To ensure proper use of this instrument as well as to avoid injury while operating instrument,
understanding this manual completelybefore use is highly recommended.
Part No: 6167500-795
Issue1.0
Printed on: May 2013
Prima DNT Trainer
Warnings & Cautions
LABOMED is not responsible for the safety and reliability of this instrument when:
•Assembly, disassembly, repair, or modification is made by unauthorized dealers or
persons.
•Instrument is not used in accordance with this User’s Guide.
WAR NING :AN INST RU CT ION THAT DRAWS AT TENTION TO RISK OF I NJURY OR DEAT H.
WARNING: UNITED STATES FEDERAL LAW AND EUROPEAN REGULATIONS
REQUIRE THAT THIS DEVICE BE PURCHASED ONLY BY A PHYSICIAN OR A
PERSON ACTING ON BEHALF OF A PHYSICIAN.
WARNING: THIS INSTRUMENT SH OULD BE USED IN STRICT ACCORDANCE WITH
THE INSTRUCTIONS OUTLINED IN THIS USER’S GUIDE. THE SAFETY OF THE
OPERATOR AND THE P ERFORMANCE OF THE INSTRUMENT CANNOT BE
GUARANTEEDIFUSEDINAMANNERNOTSPECIFIEDBYLABOMED
TECHNOLOGIES.
WARNING: DO NOT REPAIR OR SERVICE THIS INSTRUMENT WITHOUT
AUTHORIZATION FROM THE MANUFACTURER. ANY REPAIR OR SERVICE TO
THIS INSTRUMENT M UST BE PERFORMED BY EXPER IENCED PERSONNEL OR
DEALERS WHO ARE TRAINED BY LABOMED OR SERIOUS INJURY TO THE
OPERATOR OR PATIENT MAY OCCUR.
WARNING: MODIFICATIONS TO THIS INSTRUMENT ARE NOT ALLOWED. ANY
MODIFICATION TO THIS UNIT MUST BE AUTHORIZED BY LABOMED OR SERIOUS
INJURY TO THE OPERATOR OR PATIENT MAY OCCUR.
WARNING: IF THIS INSTRUMENT IS MODIFIED, APPROPRIATE INSPECTION AND
TESTING MUST BE CONDUCTED TO ENSURE CONTINUED SAFE USEOF THIS
INST RU M ENT.
WARNING : T O AVOI D R ISK OF ELECT RI C SHOCK, THIS EQUIPMENT M UST ON LY
BE CONNECTED TO A SUPPLY MAIN WITH PROTECTIVE EARTH OR DAMAGE TO
THIS I NST RUMENT AN D/ OR I NJURY TO THE OPER ATOR OR PATIEN T MAY OCC UR.
WARNING : ENSURE THAT THE VOLTAGE APPLIED TO THE UNIT IS THE SAME AS
THE VOLTAGE THAT IS INDICATED ON THE DATA PLATE OR DAM AGE TO THE UNIT
MAYOCCUR.
WARNING : THIS INSTRUMENT MUST BE PLUGGED INTO AN OUTLET WITH AN EARTH
GROUND. DO NOT REMOVE OR DEFEAT THE EARTH GROUND CONNECTION ON
POWER INPUT CONNECTOR OR THE UNIT’S POWER CORD OF THIS INSTRUMENT OR
DAMAGE TO IT AND/OR INJURY TO THE OPERATOR OR PATIENT MAY OCCUR.
WARNING : THE E QUIP MENT OR S YS TEM SHOULD NOT BE USED ADJACENT TO OR
STACKED W ITH OTHER EQUIPMENT AND THAT IF ADJACENT OR STACKED U SE IS
NECESSARY, THE EQUIPM ENT OR SYSTEM SHOULD BE OBSERVED TO VERIFY
NORMAL OPERATION IN THE CONFIGURATION IN WHICH IT W ILL BE USED.
WARNING : THIS INSTRUMENT IS NOT SUITABLE FOR USE IN THE
PRESENCE OF FLAMMABLE ANESTHETIC MIXTURES, SUCH AS OXYGEN OR
NITROUS OXIDE.
WARNING : B ECAUSE PROLONGED INTENSE LIGHT EXPOSURE CAN DAMAGE THE
RETINA, THE USE OF THE DEVICE FOR OCULAR EXAMINATION SHOULD NOT BE
UNNECESSARILY PROLONGED, AND THE BRIGHTNESS SETTING SHOULD N OT
EXCEED WHAT IS NEEDED TO PROVIDE CL EAR VISUALIZATION OF THE TAR GET
STRUCTURES. THIS DEVICE SHOULD BE USED WITH FILTERS THAT ELIMINATE UV
RADIATION <400 NM) AND, W HENEVER POSSIBLE, F ILTERS THAT ELIMINATE
Issue 1.0
Prima DNT Trainer
Warnings& Cautions (continued)
CAUTION:AN INST RUCT ION THAT DRAWS ATTENTION TO TH E RI SK OF D AMAG E T O TH E
PRODUCT.
CAUTION :TH EINTERN ALCIRC UITR YO FTH EINSTR UMENTCONTAIN S
ELECTROSTATIC DISCHARGE SENSITIVE DEVICES (ESDS) THAT MAY BE SENSITIVE
TO STATIC C HARGES PRODUCED BY THE HUMAN BODY. DO NOT REMOVE THE
COVERS WITHOUT TAKING PROPER ESDS PRECAUTIONS.
CAUTION: DO NOT USE SOLVENTS OR STRONG CLEANING SOLUTIONS ON ANY PART
OF TH IS INSTRUMEN T AS DA MAGE TO THE U NI T M AY OCCUR. SEE MAINTENANCE
SECTION FOR DETAILED CL EANING INSTRUCTION.
CAUTION: MEDICAL ELECTRONIC EQUIPMENT NEEDS SPECIAL PRECAUTIONS
REGARDING EMC AND NEEDS TO BE INSTALLED AND PUT INTO SERVICE
ACCORDING TO THE EM C I NFORMATION PROVIDED IN THE ACCOMPANYING
DOCU MENTS.
CAUTION: PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT C AN
AFFECT MEDICAL ELECTRICAL EQUIPMENT.
CAUTION: T HI S I NSTRUMEN T IS NOT TO BE US ED N EAR HIGH-FREQUEN CY
EMITTING SURGICAL EQUIPMENT.
CAUTION: T HI S I NSTRUMEN T I S N OT IN TENDED TO B E CONNEC TED TO
EQUIPMENT OUTSIDE THE CONTROL OF LABOMED OR M UST BE TESTED TO AN
APPLICABLE IEC OR ISO STANDARDS.
Issue 1.0
CONTENTS
-Key to symbols
Prima DNT Trainer
-
-Safety Instructions2
-Special Instruction for Installation & Use3-4
-
-
-Description of Parts7
-System Diagram8-9
-
Introduction1
W
arning labels and notes5
Unpacking6
Installation & Instruction
Installation
Electric Connection
Controls
Instruction for using the microscope
Changing the Objectives & Eyepieces
Replacing the Illumination source
Adjustment of tension
10-11
12
12
13-14
15
15
16
-
-
-Disposal21
-Specification22
-Dimensions23
Cleaning & Servicing17
Ambient Requirement18
Troubleshooting T
able19-20
Issue 1.0
Prima DNT Trainer
1
The LABOMED Prima DNT Trainer is designed to be used during teaching program in restorative
dentistry, which is adaptable for different surgical needswithout compromise to performance.
The microscope provides extremely high optical image quality, good depth of focus and wide field
of view for precise surgery. Illumination control, inbuilt tilt, adjustment of the observation head help
to reduce the surgeon’s work fatigue and allow comfortable use over long period.
Salient features of this Microscope are:
1. The observation head can easily be positioned with the help of suspension arm .
2. An advanced 5-step magnification changer allows an optimal magnification for a particular
surgery from five different magnifications.
3. Cold light illumination with a high intensity 50W LED lamp is provided using a fiber optic guide
for proper illumination. The illumination is further adjustable up to its most suitable brightness
using intensity control knob suitably located at the suspension arm, and is easily approachable
to the surgeon.
INTRODUCTION
4. When the microscope is not in use, the suspension arm can be folded over the main body to
store it compactly.
5. The microscope provides extremely high optical image quality, good depth of focus and wide
field of view for precise surgery. Illumination control, inbuilt tilt, adjustment of the observation
head help to reduce the surgeon’s work fatigue and allow comfortable use over long period.
1
Issue 1.0
Prima DNT Trainer
SAFETY INSTRUCTIONS
2
Safety
This instrument described in this manual has been developed and tested in accordance with
LABOMED safety standards and with national and international regulations. A high degree on
instrument safety is thus ensured.
This manual contains information on safety aspects governing the use of this product and brief
summary of important precaution for this user.
Additional safety notes are also contained in other parts of this user’s manual; they are marked
with a warning triangle containing an exclamation mark as shown here. Please pay special
attention to these safety notes.
Safety is only ensures when this instrument is operated properly. Please read this manual
carefully before turning the instrument on. Also read through the user’s manuals on the other
equipment used with this instrument. You may obtain further information from our service
organization or authorization representative.
Directives and standards
The instrument in this manual has been designed in the compliance with the following standards
- EN
- IEC
- UL
- CSA
-As per Directive 93/42/EEC, the unit is a Class I intrument
- For USA: FDA classification Class I
- IEC 60601-1:2007(3rd edition) Compliance
Please observe all applicable accidental prevention regulations.
2
Issue 1.0
Prima DNT Trainer
SPECIAL INSTRUCTIONS for INSTALLATION & USE
3
Safe Working Order
Do not operate the equipment contained in the delivery package in
-Areas having explosion or inflammability, risk the presence of inflammable anesthetics or
volatile solvents such as alcohol, benzene or similar chemicals.
Do not install, store or use the instrument in damp rooms. It is important and cautionary
to protect the instrument from being exposed to humidity, wet conditions, water splashes
or water sprays.
Do not use the instrument if you notice any electrical arcing, sparks, abnormal noises,
smokes or fumes. Unplug it immediately and contact authorized LABOMED service
representative.
Ensure that the power point is properly grounded
Do not force cable connections. If the male and female parts do not readily connect,
make sure that they are appropriate for one another. If any of the connectors and
damaged, have our service representative repair them.
The effects of Radio waves and interference on medical imaging systems and equipment
are unpredictable. It is cautionary to mention that mobile phones or radios not be used in
the proximity of the equipment.
Anything concerning repairs, customization and upgradation this instruction must be
carried out by LABOMED authorized representative failing which, LABOMED product
warranty will not be applicable.
The instrument must be operated for its intended use and only after duly understanding
the operation instructions or undergoing training on use of the instrument by a LABOMED
authorized representative.
Only use the instrument with the accessories supplied. Should you wish to use other
accessories equipment, make sure that LABOMED or the equipment manufacturer has
certified that its use will not repair the safety of the instrument.
It is recommended that the user’s manual be kept accessible always.
Do not pull at the light source cable, at the power cord or at other cable connections.
It is recommended to get the instrument checked once every 12 months to ensure its
optimum performance.
3
Issue 1.0
Prima DNT Trainer
Requirement for operation
Installation of the instrument must only be carried out by an authorized LABOMED agent or
representative. Please make sure that the following requirements for operation remain fulfilled
in the future:
- All mechanical connections(details in the user’s manual) which are relevant to safety
and properly connected and screw connections tightened.
- All cable and plugs are in good working condition.
- The voltage setting on the instruments conforms to the rated voltage of the line
supply on site.
-The power cord being used in the one designed for use with this instrument.
Before every use after re-equipping the instrument
Make sure that all “Requirement for operation” are fulfilled.
Go through the checklist
Re- attach or close any covers, panels or caps which have been removed or opened.
Pay special attention to warning symbols on the instrument (triangular warning signs
with exclamation marks), labels and any parts such as screws or surfaces painted red).
For every use of the instrument
Avoid looking directly into the light source, e.g. into the microscope objective lens or a
light Guide.
All light radiation have an effect on skin and biological tissues. For your own protection
and advices to keep the light level to the absolute minimum required for your procedure.
Warning!
Only the right configuration of LABOMED Prima DNT Trainer surgical microscope must
be used for related procedure.
4
Issue 1.0
safety compliance
brand name
Prima DNT Trainer
for the suspension system about the admissible maximum load.
Com pliance to Medical Device Direct ive 93/42/EE C
Accompanying Document must be consulted
REF
Catalog Number
Protective Earth
5
of 7.5 kg. Please consult the user manual
Issue 1.0