LaMotte 2050-DW6 User manual

Page 1
ColorQ Photometer for Drinking Water
For software Version 5.73 or higher
To insure proper testing results.
Insert tubes into ColorQ with line
and arrow facing forward.
Sun = Shade meter from bright
sunlight. Press/Hold button to turn o .
Er7 =Tube is misaligned or
Brush and rinse all tubes
promptly after use.
Rinse tablet crusher between test
factors.
Allow tubes to dry before
returning to case.
Replace reagent caps securely.
Use only LaMotte replacement
reagents and ColorQ test tubes Code 0201.
If unit should auto-o during
testing rinse and Þ ll one tube to Blank and return to testing.
To skip test in the sequence press
button to advance past that test factor.
To turn meter o at any time,
press and hold the button until o is displayed.
bAt = battery is low.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
1. Fill sample bottle with water sample. Replace cap.
2. Fill clean tube (#0201) to the 5mL line with water sample.
3. Insert tube into ColorQ as shown.
4. Press button to turn meter on. When bLA appears press button to Blank the meter, and go to pH. Remove tube.
pH - Code 7059-E
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 3
+
*Hardness 1 Buffer: Code 7045-G
Hardness 2 Indicator: Code 7046-G
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 4
*Iron IG- Code 3725A-H
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 5
DPD#1 IG - Code 6903A-H DPD#3 IG - Code 6197A-H
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 6
*Sulfide Reagent A: Code V-4458LWT-G
*Sulfide Reagent B: Code V-4459-E
Sulfide Reagent C: Code 4460LWT-G
+
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 7
*Nitrate IG - Code 3884A-H
18
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 8
Read all instructions before testing.
If not testing all tests in the sequence, follow
blanking procedure and then press button until desired test factor is shown. Follow test procedure.
In one complete inversion, the tube will be
turned cap down and then right side up. The air bubble will move slowly to the bottom of the tube and back again to the cap end.
If small bubbles form after adding reagents,
tap bottom of tube sharply once or twice to dislodge bubbles. Bubbles will interfere with the test results.
Samples with concentrations greater than the
range of the reagent system must be diluted. Add 2.5 mL of sample water to a test tube. Fill to the 5 mL line with distilled or deionized water. Cap and mix. Blank with the diluted sample. Follow the test procedure with the diluted sample. Multiply the displayed result by
2. (This will not work for the pH test).
Be sure the outside of the tubes are dry before
putting them into the ColorQ. This is best accomplished by dispensing the water sample from the sample bottle.
Clean ColorQ optics with a damp cotton swab.
Avoid abrasive cleaners that can damage plastic.
Do not use the test tube brush to clean the
ColorQ meter chamber.
Replace stained or scratched tubes (0201).
Always rinse tubes and caps after testing.
Reagent left behind can a ect the next test or stain the tube.
Replace liquid reagents annually. To learn more
about reagent storage guidelines and shelf-life go to Support Section then Reagent ReÞ lls/ Shelf Life at www.lamotte.com.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com