Labnet Z 300 User manual

INSTRUCTION MANUALINSTRUCTION MANUAL
INSTRUCTION MANUAL
INSTRUCTION MANUALINSTRUCTION MANUAL
FORFOR
FOR
FORFOR
BENCH BENCH
BENCH
BENCH BENCH
TT
T
OP CENTRIFUGE
TT
Z 300Z 300
Z 300
Z 300Z 300 Z 400Z 400
Z 400
Z 400Z 400
Labnet International
PO Box 841
Woodbridge, NJ 07095
Phone: 732-417-0700
Fax: 732-417-1750
Z 300 / Z 400 / Jan. 2002
INDEXINDEX
INDEX
INDEXINDEX
1.1.
1.
1.1.
2.2.
2.
2.2.
3.3.
3.
3.3.
4.4.
4.
4.4.
5.5.
5.
5.5.
6.6.
6.
6.6.
7.7.
7.
7.7.
8.8.
8.
8.8.
9.9.
9.
9.9.
........................................................................................................................................................................................................................................................................
....................................................................................................................................
........................................................................................................................................................................................................................................................................
General InformationGeneral Information
General Information
General InformationGeneral Information
1.1 Precautions and hazards .............................................................................. 4 E
1.2 Description .................................................................................................... 5 E
1.3 Safety standards ........................................................................................... 5 E
1.4 Technical data Z 300 .................................................................................... 6 E
Technical data Z 400 .................................................................................... 7 E
1.5 Accessories supplied with each centrifuge unit ............................................ 8 E
1.6 Warranty........................................................................................................ 8 E
InstallationInstallation
Installation
InstallationInstallation
2.1 Unpacking the centrifuge .............................................................................. 8 E
2.2 Transport ....................................................................................................... 8 E
2.2 Required space ............................................................................................. 8 E
2.3 Installation ..................................................................................................... 9 E
How to install and load a rotorHow to install and load a rotor
How to install and load a rotor
How to install and load a rotorHow to install and load a rotor
3.1 Mounting and securing a swing out rotor...................................................... 9 E
3.2 Mounting and securing an angle rotor .......................................................... 10 E
3.3 Mounting and securing a hematocrit rotor .................................................... 10 E
3.4 Overloading of rotors ..................................................................................... 11 E
3.5 Removing of the rotor .................................................................................... 11 E
OperationOperation
Operation
OperationOperation
4.1 Power switch ................................................................................................. 12 E
4.2 Lid release..................................................................................................... 12 E
4.3 Lid lock .......................................................................................................... 12 E
4.4 Preselection of speed ................................................................................... 13 E
4.5 Preselection of operating time ...................................................................... 14 E
4.6 Keyboard - Starting the centrifuge - "quick"-key........................................... 14 E
4.7 Key "stop"...................................................................................................... 14 E
TemperatureTemperature
Temperature
TemperatureTemperature
5.1 Temperature.................................................................................................. 15 E
Safety facilitiesSafety facilities
Safety facilities
Safety facilitiesSafety facilities
6.1 Imbalance...................................................................................................... 15 E
Service and MaintenanceService and Maintenance
Service and Maintenance
Service and MaintenanceService and Maintenance
7.1 Service and inspection of the centrifuge ....................................................... 16 E
7.2 Cleaning the centrifuge ................................................................................. 16 E
7.3 Cleaning the centrifuge after breakage of glass tubes/bottles ..................... 17 E
7.4 Disinfection.................................................................................................... 17 E
BreakdownBreakdown
Breakdown
BreakdownBreakdown
8.1 Emergency lid release ................................................................................... 17 E
8.2 Check list / Trouble shooting......................................................................... 18 E
AppendixAppendix
Appendix
AppendixAppendix
9.1 Chart for determining g-values...................................................................... 22 E
PagePage
Page
PagePage
1.1.
General InformationGeneral Information
1.
General Information
1.1.
General InformationGeneral Information
1.1 Precautions and hazards
Before putting the centrifuge into operation, please read this instructionBefore putting the centrifuge into operation, please read this instruction
Before putting the centrifuge into operation, please read this instruction
Before putting the centrifuge into operation, please read this instructionBefore putting the centrifuge into operation, please read this instruction manual carefully.manual carefully.
manual carefully.
manual carefully.manual carefully.
The centrifuge must not be operated by unqualified persons not familar with the correct use and intended purpose of the machine. Please use only the original spare parts. For personal and enviromental safety, special attention has to be paid to the following precautions:
The Hermle Centrifuges are neither explosion proof nor intert gas shielded and should therefore never be operated in explosion-hazardais locations. Never stay in the safety zone of 30 cm around the centrifuge or deposite dangerous goods inside this zone during centrifugation. The centrifugation of the flammable or explosive or radioactive samples is not allowed. Further more do not spin samples reacting chemically with each other with high energy when explosed to air. Never spin toxical or phatological material without adequate safety precautions i.e. centrifuging of buckets/tubes without or with defective hermetic sealing is not allowed. The enduser is engaged to performe appropriate disinfection procedures in case dangerous goods have contaminated the centrifuge or its accessories. The general universal laboratory precautions should be observed in case infection materials are centrifuged. If necessary, please contact your local security officer! It is prohibited to run the centrifuge with rotors not suited for this centrifuge model. Under no circumstances, the lid at the centrifuge should be opened while the rotor is still turning respectively running with a speed of more than 2 meter per second.
The following rules must be strictly adhered to:The following rules must be strictly adhered to:
The following rules must be strictly adhered to:
The following rules must be strictly adhered to:The following rules must be strictly adhered to:
Do not operate the centrifuge in case it is not installed correctly. Never operate the centrifuge in a dismounted state ( e.g. without covering sheet metal) Do not run the centrifuge when electrical or mechanical assembly groups have been tampered with by unauthorized persons. Never use accessories such as rotors and buckets which are not exclusively approved by centrifuge tubes of glas or plastic. Do not spin corrosive samples which may cause damages in material and impair the mechanic resistance. never operate the centrifuge with rotors or buckets which show any signs of corrosion or mechanical damage
The manufacturer is only responsible for the security and reliability of theThe manufacturer is only responsible for the security and reliability of the
The manufacturer is only responsible for the security and reliability of the
The manufacturer is only responsible for the security and reliability of theThe manufacturer is only responsible for the security and reliability of the centrifuge if:centrifuge if:
centrifuge if:
centrifuge if:centrifuge if:
The unit is operated according to the instruction manual. Modifications, repairs and new adjustments are performed by HERMLE authorized persons and the electrical installation of the location where the centrifuge is operated corresponds to the IEC-regulations.
Hermle Labortechnik GmbHHermle Labortechnik GmbH
Hermle Labortechnik GmbH, except commercially available
Hermle Labortechnik GmbHHermle Labortechnik GmbH
1.1 Description
The model Z 300 and Z 400 are a universal centrifuge which covers many fields of applications by offering a wide range of accessories. The centrifuges are equipped with the Hermle Standart-Control panel. Speed and running time are set with easy to use control knobs. The precise parameters selected are shown on the large digital LED display.
1.3 Safety standards
The centrifuge corresponds to the general requirements set by German law for medical apparatus, "MedGV" group 3. The following standards have been considered for the production of our centrifuges:
- Accident prevention rules for centrifuges, UVV-VBG 7z.
- Accident prevention rules for electrical equipment & installations, UVV-VBG 4.
- International Standart IEC 1010-1 and IEC 1001-2-D
- European Standart PR EN 61010-1 and PR EN 61010-2-2
- Electrical interference suppression according to interference degree B as per VDE 0871.
1.41.4
Technical dataTechnical data
1.4
Technical data
1.41.4
Technical dataTechnical data
______________________________________________________________________
Manufacturer Hermle Labortechnik GmbH
______________________________________________________________________
Type Z 300
______________________________________________________________________
Dimensions: Width 35,5 cm Depth 33 cm Height 47,4 cm
______________________________________________________________________
Weight 27 kg
______________________________________________________________________
Noise level 65 dB(A)
______________________________________________________________________
Max. speed 13.500 rpm Max. Volume 400 ml Max. RCF 17310 x g Admiss. density 1,2 kg/dm
3
Admiss. kinetic energy 3 052 Nm
______________________________________________________________________
Electrical connection 230 V/50 Hz 1Ph 120 V/60 Hz 1Ph Current 3,8 A 4,0 A Connected load 580 Watt 460 Watt
______________________________________________________________________
Interference suppression VDE 0871, interference degree B
______________________________________________________________________
Service dept. at Hermle 0 74 26 / 96 22 55
______________________________________________________________________
Address of service: _________________________________
_________________________________ _________________________________
______________________________________________________________________
Address of agent: _________________________________
_________________________________ _________________________________
1.41.4
Technical dataTechnical data
1.4
Technical data
1.41.4
Technical dataTechnical data
______________________________________________________________________
Manufacturer Hermle Labortechnik GmbH
______________________________________________________________________
Type Z 400
______________________________________________________________________
Dimensions: Width 41 cm Depth 37,5 cm Height 56,5 cm
______________________________________________________________________
Weight 33 kg
______________________________________________________________________
Noise level 65 dB(A)
______________________________________________________________________
Max. speed 13.500 rpm Max. Volume 1 000 ml Max. RCF 17310 x g Admiss. density 1,2 kg/dm
3
Admiss. kinetic energy 4 000 Nm
______________________________________________________________________
Electrical connection 230 V/50 Hz 1Ph 120 V/60 Hz 1Ph Current 3,8 A 4,0 A Connected load 650 Watt 480 Watt
______________________________________________________________________
Interference suppression VDE 0871, interference degree B
______________________________________________________________________
Service dept. at Hermle 0 74 26 / 96 22 55
______________________________________________________________________
Address of service: _________________________________
_________________________________ _________________________________
______________________________________________________________________
Address of agent: _________________________________
_________________________________ _________________________________
Loading...
+ 15 hidden pages