QUANTITY CONTENTS CODE
2 x 120 mL *Ammonia Nitrogen Absorbing Solution *7737-J
30 mL Ammonia Nitrogen Reagent #1 4797WT-G
30 mL *Ammonia Nitrogen Reagent #2 *4798WT-G
2 Test Tubes, 5 mL, glass, w/caps 0230
1 Ammonia In Air Comparator 7736
*WARNING: Reagents marked with an * are considered to be potential
health hazards. To view or print a Material Safety Data Sheet (MSDS) for
these reagents go to www.lamotte.com. To obtain a printed copy, contact
LaMotte by e-mail, phone or fax.
To individual reagents or test kit components, use the specified code numbers.
NOTE: The sampling period specified in
the instructions corresponds to
concentrations which may be
encountered in ambient air conditions.
High concentrations of pollutants in the
atmosphere will require shorter sampling
periods, while low concentrations will
require longer sampling periods.
USE OF THE OCTET COM PARA TOR
The Octet Comparator contains eight
permanent color standards. A test sample is
inserted into the openings in the top of the
comparator. The sample can then be compared
to four color standards at once, and the value
read off the comparator. For optimum color
comparison, the comparator should be
positioned between the operator and a light
source, so that the light enters through the
special light-diffusing screen in the back of the
comparator. Avoid viewing the comparator
against direct sunlight or an irregularly lighted
background.
PRO CE DURE
1. Pour 10 mL of *Ammonia Nitrogen Absorbing Solution (7737)
into impinging apparatus. Connect impinging apparatus to intake of
air sampling pump.
NOTE: Make sure the long tube is immersed in the absorbing
solution.
2. Set flowmeter to sample at a rate of 1.0 Lpm for 10 minutes or until
a measurable amount of ammonia is absorbed.
3. At the end of the sampling period, add contents of impinger to the
5 mL line of test tube (0230).
4. Add 2 drops of Ammonia Nitrogen Reagent #1 (4797WT). Cap
and mix.
5. Add 8 drops of *Ammonia Nitrogen Reagent #2 (4798WT). Cap
and mix. Solution will turn yellow to brown color if ammonia is
present.
6. Place test tube into Ammonia In Air Comparator (7736). Match
sample color to index of color standards. Note the index number
which gives the proper color match. Record results as ppm
Ammonia Nitrogen.
NOTE: To convert ammonia nitrogen to ammonia, multiply chart
value by 1.2. Record as ppm Ammonia.
AMMONIA IN AIR CALIBRATION CHART**
TIME COMPARATOR INDEX NUMBER
(min) 1 2 3 4 5 6 7 8
5
3.43 6.86 10.30 13.73 17.16 20.59 24.02 27.46
10
1.72 3.43 5.15 6.86 8.58 10.30 12.01 13.73
15
1.15 2.28 3.43 4.58 5.71 6.86 8.02 9.14
20
0.86 1.72 2.58 3.43 4.30 5.15 6.01 6.86
**Values in ppm