To replace rotor, first make sure the motor shaft and rotor
mounting hole are clean. Place the rotor on the motor shaft.
Reinstall the rotor securing screw on the motor shaft by turning it
clockwise. Hold the rotor with one hand and tighten the rotor
securing screw, using the rotor wrench.
3.4 Loading the rotor
Tubes to be loaded should be filled equally by eye. The differ-
ence in the weight between the tubes should not exceed 0.1
gram. Tubes should always be loaded so that there is equal
spacing between all tubes. One or two additional loaded tubes
may need to be added to achieve this. Refer to Figure 1. to see
one typical balancing scheme.
3.5 Overloading rotors
The maximum load of the rotor and the maximum speed have
been established by the manufacturer. Do not attempt to exceed
these values. The maximum speed of the rotor has been estab-
lished for liquids having a homogeneous density of 1.2g/ml or
less. In order to centrifuge liquids with a higher density it is
necessary to reduce the speed. Failure to reduce the speed
may result in damage to the rotor and centrifuge. The
revised maximum speed can be calculated with the following
formula:
Reduced speed (n
red
) = x max speed (n
max
)
Example:
Where the density of the liquid is 1.7, the new maximum speed
would be calculated as follows:
n
red
= x 13,300 = 9,882 rpm
If in doubt concerning maximum speeds, please contact the
manufacturer for assistance.
6
1.2
higher density value
1.2
1.7
9
5. Service and Maintenance
5.1 Centrifuge service
The brushless motor in the Spectrafuge 24D requires no routine
maintenance. Any required service should be performed by
authorized, qualified personnel only. Repairs performed by
unauthorized personnel may void the warranty.
5.2 Cleaning the centrifuge
Always keep the centrifuge housing, rotor chamber, rotor and
rotor accessories clean. All parts should be wiped down periodi-
cally with a soft cloth. For more thorough cleaning, use a neutral
cleaning agent (pH between 6 and 8) applied with a soft cloth.
Excessive amounts of liquid should be avoided. Liquid should not
come into contact with the motor. After cleaning, ensure that all
parts are dried thoroughly by hand or in a warm air cabinet
(maximum temperature 50ºC)
5.3 Cleaning the rotor
The rotor should be cleaned after each use. When spinning
samples containing phenol or phenol chloroform, the rotor should
be cleaned immediately after use.
5.4 Disinfection
Should a spill of infectious materials occur within the rotor or
chamber, the unit should be disinfected. This should be per-
formed by qualified personnel with proper protective equipment.
5.5 Replacing fuses
Check the fuse when it is recommended in the Troubleshooting
Guide located in this manual. The fuse holder is located in the
power inlet on the rear of the unit. Disconnect the power cord
from the power inlet. Open the fuse holder drawer by inserting a
small screwdriver under the tab and prying it open. Remove the
innermost (operative) fuse from its retaining tabs and replace the
fuse if necessary. A spare fuse is located in the outermost
chamber of the fuse drawer. Replace only with a fuse of exactly
the same value as the original. (Fuse type may be found in the
Technical data section of this manual.)