Congratulations on your purchase of the SAL/EC/TDS (Conductivity/Total
Dissolved Solids/Salinity) TRACER PockeTester. The TRACER
direct reading
and maintenance
of salinity, conductivity, and TDS with one electrode. Careful use
will provide years of reliable service.
offers
SPECIFICATIONS
Display2000 count LCD with Bar Graph
Conductivity Range0 to 199.9 μS
200 to 1999 μS
2.00 to 19.99 mS
TDS Range0 to 9,990 ppm (variable ratio)
Salinity Range0 to 9,990 ppm Salt(fi xed ratio of 0.5)
TDS Ratio0.4 to 1.0, adjustable
Salinity Ratio0.5 fi xed
Conductivity ATC2.0% per °C
Temperature Range32.0 to 149.0 °F, 0.0 to 65.0 °C
Temperature Resolution0.1 up to 99.9, 1>100
Temperature Accuracy±1.8°F, 1°C (from 32 to 122°F, 0 to 50°C)± 5.4°F,
Conductivity ATC Range32.0°F to 140°F, (0.0°C to 60.0°C)
AccuracyConductivity: ± 2% FS
Measurement Storage25 numbered readings
Low Battery Indication‘BAT’ appears on the display
PowerFour CR 2032 Lithium Ion batteries
Auto Power OffAfter 10 minutes of no button presses
Operating Conditions23 to 122° F, –5 to 50° C
Dimensions1.6 X 7.4 X 1.6 inches, 40 X 187 X 40 mm
The Tracer uses four CR2032 Lithium Ion batteries. If the batteries are weak,
BAT indicator will appear on the display. Press the ON/OFF key to turn
the
the TRACER on or off. The auto power off feature will shut the TRACER off
automatically after ten minutes of inactivity.
Automatic Calibration
When the TRACER is turned on, it will enter the Automatic Calibration
SELF and CAL will appear while the calibration is in progress. After the
mode.
calibration is completed, the
TDS Compensation Ratio
The TDS value is determined by multiplying the conductivity measurement
by a known conversion ratio factor. The meter allows the selection of a
conversion ratio factor in the range of 0.4 to 1.0. The selected ratio will vary
with application but is typically set between 0.5 and 0.7. In the salinity mode,
the ratio is fi xed at 0.5. The stored ratio factor will briefl y appear in the lower
temperature display when the meter is fi rst turned on or when changing the
measurement function to TDS.
To Change the TDS Conversion Ratio:
1. Turn the TRACER on.
2. Press and release the CAL button twice.
The stored ratio will appear in the display.
3. Press the MODE button to change the ratio
value in steps of 0.1.
4. When the desired ratio is displayed, press and release
the CAL button to store the value and return
to the normal mode.
5. If no buttons are pressed for 5 seconds, the meter
will return to the measurement mode.
To change the displayed temperature units between °F or °C:
1. With the TRACER off, press and hold the CAL button.
2. With the CAL button pressed, momentarily press the ON/OFF button.
Data Hold
Press the MODE button to freeze the current reading. The HOLD icon will
appear, the reading will be stored. Press the MODE key to return to normal
operation.
Auto Power Off
The auto-power off feature will automatically shut the meter off 10 minutes
after the last button was pressed. To disable the auto-off feature:
1. Press the ON/OFF button to turn the meter on.
2. Wait for the
3. Press CAL button once quickly.
4. Press MODE and ON/OFF buttons simultaneously for approximately two
SELF CAL appears in the display, release the CAL button. The
When
TRACER will return to the operational mode with the temperature displayed in the new units.
SELF display screen to appear. This is the second screen to
appear after turning on the meter. IMPORTANT: Review the instructions
for the next 3 steps before proceeding. Step 3 must be followed immediately by Step 4.
seconds. Quickly release buttons.
5. oFF will be displayed on the screen. Watch carefully. It will disappear
quickly.
6. The auto power off feature will be restored automatically when the meter
is turned off. Auto-off is the default function when the meter is turned on.
Low Battery Indicator
The “BAT” indicator will be displayed when the batteries become weak. Refer
to the maintenance section for battery replacement information
1. Remove the cap from the bottom of the TRACER to expose the electrode.
2. Before the fi rst use, rinse the electrode in deionized water and dry.
3. For best results, calibrate for conductivity with a standard in the expected
range of the sample. For maximum accuracy calibrate from low conductivity
value standards to high conductivity value standards.
4. Store dry.
Changing the Measurement Function
The meter can be set to measure Conductivity, TDS (ppm), TDS (mg/L) or
Salinity (ppm). To change the mode:
1. Turn the TRACER on.
2. Press and hold the MODE button for 2 seconds. The display will begin to
scroll through the units.
μS or mS (Conductivity)
ppm (TDS)
ppm (Salinity “S”)
3. Release the MODE key when the desired mode is displayed.
4. Note that the “HOLD” function can not be on when changing functions.
5. Filter the contents of the fl ask. Collect the fi ltrate in a beaker.
6. Rinse the electrode with distilled or deionized water to remove impurities.
7. Press the ON/OFF button to turn the TRACER on. Make sure the meter is
in the TDS mode.
8. Immerse the electrode in the fi ltrate. Make sure the tip of the electrode is
completely immersed.
9. Stir the fi ltrate with the electrode to create a homogeneous solution.
10. Gently stir the fi ltrate with the electrode. Read the TDS value of the fi ltrate from the display.
11. Rinse the electrode in distilled water. Replace the cap.
Storing Readings
1. After the reading is displayed press and hold the
MODE button to store the current reading. The
meter will enter the HOLD mode and HOLD will
be displayed. The storage location number will
be displayed on the lower display followed by the
reading being stored.
2. Press the MODE button to exit the HOLD mode
and return to normal operation.
3. If an attempt is made to store more than 25 readings,
the stored readings will be overwritten starting
with the fi rst reading.
Recalling Stored Readings
NOTE: First ensure that the HOLD symbol is not displayed. If it is, exit the HOLD
function by pressing the MODE button.
1. Press the CAL button and then press the MODE button; the location
number (1 through 25) will briefl y appear and then the value
that location will appear. The displayed units will fl ash, indicating
2. The last stored reading taken will be displayed fi rst. To advance to the
previously stored readings, press the MODE button. The location number
is displayed fi rst, followed by the reading stored in that location.
3. To exit the storage mode, press the CAL button and the TRACER will
return to normal operation after displaying “End”.
Clearing the Stored Memory
Turn the TRACER on. Press and hold the ON/OFF button for 4 second the display
will briefl y display “clr” when the memory is cleared.
CALIBRATION
For the most accurate results, allow suffi cient time for the temperature of the
probe to reach the temperature of the sample before calibrating. This will be
indicated by a stable temperature reading on the display
Meter accuracy verifi cation should be performed on a periodic basis as needed.
If calibration is required, the meter must be in the salinity mode to perform a
calibration for salinity, or in the conductivity mode to perform calibrations for
conductivity, and TDS. The meter can perform a calibration and store the data
for one range in the salinity mode and each of the three ranges – low, medium
and high for the conductivity and TDS modes. The automatic calibration
recognition procedure will recognize salinity standard 3,000 ppm (medium)
and conductivity standards of 84 μS (Low), 1413 μS (Medium) or 12,880 μS
(12.88 μS) (High). (See Page 5). Always calibrate in the range closest to the
expected measurement value. For salinity samples within the range of 1,000 to
9,999 ppm salinity, calibrate with a 3,000 ppm calibration standard.
1. Fill a sample cup to 20 mL line with a calibration standard.
NOTE: The meter allows for a 1, 2, or 3 point calibration for conductivity
and TDS. If calibration is done for more than one point, the lowest
concentration should be done fi rst to obtain the best accuracy. Calibrate
the ranges from low to high
2. Press the ON/OFF button to turn the TRACER on. Insert the electrode
into the standard. Tap or stir the sample with the Tracer to dislodge air
bubbles.
3. Press and hold the CAL button for approximately 2 seconds. “CAL” will
appear and the display will fl ash.
4. The meter will automatically recognize and calibrate to the calibration
standard. The display will briefl y indicate “SA” and “End” and then
return to the measurement mode.
NOTE: “SA” will not appear if the calibration fails.
5. The calibration range indicator will appear on the display for each range
that is calibrated during a power on cycle.
Low Range, 84 μS/cm
L
Medium Range, 1413 μS/cm or 3,000 ppm SaH
M
High Range, 12.88 mS/cm (12,880 μS/cm)
H
NOTE: Each time the calibration mode is entered all calibration range indicator
will be cleared, but only the calibration data for the currently selected range will
be replaced. In the conductivity/TDS modes, the calibrations for the other two
ranges will be saved even though the indicators for those ranges are no longer
displayed. Calibration of all three ranges must be performed during one power
on period for all three calibration range indicators to be displayed.
1. Rinse the electrode in distilled or deionized water.
2. Store the electrode dry with the cap on.
3. Always rinse the electrode in deionized water between measurements to
avoid cross contamination. Double rinsing is recommended when high
accuracy is required.
Battery Replacement
1. Twist off the battery compartment cap.
2. Hold the battery housing in place with one fi nger. Remove the battery
carrier by pulling on the small tabs.
3. Replace the four CR2032 batteries. Observe polarity.
4. Replace the battery compartment cap.
You, as the end user, are legally bound (EU Battery ordinance)
to return all used batteries, disposal in the household garbage is
prohibited! You can hand over your used batteries / accumulators
at collection points in your community or wherever batteries /
accumulators are sold!
Disposal: Follow the valid legal stipulations in respect of the disposal of the
device at the end of its lifecycle.
Electrode Care
1. Always rinse the electrode in distilled or deionized water between
measurements to avoid cross-contamination of the samples. Double
rinsing is recommended when high accuracy is required.
2. Do not touch the electrodes. Touching the surface of the platinized
electrodes may damage and reduce the life of the electrodes.
1. Unscrew and remove the electrode collar. Turn collar counter-clockwise.
2. Gently rock the electrode side to side, while pulling it away from the
meter, until it disconnects from the electrode socket.
3. To attach an electrode, align the slots and carefully plug the electrode into
the meter socket. CAUTION: Take care to align pins carefully. Bent or
broken pins will cause the meter to malfunction.
4. Firmly tighten the electrode collar to create a seal with the rubber gasket
between the electrode and the meter.
Electrode Cleaning Recommendations
Do not soak the electrode in the solutions for longer than the recommended
length of time. To do so may cause a reference potential shift which will cause
a degradation in performance or failure. When cleaning the probe, do not
scratch or damage the platinized electrode surfaces.
ContaminantCleaning SolutionProcedure
Water soluble
substances
Grease and oilWarm water and
Heavy greaseAlcoholSoak for a maximum of 5
Lime and
hydroxide
coatings and oil
Deionized waterSoak or scrub with a soft
brush. Rinse thoroughly with
DI water and dry.
Soak or scrub with a soft
household detergant
10% Acetic acidSoak until coating dissolves,
brush, maximun 10 minutes.
Rinse thoroughly with DI
water and dry.
minutes. Scrub with a soft
brush. Rinse thoroughly with
DI water and dry.
maximum 5 minutes. Rinse
thoroughly with DI water and
dry.
LaMotte Company warrants this instrument to be free of defects in parts and
workmanship for 1 year from the date of shipment and the probe to be free of
defects in parts and workmanship for 6 months from the date of shipment. If it
should become necessary to return the instrument for service during or beyond
the warranty period, contact our T
for troubleshooting help. The sender is responsible for shipping charges, freight,
insurance and proper packaging to prevent damage in transit. This warranty does
not apply to defects resulting from action of the user such as misuse, improper
wiring, operation outside of specifi cation, improper maintenance or repair, or
unauthorized modifi cation. LaMotte Company specifi cally disclaims any implied
warranties or merchantability or fi tness for a specifi c purpose and will not be
liable for any direct, indirect, incidental or consequential damages. LaMotte
Company’s total liability is limited to repair or replacement of the product. The
warranty set forth above is inclusive and no other warranty, whether written or
oral, is expressed or implied.