. Press button to go to “tCL” Total Chlorine.
2. Remove cap from reacted FCL tube.
3. Add 1 DPD 3(6197A) Tablet. Cap. Shake for TEN seconds. Invert slowly 5 times.
4. Insert tube into ColorQ.
5. Press button to read “tCL” Total Chlorine in ppm. Remove tube.
C
+
Color
L
Q
16
Total Chlorine Only
1. Press button past “FCL” Free Chlorine to go to “tCL” Total Chlorine.
2. Fill a clean tube (0201)to the 5 mL line with water sample.
3. Add 1 DPD 1(6903A) Tablet. Cap. Shake for TEN seconds. Invert slowly 5 times.
4. Remove cap. Add 1 DPD 3(6197A) Tablet. Cap. Shake for TEN seconds. Invert slowly 5 times.
5. Insert tube into ColorQ.
6. Press button to read “tCL” Total Chlorine in ppm. Remove tube.
*Reagent is a potential health hazard. Read MSDS at www.lamotte.com.
Page 17
FREE / TOTAL CHLORINE ColorQ Range: 0.0-10.0 ppm Chlorine
*Sulfi de Reagent A: Code V-4458LWT-G
6
Color
Q
*Sulfi de Reagent A: Code V-4458LWT-G *Sulfi de Reagent A: Code V-4458LWT-G
+
5
3
Color
Q
5
1. Press button to go to “SuL” Sulfi de.
2. Fill a clean tube (0201) to the 5 mL line with water sample.
3. Add 5 drops of SULFIDE REAGENT A (4458).
4. Add 3 drops of SULFIDE REAGENT B (4459).
5. Cap and invert 3 times to mix. Allow tube to sit for ONE minute.
6. Add 5 drops of SULFIDE REAGENT C (4460).
7. Cap. Invert 3 times to mix. Insert tube into ColorQ.
8. Press button to read “SuL” Total Sulfi de in ppm. Remove tube.
DILUTION
Samples with sulfi de concentrations above 3 ppm must be diluted. The meter will display “Hi”.
A. Use the 1 mL pipet (0354) to add 1 mL water sample to a test tube. B. Dilute to the 5 mL line with distilled or deionized water. Cap and mix. C. Blank with the diluted sample from Step B. Follow the test procedure,
steps 3 - 8, with the diluted sample. Multiply the displayed result by 5.
* Reagent is a potential health hazard. Read MSDS at www.lamotte.com.
17
Page 18
NITRATE-NITROGEN ColorQ Range: 0-15 ppm
7
Color
Q
*Nitrate - Code 3884A-H
1
1. Press button to go to “nit” Nitrate.
2. Fill a clean tube (0201) to the 5 mL line with water sample.
3. Add 1 NITRATE (3884A) Tablet. Crush with tablet crusher.
4. Cap. Invert 30 times per minute for TWO minutes.
5. Wait FIVE minutes. Insert tube into Color Q.
6. Press button to read “nit” Nitrate-Nitrogen in ppm. Remove tube.
Color
Q
18
*Reagent is a potential health hazard. Read MSDS at www.lamotte.com.
Page 19
Color
Q
User Tips
Testing Tips
• Read all instructions before testing.
• If not testing all tests in the sequence, follow blanking procedure and then press button until desired test factor is shown. Follow test procedure.
• In one complete inversion, the tube will be turned cap down and then right side up. The air bubble will move slowly to the bottom of the tube and back again to the cap end.
• If small bubbles form after adding reagents, tap bottom of tube sharply once or twice to dislodge bubbles. Bubbles will interfere with the test results.
• Be sure the outside of the tubes are dry before putting them into the ColorQ. This is best accomplished by dispensing the water sample from the sample bottle.
Low pH Low pH levels can signifi cantly
lower the hardness results.
Error Messages
Er 3 = Insuffi cient light is reaching the detector.
Sample may be too dark, or tube may be in sideways. Turn off meter and retest.
Er 7 = The blank was not properly set. Turn
off meter and retest blank making sure the black arrow on the tube is facing forward, toward the user (not left or right).
SUN = Extremely bright sunlight is penetrating
the chamber. Turn your back to the sun to shade the meter from bright sunlight.
Lo = Result is below the effective measuring
range for the system (e.g. < 4.0 pH for Code 7059 reagent.)
Hi = Result is above the effective measuring
range for the system (e.g. >9.0 pH for Code 7059 reagent.)
bAt = Battery is low.
Maintenance
• Clean ColorQ optics with a damp cotton swab. Avoid abrasive cleaners that can damage plastic.
• Replace stained or scratched tubes (0201).
Battery Replacement
Use a small Phillips head screwdriver to remove all 5 screws in the base. Gently pull the battery holder from the inside and replace TWO AA style batteries. Replace screws securely but do not over-tighten screws which can tear the rubber base.
8/11 62059-01
19
Page 20
TDS • CODE 5-0080 • Dual Range TDS Pocketester
BEFORE FIRST USE: Soak electrodes for a few minutes in alcohol to remove oils. Rinse well with distilled water and shake dry.
CALIBRATION
(To Be Done Once a Month)
Use a calibration standard between 900 ppm and 5,000 ppm for low range testing. LaMotte conductivity stan­dard #6354 (989 ppm TDS) is most commonly used.
Pour ½” to 1”
1. 2. 3.
of a Calibration Standard into 2 containers, and tap or deionized water into a third container.
4. 5. 6.
Rinse electrode in water. Then rinse in fi rst container of standard, then dip into the second.
Open battery compartment lid (end with lanyard loop). The two white buttons are Increment (INC) and Decrement (DEC) calibration keys.
Turn the meter on, and wait several minutes to allow the display to stabilize.
DEC Key
INC Key
Press in INC or DEC keys to adjust reading to match the calibration standard value. After 3 seconds without a key
positive
+
negative
batteries
Figure #1
plastic ribbon
press, the display fl ashes 3 times, then shows “ENT”. The tester accepts calibration value; returns to measurements mode.
7.
Replace battery cap. Rinse the electrode in the water and proceed with test procedure.
20
Page 21
PROCEDURE
1. 2. 3.
Remove the cap.
Press ON/OFF button to turn on.
Dip the electrode end into the test solution. Make sure the sensor is fully covered. Stir once. Let the reading stabilize.
4. 5. 6.
7.
Allow time for the Automatic T emperature Compensation to correct the readings for solution temperature changes.
Replace the cap.
You can replace the electrode module at the fraction of the cost of a new Testr. When the Testr fails to calibrate or gives fl uctuating readings in calibration standards, you need to change the electrode.
1. With dry hands, grip the ribbed Testr collar with electrode facing you. Twist the collar counter clockwise. Save the ribbed T estr collar and O-ring for later use.
2. Pull the old electrode module away from the Testr.
3. Align the four tabs on the new module so they match the four slots on the Testr.
4. Gently push the module onto the slots to seat in position. Push the smaller O-ring fully onto the new electrode module. Push the collar
over the module and thread it onto place by fi rmly twisting it clockwise.
Note the reading once the display stabilizes. If the solution is above the range of the TDST estr, the display will show:
Press the ON/OFF button to turn off Testr. Note: Testr will automatically shut off after 8.5 minutes of nonuse.
HOLD FUNCTION: Press HOLD key to freeze display. Press HOLD again to release. MAINTENANCE: To improve performance, clean the stainless steel electrodes by periodically rinsing them
in alcohol for 10-15 minutes. Replace all 4 batteries if the display becomes faint or disappears, or if the readings are unstable or constant. If drift is detected while electrodes are continuously exposed to solution for longer than one hour, allow electrode to fully dry off periodically.
CHANGING BATTERIES: Open the battery compartment lid. Remove old batteries and replace with fresh ones noting polarity as shown in the battery compartment. Recalibrate Testr after battery change.
21
Page 22
Replacement Parts
CODE DESCRIPTION CODE DESCRIPTION
1002 Model “S” softener (AT-38Q) 0688 Sample bottle, 60 mL with fl ip top 1022 DuoSoft softener (AT-40Q) S-0080 Dual Range TDS PockeTester 2050-DW6 ColorQ DW photometer S-0084 Replacement electrode for Dual
Range TDS PockeTester
0452 Flask, 250 mL, “HARD” with cap 6354-J Optional Calibration standard 989
ppm; 100 mL 0453 Flask, 250 mL, “SOFT” with cap 7059-E WR pH ColorQ reagent 0672 Stopper for 250 mL fl ask *7045-G * Hardness 1 Buffer *4767-H *Soap reagent #4 60 mL *7046-G * Hardness 2 Indicator *4767-L *Soap reagent #4 500 mL 0354 Pipet, 1 mL plastic 0392 Pipet, plain plastic with cap 3725 A-H * Iron tablet box of 50 0297 Test tube, “HARD” with cap *3725 A-J *Iron tablet box of 100 0298 Test tube, “SOFT” with cap *6903 A-H *DPD #1, instrument grade box of 50 0651 Stopper for test tubes above *6903 A-J *DPD#1, instrument grade box of 100 0879 Precipitation rack, acrylic *6197 A-H *DPD #3, instrument grade box of 50 *4542WT-H *Precipitation reagent A, 60 mL *6197 A-J *DPD #3, instrument grade box of 100 *4542-L *Precipitation reagent A, 500 mL *V-4458 LWT-G *Sulfi de Reagent A, 60 mL *4543WT-H *Precipitation reagent B, 60 mL *V-4459-E *Sulfi de Reagent B, 15 mL *4543-L *Precipitation reagent B, 500 mL 4460 LWT-G4 Sulfi de Reagent C, 60 mL 0392 Pipet, plain plastic with cap * 3384A-H *Nitrate ColorQ TesTab, box of 50 0201 ColorQ test tube with cap 0175 Tablet crusher
4-3051CS Case & foam insert
22
WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
Page 23
AT-38 & AT-40 Diagram Layout
SOFT
SOAP
4767-H
Iron
3725A-H
0201
-NC
0201
-NC
DPD #3
6197A-H
0201
-NC
WT-G
4460L
Sulfide C
6903A-H
0201
-NC
Sulfide B
V-4459-E
DPD #1
3884A-H
0201
-NC
WT-G
Sufide A
V-4458L
Nitrate
0201
-NC
CH2
7046-G
0809-C c addy
& 0354 pipette (1mL)
CH 1
WR pH
7059-E
Single Softener 1002
0688
0879
0298
0297
4543WT-H
7045-G
26905
0688
0453
0452
4542WT-H
HARD
PRECIPITATION
23
Page 24
PO Box 329 • Chestertown • MD • 21620
800-344-3100 • 410-778-3100 • f 410-778-6394
www.lamotte.com
64-3051-WT-01 8.11
Loading...