LaMotte 1766 User manual

Page 1
Tracer pH/TDS/Salt
TM
PockeTester
Code 1766
Do Not allow pH sensor to dry out.
¬ Store Upright
¬ Before fi rst use, tap meter against
See pages 10 and 19.
pH/Conductivity/TDS/Salinity Meter!
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
TRACER
pH/TDS/SALT POCKETESTER
Introduction ........................................................................................4
Specifications ................................................................................5
Contents .........................................................................................6
Parts & Accessories ...........................................................................5
Meter Description
Front Panel Description .................................................................. 7
TRACER Display ............................................................................. 7
Basic Operation
Powering the TRACER .................................................................... 8
Electrode Recognition .................................................................... 8
Automatic Calibration ..................................................................... 8
TDS Conversion Ratio .................................................................... 8
Changing the Display Temperature Units ...................................... 9
Data Hold ........................................................................................ 9
Auto-Power Off ............................................................................... 9
Low Battery Indicator ..................................................................... 9
Measurement and Display............................................................. 10
Technique ....................................................................................... 10
Testing
Getting Started ............................................................................... 11
Changing the Measurement Function .......................................... 11
Measurement ................................................................................. 12
Measuring the TDS of Soil ............................................................ 13
Storing Readings ........................................................................... 14
Recalling Stored Readings ............................................................ 14
Clearing the Memory ..................................................................... 14
Calibration
pH ................................................................................................... 15
Conductivity ................................................................................... 16
Operational Matrix ............................................................................. 18
Maintenance
Storage ........................................................................................... 20
Battery Replacement ..................................................................... 20
Electrode Care ............................................................................... 20
Replacing the Electrode ................................................................ 21
Electrode Cleaning Recommendations ........................................ 21
Troubleshooting ................................................................................ 22
Warranty .............................................................................................23
TM
CODE 1766
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 3
INTRODUCTION
Congratulations on your purchase of the pH/TDS/SALT (pH/ conductivity/TDS/Salinity) TRACER PockeTester. The TRACER is a revolutionary, fi rst of its kind measurement device that offers direct reading of pH, conductivity, TDS and salinity with one electrode. Careful use and maintenance will provide years of reliable service.
4 LaMotte TRACER pH/TDS/Salt PockeTesterTM 1766 Instruction
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 4
SPECIFICATIONS
Display 2000 count LCD with Bar Graph
pH Range 0.00 to 14.00
pH Accuracy ± 0.01 pH typical
pH ATC Range 32 to 194 °F (0 to 90 °C)
pH Measurement Storage 25 tagged (numbered) readings
pH Reference Junction
Conductivity Range 0 to 199.9 μS/cm
TDS Range 0 to 99.9 ppm and mg/L
Salinity Range 0 to 99.9 ppm S
TDS Ratio 0.4 to 1.0, adjustable
Salinity Ratio 0.5 fi xed
Conductivity ATC 2.0% per °C
Conductivity ATC Range 32.0 to 140.0 °F (0.0 to 60.0 °C)
Temperature Range 23.0°F to 194°F (–5.0°C to 90.0°C)
Temperature Resolution 0.1 up to 99.9, 1 >100
Temperature Accuracy ±1.8°F; 1°C; (from 23 to 122°F; -5 to 50°C)
Conductivity ATC Range 32.0°F to 140°F (0.0°C to 60.0°C)
Accuracy Conductivity: ±2% FS
Measurement Storage 25 numbered readings
Low Battery Indication ‘BAT’ appears on the LCD
Power Four CR2032 Lithium Ion batteries
Auto Power Off After 10 minutes of no button presses
Operating Conditions 23 to 122 °F (–5 to 50 °C)
Permanent gel, non-refi llable
200 to 1999 μS/cm
2.00 to 19.99 μS/cm
100 to 999 ppm and mg/L
1.00 to 9.99 ppt and g/L (variable ratio)
100 to 999 ppm S
1.00 to 9.99 ppt S (fi xed ratio of 0.5)
±5.4°F; 3°C; (from 122 to 194°F; 50 to 90°C)
TDS: ±2% FS Salinity: ±2% FS
(override available)
NOTE: 1 part per thousand (ppt) equals 1000 parts per million (ppm). Example: 3.1ppt=3,100 ppm
LaMotte TRACER pH/TDS/Salt PockeTesterTM 1766 Instruction 5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 5
CONTENTS
pH/CON TRACER PockeTester Kit Code 1766 Includes: Sample Cup w/cap * Tablet Crusher Code 0175 Buffer Tablets, pH 4.0, 7.0 & 10.0 * *Not sold in this quantity. See below.
PARTS & ACCESSORIES
pH/TDS/SALT Replacement Electrode Code 1755 Weighted Stand w/Sample Cups (5) Code 1746 Sample Cups w/caps (24) Code 1745 pH 4.0 Mini Buffer Tablets (100) Code 3983A-J pH 7.0 Mini Buffer Tablets (100) Code 3984A-J pH 10.0 Mini Buffer Tablets (100) Code 3985A-J Conductivity Standard, 84 μS, 30 mL, 500 mL Code 6312-G, L Conductivity Standard, 1413 μS, 30 mL, 500 mL Code 6354-G, L Conductivity Standard, 12,880 μS, 30 mL, 500 mL Code 6317-G, L
6 LaMotte TRACER pH/TDS/Salt PockeTesterTM 1766 Instruction
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 6
METER DESCRIPTION
Front Panel Description
1. Battery compartment cap
2. LCD Display
3. MODE/HOLD button - change mode, hold data, store data
4. CAL/RECALL button ­calibration, change temperature units, recall data
5. ON/OFF button
6. Electrode Collar
7. Electrode
NOTE: The Electrode cap is not shown.
TRACER DISPLAY
1. Bar graph display
2. Measurement units
3. Main display
4. Range calibration and battery indicators
1 2
5. Temperature display
6. Renew indicator
7. HOLD indicator
6 7
LaMotte TRACER pH/TDS/Salt PockeTesterTM 1766 Instruction 7
5
10
7
0 0
m
t
mg/L
S
pp
2
μ
m
88.88
BAT pH
HML
RENEW
-8.8.8
HOLD
10 20 12
3 4
°C
5
°F
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 7
BASIC OPERATION
Powering the TRACER
The Tracer uses four CR 2032 Lithium Ion batteries. If the batteries are weak, the BAT indicator will appear on the display. Press the ON/OFF button to turn the TRACER on or off. The auto power off feature will shut the TRACER off automatically after 10 minutes after the last button push.
Automatic Calibration
When the TRACER is turned on, it will enter the Automatic Calibration mode. SELF and CAL will appear while the calibration is in progress. After the calibration is completed, the SELF and CAL display icons will extinguish.
TDS Conversion Ratio
The TDS value is determined by multiplying the conductivity measurement by a known conversion ratio factor. The meter allows the selection of a conversion ratio factor in the range of 0.4 to 1.0. The selected ratio will vary with application but is typically set between 0.5 and 0.7. In the salinity mode, the ratio is fi xed at 0.5. The stored ratio factor will briefl y appear in the lower temperature display when the meter is fi rst turned on or when changing the measurement function to TDS.
To change the ratio while in the TDS measurement mode:
1. Turn the TRACER on.
2. Press and release the CAL/RECALL button twice. The stored ratio will appear in the display.
3. Press the MODE/HOLD button to
10
0
20
change the ratio value in steps of 0.1.
4. When the desired ratio is displayed, press and release the CAL/RECALL
0.7
L
button to store the value and return to the normal mode.
5. If no buttons are pressed for 5 seconds, the meter will return to the measurement mode.
8 LaMotte TRACER pH/TDS/Salt PockeTesterTM 1766 Instruction
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 8
Changing the Displayed Temperature Units
To change the displayed temperature units between °C or °F:
1. With the TRACER off, press and hold the CAL/RECALL button.
2. With the CAL/RECALL button pressed, momentarily press the ON/ OFF button. When SELF CAL appears in the display, release the CAL/RECALL button. The TRACER will return to the operational mode with the temperature displayed in the new units.
Data Hold
Press the MODE/HOLD button to freeze the current reading. The HOLD icon will appear. The reading will be stored. Press the MODE/HOLD key to return to normal operation.
Auto-Power Off
The auto-power off feature will automatically shut the meter off 10 minutes after the last button was pressed. To disable the auto-off feature:
1. Press the ON/OFF button to turn the meter on.
2. Wait for the SELF display screen to appear. This is the second screen to appear after turning on the meter. IMPORTANT: Review the instructions for the next 3 steps before proceeding. Step 3 must be followed immediately by Step 4.
3. Press CAL/RECALL button once.
4. Press MODE/HOLD and ON/OFF buttons simultaneously, for approximately 2 seconds.
5. oFF will be displayed on the screen. Watch carefully. It will disappear quickly.
6. To re-enable the auto-off feature, repeat steps 3 and 4. ON will be briefl y displayed on the screen.
7. The auto power off feature will be restored automatically when the meter is turned off. Auto-off is the default function when the meter is turned on.
Low Battery Indicator
The “BAT” indicator will be displayed when the batteries become weak. Refer to the Maintenance section for battery replacement information.
LaMotte TRACER pH/TDS/Salt PockeTesterTM 1766 Instruction 9
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 9
Measurement and Display
• If the unit appears to be locked (display frozen). It is possible that
the Data Hold mode has been inadvertently accessed by pressing the MODE button. (“HOLD” will be displayed in the bottom left of the display.) Press the MODE button again or turn the meter off and then on.
• For maximum accuracy, allow suffi cient time for the temperature of
the probe to reach the temperature of the sample before calibrating. This will be indicated by a stable temperature reading on the display.
Technique
• Do not touch the inner surfaces of the conductivity electrodes.
Touching the surface of the platinized electodes may damage and reduce the life of the probe.
• Store the electrode in the wetting cap with the sponge moistened with
pH 4.0 buffer solution.
• Always rinse the electrode in deionized or distilled water between
measurements to avoid cross contamination of the sample. Double rinsing is recommended when high accuracy is required.
• Periodically, accumulated salt deposits from the reference electrode
may build up in the storage cap, and should be rinsed away. These deposits could affect measured values of low conductivity samples.
• When measuring low conductivity samples, extra care is recommended
in rinsing the probe to avoid contamination of the sample with electrolyte from the pH reference electrode. This will only be a factor when the volume of the sample. (Example: Try a 200 to 500 mL sample.)
• When a 20mL sample cup is usedthe electrode should not be allowed
to sit in the sample for any longer than necessary, to avoid electrolyte leakage into the sample, raising the conductivity value.
10 LaMotte TRACER pH/TDS/Salt PockeTesterTM 1766 Instruction
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 10
TESTING
Getting Started
• Remove the cap from the bottom of the Tracer to expose the pH
electrode, reference junction and conductivity electrodes.
• Before the fi rst use or after storage, soak the electrode in tap water or
pH 4 buffer solution for about 10 minutes.
• Before fi rst use, hold the meter by the top battery compartment
and swiftly tap the back of the meter downward into your palm (not a hard surface). This assures that the internal electrolyte moves to the very tip of the electrode. The electrolyte should fi ll the circular junction window at the tip of the electrode.
• White KCL crystals may be present in the cap or on the electrode.
This is to be expected depending on the length of time in storage. These crystals will dissolve while soaking the electrode or they can be rinsed away with tap water.
• For best results calibrate with pH 7 buffer solution fi rst, then calibrate
with the buffer solution closest to the expected pH value of the solution or material to be tested.
• To preserve the pH electrode life keep the sponge in the protective cap
soaked with tap water or pH 4 buffer solution.
• For best results, calibrate for conductivity with a standard in the
expected range of the sample. For maximum accuracy calibrate from low conductivity value standards to high value standards.
• When the meter is calibrated for conductivity, salinity or TDS the
meter must be in the conductivity mode. See page 15. For storage see page 19.
Changing the Measurement Function
The meter can be set to measure Conductivity, pH, Salinity (ppm), TDS (ppm), or TDS (mg/L). To change the mode:
1. Turn the TRACER on.
2. Press and hold the MODE/HOLD button for 2 seconds. The display will begin to scroll through the units.
μS or mS (Conductivity) pH ppm S or ppt S (Salinity) ppm or ppt (TDS) mg/L (TDS)
LaMotte TRACER pH/TDS/Salt PockeTesterTM 1766 Instruction 11
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 11
NOTE: 1 part per thousand (ppt) equals 1000 parts per million (ppm). Example: 3.1 ppt = 3,100 ppm
3. Release the MODE/HOLD key when the desired mode is displayed.
NOTE: The “HOLD” function cannot be used when changing the measurement function. If “HOLD” is displayed in the lower left corner of the display, briefl y press the MODE/HOLD button to turn it off.
MEASUREMENT PROCEDURE
Measurement
1. For small samples fi ll a sample cup to the 20 mL line with the test sample. Sample depth must be greater than or equal to 1.5 inches. pH can also be measured by contact of the electrode tip with a wet surface. For larger samples, like pools, go to step 2
2. Press the ON/OFF button. (8888 and then SELF CAL will appear in the display during the initial diagnostics).
3. Press and hold the MODE/HOLD button to scroll to the desired measurement mode.
4. Immerse the TRACER electrode in the sample. Make sure the electrode is completely submersed.
5. Slowly stir the sample with the TRACER to remove air bubbles if in the Conductivity, TDS or Salinity mode.
6. If the TRACER is in the Conductivity, TDS, or Salinity mode, the meter will automatically auto-range to the proper range and then display the reading. The display will fl ash “0000” while auto-ranging.
7. If the TRACER is in the pH mode, the reading will fl ash until it has stabilized. This may take several seconds depending on the buffer capacity of the sample.
8. Rinse the electrode in distilled water. Replace the cap.
NOTE: 1 part per thousand (ppt) equals 1000 parts per million (ppm). Example: 3.1ppt=3,100 ppm
12 LaMotte TRACER pH/TDS/Salt PockeTesterTM 1766 Instruction
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 12
Conductivity
TDS (ppm)
TDS (mgL)
pH
Salinity
0
μS
100.9
25.0
1000
0
ppm
1000
0
790
°C
25.0
°C
mgL
790
25.0
1000
°C
2
7
790
pH
HML
78.6
12
0
1000
Sppm
790
°F
25.0
NOTE: When measuring samples with low conductivity, thoroughly rinse the probe with distilled or deionized water before placing it in the sample to avoid contamination of the sample with electrolyte from the pH reference electrode. Also increase the sample size to 200 to 500 mL to decrease the chance of contamination. If a sample cup is used, do not allow the probe to sit in the sample for any longer than necessary to avoid electrolyte leakage into the sample. Leakage will raise the conductivity level.
Measuring the TDS of Soil
1. Fill a 50 mL beaker with the soil sample. Tap the beaker lightly on a hard surface to remove trapped air. Remove excess soil from the surface.
2. Empty the soil into a 250 mL wide-mouth fl ask.
3. Add 100 mL of distilled water. Stopper and shake vigorously.
4. Wait 30 minute. (Shake the fl ask vigorously three or four times during this period.)
5. Filter the contents of the fl ask. Collect the fi ltrate in a beaker.
6. Rinse the electrode with distilled or deionized water to remove impurities.
7. Press the ON/OFF button to turn the TRACER on. Make sure the meter is in the TDS mode.
8. Immerse the electrode in the fi ltrate. Make sure the tip of the electrode is completely immersed.
9. Stir the fi ltrate with the electrode to create a homogeneous solution.
10. Gently stir the fi ltrate with the electrode. Read the TDS value of the ltrate from the display.
11. Rinse the electrode in distilled water. Replace the cap.
°C
LaMotte TRACER pH/TDS/Salt PockeTesterTM 1766 Instruction 13
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 13
Storing Readings
1. After the reading is displayed press and hold the MODE/HOLD button to store the current reading. The meter will enter the HOLD mode and HOLD will be displayed. The storage location number (1 through 25) will be displayed on the lower display followed by the reading being stored.
2
1563
L
HOLD
10
20
μS
4
2. Press the MODE/HOLD button to exit the HOLD mode and return to normal operation.
3. If an attempt is made to store more than 25 readings, the stored readings will be overwritten starting with the fi rst reading.
Recalling Stored Readings
NOTE: First ensure that the HOLD symbol is not displayed. If it is, exit the HOLD function by pressing the MODE/HOLD button.
1. Press the CAL/RECALL button and then press the MODE/HOLD button. The location number (1 through 25) will briefl y appear and then the value stored in that location will appear. The displayed units will fl ash, indicating that the storage recall mode is active.
10
2
20
15
2
1423
L
10
μS
72.3
20
°F
2. The last stored reading taken will be displayed fi rst. To advance to the previously stored readings, press the MODE/HOLD button. The location number is displayed fi rst, followed by the reading stored in that location.
3. To exit the storage mode, press the CAL/RECALL button and the TRACER will return to normal operation after displaying “End”.
Clearing Stored Memory
Turn the TRACER on. Press and hold the ON/OFF button for 4 seconds. The display will briefl y display “clr” when the memory is cleared.
14 LaMotte TRACER pH/TDS/Salt PockeTesterTM 1766 Instruction
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 14
CALIBRATION
For the most accurate results, allow suffi cient time for the temperature of the probe to reach the temperature of the sample before calibrating. This will be indicated by a stable temperature reading on the display.
pH
The TRACER can be calibrated at 1, 2 or 3 points. For the most accurate results with a two point calibration, calibrate the TRACER with a pH 7 buffer fi rst, then calibrate with either a pH 4 or pH 10 buffer whichever is closest to the pH value of the sample to be tested. When performing a three point calibration, calibrate with the pH 7 buffer fi rst, followed with the pH 4 buffer and then the pH 10 buffer.
Preparation of Buffers
1. Fill a sample cup with 20 mL of distilled or deionized water.
2. Add one buffer tablet: pH 4.0 Code 3983A
pH 7.0 Code 3984A pH 10.0 Code 3985A
3. Use the tablet crusher (0175) to crush the tablet. Stir until the tablet has disintegrated.
NOTE: Buffers should be prepared fresh daily.
Calibration
1. Fill a sample cup to the 20 mL line with a buffer solution.
2. Press the ON/OFF button to turn the TRACER ON.
3. Place the electrode in the buffer solution. Press and hold the CAL/ RECALL button until “CAL” appears in the lower display. The meter will automatically recognize the buffer and calibrate itself to that value. The circled number on the display will match the value of the buffer.
NOTE: If the buffer solution is more than 1 pH unit off from 4, 7, or 10, or the electrode slope is low, the meter will assume that there is an error and the calibration will be terminated. END will be displayed.
4. During the calibration the pH value on the display will fl ash. When the calibration is complete, the TRACER will display “SA” and “End” and return to normal operation.
LaMotte TRACER pH/TDS/Salt PockeTesterTM 1766 Instruction 15
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 15
5. The appropriately circled indicator (L, M or H) will appear on the display when a calibration has been completed within one power on cycle.
6. For a two or three point calibration, repeat steps 1-5 with the remaining buffers.
7. When the TRACER is turned off, the circled indicator confi guration and the calibration data will be memorized.
Calibration Reminder
The “CAL” indicator will appear when the TRACER is in the pH mode and a calibration is required. The “CAL” indicator will appear if the meter has not been calibrated after 15 on/off cycles of the meter. Some applications may require more frequent calibrations than others. The “CAL” indicator is only a reminder and will not affect function in any way. The indicator will no longer be displayed when the pH electrode is recalibrated.
RENEW Indicator
A fl ashing “RENEW” indicator will appear on the display to warn that the electrode is not performing to expected specifi cations. If cleaning or recalibration does not cause the RENEW indicator to disappear, the electrode should be replaced. The RENEW indicator will appear when the pH electrode slope falls below 70% of a nominal slope.
Conductivity
Meter accuracy verifi cation should be performed on a periodic basis as needed. If calibration is required, the meter must be in the
conductivity mode to perform all calibrations for conductivity, TDS and salinity. The meter can perform calibrations and store the
data for each of the three ranges–low, medium and high. The automatic calibration recognition procedure will recognize conductivity standards of 84μS (Low), 1413μS (Medium) or 12,880μS (12.88mS) (High). (See Page 6). Always calibrate in the range closest to the expected measurement value. For salinity samples within the range of 1.00 to
9.99 ppt salinity, calibrate with a 12,880μS calibration standard.
1. Fill a sample cup to 20 mL line with a conductivity standard. NOTE: The meter allows for a 1, 2, or 3 point calibration. If
calibration is done for more than one point, the lowest concentration should be done fi rst to obtain the best accuracy. Calibrate the ranges from low to high
16 LaMotte TRACER pH/TDS/Salt PockeTesterTM 1766 Instruction
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 16
2. Press the ON/OFF button to turn the TRACER on. Insert the electrode into the standard. Tap or stir the sample with the Tracer to dislodge air bubbles.
3. Press and hold the CAL/RECALL button for approximately 2 seconds until the display begins to fl ash.
4. The meter will automatically recognize and calibrate to the conductivity standard. The display will briefl y indicate “SA” and “End” and then return to the measurement mode.
NOTE: “SA” will not appear if the calibration fails.
5. The calibration range indicator will appear on the display for each range that is calibrated during a power on cycle.
L
Low Range, 84 μS/cm
M
Medium Range, 1413 μS/cm
H
High Range, 12.88 mS/cm (12,880 μS/cm)
NOTE: Each time the calibration mode is entered all calibration range indicators will be cleared, but only the calibration data for the currently selected range will be replaced. The calibrations for the other two ranges will be saved even though the indicators for those ranges are no longer displayed. Calibration of all three ranges must be performed during one power on period for all three calibration range indicators to be displayed.
NOTE: As with all combination pH/conductivity probes, at low conductivity levels the fl ow of the pH electrolyte may affect the readings and result in artifi cially high conductivity readings. To eliminate the interference, low level conductivity or TDS measurements should be taken with a meter that measures conductivity only.
LaMotte TRACER pH/TDS/Salt PockeTesterTM 1766 Instruction 17
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 17
OPERATIONAL MATRIX
Function/ Resulting Action Power Mode Key Press Sequence
On/Off On or Off Any Momentary press of ON/
Calibration On pH or Con Press & hold CAL/
Store Reading On Any Momentary press of
Hold Release On Hold Momentary press of
Enter Memory Retrieval
Scroll Stored Readings
Exit Memory Retrieval
Clear Stored Memory
Change Measurement Mode
Enter CON/ TDS Ratio
Change CON/ TDS Ratio
Exit CON/TDS Ratio
On Any
Measurement Mode
On Memory Recall Momentary press of
On Memory Recall Momentary press of CAL/
On Any Memory
Mode
On Any Press and hold the MODE/
On TDS (ppm or
mg/L)
On TDS (ppm or
mg/L)
On TDS (ppm or
mg/L)
OFF button
RECALL button for 2 seconds until CAL is displayed
MODE/HOLD button
MODE/HOLD button
Momentary press of CAL/ RECALL button followed by a momentary press of MODE/HOLD button within 4 seconds.
MODE/HOLD button. Displays last in fi rst out.
RECALL button
Press and hold the ON/OFF button for 4 seconds until “clr” is displayed.
HOLD button for at least 2 seconds Modes will scroll by until button is released
Press and release the CAL/ RECALL button twice in succession
Momentary press of MODE/HOLD button. Each press increases ratio by 0.1 from 0.4 to 1.0.
Momentary press of CAL/ RECALL button.
18 LaMotte TRACER pH/TDS/Salt PockeTesterTM 1766 Instruction
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 18
Function/ Resulting Action Power Mode Key Press Sequence
Change Temperature Units
Override Auto Power Off
Default Reset Off N/A Simultaneously press ON/
Off Off Press and hold CAL/
On Any
Measurement Mode
RECALL button then momentarily press ON/ OFF button. Release CAL/ RECALL button after “Self Cal” is displayed.
Momentarily press CAL/ RECALL button then simultaneously press and hold ON/OFF and MODE/HOLD buttons for 2 seconds until “oFF” is displayed.
OFF, CAL/RECALL and MODE/HOLD buttons momentarily. “dFlt” will be displayed.
LaMotte TRACER pH/TDS/Salt PockeTesterTM 1766 Instruction 19
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 19
MAINTENANCE
Storage
1. Rinse the electrode in distilled or deionized water.
2. Store the electrode with the cap on. Keep the sponge in the cap
soaked with tap water or pH 4 buffer solution. Do not allow sensor to dry out. Store upright.
3. Always rinse the electrode in deionized water between measurements to avoid cross contamination. Double rinsing is recommended when high accuracy is required.
4. Salt deposits may build up in the storage cap and should periodically be rinsed away. These deposits could affect measurements at low conductivity.
Battery Replacement
1. Twist off the battery compartment cap.
2. Hold the battery housing in place with one fi nger. Remove the battery carrier by pulling on the small tabs.
3. Replace the four CR2032 batteries. Observe polarity.
4. Replace the battery compartment cap.
Electrode Care
1. Always rinse the electrode in distilled or deionized water between measurements to avoid cross-contamination of the samples. Double rinsing is recommended when high accuracy is required.
2. Do not touch the electrodes. Touching the surface of the platinized electrodes may damage and reduce the life of the electrodes.
20 LaMotte TRACER pH/TDS/Salt PockeTesterTM 1766 Instruction
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 20
REPLACING THE ELECTRODE
1. Unscrew and remove the electrode collar. Turn collar counter­clockwise.
2. Gently rock the electrode side to side, while pulling it away from the meter, until it disconnects from the electrode socket.
3. To attach an electrode, align the slots and carefully plug the electrode into the meter socket.
4. Firmly tighten the electrode collar to create a seal with the rubber gasket between the electrode and the meter.
Electrode Cleaning Recommendations
Do not soak the electrode in the solutions for longer than the recommended length of time. To do so may cause a reference potential shift which will cause degradation in performance or failure. When cleaning the electrode, take care not to scratch or damage the sensing surface or the platinized electrode surfaces.
Contaminant Cleaning Solution Procedure
Water soluble Deionized water Soak or scrub with a soft brush.
Grease and oil Warm water and
Heavy grease Alcohol Soak for a maximum of
Lime and hydroxide coatings
household detergent
10% Acetic acid Soak until coating dissolves,
Recondition in pH 4 or 7 buffer for 1 hour.
Soak or scrub with a soft brush, maximum 10 minutes. Rinse thoroughly with DI water. Recondition in pH 4 or 7 buffer for 1 hour.
5minutes. Scrub with a soft brush. Rinse thoroughly with DI water. Recondition in pH 4 or 7 buffer for 1 hour.
maximum 5 minutes. Rinse thoroughly with DI water. Recondition in pH 4 or 7 buffer for 1 hour.
LaMotte TRACER pH/TDS/Salt PockeTesterTM 1766 Instruction 21
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 21
TROUBLESHOOTING
Problem Possible Cause Action
Reading is frozen
"BAT" message
Meter will not calibrate in pH
Meter will not calibrate in conductivity mode
Unit will not turn on
“RENEW” message
Unit will not respond to any key presses
Unit is in "HOLD" mode
Batteries are low Replace batteries.
Low pH slope Replace electrode.
Clogged or contaminated reference junction
Damaged or worn sensing membrane
Contaminated pH buffers
Trapped air bubbles Tap or stir to release air bubbles.
Dirty probe Clean conductivity probe.
Damaged conductivity probe
Contaminated conductivity standards
Batteries low or dead Replace batteries.
Battery installed with incorrect polarity
pH sensor needs recalibration
pH sensor slope has fallen below acceptable limits
Internal fault Perform hard reboot: Remove
Press MODE button to exit "HOLD" mode.
Clean junction. See cleaning instructions.
Replace electrode.
Use fresh buffers.
See cleaning instructions.
Replace electrode.
Use fresh standards.
Replace batteries, observe polarity.
Recalibrate unit. Use fresh buffers.
Replace electrode.
batteries, hold ON/OFF switch down for 5 seconds, replace batteries.
22 LaMotte TRACER pH/TDS/Salt PockeTesterTM 1766 Instruction
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 22
WARRANTY
LaMotte Company warrants this instrument to be free of defects in parts and workmanship for 1 year from the date of shipment and warrants this probe be free of defects in parts and workmanship for 6 months from the date of shipment. If it should become necessary to return the instrument for service during or beyond the warranty period, contact our Technical Service
shipping charges, freight, insurance and proper packaging to prevent damage in transit. This warranty does not apply to defects resulting from action of the user such as misuse, improper wiring, operation outside of specifi cation, improper maintenance or repair, or unauthorized modifi cation. LaMotte Company specifi cally disclaims any implied warranties or merchantability or fi tness for a specifi c purpose and will not be liable for any direct, indirect, incidental or consequential damages. LaMotte Company’s total liability is limited to repair or replacement of the product. The warranty set forth above is inclusive and no other warranty, whether written or oral, is expressed or implied.
Find Quality Products Online at: sales@GlobalTestSupply.com