Congratulations on your purchase of the Total Dissolved Solids/Salinity TRACER
PockeTester. The TRACER is a revolutionary, fi rst of its kind measurement
device that offers direct reading
use and maintenance
will provide years of reliable service.
of salinity and TDS with one electrode. Careful
SPECIFICATIONS
Display2000 count LCD with Bar Graph
Conductivity Range0 to 199.9 μS
200 to 1999 μS
2.00 to 19.99 mS
TDS Range0 to 9,990 ppm (variable ratio)
Salinity Range0 to 9,990 ppm Salt(fi xed ratio of 0.5)
TDS Ratio0.4 to 1.0, adjustable
Salinity Ratio0.5 fi xed
Conductivity ATC2.0% per °C
Temperature Range32.0 to 149.0 °F, 0.0 to 65.0 °C
Temperature Resolution0.1 up to 99.9, 1>100
Temperature Accuracy±1.8°F, 1°C (from 32 to 122°F, 0 to 50°C)± 5.4°F,
Conductivity ATC Range32.0°F to 140°F, (0.0°C to 60.0°C)
AccuracyConductivity: ± 2% FS
Measurement Storage25 numbered readings
Low Battery Indication
PowerFour CR 2032 Lithium Ion batteries
Auto Power OffAfter 10 minutes of no button presses
The Tracer uses four CR2032 Lithium Ion batteries. If the batteries are weak,
BAT indicator will appear on the display. Press the ON/OFF key to turn
the
the TRACER on or off. The auto power off feature will shut the TRACER off
automatically after 10 minutes after the last button push.
Automatic Calibration
When the TRACER is turned on, it will enter the Automatic Calibration
SELF and CAL will appear while the calibration is in progress. After the
mode.
calibration is completed, the
TDS Conversion Ratio
The TDS value is determined by multiplying the conductivity measurement
by a known conversion ratio factor. The meter allows the selection of a
conversion ratio factor in the range of 0.4 to 1.0. The selected ratio will vary
with application but is typically set between 0.5 and 0.7. In the salinity mode,
the ratio is fi xed at 0.5. The stored ratio factor will briefl y appear in the lower
temperature display when the meter is fi rst turned on or when changing the
measurement function to TDS.
To Change the TDS Conversion Ratio:
1. Turn the TRACER on.
2. Press and release the CAL/RECALL button twice.
The stored ratio will appear in the display.
3. Press the MODE/HOLD button to change the ratio
value in steps of 0.1.
4. When the desired ratio is displayed, press and release
the CAL/RECALL button to store the value and return
to the normal mode.
5. If no buttons are pressed for 5 seconds, the meter
will return to the measurement mode.
To change the displayed temperature units between °F or °C:
1. With the TRACER off, press and hold the CAL/RECALL button.
2. With the CAL/RECALL button pressed, momentarily press the ON/OFF
button. When
button. The TRACER will return to the operational mode with the temperature
displayed in the new units.
Data Hold
Press the MODE/HOLD button to freeze the current reading. The HOLD icon
will appear, the reading will be stored. Press the MODE/HOLD key to return
to normal operation.
Auto Power Off
The auto-power off feature will automatically shut the meter off 10 minutes
after the last button was pressed. To disable the auto-off feature:
1. Press the ON/OFF button to turn the meter on.
2. Wait for the
appear after turning on the meter. IMPORTANT: Review the instructions
for the next 3 steps before proceeding. Step 3 must be followed immediately by Step 4.
3. Press CAL/RECALL button once.
4. Press MODE/HOLD and ON/OFF buttons simultaneously. Quickly
release buttons.
SELF CAL appears in the display, release the CAL/RECALL
SELF display screen to appear. This is the second screen to
5. oFF will be displayed on the screen. Watch carefully. It will disappear
quickly.
6. The auto power off feature will be restored automatically when the meter
is turned off. Auto-off is the default function when the meter is turned on.
Low Battery Indicator
The “BAT” indicator will be displayed when the batteries become weak. Refer
to the maintenance section for battery replacement information
1. Remove the cap from the bottom of the TRACER to expose the electrode.
2. Before the fi rst use, rinse the electrode in deionized water and dry.
Changing the Measurement Function
The meter can be set to measure Conductivity, TDS (ppm), TDS (mg/L) or
Salinity (ppm). To change the mode:
1. Turn the TRACER on.
2. Press and hold the MODE/HOLD button for 2 seconds. The display will
begin to scroll through the units.
μS or mS (Conductivity)
ppm (TDS)
ppm (Salinity “S”)
3. Release the MODE/HOLD key when the desired mode is displayed.
4. Note that the “HOLD” function can not be on when changing functions.
1. Fill a sample cup to the 20 mL line with the test sample. Sample depth
must be greater than or equal to 1.5 inches.
2. Immerse the TRACER electrode in the sample. Make sure the electrode is
completely submersed.
3. Press the ON/OFF button. (8888 and then SELF CAL
will appear in the display during the initial diagnostics).
4. Press and hold the MODE/HOLD button to scroll to
the desired measurement mode.
5. Slowly stir the sample with the TRACER to remove
air bubbles.
6. The meter will autorange to the proper range and the
100.9
72.3
MODE
CAL
ON
OFF
SAL/EC/TDS
∞F
reading will be displayed.
7. Rinse the electrode in distilled water. Replace the cap.
Storing Readings
1. After the reading is displayed press and hold the MODE/HOLD button
to store the current reading. The meter will enter the HOLD mode and
HOLD will be displayed. The storage
location number will be displayed
on the lower display followed by the
reading being stored.
2. Press the MODE/HOLD button to exit
the HOLD mode and return to normal
operation.
3. If an attempt is made to store more than
25 readings,
the stored readings will
be overwritten starting with the fi rst
reading.
NOTE: First ensure that the HOLD symbol is not displayed. If it is, exit the HOLD
function by pressing the MODE/HOLD button.
1. Press the CAL/RECALL button and then press the MODE/HOLD button;
the location number (1 through 25) will briefl y appear and then the value
stored in that location will appear. The displayed units will fl ash, indicat-
ing that the storage recall mode is active.
10
2
20
15
2. The last stored reading taken will be displayed fi rst. To advance to the
previously stored readings, press the MODE/HOLD button. The location
number is displayed fi rst, followed by the reading stored in that location.
3. To exit the storage mode, press the CAL/RECALL button and the
TRACER will return to normal operation after displaying “End”.
Clearing the Stored Memory
Turn the TRACER on. Press and hold the ON/OFF button for 4 second the display
will briefl y display “clr” when the memory is cleared.
For the most accurate results, allow suffi cient time for the temperature of the
probe to reach the temperature of the sample before calibrating. This will be
indicated by a stable temperature reading on the display
Meter accuracy verifi cation should be performed on a periodic/monthly basis
as needed. The frequency of the verifi cation will depend on the storage and
maintenance conditions and amount of meter use. If calibration is required,
the meter must be in the salinity mode to perform calibration for salinity
and TDS. The meter can perform a calibration and store the data. The automatic
calibration recognition procedure will recognize a conductivity standard of
3,000 ppm (3ppt). Always calibrate in the range closest to the expected
measurement value. For salinity samples within the range of 1,000 to 9,999 ppm
salinity, calibrate with a 3,000 ppm calibration standard.
1. Fill a sample cup to 20 mL line with a calibration standard.
2. Press the ON/OFF button to turn the TRACER on. Insert the electrode
into the standard. Tap or stir the sample with the Tracer to dislodge air
bubbles.
3. Press and hold the CAL/RECALL button for approximately 2 seconds.
“CAL” will appear and the display will fl ash.
4. The meter will automatically recognize and calibrate to the calibration
standard. The display will briefl y indicate “SA” and “End” and then return
to the measurement mode.
NOTE: “SA” will not appear if the calibration fails.
5. The calibration range indicator will appear on the display for each range
that is calibrated during a power on cycle.
Medium Range, 3,000 ppm (3ppt)
M
NOTE: Each time the calibration mode is entered all calibration range indicator
will be cleared, but only the calibration data for the currently selected range will
be replaced.
Press and release
the CAL/RECALL
button twice in quick
succession.
Press and release
the CAL/RECALL
button twice in quick
succession. Momentary
press of MODE/HOLD
button. Each press
increases ratio by 0.1
from 0.4 to 1.0.
Momentary press of
CAL/RECALL button.
RECALL button then
momentarily press ON/
OFF button. Release
CAL/RECALL button
after“Self Cal” is
displayed.
CAL/RECALL button
then simultaneously
press and hold CAL/
RECALL and MODE/
HOLD buttons for 2
seconds until “oFF” is
displayed.
ON/OFF, CAL/
RECALL and MODE/
HOLD buttons
momentarily. “dFlt” will
be displayed.
1. Always rinse the electrode in distilled, tap or deionized water between
measurements to avoid cross contamination. Double rinsing is recommended
when high accuracy is required.
2. Store the electrode dry with the cap on.
3. Do not touch the electrodes. Touching the surface of the platinized
electrodes may damage and reduce the life of the electrodes.
Electrode Cleaning Recommendations
Do not soak the electrode in the solutions for longer than the recommended length
of time. To do so may cause a reference potential shift which will cause a
degradation in performance or failure. When cleaning the probe, do not scratch
or damage the platinized electrode surfaces. To remove water soluble contaminants
from the electrode, soak the electrode in deionized water and scrub with a soft brush.
Battery Replacement
1. Twist off the battery compartment cap.
2. Hold the battery housing in place with one fi nger. Remove the battery
carrier by pulling on the small tabs.
3. Replace the four CR2032 batteries. Observe polarity.
4. Replace the battery compartment cap.
Replacing the electrode
1. Unscrew and remove the electrode collar. Turn collar counter-clockwise.
2. Gently rock the electrode side to side, while pulling it away from the
meter, until it disconnects from the electrode socket.
3. To attach an electrode, align the slots and carefully plug the electrode into
the meter socket. CAUTION: Take care to align pins carefully. Bent or
broken pins will cause the meter to malfunction.
4. Firmly tighten the electrode collar to create a seal with the rubber gasket
between the electrode and the meter.
LaMotte Company warrants this instrument to be free of defects in parts and
workmanship for 1 year from the date of shipment and the probe to be free of
defects in parts and workmanship for 6 months from the date of shipment. If it
This warranty does
not apply to defects resulting from action of the user such as misuse, improper
wiring, operation outside of specifi cation, improper maintenance or repair, or
unauthorized modifi cation. LaMotte Company specifi cally disclaims any implied
warranties or merchantability or fi tness for a specifi c purpose and will not be
liable for any direct, indirect, incidental or consequential damages. LaMotte
Company’s total liability is limited to repair or replacement of the product. The
warranty set forth above is inclusive and no other warranty, whether written or
oral, is expressed or implied.