HYDROGEN PEROXIDE &
PERACETIC ACID KIT
DROP COUNT
CODE 7191-CN-02
QUANTITY CONTENTS CODE
30 mL *Sulfuric Acid, 1:1 *6141-CNWT-G
15 mL Ferroin Indicator 6410-CN-E
2 x 30 mL *Hydrogen Peroxide Titrant *5650-CNLWT-G
30 mL Potassium Iodide 20% Solution 6521-CN-G
60 mL Peracetic Acid Titrant S-6155-CN-H
1 Test Tube, 5-10-25 mL, plastic, w/cap 0715
1 Pipet, 0.5 mL, plastic 0353
*WARNING: Reagents marked with an * are considered to be potential health hazards.
Emergency information for all LaMotte reagents is available from Chem-Tel:
(US, 1-800-255-3924) (International, call collect, 813-248-0585)
To order individual reagents or test kit components, use the specifi ed code number.
NOTE: Do not use for tapwater or less than 20 ppm peracetic acid.
PROCEDURE
HYDROGEN PEROXIDE
1. Fill test tube (0715) to 10 mL line with sample water.
2. Add 10 drops of *Sulfuric Acid, 1:1 (6141-CNWT). Swirl to mix. Make sure
solution is thoroughly mixed.
3. Add 1 drop of Ferroin Indicator (6410-CN). Swirl to mix. Make sure solution is
thoroughly mixed. Solution will turn pale orange-pink if hydrogen peroxide is
present.
4. While gently swirling tube, add *Hydrogen Peroxide Titrant (5650-CNLWT) one
drop at a time until pale orange-pink color changes to colorless or pale yellow.
Make sure solution is thoroughly mixed. Count the number of drops added. Hold
bottle vertically.
5. Discard titrated sample and rinse tube out with sample water.
6. Multiply the number of drops used in Step 4 by 50 (fi fty). Record as ppm
Hydrogen Peroxide.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com

PERACETIC ACID
NOTE: This method may be used to test low to high concentration of peracetic acid.
Choose the procedure that is most appropriate for the sample being tested.
1 drop = 15 ppm
1. Fill test tube (0715) to 10 mL line with sample water.
2. Add 10 drops of *Sulfuric Acid, 1:1 (6141-CNWT). Swirl to mix. Make sure
solution is thoroughly mixed.
3. Add 1 drop of Ferroin Indicator (6410-CN). Swirl to mix. Make sure solution is
thoroughly mixed.
4. Add 1 drop of Potassium Iodide 20% Solution (6521-CN). Swirl to mix. Make
sure solution is thorougly mixed. Solution will turn a cloudy brown if peracetic
acid is present.
5. While gently swirling tube, add Peracetic Acid Titrant (S-6155-CN), one drop at
a time until brown color changes to a clear, pale orange-pink which persists for
20 seconds. Make sure solution is thoroughly mixed. Count the number of drops
added. Hold bottle vertically.
6. Multiply the number of drops used in Step 5 by 15 (fi fteen). Record as ppm
Peracetic Acid.
For example:
10 drops x 15 = 150 ppm Peracetic Acid
Low Range (20-90 ppm) 1 drop = 6ppm
Fill the test tube to the 25 mL line. Follow the test procedure and use 25 drops of
*Sulfuric Acid, 1:1 (6141-CNWT), 3 drops of Ferroin Indicator (6410-CN) and 3
drops of Potassium Iodide 20% Solution (6521-CN). In Step 6 multiply the number of
drops used by 6 (six). Record as ppm Peracetic Acid.
For example:
10 drops x 6 = 60 ppm Peracetic Acid
High Range (300-500 ppm) 1 drop = 300 ppm
For higher concentrations of peracetic acid use the 0.5 mL pipet (0353) to add 0.5 mL
of the sample to the test tube (0715). Fill the tube to the 10 mL line with hydrogen
peroxide/peracetic acid-free water. Follow Steps 2 - 5. In Step 6 multiply the number
of drops used by 300 (three hundred). Record as ppm Peracetic Acid.
For example:
10 drops X 300 = 3000 ppm Peracetic Acid
LaMOTTE COMPANY
Helping People Solve Analytical Challenges