Lakewood Instruments 543, 543-M, 543-L, 543-LL Operation Manual

LAKEWOOD INSTRUMENTS
MODEL 543
4-ELECTRODE CONDUCTIVITY
SENSORS
INSTALLATION & OPERATION MANUAL
543-L
543-LL
Lakewood Instruments
7838 North Faulkner Road, Milwaukee, WI 53224 USA Phone (800) 228-0839 • Fax (414) 355-3508 http://www.lakewoodinstruments.com
Lakewood Instruments
Congratulations on your purchase of a Lakewood Instruments product. We would like to take this opportunity to welcome you to the Lakewood Instruments product family.
With proper care and maintenance, your product should give you trouble-free service. Please take the time to read and understand the operation manual, paying special attention to the sections on INSTALLATION and MAINTENANCE.
If, in the future, any parts or repairs are required, we strongly recommend that only original replacement parts be used. Our Customer Service Department would be happy to assist you with your parts or service requests.
We thank you for your selection and purchase of an Lakewood Instruments product.
1
2
MODEL 543
Table of Contents
INTRODUCTION 5
Features, Benefits, Specifications 5 Ordering Information 6
INSTALLATION 7
Installation 7 Storage 8 Sensor Life 8 Cleaning 8
REPLACEMENT PARTS 9
Replacement tips 9 Accessories 9
MAINTENANCE AND TECHNICAL SERVICE 10 DRAWINGS 11
3
4
Features, Benefits, Specifications
4-ELECTRODE CONDUCTIVITY SENSORS
543 SERIES
543-L
543-LL
Body Specifications
Option Body Material Max Temp Pressure Wetted Materials
-5x CPVC 150°F (66°C) 100 psi (7 bar) CPVC, Glass, Carbon, Viton
-8x 316 SS 260°F (127°C) 150 psi (10 bar) 316 SS, Glass, Viton
Electrode Specifications
Electrode Conductivity Range Temp Compensator Max Temp Max Pressure
543-M
543-L
543-LL
500-100,000 µS
100-1,000 µS
10-500 µS
4K NTC 392°F (200°C) 250 psi (17 bar) 4K NTC 392°F (200°C) 250 psi (17 bar) 1K NTC 392°F (200°C) 250 psi (17 bar)
5
Enclosures
-1 NEMA-4X w/amplifier* -2 Cl I, II w/amplifier*
-5 NEMA-4X -3 Cl I, II
In-line Mounting Options
(All sensors have ¾ NPT on sensor tip)
-8I 1 inch NPT 316 SS Swagelock
-8R 1½ inch NPT 316 SS Ball valve
Ordering Information
Cond
Sensor
543-M -0 ½ in NPT adapter -4T 4 in steel adapter (must use 5.5 Option) -5.5 in -STD 543-L -1 NEMA-4X
543-LL -2 Aluminum encl w/amp* -8S 316 SS no fitting -18 in
543-M -4 -8R -18 -STD
* Amplifier options -1 and -2 are only available with Model 175-RP, 843 and 1040 controllers.
= 10 in body not available with -8R Option.
Cable Length or
Enclosure
-5S CPVC no fitting
w/amplifier*
-3 Aluminum enclosure -8I 316 SS w/Swagelock fitting 1 in NPT -48 in
-4 240 in of cable -8R 316 SS w/ball valve 1½ in NPT
-5 NEMA-4X enclosure
Material, In-line Fitting Body
Length
-10 in =
Example
Electrode
6
INSTALLATION
Installation
The Conductivity sensor may be mounted in any position as long as water is always in contact when measuring process water. Avoid connections in “dead leg” sections of pipe. An air pocket around the electrode tips will cause erroneous readings. The sensor electrodes should be in direct contact with the process flow (see DWG #04259 in the back of this manual).
The 543 series sensors have 6 wires. They are as follows:
Black wire Electrode
Red wire Electrode
White wire Electrode
Green wire Electrode
Black wire Temp compensation 4K NTC or 1K PTC
Black wire Temp compensation
Cable length of Conductivity sensors is measured from the electrode tips to the end of the wire. Twenty (20) feet is the maximum cable length. Lakewood Instruments guarantees operation up to 20 feet. If a cable extension is used, Lakewood Instruments will not guarantee operation of the sensor.
7
Storage
Conductivity sensors may be stored for years. No special handling procedures are required for storage.
Sensor Life
Conductivity sensor life varies with the following:
Type of process water and chemicals used.
Temperature of water.
Pressure of water.
Ambient temperatures, low pressures, and water with no chemicals are ideal for longest life.
Cleaning
Conductivity sensors may be cleaned with a nylon brush or clean paper towel. Cleaning frequency will vary with the type of process water. Cleaning should be done at least once a month or if conductivity readings become unstable.
8
REPLACEMENT PARTS
Replacement Tips
543 series sensors may be replaced by simply unscrewing the tip of the sensor. The body, fittings, and other options may be reused with the new sensor tip. Replacement numbers are a follows:
All sensor tips come with 20 ft cable. Cable can be cut to fit.
543M-STD Standard replacement titanium/316 SS sensor tip 543L-STD Standard replacement titanium/316 SS sensor tip 543LL-STD Standard replacement titanium/316 SS sensor tip - (for Model 1743 only)
Accessories
ADAPTER
68893 Adapter from sensor cable end to ¾ NPT female 69168 316 SS compression fitting, 1 in NPT connection 69058 316 SS Ball valve 1½ in NPT connection
PREAMP 67230 Remote conductivity preamp 68900 Manual for remote conductivity preamp 66317 Preamp cable. 4 conductor shielded. per ft
MAINTENANCE AND TECHNICAL SERVICE
9
Technical Service/Return Material Procedure
Technical Support for Lakewood Instruments can be reached by calling (800) 228-0839
or faxing (414) 355-3508, Monday through Friday, 7:30 a.m. - 5:00 p.m. CST.
Mail and returns should be sent to:
Lakewood Instruments 7838 North Faulkner Road Milwaukee, WI 53224 USA
When any merchandise is returned to the factory, please call and obtain a return goods authorization (RGA) number and have the following information available:
Customer’s name, address, phone and fax numbers (shipping and billing).
A hard copy purchase order number (no exceptions) for cases where repairs or parts are
required that are not under warranty.
A contact person’s name and phone number to call if the equipment is beyond repair or to
discuss any other warranty matter.
Equipment model and serial numbers.
Reason for return (i.e., repair, warranty, incorrect part, etc.).
We will then fax to your attention an RGA form that must accompany the returned item.
NOTE: THE RGA NUMBER MUST BE CLEARLY WRITTEN ON THE OUTSIDE OF THE PACKAGE(S) BEING RETURNED.
ANY ITEMS SENT BACK TO THE FACTORY
WITHOUT AN RGA NUMBER WILL BE REFUSED
AND RETURNED TO SENDER
10
DRAWINGS
11
For more information call toll free in the USA (800) 228-0839
Manufactured in the USA
Lakewood Instruments
7838 North Faulkner Road, Milwaukee, WI 53224 USA Phone (800) 228-0839 • Fax (414) 355-3508 http://www.lakewoodinstruments.com
© Copyright 1997 Lakewood Instruments, LLC.
Printed in USA, P/N 1109982 Rev. A