LaMotte 3590-03 User Manual

Page 1
Combination Outfi t
For General Water Analysis
Test Kit Instruction Manual
Code 4783-03 Code 3590-03
Page 2
USE OF THE OCTA-SLIDE 2 VIEWER
The Octa-Slide 2 Viewer should be held so non-direct light enters through the back of the Viewer. Insert the reacted sample into the top of the Viewer. Slide the Octa-Slide 2 Bar into the Viewer and match the color of the reaction to the color standards.
1
2
3
Insert the Octa­Slide 2 Bar into the Octa-Slide 2 Viewer (1101).
Insert test tube into
the Octa-Slide
2
Viewer (1 101).
Match sample color to a color standard. Record results.
Page 3
COMBINATION WATER OUTFIT
CODE 4783-03 CODE 3590-03
QUANTITY CONTENTS CODE
15 mL *Hardness Reagent #5 *4483-E 100 Hardness Reagent #6 Tablets 4484-J 60 mL Hardness Reagent #7 4487WT-H 30 mL *Iron Reagent #1 *4450-G
4.5 g *Iron Reagent #2 Powder *4451-S 30 mL *Wide Range Indicator *2218-G 30 mL *Sulfi de Reagent A *4458-G† 15 mL *Sulfi de Reagent B *4459-E† 60 mL Sulfi de Reagent C 4460-H† 1 Test Tube, w/cap 4488 1 Spoon, 0.05 g, plastic 0696 4 (6†) Test Tubes, plastic, 2.5-5-10 mL, w/caps 0106 1 Pipet, 1.0 mL, plastic 0354 1 Iron Octa-Slide 2 Bar, 0.5-10.0 ppm 4448-01 1 Wide Range pH Octa-Slide 2 Bar, 5.0-10.0 ppm 3483-01 1 Sulfi de Octa-Slide 2 Bar, 0.2-20 ppm 4457-01 1 Octa Slide 2 Viewer 1101
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents go to www. lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.To order individual reagents or test kit components, use the specifi ed code number.
Code 3590-03 Only
Page 4
TOTAL HARDNESS TEST PROCEDURE
1
Fill the test tube (4488) to the desired line with the sample water, upper line: 1 drop = 10 ppm CaCO middle line: 1 drop = 1 gpg CaCO3 lower line: 1 drop = 20 ppm CaCO
3
.
3
2
4
6
Add 5 drops of *Hardness Reagent #5 (4483).
Add 1 Hardness Reagent #6 Tablet (4484).
While gently swirling the tube, add Hardness Reagent #7 (4487WT) one drop at a time until the red color changes to blue. Count the number of drops added. Hold bottle vertically.
3
5
Swirl to mix.
Swirl until tablet has disintegrated. Solution will turn Red if hardness is present. Proceed to next step. If solution is Blue, there is no measurable amount of hardness.
7
Multiply the number of drops used in Step 6 as follows: Tube fi lled to: upper line: each drop equals 10 ppm Hardness as CaCO middle line: each drop equals 1 gpg Hardness as CaCO lower line: each drop equals 20 ppm Hardness as CaCO
3
3
3
Page 5
IRON TEST PROCEDURE
1
3
5
Cap and shake until powder dissolves.
Rinse test tube (0106) with sample water. Fill to 5 mL line.
Cap and mix.
2
4
6
Add 5 drops of *Iron Reagent #1 (4450).
Use the 0.05g spoon (0690) to add one level measure of *Iron Reagent #2 Powder (4451).
Wait 3 minutes.
3
MINUTES
7
9
Insert the Octa-Slide 2 Bar (4448-01) into
the Octa-
Slide 2
Viewer
(1101).
Match sample color to a color standard. Record results as ppm iron.
8
Insert test tube into
the Octa-
Slide 2
Viewer
(1101).
Page 6
pH TEST PROCEDURE
1
3
5
Fill a test tube (0106) to the 10 mL line with sample water.
Cap and mix.
Insert test tube into
the Octa-
Slide 2
Viewer
(1101).
2
4
6
Add 8 drops of *Wide Range pH Indicator (2218).
Insert Wide Range pH Octa-Slide 2 Bar (3483-01) into the Octa-Slide 2 Viewer (1101).
Match sample color to a color standard. Record as pH.
Page 7
TOTAL SULFIDE TEST PROCEDURE
Supplied only with Code 3590-03 Note: Collect sample with a minimum of aeration and analyze promptly.
1
3
5
Fill a test tube (0106) to the 5 mL line with the sample water.
Cap and gently invert to mix.
Wait 1 minute.
1
MINUTE
2
Add 15 drops of *Sulfi de Reagent A (4458).
4
Add 3 drops of *Sulfi de Reagent B (4459).
Cap and mix.
6
Use the pipet (0354) to add
1.0 mL of Sulfi de Reagent C (4460).
Cap and mix
7
9
Insert the Sulfi de Octa-Slide 2 Bar (4457-01) into the Octa-Slide 2 Viewer (1101).
Match sample color to a color standard. Record results as ppm Sulfi de.
8
EPA accepted procedure
A known concentration of sulfi de in the range of 0.2-20.0 ppm S= may be prepared as a check standard by referring to APHA Standard Methods 4500-S2-D, 17th Ed., 1989. Run this test on the check standard. If the result is incorrect, discard the old reagents and order the reagent refi ll package.
Insert test tube into the Octa-Slide 2 Viewer (1101).
Page 8
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside USA) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
63590-03 4/12