Labomed TCM 400 FLR User Manual

TCM 400FLR
User Manual
R
Inverted Microscopy
To ensure proper use of this instrument as well as to avoid injury while operating instrument, understanding this manual completely before use is highly recommended.
________________________________________________________________________________________________________
7127000-795
TCM400Flr
CONTENTS
1 INTRODUCTION 1
2 PRINCIPLE 1
3 UNPACKING 2
4 SAFETY INSTRUCTIONS 3
5 INSTALLATION 4
6 MAINTAINING OPERATING LIFE 6
7 CENTERING THE LIGHT BEAM 7
8 OBSERVING A SPECIMEN 7
9 TROUBLESHOOTING 8
10 TECHNICAL SPECIFICATION 9
PACKAGING CONTENTS
1. FL attachment in 2 parts
2. 2 Filter cube
3. Power Supply
4. Power cord
5. UV protection shield
Fluorescence Attachment User Guide 2
SAFETY INSTRUCTIONS
Fluorescent attachment is manufactured according to CE safety norms and regulations.
Fluorescent attachment is intended for use only as prescribed in this manual.
Make sure the main switch on the power supply unit is set to OFF (O) before supplying power to the unit.
The power supply unit contains high voltage components. Do not attempt to disassemble the unit when turned ON (I).
To avoid injury, do not remove the lamp housing while the lamp is lit. Also, do not turn the lamp ON while the lamp housing is removed to avoid damaging the instrument.
Before opening lamp housing for replacement of the lamp, set the main switch to OFF (O) and unplug the lamp housing's connecting cord plug from the output connector on the power supply unit. Wait at least 10 minutes or until the lamp housing cools down before any changes are made.
Do not open the lamp housing while the system is turned ON, for any user contact with internal parts may cause injury.
A used mercury burner should be disposed of in compliance with the ordinances or regulations of your national or local government.
The manufacturer will not accept any liability for damage caused by any user following unsafe practice of the instrument.
Avoid using the lamp beyond its service life. The mercury lamp seals high pressure gasses. If the lamp is used beyond its service life, stress may accumulate inside the lamp and in the worst but very rare case, the lamp may burst.
Fluorescence Attachment User Guide 3
INSTALLATION
Remove all parts from the packing carefully.
1. Remove the plastic cover (1) from the microscope to make room for
the fluorescence attachment.
Bring the nosepiece to the maximum height by rotating the coarse knob (2) anti- clockwise. Remove the illumination power cable before removing the cover. Apply similar pressure by thumbs on the inner
sides of the panel (3 & 4) to push it out. Refer to figure 1 and 2.
2. Join the Optical assembly and illumination assembly to complete the FL attachment. Guide the illumination assembly to the optical assembly. Secure both parts by the allen screws (1) provided. Refer
Fig. 3.
3. Assemble attachment in the microscope as shown in Fig. 4 and
secure it with 3 Nos of Screws. Then install filter in the guide slot as shown in Fig. 5.
1
Fig. 1
3
4
2
Fig. 2
Fig. 3
Fig. 4
1
Fig. 5
Fluorescence Attachment User Guide 4
INSTALLATION
Open the illumination house to install the mercury lamp. Loosen the
A
B
thumbscrew housing.
To avoid any damage to the mercury vapor lamp, a plastic rod (1) is fitted in
place of lamp during transit. Remove this rod by loosening the two screws
indicated by the two arrows A and B shown in Fig. 6.
Mount the mercury vapor lamp in place of plastic rod and tighten the screws
to secure the lamp. install as shown in Fig. 6.A.
Avoid touching the lamp in order to prevent leaving finger prints on the
glass. If there is any residue on the lamp's surface, gently wipe it off with
soft tissue paper.
Install the power cable coming from the illumination house to the power unit
indicated by arrow A in Fig. 7.
(1). Connect the power cord to the AC Inlet indicated by arrow B in Fig. 7.
indicated by the arrow and gently pull out lamp door from lamp
Fig. 6
. 6A
Fig
B
A
Fig. 7
Fluorescence Attachment User Guide 5
INSTALLING THE FILTER B & G
Select the filter to be used on the fluorescence attachment.
Engage the slider to the filter house and gently slide it in.
It will stop at the Iindex position.
Fig. 8
MAINTAINING OPERATING LIFE
Ensure that the power supply unit has been connected to proper power supply voltage.
Set the main switch on the power supply unit to ( | ). After the lamp is ignited, at least 5 minutes required to
stabilize light..
To avoid shortening the power supply unit's operating life, avoid turning power ON while the lamp is not mounted
Avoid turning power ON and OFF within short intervals for this may reduce working life of the power supply.
After the lamp is turned OFF, it should not be re-ignited before the mercury vapors cool and condense. Wait
approximately 10 minutes before turning power to the fluorescent attachment O N again.
Fluorescence Attachment User Guide 6
CENTERING THE LIGHT BEAM
Turn the power supply ON. Allow the lamp at least 5 minutes to fully illuminate for optimal use.
The halogen illumination of the microscope should be kept off while working with FLR.
Place the specimen slide on stage. Select the excitation light by installing the filter block as follows :
For G-excitation - Install red sliding block to its outermost position.
For B-excitation - Push blue sliding block in to its innermost position.
For O light (non-excited light) - Slide out the filer to parking point.
Observe the specimen slide through the microscope's eyepieces.
When no fluorescent light is required, turn the attachment's power
supply OFF and set the sliding filter block to its middle / neutral
position. To view the specimen in normal condition under the
microscope's light, turn the microscope ON.
OBSERVING A SPECIMEN
First allow the mercury vapor lamp at least five
minutes to fully illuminate for optimal use (it is very
important to allow the system enough time to fully ignite in order to achieve a fully illuminated centering beam). Look into the eyeppiece aanddmanipulate
the ccollector focusing knobb to make the field asd
bright as regular asapossible, then, turn the knobb
clockwise to tighten as shown in Fig.#9/Part 1.
When the lamp is fully illuminated, a beam of light is
visible through the eyepieces in the field of view.
To center this beam of light in the field of view, two
centering knobs are located on the lamp housing as
indicated by the two arrows (A & B) in Fig. 9. By
1
A
B
Fig. 9
adjusting these two knobs, the focused beam will shift. Adjust the beam position to the center of the
field of view as shown in Fig. 9. A.
Fluorescence Attachment User Guide 7
Fig. 9A.
TROUBLESHOOTING
Problem Cause Remedy
Optical System
1) The mercury lamp lights up, Filter slider is not engaged properly Adjust filter to required setting. but field of view remains dark in light path.
2) Image is unclear, blurred or a) Objectives and/or filters are dirty. Clean them with tissue/lens paper. has poor contrast. b) Field iris diaphragm is not Adjust field iris diaphragm so
fully opened. that the image circumscribes
the field of view.
3) Field of view is vignetted or a) Objective is not engaged properly Be sure when rotating nosepiece it is not evenly illuminated. in light path. that objective clicks into position.
b) Filter slider is not engaged properly Engage required filter
in light path. properly in light path.
c) Mercury lamp is not centered Center mercury lamp correctly.
properly. This process is shown on Page 7
d) Beam focus deviates from correct Adjust the condenser knob.
position. This process is shown on Page 7
4) Mercury lamp is not illuminating Some mercury lamps may not ignite Press the Trigger button provided after switching power supply ON. the first time the power supply is on the front panel of power supply
turned ON. unit and wait for lamp to ignite.
Electrical System
1) Lamp housing / Main switch a) Power cord is not connected Make sure power cord is firmly indicator does not light up. properly. in place.
b) The fuse is blown. Replace with fuse of same type
and rating.
2) Main switch indicator lights up, a) Connector cable is not engaged Verify connection is secure. yet mercury lamp remains OFF. properly.
b) Mercury lamp is not mounted. Install mercury lamp as
indicated on Page 5
3) Light demonstrates partial a) Sufficient time has not been Allow at least 5 minutes after flickering. allowed for the lamp to fully turning power supply ON for the
illuminate. lamp to fully illuminate. If flickering
persist press Trigger button.
b) Lamp's service life has expired. Replace the mercury lamp
Fluorescence Attachment User Guide 8
TECHNICAL SPECIFICATIONS
Type : Transmitted Fluorescent illumination based on switching of dichroic mirror cubes
Observation modes : B-excitation and G-excitation Lamp type : HBO 50 W, high pressure Mercury vapor lamp Supply Voltage : 110V AC, 60 Hz / 220V AC, 50Hz Operating environment : Indoor use
Ambient temperature 5 C to 40 C Maximum relative humidity 80% at 40 C
0 0
0
FILTER SPECIFICATIONS
B-excitation Excitation filter : Transmission 90% at wavelength 420 to 480 nm.
Transmission 0% at wavelength 485 nm onwards.
Dichroic mirror : Transmission 93% at wavelength 540 nm onwards .
Transmission 0% at wavelength 400 to 530 nm.
Barrier filter : Transmission 95% at wavelength 535 nm onwards
Transmission 0% at wavelength 400 to 500 nm. Recommended Fluorescent Dye : Acridine Orange, both DNA & RNA G-excitation Excitation filter : Transmission 85% at wavelength 490 to 525 nm.
Transmission 0% at wavelength 530 nm onwards. Dichroic mirror : Transmission 90% at wavelength 600 nm onwards .
Transmission 0% at wavelength 400 to 595 nm. Barrier filter : Transmission 92% at wavelength 615 nm onwards
Transmission 0% at wavelength 400 to 575 nm. Recommended Fluorescent Dye : Acridine Red FM 5-95
TM
Fluorescence Attachment User Guide 9
www.laboamerica.com
Our policy is one of continuous development. Labo America, Inc., reserves the right to change design and specifications without prior notice.
Labo America Inc.
920 Auburn Court Fremont CA 94538
LABOMED and TCM400 FLR are registered trademarks of Labo America, Inc. With a policy of continuous development, Labo America, Inc. reserves the right to change design and specifications without prior notice.
© 2011 Labo America, Inc. | 7127000-795 08-2018
U.S.A. Telephone: 510 445 1257 Fax: 510 445 1317 sales@laboamerica.com
A28240
ISO 9001
LABOMED
Loading...