The Keypad ····················································································· 22
Sample Holders ················································································ 22
The Display & the Menus ······························································· 23
Looping Menus ················································································ 25
TESTING
Testing Menu ·················································································· 26
Sequences of Tests ··········································································· 27
General Testing Procedures ····························································· 28
Testing With the LaMotte Pre-Programmed Tests ·························· 28
Calibrating With the LaMotte Pre-Programmed Tests····················· 30
Measuring in the %T/ABS Mode ···················································· 32
COD PLUS COLORIMETER 4/033
T ABLE OF CONTENTS (cont.)
EDITING MENU
Edit a Sequence ·············································································· 34
Adding or Deleting Tests ································································· 35
Edit User Tests ················································································ 38
Naming the Test ·············································································· 39
Selecting the Vial and Wavelength ················································· 41
Entering a New Calibration ···························································· 42
Selecting the Numerical Format of the Result ································ 44
Selecting Units of Concentration····················································· 45
Setting the Clock ············································································· 46
Turning the Data Logger On and Off ·············································· 47
Factory Setup ··················································································· 48
Setting the Power Saver Function···················································· 48
Experienced packaging personnel at LaMotte Company assure adequate
protection against normal hazards encountered in transportation of shipments.
After the product leaves the manufacturer, all responsibility for its safe delivery
is assured by the transportation company. Damage claims must be filed
immediately with the transportation company to receive compensation for
damaged goods.
Should it be necessary to return the instrument for repair or servicing, pack
instrument carefully in suitable container with adequate packing material. A
return authorization number must be obtained from LaMotte Company by
calling 1-800-344-3100. Attach a letter with the authorization number to the
shipping carton which describes the kind of trouble experienced. This valuable
information will enable the service department to make the required repairs
more efficiently.
GENERAL PRECAUTIONS
n
Before attempting to set up or operate this instrument it is important to read
the instruction manual. Failure to do so could result in personal injury or
damage to the equipment.
The COD PLUS Colorimeter should not be stored or used in a wet or corrosive
environment. Care should be taken to prevent water or reagent chemicals from
wet colorimeter tubes from entering the colorimeter chamber.
NEVER PUT WET TUBES IN COLORIMETER.
SAFETY PRECAUTIONS
n
Read the labels on all LaMotte reagent containers prior to use. Some
containers include precautionary notices and first aid information. Certain
reagents are considered hazardous substances and are designated witha*inthe
instruction manual. Material Safety Data Sheets (MSDS) are supplied for these
reagents. Read the accompanying MSDS before using these reagents.
Additional information for all LaMotte reagents is available in the United
States from the Poison Control Center listed in the front of the phone book or
by calling 1-800-222-1222. Call 1-813-248-0585 to contact the International
Poison control Center. LaMotte reagents are registered with a computerized
poison control information system available to all local poison control centers.
Be prepared to supply the name and four-digit LaMotte code number found on
the container label , listed in the test procedures, or at the top of the MSDS.
Keep equipment and reagent chemicals out of the reach of young children.
Protect Yourself and Equipment: Use Proper Analytical Techniques
LIMITS OF LIABILITY
n
COD PLUS COLORIMETER 6/025
Under no circumstances shall LaMotte Company be liable for loss of life,
property, profits, or other damages incurred through the use or misuse of its
products.
SPECIFICATIONS
n
n
INSTRUMENT TYPE: Colorimeter
ReadoutGraphical 4 line, 16 character per line LCD
Wavelengths430nm, 620 nm
Wavelength Accuracy±2nm
Readable ResolutionDetermined by reagent system
Wavelength
10 nm typical
Bandwidth
Photometric Range-2to+2A
Photometric Precision± 0.001A
Sample ChamberAccepts 25 mm diameter flat-bottomed test tubes, 10
mm square cuvettes, 16 mm COD test tubes
Light Sources2 LEDs
Detectors2 silicon photodiodes with integrated interference
filters
ModesAbsorbance, pre-programmed tests
Pre-Programmed TestsYES, with automatic wavelength selection
User Defined TestsUp to 10 user tests can be input
RS232 Port8 pin mini-DIN, 9600b, 8, 1, n
Power RequirementsBattery Operation: 9 volt alkaline
Line Operation: 110/220V AC;
50/60 Hz with adapter, 6V 500 mA DC
Dimensions (LxWxH)8.5 x 16.2 x 16.7 cm, 3.4 x 6.4 x 2.6 inches
Weight312 g, 11 oz (meter only)
Data Logger350 test results stored for download to a PC
SMARTLink 2 Program & Interface Cable (3.5 disk) Code 1912-3
SMARTLink 2 Program & Interface Cable (CD)Code 1912-CD
EPA COMPLIANCE
n
The COD PLUS Colorimeter is an EPA-Accepted instrument. EP A-Accepted
means that the instrument meets the requirements for instrumentation as
found in test procedures that are approved for the National Primary Drinking
Water Regulations (NPDWR) or National Pollutant Discharge Elimination
System (NPDES) compliance monitoring programs. EPA-Accepted
instruments may be used with approved test procedures without additional
approval.
COD PLUS COLORIMETER 6/027
CE COMPLIANCE
n
The COD PLUS Colorimeter has earned the European CE Mark of
Compliance for electromagnetic compatibility and safety.
I, the undersigned, hereby declare that the equipment specified above
conforms to the above Directive and Standards.
Signature
1/15/02
Date
8COD PLUS COLORIMETER 6/02
Scott H. Steffen
Name
VP New Products & Quality
Position
CHEMICAL TESTING
WATER SAMPLING FOR CHEMICAL ANALYSIS
n
n
Taking Representative Samples
The underlying factor to be considered for any type of water sampling is
whether or not the sample is truly representative of the source. To properly
collect a representative sample:
l
Sample as frequently as possible.
l
Collect a large sample or at least enough to conduct whatever tests are
necessary.
l
Make a composite sample for the same sampling area.
l
Handle the sample in such a way as to prevent deterioration or
contamination before the analysis is performed.
l
Perform analysis for dissolved gases such as dissolved oxygen, carbon
dioxide, and hydrogen sulfide immediately at the site of sampling. These
factors, as well as samples for pH, cannot be stored for later examination.
l
Make a list of conditions or observations which may affect the sample.
Other considerations for taking representative samples are dependent
upon the source of the sample. Taking samples from surface waters
involves different considerations than taking samples from impounded and
sub-surface waters.
n
Sampling of Open Water Systems
Surface waters, such as those found in streams and rivers, are usually well
mixed. The sample should be taken downstream from any tributary, industrial
or sewage pollution source. For comparison purposes samples may be taken
upstream and at the source of the pollution before mixing.
In ponds, lakes, and reservoirs with restricted flow, it is necessary to collect a
number of samples in a cross section of the body of water, and where possible
composite samples should be made to ensure representative samples.
To collect samples from surface waters, select a suitable plastic container with a
tight fitting screw cap. Rinse the container several times with the sample to be
tested, then immerse the container below the surface until it is filled to
overflowing and replace the cap. If the sample is not to be tested immediately,
pour a small part of the sample out and reseal. This will allow for any
expansion. Any condition which might affect the sample should be listed.
Sub-surface sampling is required to obtain a vertical profile of streams, lakes,
ponds, and reservoirs at specific depths. This type of sampling requires more
sophisticated sampling equipment.
For dissolved oxygen studies, or for tests requiring small sample sizes, a Water
Sampler (LaMotte Code 1060) will serve as a subsurface or in-depth sampler.
COD PLUS COLORIMETER 6/029
This weighted device is lowered to the sampling depth and allowed to rest at
this depth for a few minutes. The water percolates into the sample chamber
displacing the air which bubbles to the surface. When the bubbles cease to rise,
the device has flushed itself approximately five times and it may be raised to
the surface for examination. The inner chamber of the sampling device is lifted
out and portions of the water sample are carefully dispensed for subsequent
chemical analysis.
A Snap-Plunger Water Sampler (LaMotte Code 1077) is another “in-depth”
sampling device which is designed to collect large samples which can be used
for a multitude of tests. Basically, this collection apparatus is a hollow cylinder
with a spring loaded plunger attached to each end. The device is cocked above
the surface of the water and lowered to the desired depth. A weighted
messenger is sent down the calibrated line to trip the closing mechanism and
the plungers seal the sample from mixing with intermediate layers as it is
brought to the surface. A special drain outlet is provided to draw off samples for
chemical analysis.
n
Sampling of Closed System
To obtain representative samples from confined water systems, such as pipe
lines, tanks, vats, filters, water softeners, evaporators and condensers, different
considerations are required because of chemical changes which occur between
the inlet and outlet water. One must have a basic understanding of the type of
chemical changes which occur for the type of equipment used. Also,
consideration should be given to the rate of passage and retaining time for the
process water.
Temperature changes play an important part in deciding exactly what test
should be performed. Process water should be allowed to come to room
temperature, 20–25°C, before conducting any tests.
When drawing off samples from an outlet pipe such as a tap, allow sample to
run for several minutes, rinsing the container several times before taking the
final sample. Avoid splashing and introduction of any contaminating material.
FILTRATION
n
When testing natural waters that contain significant turbidity due to suspended
solids and algae, filtration is an option. Reagent systems, whether EPA,
Standard Methods, LaMotte or any others, will generally only determine
dissolved constituents. Both EPA and Standard Methods suggest filtration
through a 0.45 micron filter membrane, to remove turbidity, for the
determination of dissolved constituents.** T o test for total constituents,
organically bound and suspended or colloidal materials, a rigorous high
temperature acid digestion is necessary.
**LaMotte offers a filtering apparatus: syringe assembly (Code 1050) and membrane
filters, 0.45 micron, (Code 1103).
10COD PLUS COLORIMETER 6/02
AN INTRODUCTION TO COLORIMETRIC ANALYSIS
n
Most test substances in water are colorless and undetectable to the human eye.
To test for their presence we must find a way to “see” them. The COD PLUS
Colorimeter can be used to measure a test substance that is itself yellow or
green to blue, or can be reacted to produce these colors. In fact a simple
definition of colorimetry is “the measurement of color” and a colorimetric
method is “any technique used to evaluate an unknown color in reference to
known colors”. In a colorimetric chemical test the intensity of the color from
the reaction must be proportional to the concentration of the substance being
tested. Some reactions have limitations or variances inherent to them that may
give misleading results. Many such interferences are discussed with each
particular test instruction. In the most basic colorimetric method the reacted
test sample is visually compared to a known color standard. However, accurate
and reproducible results are limited by the eyesight of the analyst,
inconsistencies in the light sources, and the fading of color standards.
To avoid these sources of error, a colorimeter can be used to photoelectrically
measure the amount of colored light absorbed by a colored sample in reference
to a colorless sample (blank).
White light is made up of many different colors or wavelengths of light. A
colored sample typically absorbs only one color or one band of wavelengths
from the white light. Only a small difference would be measured between white
light before it passes through a colored sample versus after it passes through a
colored sample. The reason for this is that the one color absorbed by the
sample is only a small portion of the total amount of light passing through the
sample. However, if we could select only that one color or band of wavelengths
of light to which the test sample is most sensitive, we would see a large
difference between the light before it passes through the sample and after it
passes through the sample.
The COD PLUS Colorimeter passes one of two colored light beams through
one of two optical filters which transmits only one particular color or band of
wavelengths of light to the photodectector where it is measured. The difference
in the amount of colored light transmitted by a colored sample is a
measurement of the amount of colored light absorbed by the sample. In most
colorimetric tests the amount of colored light absorbed is directly proportional
to the concentration of the test factor producing the color and the path length
through the sample. However, for some tests the amount of colored light
absorbed is inversely proportional to the concentration.
The choice of the correct wavelength for testing is important. It is interesting
to note that the wavelength that gives the most sensitivity (lower detection
limit) for a test factor is the complementary color of the test sample. For
example the Nitrate-Nitrogen test produces a pink color proportional to the
nitrate concentration in the sample (the greater the nitrate concentration, the
darker the pink color). A wavelength in the green region should be selected to
analyze this sample since a pinkish-red solution absorbs mostly green light.
COD PLUS COLORIMETER 6/0211
REAGENT BLANK
n
Some tests will provide greater accuracy if a reagent blank is determined to
compensate for any color or turbidity resulting from the reagents themselves. A
reagent blank is performed by running the test procedure on 10 mL of
demineralized water. Use sample water to SCAN BLANK. Insert the reagent
blank in the colorimeter chamber and select SCAN SAMPLE. Note result of
reagent blank. Perform the tests on the sample water as described. Subtract
results of reagent blank from all subsequent test results. NOTE: Some tests
require a reagent blank to be used to SCAN BLANK.
COLORIMETER TUBES
n
Colorimeter tubes which have been scratched through excessive use should be
discarded and replaced with new ones. Dirty tubes should be cleaned on both
the inside and outside. Fingerprints on the exterior of the tubes can cause
excessive light scattering and result in errors. Handle the tubes carefully,
making sure the bottom half of the tube is not handled.
LaMotte Company makes every effort to provide high quality colorimeter
tubes. However, wall thicknesses and diameter of tubes may still vary slightly.
This may lead to slight variations in results (e.g. if a tube is turned while in the
sample chamber, the reading will likely change slightly). To eliminate this error
put the tubes into the sample chamber with the same orientation every time.
The tubes that are included with the colorimeter have an index mark to
facilitate this. If possible, use the same tube to SCAN BLANK and SCANSAMPLE.
SELECTING AN APPROPRIATE WAVELENGTH
n
The most appropriate wavelength to use when creating a calibration curve is
usually the one which gives the greatest change from the lowest reacted
standard concentration to the highest reacted standard concentration.
However, the absorbance of the highest reacted standard concentration should
never be greater than 2.0 absorbance units. Scan the lowest and highest reacted
standards at different wavelengths using the absorbance mode to find the
wavelength which gives the greatest change in absorbance without exceeding
2.0 absorbance units. Use this wavelength to create a calibration curve.
12COD PLUS COLORIMETER 6/02
Below is a list of suggested wavelengths for the color of the reacted samples.
Use these as a starting point.
Sample
Color
Yellow430
Pink520
Red570
Green and Blue620
Wavelength
Range
NOTE: Available wavelengths in the COD PLUS are 430 nm and 620 nm
only.
CALIBRATION CURVES
n
The COD PLUS Colorimeter contains precalibrated tests for the LaMotte
reagent systems (p. 49). The first step in using a non-LaMotte reagent system
with your COD PLUS Colorimeter is to create a calibration curve for the
reagent system. To create a calibration curve, prepare standard solutions of the
test factor and use the reagent system to test the standard solutions with the
COD PLUS Colorimeter. Select a wavelength for the test as described above.
Plot the results (in ABS or %Transmittance) versus concentration to create a
calibration curve. The calibration curve may then be used to identify the
concentration of an unknown sample by testing the unknown, reading
Absorbance or %T, and finding the corresponding concentration from the
curve. The linear range of the reagent system can be determined and this
information can be used to input a User Test into the COD PLUS Colorimeter
(see EDIT USER TESTS, p. 36).
n
PROCEDURE
1. Prepare 5 or 6 standard solutions of the factor being tested. The
concentration of these standards should be evenly distributed throughout
the range of the reagent system, and should include a 0 ppm standard
(distilled water). For instance, the solutions could measure 0, 10%, 30%,
50%, 70%, and 90% of the system’s maximum range.
2. Turn on the COD PLUS Colorimeter . Select the appropriate wavelength
from the absorbance mode. Be sure to select the appropriate wavelength for
the color produced by the reagent system.
3. Use the unreacted 0 ppm standard to standardize the colorimeter by using it
to scan blank.
4. Following the individual reagent system instructions, react each standard
solution beginning with 0 ppm. Continue with standards in increasing
concentration. Record the reading and the standard solution concentration
on a chart. Readings can be recorded as percent transmittance (%T) or
absorbance (A).
COD PLUS COLORIMETER 6/0213
5. Plot results on graph paper or computer using any available plotting
program. If results are as %T versus concentration, semilog graph paper
must be used. Plot the standard solution concentrations on the horizontal,
linear axis, and the %T on the vertical, logarithmic axis. If results are as
absorbance versus standard solution concentration, simple linear graph
paper can be used. Plot the standard solution concentration on the
horizontal axis, and the absorbance on the vertical axis.
6. After plotting the results, draw a line, or curve, of best fit through the
plotted points. The best fit may not connect the points. There should be
approximately an equal number of points above the curve as below the
curve. Some reagent systems will produce a straight line, while others
produce a curve. Many computer spreadsheet programs can produce the
curve of best fit by regression analysis of the standard solution data.
NOTE: Only reagent systems which produce a straight line can be used for a
User Test.
A sample of each type of graph appears below:
CALIBRATION CURVE
Absorbance vs. Concentration
2.0
1.8
1.6
1.4
1.2
1.0
0.8
0.6
Absorbance
0.4
0.2
0.0
123456
Concentration in ppm
CALIBRATION CURVE
%T vs. Concentration
100
10
%T Transmission
1
0
89107
123456
Concentration in ppm
891070
14COD PLUS COLORIMETER 6/02
n
PREPARING DILUTE STANDARD SOLUTIONS
Standard solutions should be prepared to create a calibration curve. Standard
solutions can be prepared by diluting a known concentrated standard by
specified amounts. A chart or computer spreadsheet can be created to
determine the proper dilutions. Use volumetric flasks and volumetric pipets for
all dilutions.
1. In Column A – Record the maximum concentration of test as determined
by the range and path length.
2. In Column B – Record the percent of the maximum concentration the
standard solution will be.
3. In Column C – Calculate the final concentration of the diluted standard
solutions by multiplying the maximum concentration (In Column A) by the
% of maximum concentration divided by 100. (C=Ax
B
).
100
4. In Column D – Record the final volume of the diluted sample (i.e. volume
of volumetric flask).
5. In Column E – Record the concentration of the original standard.
6. In Column F – Calculate the milliliters of original standard required
D
(C x
= F).
E
A sample chart appears below:
ABC=Ax
B
I00
DEF=Cx
Final
Maximum
concentration
of test
%of
Maximum
concentration
concentration
of Diluted
Standard
Volume of
Standard
Concentration
of Original
Standard
Standard
Required
10.0 ppm909.0 ppm100 mL1000 ppm0.90 mL
10.0 ppm707.0 ppm100 mL1000 ppm0.70 mL
10.0 ppm505.0 ppm100 mL1000 ppm0.50 mL
10.0 ppm303.0 ppm100 mL1000 ppm0.30 mL
10.0 ppm101.0 ppm100 mL1000 ppm0.10 mL
10.0 ppm00 ppm100 mL1000 ppm0 mL
mL of
Original
D
E
STANDARD ADDITIONS
n
A common method to check the accuracy and precision of a test is by standard
additions. In this method a sample is tested to determine the concentration of
the test substance. A second sample is then “spiked” by the addition of a
known quantity of the test substance. The second sample is then tested. The
determined concentration of the spiked sample should equal the concentration
of the first plus the amount added with the spike. The procedure can be
repeated with larger and larger “spikes.” If the determined concentrations do
not equal the concentration of the sample plus that added with the “spike”,
then an interference may exist.
COD PLUS COLORIMETER 6/0215
For example, a 10.0 mL water sample was determined to contain 0.3 ppm iron.
To a second 10.0 mL sample, 0.1 mL of 50 ppm iron standard was added. The
concentration of iron due to the “spike” was (0.10 mL x 50 ppm)/10.0 mL =
0.50 ppm. The concentration of iron determined in the spiked sample should
be 0.3 + 0.5 = 0.8 ppm iron. (Note: any error due to the increased volume from
the “spike” is negligible).
LaMotte offers a line of calibration standards which can be used to generate
calibration curves and perform standard additions.
SAMPLE DILUTION TECHNIQUES
n
& VOLUMETRIC MEASUREMENTS
If a test result using the COD PLUS Colorimeter gives an OVERRANGE message
then the sample concentration could be over range or under range. If it is over
range, the sample must be diluted. Then the test should be repeated on the
diluted sample to obtain a reading which is in the concentration range for the
test. (Note: This is not true for colorimetric determination of pH.)
Example:
Measure 5 mL of the water sample into a graduated cylinder. Add
demineralized water until the cylinder is filled to the 10 mL line. The sample
has been diluted by one-half, and the dilution factor is therefore 2. Perform the
test procedure, then multiply the resulting concentration by 2 to obtain the
test result.
The following table gives quick reference guidelines on dilutions of various
proportions. All dilutions are based on a 10 mL volume, so several dilutions
will require small volumes of the water sample. Graduated pipets should be
used for all dilutions.
Size of Sample
10 mL0 mL1
5mL5mL2
2.5 mL7.5 mL4
1 mL9 mL10
0.5 mL9.5 mL20
Deionized W ater to Bring
Volume to 10 mLMultiplication Factor
If the above glassware is not available, dilutions can be made with the
colorimeter tube. Fill the tube to the 10 mL line with the sample then transfer
it to another container. Add 10 mL volumes of demineralized water to the
container and mix. T ransfer back 10 mL of the diluted sample to the tube and
follow the test procedure. Continue diluting and testing until a reading, which
is in the concentration range for the test, is obtained. Be sure to multiply the
concentration found by the dilution factor (the number of total 10 mL volumes
used).
Example:
10 mL of sample is diluted with three 10 mL volumes of demineralized water;
the dilution factor is four.
16COD PLUS COLORIMETER 6/02
INTERFERENCES
n
LaMotte reagent systems are designed to minimize most common interferences.
Each individual test instruction discusses interferences unique to that test. Be
aware of possible interferences in the water being tested.
The reagent systems also contain buffers to adjust the water sample to the ideal
pH for the reaction. It is possible that the buffer capacity of the water sample
may exceed the buffer capacity of the reagent system and the ideal pH will not
be obtained. If this is suspected, measure the pH of a reacted distilled water
reagent blank using a pH meter. This is the ideal pH for the test. Measure the
pH of a reacted water sample using the pH meter. If the pH is significantly
different from the ideal value, the pH of the sample should be adjusted before
testing.
Interferences due to high concentration of the substance being tested, can be
overcome by sample dilution (p. 16).
STRAY LIGHT INTERFERENCE
n
When scanning samples in 16 mm tubes, such as COD, the sample chamber lid
can not be closed. The COD adapter minimizes stray light. To further reduce
stray light interference, do not scan sample in direct sunlight.
COD PLUS COLORIMETER 6/0217
OPERA TION OF THE
COD PLUS COLORIMETER
OVERVIEW
n
The COD PLUS Colorimeter is a portable, microprocessor controlled, direct
reading colorimeter. It has a graphical 4 line, 16 character liquid crystal display
for graphical, alphabetical and numerical messages. The operation is controlled
with the keypad through menu driven software in response to selections shown
on the display.
The test library consists of 29 LaMotte tests and 10 “User Tests”. The LaMotte
tests are precalibrated for LaMotte reagent systems. The colorimeter displays
the results of these tests directly in units of concentration. The 10 “User Tests”
may be used to enter additional calibrations. All of these tests may be arranged
in any of 3 sequences. These sequences can be modified a limitless number of
times to meet changing testing needs.
The optics feature 2 different colored LEDs. Each LED has a corresponding
silicon photodiode with an integrated interference filter. The interference
filters select a narrow band of light from the corresponding LED for the
colorimetric measurements. The microprocessor automatically selects the
correct LED/photodiode combination for a test.
A RS-232 serial port on the back of the colorimeter, and optional software,
allows the COD PLUS to be interfaced with an IBM compatible personal
computer for real time data acquisition and data storage. This port also allows
an interface with a RS-232 serial printer.
Due to its portability, alternate power sources, and rugged construction, the
COD PLUS Colorimeter is ideal for lab and field use.
POWER SOURCE
n
The COD PLUS Colorimeter uses a 6V 500 mA AC adapter. Please refer to
the Parts List for the code number for the correct adapter.
USE OF ANY AC ADAPTER OTHER THAN THE ONE SPECIFIED
FOR USE WITH THE COD PLUS COLORIMETER MAY DAMAGE
THE METER AND WILL VOID THE WARRANTY.
To use the adapter, slide the connector pin from the AC adapter into the small
hole on the left side of the meter. Plug the AC adapter into an appropriate wall
socket or power source.
18COD PLUS COLORIMETER 6/02
COMPONENTS
n
Figure 1 shows a diagram of the COD PLUS Colorimeter and its components.
Top View
COD PLUS
••••••••••••••••••
ENTER
OFFEXIT
Lid
*
ON
Side Views
Bottom View
Serial
Number
Battery
Compartment
Rs232
Serial Port
AC
Adapter Socket
Figure 1
COD PLUS COLORIMETER 6/0219
QUICK START
n
Some quick instructions to get into testing.
1. Press ON to turn on the COD PLUS. The
LaMotte logo screen will appear for about 2
seconds and then the Start screen appears. Press
Q/ENTER to start testing.
2. The Main Menu will appear. Press
Q/ENTER to select TESTING MENU.
3. Press Q/ENTER to select All Tests.
4. Press t or s to move the * to the desired
test.
5. Press Q/ENTER to select test.
VER 1.0
COD PLUS
* Start
MAIN MENU
* Testing Menu
Editing Menu
PC Link
TESTING MENU
* All Tests
Sequence 1
Sequence 2
ALL TESTS
* 001 COD LR
002 COD SR
003 COD HR
ALL TESTS
* 002 COD SR
003 COD HR
004 Ammonia-N LF
6. Insert blank, press Q/ENTER to scan blank.
7. The screen will display Blank Done for about
1 second.
20COD PLUS COLORIMETER 6/02
002 COD SR
* Scan Blank
002 COD SR
Blank Done
* Scan Blank
8. Insert the reacted sample. Press Q/ENTER to
scan sample. The COD PLUS will scan the
sample and display the concentration.
002 COD SR
* Scan Sample
9. After recording test result, scroll with t or s
and make another selection with Q/ENTER.
Press EXIT to escape to previous menus.
002 COD SR
722 mgL
* Scan Sample
COD PLUS COLORIMETER 6/0221
GENERAL OPERATING PROCEDURES
The operation of the COD PLUS Colorimeter is controlled by a
microprocessor. The microprocessor is programmed with menu driven software.
A menu is a list of choices. This allows a selection of various tasks for the
colorimeter to perform, such as, scan blank, scan sample, and edit test
sequences. The keypad is used to make menu selections which are viewed in
the display. There are three selections accessible from the MAIN MENU:
Testing Menu, Editing Menu and PC Link.
THE KEYPAD
n
The keypad has 6 buttons which are used to perform specific tasks.
ON
tThis button will cause the display to scroll down through a list of
sThis button will cause the display to scroll up in a list of menu
ENTER
Q
EXIT
OFF
SAMPLE HOLDERS
n
The sample chamber is designed for 25 mm round tubes, and a sample holder
for 16 mm COD tubes is included. A 1 cm square UDV cuvette sample holder
is available for the COD PLUS Colorimeter.
Position the COD adapter in the SMART 2 chamber so that the grooves in the
adapter are aligned with the ridges located at the rear of the chamber. The
adapter should be inserted with the small hole, containing the ball plunger, at
the top. The ball plunger can be adjusted with a small screwdriver to control
the tightness of the fit of the tube in the adapter.
This button is used to turn the colorimeter on.
menu choices. It will move through a list viewed in the display . It
will auto scroll when held down.
choices. It will move through a list viewed in the display. It will
auto scroll when held down.
This button is used to select the menu choice adjacent to the “*”in
a menu viewed in the display.
This button is an exit or escape button. When pressed, the display
will exit from the current menu and go to the previous menu.
This button turns the colorimeter off.
22COD PLUS COLORIMETER 8/02
THE DISPLAY & THE MENUS
n
The display allows menu selections to be viewed and chosen. These choices
instruct the colorimeter to perform specific tasks. The menus are viewed in the
display using two general formats which are followed from one menu to the
next. Each menu is a list of choices or selections.
There are four lines in the display. The top line in each menu is a title or
pertinent instruction. The top line does not change unless a new menu is
selected. The second and third lines are used in two ways. One way is to display
menu choices. The second way takes advantage of the graphical capabilities of
the display. Both lines are used to display important messages, such as test
results, in a large, easy to read format. The fourth line is used for menu choices.
DISPLAY
TESTING MENU
* FIRST CHOICE
SECOND CHOICE
ANOTHER
AND ANOTHER
AND SO ON
TITLE or INSTRUCTION
MENU CHOICE WINDOW
Think of the menu choices as a vertical list in the display which moves up or
down each time an arrow button is pressed. This list or menu is viewed through
a window, the menu choice window, in the display. The menu choice window
is the lower 2 or 3 lines of the display. Pushing the arrow buttons brings
another portion of the menu into menu choice window. This is referred to as
scrolling through the menu.
TESTING MENU
* FIRST CHOICESECOND CHOICEANOTHER
SECOND CHOICE* ANOTHERAND ANOTHER
ANOTHERAND ANOTHER* AND SO ON
AND ANOTHERAND SO ONLAST CHOICE
AND SO ONLAST CHOICE
LAST CHOICE
t
TESTING MENU
t
TESTING MENU
An asterisk, “*”, will start in the far left position of the top line in the menu
choice window. As the menu is scrolled through, different choices appear next
to the “*”. The “*” in the display corresponds with the Q/ENTER button.
Pushing the Q/ENTER button selects the menu choice which is adjacent to
the “*” in the menu choice window.
COD PLUS COLORIMETER 6/0223
The second general format of the display takes advantage of the graphics
capabilities of the display. The top line of the display is still a title line. The
middle two lines of the display are used to display important messages, results or
graphics in a large, easy to read format. The menus work in the same way as
described previously but only one line of the menu is visible at the bottom of
the display.
TESTING MENU
TESTING MENU
t
TESTING MENU
t
Result or MessageResult or MessageResult or Message
* ANOTHER* AND ANOTHER* AND SO ON
AND ANOTHERAND SO ONLAST CHOICE
AND SO ONLAST CHOICE
LAST CHOICE
As described previously, the EXIT button allows an exit or escape from the
current menu and a return to the previous menu. This allows a rapid exit from
an inner menu to the main menu by repeatedly pushing the
Pushing
OFF at any time will turn the colorimeter off.
EXIT button.
24COD PLUS COLORIMETER 6/02
LOOPING MENUS
n
Long menus, such as All Tests, incorporate a looping feature which allow
the user to quickly reach the last choice in the menu from the first choice. In a
looping menu the last choices in the menu are above the first choice and
scrolling upward moves through the menu in reverse order. Scrolling downward
moves through the menu from first choice to last but the menu starts over
following the last choice. So all menu choices can be reached by scrolling in
either direction. The diagrams below demonstrate a looping menu.
AND SO ONAND ANOTHERANOTHER
: : :AND SO ONAND ANOTHER
: : :: : :AND SON ON
THIRD TO LAST: : :: : :
SECOND TO LASTTHIRD TO LAST: : :
LAST CHOICESECOND TO LASTTHIRD TO LAST
TESTING MENU
* FIRST CHOICE* LAST CHOICE* SECOND TO LAST
SECOND CHOICEFIRST CHOICELAST CHOICE
ANOTHERSECOND CHOICEFIRST CHOICE
AND ANOTHERANOTHERSECOND CHOICE
AND SO ONAND ANOTHERANOTHER
: : :AND SO ONAND ANOTHER
: : :: : :AND SO ON
LAST CHOICE: : :: : :
TESTING MENU
s
TESTING MENU
s
COD PLUS COLORIMETER 6/0225
TESTING
TESTING MENU
n
The Testing Menu is used to run all LaMotte pre-programmed tests, USER
TESTS and Absorbance test at two wavelengths. Testing from any of three
sequences can also be done.
1. Press the ON button to turn on the
COD PLUS Colorimeter. The LaMotte logo will
appear for about 2 seconds and the the Start screen
appears. Press the Q/ENTER button to begin testing.
2. The MAIN MENU will appear. Press the
Q/ENTER button to select Testing Menu.
3. Scroll with the t or s buttons and make
a selection with the Q/ENTER button. AllTests has all the available tests. The three
sequences have selected tests and
Absorbance has %T/ABS tests.
VER 1.0
COD PLUS
* Start
MAIN MENU
* Testing Menu
Editing Menu
PC Link
TESTING MENU
* All Tests
Sequence 1
Sequence 2
Sequence 3
Absorbance
26COD PLUS COLORIMETER 6/02
SEQUENCES OF TESTS
n
SEQUENCE 1, SEQUENCE 2, and SEQUENCE 3 are alterable sequences. They
may be edited using the Editing Menu. Any of the LaMotte
pre-programmed tests or User T ests may be placed in these sequences in
whatever testing order that is preferred. Some examples of typical sequences are
given below.
SEQUENCE 1SEQUENCE 2SEQUENCE 3
* 004 Ammonia-N LF* 008 Cobalt* 001 COD LR
005 Ammonia-N LS010 Cu Cuprizone002 COD SR
006 Ammonia-N H011 Cu - DDC003 COD HR
007 Boron015 Hydrazine
001 COD LR016 Moly HR
002 COD SR017 Nickel
003 COD HR018 Ozone-LR
009 Color021 Phosphate-H
013 DO022 Potassium
014 Fluoride023 Silica-Lo
018 Ozone-LR024 Silica-Hi
020 Phosphate-L027 Tannin
025 Sulfate-HR029 Zinc-LR
026 Sulfide-LR
028 Turbidity
These alterable sequences allow a series of tests to be setup that are run
frequently. The order of the individual tests in the sequence is determined by
the user. After running a test, use the t button to scroll to the next test and
press the Q/ENTER button to select the next test in the sequence. Continue
this pattern until the entire sequence has been completed.
All Tests is a fixed sequence containing the LaMotte pre-programmed tests,
User Tests, and Absorbance tests.
Modification of the alterable sequences is accomplished through the EditingMenu. This menu is explained in greater detail in EDITING MENU (p. 32).
Pressing the EXIT button while in a sequence menu will escape back to the
Testing Menu.
Pressing the OFF button at any time will turn the colorimeter off.
COD PLUS COLORIMETER 6/0227
GENERAL TESTING PROCEDURES
n
The following are some step by step examples of how to run tests from the
Testing Menu. These test procedures are designed to be used with LaMotte
SMART Reagent Systems.
TESTING WITH THE LaMOTTE
n
PRE-PROGRAMMED TESTS
Press ON to turn on the COD PLUS Colorimeter.
The LaMotte logo will appear for about 2 seconds
and then the Start screen appears. Press the
Q/ENTER button to start testing.
The MAIN MENU will appear. Press the Q/ENTER
button to select Testing Menu.
Press the Q/ENTER button to select All Tests.
Press the t button to move to the 001 COD LR to
*.
VER 1.0
COD PLUS
* Start
MAIN MENU
* Testing Menu
Editing Menu
PC Link
TESTING MENU
* All Tests
Sequence 1
Sequence 2
ALL TESTS
* 001 COD LR
002 COD SR
003 COD HR
Press the Q/ENTER button to select 001 COD LR.
28COD PLUS COLORIMETER 6/02
ALL TESTS
* 001 COD LR
002 COD SR
003 COD HR
The COD PLUS Colorimeter is ready to scan at the
correct wavelength. Place the blank in the sample
chamber, close the lid and press the Q/ENTER
button to scan blank.
NOTE: Do not keep the button depressed.
001 COD LR
* Scan Blank
The screen will display Blank Done for about 1
second. Scan Sample will be positioned next to *.
Place the reacted sample in the chamber, close the
lid and press the Q/ENTER button to scan sample.
The colorimeter will scan the sample and the results
screen will appear.
Record test result. To repeat the test, press the
Q/ENTER button to scan the sample again. The
last blank scanned is used to zero the colorimeter for
repeated scans. A different blank can be used by
pressing the s button to scroll back to ScanBlank and then scanning another blank. Scroll with
the t or s buttons and make another selection
with the Q/ENTER button. The %T or Absorbance
of the last test can be viewed by choosing %T/Abs.
Press the EXIT button to escape to previous menus.
NOTE: The menus loop in this screen so either the
s or t buttons will lead to the menu selection
needed.
001 COD LR
Blank Done
* Scan Blank
001 COD LR
* Scan Sample
001 COD LR
100 mgL
* Scan Blank
Next Test
Previous Test
%/Abs
Scan Blank
COD PLUS COLORIMETER 6/0229
CALIBRATING LaMOTTE PRE-PROGRAMMED TESTS
n
The LaMotte Pre-Programmed T ests have been pre-calibrated. Recalibration of
the pre-programmed tests by the user is not possible. However, a procedure to
standardize the calibration can be performed to obtain the most accurate
readings or to meet regulatory requirements.
The LaMotte Pre-Programmed tests are standardized with one standard
solution. To standardize over the full range of the test, the concentration of the
standard should be chosen from the high end of the range. Alternatively , if
samples do not cover the full range of the test, a standard should be chosen that
is close to the concentration of the samples.
The standardization procedure should be followed as often as required by
regulations and laws for compliance monitoring.
In the example below the Aluminum calibration will be standardized.
Prepare a standard solution to be tested. Use 0.10 ppm aluminum.
Use the s or t button to scroll to 002 Aluminum.
Follow instructions in the COD PLUS Manual for
testing the aluminum standard. Scan the blank.
The screen will display Blank Done for about 1
second. Scan Sample will be positioned next to *.
Place the reacted sample in the chamber, close the
lid and press Q/ENTER to scan sample. The result
will be displayed.
The displayed result can now be standardized. Use
the s or t buttons to scroll to Calibrate. Press
Q/ENTER to select.
002 Aluminum
* Scan Blank
002 Aluminum
Blank Done
* Scan Sample
002 Aluminum
* Scan Sample
002 Aluminum
0.09 ppm
* Scan Sample
Next Test
Previous Test
%T/Abs
Calibrate
Scan Blank
30COD PLUS COLORIMETER 4/03
Loading...
+ 103 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.