This booklet provides step-by-step instructions for each of the tests
L
M
o
tt
provided in the AG-104 test kit. Please refer to the separate sheets for
proper use of the Direct Reading Titrator and Axial Reader
Comparator where noted.
Use of the pH Oc tet Com para tor
The Octet Comparator contains eight permanent
color standards. A test sample is inserted into the
openings in the top of the comparator. The sample
can then be compared to four color standards at
once, and the value read off the comparator. For
optimum color comparison, the comparator should
be positioned between the operator and a light
source, so that the light enters through the special
light-diffusing screen in the back of the comparator.
Avoid viewing the comparator against direct
sunlight or an irregularly lighted background.
Use of the Octa-Slide Viewer
The Octa-Slide Viewer should be held so
non-direct light enters through the back of the
comparator. With sample tube inserted at top,
slide the Octa-Slide bar thorough the viewer
and march with color standard.
*WARNING: Reagents marked with a * are considered to be potential health hazards.
To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS
CD or www.lamotte.com. To obtain a printed copy, contact LaMotte by email, phone
or fax.
NOTE: Read Direct Reading Titrator Manual before performing test.
Pro ce dure
1. Fill the test tube (0778) to 5 mL line with sample water.
2. Add one BCG-MR Indicator Tablet (T-2311). Cap and shake until
dissolved. A blue-green color will develop.
3. Fill the Direct Reading Titrator (0382) with *Alkalinity Titration
Reagent B (4493DR). Insert Titrator into the center hole of the test
tube cap.
4. While gently swirling the tube, slowly depress plunger to titrate until
the blue-green color changes to pink. Read the test result directly
from the scale where the large ring on the Titrator meets the
Titrator barrel. Record results as Total Alkalinity in ppm (CaCO3).
Each minor division on Titrator scale = 4 ppm Total Alkalinity as
CaCO3.
EXAMPLE: Plunger tip is 3 minor divisions below line 140. The
test result is 140 plus (3 divisions x 4) equals 152 ppm.
5. If the plunger tip reaches the bottom line on the titrator scale (200
ppm) before the endpoint color change occurs, refill the titrator and
continue titration. When recording the test result, be sure to include
the value of the original amount of reagent dispensed (200 ppm).
When testing salt water aquariums, alkalinity may be called carbonate
hardness. Multiply ppm Alkalinity by 0.056 to convert to German
degrees of Hardness.
This test set provides Total Alkalinity readings only. For
comprehensive analysis of Phenolphthalein (P), Total (T), Hydroxide,
Carbonate, and Bicarbonate Alkalinity the Model WAT-MP-DR Test
Set, Code 4533-DR is recommended.
*WARNING: Reagents marked with a * are considered to be potential health hazards.
To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS
CD or www.lamotte.com. To obtain a printed copy, contact LaMotte by email, phone
or fax.
Pro ce dure
1. Fill a test tube (0106) to the 5 mL line with the water sample.
2. Add 10 drops of Salicylate Ammonia #1 (3978LWT). Cap and mix.
3. Add 7 drops of *Salicylate Ammonia #2 (3979WT). Cap and mix.
Wait one minute.
4. Add 7 drops of Salicylate Ammonia #3 (3982). Cap and mix. Wait
20 minutes.
5. Insert Ammonia Nitrogen Octa-Slide Bar (3441) into the
Octa-Slide Viewer (1100). Insert test tube into Octa-Slide Viewer.
Match sample color to a color standard. Record as ppm Ammonia
Nitrogen (NH
3
_
N).
To express results as Unionized Ammonia (NH3):
ppm Unionized Ammonia (NH3) =
ppm Ammonia Nitrogen (NH3–N) x 1.2
To express results as Ionized Ammonia (NH4):
ppm Ionized Ammonia (NH
+
) =
4
ppm Ammonia Nitrogen (NH3–N) x 1.3
Ammonia in water occurs in two forms: toxic unionized ammonia
(NH3) and the relatively non-toxic ionized form, ammonium ion
+
(NH
). This test method measures both forms as ammonia nitrogen
4
+
(NH
–N) to give the total ammonia nitrogen concentration in water.
4
The actual proportion of each compound depends on temperature,
salinity, and pH. A greater concentration of unionized ammonia is
present when the pH value and salinity increase.
4
3441
Consult the table to find the percentage that corresponds to the
temperature, pH, and salinity of the sample.
2. To express the test result as ppm Unionized Ammonia Nitrogen
(NH3–N), multiply the total ammonia-nitrogen test result by the
percentage from the table.
3. To express the test result as ppm Ionized Ammonia Nitrogen
+
(NH
–N), subtract the unionized ammonia-nitrogen determined in
4
Step 2 from the total ammonia-nitrogen.
10°C15°C20°C25°C
pHFW1SW
2
FWSWFWSWFWSW
7.00.190.270.400.55
7.10.230.340.500.70
7.20.290.430.630.88
7.30.370.540.791.10
7.40.470.680.991.38
7.50.590.4590.850.665 1.240.9631.731.39
7.60.740.5771.070.836 1.561.212.171.75
7.70.920.7261.351.051.961.522.722.19
7.81.160.9121.691.322.451.903.392.74
7.91.461.152.121.663.062.394.243.43
8.01.831.442.652.073.832.985.284.28
8.12.291.803.322.604.773.736.555.32
8.22.862.264.143.255.944.658.116.61
8.33.582.835.164.067.365.7810.008.18
8.44.463.546.415.059.097.1712.2710.10
8.55.554.417.986.2811.188.8714.9712.40
1
Freshwater data from Trussel (1972).
2
Seawater values from Bower and Bidwell (1978). Salinity for Seawater values = 34% at an
ionic strength of 0.701m.
EXAMPLE:
A fresh water sample sample at 20°C has a pH of 8.5 and the test
result is 1.0 ppm as Total Ammonia Nitrogen.
1. The percentage from the table is 11.18% (or 0.1118).
2. 1.0 ppm Total Ammonia Nitrogen x 0.1118 = 0.1118 ppm
Unionized Ammonia Nitrogen.