Chlorine dioxide is used as a substiture for and an adjunct to chlorine in
water treatment. It is better than chlorine in eliminating taste and odor in
certain cases. Chlorine dioxide, unlike chlorine, does not produce carcinogenic
chlorinated organic compounds when reacted with organic materials. A
disadvantage is the higher cost of producing chlorine dioxide compared to
chlorine.
APPLICATION
Drinking and pool waters; domestic and industrial wastewater.
RANGE
0 to 7.0 ppm Chlorine Dioxide
METHOD
Chlorine dioxide reacts with DPD to form a red color in proportion to the
concentration.
HANDLING & PRESERVATION
Test as soon as possible to avoid loss of chlorine dioxide.
INTERFERENCES
Chlorine interference is eliminated by the addition of glycine to the sample
before the indicator.
6. Insert tube into chamber, close
lid and select Scan Blank.
Analysis
7. Remove tube from colorimeter.
Add 5 drops of Glycine Solution
(6811). Cap and mix.
8. Add one *Chlorine DPD #1
Instrument Grade Tablet
(6903A).
9. Cap and shake for 10 seconds.
Invert slowly 5 times. Make
reading within 30 seconds.
10. Insert tube into chamber. Close
lid. Select Scan Sample. Record
result as Chlorine Dioxide.
NOTE: The meter will remember the last scanned blank reading. It is not
necessary to scan a blank each time the test is performed. To use the
previous blank reading, instead of scanning a new one, scroll to Scan Sample
and proceed. For the most accurate results, the meter should be blanked
before each test and the same tube should be used for the blank and the
reacted sample.
NOTE: For the best possible results, carry a reagent blank through the
procedure. After scanning the blank, perform the test procedure on clear,
colorless,
subsequent test results.
CAUTION: DO NOT leave reacted DPD samples in test tubes (0290). Discard
sample and thoroughly rinse tubes. If allowed to remain, DPD will stain tubes,
signifi cantly impairing the operation of the colorimeter. If necessary, acid
wash, and vigorously clean glassware with test tube brush and detergent.
distilled or deioized water. Subtract results of regent blank from all
CALIBRATION
STANDARDS
The meter should be calibrated with free chlorine standards. The calibration
should be done with a distilled or deionized water blank and one chlorine
standard of known concentration. The concentration of the calibration
standard should be similar to the expected concentration of the sample that
will be tested.
Prepare a chlorine standard with a concentration that is equivenlent to a
chlorine dioxide standard that is within the range of the chlorine dioxide test.
(1 ppm chlorine = 1.9 ppm chlorine dioxide).
CALIBRATION PROCEDURE
Calibration
1. Press and briefl y hold to
turn the meter on. The LaMotte
logo screen will appear for about
3 seconds and the Main Menu
will appear.
3. Scroll to and select 022 Chlorine
Dioxide from menu.
Calibration
4. Rinse a clean tube (0290) three
times with the free chlorine
standard. Fill to the 10 mL line
with the free chlorine standard.
Cap the tube. Dry the tube with a
lint-free cloth.
5. Insert tube into chamber, close
lid and select Scan Blank.
6. Remove tube from colorimeter.
Add one *Chlorine DPD #1
Instrument Grade Tablet
(6903A).
022 Chlorine Dioxide
Scan Blank
Scan Sample
12:00:00 001/500
7. Cap and shake for 10 seconds.
Invert slowly 5 times. Make
reading within 30 seconds.
The default setting for the data logger is enabled. The meter will log the last
500 data points. The counter in the center bottom of the display will show how
many data points have been logged. The display will show 500+ when the data
logger has exceeded 500 points and the data points are being overwritten.
1. From the Editing Menu, press
or to scroll to
Logging.
2. Press
ENTER
to select Logging.
3. Press or to scroll to
desired function.
Editing Menu
Set Clock
Logging
Factory Setup
Set Power Save
12:00:00 001/500
Logging
Display Test Log
Logging Enabled
ogging Disabled
L
Erase Log
12:00:00 001/500
Logging
Display Test Log
Logging Enabled
Logging Disabled
rase Log
E
12:00:00 001/500
4. Press
ENTER
. The screen will
display Storing... for about 1
second and return to the EditingMenu.
The Factory Setup menu is used in manufacturing of the colorimeter. This
menu is not for use by the operator in the fi eld.
SETTING POWER SAVE
The power saving Auto Shutoff feature will turn the meter off when a button
has not been pushed for a set amount of time. The default setting is disabled.
To change the setting:
1. From the Editing Menu, press
or to scroll to Set
Power Save.
2. Press
ENTER
to select Set Power
Save.
3. Press or to scroll to
desired function.
Editing Menu
Set Clock
Logging
Factory Setup
Set Power Save
12:00:00 001/500
Set Power Save
Disable
5 Minutes
15 Minutes
30 Minut
12:00:00 001/500
es
Set Power Save
Disable
5 Minut
es
15 Minutes
30 Minutes
12:00:00 001/500
4. Press
ENTER
. The screen will
display Storing.... for about 1
second and the meter will retun
to the Editing Menu.
The backlight illuminates the display for enhanced viewing. The default setting
is 10 seconds. If Button Control is chosen the backlight button on the key pad
will act as an on/off switch and the backlight will remain on or off when the
meter is being used. When one of the other settings – 10, 20 or 30 seconds –
is chosen, the display will be illuminated for the specifi ed amount of time after
any button is pressed.
NOTE: The backlight feature uses a signifi cant amount of power. The longer
the backlight is on, the more frequently the battery will have to be charged if
the USB/Wall Adapter is not being used.
1. From the Editing Menu, press
or to scroll to
Backlight Time.
2. Press
ENTER
to select Set
Backlight Time.
3. Press or to scroll to
desired option.
Editing Menu
Logging
Factory Setup
Set Power Save
Set Backlight Time
12:00:00 001/500
Set Backlight Time
Button Control
10 seconds
20 seconds
30 seconds
12:00:00 001/500
Set Backlight Time
Button Control
10 seconds
20 seconds
30 seconds
12:00:00 001/500
4. Press
ENTER
. The screen will
display Storing... for about 1
second and the meter will return
to the Editing Menu.
There are seven languages available: English, Spanish, French,
Portuguese, Italian, Chinese, and Japanese.
1. From the Editing Menu, press
or to scroll to
Select Language.
2. Press
ENTER
to select Select
Language.
3. Press or to scroll to
desired language.
Editing Menu
Factory Setup
Set Power Save
Set Backlight Time
Select Language
12:00:00 001/500
Select Language
English
Spanish
rench
F
Portugese
12:00:00 001/500
Select Language
English
Spanish
rench
F
Portugese
12:00:00 001/500
4. Press
ENTER
. The screen will
display Storing... for about 1
second and the meter will return
to the Editing Menu.
Set Power Save
Set Backlight Time
Editing Menu
Bluetooth Menu
Select Language
12:00:00 001/500
NOTE: If meter unintentionally switches to another language, use the
procedure above to reset the meter to the desired language. For example, to
reset the meter to English:
Long menus, such as the Tsting Menu, incorporate a looping feature which
allows the user to quickly reach the last choice in the menu from the fi rst
choice. In a looping menu the last choices in the menu are above the fi rst
choice and scrolling upward moves through the menu in reverse order.
Scrolling downward moves through the menu from fi rst choice to last but
the menu starts over following the last choice. So all menu choices can be
reached by scrolling in either direction. The diagrams below demonstrate a
looping menu.
The meter may be interfaced with any Windows-based 64 bit computer
by using the LaMotte WaterLink Connect program and a USB cable. The
program will store test information and results in a local database, and allow
for exporting this data to a comma separated value (CSV) fi le. The meter
will send the following data: time/date stamp, name of test, sample value,
sample units, meter name, and location. To download WaterLink go to
BATTERY
BATTERY/AC OPERATION
The colorimeter may be operated on battery power using a USB wall adapter
or USB computer connection. If using the meter as a bench top unit, use the
wall adapter if possible to extend the battery life. The meter will remain on
when the USB adapter is used.
To charge the battery with the wall adapter, plug the smaller end of the USB
cable (USB mini B connector) into the meter and the larger end of the USB
cable (USB type A connector) into the wall adapter. Plug the wall adapter into
an AC outlet. Reinsert the rubber USB port plug after charging. To charge the
battery from a computer, plug the smaller end of the USB cable (USB mini B
connector) into the meter and the larger end of the USB cable (USB Type A
connector) into a USB port on a computer. The battery icon will show no bars
and fl ash when the unit fi rst turns on. Then the indicator will indicate the
battery status by showing 0, 1, 2, 3, or 4 bars.
It will take 6 hours to fully charge a low battery. The battery icon will fl ash
when the battery is charging. The battery icon will show four bars and stop
fl ashing when it is fully charged. The charging circuit will automatically switch
to a fl oat charge when the battery is fully charged. The charger may remain
connected. Some computers will NOT supply power to their USB ports during
standby operation. The wall adapter will charge the unit continuously.
Storing the meter above ambient room temperature will decrease the battery
charge more quickly than storage at room temperature. If the meter does
not turn on, it means that the battery is at a very low charge. Charging the
battery with the wall adapter in this state may take up to 10 hours. At low
temperatures, approaching 0 °C, the battery will charge more slowly. It will not
charge at all below 0 °C. The battery icon will show no bars and continuously
fl ash if the battery is getting low but the unit will still operate normally. A
“Low Battery” message on the status bar of the display will replace the time
when the battery voltage is too low for proper operation and accuracy may
be degraded. A “Shutdown Low Batt” message on the display will appear for
a few seconds before the power is switched off when the battery is too low to
operate the unit. When the battery icon simultaneously fl ashes bars 1 and 2
followed by bars 3 and 4, it is an indication that the battery is damaged and
technical support should be contacted.
To extend the battery life:
• Shut down the unit with the power switch when not taking
measurements or use the power save option to have the unit
automatically turn off after 5 minutes.
• Store the unit at a moderate temperature.
• Fully charge the battery before storing the unit for extended periods
of time.
• Fully charge the battery at least once per year. Failure to do so may
result in a permanently drained battery.
• Limit backlight use. The unit consumes 3X normal power with the
backlight on. Set the backlight time option to 10 seconds, or select
“Button Control” and keep the backlight off .
BATTERY REPLACEMENT
The lithium ion battery used in this unit should last for many years with
normal use. When it no longer powers the unit long enough to meet testing
requirements it will need to be replaced. Lithium ion batteries that are
properly charged and stored do not usually lose all capacity; they just have
less capacity after hundreds of charge cycles. This unit uses a custom battery
assembly that is only available from LaMotte Company. Battery replacement
must be performed at a LaMotte authorized repair facility. The water resistant
housing of this meter should not be opened by the user
MAINTENANCE
CLEANING
Clean the exterior housing with a damp, lint-free cloth. Do not allow water
to enter the light chamber or any other parts of the meter. To clean the
light chamber and optics area, point a can of compressed air into the light
chamber and blow the pressurized air into the light chamber. Use a cotton
swab dampened with Windex® window cleaner to gently swab the interior of
the chamber. Do not use alcohol; it will leave a thin residue over the optics
when dry.
Waste Electrical and Electronic Equipment (WEEE)
Natural resources were used in the production of this equipment. This
equipment may contain materials that are hazardous to health and the
environment. To avoid harm to the environment and natural resources, the
use of appropriate take-back systems is recommended. The crossed out
wheeled bin symbol on the meter encourages the use of these systems when
disposing of this equipment.
Take-back systems will allow the materials to be reused or recycled in a
way that will not harm the environment. For more information on approved
collection, reuse, and recycling systems contact local or regional waste
administration or recycling services.
GENERAL OPERATING INFORMATION
OVERVIEW
The meter is a portable, microprocessor controlled, direct reading colorimeter.
It has a graphical liquid crystal display and 6 button keypad. These allow the
user to select options from the menu driven software, to directly read test
results or to review stored results of previous tests in the data logger. The
menus can be displayed in seven diff erent languages.
The LaMotte tests are precalibrated for LaMotte reagent systems.
The colorimeter displays the result of these tests directly in units of
concentration.
The optics feature a colored LED. The LED has a corresponding silicon
photoiode with an integrated interference fi lter. The interference fi lter selects
a narrow band of light from the corresponding LED for the colorimetric
measurements.
A USB wall adapter, USB computer connection or lithium battery powers the
colorimeter.
GENERAL OPERATING INFORMATION
The operation of the colorimeter is controlled by the menu driven software
and user interface. A menu is a list of choices. This allows a selection of
various tasks for the colorieter to perform, such as scan blank and scan
sample. The keypad is used to make menu selections that are viewed on the
display.
The keypad has 6 buttons which are used to perform specifi c tasks.
This button will scroll up through a list of menu selections.
ENTER
The button is used to select choices in a menu viewed on the
display.
This button controls the backlight on the display.
This button will scroll down through a list of menu selections.
EXIT
This button exits to the previous menu.
This button turns the meter on or off .
ENTER
EXIT
THE DISPLAY AND MENUS
The display allows menu selections to be viewed and selected. These
selections instruct the colorimeter to perform specifi c tasks. The menus are
viewed in the display using two general formats that are followed from one
menu to the next. Each menu is a list of choices or selections.
The display has a header line at the top and a footer line at the bottom. The
header displays the title of the current menu. The footer line displays the
time and the date, the data logger status and the battery status. The menu
selection window is in the middle of the display between the header and the
footer.
The menu selection window displays information in two general formats. In
the fi rst format only menu selections are displayed. Up to 4 lines of menu
selections may be displayed. If more selections are available they can be
viewed by pressing the arrow buttons to scroll the other menu
selections into the menu selection window. Think of the menu selections as
a vertical list in the display that moves up or down each time an arrow button
is pressed. Some menus in the colorimeter are looping menus.
The top and bottom menu choices are connected in a loop. Scrolling down
past the bottom of the menu will lead to the top of the menu. Scrolling up
past the top of the menu will lead to the bottom of the menu.
A light bar will indicate the menu choice. As the menu is scrolled through, the
light bar will highlight diff erent menu choices. Pressing the
select the menu choice that is indicated by the light bar.
In the second format the menu choice window takes advantage of the
graphical capabilities of the display. Large format graphic information, such
as test results or error messages or the LaMotte logo is displayed. The top
two lines of the display are used to display information in a large, easy to read
format. The menus work in the same way as previously described but two lines
of the menu are visible at the bottom of the display.
HeaderMenu Title
Message or Result Window
ENTER
button will
Result or
Message
Another
Main Window Selection And Another
Footer
As described previously, the
current menu and a return to the previous menu. This allows a rapid exit from
an inner menu to the main menu by repeatedly pushing the
Pushing at any time will turn the colorimeter off .
More choices are available and can be viewed by scrolling
up and/or down through the display.
HeaderIdentifi es the current menu and information on units and
reagent systems if applicable.
FooterIn the data logging mode the number of the data point
is displayed and the total number of data points in the
memory will be shown. The footer also shows current time
and battery status
TUBES AND CHAMBERS
The colorimeter uses one type of tube (Code 0290) for all test factors.
The handling of the tubes is of utmost importance. Tubes must be clean and
free from lint, fi ngerprints, dried spills and signifi cant scratches, especially
the central zone between the bottom and the sample line.
Scratches, fi ngerprints and water droplets on the tube can cause stray light
interference leading to inaccurate results. Tubes that have been scratched in
the light zone through excessive use should be discarded and replaced with
new ones.
Tubes should always be washed on the inside and outside with mild detergent
prior to use to remove dirt or fi ngerprints. The tubes should be allowed to airdry in an inverted position to prevent dust from entering the tubes. Dry tubes
should be stored with the caps on to prevent contamination.
After a tube has been fi lled and capped, it should be held by the cap and the
outside surface should be wiped with a clean, lint-free absorbent cloth until it
is dry and smudge-free. Handling the tube only by the cap will avoid problems
from fi ngerprints. Always set the clean tube aside on a clean surface that will
not contaminate the tube. It is imperative that the tubes and light chamber be
clean and dry. The outside of the tubes should be dried with a clean, lint-free
cloth or disposable wipe before they are placed in the meter chamber.
Tubes should be emptied and cleaned as soon as possible after reading a
sample to prevent deposition of particulates on the inside of the tubes.
Variability in the geometry of the glassware and technique is the predominate
cause of variability in results. Slight variations in wall thickness and the
diameter of the tubes may lead to slight variations in the test results. To
eliminate this error the tubes should be placed in the chamber with the same
orientation each time.
Chambers which have been scratched through excessive use should be
discarded and replaced with a new one.
If a test result is out of the range of the meter, it must be diluted. The test
should then be repeated on the diluted sample. The following table gives quick
reference guidelines for dilutions of various proportions.
Amount of SampleDeionized Water to Bring
Final Volume to 10 mL
10 mL0 mL1
5 mL5 mL2
2.5 mL7.5 mL4
1 mL9 mL10
0.5 mL9.5 mL20
All dilutions are based on a fi nal volume of 10 mL, so several dilutions will
require small volumes of the water sample. Graduated pipets should be
used for all dilutions. If volumetric glassware is not available, dilutions can
be made with the colorimeter tube. Fill the tube to the 10 mL line with the
sample and then transfer it to another container. Add 10 mL volumes of
deionized water to the container and mix. Transfer 10 mL of the diluted
sample to the colorimeter tube and follow the test procedure. Repeat the
dilution and testing procedures until the result falls within the range of the
calibration. Multiply the test result by the dilution factor. For example, if 10 mL
of the sample water is diluted with three 10 mL volumes of deionized water,
the dilution factor is four. The test result of the diluted sample should be
multiplied by four.
Multiplication Factor
GENERAL INFORMATION
PACKAGING AND DELIVERY
Experienced packaging personnel at LaMotte Company assure adequate
protection against normal hazards encountered in transportation of
shipments.
After the product leaves LaMotte Company, all responsibility for safe delivery
is assured by the transportation company. Damage claims must be fi led
immediately with the transportation company to receive compensation for
damaged goods.
GENERAL PRECAUTIONS
READ THE INSTRUCTION MANUAL BEFORE ATTEMPTING TO SET UP OR
OPERATE THE METER. Failure to do so could result in personal injury or
damage to the meter. The meter should not be used or stored in a wet or
corrosive environment. Care should be taken to prevent water from wet tubes
from entering the meter chamber.
Read the labels on all LaMotte reagent containers prior to use. Some
containers include precautionary notices and fi rst aid information. Certain
reagents are considered hazardous substances and are designated with
a * in the instruction manual. Safety Data Sheets (SDS) can be found at
Keep equipment and reagent chemicals out of the reach of young children.
Ensure that the protection provided by this equipment is not impaired. Do not
intall or use this equipment in a manner that is not indicated in this manual.
LIMITS OF LIABILITY
Under no circumstances shall LaMotte Company be liable for loss of life,
property, profi ts, or other damages incurred through the use or misuse of its
products.
The colorimeter has been independently tested and has earned the European
CE Mark of compliance for electromagnetic compatibility and safety. To view
certifi cates of compliance, go to
IP67 Certifi cation
The meets IP67 standards for protection against dust and immersion only
when the USB port plug is in place. Documentation is av
If the message OVERRANGE is displayed when scanning a sample, the
sample may be over range or under range. If the sample is over range the
sample should be diluted and tested again (see Sample Dilution Techniques
and Volumetric Measurements, page 26).
If overrange is displayed, press
ENTER
to continue testing on diluted
samples.
Note: After pressing
overrange cncentration will be
displayed. This concentration is an
approximation only.
ENTER
, the
Overrange
<Enter> continue
Scan Blank
Scan Sample
12:00:00 001/500
022 Chlorine Dioxide
CALIBRATION
As with all pre-calibrated meters, it is highly recommended, even if not
required by regulations, that the user periodically verify the performance of
the meter by running standards with a predetermined concentration. Results
outside of specifi cation are an indication that the meter needs to be adjusted.
This can be done following the user calibration described on page 10. If the
user calibration fails to properly adjust the meter then the meter should be
returned to LaMotte Company for recalibration. (See page 21).
STRAY LIGHT
The colorimeter should have no problems with stray light. Make sure that the
sample compartment lid is always fully closed.