To ensure proper use of this instrument and to avoid injury during operation, reading this manual before use is highly
recommended.
6137000-795 PRIMA ENT/DNT Issue 1.4
Printed March, 2016
Prima ENT/DNT
Prima ENT/DNT is a trade name for LABOMED Products.
LABOMED is a registered trademark of Labo America, Inc.
All other trademarks are the property of their respective owners.
The information contained in this document was accurate at the time of publication. Specifications
are subject to change without notice. LABOMED reserves the right to make changes to the product
described in this user manual without notice and without incorporating those changes in any
products already sold.
ISO 9001/13485 Certified – LABOMED products are designed and manufactured under quality
processes that meet ISO 9001/13485 requirements.
No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any
form or by any means: electronic, mechanical, recording, or otherwise, without the prior written
permission from LABOMED.
2
Prima ENT/DNT
List Of Contents
1. Introduction and Intended Use4
2. Product Description 5-6
3. Warnings and Cautions 7-8
4. Explanation of Symbols 9
5. Standards and Directives 10
6. Unpacking 11
7. Installation Instructions
- 7A Installation of Base (Mobile Stand) 12-13
- 7B Microscope Installation 14-15
8. Electrical Connections
- Replacing the Illumination Source 16
- Replacing the Fuse 16
- Data Plate Details 16
- Prima Wir ing Coding Diagram 17
9. Control Elements 18
10. Using the Microscope 19-20
11. Changing the Objectives and Eyepieces 21
12. Use of Accessories 22-23
13. Thermal Cut-Off 24
14. Tension Adjustment 24
15. Moving the Instrument 25
16. Care and Maintenance 26-27
17. Ambient Requirements 28
18. Disposal 29
19. Troubleshooting Table 30-31
20. Technical Specifications 32
21. Guidance Tables 33-36
22. Dimensions and Weight 37
23. Glossary 38
24. Warranty 39
3
Prima ENT/DNT
1
LABOMED Prima ENT/DNT is a surgical and diagnostic microscope that is adaptable for different
surgical needs for consistent visualization during all intraoperative phases of ENT and dentistry
surgeries by providing a magnified view of the surgical field without compromising performance.
The microscope provides extremely high optical image quality, good depth of focus, and a wide field
of view for precise surgery.Illumination control and high eye point of the observation head helps
reduce the user’s work fatigue and allows comfortable use over a long period.
Salient features of this microscope are:
1. The observation head can easily be positioned with the help of a suspension arm.
2. An advanced 5-step magnification changer allows optimal magnification for various needs at
INTRODUCTION AND INTENDED USE
five different magnification levels.
3. Cold light illumination with a high-intensity 50W LED lamp is provided using a fiber optic
guide for proper illumination. Illumination can be adjusted using the intensity control knob
conveniently located at the suspension arm and easily accessible to the user.
4. When the microscope is not in use, the suspension arm can be folded over the main body for
compact storage.
5. The rigid H-shape base with caster wheels provides the instrument with stability and
mobility.
Intended Use
The Prima ENT/DNT microscope is an AC powered device intended for use during diagnosis and
surgery to provide a magnified view of the region of interest.
NOTE: Use Prima ENT/DNT microscopes as specified in the intended use.
Configurations
Microscope Catalogue no.
Prima ENT 6137000-000
Prima DNT 6138000-000
4
Prima ENT/DNT
6
24
25
3
22
23
13
4
7
5
11
19
14
15
20
17
16
18
12
10
8
21
9
1
2
PRODUCT DESCRIPTION
Prima DNT
1. Wheel with brake 14. Ergo Head 0-210° adjustable
2. H shape metal base 15. Magnification changer (magnichanger)
3. Column 16. Handles
4. Swivel arm 17. Common main objective (CMO)
5. Suspension arm 18. Magnichanger inclination
6. Swivel arm locking knob movement lock
7. Suspension arm movement locking knob 19. Illumination control knob
8. Coupling movement locking knob 20. Fine focusing knob
9. Inclined coupling 21. iVu Camera inlet
10. Inclined coupling movement knob 22. On/Off switch
11. Suspension arm hydraulic movement lock 23. Power inlet
12. Suspension arm spring tension adjustment 24. Eyepieces
13. Swivel arm covers 25. Counter balancing knob
5
Prima ENT
Prima ENT/DNT
24
13
23
6
3
4
7
11
19
22
8
18
16
17
9
21
25
14
15
12
5
26
10
2
20
1
1. Wheel with brake 14. Binocular head
2. H shape base 15. Magnification changer
3. Column 16. Handles
4. Swivel arm 17. Common main objective (CMO)
5. Suspension arm 18. Magnichanger inclination movement lock
6. Swivel arm locking knob 19. Illumination control knob
7. Suspension arm movement locking knob 20. Foot pedal for Z-axis (optional)
8. Coupling movement locking knob 21. Fine focusing knob
9. Inclined coupling 22. iVu Camera inlet
10. Inclined coupling movement knob 23. On/Off switch
11. Suspension arm hydraulic movement lock 24. Power inlet
12. Suspension arm spring tension adjustment 25. Eyepieces
13. Swivel arm covers 26. Counter balancing knob
6
Prima ENT/DNT
3
LABOMED is not responsible for the safety and reliability of this instrument when:
- Assembly, disassembly, repair, or modification is made by unauthorized dealers or persons.
- The instrument is not used in accordance with this user manual.
A WARNING is an instruction that draws attention to the risk of injury or death.
WARNING: USERS OF THIS EQUIPMENT SHOULD BE THOROUGHLY TRAINED IN THE APPROPRIATE
WARNING: THIS INSTRUMENT SHOULD BE USED IN STRICT ACCORDANCE WITH THE INSTRUCTIONS
WARNINGS AND CAUTIONS
MEDICAL PROCEDURES. FURTHERMORE, THEY SHOULD TAKE THE TIME TO READ AND
UNDERSTAND THESE INSTRUCTIONS BEFORE PERFORMING ANY PROCEDURE. THEY SHOULD
ALSO READ AND UNDERSTAND THE INSTRUCTIONS FOR ANY OTHER EQUIPMENT USED IN
CONJUNCTION WITH THIS MICROSCOPE (i.e. ELECTROSURGICAL GENERATORS). FAILURE TO
DO SO MAY RESULT IN INJURY TO THE PATIENT AND/OR DAMAGE TO THE MICROSCOPE.
OUTLINED IN THIS USER MANUAL. THE SAFETY OF THE OPERATOR AND THE PERFORMANCE
OF THE INSTRUMENT CANNOT BE GUARANTEED IF USED IN A MANNER NOT SPECIFIED BY
LABOMED.
WARNING: DO NOT REPAIR OR SERVICE THIS INSTRUMENT WITHOUT AUTHORIZATION FROM THE
MANUFACTURER. ANY REPAIR OR SERVICE TO THIS INSTRUMENT MUST BE PERFORMED BY
EXPERIENCED PERSONNEL OR DEALERS WHO ARE TRAINED BY LABOMED, OR SERIOUS
INJURY TO THE OPERATOR OR PATIENT MAY OCCUR.
WARNING: MODIFICATIONS TO THIS INSTRUMENT ARE NOT ALLOWED. ANY MODIFICATION TO THIS UNIT
MUST BE AUTHORIZED BY LABOMED, OR SERIOUS INJURY TO THE OPERATOR OR PATIENT MAY
OCCUR.
WARNING: IF THIS INSTRUMENT IS MODIFIED, APPROPRIATE INSPECTION AND TESTING MUST BE
CONDUCTED TO ENSURE CONTINUED SAFE USE OF THIS INSTRUMENT.
WARNING: TO AVOID RISK OF ELECTRIC SHOCK, THIS EQUIPMENT MUST ONLY BE CONNECTED TO A
SUPPLY MAIN WITH PROPER EARTHING, OR DAMAGE TO THIS INSTRUMENT AND/OR INJURY TO
THE OPERATOR OR PATIENT MAY OCCUR.
WARNING: ENSURE THAT THE VOLTAGE APPLIED TO THE UNIT IS THE SAME AS THE VOLTAGE INDICATED
ON THE DATA PLATE, OR DAMAGE TO THE UNIT MAY OCCUR.
WARNING: THIS INSTRUMENT MUST BE PLUGGED IN TO AN OUTLET WITH AN EARTH GROUND. DO NOT
REMOVE OR DAMAGE THE EARTH GROUND CONNECTION ON THE POWER INPUT
CONNECTOR OR THE UNIT’S POWER CORD, OR DAMAGE TO THE INSTRUMENT AND/OR INJURY
TO THE OPERATOR OR PATIENT MAY OCCUR.
WARNING: THE EQUIPMENT OR SYSTEM SHOULD NOT BE USED ADJACENT OR STACKED WITH OTHER
EQUIPMENT LIKE MRI MACHINES, RFID, SECURITY SYSTEMS LIKE METAL DETECTORS, OR
SIMILAR DEVICES. IF ADJACENT OR STACKED USE IS NECESSARY, THE EQUIPMENT OR SYSTEM
SHOULD BE OBSERVED TO VERIFY NORMAL OPERATION IN THE CONFIGURATION IN WHICH IT
WILL BE USED.
7
Prima ENT/DNT
WARNING: THIS INSTRUMENT IS NOT SUITABLE FOR USE IN THE PRESENCE OF FLAMMABLE ANESTHETIC
MIXTURES SUCH AS OXYGEN OR NITROUS OXIDE.
WARNING: LED RADIATION - DO NOT STARE DIRECTLY INTO THE BEAM WHEN THE MICROSCOPE IS IN THE
ON POSITION.
WARNING: THE USE OF ACCESSORIES OR CABLES OTHER THAN THOSE SPECIFIED, WITH THE EXCEPTION
OF THOSE SOLD BY THE MANUFACTURER AS REPLACEMENT PARTS FOR THE INTERNAL
COMPONENTS, MAY RESULT IN INCREASED EMISSIONS OR DECREASED IMMUNITY OF THE
EQUIPMENT OR SYSTEM.
A CAUTION is an instruction that draws attention to the risk of damage to the product.
CAUTION: THE INTERNAL CIRCUITRY OF THE INSTRUMENT CONTAINS ELECTROSTATIC SENSITIVE
DEVICES (ESD) THAT MAY BE SENSITIVE TO STATIC CHARGES PRODUCED BY THE HUMAN BODY.
DO NOT REMOVE THE COVERS WITHOUT TAKING PROPER ESD PRECAUTIONS.
CAUTION: DO NOT USE SOLVENTS OR STRONG CLEANING SOLUTIONS ON ANY PART OF THIS
INSTRUMENT, AS DAMAGE TO THE UNIT MAY OCCUR. SEE THE CARE AND MAINTENANCE
SECTION FOR DETAILED CLEANING INSTRUCTIONS.
CAUTION: MEDICAL ELECTRONIC EQUIPMENT NEEDS SPECIAL PRECAUTIONS WITH RESPECT TO
ELECTROMAGNETIC CHARGE (EMC) AND NEEDS TO BE INSTALLED AND SERVICED ACCORDING
TO THE EMC INFORMATION PROVIDED IN THE ACCOMPANYING DOCUMENTS.
CAUTION: PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT CAN AFFECT MEDICAL
ELECTRICAL EQUIPMENT.
CAUTION: THIS INSTRUMENT IS NOT TO BE USED NEAR HIGH-FREQUENCY EMITTING SURGICAL
EQUIPMENT.
CAUTION: DO NOT CONNECT ANY EQUIPMENT TO THE DEVICE OTHER THAN THOSE INTENDED FOR USE
WITH THE DEVICE.
CAUTION: DO NOT USE A CONVERTER ADAPTER THAT WILL CONVERT THE THREE-PRONG AC PLUG TO A
TWO-PRONG LINE PLUG. THE POWER SUPPLY IN THIS MICROSCOPE WILL NOT BE PROPERLY
GROUNDED, AND ELECTRIC SHOCK MAY RESULT.
CAUTION: REMOVE THE AC POWER PLUG FROM THE WALL SOCKET WHILE CHECKING FOR A BLOWN
FUSE.
CAUTION: DO NOT ROLL THE MICROSCOPE OVER CABLES OR HOSES.
CAUTION: DO NOT REMOVE FERRITE BEADS IF APPLIED TO CABLES.
8
Prima ENT/DNT
4
EXPLANATION OF SYMBOLS
Caution: Observe all warning labels and notes! If any label is missing on your instrument or
has become illegible, please contact the manufacturer to obtain new labels.
Brightness control: After the illumination has been switched on, the user can continuously
adjust brightness by turning the appropriate knob.
Accompanying documents must be consulted.
Compliance to medical device directive 93/42/EEC.
Protective earthing
This way up - indicates the correct upright position of the transport package.
Keep dry - the transport package shall be kept away from rain.
Year of manufacture used on product data plate.
Fragile - contents of the transport package are fragile and should be handled with care.
Electromagnetic interference can occur in the vicinity of devices carrying this symbol.
Alternating Current
Permissible humidity range during transport and storage.
Permissible pressure range during transport and storage.
Recycling of packaging materials.
The product complies with US and Canadian safety requirements.
Permissible temperature range during transport and storage.
Separate disposal of waste electrical equipment.
Unsafe for use with MR (Magnetic Resonance).
Do not over balance more than 5°.
9
Prima ENT/DNT
5
The instrument described in this user manual has been designed in compliance with the
following standards:
·ISO 8600-3 First edition 1997-07-01 AMENDMENT 1 Optics and Optical instruments-
· ISO 8600-3 First edition 1997-07-01 Optics and Optical instruments-Medical
·ISO 8600-5 First edition 2005-03-15 Optics and photonics-Medical endoscopes and
· ISO 9001/13485 Quality management systems approved by UL (notified body).
· ISO 14971-2007 Risk management to medical devices.
STANDARDS AND DIRECTIVES
Medical endoscopes and endoscopic accessories Part 3: Determination of field of view
and direction of view of endoscopes with optics.
endoscopes and endoscopic accessories Part 3: Determination of field of view and
direction of view of endoscopes with optics.
endotherapy devices Part 5: Determination of optical resolution of rigid endoscopes
with optics.
Directives Used
Directives 93/42/EEC, Article II, Section 5, Annex VII
·
· IEC 60601-1-3rd edition (2005)
· IEC 60601-1-2
· EN 55011:2007
Classifications
· For Europe, per Directive 93/42/EEC, the unit is a Class I instrument, per rule 12,
Annex IX.
· For the United States, the FDA classification is Class I.
· Please observe all applicable accident prevention regulations.
10
Prima ENT/DNT
6
The appliance is delivered in sub-assembled groups along with one Installation Kit and one
user manual.
Please check for the following when unpacking the device:
1. Mobile supporting base with brakes on caster wheels or other type of mounting
2. Column, depending on the type of mount ordered
3. Swivel arm and suspension arm assembly with fiber optic cable
4. Inclined coupling with magnichanger assembly and common main objective with
5. Observation head (straight or ergo tiltable 0-210°) as ordered
UNPACKING
focal length as ordered
6. Pair of eyepieces as ordered
7. Power cord
8. Installation Kit
a) Allen wrench 5 mm
b) Allen wrench 8 mm
9. User manual
11
Prima ENT/DNT
7A
1. Open the Center Support assembly box
2. Remove the tool kit and foam sheets as
3. Lift up the Center Support assembly as
INSTALLATION OF BASE (MOBILE STAND)
as shown in Fig. 1.
shown in Fig. 2.
shown in Fig. 3.
Fig 1
Fig 2
4. Place the Center Support assembly on
the floor as shown in Fig. 4.
5. Open one of the Wheel Support
assembly boxes as shown in Fig. 5.
6. Remove foam sheets as shown in Fig. 6.
Fig 3
Fig 4
Fig 5
12
Fig 6
7. Lift the Wheel Support assembly
and cover as shown in Fig. 7. Place it
(wheels down) under one end of the
Center Support as shown in Fig 8.
Repeat procedure for the other Wheel
Support assembly.
8. Align holes of the Center Support with
the holes of both Wheel Supports and
tighten the screws as shown in Fig. 8
using the 8 mm allen wrench.
Prima ENT/DNT
Fig 7
9. Remove 2 screws at center as shown
in Fig. 9 to remove the wooden block.
Unscrew with screw driver as shown in
Fig. 9.
10. Remove the wooden block as shown in
Fig. 10.
11. Tighten all screws using the 8 mm allen
wrench.
Fig 8
Fig 9
Fig 10
12. Place top cover on both Wheel Support
assemblies as shown in Fig. 12.
Fig 11
Fig 12
13
Prima ENT/DNT
7B
1. Open the microscope box. Remove pillar (column)
2. Make sure that during assembly of the column,
MICROSCOPE INSTALLATION
from the box. Engage this column to the base
through indexing pin, shown as B in Fig 2. Align
the tapped holes of the column with the base and
tighten two allen screws from the bottom, shown
as A in Fig 2.
ring (D) is fixed in place, as shown in Fig 2.
B
D
A
Fig 2
3. Remove the arm assembly from the box and
follow the instructions below (see Fig. 3).
A. Ensure that the swivel arm lock knob (A) is
loosened.
B. Place the arm assembly on the column shaft
(1) shown in Fig. 3.
C. Screw in the threaded plug from the top (2).
D. Put the protective cap (3) in place, and loosen
the suspension arm locking knob (B) so that it
can be rotated.
Do not unscrew knob A completely (see Fig. 3).
Loosen only a few threads; otherwise the lock
ring inside the column guide may disconnect and
installation may not possible.
3
2
B
A
1
Fig 3
14
Prima ENT/DNT
A
4. Remove the inclined coupling assembly from its
packaging and follow the instructions below (see
Fig. 4.)
A. Install the coupling by sliding the shaft (1)
into the suspension arm.
B. Lock the inclined coupling with the threaded
knob (3).
C. Make sure the safety screw is installed and
tightened at position (2) so the coupling
does not detach if knob (3) is accidentally
unscrewed.
5. Mount the light guide by completing the steps
below.
A. Switch off the illumination system.
The light guide comes pre-routed through
the swivel arm (which houses the LED
illumination system) and the suspension arm.
B. Remove the dust cap from the receptacle.
C. Insert the light guide into the receptacle in
the microscope until it clicks into position as
shown in A in Fig. 5.
D. Make sure the light guide has been routed
in such a way that the carrier system and
the surgical microscope are not obstructed
and can be moved in their entire range of
movement without stretching, kinking, or
twisting the light guide.
Safety Screw
2
3
1
Fig 4
Fig 5
6. Install the binocular head and eyepieces on the
magnichanger. Secure the binocular head with
the head locking screw as shown in A in Fig. 6.
A
Fig 6
15
Prima ENT/DNT
8
ELECTRICAL CONNECTIONS
Connect the power cable to the AC inlet socket (1)
located at the back of the swivel arm (see Fig. 7).
Switch on the power using the on/off switch (2).
Note: The power supply is designed with universal input
100V-240V AC, 50/60Hz. To plug in the power supply,
follow the instructions on the electrical label located at
the bottom of the arm (see 3 in Fig. 7).
1. Replacing the Illumination Source
A. Detach the swivel arm covers.
B. Detach the fiber optic cable and replace the
illumination assembly (see A in Fig. 8) with a
new assembly.
C. Resecure the arm covers.
1
2
3
Fig 7
2. Replacing the Fuse
A. Locate the fuse at the AC inlet (see Fig. 9).
B. Use a flat head screw driver to open the fuse
compartment. Two fuses are provided: one is
a live fuse and the second is a spare fuse.
C. Replace the blown fuse with the spare fuse
and close the fuse compartment.
For fuse replacement, refer to the label shown in
Fig. 9.
3.Data Plate Details
UL Mark
Safety Certification
Device Name
Model Number
Serial Number
Input Voltage
A
Fig 8
ON/OFF Switch
A/C Inlet
Fig 9
Accompanying Document
Must Be Consulted
IEC60601-1:2005
Labeling for USA and Europe
IEC60601-1:2005
Labeling for Canada
Fuse Rating
16
Frequency
Max Load
Waste Electric Device
Label Control
Number
Company
ManufacturerSeparate Disposal
Trademark
PRIMA WIRING CODING DIAGRAM
Prima ENT/DNT
17
Prima ENT/DNT
9
1. ON/OFF Switch (see 2 in Fig. 7)
The ON/OFF switch is located at the back of the swivel
arm. In the ON position, a green LED glows, and the
cooling fan starts running. Set the intensity control knob
at the minimum intensity level before switching the
system on.
To preserve the life of the LED, turn off the instrument
when the microscope is not in use.
2. Intensity Control Knob
This knob is located at the front of the suspension arm
(see A in Fig. 10). The brightness of the field of view can
be adjusted as desired using the intensity control knob.
CONTROL ELEMENTS
B
Fig 10
A
3. Swivel Arm Locking Knob
This knob (see B in Fig. 10) locks the movement of the
swivel arm at the desired position by turning it clockwise.
4. Brakes
Lock the stand to prevent unwanted movement by
pressing down on the two brakes provided on the caster
wheels (see Fig. 11). To unlock, press upward on the
brakes.
5. Adjusting Tension on Carrier Rotation
If any optional accessories are added to the microscope,
the tension on the carrier rotation must be adjusted to
compensate for the added weight by completing the
steps below.
A. Unscrew the black cap (see 1 in Fig. 12).
B. Use a 19 mm wrench to turn the screw (see 2
in Fig. 12) to achieve the desired tension.
C. Once the desired tension is achieved, screw
on the black cap (see 1 in Fig. 12).
Fig 11
1
2
Fig 12
18
Prima ENT/DNT
14b
10
WARNING: INSTRUMENT IS UNSAFE FOR MRI ENVIRONMENTS.
Setting Up the Microscope
1. For stability, lock both brakes on the caster
2. Tension in the up and down motion of the
USING THE MICROSCOPE
wheels.
microscope is preset at the factory. However, the
user can adjust the force required to move the
microscope up and down by turning the Tension
Adjustment Screw (see Fig. 13) clockwise or
counterclockwise using an 8 mm allen wrench.
Refer to section 14 for more details.
Adjustment Screw
(A)
IPD Knob
Tension
Fig 13
3. Lock the up and down movement of the
suspension arm using the locking knob (see B
in Fig. 10 above) after focusing on the area of
interest.
4. Adjust the interpupillary distance (IPD) of the
eyepieces as needed, using the IPD adjustment
knob.
5. The illumination is controlled through the
intensity control knob (see A in Fig. 13). Rotate
it clockwise or counterclockwise to achieve the
desired illumination level.
Setting Up Magnification (See Fig. 14)
1 Use the rotating knobs (14a) provided on the
magnification changer to select the desired
magnification level.
2 Use the fine focusing knob or foot pedal to achieve
fine focus.
14a
Fig 14
3 Use the manual handle (14b) to center the area of
interest in the field of view.
19
Make sure that the magnification changer is engaged at
a click stop position (not between click stops).
Main Microscope
Objective lenses with focal lengths of 250 mm, 300
mm, 400 mm, or variable working distance are available, depending on the user’s desired working distance.
A straight binocular head (Prima ENT) or ergo/tilting
binocular head (Prima DNT) is provided.
Standard configurations include eyepieces with a
magnification factor of 10x (12.5x eyepieces are
optional).
Prima ENT/DNT
Fig 15
20
Prima ENT/DNT
11
1. The Common Main Objective (CMO) can be
2. To install the eyepieces, insert them in the eye
3. A range of objectives and eyepieces are
CHANGING THE OBJECTIVES AND EYEPIECES
removed by rotating it in a counterclockwise
direction. It can be attached by rotating it in a
clockwise direction.
reducing strain and fatique.Assembled inbetween magnichanger and observation
head.
2. Rotoplate Rotoplate eliminates neck craning ,making it comfortable to view hard to see
region of mouth from 9 and 3’O clock seated position.
3. DBS and DBSi A beam splitter is inserted between magnichanger and observation head.
Light is divided into two beams. It can be 50:50 or 80:20. In case of 80:20 ratio, 80%
of light goes to observation head and 20% light goes to camera accessories.
4. Double Iris attachment An adjustable diaphragm attachment installed between
magnichanger and observation head /DBS/DBSi. It reduces incoming light and
increased depth of field, it is very useful for photography.
5. The following adapters can be attached to DBS or DBSi
- DSLR camera adapter for Nikon, Canon, and Sony cameras
- CCD camera adapter for Mintron camera
- Adapter for Sony Handy-Cam
- Assistoscope for assistant viewing
6. Attach the accessory to the left or right side of the beam splitter.
22
CMO 250 CMO 350
CMO 400
STRAIGHT
BINOCULAR TUBE
Prima ENT/DNT
Part No.: 6137000-000
Model: Prima ENT
DBS
FLOOR MOUNT
FOOT CONTROL Z AXIS
Fig 18
CCD CAMERA
ADAPTER
ASSISTOSCOPE
DOUBLE IRIS
SONY HANDYCAM
ADAPTER
DSLR CAMERA
ADAPTER
1. The following accessories are provided
- DSLR camera adapter for Nikon, Canon, and Sony cameras
- Adapter for Mintron CCD camera
- Adapter for Sony Handy-Cam
- Assistoscope for assistant viewing
2. Attach the accessory to the left or right side of the beam splitter.
3. DBS A beam splitter is inserted between magnichanger and observation head. Light is
divided into two beams. It can be 50:50 or 80:20. In case of 80:20 ratio, 80% of light
goes to observation head and 20% light goes to camera accessories.
4. Double Iris attachment An adjustable diaphragm attachment installed between
magnichanger and observation head /DBS/DBSi. It reduces incoming light and
increased depth of field, it is very useful for photography.
23
Prima ENT/DNT
13
THERMAL CUT-OFF
The instrument is designed with safety provisions. Fans in the electrical box provide freeand forced-air circulation to cool the electronic components. The instrument also includes a
built-in safety mechanism called “auto thermal cut-off.” This mechanism is activated when
the LED temperature rises above 70°C, at which time power to the LED will be cut off. The
LED indicator light will glow orange when auto thermal cut off has been initiated. Allow the
LED to cool down before switching it on again.
14
TENSION ADJUSTMENT
After any optional accessories are added to the
microscope, suspension arm tension must be adjusted
to compensate for the additional weight, as follows.
1. Remove the plate as shown by removing two
screws (see Fig. 19).
2. Loosen the two allen screws as much as
possible using the 4 mm allen wrench (see Fig.
20). Note that these two screws will not come
out.
3. Use an 8 mm allen wrench in bolt (A) (see Fig.
21). Rotate it clockwise to increase the tension
as desired. Rotate it counterclockwise to
decrease the tension.
Fig 19
Fig 20
E
U
Q
R
O
T
Loosen two screws
4. After achieving the desired tension, retighten
the two screws shown in Fig. 20.
5. Reinstall the plate shown in Fig. 19.
24
A
Fig 21
E
U
Q
R
O
T
Prima ENT/DNT
15
MOVING THE INSTRUMENT
Fold the system
as shown before
moving
Moving the Stand
1. Turn off the unit at the power switch.
2. Disconnect the power cable.
3. Remove the video cable from the video modules (e.g., video monitor, USB monitor) and
the camera control unit.
4. Release the brakes.
5. Be careful when passing through doorways to avoid collisions of any kind. Do not move
the instrument over steps or cables as the stand can topple. Be extremely careful when
moving the instrument over slopes. Do not park the instrument on a slope.
6. Press the brakes down to lock the wheels.
25
Prima ENT/DNT
16
To ensure optimal performance of your instrument, we recommend annual preventative
maintenance by an authorized LABOMED dealer. If a failure occurs which cannot be
corrected using the troubleshooting table, attach a sign to the instrument stating that it is
“out of order” and contact your dealer for assistance.
Care Instructions
· Keep accessories protected from dust when not in use (e.g., use a dust cover).
· Use a microfiber cloth and pure 99% isopropyl alcohol for cleaning lenses and eyepieces.
· Protect your microscope from moisture, fumes, acids, and cosmetic materials. Do not
store chemicals close to the instrument.
· Protect the instrument from improper handling. Never install other devices or unscrew
the optical system and mechanical parts unless explicitly instructed to do so in this user
manual.
· Protect the microscope from oil and grease. Never apply oil or grease to surfaces or
mechanical parts.
· Use disinfectants based on the following ingredients: aldehydes, alcohols, and quaternary
ammonium compounds.
· Camera: Keep the optical components clean using a microfiber cloth. Soak the cloth using
methanol or glass cleaner. Do not use ethanol or 99% isopropyl alcohol.
· Do not clean products with optical components in a cleaning/disinfecting device or
ultrasound bath.
CARE AND MAINTENANCE
Tropical Environment/Fungus
LABOMED employs certain safety precautions in its manufacturing techniques and
materials. Other preventive measures include:
· Keeping optical parts clean.
· Using and storing the instrument in a clean environment only.
· Storing the instrument under UV light when not in use.
· Using the microscope in continuously climate-controlled rooms only.
· Keeping moisture away from the instrument, and covering it using a plastic cover filled
with silica gel.
Occupational Safety and Health Protection
· Observe the work safety and health protection of individuals responsible for processing
contaminated products.
· Current regulations of hospital hygiene and prevention of infection must be observed in
the preparation, cleaning, and disinfection of the products.
Workplace
· Remove surface contamination with a paper towel.
26
Prima ENT/DNT
Cleaning and Servicing
· Wipe the instrument and stand (no optics) with a damp cloth regularly. Use standard mild
household detergents, diluted as necessary. Then wipe dry with a soft cloth.
·Do not use harsh or corrosive cleaning agents. They can cause damage to the paint and
plastic parts.
·Avoid splashing water, as it may cause damage to the device (per the safety clause in
IPXO).
· Use 99% isopropyl alcohol to clean the optical components.
· Dry the optical components using a microfiber cloth; dry the rest of the product using a
paper towel.
· Contact LABOMED or authorized service personnel for any service needs.
27
Prima ENT/DNT
17
For operationTemperature
For transportation and
storage
The unit meets the essential requirements stipulated in Annex I of the 93/42/EEC
directive governing medical devices. The unit is marked with CE and is compliant to
ANSI / AAMI EC 60601-1:2005.
AMBIENT REQUIREMENTS
Rel. humidity
(without condensation)
Air pressure
Temperature
Rel. humidity
(without condensation)
Air pressure
+10°C....+40° C
30%......90%
700hPa......1,060hPa
-40°C......+70°C
10%.......100%
500hPa.......1,060hPa
28
Prima ENT/DNT
18
Disposal of the instrument must comply with locally applicable laws and regulations.
DISPOSAL
29
Prima ENT/DNT
19
ProblemPossible CauseRemedy
No illuminationPower cable not plugged inPlug in power cable
TROUBLESHOOTING TABLE
Power switch in OFF positionPress the power switch to
Defective instrument fuseChange the fuse
Defective power cableChange the power cable
Line power failureContact in-house technician
Failure of suspension system
electronics
Light guide not properly
inserted in arm or microscope
ON position
Contact the service
department
Insert the light guide
properly to get maximum
illumination
Insufficient illuminationBrightness level set too lowAdjust the brightness
control knob
Inoperative surgical
field illumination
Light guide not properly
inserted in arm of
microscope
Defective light guide
(illumination not uniform)
Inoperative surgical field
illumination
Failure of electronicsIlluminate the surgical
Switch off via microswitch on
suspension system
Insert the light guide
properly to get maximum
illumination
Change the light guide
Insert the light guide as far
as it will go
field using an alternate
illuminator, and contact the
service department
Move the suspension arm of
the suspension system into
the working position
30
ProblemPossible CauseRemedy
Prima ENT/DNT
Insufficient illumination
(continued)
Up and down motion of the
suspension system is stiff
Stand is unstable The brakes on the wheels are
No image visible in the
field of view
The thermal cut-off in the
lamp housing is contaminated
Defective fan; failure of
system electronics
The friction adjustment screw
on the suspension system is
tightened too firmly
not in use
Magnichanger is not indexed
properly.
Clean the thermal cut-off
with a dry brush or blow it
clean, with compressed air
Contact the service
department
Loosen the friction
adjustment screw on the
suspension system as
needed
Group 1 The Prima ENT/DNT uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
Class AThe Prima ENT/DNT is suitable for use in all
establishments, other than domestic, and those directly
connected to the public low voltage power network
that supplies buildings used for domestic purposes
Complies
33
Prima ENT/DNT
CONTINUED...
Guidance and Manufacturer’s Declaration
Electromagnetic Emissions
All Equipment and systems
Guidance and Manufacturer’s Declaration -Electromagnetic Emissions
The Prima ENT/DNT is intended for use in the electromagnetic environment specified
below. The customer or user of the Prima ENT/DNT should ensure that it is used in such an
environment.
Immunity testIEC 60601Test
Level
ESD IEC 61000-
4-2
EFT IEC 61000-
4-4
Surge IEC
61000-4-5
Voltage Dips/
Dropout IEC
61000-4-11
±6kv Contact
±8kv Air
±2kv Mains ±1kv
I/Os
±1kv Differential
±2kv Common
>95% Dip for 0.5
Cycle 60% Dip
for 5 cycles 30%
Dip for 25 Cycles
> 95% Dip for 5
Second
Compliance
Level
±6kv Contact
±8kv Air
±2kv Mains ±1kv
I/Os
±1kv Differential
±2kv Common
>95% Dip for 0.5
Cycle 60% Dip
for 5 cycles 30%
Dip for 25 Cycles
> 95% Dip for 5
Second
Electromagnetic Environment-
Guidance
Floors should be wood, concrete, or
ceramic tile. If floors are synthetic,
the R/H should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
Prima ENT/DNT requires continued
operation during power mains
interruptions, it is recommended
that the Prima ENT/DNT be
powered from an uninterruptible
power supply or battery.
Power Frequency
50/60 Hz
Magnetic Field
IEC 61000-4-8
34
3A/m3A/mPower frequency magnetic
fields should be that of a
typical commercial or hospital
environment.
Prima ENT/DNT
CONTINUED...
Guidance and manufacturer’s Declaration
Electromagnetic Immunity
Equipment and Systems that are NOT Life-supporting
Guidance and manufacturer’s Declaration - Electromagnetic Immunity
The Prima ENT/DNT is intended for use in the Electromagnetic environment specified below. The customer or
user of the Prima ENT/DNT should ensure that it is used in such an environment.
Immunity TestIEC 60601 Test Level Compliance
Level
Conducted RF IEC
61000-4-6
Radiated RF IEC
61000-4-3
3 vrms 150 kHz to 80
MHz
80MHz to 2.5 GHz @
3V/m
(v1) = 3 vrmsPortable and mobile RF Communications equipment
(E1) = 3 vrms
Electromagnetic environment Guidance
should be used no closer to any part of the Prima
ENT/DNT, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended Separation Distance:
d=(3.5/v1)(Sqrt P)
d=(3.5/E1)(Sqrt P)
80 to 800 MHz
d=(7/E1)(Sqrt P)
800 MHz to 2.5 GHz
Where P is the max output power rating of
the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance levels in each
frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol
Note 1: At 80 MHz to 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects, and people.
* Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast, and Tv broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. The measured field strength in the
location in which the ME Equipment or ME System should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the ME Equipment or ME System.
* Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
35
Prima ENT/DNT
CONTINUED...
Recommended Separation Distances between Portable and Mobile RF Communications Equipment
and the Prima ENT/DNT for ME Equipment and ME systems that are NOT Life-supporting.
Guidance and Manufacturer’s Declaration -Electromagnetic Immunity
Recommended Separation Distances for between
Portable and Mobile RF Communications Equipment and the Prima ENT/DNT
The Prima ENT/DNT is intended for use in electromagnetic environments in which radiated
RF disturbances are controlled. The customer or user of the Prima ENT/DNT can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF Communications Equipment (transmitters) and the Prima ENT/DNT as recommended below,
according to the maximum output power of the communications equipment.
Max Output Power
of Transmitter(W)
0.010.11660.11660.2333
0.10.36890.36890.7378
11.16661.16662.3333
103.68933.68937.3786
10011.666611.666623.3333
For transmitters rated at a maximum output power not listed above, the recommended separation
distance (d) in meters(m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts(w) according to
Note 1: At 80 MHZ, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
Separation
(m) 150kHz
to 80 MHZ
d=(3.5/v1)
(Sqrt P)
Separation(m)
80 to 800 MHZ
d=(3.5/E1)(Sqrt P)
the transmitter manufacturer.
Separation (m) 800 MHZ to 2.5Ghz d=(7/E1)
(Sqrt P)
36
Prima ENT/DNT
Ceiling Mount
Wall Mount
Floor Mount
135
135
150
180
90
150
180
90
90
150
180
135
22
DIMENSIONS AND WEIGHT
Prima ENT/DNT- Microscope
with Mobile Stand
Total Weight: Approx. 98 kgs.
37
Prima ENT/DNT
23
Ametropia
Compensation
Working Distance Distance from the common main objective (CMO) to the object or area of
Color TemperatureRefers to the color characteristic of a light source. Using color
Field of View Diameter The visible area of an object that can be seen through the microscope.
Green Filter A color filter that darkens red and blue light and brightens green light. This
GLOSSARY
Compensation of short- or far-sightedness. This can be done for each eye
using the two individual eyepieces (range: +5 to -5 diopters).
interest.
temperature, one can set the color of a light source to ‘warm’ or ‘cold’
light relative to the color of natural light. The unit of measure for color
temperature is Kelvin (K).
The greater the magnification level, the smaller the field of view and vice
versa.
improves the contrast of an image with red tones, improving the visibility
of blood vessels.
Illumination IntensitySpecifies the luminous flux from a light source onto a certain area. The
unit of measure for illumination intensity is lux (lx).
LED Light Emitting Diode. Electronic semiconductor device that emits light
when an electrical current passes through it.
EyepieceThe optical tubes through which the enlarged image produced by the
microscope can be viewed.
Convergent Beam PathThe two light beams for the right and left eye run together at a point that
lies at a distance from the common main objective (CMO). This distance is
equal to the working distance.
White Balance White balance is used to calibrate the camera to the color temperature of
the light at the location.
DSLR Camera(Digital single lens reflex) Camera with a digital imaging sensor.
CCD Camera(Charge-coupled device) Camera with a technology to store a charge and
move this charge out of the photo sensor in an organized way.
CMOS Camera(Complementary Metal Oxide Semiconductor) Camera in which image
sensor is a silicon chip that captures and reads light.
DBS (Double Beam Splitter) Splits light beam into two directions (one to eye
and one to side port) for simultaneous user viewing and photography,
videography, or co-observation. Ratio of light distribution is 70% for eyes
and 30% to side ports for photography, videography, or co-observation.
RotoplateThe rotoplate reduces neck craning, making it comfortable to view hard-to-
see regions of the mouth from the 9 and 3 o’clock seated positions.
Double Iris Diaphragm
Attachment
IPDInter pupillary distance
The double Iris diaphragm kit allows greater depth of field, which is
particularly valuable for photography.
38
Prima ENT/DNT
24
This product is warranted by Labotech Microscopes India Pvt. Ltd. against defective material and workmanship
under normal use for a period of one year from the date of invoice to the original purchaser. (An authorized
dealer shall not be considered an original purchaser). Under this warranty, Labotech Microscopes India Pvt.
Ltd. sole obligation is to repair or replace the defective part or product at Labotech/LABOMED discretion.
This warranty applies to new products and does not apply to a product that has been tampered with, altered in
any way, misused, damaged by accident or negligence, or which has had the serial number removed, altered, or
effaced. Nor shall this warranty be extended to a product installed or operated in a manner not in accordance
with the applicable LABOMED instruction manual, nor to a product which has been sold, serviced, installed,
or repaired other than by a Labotech Microscopes India Pvt. Ltd. factory or authorized LABOMED Dealer.
Charts, cards, and other expendable items are not covered by this warranty.
All claims under this warranty must be in writing and directed to the LABOMED factory or the authorized
instrument dealer making the original sale and must be accompanied by a copy of the purchaser’s invoice.
WARRANTY
This warranty is in lieu of all other warranties implied or expressed. All implied warranties of merchantability or
fitness for a particular use are hereby disclaimed. No representative or other person is authorized to make any
other obligations for a LABOMED product. Labotech/LABOMED shall not be liable for any special, incidental,
or consequent damages for any negligence, breach of warranty, strict liability, or any other damages resulting
from or relating to design, manufacture, sale, use, or handling of the product.
PRODUCT CHANGES
Labotech Microscopes India Pvt. Ltd. reserves the right to make changes in design or to make additions to or
improvements in its products without obligation to add such to products previously manufactured.
CLAIMS FOR SHORTAGES
We use extreme care in selection, checking, rechecking, and packing to eliminate the possibility of error. If any
shipping errors are discovered:
1. Carefully go through the packing materials to be sure nothing was inadvertently overlooked when the unit
was unpacked.
2. Call the dealer you purchased the product from and report the shortage. The materials are packed at the
factory and none should be missing if the box has never been opened.
3. Claims must be filed within 30 days of purchase.
CLAIMS FOR DAMAGES IN TRANSIT
Our shipping responsibility ceases with the safe delivery in good condition to the transportation company.
Claims for loss or damage in transit should be made promptly and directly to the transportation company.
If, upon delivery, the outside of the packing case shows evidence of rough handling or damage, the transportation
company’s agent should be requested to make a “Received in Bad Order” notation on the delivery receipt. If
within 48 hours of delivery, concealed damage is noted upon unpacking the shipment and no exterior evidence
of rough handling is apparent, the transportation company should be requested to make out a “Bad Order”
report. This procedure is necessary in order for the dealer to maintain the right of recovery from the carrier.
39
www.laboamerica.com
Our policy is one of continuous development. Labo America, Inc., reserves the right to change design and specifications
without prior notice.
Labo America Inc.
920 Auburn Court
Fremont, CA 94538
T 510 445 1257
F 510 445 1317
sales@laboamerica.com
U.S.A.
LABOMED and Prima ENT/DNT are registered trademarks of Labo America, Inc.
With a policy of continuous development, Labo America, Inc. reserves the right to change design and specifications without prior notice.