LaMotte 3635-05 User manual

Page 1
Salt Water Aquaculture
Test Kit Instruction Manual
Code 3635-05
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
This booklet provides step-by-step detailed instructions for the Code 3635-05 test kit. It is important to review these instructions thoroughly tests by the short-form instructions contained in the case lid.
To order individual reagents or test kit components, use the specifi ed code number.
Safety / Testing Hints / Reagent Care .......................................... 4
Analytical Technique ..................................................................... 5
General Safety Information ........................................................... 6
Test Methods ................................................................................. 7
Test Procedures
Introduction ......................................................................................... 9
Alkalinity ............................................................................................. 10
Ammonia Nitrogen ............................................................................. 12
Carbon Dioxide .................................................................................. 15
Dissolved Oxygen .............................................................................. 16
Nitrate Nitrogen .................................................................................. 21
Nitrite Nitrogen ................................................................................... 22
pH ....................................................................................................... 23
Salinity ................................................................................................ 24
Kit Diagram ................................................................................... 26
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 3
SAFETY/ TESTING HINTS / REAGENT CARE
*WARNING: Reagents marked with an * are considered to be potential health hazards.
Tightly close all reagent immediately after use. Be sure not to interchange caps and pipets from different containers.
Wipe up any reagent chemical spills, liquid or powder, as soon as they occur. Refer to label and SDS for proper reagent disposal.
containers
Use care when dispensing or handling all reagents. Some reagents also may cause permanent stains if spilled.
Avoid prolonged exposure of equipment and reagents to direct sunlight. Protect reagents and components from extreme heat and cold.
4
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 4
ANALYTICAL TECHNIQUE
1. Clean glassware is a must for accurate results. Thoroughly rinse test tubes before and after each test. Caps and stoppers should also be cleaned after each use.
2. Use test tube caps or stoppers, not your fi ngers, to cover test tubes and fl asks during shaking or mixing.
3. When adding sample to calibrated test tube, be sure vial is fi lled to the appropriate mark. The bottom of the liquid (meniscus) should be level with the desired mark. (Fig. 1)
4. When dispensing reagents from bottles fi lled with dropper plug and cap, be sure to hold bottle vertically and gently squeeze to dispense the appropriate number of uniform drops. (Fig. 2)
5. For those reagents to be added with the screwcap pipet assemblies on bottle and replace with the screwcap pipet. NOTE: Place the polyseal caps back on the reagent bottles for longer periods of storage. Be sure that both pipet assemblies and polyseal caps are thoroughly cleaned before placing on bottles to avoid contamination.
6. When dispensing reagents from pipets, hold pipet vertically to assure uniform drop size. This is extremely important when performing drop count titrations. (Fig. 3)
7. To fi ll pipets, squeeze rubber bulb and immerse into reagent. Release bulb to fi ll. (Fig. 4)
8. To accurately dispense powdered reagents with spoon, tap spoon on edge of reagent container to remove excess reagent. (Fig. 5)
9. When performing tests that include Octa-Slide 2 Comparators, the comparator should be positioned between the operator and non-direct sunlight. This allows the light to enter through the
light-diffusing screen at the back of the
comparator for optimum
enclosed, remove polyseal cap
color comparison.
Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 5
GENERAL SAFETY PRECAUTIONS
1
12
Store the test kit in a cool dry area.
3
34
Read the labels on all reagent bottles. Note warnings and rst aid information. Reagents marked with a * are considered possible health hazards.
5
Avoid contact between reagent chemicals and skin, eyes, nose, and mouth. Wear safety glasses when
2
4
6
Read all instructions and note precautions before performing the test procedure. Read all Safety Data Sheets (SDS). SDS
Safety
Data
are available at
Sheet
65
performing test procedures.
Keep all equipment and reagent chemicals out of the reach of young children.
7.
6
Find Quality Products Online at: sales@GlobalTestSupply.com
In the event of an accident or suspected poisoning, immediately call the Poison Center phone number in the front of your local telephone directory or call a physician. Additional information for all LaMotte reagents is available in the United States, Canada, Puerto Rico, and the US Virgin Islands from Chem-Tel by calling 1-800-255-3924. For other areas, call 813-248-0585 collect to contact Chem-Tel’s International access number. Each reagent can be identifi ed by the four digit number listed on the upper left corner of the reagent label, in the contents list and in the test procedures.
www.GlobalTestSupply.com
Page 6
TEST METHODS
This test kit uses two basic analytical procedures common to fi eld test kits. A brief explanation of each follows:
COLORIMETRIC: OCTA-SLIDE 2 VIEWER
In a visual colorimetric test, a sample is treated with reagent(s) to produce a color reaction, generally in proportion to the amount of test factor present. The sample color is then compared against color standards representing known concentrations of the factor being tested over a specifi c range.
1
123
Insert Octa­Slide 2 into the Octa­Slide 2 (1101).
Bar Viewer
2
Insert test tube containing reacted sample into the Octa­Slide 2 (1101).
Viewer
3
Hold the Octa-Slide 2 Viewer so that non-direct light enters through the back of the viewer.
4
4
Match sample color to color standard. Record results.
Note:
NOTE: If sample color is between two standards, the midpoint is taken as the result.
If the sample is darker than the highest standard, a dilution may be performed on a fresh sample, and the test repeated to bring the concentration within range.
DILUTIONS
The calibrated test tubes (0106) included in this kit may be used to perform dilutions for the Ammonia Nitrogen and Nitrite Nitrogen tests. Distilled or deionized water is needed to perform dilutions. The following table provides a quick reference guide for dilutions of various proportions. Once the dilution is prepared, use this diluted sample to perform the test, and multiply the result by the dilution factor to obtain the actual concentration.
Sample Size Distilled Water
to Bring to 10 mL
5.0 mL 5.0 mL 2
2.5 mL 7.5 mL 4
Dilution Factor
7
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 7
TITRIMETRIC: DIRECT READING TITRATOR
In a titrimetric method, titrating solution (or titrant) is added to a treated sample until a color change occurs. The volume of titrant required to reach this endpoint is proportional to the concentration of the factor being tested. Direct Reading Titrators provide results directly in the appropriate concentration for the test—no counting of drops, no calculations.
The Titrator consists of a plastic barrel, a plastic plunger, and a plastic adapter tip. The adapter tip reduces the size of the drops that are dispensed, increasing the precision of the test results. DO NOT the plunger or adapter tip from the Titrator.
123
Fill the test tube to the appropriate line with sample water.
4
To fi ll Titrator, slowly withdraw the plunger until the bottom of the plunger is opposite the zero mark on the scale.
6
6
Insert the Titrator into the center hole of the test tube cap. While gently swirling tube, add titrating solution one drop at a time until the desired color changes occur.
Add reagents as specifi ed in the instructions for the individual test method. Cap with the special test tube cap. Mix by swirling gently.
NOTE: If small air bubbles appear in the barrel, expel them by partially fi lling the barrel and pumping the titration solution back into the reagent container. Repeat until bubble disappears.
7
7
Read the test result directly from the scale where the large ring on the Titrator meets the Titrator barrel.
Depress the Titrator plunger to expel air. Insert Titrator into the plastic fi tting of the titrating solution bottle and invert.
5
5
Turn the bottle right-side-up and remove the Titrator.
8
8
When testing is complete, discard titrating solution in Titrator. Rinse Titrator and titration tube thoroughly.
8
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 8
TEST PROCEDURES
INTRODUCTION
Proper control of water quality is an essential part of successful aquaculture operation. Immediate test results provided by on-site water analysis equipment can confi rm a healthy environment, or give early warning signals for required treatment.
1. Develop a routine testing schedule.
2. Keep records! Historical data is extremely important if treatments are required. Note environmental conditions, fi sh activity, feeding habits, etc.
3. Observe fi sh to note any particular behavior or feeding rates, as this may be a sign of stress.
4. Stable characteristics, such as alkalinity and hardness, do not have to be tested as frequently as ones that fl uctuate, such as ammonia nitrogen, nitrite nitrogen, pH, dissolved oxygen and temperature. Keep in mind that these factors fl uctuate throughout the day and in some cases are interdependent.
5. Be alert to sudden changes in one factor, as it may be a clue to perform further analysis.
9
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 9
Alkalinity
DESCRIPTION CODE
*BCG/MR Indicator *2311-PG-E *Alkalinity Titration Reagent B *4493DR-H Test Tube, 5-10-12.9-15-20-25 mL, glass, w/cap 0608 Direct Reading Titrator, 0-200 Range 0382
*WARNING: Reagents marked with an * are considered to be potential health hazards. For futher information see page 4.
The Direct Reading Titrator is calibrated in terms of total alkalinity expressed as parts per million (ppm) Calcium Carbonate (CaCO3). Each minor division on the Titrator scale equals 4 ppm CaCO3.
10
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 10
ALKALINITY TEST PROCEDURE
123
12 3
Fill the test tube (0608) to the 5 mL line with the sample water.
4
4
Fill Direct Reading Titrator (0382) with *Alkalinity Titration Reagent B (4493DR).
7
7
Read the test result directly from the scale where the large ring on the Titrator meets the Titrator barrel. Record Total Alkalinity as ppm Calcium Carbonate (CaCO3).
Add 4 drops of *BCG-MR Indicator (2311-PG).
5
5
Insert the Titrator into the center hole of the test tube cap.
8
NOTE: If the plunger tip reaches the bottom line on the scale (200 ppm) before the endpoint color change occurs, refi ll the Titrator and continue the titration.
When recording the test result, be sure to include the value of the original amount of reagent dispensed (200 ppm).
Cap and mix. Solution will turn blue-green.
6
6
While gently swirling the tube, slowly press the plunger to titrate until blue-green color changes to pink.
11
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 11
Ammonia Nitrogen
DESCRIPTION CODE
*Salicylate Ammonia #1 *3978L WT -G *Salicylate Ammonia #2 *3979WT -G Salicylate Ammonia #3 3982WT-G Test Tube, 2.5-5-10 mL, plastic, w/cap 0106 Octa-Slide 2 Viewer 1101 Ammonia Nitrogen Octa-Slide 2 Bar, 0-2.0 ppm, Salt Water 3441-01-SW
*WARNING: Reagents marked with an * are considered to be potential health hazards. For futher information see page 4.
12
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 12
AMMONIA NITROGEN TEST PROCEDURE
12 3
4
Cap and mix.
7
Insert Ammonia Nitrogen Octa­ Slide 2 Bar (3441-01-SW) into the Octa-Slide 2
Viewer (1101).
Wait 1 minute.
1
MINUTE
5
8
Fill a test tube (0106) to the 5 mL line with the water sample.
Add 7 drops of *Salicylate Ammonia #2 (3979WT).
Add 7 drops of Salicylate Ammonia #3 (3982WT).
Add 10 drops of *Salicylate Ammonia #1 (3978WT).
6
Cap and mix.
9
Cap and mix.
10
20
MINUTES
Wait 20
11
minutes.
Insert test tube into Octa-Slide 2 Viewer.
Calculations:
To express results as Ammonia (NH3):
Ammonia (NH
To express results as Ammonium (NH
Ammonium (NH
) = ppm Ammonia Nitrogen (NH3-N) x 1.2
3
+
) = ppm Ammonia Nitrogen (NH3-N) x 1.3
4
12
Match sample color to a color standard. Record as ppm Ammonia Nitrogen.
+
):
4
13
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 13
Ammonia in water occurs in two forms: toxic unionized ammonia (NH3) and the relatively non-toxic form, ammonium ion (NH
+
). This test method measures both
4
forms as ammonia-nitrogen (NH3-N) to give the total ammonia-nitrogen concentration in water. The actual proportion of each compound depends on temperature, salinity, and pH. A greater concentration of unionized ammonia is present when the pH value and salinity increase.
1. Consult the table below to fi nd the percentage that corresponds to the temperature, pH and salinity of the sample.
2. To express the test result as ppm Unionized Ammonia Nitrogen (NH
-N), multiply
3
the total ammonia-nitrogen test result by the percentage from the table.
3. To express the test result as ppm Ionized Ammonia Nitrogen (NH
+
-N), subtract the
4
unionized ammonia nitrogen from the total ammonia-nitrogen.
10°C 15°C 20°C 25°C
pH FW1 SW2 FW SW FW SW FW SW
7.0 0.19 0.27 0.40 0.55
7.1 0.23 0.34 0.50 0.70
7.2 0.29 0.43 0.63 0.88
7.3 0.37 0.54 0.79 1.10
7.4 0.47 0.68 0.99 1.38
7.5 0.59 0.459 0.85 0.665 1.24 0.963 1.73 1.39
7.6 0.74 0.577 1.07 0.836 1.56 1.21 2.17 1.75
7.7 0.92 0.726 1.35 1.05 1.96 1.52 2.72 2.19
7.8 1.16 0.912 1.69 1.32 2.45 1.90 3.39 2.74
7.9 1.46 1.15 2.12 1.66 3.06 2.39 4.24 3.43
8.0 1.83 1.44 2.65 2.07 3.83 2.98 5.28 4.28
8.1 2.29 1.80 3.32 2.60 4.77 3.73 6.55 5.32
8.2 2.86 2.26 4.14 3.25 5.94 4.65 8.11 6.61
8.3 3.58 2.83 5.16 4.06 7.36 5.78 10.00 8.18
8.4 4.46 3.54 6.41 5.05 9.09 7.17 12.27 10.10
8.5 5.55 4.41 7.98 6.28 11.18 8.87 14.97 12.40
1
Freshwater data from Trussel (1972).
2
Seawater values from Bower and Bidwell (1978). Salinity for Seawater values = 34 ppt at an ionic strength
of 0.701 m.
FOR EXAMPLE: A fresh water sample at 20°C has a pH of 8.5 and the test result is 1.0 ppm as Total Ammonia-Nitrogen.
1. The percentage from the table is 11.18% (or 0.1118).
2. 1 ppm total Ammonia-Nitrogen x 0.1118 = 0.1118 ppm Unionized Ammonia­Nitrogen
3. Total Ammonia-Nitrogen 1.0000 ppm
Unionized Ammonia-Nitrogen – 0.1118 ppm Ionized Ammonia-Nitrogen = 0.8882 ppm
14
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 14
Carbon Dioxide
DESCRIPTION CODE
*Phenolphthalein Indicator, 1% *2246-E *Carbon Dioxide Reagent B *4253DR-H
Direct Reading Titrator, 0-50 Range 0380 Test Tube, 5-10-12.9-15-20-25 mL, glass, w/cap 0608
*WARNING: Reagents marked with an * are considered to be potential health hazards. For futher information see page 4.
The Direct Reading Titrator is calibrated in terms of carbon dioxide expressed as parts per million (ppm) free CO2. Each minor division on the Titrator scale equals 1.0 ppm CO2.
12
1
NOTE: results, test a freshly obtained sample, and avoid splashing or prolonged contact with air.
3
3
Fill Direct Reading Titrator (0380) with Carbon Dioxide Reagent B (4253DR).
For best
Note:
Fill the test tube (0608) to the 20 mL line with the sample water.
4
4
Insert the Titrator into the center hole of the test tube cap.
water
sample
2
Add 2 drops of Phenolphthalein Indicator, 1% (2246). If sample turns red, no free carbon dioxide is present. If sample is colorless, proceed to Step 3.
5
5
While gently swirling the tube, slowly press the plunger to titrate until a faint pink color develops and persists for 30 seconds.
6
6
Read the test result directly from the where the large ring Titrator meets the Titrator barrel. Record as Carbon Dioxide.
scale
on the
Find Quality Products Online at: sales@GlobalTestSupply.com
Result:
40 ppm
www.GlobalTestSupply.com
NOTE: The Titrator is calibrated in terms of carbon dioxide expressed as minor division on the Titrator scale equals 1.0 ppm CO2.
ppm Free CO2. Each
15
Page 15
Dissolved Oxygen
DESCRIPTION CODE
*Manganous Sulfate Solution *4167-G *Alkaline Potassium Iodide Azide Reagent *7166-G *Sulfuric Acid, 1:1 *6141WT -G
Sodium Thiosulfate, 0.025N 4169-H Starch Indicator Solution 4170WT-G Direct Reading Titrator, 0-10 Range 0377 Test Tube, 5-10-12.9-15-20-25 mL, glass, w/cap 0608 Pipet, plain, plastic, w/cap 0392 Water Sampling Bottle, 60 mL, glass 0688-DO
*WARNING: Reagents marked with an * are considered to be potential health hazards. For futher information see page 4.
The Titrator is calibrated in terms of Dissolved Oxygen expressed as ppm Dissolved Oxygen. Each minor division on the Titrator scale equals 0.2 ppm Dissolved Oxygen.
16
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 16
DISSOLVED OXYGEN TEST PROCEDURE
Part 1 - Collecting the Water Sample
1
12
Rinse the Water Sampling Bottle (0688-DO) with the sample water.
3
3
Remove the cap and allow the bottle to fi ll.
56
5
2
Tightly cap the bottle, and submerge it to the desired depth.
4
4
Tap the sides of the bottle to dislodge any air bubbles.
6
Replace the cap while the bottle is still submerged.
Retrieve the bottle and make sure that no air bubbles are trapped inside.
17
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 17
Part 2 - Adding the Reagents
NOTE: Be careful not to introduce air into the sample while adding the reagents.
12
1
Remove the cap from the bottle.
3
3
5
Cap the bottle and mix by inverting several times. A precipitate will form.
56
Add 8 drops of *Sulfuric Acid, 1:1 (6141WT).
2
Immediately add 8 drops of *Manganous Sulfate Solution (4167) and add 8 drops of *Alkaline Potassium Iodide Azide (7166).
4
4
Allow the precipitate to settle below the shoulder of the bottle.
6
Cap and gently invert the bottle to mix the contents until the precipitate has totally dissolved. The solution will be clear yellow to orange if the sample contains dissolved oxygen.
NOTE: At this point the sample has been “fi xed” and contact between the sample and the atmosphere will not affect the test result. Samples may be held at this point and titrated later.
18
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 18
Part 3 - The Titration
12
12
0
Fill the titration tube (0608) to the 20 mL line with the fi xed sample. Cap the tube.
0
0.1
0.1
0.2
0.2
0.3
0.3
0.4
0.4
0.5
0.5
0.6
0.6
0.7
0.7
0.8
0.8
0.9
0.9
1.0
1.0
Depress plunger of the Titrator (0377).
3
3
Insert the Titrator into the plug in the top of the *Sodium Thiosulfate,
0.025N (4169) titrating solution.
0
.
1
0
0
.
2
.
3
0
0
.
4
5
0
.
0
.
6
.
7
0
8
0
.
.
9
0
0
1
.
4
4
Invert the bottle and slowly withdraw the plunger until the large ring on the plunger is opposite the zero (0) line on the scale.
NOTE: If small air bubbles appear in the titrator barrel, expel them by partially fi lling the barrel and pumping the titration solution back into the reagent container. Repeat until bubble disappears.
5
5
Turn the bottle upright and remove the Titrator.
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
NOTE: If the sample is a very pale yellow, go to Step 9.
1.0
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
continued . . .
19
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 19
6
67
Insert the tip of the Titrator into the opening of the titration tube cap.
7
Slowly depress the plunger to dispense the titrating solution until the yellow-brown color changes to a very pale yellow. Gently swirl the tube during the titration to mix the contents.
8
89
Carefully remove the Titrator and cap. Do not disturb the Titrator plunger.
10
Cap the titration tube. Insert the tip of the Titrator into the opening of the titration tube cap.
12
12
Read the test result directly from the scale where the large ring on the Titrator meets the Titrator barrel. Record as ppm Dissolved Oxygen. Each minor division on the Titrator scale equals 0.2 ppm.
9
Add 8 drops of Starch Indicator Solution (4170WT). The sample should turn blue.
11
Continue titrating until the blue
1110
color disappears and the solution becomes colorless.
NOTE: If the plunger ring reaches the bottom line on the scale (10 ppm) before the
endpoint color change occurs,
refi ll the Titrator and continue the titration. Include the value of the original reagent dispensed
when recording the test result.
amount of
(10 ppm)
NOTE: When testing is complete, discard the titrating solution in the Titrator. Rinse Titrator and titration tube thoroughly. DO NOT remove plunger or adapter tip.
20
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 20
Nitrate Nitrogen
DESCRIPTION CODE
*Mixed Acid Reagent *V -6278-J *Nitrate Reducing Reagent *V-6279-C Spoon, 0.1 g, plastic 0699 Test Tube, 2.5-5-10 mL, plastic, w/cap 0106 Dispenser Cap 0692 Octa-Slide 2 Viewer 1101 Nitrate Nitrogen Octa-Slide 2 Bar, 0.25-10.0 ppm 3109 -01
*WARNING: Reagents marked with an * are considered to be potential health hazards. For futher information see page 4.
NOTE: Place Dispenser Cap (0692) on *Mixed Acid Reagent (V-6278). Save this cap for refi ll reagents.
123
4
7
Insert Nitrate­Nitrogen Octa­Slide 2 Bar (3109-01) into Octa-Slide 2 Viewer (1101).
Wait 2 minutes.
2
MINUTES
Wait 10 minutes.
10
MINUTES
5
8
Fill a test tube (0106) to the
2.5 mL line with the water sample.
Use the 0.1g spoon (0699) to add 0.1g of *Nitrate Reducing Reagent (V-6279).
Cap and invert one time before inserting test tube into Octa-Slide 2 Viewer (1101).
6
9
Attach dispenser cap (0692) to the *Mixed Acid Reagent (V-6278). Fill tube to the 5mL
line.
Cap and mix.
Cap and invert 50-60 times for one minute.
Insert test tube into Octa-Slide 2 Viewer (1101).
10
Find Quality Products Online at: sales@GlobalTestSupply.com
Match sample color ppm Nitrate Nitrogen (NO
Conversions: ppm Nitrate-N (NO
www.GlobalTestSupply.com
to color standard. Record results as
-N).
3
-N)
x 4.4 = ppm Nitrate (NO3)
3
21
Page 21
Nitrite Nitrogen
DESCRIPTION CODE
*Mixed Acid Reagent *V -6278-J *Color Developing Reagent *V-6281-D Spoon, 0.1 g, plastic 0699 Test Tube, plastic, w/cap 0106 Dispenser Cap 0692 Octa-Slide 2 Viewer 1101 Nitrite Nitrogen Octa-Slide 2 Bar, 0.05-0.8 ppm 3437-01
*WARNING: Reagents marked with an * are considered to be potential health hazards. For futher information see page 4.
NOTE: Place Dispenser Cap (0692) on *Mixed Acid Reagent (V-6278-J). Save this cap for refi ll reagents.
1
4
Insert Nitrite­Nitrogen Octa-Slide 2 Bar (3437-01) into
the Octa-Slide 2 Viewer (1101).
Use the 0.1 g spoon (0699) to add 0.1 g of
*Color Developing Reagent (V-6281).
23
Fill a test tube (0106) to the
5
2.5 mL line with the water sample.
Cap and mix by inverting for one minute.
6
Dilute to the *Mixed Acid
Reagent (V-6278).
5mL line with
Wait 5 minutes.
5
MINUTES
78
22
Find Quality Products Online at: sales@GlobalTestSupply.com
Insert test tube into Octa-Slide 2 Viewer (1101).
www.GlobalTestSupply.com
Match sample color to color standard. Record results as ppm Nitrite Nitrogen.
9
To convert to multiply results by 3.3. Record as ppm
Nitrite-N (NO3-N) ppm Nitrite (NO
Nitrite,
Nitrite.
x 3.3 =
)
2
Page 22
pH
DESCRIPTION CODE
*Wide Range Indicator *2218-G Test Tube, plastic, w/cap 0106 Octa-Slide 2 Viewer 1101 Wide Range pH Octa-Slide 2 Bar, 5.0-10.0 3483-01
*WARNING: Reagents marked with an * are considered to be potential health hazards. For futher information see page 4.
123
4
Insert Wide
Range pH
Octa-Slide 2
Bar (3483-01)
into the Octa-
Slide 2 Viewer
(1101).
Cap and mix.
5
Fill a test tube (0106) to the 10 mL line with the water sample.
Insert test tube into Octa-Slide 2 Viewer (1101).
6
Add 8 drops of *Wide Range pH Indicator (2218).
Match sample color to color standard. Record as pH.
23
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 23
Salinity
DESCRIPTION CODE
*Salinity Indicator Reagent A *7460-E *Salinity Titration Reagent B *7461-H
Demineralizer Bottle 1151 Test Tube, 5-10-12.9-15-20-25 mL, glass, w/cap 0608 Direct Reading Titrator, 0-20 Range 0378 Direct Reading Titrator, 0-1.0 Range 0376
*WARNING: Reagents marked with an * are considered to be potential health hazards. For futher information see page 4.
Deminerlizer Bottle (Code 1151)
The Demineralizer Bottle will be the source of demineralized water for the salinity test. It is partially fi lled with an ion exchange resin containing an indicator to show when the exchange resin has been exhausted. The indicator will change from dark green to yellow amber at which point the resin should be replaced.
Instructions For Use
1. Remove cap and fi ll bottle with water to be demineralized.
2. Recap, make sure spout is closed, and shake vigorously for 30 seconds.
3. Open spout, invert bottle, and gently squeeze to dipense the demineralized water.
Care and Storage
1. Do not heat water over 100°F.
2. Keep resin covered with water at all times.
3. Store bottle fi lled with enough water to cover resin completely.
The Titrator is calibrated in terms of Salinity expressed as ppt Salinity. Each minor division on the Titrator scale equals 0.4 ppt Salinity.
24
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 24
SALINITY TEST PROCEDURE
12
Fill the test tube (0608) to the 10 mL line with Demineralized Water (1151).
4
Add 3 drops of *Salinity Indicator Reagent A (7460) to test tube.
7
Insert the Titrator into the center hole of the test tube cap.
Fill the 1.0 mL Direct Reading Titrator (0376) to the 0 mark with sample water. Wipe off excess water.
5
Cap and gently swirl to mix. Sample will turn yellow.
89
While gently swirling the tube, press the plunger to add Salinity Titration Reagent B until yellow color changes to pink-brown.
3
Dispense 0.5 mL of sample water into test tube by depressing plunger until tip is at 0.5 mark. Discard remaining sample in Titrator.
6
Fill Direct Reading Titrator (0378) with *Salinity Titration Reagent B (7461).
Read the test result directly from the scale where the large ring on the Titrator meets the Titrator barrel. Record as ppt salinity.
Note: Each Titration Division = 0.4 ppt Salinity
25
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 25
3635-05
0377
0378
0380
4170
PS-G
6141
WT-G
4169-H
7461-H
4253DR-H
3437-01
V-6279-C
7460-E
2246-E
1066
7166-G
1151
V-6281-C
0382
4167-G
0376
4493DR-H
V-6278-J
0608
-DO
0688
0608
2311-PG-E
V-6278-J
3441-01-SW
1101
3109-01
3483-01
0106
0106 01060106
01060106
3982
WT-G
2218-G
3979
WT-G
0688
3978
LWT-H
26
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Loading...