LaMotte 2020i User Manual

e
/i
2020e • Code 1979-EPA
2020i • Code 1979-ISO
Version 1.5 • Code 1979-MN • 12-07
WARNING! This set contains chemicals
that may be harmful if misused. Read
cautions on individual containers
except under adult supervision
CONTENTS
Introduction............................6
Turbidity ...................................6
WhatisTurbidity?............................6
HowisTurbidityMeasured?........................7
TakingTurbidityWaterSamples......................7
Sample Dilution Techniques .........................7-8
OptionsandSetUp........................8
FactoryDefaultSettings ............................8
ResettoFactoryDefaultSettings ........................8
Turbitity....................................8
SelectingTurbidityUnits ........................8-9
SelectingaTurbidityCalibrationCurve................10-11
Averaging................................11-12
SettingDate&Time............................12-13
SelectingaLanguage.............................14
SettingAutoShutdown............................15
DataLogging.........................16-17
TubePositioningRing.......................18
Analysis&Calibration ......................18
Analysis ..............................19-21
Dilution Procedures ...........................22
PreparationofTurbidity-FreeWater..................22-23
TurbidityStandards...........................24
CalibrationProcedure........................24-28
TestingTips...............................29
Troubleshooting Guide ..........................30
ErrorMessages ...............................30
Troubleshooting...............................30
StrayLight..................................31
GeneralOperatingInformation..................30
Overview..................................31
TheKeypad ...............................32-33
The Display & Menus ...........................34-36
NegativeResults...............................35
Tubes....................................36
PCLINK.............................37
Output...................................37
4 CONTENTS
Computer Connection ............................37
BatteryOperation ........................37
ReplacingtheBattery.............................37
Maintenance...........................38
Cleaning................................38
Repairs ................................38
MeterDisposal.............................38
GeneralInformation.......................39
PackagingandDelivery............................39
General Precautions .............................39
SafetyPrecautions..............................39
Limits of Liability ...............................39
Specifications................................40
StatisticalandTechnicalDefinitionsRelatedtoProductSpecifications.......41
ContentsandAccessories...........................42
EPACompliance...............................42
ISOCompliance .............................. 42
CEComplicance...............................42
Warranty ..................................42
CONTENTS 5
INTRODUCTION
TURBIDITY
What is Turbidity?
Turbidity is an aggregate property of the solution, water in most cases. Turbidity is not specific to the types of particles in the water. They could be suspended or colloidal matter and they can be inorganic, organic or biological. At high concentrations turbidity is perceived as cloudiness or haze or an absence of clarity in the water. Turbidity is an optical property that results when light passing through a liquid sample is scattered. The scattering of light results in a change in the direction of the light passing through the liquid. This is most often caused when the light strikes particles in solution and is scattered backward, sideways and forward. If the turbidity is low much of the light will continue in the original direction. Light scattered by the particles allows the particle to be ”seen” or detected in solution. Just as sunlight passing through a window is a scattered by dust particles in the air, allowing them to be seen.
In the past 10 years, turbidity has become more than just a measure of water clarity. Because of the emergence of pathogens such as Cryptosporidium and Giardia, turbidity now holds the key to assuring proper water filtration. In 1998, the EPA published the IESWTR (interim enhanced surface water treatment rule) mandating turbidities in combined filter effluent to read at or below 0.3 NTU. By doing so, the EPA hoped to achieve a 2 log (99%) removal of Cryptosporidium. There is presently consideration to lower this to 0.1 NTU. The trend has been to check the calibration of on-line turbidimeters with hand-held field units. The optical design and low detection limit of the 2020 allow very accurate readings for such calibrations.
The meter also allows the user to choose the units of measure for expressing turbidity. While nephelometric turbidity unit (NTU) has been the standard for years, FNU (formazin nephelometric unit) and FAU (formazin attenuation unit) are now being used in ISO 7027 units. American Society of Brewing Chemists (ASBC) units and European Brewery Convention (EBC) units allow the brewing industry to check process waters.
How is Turbidity Measured?
Turbidity is measured by detecting and quantifying the scattering of light in water (solution). Turbidity can be measured in many ways. There are visual methods and instrumental methods. Visual methods are more suitable for samples with high turbidity. Instrumental methods can be used on samples with both high and low levels of turbidity.
Two visual methods are the Secchi Disk method and the Jackson Candle method. The Secchi Disk method is often used in natural waters. A black and white Secchi Disk is lowered into the water until it can no longer be seen. It is then raised until it can be seen again. The average of these two distances is known as the “Secchi Depth”. The Jackson Candle method uses a long glass tube over a standard candle. Water is added or removed from the tube until the candle flame becomes indistinct. The depth of the water measured with a calibrated scale is reported as Jackson Turbidity Units (JTU). The lowest turbidity that can be determined with this method is about 25 NTU. There are two common methods for instruments to measure turbidity. Instruments can measure the attenuation of a light beam passing through a sample and they can measure the scattered light from a light beam passing through a sample. In the attenuation method, the intensity of a light beam passing through turbid sample is
6 INTRODUCTION
compared with the intensity passing through a turbidity-free sample at 180° from the light source. This method is good for highly turbid samples. The most common instrument for measuring scatter light in a water sample is a nephelometer. A nephelometer measures light scattered at 90° to the light beam. Light scattered at other angles may also be measured, but the 90° angle defines a nephelometric measurement. The light source for nephelometric measurements can be one of two types to meet EPA or ISO specifications. The EPA specifies a tungsten lamp with a color temperature of 2,200–3,000 K. The units of measurement for the EPA method are nephelometric turbidity units (NTU). The ISO specifies a light emitting diode (LED) with a wavelength of 860 nm and a spectral bandwidth less than or equal to 60 nm. The units of measurement for the ISO method are formazin nephelometric units (FNU). The 2020e meets the EPA specification and the 2020i meets the ISO specification. The nephelometric method is most useful for low turbidity.
The 2020 is a nephelometer that is capable of measuring turbidity by both the attenuation method and the nephelometric method. It uses a detector placed at 180° to the light source for the attenuation method. It uses a detector placed at 90° to the light source for the Nephelometric method. The 2020 also has a third detector that monitors the intensity of the light source. It uses this detector to improve instrumental stability and minimize calibration drift. The 2020 also has a signal averaging option to improve the stability of readings on low turbidity samples.
The 2020 has two different turbidity calibrations, formazin and Japan Standard. The formazin calibration is the EPA and ISO approved method of calibrating nephelometers. This calibration can be used with user prepared formazin standards or commercially purchased formazin standards. LaMotte Company approved AMCO
TM
standards labeled for use with the 2020 can also be used with the formazin calibration. Stablcal
â
standards below 50 NTU should not be used to calibrate the
2020. The Japan Standard calibration is a calibration for a Japanese Water Works standard.
It is based on Japanese formulated polystyrene turbidity standards. This calibration should only be use to meet Japanese Water Works requirements. The Japanese polystyrene standards can only be purchased in Japan. Formazin, AMCO and Stablcal
â
standards cannot be used with this calibration.
Taking Turbidity Water Samples
Clean plastic or glass containers may be used for turbidity samples. Ideally, samples should be tested soon after collection and at the same temperature as when collected.
SAMPLE DILUTION TECHNIQUES
If a test result is out of the range of the meter, it must be diluted. The test should then be repeated on the diluted sample. The following table gives quick reference guidelines for dilutions of various proportions.
Amount of Sample Deionized Water to Bring
Final Volume to 10 mL 10 mL 0 mL 1 5mL 5mL 2
2.5 mL 7.5 mL 4 1mL 9mL 10
0.5 mL 9.5 mL 20
INTRODUCTION•SAMPLE DILUTION TECHNIQUES 7
Multiplication Factor
All dilutions are based on a final volume of 10 mL so several dilutions will require small volumes of the water sample. Graduated pipets should be used for all dilutions. If volumetric glassware is not available, dilutions can be made with the colorimeter tube. Fill the tube to the 10 mL line with the sample and then transfer it to another container. Add 10 mL volumes of deionized water to the container and mix. Transfer 10 mL of the diluted sample to the colorimeter tube and follow the test procedure. Repeat the dilution and testing procedures until the result falls within the range of the calibration. Multiply the test result by the dilution factor. For example, if 10 mL of the sample water is diluted with three 10 mL volumes of deionized water, the dilution factor is four. The test result of the diluted sample should be multiplied by four.
OPTIONS & SET UP
FACTORY DEFAULT SETTINGS
Settings that have user options have been set at the factory to default settings.
The factory default settings are:
Turbidity Units NTU Turbidity Calibration formazin Averaging Disabled Date Format MM/DD/YY Language English Auto Shutdown Disabled
RESET TO FACTORY DEFAULT SETTINGS
To return the meter to the factory settings, turn the meter off. Hold down *IOK button. Press ON. Release both buttons. Press *IOK button to select the default settings. Meter will turn off and the factory settings will be restored. Restoring the factory settings will remove the user-level calibration but not the zeroing. To change the default settings follow the instructions in the following sections.
TURBIDITY
The default units are NTU and the default calibration curve is formazin. To change the settings:
SELECTING TURBIDITY UNITS
1. Press ON to turn the meter on.
1.3
8 OPTIONS & SETUPS • TURBIDITY UNITS
2. Scroll down and then press *IOK
to select Options.
Main Menu
Measure Data Logging
* Options
16:02:19 01/04/05
3. Scroll down and then press *IOK
to select Turbidity.
Options
Averaging
* Turbidity
Date/Time Language
16:02:19 01/04/05
4. Press the *IOK to select Units. Turbidity
*Units
Calibration
16:02:19 01/04/05
5. Scroll down and then press *IOK
to select units. Available units are: NTU
(Nephelometric Turbidity Units); FNU (Formazin Nephelometric Units); ASBC (American Society of Brewing Chemists); EBC (European
Units
NTU FNU ASBC EBC
16:02:19 01/04/05
Brewery Convention)
Note: If Attenuation is chosen as a calibration curve. The result will be reported in
FAU (Formazin Attenuation Units).
6. Press to exit to a previous menu or make another menu selection or press OFF to turn the meter off.
OPTIONS & SETUP • TURBIDITY UNITS 9
SELECTING A TURBIDITY CALIBRATION CURVE
1. Press ON to turn the meter on.
1.3
2. Scroll down and then press *IOK to
select Options.
3. Scroll down and then press *IOK
to select Turbidity.
4. Scroll down and then press *IOK
to select Calibration.
5. Scroll down and then press *IOK
to select a Calibration curve. Select a calibration option based on the composition of the standards that will be used to calibrate the meter. Available options are: Formazin,
Japan Standard, Attenuation.
Main Menu
Measure Data Logging
* Options
16:02:19 01/04/05
Options
Averaging
* Turbidity
Date/Time Language
16:02:19 01/04/05
Turbidity
Units
* Calibration
16:02:19 01/04/05
Turbidity
Units
* Calibration
16:02:19 01/04/05
For the most accurate results, the Attenuation option should be chosen when samples are over 500 NTU. The range for the Attenuation option is 40–4000 NTU.
Calibration * Formazin
Japan Standard Attenuation
16:02:19 01/04/05
10 OPTIONS & SETUP • TURBIDITY CALIBRATION
The Japan Standard calibration mode should be used only with Japanese Polystyrene Standards (0–100 NTU).
Note: StablCal
®
standards below 50 NTU should not be used to calibrate the
2020. The diluent has a different refractive index than traditional formazin standards and will affect the results.
6. Press to exit to a previous menu or make another menu selection or press OFF to turn the meter off.
AVERAGING
The averaging option allows the user to average multiple readings. This option will improve the accuracy of samples with readings that may tend to drift with time. When the two, five or ten reading options have been selected, the meter will show a running average of the readings that have been taken until the final average is displayed. The default setting is disabled. To change the setting:
1. Press ON to turn the meter on.
1.3
2. Scroll down and then press *IOK to select Options.
Main Menu
Measure Data Logging
* Options
16:02:19 01/04/05
3. Press *IOK to select Averaging. Options
* Averaging
Turbidity Date/Time Language
16:02:19 01/04/05
OPTIONS & SETUP • AVERAGING 11
4. Scroll down and then press *IOK to
select an averaging option. Available options are:
Disabled, 2 Measurements, 5 Measurements, 10 Measurements.
Note: The current setting.
5. Press to exit to a previous menu or make another menu selection or press OFF to turn the meter off.
Note: When the Averaging option is selected, it will take longer to get the final result and more power will be used.
is displayed next to the
*
SETTING THE DATE AND TIME
1. Press ON to turn the meter on.
Averaging
Disabled
* 2 Measurements
5 Measurements 10 Measurements
16:02:19 01/04/05
1.3
2. Scroll down and then press *IOK to select Options.
3. Scroll down and then press *IOK to select Date/Time.
12 OPTIONS & SETUP • SET DATE & TIME
Main Menu
Measure Data Logging
* Options
16:02:19 01/04/05
Options
Averaging Turbidity
* Date/Time
Language
16:02:19 01/04/05
4. Press *IOK to select Set Date or scroll down and then press *IOK to select Set Time or Date Format.
Date/Time *Set Date
Set Time Date Format
16:02:19 01/04/05
5. When setting the time or the date, use the or to change the highlighted number on the display. Press *IOK to accept the value and move to the next value.
When choosing a date format, use the or to select a date format. Press *IOK to accept format
Set Date
01/0/4/05
q , * , p
16:02:19 01/04/05
Set Date
01/04/05
q , * , p
16:02:19 01/04/05
Set Date
01/04/05
q , * , p
16:02:19 01/04/05
Date/Time *Set Date
Set Time Date Format
16:02:19 01/04/05
6. Press to exit to a previous menu or make another menu selection or press OFF to turn the meter off.
OPTIONS & SETUP • SET DATE & TIME 13
SELECTING A LANGUAGE
The default setting is English. To change the setting:
1. Press ON to turn the meter on.
1.3
2. Scroll down and then press *IOK
to select Options.
3. Scroll down and then press *IOK
to select Language.
4. Scroll down and then press *IOK
to select a language. Available languages are: English,
French, Spanish, Japanese (Kana), Portuguese, Italian.
5. Press to exit to a previous menu or make another menu selection or press OFF to turn the meter off.
Main Menu
Measure Data Logging
* Options
16:02:19 01/04/05
Options
Averaging Turbidity
Date Time * Language 16:02:19 01/04/05
Language * English
Frances
Español
Japanese 16:02:19 01/04/05
14 OPTIONS & SETUP • SELECTING A LANGUAGE
Loading...
+ 30 hidden pages