Mechanical Calibration (required only on model 822):
1. With the power turned off
, install a jumper on terminal strip TB1 between terminals 0 and 1.
2. Adjust the 822's mechanical zero (needle pointing to zero with power off) by turning the screw located directly
behind the meter movement.
Power Supply Board Calibration:
1. Remove the previously installed jumper (Mechanical Calibration step 1).
2. Place the common lead (-) of a Digital Volt Meter (DVM) on TB1 terminal 1.
3. Place the positive lead (+) of the DVM on pin 1 of IC U1.
4. Turn on the power.
5. Adjust R13 on the power supply board until the DVM reads 10.20 VDC ± 0.01 VDC.
Analog Board Calibration:
1. Connect the calibrator/simulator via the patch cord or fixed resistors to TB1 terminals 1 thru 4.
2. Dial X1 on 822/832's range switch .
3. Place the DVM's common lead (-) on TB1 terminal 1, and the positive lead (+) on TB1 terminal 0.
4. Adjust R1 on analog PCB until DVM reads 5.00 VDC ± 0.05 VDC.
5. Turn the 822/832's range switch to "CHECK" and hold it at this position.
6. Adjust R29 on analog PCB until DVM reads:
For 822: 10.00 VDC ± 0.05 VDC
For 832: 5.00 VDC ± 0.05 VDC
Digital Board Calibration (required only on model 832):
1. While holding the 832's range switch in the "CHECK" position, turn R1 on the digital PCB until the 832's
digital display reads 10.0 ± 0.1.
Calibration & Accuracy Verification Procedures
For 822 & 832 Meters
THORNTON
Leading Pure Water Analytics
www.thorntoninc.com
For the most current product information visit:
email: info@thorntoninc.com
www.thorntoninc.com
822_cal.doc Rev C.
Updated logo & address 09/03
Mettler-Toledo Thornton, Inc.
36 Middlesex Turnpike
Bedford, MA 01730 USA
Telephone: +1-781-301-8600
Toll-Free: 1-800-510-PURE
Customer/Technical Service
Telephone: +1-781-301-8690
Toll-Free: 1-800-642-4418
Cust Service Fax: +1-781-271-0214
Tech Service Fax: +1-781-271-0675
Mettler-Toledo Thornton, Inc. – Subject to Technical Changes
10 MΩΩ Calibrator/simulator Pinout:
100.0 KΩΩ between pins 1 and 2
10.0 KΩΩ between pins 2 and 3
180.0 KΩΩ between pins 2 and 4
NOTE: It may be necessary to connect a ground wire internally
between earth ground (Pin 9, TB2) and cell ground (Pin 1,
TB1) to avoid ground pickup problems.