LaMotte 4491-DR-01 User manual

Page 1
ALKALINITY KIT, TOTAL
DIRECT READING TITRATOR, 0-200 ppm
CODE 4491-DR-01
QUANTITY CONTENTS CODE
50 BCG-MR Indicator Tablets T-2311-H 60 mL *Alkalinity Titration Reagent B *4493DR-H 1 Test Tube, 5-10-15 mL, w/cap 0778 1 Direct Reading Titrator, 0-200 Range 0382 1 Alkalinity Endpoint Color Chart 4491-CC
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specifi ed code number. This test set provides total alkalinity readings only.
Read LaMotte Direct Reading Titrator Manual before proceeding. The Titrator is calibrated in terms of total alkalinity expressed as parts per million (ppm) Calcium Carbonate (CaCO3). Each minor division on the Titrator scale equals 4 ppm CaCO3.
WARNING! This set contains chemicals
that may be harmful if misused. Read
cautions on individual containers
carefully. Not to be used by children
except under adult supervision
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
PROCEDURE
1.
Fill the titration tube (0778) to the 5 mL line with the sample water.
4.
Fill Direct Reading Titrator (0382) with *Alkalinity Titration Reagent B (4493).
1.0
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
7.
Result:
Read the test result directly from the scale where the large ring on the Titrator meets the Titrator barrel. Record as ppm Total Alkalinity in ppm Calcium Carbonate (CaCO3).
140 ppm
2.
Add one BCG-MR Indicator Tablet (T-2311).
5.
Insert the Titrator into the center hole of the test tube cap.
140
160
3.
Cap and swirl to mix until tablet dissolves. Solution will turn blue-green.
6.
While gently swirling the tube, slowly press the
0
0.1
0.2
0.3
plunger to titrate
0.4
0.5
0.6
until the solution
0.7
0.8
0.9
1.0
color changes from blue-green to purple. Consult Alkalinity Endpoint Color Chart (4491-CC).
NOTE :
If the plunger tip reaches the bottom line on the scale (200 ppm) before the endpoint color change occurs, refi ll the Titrator and continue the titration.
When recording the test result, be sure to include the value of the original amount of reagent dispensed (200 ppm).
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
NOTE: To convert the results to degrees of carbonate hardness dKH:
ppm CaCO
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
x 0.056 = dKH CaCO
3
64491-DR-01 2/12
3
®
Loading...