LaMotte AQ-5 User Manual

SALT WATER AQUACULTURE
TEST KIT
MODEL AQ-5 · CODE 3636-03
NOTE: It is important to read the instruction manual before attempting to perform the tests with the short form instructions provided below.
*WARNING: Reagents marked with a * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specified code number.
AM MO NIA NI TRO GEN
1. Fill test tube (0106) to 5 mL line with sample water.
2. Add 10 drops *Salicylate Ammonia #1 (3978LWT).
Cap and mix.
3. Add 7 drops of *Salicylate Ammonia #2 (3979WT). Cap and mix. Wait 1 minute.
4. Add 7 drops of Salicylate Ammonia #3 (3982WT). Cap and mix. Wait 20 minutes.
5. Insert test tube into Octa-Slide Viewer (1100) with Ammonia Nitrogen Octa-Slide Bar (3438) inserted.
6. Record as ppm Ammonia Nitrogen (NH3-N).
NI TRITE NI TRO GEN
1. Fill test tube (0106) to 2.5 mL line with sample water.
2. Dilute to 5 mL line with *Mixed Acid Reagent
(V-6278).
3. Use 0.1g spoon (0699) to add 0.1g of *Color Developing Reagent (V-6281). Cap and mix for 1 minute. Wait 5 minutes.
4. Insert test tube into Octa-Slide Viewer (1100) with Nitrite Nitrogen Octa-Slide Bar (3437) inserted.
5. Record as ppm Nitrite Nitrogen (NO2-N).
NI TRATE NI TRO GEN
1. Fill test tube (0106) to 2.5 mL line with sample water.
2. Dilute to 5 mL line with *Mixed Acid Reagent
(V-6278). Cap and mix. Wait 2 minutes.
3. Use 0.1g spoon (0699) to add one level measure of *Nitrate Reducing Reagent (V-6279). Cap and invert tube 50-60 times in one minute. Wait 10 minutes.
4. Invert sample one time and insert test tube into Octa-Slide Viewer (1100) with Nitrate Nitrogen Octa-Slide Bar (3413) inserted.
5. Record as ppm Nitrate Nitrogen (NO3 -N).
AL KA LIN ITY
1. Fill titration tube (0608) to 5 mL line with sample water.
2. Add 4 drops of *BCG-MR Indicator (2311-EG). Cap and mix. Sample will turn blue-green.
3. Fill Direct Reading Titrator (0382) with *Alkalinity Titration Reagent B (4493DR).
4. Titrate sample until blue-green color changes to pink.
5. Record as ppm Alkalinity (CaCO3).
CARBON DI OX IDE
1. Fill titration tube (0608) to 20 mL line with sample water.
2. Add 2 drops *Phenolphthalein Indicator, 1% (2246). If sample turns red, no free carbon dioxide is present. If colorless, proceed to Step 3.
3. Fill Direct Reading Titrator (0380) with *Carbon Dioxide Reagent B (4253DR).
4. Titrate sample until faint pink color persists for 30 seconds.
5. Record as ppm Carbon Dioxide (CO2).
LaMOTTE COM PANY
Helping Peo ple Solve An a lyt i cal Challenges
SM
800-344-3100 · 410-778-3100 (Out side U.S.A.) · Fax 410-778-6394 · www.lamotte.com
PO Box 329 · Chestertown · Mary land · 21620 · USA
3636-03-SF · 02/09