ALKALINITY KIT, TOTAL
DIRECT READING TITRATOR, 0-200 ppm
CODE 4491-DR-01
QUANTITY CONTENTS CODE
50 BCG-MR Indicator Tablets T-2311-H
60 mL *Alkalinity Titration Reagent B *4493DR-H
1 Test Tube, 5-10-15 mL, w/cap 0778
1 Direct Reading Titrator, 0-200 Range 0382
1 Alkalinity Endpoint Color Chart 4491-CC
*WARNING: Reagents marked with an * are considered to be potential health hazards.
To view or print a Material Safety Data Sheet (MSDS) for these reagents go to
www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specifi ed code number.
This test set provides total alkalinity readings only.
Read LaMotte Direct Reading Titrator Manual before proceeding. The Titrator is
calibrated in terms of total alkalinity expressed as parts per million (ppm) Calcium
Carbonate (CaCO3). Each minor division on the Titrator scale equals 4 ppm CaCO3.
WARNING! This set contains chemicals
that may be harmful if misused. Read
cautions on individual containers
carefully. Not to be used by children
except under adult supervision
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com

PROCEDURE
1.
Fill the titration
tube (0778) to
the 5 mL line
with the
sample water.
4.
Fill Direct
Reading Titrator
(0382) with
*Alkalinity
Titration
Reagent B
(4493).
1.0
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
7.
Result:
Read the test
result directly from
the scale where the
large ring on the
Titrator meets the
Titrator barrel. Record
as ppm Total Alkalinity
in ppm Calcium
Carbonate (CaCO3).
140 ppm
2.
Add one
BCG-MR
Indicator
Tablet
(T-2311).
5.
Insert the Titrator
into the center
hole of the test
tube cap.
140
160
3.
Cap and swirl to mix until
tablet dissolves. Solution
will turn blue-green.
6.
While gently
swirling the tube,
slowly press the
0
0.1
0.2
0.3
plunger to titrate
0.4
0.5
0.6
until the solution
0.7
0.8
0.9
1.0
color changes from
blue-green to purple.
Consult Alkalinity
Endpoint Color
Chart (4491-CC).
NOTE :
If the plunger tip reaches the
bottom line on the scale
(200 ppm) before the endpoint
color change occurs, refi ll
the Titrator and continue the
titration.
When recording the test result,
be sure to include the value of
the original amount of reagent
dispensed (200 ppm).
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
NOTE: To convert the results to degrees of carbonate hardness dKH:
ppm CaCO
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
x 0.056 = dKH CaCO
3
64491-DR-01 2/12
3
®