LaMotte 1200 User Manual

Page 1
12
0
0
COL ORIMETER
.................
Instruction
MANUAL
Page 2
Page 3
TABLE OF CONTENTS
PackagingandDelivery...............................4
General Precautions ................................4
Safety Precautions .................................4
Limits of Liability ..................................4
Specifications....................................5
Parts&Accessories.................................6
EPACompliance..................................7
CEStatement....................................7
CHEMICALTESTING........................8
WaterSamplingforChemicalAnalysis.....................8-9
Filtration ......................................9
AnIntroductionToColorimetricAnalysis..................10-11
Reagent Blank ...................................11
ColorimeterTubes.................................11
SampleDilutionTechniques&VolumetricMeasurement...........12
Interferences....................................13
StrayLightInterference..............................13
GENERALOPERATINGINFORMATION.............14
Overview....................................14-15
TheKeypadandDisplay .............................16
TestingTips....................................17
CALIBRATIONPROCEDURE..................18-19
ANALYSISPROCEDURE....................20-21
AVAILABLETESTKITS ......................22
TROUBLESHOOTINGGUIDE...................23
RS232PORT ...........................24
MAINTENANCE OF THE 1200 ..................25
ReplacingtheBattery...............................25
Repairs.......................................25
MeterDisposal...................................25
WARRANTY............................26
Limitations.....................................26
Page 4
GENERAL INFORMATION
PACKAGING & DELIVERY
Experienced packaging personnel at LaMotte Company assure adequate protection against normal hazards encountered in transportation of shipments. After the product leaves the manufacturer, all responsibility for safe delivery is assured by the transportation company. Damage claims must be filed immediately with the transportation company to receive compensation for damaged goods.
Should it be necessary to return the instrument for repair or servicing, pack the instrument carefully in a suitable container with adequate packing material. A return authorization number must be obtained from LaMotte Company by calling 1-800-344-3100 or faxing 1-410-778-6394. Attach a letter with the authorization number to the shipping carton which describes the reason for the return. This information will enable the service department to make the required repairs more efficiently.
GENERAL PRECAUTIONS
Read the instruction manual before attempting to set up or operate this instrument. Failure to do so could result in personal injury or damage to the instrument.
The 1200 Colorimeter should not be stored or used in a wet or corrosive environment. Care should be taken to prevent water from wet colorimeter tubes from entering the colorimeter light chamber.
NEVER PUT WET TUBES IN THE COLORIMETER.
SAFETY PRECAUTIONS
Read labels on all reagent containers. Some labels include precautionary notices and first aid information. Certain reagents are considered hazardous and are designated with a * in the instruction manual. To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or our web site. To obtain a printed copy, contact us by e-mail, phone or FAX. Additional information for all LaMotte reagents is available in the United States, Canada, Puerto Rico, and the US Virgin Islands from Chem-Tel by calling 1-800-255-3924. For other areas, call 813-248-0585 collect to contact Chem-Tel’s International access number. Each reagent can be identified by the four digit number listed on the upper left corner of thereagent label, in the contents list and in the test procedures.
LIMITS OF LIABILITY
Under no circumstances shall LaMotte Company be liable for loss of life, property, profits, or other damages incurred through the use or misuse of their products.
4
Page 5
SPECIFICATIONS
Instrument Type Single wavelength, direct reading colorimeter
Measurement Wavelengths: 420 nm, 460 nm, 510 nm, 530 nm, 570 nm or
605 nm
Readable Resolution: Determined by test factor
Photometric Precision: ±0.001 Absorbance Unit
Range: 0 - 2.00 absorbance units
Display: 3 ½ digits
Response Time: 2 seconds
Warm-up Time: Not required
Lamp: LED
Detector: Silicon Photodiode
Sample: 10 mL in capped tube
Sample Chamber: Accepts 25mm diameter flat-bottomed tubes
(capped)
Power Source: Battery Operation: 9 Volt Alkaline
Line Operation: 110V/60Hz, 220V/50Hz, with available adapter
Size (L X W X H): 8.5 x 16.2 x 6.7 cm, 3.4 X 6.4 X 2.6 inches
Shipping Weight: 11 oz (312 g) meter only
Serial Interface: RS232, 8 pin mDIN, 9600b, 8, 1, n
5
Page 6
PARTS & ACCESSORIES
Included in the Model 1200 Colorimeter Kit :
1200 Colorimeter
Colorimeter tubes, set of 6
Water Sample Collecting Bottle
Reagent System
Optional Accessories
AC Adapter, 110V/60Hz Code 1726-110
AC Adapter, Universal Code 1754
Colorimeter Tubes, set of 6 Code 0290-6
6
Page 7
EPA COMPLIANCE
CE
EUROPEAN
MARK
Application of Council Directives:
Standards to which Conformity Declared:
Manufacturer's Name:
Manufacturer's Address:
Importer's Name:
Importer's Address:
Type of Equipment:
Model Number:
Year of Manufacture:
89/336/EEC
EN55022, EN50082-1, En600950
802 Washington Avenue PO Box 329 Chestertown, MD 21620
Reagecon Diagnostics Ltd
Water Quality Meters
2020/1200
1997
13 A/D Shannon Free Zone Shannon, Co. Clase. Ireland
LaMotte Company
I, the undersigned, hereby declare that the equipment specified above
conforms to the above Directive and Standards.
Place
Chestertown, Maryland
Date
3-19-97
Signature
Name
James K. Trumbauer
Position
V.P., Director of Research & Development
DECLARATION OF CONFORMITY
NOTE: The device complies to the product specifications for the
Low Voltage Directive when furnished with the 220V AC Adapter (Code 1774).
The 1200 Chlorine Colorimeter is an EPA Accepted instrument. EPA Accepted means that the instrument meets requirements for colorimeters as found in test procedures that are approved for the National Primary Drinking Water Regulations (NPDWR) or National Pollutant Discharge Elimination Systems (NPDES) compliance monitoring programs. EPA Accepted instruments may be used with approved test procedures without additional approval.
CE COMPLIANCE
The 1200 Colorimeter has been independently tested and has earned the European CE Mark of Compliance for electromagnetic compatibility and safety.
7
Page 8
CHEMICAL TESTING
WATER SAMPLING FOR CHEMICAL ANALYSIS
Taking Representative Samples
The underlying factor to be considered for any type of water sampling is whether or not the sample is truly representative of the source. To properly collect a representative sample:
Sample as frequently as possible.
Collect a large sample or at least enough to conduct whatever tests are
necessary. Makeacompositesampleforthesamesamplingarea.
Handle the sample in such a way as to prevent deterioration or contamination
before the analysis is performed. Perform analysis for dissolved gases such as dissolved oxygen, carbon dioxide,
and hydrogen sulfide immediately at the site of sampling. These factors, as well as samples for pH, cannot be stored for later examination.
Make a list of conditions or observations which may affect the sample. Other
considerations for taking representative samples are dependent upon the source of the sample. Taking samples from surface waters involves different considerations than taking samples from impounded and sub-surface waters.
SamplingofOpenWaterSystems
Surface waters, such as those found in streams and rivers, are usually well mixed. The sample should be taken downstream from any tributary, industrial or sewage pollution source. For comparison purposes samples may be taken upstream and at the source of the pollution before mixing.
In ponds, lakes, and reservoirs with restricted flow, it is necessary to collect a number of samples in a cross section of the body of water, and where possible composite samples should be made to ensure representative samples.
To collect samples from surface waters, select a suitable plastic container with a tight fitting screw cap. Rinse the container several times with the sample to be tested, then immerse the container below the surface until it is filled to overflowing and replace the cap. If the sample is not to be tested immediately, pour a small part of the sample out and reseal. This will allow for any expansion. Any condition which might affect the sample should be listed.
Sub-surface sampling is required to obtain a vertical profile of streams, lakes, ponds, and reservoirs at specific depths. This type of sampling requires more sophisticated sampling equipment.
8
Page 9
For dissolved oxygen studies, or for tests requiring small sample sizes, a Water Sample Bottle (LaMotte Code 1060) will serve as a subsurface or in-depth sampler. This weighted device is lowered to the sampling depth and allowed to rest at this depth for a few minutes. The water percolates into the sample chamber displacing the air which bubbles to the surface. When the bubbles cease to rise, the device has flushed itself approximately five times and it may be raised to the surface for examination. The inner chamber of the sampling device is lifted out and portions of the water sample are carefully dispensed for subsequent chemical analysis.
A Snap-Plunger Water Sampler (LaMotte Code 1077) is another “in-depth” sampling device which is designed to collect large samples which can be used for a multitude of tests. Basically, this collection apparatus is a hollow cylinder with a spring loaded plunger attached to each end. The device is cocked above the surface of the water and lowered to the desired depth. A weighted messenger is sent down the calibrated line to trip the closing mechanism and the plungers seal the sample from mixing with intermediate layers as it is brought to the surface. A special drain outlet is provided to draw off samples for chemical analysis.
Sampling of Closed System
To obtain representative samples from confined water systems, such as pipe lines, tanks, vats, filters, water softeners, evaporators and condensers, different considerations are required because of chemical changes which occur between the inlet and outlet water. One must have a basic understanding of the type of chemical changes which occur for the type of equipment used. Also, consideration should be given to the rate of passage and retaining time for the process water.
Temperature changes play an important part in deciding exactly what test should be performed. Process water should be allowed to come to room temperature, 20–25°C, before conducting any tests.
When drawing off samples from an outlet pipe such as a tap, allow sample to run for several minutes, rinsing the container several times before taking the final sample. Avoid splashing and introduction of any contaminating material.
FILTRATION
When testing natural waters that contain significant turbidity due to suspended solids and algae, filtration is an option. Reagent systems, whether EPA, Standard Methods, LaMotte or any others, will generally only determine dissolved constituents. Both EPA and Standard Methods suggest filtration through a 0.45 micron filter membrane, to remove turbidity, for the determination of dissolved constituents.** To test for total constituents, organically bound and suspended or colloidal materials, a rigorous high temperature acid digestion is necessary.
**LaMotte offers a filtering apparatus: Syringe Assembly (Code 1050) and Membrane Filters, 0.45 micron (Code 1103).
9
Page 10
AN INTRODUCTION TO COLORIMETRIC ANALYSIS
Most test substances in water are colorless and undetectable to the human eye. To test for their presence we must find a way to “see” them. The LaMotte colorimeter can be used to measure any test substance that is itself colored or can be reacted to produce a color. In fact a simple definition of colorimetry is “the measurement of color” and a colorimetric method is “any technique used to evaluate an unknown color in reference to known colors”. In a colorimetric chemical test, the intensity of the color from the reaction must be proportional to the concentration of the substance being tested. Some reactions have limitations or variances inherent to them that may give misleading results. Many such interferences are discussed with each particular test instruction. In the most basic colorimetric method the reacted test sample is visually compared to a known color standard. However, accurate and reproducible results are limited by the eyesight of the analyst, inconsistencies in the light sources, and the fading of color standards.
To avoid these sources of error, a colorimeter can be used to photoelectrically measure the amount of colored light absorbed by a colored sample in reference to a colorless sample (blank).
White light is made up of many different colors or wavelengths of light. A colored sample typically absorbs only one color or one band of wavelengths from the white light. Only a small difference would be measured between white light before it passes through a colored sample versus after it passes through a colored sample. The reason for this is that the one color absorbed by the sample is only a small portion of the total amount of light passing through the sample. However, if we could select only that one color or band of wavelengths of light to which the test sample is most sensitive, we would see a large difference between the light before it passes through the sample and after it passes through the sample.
A colorimeter passes a white light beam through an optical filter which transmits only one particular color or band of wavelengths of light to the photodetector where it is measured. The difference in the amount of colored light transmitted by a colorless sample (blank) and the amount of colored light transmitted by a colored sample is a measurement of the amount of colored light absorbed by the sample. In most colorimetric tests the amount of colored light absorbed is directly proportional to the concentration of the test factor producing the color and the path length through the sample. However, for some tests the amount of colored light absorbed is inversely proportional to the concentration.
The choice of the correct optical filter and therefore the correct color or wavelength of light is important. It is interesting to note that the filter that gives the most sensitive calibration for a test factor is the complementary color of the test sample. For example, the Nitrate-Nitrogen test produces a pink color proportional to the nitrate concentration in the sample (the greater the nitrate concentration, the darker the pink color). A green filter is used since a pinkish-red solution absorbs mostly green light.
10
Page 11
The 1200 Colorimeter has been specially calibrated to read test results directly in parts per million (ppm) of the test factor. The sensitivity of the response has been maximized by using a specific, complementary light source for each test.
REAGENT BLANK
Some tests will provide greater accuracy if a reagent blank is determined, to compensate for any color or turbidity resulting from the reagents themselves. A reagent blank is performed by running the test procedure on 10 mL of demineralized water. Use sample water to zero the meter. Insert the reagent blank in the colorimeter chamber and select READ. Note result of reagent blank.
Perform the tests on the unknown as described. Subtract results of reagent blank from all subsequent test results.
COLORIMETER TUBES
The handling of the colorimeter tubes is of utmost importance. Scratches, fingerprints and water droplets on the tube or inside the light chamber can cause stray light interference leading to inaccurate results. It is imperative that the tubes and light chamber be clean and dry. The glassware must be clean and defect-free. Scratches and abrasions will permanently affect the accuracy of the readings. Tubes can be acid washed periodically. After a tube has been filled and capped, it should be held by the cap and the outside surface should be wiped with a clean, lint-free absorbent cloth until it is dry and smudge-free. Handling the tube only by the cap will avoid problems from fingerprints. Always set the clean tube aside on a clean surface that will not contaminate the tube.
Variability in the geometry and quality of the glassware is the predominate cause of variability in results. Orientation of the tube in the chamber will greatly affect the test results. To obtain the most accurate results, the tubes must be positioned so that the index line on the tube is aligned with the arrow-shaped index mark t molded into the housing in front of the light chamber. This will ensure that the most accurate results are obtained.
11
Page 12
SAMPLE DILUTION TECHNIQUES & VOLUMETRIC MEASUREMENTS
If a test result using the 1200 Colorimeter gives an ER2 (over range) message, the sample must be diluted. Then the test should be repeated to obtain a reading which is in the concentration range for the test. (Note: This is not true for colorimetric determination of pH.)
Example:
Measure 5 mL of the water sample into a graduated cylinder. Add demineralized water until the cylinder is filled to the 10 mL line. The sample has been diluted by one-half, and the dilution factor is therefore 2. Perform the test procedure, then multiply the resulting concentration by 2 to obtain the test result.
The following table gives quick reference guidelines on dilutions of various proportions. All dilutions are based on a 10 mL volume, so several dilutions will require small volumes of the water sample. Graduated pipets should be used for all dilutions.
Deionized Water to Bring Volume to 10
Size of Sample
10 mL 0 mL 1
5mL 5mL 2
mL
Multiplication
Factor
2.5 mL 7.5 mL 4
1mL 9mL 10
0.5 mL 9.5 mL 20
If the above glassware is not available, dilutions can be made with the colorimeter tube. Fill the colorimeter tube to the 10 mL line with the sample then transfer it to another container. Add 10 mL volumes of demineralized water to the container and mix. Transfer back 10 mL of the diluted sample to the colorimeter tube and follow the test procedure. Continue diluting and testing until a reading, which is in the concentration range for the test, is obtained. Be sure to multiply the concentration found by the dilution factor (the number of total 10 mL volumes used).
Example:
10 mL of sample is diluted with three 10 mL volumes of demineralized water; the dilution factor is four.
12
Page 13
INTERFERENCES
LaMotte reagent systems are designed to minimize most common interferences. Each individual test discusses interferences unique to that test. Be aware of possible interferences in the water being tested.
The reagent systems also contain buffers to adjust the water sample to the ideal pH for the reaction. It is possible that the buffer capacity of the water sample may exceed the buffer capacity of the reagent system and the ideal pH will not be obtained. If this is suspected, measure the pH of a reacted distilled water reagent blank using a pH meter. This is the ideal pH for the test. Measure the pH of a reacted water sample using the pH meter. If the pH is significantly different from the ideal value, the pH of the sample should be adjusted before testing.
Interferences due to high concentration of the substance being tested, can be overcome by sample dilution (see page 12).
STRAY LIGHT INTERFERENCE
Stray light interference can be minimized by always zeroing the meter and reading a sample with the lid closed. Turbidimetric determinations (i.e. sulfate) are most likely to exhibit a stray light interference. Colorimetric tests are less likely to have this problem unless the samples are turbid. If sample turbidity is causing a stray light interference, filtration may be needed.
13
Page 14
GENERAL OPERATING INFORMATION
OVERVIEW
The Model 1200 Colorimeter is a portable, microprocessor-controlled, direct reading colorimeter. The microprocessor enables factory programmed calibrations to optimally match non-linear curves. The auto zero feature eliminates the need to dial in the zero manually. A sealed keypad controls the operation. The large display presents measurements in ppm concentration and indicates low battery warnings. The 1200 will turn off automatically after 5 minutes to prolong battery life. This will not affect the user calibration and the meter will not have to be re-zeroed. The calibration and the last zero reading will be stored in memory.
The 1200 is supplied with a 9 volt alkaline battery. An optional AC power adapter is available.
A RS-232 serial port on the back of the meter allows an interface of the colorimeter with a computer for real time data acquisition and data storage using the PC. This port also allows an interface with a RS-232 serial printer.
14
Page 15
READ
ZERO
Top
View
Bottom
View
2.55
Serial Number
Battery Compartment
I200
••••••••••••••••••
15
Page 16
THE KEYPAD and DISPLAY
The button is used to turn the meter ON and to take readings. Pressing the button for 2 seconds will cause the meter to display OFF. Releasing the button when OFF is displayed turns the meter OFF.
The button is used to zero the meter and for calibration procedures.
The DOWN ARROW will
the numerical value of the display.
The UP ARROW will
the numerical valueofthe display.
See
TROUBLE
SHOOTING
GUIDE
page 23
The DISPLAY :
When the button is first pushed, a number will be briefly displayed that indicates the software version number.
A walking dash ( ) will be displayed when reading is taking place.
(blank) will be displayed after the button has been pushed and held for 2 seconds. If the button is released while is displayed, a zero or blank reading will be taken.
(calibrate) will be displayed after the zero button has been pushed and held for 5 seconds. If the button is released while is displayed, the calibration mode will be entered as indicated by a flashing display.
will be displayed after the button has been held down for 2 seconds. The meter will turn off when the button is released.
will be displayed when the battery voltage is very low.
will be displayed when concentration is over range.
will be displayed when the bulb has burned out.
will be displayed when the battery voltage is getting low. Readings are reliable. Replace battery as soon as possible.
-
READ
ZERO
I200
••••••••••••••••••
L
Mott
DECREASE
READ
ZERO
INCREASE
READ
ZERO
ZERO
ZERO
READ
Bla
BLA
cal
CAL
OFF
Er1 Er2 Er3
BAT
16
Page 17
TESTING TIPS
Samples should be collected in a clean glass or polyethylene container.
Observe the shelf-life
recommendations for reagents.
Samples should be analyzed as soon as possible after collection.
5.
3.
1. 2.
4.
6.
Protect reagents and components from extreme heat and cold.
Never put wet tubes in colorimeter.
Discard tubes that are badly scratched.
ONE
year
SHELF
LIFE
17
Page 18
CALIBRATION PROCEDURE
Prepare standard solutions to be tested.
Insert the tube into the chamber, being sure to align the index line with the arrow on the meter. Close the lid. This tube is the blankorzero.
Rinse and fill a colorimeter tube (0290) to the 10 mL line with standard solution. Cap and wipe dry.
7. 8.
5.
3.
1. 2.
4.
6.
React a standard that is in the range of the reagent system, and similar to the expected range of the sample. Follow the individual test procedure.
Wipe the tube clean with a lint-free cloth.
2.50 ppm
Push the button to turn the meter on. Press the
button andholditfor2 seconds until is displayed. Release the button to take a zero reading (0 ppm).
Insert the tube into the chamber, being sure to align the index line with the arrow on the meter. Close the lid.
Push the button. If the displayed value is not the same as
thevalueofthe reacted standard (within the specification limits), continue with the calibration procedure.
I200
••••••••••••••••••
ZERO
COLORIMETER
READ
2.55
I200
••••••••••••••••••
ZERO
COLORIMETER
READ
bla
READ
READ
ZERO
bla
The 1200 has been pre-calibrated. Recalibration of the 1200 by the user is not required. However, a procedure to standardize the calibration (shown below) should be performed to obtain the most accurate readings.
18
Page 19
The 1200 can be calibrated with two prepared standards. The concentrations of
Push the button again to memorize the calibration. The 1200 display will stop flashing. Calibration is complete.
ZERO
Turn the unit off by holding the button down for at least 2 seconds, or proceed to measure the test samples following the procedure on page
20.
READ
9. 10.
11 .
Push and hold the
button for 5 seconds until is displayed. Release the button. The display will flash. Adjust the display with the and buttons until the value of the standard is displayed.
ZERO
CAL
I200
••••••••••••••••••
ZERO
COLORIMETER
READ
2.55
I200
••••••••••••••••••
ZERO
COLORIMETER
READ
2.50
I200
••••••••••••••••••
ZERO
COLORIMETER
READ
the standards should be chosen from the low and high ends of the range of the meter. The low standard must be less than 5% of over range. To calibrate the meter with two standards, follow steps 1-11 of the calibration procedure for one reacted sample. Repeat steps 1-11 of the calibration procedure with the second reacted standard.
The calibration procedure should be followed as often as required by regulations and laws for compliance monitoring.
To reset the calibration to the factory calibration: with the meter off, hold down the t and push READ. The meter will turn on and the calibration
will be reset.
19
Page 20
ANALYSIS PROCEDURE
Fill the Water Sample Collecting Bottle (0688) with sample water. This will be used to dispense sample water for the tests.
Push the button to turn the meter on. Press the
button and hold it for 2 seconds until is displayed. Release the button to take a zero reading (0 ppm).
READ
ZERO
bla
Rinse and fill a colorimeter tube (0290) to the 10 mL line with sample water. Cap and wipe dry.
7. 8.
5.
3.
1. 2.
4.
6.
React a sample following the individual test procedure.
Wipe the tube clean with a lint-free cloth.
Insert the tube into the chamber, being sure to align the index line with the arrow on the meter. Close the lid. This tube is the blankorzero.
Insert the tube into the chamber, being sure to align the index line with the arrow on the meter. Close the lid.
Push the button. The
concentration in ppm will be displayed within 2 seconds.
READ
I200
••••••••••••••••••
ZERO
COLORIMETER
READ
2.55
I200
••••••••••••••••••
ZERO
COLORIMETER
READ
bla
20
Page 21
9.
The 1200 will turn off automatically 5 minutes after the last button push. To turn the meter off manually, hold the
button for 2 seconds. Release the button when is displayed.
READ
OFF
If is displayed, the concentration is over range. The sample must be diluted and re-tested. See page 12.
Er2
Note
I200
••••••••••••••••••
ZERO
COLORIMETER
READ
21
Page 22
AVAILABLE TEST KITS
TEST FACTOR
Ammonia Nitrogen
ORDER CODE/MODEL
3680-01 DC 1200-NH
Bromine 3672-01
DC 1200-BR
Chlorine (Free &
3670-01 DC1200-CL
Total)
Chlorine (Free &
3670-01-LI DC1200-CL-LI
Total)
Chlorine Dioxide
3671-01 DC1200-CLO
Copper 3673-01
DC1200-CO
Fluoride 3674-01
DC1200-FL
Iron 3681-01
DC1200-FE
RANGE
(PPM)
DETECTION LIMIT
TEST METHOD
(# OF REAGENTS)
0-5.00 0.05 Nessler (2)
0-7.00 0.05 DPD Tablets (1)
0-4.00 0.05 DPD Tablets (2)
0-4.00 0.05 DPD Liquid (3)
0-7.00 0.05 DPD with Glycine
Solution (2)
0-8.00 0.03 Diethyldithio-
carbamate (1)
0-2.00 0.03 Aliizarin-Zirconyl
(2)
0-4.00 0.25 1-10
Phenanthroline (2)
Manganese 3682-01
0-0.90 0.01 PAN (3)
DC1200-MN
Molybdenum 3676-01
0-50.0 0.5 Thioglycolate (3)
DC1200-MO
Nitrate Nitrogen
Ozone 3678-01
3677-01 DC1200-NA
0-3.00 0.05 Cadmium
0-0.40 0.04 Indigo Blue (3)
DC1200-OZ
Phosphate 3679-01
0-3.00 0.07 Ascorbic Acid (2)
DC1200-PLR
Sulfate 3683-01
0-100.0 1.0 Barium Chloride
DC1200-SU
Test kit includes 1200 meter and reagent system.
22
Reduction (2)
(1)
Page 23
TROUBLESHOOTING
er1
er2
er3
PROBLEM CHECK ACTION
Meter won’t turn on Battery Replace
AC Adapter Plug in
AC Wall Outlet Verify power source
Contact LaMotte for Return Authorization
Suspect Calibration Check calibration with
standards
Verify standard concentration
Verify with another meter
Check tube alignment Re-align tube
Check sample tubes for dirt and scratches
Check to see if internal meter components are wet
Reset meter to factory calibration
Contact LaMotte for Return Authorization
Return to LaMotte for repair
Use new standards
Runtestwithalternative reagent system
Check other meter calibrations
Check, clean and/or replace if necessary
Always dry tubes before inserting. Examine chamber for visible moisture.
With meter off, hold down t and press READ
Return for calibration check
Very low battery Change battery
Over range Dilute sample
Burnt out bulb Call LaMotte
BAT Low battery Change battery
23
Page 24
RS232 PORT
The 1200 Colorimeter may be interfaced with any IBM compatible computer using an Interface cable (Code 1772). The meter may also be interfaced with an RS-232 serial printer, using an appropriate cable and setting the printer configuration to the output below.
Output: RS232 compatible, asynchronous serial, 9600 baud, no parity, 8 data bits, 1 stop bit.
Computer Connection: RS232 (Code 1772) interface connection, 8 pin mini-DIN/9 pin F D-submin.
Pin out:
5 RS-232 TxD
3 RS-232 RxD
4, 6, 8 Digital ground
24
Page 25
MAINTENANCE
REPLACING THE BATTERY
The LaMotte 1200 uses a standard 9-volt alkaline battery that is available worldwide. The battery compartment is located on the bottom of the case. To replace the battery:
1. Open the battery compartment lid.
2. Remove the battery and disconnect the battery from the polarized plug.
3. Carefully connect the new battery to the polarized plug and insert it into the compartment.
4. Close the battery compartment lid.
REPAIRS
If it is necessary to return the instrument for repair, telephone LaMotte Company at 1-800-344-3100 or fax 1-410-778-6394 for a return authorization number.
METER DISPOSAL
Waste Electrical and Electronic Equipment (WEEE)
Natural resources were used in the production of this equipment. This equipment may contain materials that are hazardous to health and the environment. To avoid harm to the environment and natural resources, the use of appropriate take-back systems is recommended. The crossed out wheeled bin symbol on the meter encourages you to use these systems when disposing of this equipment.
Take-back systems will allow the materials to be reused or recycled in a way that will not harm the environment. For more information on approved collection, reuse, and recycling systems contact your local or regional waste administration or recycling service.
25
Page 26
WARRANTY
This Instrument is guaranteed to be free from defects in material and workmanship for a period of one (1) year from the original purchase date. In the event that a defect is found during the warranty time frame, LaMotte Company agrees that it will be repaired or replaced without charge except for the transportation costs. This guarantee does not cover batteries.
This product can not be returned without a return authorization number from LaMotte Company. For warranty support or a Return Authorization Number, contact LaMotte Company at 1-800-344-3100 or tech@lamotte.com.
Limitations
This guarantee is void under the following circumstances:
Damage due to operator negligence, misuse, accident or improper application.
Damage or alterations from attempted repairs by an unauthorized
(non-LaMotte) service. Damage due to improper power source, AC adapter or battery.
Damage caused by acts of God or natural disaster.
Damage occurred while in transit with a shipping carrier.
LaMotte Company will service and repair out-of-warranty products at a nominal charge.
26
Page 27
27
Page 28
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside U.S.A.) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
®
626728 • 4.09
Page 29
1200 COLORIMETER
CHLORINE
DPD TABLET TEST
MODEL 1200-CL · CODE 3670-01
QUANTITY CONTENTS CODE
100 *Chlorine DPD #1 Instrument Grade Tablets *6903A-J 100 *Chlorine DPD #3 Instrument Grade Tablets *6197A-J 1 Colorimeter Tubes, w/caps,set of 6 0290-6 2 Tablet Crushers 0175 1 Water Sample Collecting Bottle 0688 1 1200 Colorimeter for Chlorine DPD 26728
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or www .lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specified code number.
INTRODUCTION
Chlorine is added to water to kill bacteria and other disease-producing organisms, control algae, and remove undesirable odors and colors. Chlorine added to water quickly forms hypochlorous acid, HClO, also known as Free Available Chlorine, the active ingredient responsible for chlorine*s sanitizing capabilities. Free Available Chlorine combines with impurities in the water to form chloramines and other organic nitrogen compounds. In combined form, its sanitizing capability diminishes, and higher levels of chlorine are necessary to achieve effective sanitation.
Therefore, it is essential to chlorinate to the point of establishing a Free Available Chlorine Residual, and then to maintain that residual at a recommended level. The LaMotte DPD tablet test method distinguishes levels of Free Available Chlorine, Combined Chlorine, and Total Residual Chlorine, using a single test sample.
Page 30
CHLORINE TEST PROCEDURE - DPD METHOD
7. 8.
5.
3.
1. 2.
4.
6.
I200
COLORIMETER
••••••••••••••••••
READ
READ
ZERO
CAL
••••••••••••••••••
bla
COLORIMETER
I200
Fill the Water Sample Collecting Bottle (0688) with sample water. This will be used to dispense sample water for the tests.
Insert the tube into the chamber, being sure to align the index line with the arrow on the meter. Close the lid. This tube is the blankorzero.
Empty all but a few drops of sample from tube. Add one *Chlorine DPD #1 IG T ablet (6903A) and crush with tablet crusher (0175). NOTE: T o insure accurate results, tablet must be crushed before filling tube.
Align the index line with the arrow on the meter, insert tube
into chamber. the lid. Push the
READ button.
Record results as ppm Free A vailable Chlorine. Save sample for the T otal Residual Chlorine test.
Rinse and fill a colorimeter tube (0290) to the 10 mL line with sample water. Cap and wipe dry .
Push the
READ
button to turn the meter on. Press the
ZERO button and
hold it for 2 seconds until is displayed.
bla
Release the button to take a zero reading (0.00 ppm).
Fill to 10 mL line with sample. Cap and mix until tablet disintegrates. Make readings within 30 seconds after disintegration of the tablet.
Add one Chlorine DPD #3 IG T ablet (6197A) to the sample from Step 6. Crush tablet.
READ
Read the 1200 Colorimeter Manual before proceeding. Carefully wipe tubes dry before inserting into the colorimeter chamber.
FREE AVAILABLE CHLORINE
Page 31
er2
TOTAL RESIDUAL CHLORINE
11.
9. 10.
12.
••••••••••••••••••
READ
2.55
ZERO
COLORIMETER
I200
Cap and mix until tablet disintegrates.**
Push the
READ button.
Record reading as ppm Total Residual Chlorine.
Insert the tube into the chamber, being sure to align the index line with the arrow on the meter. Close the lid.
Subtract the Free Available Chlorine reading from the Total Residual Chlorine reading to determine ppm Combined Chlorine (Monochloramine, Dichloramine, and Nitrogen Trichloride) present in the water sample.
If either chlorine reading displays , repeat procedure on diluted sample, and multiply the result by the appropriate dilution factor. See 1200 Colorimeter Instruction Manual for procedure.
Levels of chlorine above 10 mg/L will cause a bleaching effect on the DPD indicator, and may give a false indication that no chlorine is present. If it is possible that the chlorine concentration is greater than 10 mg/L (e.g. after shock treatment), perform test on a diluted sample and multiply the result by appropriate dilution factor.
CAUTION: DO NOT leave reacted DPD samples in test tubes. Discard sample and thoroughly rinse tubes. If allowed to remain, DPD will stain tubes, significantly impairing the operation of the 1200 Colorimeter. If necessary, acid wash, and vigorously clean glassware with test tube brush and detergent.
**For wastewater samples, Standard Methods for the Examination of Water and Wastewater recommends waiting 2 minutes for full color development.
Page 32
DPD CHLORINE TEST METHOD SPECIFICATIONS
APPLICATION
Drinking water supplies and distribution systems, swimming pool and spas, sewage and chlorinated waste waters, process waters and sanitizing solutions.
RANGE
0 to 4.0 mg/L Chlorine (may be extended by dilution)
METHOD
In the absence of Iodide, Free Available Chlorine reacts instantly with the buffered diethyl-p-phenylenediamine indicator (DPD) to produce a red color in proportion to the amount of chlorine present. Subsequent addition of potassium iodide produces a rapid color response from the combined forms of chlorine (chloramines).
HANDLING & PRESERVATION
Chlorine in aqueous solutions, particularly weak solutions, is not stable. Exposure to sunlight or agitation will accelerate the reduction of chlorine. Fill sample containers to the top and cap tightly. Analyze samples as soon as possible after collection.
INTERFERENCES
The only interfering substance likely to be encountered is oxidized manganese. The extent of this interference can be determined by treating a sample with sodium arsenite to destroy the chlorine present, so that the amount of interference can be measured.
CALIBRATION
The single test colorimeter is precalibrated. In order to comply with NPDWR or NPDES reporting regulations, the calibration should be checked periodically by using a set of reference standards including a 0 mg/L blank and 0.3, 1.0, and 3.5 mg/L chlorine. To prepare these standards, a LaMotte 1000 mg/L standard chlorine equivalent solution (Code 3858) is available. Consult with your local regulatory agency to determine standardization frequency.
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside USA) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
SM
63670-01 · 7/08
Page 33
1200 COLORIMETER
CHLORINE
DPD LIQUID TEST
MODEL 1200-CL-LI • CODE 3670-01-LI
QUANTITY CONTENTS CODE
30 mL DPD #1A Free Chlorine Reagent P-6740-G 30 mL *DPD #1B Free Chlorine Reagent *P-6741-G 30 mL *DPD #3 Total Chlorine Reagent *P-6743-G 1 Colorimeter Tubes, with caps 0290-6 1 Water Sample Collecting Bottle 0688 1 1200 Colorimeter for Chlorine DPD 26728
*WARNING: Reagents marked with a * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or www .lamotte.com. To obtain a printed copy, contact LaMotte by email, phone or fax.
To order individual reagents or test kit components, use the specified code number.
INTRODUCTION
Chlorine is added to water to kill bacteria and other disease-producing organisms, control algae, and remove undesirable odors and colors. Chlorine added to water quickly forms hypochlorous acid, HClO, also known as Free Available Chlorine, the active ingredient responsible for chlorine's sanitizing capabilities. Free Available Chlorine combines with impurities in the water to form chloramines and other organic nitrogen compounds. In combined form, its sanitizing capability diminishes, and higher levels of chlorine are necessary to achieve effective sanitation.
Therefore, it is essential to chlorinate to the point of establishing a Free Available Chlorine Residual, and then to maintain that residual at a recommended level. The LaMotte DPD method distinguishes levels of Free Available Chlorine, Combined Chlorine, and Total Residual Chlorine, using a single test sample.
Page 34
CHLORINE TEST PROCEDURE - DPD METHOD
Insert the tube into the chamber, being sure to align the index line with the arrow on the meter. Close the lid. This tube is the blank or zero.
8.
5.
3. 4.
Fill the Water Sample Collecting Bottle (0688) with sample water. This will be used to dispense sample water for the tests.
Rinse and fill a colorimeter tube (0290) to the 10 mL line with sample water. Cap and wipe dry .
1. 2.
I200 COLORIMETER
••••••••••••••••••
READ
bla
ZERO
7.
CAL
I200COLORIMETER
••••••••••••••••••
READ
Align the index line with the arrow on the meter, insert tube into chamber. Close the lid. Push the button. Record results as ppm Free Available Chlorine.
READ
Remove the tube and add 5 drops DPD #1A Free Chlorine Reagent (6740), and 5 drops of *DPD #1B Free Chlorine Reagent (6741).
6.
Cap and invert to mix. Wipe tube dry . Make readings within 30 seconds.
Push the button to turn the meter on. Press the
button and hold it for 2 seconds until is displayed. Release the button to take a zero reading
(0.00 ppm).
READ
ZERO
bla
Save sample for the Total Residual Chlorine T est. Proceed to Step 9.
Read the 1200 Colorimeter Manual before proceeding. Carefully wipe tubes dry before inserting into the colorimeter chamber.
FREE AVAILABLE CHLORINE
Page 35
er2
Subtract the Free Available Chlorine reading from the Total Residual Chlorine reading to determine ppm Combined Chlorine (Monochloramine, Dichloramine, and Nitrogen T richloride present in the water sample).
11.
9. 10.
12.
Push the
button. Record reading as ppm Total Residual Chlorine.
READ
Cap and invert to mix.**
Insert the tube into the chamber, being sure to align the index line with the arrow on the meter. Close the lid.
I200 COLORIMETER
••••••••••••••••••
READ
2.55
ZERO
Add 5 drops *DPD #3 T otal Chlorine Reagent (6743) to the sample from Step 7.
13.
TOTAL RESIDUAL CHLORINE
If either chlorine reading displays , repeat procedure on diluted sample, and multiply the result by the appropriate dilution factor. See 1200 Colorimeter Instruction Manual for procedure.
Levels of chlorine above 10 mg/L will cause a bleaching effect on the DPD indicator, and may give a false indication that no chlorine is present. If it is possible that the chlorine concentration is greater than 10 mg/L (e.g. after shock treatment), perform test on a diluted sample and multiply the result by appropriate dilution factor.
CAUTION: DO NOT leave reacted DPD samples in test tubes (0290). Discard sample and thoroughly rinse tubes. If allowed to remain, DPD will stain tubes, significantly impairing the operation of the 1200 Colorimeter. If necessary, acid wash, and vigorously clean glassware with test tube brush and detergent.
**For wastewater samples, Standard Methods for the Examination of Water and Wastewater recommends waiting 2 minutes for full color development.
Page 36
DPD CHLORINE TEST METHOD SPECIFICATIONS
APPLICATION
Drinking water supplies and distribution systems, swimming pool and spas, sewage and chlorinated waste waters, process waters and sanitizing solutions.
RANGE
0 to 4.0 mg/L Chlorine (may be extended by dilution)
METHOD
In the absence of Iodide, Free Available Chlorine reacts instantly with the buffered diethyl-p-phenylenediamine indicator (DPD) to produce a red color in proportion to the amount of chlorine present. Subsequent addition of potassium iodide produces a rapid color response from the combined forms of chlorine (chloramines).
HANDLING & PRESERVATION
Chlorine in aqueous solutions, particularly weak solutions, is not stable. Exposure to sunlight or agitation will accelerate the reduction of chlorine. Fill sample containers to the top and cap tightly. Analyze samples as soon as possible after collection.
INTERFERENCES
The only interfering substance likely to be encountered is oxidized manganese. The extent of this interference can be determined by treating a sample with sodium arsenite to destroy the chlorine present, so that the amount of interference can be measured.
CALIBRATION
The single test colorimeter is precalibrated. In order to comply with NPDWR or NPDES reporting regulations, the calibration should be checked periodically by using a set of reference standards including a 0 mg/L blank and 0.3, 1.0, and 4.0 mg/L chlorine. To prepare these standards, a LaMotte 250 ppm standard chlorine equivalent solution (Code 6973) is available. Consult with your local regulatory agency to determine standardization frequency.
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside U.S.A.) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
®
3670-01-LI · 08/08
Page 37
1200 COLORIMETER CHLORINE DIOXIDE
DPD LIQUID TEST
CODE 3671-01-LI
QUANTITY CONTENTS CODE
30 mL DPD #1A Free Chlorine Reagent P-6740-G 30 mL *DPD #1B Free Chlorine Reagent *P-6741-G 15 mL Glycine Solution 6811-E 1 Colorimeter Tubes, w/caps 0290-6 1 Water Sample Collecting Bottle 0688 1 1200 Colorimeter for Chlorine Dioxide 26731
*WARNING: Reagents marked with a * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or www .lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specified code number.
INTRODUCTION
Chlorine dioxide is used as a substiture for and an adjunct to chlorine in water treatment. It is better than chlorine in eliminating taste and odor in certain cases. Chlorine dioxide, unlike chlorine, does not produce carcinogenic chlorinated organic compounds when reacted with organic materials. A disadvantage is the higher cost of producing chlorine dioxide compared to chlorine.
1
Page 38
CHLORINE DIOXIDE TEST PROCEDURE - DPD METHOD
Remove tube. Add 5 drops Glycine Solution (6811). Cap and mix.
Cap and invert to mix. Wipe tube dry. Make readings within 30 seconds.
Fill the Water Sample Collecting Bottle (0688) with sample water. This will be used to dispense sample water for the tests.
Insert the tube into the chamber, being sure to align the index line with the arrow on the meter. Close the lid. This tube is the blank or zero.
Rinse and fill a colorimeter tube (0290) to the 10 mL line with sample water. Cap and wipe dry .
5.
7.
3.
1. 2.
4.
6.
I200 COLORIMETER
••••••••••••••••••
READ
bla
ZERO
Add 5 drops DPD #1A Free Chlorine Reagent (6740), and 5 drops of *DPD #1B Free Chlorine Reagent (6741).
Push the button to turn the meter on. Press the
button andholditfor2 seconds until is displayed. Release the button to take a zero reading
(0.00 ppm).
READ
ZERO
bla
8.
CAL
I200COLORIMETER
••••••••••••••••••
READ
Align the index line with the arrow on the meter, insert tube into chamber. Close the lid. Push the button. Record results as ppm Chlorine Dioxide.
READ
Read the 1200 Colorimeter Manual before proceeding. Carefully wipe tubes dry before inserting into the colorimeter chamber.
CHLORINE DIOXIDE
2
Page 39
NOTE: If reading displays , repeat procedure on diluted sample, and multiply the
er2
result by the appropriate dilution factor. See 1200 Colorimeter Instruction Manual for procedure.
CAUTION: DO NOT leave reacted DPD samples in test tubes (0290). Discard sample and thoroughly rinse tubes. If allowed to remain, DPD will stain tubes, significantly impairing the operation of the 1200 Colorimeter. If necessary, acid wash, and vigorously clean glassware with test tube brush and detergent.
CHLORINE DIOXIDE TEST METHOD SPECIFICATIONS
APPLICATION
Drinking and pool waters; domestic and industrial wastewater.
RANGE
0 to 7.0 ppm Chlorine Dioxide
METHOD
Chlorine dioxide reacts with DPD to form a red color in proportion to the concentration.
HANDLING & PRESERVATION
Test as soon as possible to avoid loss of chlorine dioxide.
INTERFERENCES
Chlorine interference is eliminated by the addition of glycine to the sample before the indicator.
3
Page 40
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
PO Box 329 • Chestertown • Maryland • 21620 • USA 800-344-3100 • 410-778-3100 • Fax 410-778-6394
Visit us on the web at www.lamotte.com
SM
63671-01-LI• 4/08
Page 41
1200 COLORIMETER
BROMINE
DPD TABLET METHOD
MODEL 1200-BR · CODE 3672-01
QUANTITY CONTENTS CODE
100 *Chlorine DPD #1 Tablets (Instrument Grade) *6903A-J 6 Colorimeter Tubes, w/caps 0290-6 1 Tablet Crusher 0175 1 Water Sample Collecting Bottle 0688 1 1200 Colorimeter for Bromine DPD 26730
*WARNING: Reagents marked with a * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or www .lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specified code number.
INTRODUCTION
Bromine is added to water to kill bacteria and other disease-producing organisms, control algae, and remove undesirable odors and colors. Bromine added to water, forms hypobromus acid, the sanitizing form of bromine. Bromine forms free available bromine and combined bromine or bromamines. Combined bromine is a very active sanitizer.
Page 42
BROMINE TEST PROCEDURE - DPD METHOD
Push the button to turn the meter on. Press the
button andholditfor2 seconds until is displayed. Release the button to take a zero reading
(0.00 ppm).
READ
ZERO
bla
Fill the W ater Sample Collecting Bottle (0688) with sample water. This will be used to dispense sample water for the tests.
Insert the tube into the chamber, being sure to align the index line with the arrow on the meter. Close the lid. This tube is the blankorzero.
Rinse and fill a colorimeter tube (0290) to the 10 mL line with sample water. Cap and wipe dry .
Align the index line with the arrow on the meter, insert tube into chamber . Close the lid. Push the button. Record results as ppm Bromine.
READ
CAL
I200COLORIMETER
••••••••••••••••••
READ
7.
5.
3.
1. 2.
4.
6.
I200
COLORIMETER
••••••••••••••••••
READ
ZERO
Empty all but a few drops of sample from tube. Add one *Chlorine DPD #1 Tablet (6903A) and crush with tablet crusher (0175).
To insure accurate results, tablet must be crushed before filling tube.
NOTE:
Fill to 10 mL line with sample. Cap and mix until tablet disintegrates. Make readings within 30 seconds after disintegration of the tablet.
bla
Read the 1200 Colorimeter Manual before proceeding. Carefully wipe tubes dry before inserting into the colorimeter chamber.
FREE AVAILABLE BROMINE
Page 43
DPD BROMINE TEST METHOD SPECIFICATIONS
APPLICATION
Drinking, surface, saline waters; swimming pool water; domestic and industrial waters and wastes.
RANGE
0to7.0mg/LBromine(maybeextendedbydilution)
METHOD
In buffered sample, bromine reacts with disthyl-p-phenylene diamine (DPD) to produce a pink-red color in proportion to the concentration of bromine present.
HANDLING & PRESERVATION
Bromine is aqueous solutions is not stable, and the bromine content of samples or solutions, particularily weak solutions, will rapidly decrease. Exposure to sunlight or agitation will acceslerater the reduction of bromine present in such solutions. For best results, start analysis immediately after sampling. Samples to be analyzed for bromine cannot be preserved or stored.
INTERFERENCES
The only interfering substance likely to be encountered in water is oxidized manganese. The extent of this interference can be dtermined by treating a sample with sodium arsenite to destroy the bromine present so that the degree of interference can be estimated.
Iodine and chlorine can also interfere, but these are not normally present unless they have been added as sanitizers.
Page 44
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside U.S.A.) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
®
63672-01 · 10/08
Page 45
1200 COLORIMETER

COPPER

MODEL 1200-CO · CODE 3673-01
QUANTITY CONTENTS CODE
30 mL *Copper 1 *6446-G
1 Colorimeter Tubes, w/caps 0290-6
1 Water Sample Collecting Bottle 0688
1 1200 Colorimeter for Copper 26732
*WARNING: Reagents marked with a * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or www .lamotte.com. To obtain a printed copy, contact LaMotte by email, phone or fax.
To order individual reagents or test kit components, use the specified code number.
INTRODUCTION
The copper content of drinking water generally falls below 0.03 parts per million, but copper levels as high as 1.0 part per million will give water a bitter taste. W aters testing as high as 1.0 part per million copper have probably been treated with a copper compound, like those used in the control of algae, or have become contaminated from untreated industrial wastes. The addition of copper sulfate to lakes causes an increase in the copper content of the sediments. Acid waters and those high in free carbon dioxide may cause the corrosion or “eating away” of copper, brass, and bronze pipes and fittings. This corrosion results in the addition of copper into the water supply.
Page 46
COPPER TEST PROCEDURE - DIETHYLDITHIOCARBAMATE
Fill the Water Sample Collecting Bottle (0688) with sample water. This will be used to dispense sample water for the tests.
Insert the tube into the chamber, being sure to align the index line with the arrow on the meter. Close the lid. This tube is the blank or zero.
Rinse and fill a colorimeter tube (0290) to the 10 mL line with sample water. Cap and wipe dry .
5.
3.
1. 2.
4.
6.
I200 COLORIMETER
••••••••••••••••••
READ
bla
ZERO
7.
Align the index line with the arrow on the meter , insert tube
CAL
I200COLORIMETER
••••••••••••••••••
READ
into chamber. Close the lid. Push the button. Record resultsasppm Copper.
READ
Cap and invert to mix. Wipe tube dry .
Remove tube from colorimeter. Add 5 drops of *Copper 1 (6446).
Push the
button to turn the meter on. Press the
button and hold it for 2 seconds until is displayed. Release the button to take a blank reading (0.0 ppm).
READ
ZERO
bla
L
Mott
L
Mott
Read the 1200 Colorimeter Manual before proceeding. Carefully wipe tubes dry before inserting into the colorimeter chamber.
COPPER
Page 47
NOTE: If test reading displays , repeat procedure on diluted sample, and multiply
er2
the result by the appropriate dilution factor. See 1200 Colorimeter Instruction Manual for procedure.
COPPER TEST METHOD SPECIFICATIONS
APPLICATION
Drinking, surface, and saline waters; domestic and industrial wastes.
RANGE
0 to 5.0 ppm Copper
METHOD
Cupric ions form a yellow colored chelate with diethyldithiocarbamate around pH 9-10 in proportion to the concentration of copper in the sample.
HANDLING & PRESERVATION
Copper has a tendency to be absorbed to the surface of the sample container. Samples should be analyzed as soon as possible after collection. If storage is necessary, 0.5 mL of 20% hydrochloric acid per 100 mL of sample will prevent “plating out.” However, a correction must be made to bring the reaction into the optimum pH range.
INTERFERENCES
Bismuth, cobalt, mercurous, nickel and silver ions and chlorine (6 ppm or greater) interfere and must be absent.
Page 48
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside USA) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
®
3673-01 · 6/08
Page 49
1200 COLORIMETER
EPA
COMPLIANT

FLUORIDE

SPADNS METHOD
MODEL 1200-FL · CODE 3674-01
QUANTITY CONTENTS CODE
2 x 100 mL *Acid-Zirconyl-SPADNS Reagent *3875-J 2 x 60 mL *Sodium Arsenite Solution *4128-H 1 Pipet, 0.5 mL, plastic 0353 1 Pipet, 1.0 mL, plastic 0354 1 Colorimeter Tubes, with caps, set of 6 0290-6 1 Water Sample Collecting Bottle 0688 1 1200 Colorimeter for Fluoride 26733
*WARNING: Reagents marked with a * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or www .lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specified code number.
NOTE: This procedure uses EPA approved SPADNS Reagent System for fluoride found in method 4500-F-D, 18th Edfition of Standard Methods, page 1-27.
FLUORIDE INTRODUCTION
Fluoride may occur naturally in some ground waters or it may be added to public drinking water supplies to maintain a 1.0 mg/L concentration to prevent dental cavities. At higher concentrations, fluoride may produce an objectionable discoloration of tooth enamel called fluorosis, though levels up to 8 mg/L have not been found to be physiologically harmful.
Page 50
FLUORIDE TEST PROCEDURE - SPADNS METHOD
Insert the tube into the chamber, being sure to align the index line with the arrow on the meter. Close the lid. This tube is the sample blank.
5.
Use the 0.5 mL pipet (0353) to add 0.5 mL of *Sodium Arsenite Solution (4128). Cap and mix.
3.
This test requires a reagent blank. Rinseaclean colorimeter tube (0290) with clear, colorless, fluoride free water. Fill to the 10 mL line with clear, colorless, fluoride free water.
2.
Fill the Water Sample Collecting Bottle (0688) with sample water. This will be used to dispense sample water for the tests.
1.
Use the 1.0 mL pipet (0354) to add 2 measures of *Acid-Zirconyl SP ADNS Reagent (3875). Cap and mix thoroughly . (This is the reagent blank.
4.
6.
Push the
button to turn the meter on. Press the
button andholditfor2 seconds until is displayed. Release the button to take a blank reading (0.0 ppm).
READ
ZERO
bla
I200 COLORIMETER
••••••••••••••••••
READ
ZERO
bla
L
Mott
Read the 1200 Colorimeter Manual before proceeding. Carefully wipe tubes dry before inserting into the colorimeter chamber.
FLUORIDE
Page 51
FLUORIDE
9.
10 .
Align the index line with the arrow on the meter, insert tube into chamber. Close the lid. Push the
button. Record results as ppm Fluoride.
READ
CAL
I200COLORIMETER
••••••••••••••••••
READ
18.0
Use the 1.0 mL pipet (0354) to add 2 measures of *Acid-Zirconyl SP ADNS Reagent (3875). Cap and mix thoroughly.
8.7.
Rinseaclean colorimeter tube (0290) with sample water. Fill to the 10 mL line with sample water.
Use the 0.5 mL pipet (0353) to add 0.5 mL of *Sodium Arsenite Solution (4128). Cap and mix.
L
Mott
er2
NOTE: Zeroing the meter with sample water or an empty chamber will result in an message. Meter must be zeroed with a reagent blank.
Page 52
FLUORIDE SPADNS TEST METHOD SPECIFICA TIONS
APPLICATION
Drinking and surface waters; domestic and industrial waters.
RANGE
0.0 to 2.0 ppm Fluoride
METHOD
Colorimetric test based upon the reaction between fluoride and zirconium dye lake. The fluoride reacts with the dye lake, dissociating a portion of it into a colorless complex ion and dye. As the fluoride concentration increases, the color produced becomes progressively lighter.
HANDLING & PRESERVATION
Samples may be stored and refrigerated in plastic containers.
INTERFERENCES
The following substances produce a positive interference at the concentration given:
Chloride (Cl
Phosphate (PO
Hexametaphosphate (NaPO
-
) 7000 mg/L
-3
)16mg/L
4
1mg/L
3)6
The following substances produce a negative interference at the concentration given:
Alkalinity (CaCO
Aluminum (Al
+3
Iron (Fe
)10mg/L
Sulfate (SO
) 5000 mg/L
3
+3
)0.1mg/L
-2
) 200 mg/L
4
Color and turbidity must be removed or compensated for in the procedure. Temperature should be maintained within 5 degrees Celcius of room temperature.
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside USA) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
®
03.08
Page 53
1200 COLORIMETER

MOL YBDENUM

MODEL 1200-MO • CODE 3676-01
QUANTITY CONTENTS CODE
60 mL *Mo Buffer *3997-H 2 x 30 mL *Molybdenum Oxidizing Reagent *6485-G
2.5 g *Molybdenum Indicator Powder *6486-S 1 Colorimeter Tubes 10 mL, w/cap, set of 6 0290-6 1 Spoon, 0.05g, plastic 0696 1 Pipets, 1.0 mL, plastic, w/cap 0372 1 Pipet, 1.0 mL 0354 1 Water Sample Collecting Bottle 0688 1 1200 Colorimeter for Molybdenum 26729
*WARNING: Reagents marked with a * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or www .lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specified code number. NOTE: After use, return *MO Buffer (3997) to zipper top storage bag to reduce exposure to
corrosive reagent.
INTRODUCTION
Molybdenum occurs naturally in the earth’s crust as molybdenite and wolfenite, and is an important element in many biochemical reactions, including nitrogen fixation. In industrial processes, such as the operation of boilers and cooling towers, molybdenum, in the form of sodium molybdate, is used as an environmentally safe corrosion inhibitor.
1
Page 54
MOLYBDENUM PROCEDURE - Thioglycolate Method
8.
Fill the Water Sample Collecting Bottle (0688) with sample water. This will be used to dispense sample water for the tests.
Rinse and fill a colorimeter tube (0290) to the 10 mL line with sample water.
Use the 0.05 spoon (0696) to add 0.05g of *Molybdenum Indicator Powder (6486).
1. 2.
Insert the tube into the chamber, being sure to align the index line with the arrow on the meter. Close the lid. This tube is the sample blank.
6.
5.
Wipe the tube dry.
Use the 1.0 mL pipet (0354) to add1mLof *Molybdenum Oxidizing Reagent (6485). Cap and mix.
4.
Use the 1.0 mL pipet (0372) to add1mLof*Mo Buffer (3997). Cap and mix.
3.
7.
Push the button to turn the meter on. Press the
button andholditfor2 seconds until is displayed. Release the button to take a blank reading (0.0 ppm).
READ
ZERO
bla
I200 COLORIMETER
••••••••••••••••••
READ
ZERO
bla
L
Mott
Read the 1200 Colorimeter Manual before proceeding. Carefully wipe tubes dry before inserting into the colorimeter chamber
MOLYBDENUM
2
Page 55
NOTE: If the molybdenum reading exceeds 50 ppm, repeat procedure on diluted sample,
9.
10 .
Wipe the tube dry .
Cap and mix until the powder dissolves. Solution will turn yellow if molybdenum is present.
Align the index line with the arrow on the meter, insert tube into chamber. Close the lid. Push the button. Record results as ppm Molybdenum.
READ
CAL
I200COLORIMETER
••••••••••••••••••
READ
18.0
11.
L
Mott
and multiply the result by the appropriate dilution factor. See 1200 Colorimeter Instruction Manual for procedure.
3
Page 56
MOLYBDENUM TEST METHOD SPECIFICATIONS
APPLICATION
Boiler and cooling water
RANGE
0 to 30 ppm Molybdenum
METHOD
Calcium thioglycolate reacts with molybdenum to give a yellow color with an intensity proportional to the amount of molybdenum present.
HANDLING & PRESERVATION
Molybdenum samples may be stored in either plastic or glass containers.
INTERFERENCES
Nickel levels less than 50 ppm do not interfere; aluminum levels less than 10 ppm do not interfere; chromate, at higher concentrations, interferes due to the intense yellow color. Ferrous iron levels below 50 ppm do not interfere, but low levels of ferric iron will cause a large blank. Highly buffered samples may exceed the capacity of the system possibly producing inaccurate results. Add an extra 1.0 mL of *Mo Buffer (3997) to adjust the pH of the sample to approximately 4.5. Multiply the result in ppm by 1.08.
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside U.S.A.) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
®
06/08
Page 57
1200 COLORIMETER

NITRATE NITROGEN

MODEL 1200-NA · CODE 3677-01
QUANTITY CONTENTS CODE
2 x 120 mL *Mixed Acid Reagent *V-6278-J
10 g *Nitrate Reducing Reagent *V-6279-D
1 Spoon, 0.1 g, plastic 0699
1 Graduated Cylinder 0416
1 Colorimeter Tubes, w/caps 0290-6
1 Water Sample Collecting Bottle 0688
1
*WARNING: Reagents marked with a * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or www.lamotte.com. To obtain a printed copy , contact LaMotte by email, phone or fax.
To order individual reagents or test components, use the specified code number.
NITRATE - NITROGEN INTRODUCTION
Nitrogen is essential for plant growth, but the presence of excessive amounts in water supplies presents a major pollution problem. Nitrogen compounds may enter water as nitrates or be converted to nitrates from agricultural fertilizers, sewage, industrial and packing house wastes, drainage from livestock feeding areas, farm manures and legumes. Nitrates in large amounts can cause “blue babies” (methemoglobinemia) in infants less than six months of age. Nitrate concentration is an important factor to be considered in livestock products, where, in addition to causing methemoglobinemia, it is responsible for many other problems. Nitrates in conjunction with phosphate stimulate the growth of algae with all of the related difficulties associated with excessive algae growth.
U.S. Public Health Service Drinking Water Standards state that 10 ppm nitrate nitrogen should not be exceeded. To the sanitary and industrial engineer, concentrations of less than 1 ppm are acceptable.
1200 Colorimeter for Nitrate Nitrogen
26734
1
Page 58
NITRATE-NITROGEN TEST PROCEDURE -
Insert the tube into the chamber, being sure to align the index line with the arrow on the meter. Close the lid. This tube is the blankorzero.
8.
5.
3. 4.
Fill the Water Sample Collecting Bottle (0688) with sample water. This will be used to dispense sample water for the tests.
Rinse and fill a colorimeter tube (0290) to the 10 mL line with sample water. Cap and wipe dry .
1. 2.
I200 COLORIMETER
••••••••••••••••••
READ
bla
ZERO
7.
Remove tube from colorimeter and pour off 5 mL into graduated cylinder or similar. Discard the remaining sample.
6.
Pour the 5 mL sample from a graduated cylinder or similar into the colorimeter tube.
Push the button to turn the meter on. Press the
button andholditfor2 seconds until is displayed. Release the button to take a zero reading
(0.00 ppm).
READ
ZERO
bla
Use the graduated cylinder or similar to measure 5 mL of *Mixed Acid Reagent (V-6278) and add to tube.
5mL
5mL
2
MINUTES
5mL
Cap and mix. Wait approximately 2 minutes before proceeding to Step 9.
L
Mott
CADMIUM REDUCTION METHOD
Read the 1200 Colorimeter Manual before proceeding. Carefully wipe tubes dry before inserting into the colorimeter chamber.
NITRATE - NITROGEN
2
Page 59
NITRATE - NITROGEN
NOTE:
11.
9. 10.
NOTE:
At the end of waiting period, an undissolved portion of *Nitrate Reducing Reagent may remain in bottom of the tube without affecting results.
Use the 0.1 g spoon (0699) to add two measures of *Nitrate Reducing Reagent (V -6279). Cap.
Hold tube by index finger and thumb and mix by
12.
inverting approximately
times a minute for
. Wait
for maximum
color development.
50-60 4 minutes 10 minutes
10
MINUTES
Wipe tube dry.
Align the index line with the arrow on the meter, insert
CAL
I200COLORIMETER
••••••••••••••••••
READ
tube into chamber. Close the lid. Push the button. Record results as ppm Nitrate Nitrogen.
READ
To convert Nitrate Nitrogen (NO -N) results to ppm Nitrate (NO ), multiply by 4.4.
3 3
L
Mott
3
Page 60
NITRATE-NITROGEN TEST METHOD SPECIFICATIONS
APPLICATION
This method determines nitrate levels in drinking, surface, saline waters, domestic, and industrial waters.
RANGE
0 - 3.0 ppm Nitrate Nitrogen (Range can be extended by dilution.)
METHOD
Powdered cadmium is used to reduce nitrate to nitrite. The nitrite that is originally present plus reduced nitrate is determined by diazotization of sulfanilamide and nitrite followed by coupling with N-(1 naphthyl) -ethylenediamine dihydrochloride to form a highly colored azo dye which is measured colorimetrically.
HANDLING & PRESERVATION
Analysis should be made as soon as possible. If analysis cannot be made within 24 hours, the sample should be preserved by refrigeration at 4°C. When samples must be stored for more than 24 hours, they can be preserved by adding 2 mL of concentrated sulfuric acid per liter of sample. For best results, the analysis for nitrate should be determined at temperatures between 20°C and 25°C.
INTERFERENCES
Nitrite interferences at all levels. Strong oxidizing and reducing substances interfere. Low results might be obtained for samples that contain high concentrations of iron and copper.
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside U.S.A.) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
®
06.08 #63677-01
Page 61
1200 COLORIMETER

OZONE

MODEL 1200-OZ · CODE 3678-01
QUANTITY CONTENTS CODE
15 mL Chlorine Inhibitor 3990-E
250 mL *Ozone Buffer *3991-K
30 mL Indigo Blue Stock Solution 3989-G
1 Bottle, HR Reagent, amber glass 0680-J
1 Sampling Apparatus 0681
1 Pipet, transfer, 1.0 mL 2-2170
1 Pipet, 5 mL, glass volumetric 0329
1 Pipet Pump, 10 mL 2-2216
1 Graduated Cylinder, 50 mL, glass 0418
1 Colorimeter Tubes, w/caps 0290-6
1 1200 Colorimeter for Ozone 26735
*WARNING: Reagents marked with a * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or www.lamotte.com. To obtain a printed copy, contact LaMotte by email, phone or fax.
To order individual reagents or test kit components, use the specified code numbers.
INTRODUCTION
Ozone is sometimes used in place of, or in conjunction with, chlorine or other halogens for disinfection of pool, spa or drinking waters. Recently, large aquatic facilities have begun using ozone as a disinfectant in many artificial habitats.
Page 62
OZONE TEST PROCEDURE - INDIGO METHOD
Use the 1.0 mL transfer pipet (2-
2170) and pipette pump (2-2216) to add 1.0 mL of HR Reagent to each of 2 clean colorimeter tubes (0967).
If chlorine is present, add 3 drops Chlorine Inhibitor (3990) to each tube. Cap tubes.
8.
5.
3. 4.
Use the 50 mL graduated cylinder (0418) to carefully add 45 mL of *Ozone Buffer (3991) to amber glass bottle marked HR Reagent (0680).
Use the 5 mL volumetric pipet (0329) and pipette pump (2-2216) to add 5 mL of Indigo Blue Stock Solution (3989) to the amber glass bottle. Cap and mix.
1. 2.
7.
Take one of the prepared colorimeter tubes and sampling apparatus (0681) to sampling site.
6.
Lower the end of the tubing of sampling apparatus to desired depth. Slowly withdraw and depress plunger several times to purge syringe and tubing. Slowly withdraw plunger to fill purged syringe.
Remove plastic tubing from syringe. Remove
Fill the second prepared tube (0967) to the 10 mL line with ozone free water. This is the Reagent Blank.
HR Reagent
cap from the prepared tube. Place tip of syringe against inside of the prepared tube. Slowly depress plunger and fill to the 10 mL line and cap. This is the Sample Tube. NOTE: Do not shake or invert the sample.
Read the 1200 Colorimeter Manual before proceeding. Carefully wipe tubes dry before inserting into the colorimeter chamber.
PROCEDURE
2
Page 63
Push the button to turn the meter on. Press the
button and hold it for 2 seconds until is displayed. Release the button to take a zero reading (0 ppm).
READ
ZERO
bla
13. Note:
11.
9. 10.
12.
Wipe both tubes clean with a lint-free cloth.
Insert the Reagent Blank into the chamber, being sure to align the index line with the arrow on the meter. Close the lid. This tube is the blank or zero.
Insert the Sample T ube into the chamber, being sure to align the index line with the arrow on the meter. Close the lid.
The HR Reagent must be made fresh each week. If reagent is refrigerated, it may be kept up to 3 weeks.
I200 COLORIMETER
••••••••••••••••••
READ
bla
ZERO
Push the
button. The concentration in ppm will be displayed within 2 seconds.
READ
ZERO
I200
COLORIIMETER
••••••••••••••••••
READ
0.15
L
Mott
L
Mott
er2
NOTE: Zeroing the meter with sample water or an empty chamber will result in an
message when reading reacted samples. Meter must be zeroed with a reagent
blank.
3
Page 64
OZONE TEST METHOD SPECIFICATIONS
APPLICATION
Drinking, pool, and aquatic waters.
RANGE
0.0 - 0.5 ppm Ozone
METHOD
Ozone rapidly and stoichiometrically decolorizes Indigo Trisulfonate under acidic conditions.
HANDLING & PRESERVATION
Ozone is extremely unstable in aqueous solutions. Test must be performed immediately and the sample must not be agitated.
INTERFERENCES
Manganese at any level will interfere.
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
PO Box 329 · Chestertown · Maryland · 21620 · USA
800-344-3100 · 410-778-3100 (Outside USA) · Fax 410-778-6394
Visit us on the web at www.lamotte.com
®
1/09
Page 65
1200 COLORIMETER
L
Mott

PHOSPHATE

MODEL 1200-PLR · CODE 3679-01
QUANTITY CONTENTS CODE
2 x 60 mL *Phosphate Acid Reagent *V-6282-H 10 g *Phosphate Reducing Reagent *V-6283-D 6 Colorimeter Tubes, 10 mL, w/cap 0290 1 Spoon, 0.1g, plastic 0699 1 Water Sample Collecting Bottle 0688 1 Pipet, 1.0 mL, plastic 0354 1 1200 Colorimeter for Phosphate 26736
WARNING: Reagents marked with a * are considered to be potential health hazards. T o view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or www .lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specified code number.
INTRODUCTION
Phosphorus is an important nutrient for aquatic plants. The amount found in water is generally not more than 0.1 ppm unless the water has become polluted from waste water sources or excessive drainage from agricultural areas. When phosphorus is present in excess of the concentrations required for normal aquatic plant growth, a process called eutrophication takes place. This creates a favorable environment for an increase in algae and weed nuisances. When algae cells die, oxygen is used in the decomposition and fish kills often result. Rapid decomposition of dense algae scums with associated organisms give rise to foul odors and hydrogen sulfide gas.
Page 66
PHOSPHATE PROCEDURE - ASCORBIC ACID METHOD
4.
Push the
button to turn the meter on. Press the
button andholditfor2 seconds until is displayed. Release the button to take a blank reading (0.0 ppm).
READ
ZERO
bla
I200 COLORIMETER
••••••••••••••••••
READ
ZERO
bla
Insert the tube into the chamber, being sure to align the index line with the arrow on the meter . Close the lid. This tube is the sample blankorzero.
3.
Rinse and fill a colorimeter tube (0290) to the
10 mL line with sample water. Cap and wipe dry .
2.
Fill the W ater Sample Collecting Bottle (0688) with sample water. This will be used to dispense sample water for the tests.
1.
Remove tube from colorimeter. Use 1.0 mL pipet (0354) to add
1.0 mL of *Phosphate Acid Reagent
(V -6282). Cap and mix.
5.
6.
Use the 0.1 g spoon (0699) to add one measure of *Phosphate Reducing Reagent (V -6283).
8.
Align the index line with the arrow on the meter, insert tube
CAL
I200COLORIMETER
••••••••••••••••••
READ
into chamber. Close the lid. Push the
button. Record results as ppm Orthophosphate.
READ
7.
Cap and shake until powder dissolves. Wait 5 minutes for full color development. Solution will turn blue if phosphates are present. Wipe tube dry.
L
Mott
L
Mott
Read the 1200 Colorimeter Manual before proceeding. Carefully wipe tubes dry before inserting into the colorimeter chamber.
PHOSPHATE
Page 67
PHOSPHATE TEST METHOD SPECIFICATIONS
APPLICATION
Drinking, surface and saline waters; domestic and industrial wastes (Method based on reactions that are specific for orthophosphate).
RANGE
0 to 3.0 ppm Orthophosphate (Range can be extended by dilution.)
METHOD
Ammonium molybdate and antimony potassium tartrate react in a filtered acid medium with dilute solution of PO
3-
to form an antimony-phosphomolybdate complex. This
4
complex is reduced to an intense blue colored complex by ascorbic acid. The color is proportional to the amount of phosphate present. (Only orthophosphate forms a blue color in this test.) Polyphosphates (and some organic phosphorus compounds) may be converted to the orthophosphate form by sulfuric acid digestion. Organic phosphorus compounds may be converted to the orthophosphate form by persulfate digestion.
HANDLING & PRESERVATION
If benthic deposits are present in the area being sampled, great care should be taken not to include these deposits. If the analysis cannot be performed the same day of collection, the sample should be preserved by the addition of 2 mL of concentrated sulfuric acid or 40 mg mercuric chloride per liter and refrigerated at 4°C.
INTERFERENCES
1. No interference from copper , iron, or silicate at concentrations many times the concentration of sea water. However, high iron concentrations can cause precipitation and subsequent loss of phosphorus.
2. Salt error for samples ranging from 5% to 20% salt content was found to be less than 1%.
3. Mercuric chloride, HgCl
, when used as the preservative, interferes when the
2
chloride levels are low (less than 50mg/L). This interference is overcome by spiking samples with a minimum of 50 mg/L of sodium chloride.
Page 68
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside USA) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
®
3679-01 • 10.08
Page 69
1200 COLORIMETER

AMMONIA-NITROGEN

MODEL 1200-NH · CODE 3680-01
QUANTITY CONTENTS CODE
30 mL Ammonia Nitrogen Reagent #1 V-4797-G 3 x 30 mL *Ammonia Nitrogen Reagent #2 *V-4798-G 1 Pipet, 1 mL, plastic 0354 1 Colorimeter Tubes, with caps 0290-6 1 Water Sample Collecting Bottle 0688 1 1200 Colorimeter for Ammonia Nitrogen 26737
*WARNING: Reagents marked with a * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or www .lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specified code number.
INTRODUCTION
Ammonia nitrogen is present in various concentrations in many surface and ground water supplies. Any sudden change in the concentration of ammonia nitrogen in a water supply is cause for suspicion. A product of microbiological activity, ammonia nitrogen is sometimes accepted as chemical evidence of pollution when encountered in natural waters.
Ammonia is rapidly oxidized in natural water systems by special bacterial groups that produce nitrite and nitrate. This oxidation requires that dissolved oxygen be available in the water. Ammonia is an additional source of nitrogen as a nutrient which may contribute to the expanded growth of undesirable algae and other forms of plant growth that overload the natural system and cause pollution.
1
Page 70
AMMONIA NITROGEN TEST PROCEDURE:
sure to align the index line with the arrow on the meter. Close the lid. This tube is the blank or zero.
3.
Rinse and fill a colorimeter tube (0290) to the 10 mL line with sample water. Cap and wipe dry.
2.
Fill the W ater Sample Collecting Bottle (0688) with sample water. This will be used to dispense sample water for the tests.
1.
6.
Use 1.0 mL pipet (0354) to add 1.0 mL of *Ammonia Nitrogen Reagent #2 (V -4798).
5.
Remove tube from colorimeter. Add 8 drops of Ammonia Nitrogen Reagent #1 (V-4797). Cap and mix.
4.
the button and hold it for 2 seconds until
is displayed. Release the button to take a blank reading (0.0 ppm).
ZERO
bla
I200
COLORIMETER
••••••••••••••••••
READ
ZERO
bla
7.
Cap and invert to mix. Wait 5 minutes for full color development. Wipe tube dry.
5
MINUTES
8.
Align the index line with the arrow on the meter, insert tube
into chamber. Close the lid. Push the button. Record results as ppm Ammonia Nitrogen (NH -N).
READ
3
CAL
I200
COLORIMETER
••••••••••••••••••
READ
L
Mott
L
Mott
Insert the tube into the chamber, being
Push the button to turn the meter on. Press
READ
NESSLER METHOD
Read the 1200 Colorimeter Manual before proceeding. Carefully wipe tubes dry before inserting into the colorimeter chamber.
AMMONIA NITROGEN
2
Page 71
NOTE: For the best possible results, carry a reagent blank through the procedure. After scanning the blank in Step 4, perform the test procedure on clear, colorless, distilled or deioized water. Subtract results of regent blank from all subsequent test results.
NOTE: If the reading displays
Er2, repeat procedure on diluted sample, and multiply
the result by the appropriate dilution factor. See 1200 Colorimeter Instruction Manual for procedure.
CALCULATIONS
To express results as Unionized Ammonia (NH
Unionized Ammonia (NH
To express results as Ionized Ammonia (NH
Ionized Ammonia (NH
) = ppm Ammonia Nitrogen (NH3-N) x 1.2
3
+
) = ppm Ammonia Nitrogen (NH3-N) x 1.3
4
Ammonia in water occurs in two forms: toxic unionized ammonia (NH relatively non-toxic ionized form, ammonium ion (NH both forms as ammonia-nitrogen (NH concentration in water. The actual proportion of each compound depends on
-N) to give the total ammonia-nitrogen
4
):
3
+
):
4
)andthe
+
). This test method measures
4
3
temperature, salinity, and pH. A greater concentration of unionized ammonia is present when the pH value and salinity increase.
1. Consult the table below to find the percentage that corresponds to the temperature, pH and salinity of the sample.
2. To express the test result as ppm Unionized Ammonia Nitrogen (NH the total ammonia-nitrogen test result by the percentage from the table.
3. To express the test result as ppm Ionized Ammonia Nitrogen (NH the unionized ammonia-nitrogen determined in Step 2 from the total ammonia
-N), multiply
3
+
-N), subtract
4
nitrogen.
10°C 15°C 20°C 25°C
pH FW
1
SW
2
FW SW FW SW FW SW
7.0 0.19 0.27 0.40 0.55
7.1 0.23 0.34 0.50 0.70
7.2 0.29 0.43 0.63 0.88
7.3 0.37 0.54 0.79 1.10
7.4 0.47 0.68 0.99 1.38
7.5 0.59 0.459 0.85 0.665 1.24 0.963 1.73 1.39
7.6 0.74 0.577 1.07 0.836 1.56 1.21 2.17 1.75
7.7 0.92 0.726 1.35 1.05 1.96 1.52 2.72 2.19
7.8 1.16 0.912 1.69 1.32 2.45 1.90 3.39 2.74
7.9 1.46 1.15 2.12 1.66 3.06 2.39 4.24 3.43
8.0 1.83 1.44 2.65 2.07 3.83 2.98 5.28 4.28
8.1 2.29 1.80 3.32 2.60 4.77 3.73 6.55 5.32
8.2 2.86 2.26 4.14 3.25 5.94 4.65 8.11 6.61
8.3 3.58 2.83 5.16 4.06 7.36 5.78 10.00 8.18
8.4 4.46 3.54 6.41 5.05 9.09 7.17 12.27 10.10
8.5 5.55 4.41 7.98 6.28 11.18 8.87 14.97 12.40
1
Freshwater data from Trussel (1972).
2
Seawater values from Bower and Bidwell (1978). Salinity for the Seawater values =
34% at an ionic strength of 0.701 m.
3
Page 72
FOR EXAMPLE:
A fresh water sample at 20°C has a pH of 8.5 and the test result is 1.0 ppm as total Ammonia-Nitrogen.
1. The percentage from the table is 11.18% (or 0.1118).
2. 1 ppm total Ammonia-Nitrogen x 0.1118 = 0.1118 ppm Unionized
Ammonia-Nitrogen
3.
Total Ammonia-Nitrogen 1.0000 ppm Unionized Ammonia-Nitrogen - 0.1118 ppm Ionized Ammonia-Nitrogen = 0.8882 ppm
AMMONIA NITROGEN TEST METHOD SPECIFICATIONS
APPLICATION
Drinking , surface, and saline waters; domestic and industrial wastes.
RANGE
0 to 5.0 ppm Ammonia Nitrogen
METHOD
Ammonia forms a colored complex with Nessler’s Reagent in proportion to the amount of ammonia present in the sample. Rochelle salt is added to prevent precipitation of calcium or magnesium in undistilled samples.
HANDLING & PRESERVATION
Preservation is accomplished by the addition of 2 mL of concentrated H
2SO4
at 4°C.
INTERFERENCES
Sample turbidity and color may interfere. Turbidity may be removed by a filtration procedure. Color interference may be eliminated by adjusting the instrument to 100%T withasampleblank.
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside U.S.A.) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
®
3680-01 · 6.07
Page 73
1200 COLORIMETER

IRON

MODEL 1200-FE · 3681-01
QUANTITY CONTENTS CODE
2 x 15 mL *Acid Phenanthroline Indicator *2776-E 2x5g *IronReducingReagent *2777-C 1 Colorimeter Tubes, with caps, set of 6 0290-6 1 Spoon, 0.1 g, plastic 0699 1 Water Sample Collecting Bottle 0688 1 1200 Colorimeter for Iron 26738
*WARNING: Reagents marked with a * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or www .lamotte.com. To obtain a printed copy, contact LaMotte by email, phone or fax.
To order individual reagents or test kit components, use the specified code number.
INTRODUCTION
Most natural waters contain some iron. The concentration may vary from small traces to very large amounts in water which is contaminated by acid mine wastes. For domestic use, the concentration should not exceed 0.2 ppm and for some industrial applications not even a trace of iron can be tolerated. There are many means available for removing or reducing the iron content. W a ter softening resins are effective for removing small amounts of iron and special ion exchange materials are selective for iron removal. High concentrations of iron can be removed by such chemical processes as oxidation and lime or lime-soda softening. Because of the many means of removing or reducing the amount of iron in water , the particular method employed will depend largely on the form of iron which is present and the end use of the treated water.
1
Page 74
IRON TEST PROCEDURE - 1,10 PHENANTHROLINE METHOD
Fill the Water Sample Collecting Bottle (0688) with sample water. This will be used to dispense sample water for the tests.
Insert the tube into the chamber, being sure to align the index line with the arrow on the meter. Close the lid. This tube is the blankorzero.
Rinse and fill a colorimeter tube (0290) to the 10 mL line with sample water. Cap and wipe dry .
5.
3.
1. 2.
4.
6.
I200 COLORIMETER
••••••••••••••••••
READ
bla
ZERO
Cap and invert the tube 15 times to mix.
Push the button to turn the meter on. Press the
button andholditfor2 seconds until is displayed. Release the button to take a zero reading
(0.00 ppm).
READ
ZERO
bla
Remove tube from colorimeter. Use the
0.1 g spoon (0699) to add one measure of *Iron Reducing Reagent (2777).
8.
7.
Remove the cap and add 6 drops of *Acid Phenanthroline Indicator (2776).
Cap and invert the tube 3 times to mix reagents.
L
Mott
Read the 1200 Colorimeter Manual before proceeding. Carefully wipe tubes dry before inserting into the colorimeter chamber.
IRON
2
Page 75
If the reading displays Er2, repeat procedure on diluted sample, and multiply the result
9. 10.
Wait 5 minutes for full color development. Solution will turn orange if iron is present.
Wipe tube dry.
11.
Align the index line with the arrow on the meter, insert
CAL
I200COLORIMETER
••••••••••••••••••
READ
tube into chamber. Close the lid. Push the button. Record results as ppm Iron.
READ
5
MINUTES
L
Mott
by the appropriate dilution factor. See 1200 Colorimeter Instruction Manual for procedure.
3
Page 76
IRON TEST METHOD SPECIFICATIONS
APPLICATION
Drinking , surface, and saline waters; domestic and industrial wastes.
RANGE
0to4.0ppmIron
METHOD
Ferric iron is reduced to ferrous iron and subsequently forms a colored complex with phenanthroline for a quantitative measure of total iron.
HANDLING & PRESERVATION
The sample container should be cleaned with acid and rinsed with deionized water. Addition of acid to adjust the sample to pH 2 - 3 will prevent deposition of iron on the container walls. Samples should be analyzed as soon as possible after collection since ferrous iron undergoes oxidation to ferric iron.
INTERFERENCES
Strong oxidizing agents, cyanide, nitrite, phosphates, chromium, and zinc in concentrations exceeding 10 times that of iron; cobalt and copper in excess of 5 mg/L, and nickel in excess of 2 mg/L. Bismuth, cadmium, mercury, molybdate, and silver precipitate phenanthroline.
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside U.S.A.) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
®
63681-01 · 4/09
Page 77
1200 COLORIMETER

MANGANESE

MODEL 1200-MN · CODE 3682-01
QUANTITY CONTENTS CODE
2 x 100 mL *Hardness Buffer Reagent *4255-J 2 x 30 mL *Manganese Indicator Reagent *3956-G 15 mL *Sodium Cyanide, 10% *6565-E 1 Pipet, 0.5 mL, plastic, w/cap 0369 1 Pipet, 1.0 mL, plastic 0354 6 Colorimeter Tubes, with caps 0290-6 1 Water Sample Collecting Bottle 0688 1 1200 Colorimeter for Manganese 26739
*WARNING: Reagents marked with a * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or www .lamotte.com. To obtain a printed copy, contact LaMotte by email, phone or fax.
To order individual reagents or test kit components, use the specified code number.
INTRODUCTION
Manganese is present in ground water in the divalent state due to the lack of oxygen. In surface waters, manganese may be in various oxidation states as soluble complexes or as suspended compounds. Manganese is rarely present in excess of 1 mg/L. It may cause an objectionable taste or cause staining problems in laundry, but manganese levels normally encountered in water seldom produce any health hazard.
Manganese is removed from water by various means including chemical precipitation, pH adjustment, aeration, superchlorination and the use of ion exchange resins.
Page 78
MANGANESE TEST PROCEDURE - PAN METHOD
4.
Push the
button to turn the meter on. Press the
button andholditfor2 seconds until is displayed. Release the button to take a blank reading (0.0 ppm).
READ
ZERO
bla
I200 COLORIMETER
••••••••••••••••••
READ
ZERO
bla
Insert the tube into the chamber, being
3.
Rinse and fill a colorimeter tube (0290) to the
10 mL line with sample water. Cap and wipe dry.
2.
Fill the Water Sample Collecting Bottle (0688) with sample water. This will be used to dispense sample water for the tests.
1.
Use the 1.0 mL pipet (0354) to add 2.0 mL of *Hardness Buffer Reagent (4255).
5.
6.
Swirl the tube to mix.
8.
7.
Add two drops of *Sodium Cyanide, 10% (6565).
Swirl the tube to mix.
sure to align the index line with the arrow on the meter. Close the lid. This tube is the blank or zero.
L
Mott
Read the 1200 Colorimeter Manual before proceeding. Carefully wipe tubes dry before inserting into the colorimeter chamber.
MANGANESE
Page 79
10.9.
Cap tube and invert to mix.
11.
Align the index line with the arrow on the meter, insert
tube into chamber. Close the lid. Push the button. Record results as ppm Manganese.
READ
Note:
For the best possible results, carry a reagent blank through the procedure. After scanning the blank in Step 4, perform the test procedure on clear, colorless, distilled or deionized water. Subtract results of reagent blank from all subsequent test results.
12.
Use the 0.5 mL pipet (0369) to add 0.5 mL of *Manganese Indicator Reagent (3956).
Wipe tube dry.
CAL
I200COLORIMETER
••••••••••••••••••
READ
L
Mott
If the reading displays Er2, repeat procedure on diluted sample, and multiply the result by the appropriate dilution factor. See 1200 Colorimeter Instruction Manual for procedure.
Page 80
MANGANESE TEST METHOD SPECIFICATIONS
APPLICATION
Drinking and surface; domestic and industrial wastewaters.
RANGE
0.0 - 0.7 mg/L Manganese
METHOD
P AN, (1-[2-Pyridylazo]-2-Naphthol), forms a red complex with Manganese (Mn
+2
)ata
pH of 8 to 10.
HANDLING & PRESERVATION
Manganese may oxidize readily in neutral water and precipitate from solution. It may adhere to or be absorbed by container walls, especially glass. Acidified sample can be stored in plastic.
INTERFERENCES
None. Test is quite specific.
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside U.S.A.) • Fax: 410-778-6394
Visit us on the web at www.lamotte.com
®
63682-01 · 10.08
Page 81
1200 COLORIMETER

SULFA TE

MODEL 1200-SU · CODE 3683-01
QUANTITY CONTENTS CODE
10 g *Sulfate Reagent *V-6277-D 1 Colorimeter Tubes, with caps 0290-6 1 Spoon, 0.1 g, plastic 0699 1 Water Sample Collecting Bottle 0688 1 1200 Colorimeter for Sulfate 26740
*WARNING: Reagents marked with a * are considered hazardous substances. Material Safety Data Sheets (MSDS) are supplied for these reagents. For your safety, read label and accompanying MSDS before using.
To order individual reagents or test kit components, use the specified code number.
SULFA TE INTRODUCTION
The most common mineral forms of sulfur are iron sulfide, lead sulfide, zinc sulfide, calcium sulfate and magnesium sulfate. In most fresh waters, the sulfate ion is the second or third most abundant anion, being exceeded only by bicarbonate and, in some cases, silicate. Sulfur, in the form of sulfate, is considered an important nutrient element. Mineral springs are rich in sulfate and feed appreciable quantities of this compound to the watershed. Acid mine water drainage is a form of pollution which may contribute extremely large amounts of sulfate content to natural waters. Other sources of sulfate include waste material from pulp mills, steel mills, food processing operations and municipal wastes. Many bacteria obtain sulfur from sulfate for the synthesis of amino acids. In lakes and streams low in oxygen, this process of sulfate reduction causes the production of hydrogen sulfide, with its characteristic offensive odor. Calcium sulfate and magnesium sulfate contribute significantly to the hardness of water. Under natural conditions, the quantities ordinarily to be expected in lakes are between 3 and 30 parts per million.
1
Page 82
SULFA TE TEST PROCEDURE
Fill the Water Sample Collecting Bottle (0688) with sample water. This will be used to dispense sample water for the tests.
Insert the tube into the chamber, being sure to align the index line with the arrow on the meter. Close the lid. This tube is the blankorzero.
Rinse and fill a colorimeter tube (0290) to the 10 mL line with sample water. Cap and wipe dry .
5.
3.
1. 2.
4.
6.
I200 COLORIMETER
••••••••••••••••••
READ
bla
ZERO
Cap and shake vigorouslly for 15 seconds. A white precipitate will develop if sulfates are present.
Push the button to turn the meter on. Press the
button andholditfor2 seconds until is displayed. Release the button to take a zero reading
(0.00 ppm).
READ
ZERO
bla
Remove tube from colorimeter. Use the
0.1 g spoon (0699) to add one measure of *Sulfate Reagent (V -6277).
8.
Invert the tube to mix again.
5
7.
Wait 5 minutes.
MINUTES
L
Mott
- BARIUM CHLORIDE METHOD
Read the 1200 Colorimeter Manual before proceeding. Carefully wipe tubes dry before inserting into the colorimeter chamber.
SULFATE
2
Page 83
9.
Note:
A white film is deposited on the inside of test tubes as a result of the sulfate test. Thoroughly clean and rinse test tubes after each test.
Align the index line with the arrow on the meter, insert
CAL
I200COLORIMETER
••••••••••••••••••
READ
tube into chamber. Close the lid. Push the button. Record results as ppm Sulfate.
READ
Note:
The Sulfate test is sensitive to temperature. Best results will be obtained if the water sample is 20° - 23°C.
L
Mott
3
Page 84
SULFA TE TEST METHOD SPECIFICATIONS
APPLICATION
Drinking and surface; domestic and industrial wastes.
RANGE
5 to 100 ppm Sulfate
METHOD
Sulfate ion is precipitated in an acid medium with barium chloride to form a barium sulfate suspension in proportion to the amount of sulfate present.
HANDLING & PRESERVATION
Sulfate samples may be preserved by refrigeration at 4°C up to 7 days in glass or plastic containers without any change in concentration.
INTERFERENCES
Suspended matter and color interference may be removed by a filtration step. Silica in excess of 500 mg/L will interfere.
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside USA)• Fax 410-778-6394
Visit us on the web at www.lamotte.com
®
63683-01 · 11/01
Loading...