Bedienungsanleitung CoaLAB 6000 Best. Nr.: 30.000.1615 German Version
Operators Manual CoaLAB 6000 Order No.: 30.000.1616 English Version
Revision History
Version Manual
Date
(mm/dd/yy)
Software-Version - Release -
Analyser
1.0 10/29/99 1.0 Final release Version
1.2 02/20/01 1.02a Software update
1.3 07/17/01 1.02d Software update/corrections
1.4 05/07/02 1.03 Software update/corrections
1.5 08/12/02 1.03 Corrections
1.6 02/06/03 1.04 Software update/corrections
1.7 08/21/03 1.04 IVD update
1.8 11/26/04 1.04 Updates/Corrections
Copyright of Software
All software by LABiTec LAbor BioMedical Technologies GmbH (in the following LABiTec-Software) is the
intellectual property of the LABiTec LAbor BioMedical Technologies GmbH. Intellectual property rights shall
remain with LABiTec LAbor BioMedical Technologies GmbH. You are entitled to use LABiTec-Software and
the printed accompanying material at a place of work that cannot be transferred. Any violation of property
rights or copyright or trademark or using conditions may be subject to legal action. LABiTec reserves the
rights to modify the software, documentation as well as this operator manual without prior written notice.
Your Distributor:
(Please do not hesitate to contact your local distributor if you have any questions or problems.)
CoaLAB 6000
The name and serial number of your system can be found on the plate located on the back of the instrument.
Legend:
1 Type
2 Product Identification
3 Input
- Voltage (V)
- Power consumption (VA)
- Frequency (Hz)
- Fuses
1
2
3
4
4 Serial Number coded*
5 IVD Manufacturer
5
* Serial number code:
e.g. C 28 2 0501
C Productions month
28 Production date
2 Production year
0501 consecutive instrument number
8.7 Adaptation of coagulation reagents................................106
8.8 Packing of the system, preparation of transport ........... 107
CoaLAB 6000 – Operators Manual 1.8 Page 2
Introduction
1 Introduction
The instrument type CoaLAB 6000 (in the following titled as
analyzer/system) is a compact and automatic blood coagulation analyzer for
routine and STAT samples in the clinical laboratory.
For in-vitro diagnostic use only!
The system can perform 3 clotting tests such as PT / aPTT and Fibrinogen
simultaneously. It can perform additional tests such as Thrombin Time,
when the ChipCARD is used to replace any of the 3 standard tests. The
system can work in batch mode (individual test handling) or in randomaccess mode simultaneous test handling). In both cases the analyzer will
manage the results automatically.
The analyzer has 6 digital measuring channels based on the
turbodensitometric FIBRINTIMER
To avoid sedimentation a Teflon stirrer (2 stirrers are included) can be
added to the reagent. The reagent positions are always at room
temperature.
TM
measuring principle.
Purpose of this instruction manual
The instruction manual describes the system and is directed to all users of
the system.
This instruction manual gives all information necessary for the installation,
operation, maintenance, storing, packing, and transport of the system.
Service information is handled separately.
Careful observation of all information, especially of hazards and precautions
will ensure the correct and safe operation of the analyzer.
Therefore it is absolutely necessary to read the instruction manual
completely.
This instruction manual was prepared with greatest care. Should you have
any questions please refer to your distributor or to the manufacturer.
Validity of this instruction manual
All information in this instruction manual refers to this system:
For all requests please indicate type and serial number of your system
which can be found on the type plate on the back of the instrument or under
‘identification’ of this manual.
CoaLAB 6000 – Operators Manual 1.8 Page 3
Introduction
1.1 Definitions
Analyzer/System/ Instrument
Manufacturer
User
Personnel
In this manual the following terms are used:
This is the CoaLAB 6000.
The manufacturer according to IVD directive 98/79/EC is the
company/dealer who distributes the system within the respective country.
User is the current operator of the instrument.
Personnel are all people authorised to work with the system and meeting
the specifications required by the manufacturer for the operation of the
system.
Skilled personnel
All people who are especially qualified and authorised to handle specific
matters in connection with the instrument.
Life phases
All phases of the system from end of manufacture to disposal of the
instrument.
In this instruction manual all phases of the system will be mentioned.
QUALIFICATION OF PERSONNEL
The different life phases of the system require different qualifications of the personnel. The minimum
requirements for qualification are described at the beginning of each chapter or at the proper place in this
manual.
CoaLAB 6000 – Operators Manual 1.8 Page 4
Introduction
130-3-12-13
TSG
weiss
blau
nc.
gr-n
schwarz
rotrot
schwarz
gr-n
nc.
blau
weiss
1.2 Description of the system
The System
No additional equipment is necessary to operate the instrument. The
integrated module and system software controls the system and manages
all data. The system can be connected to a host computer for the
management of results. All measured results, reagent and test-specific data
can be printed on the internal thermal printer.
The system will be loaded with samples, reagents, and cuvette bars. The
samples in primary tubes (Becton Dickinson Vacutainer, Sarstedt
Monovette or specially designed sample cups) are loaded into the sample
rotor. At any time STAT samples can be loaded for priority handling. The
system can either work sample-orientated or test-selective.
Dilutor
Big loading lid
Printer
ChipCARD
F1
F2
F3
ESC
F4
Key pad
ENTER
7
8
4
0
9
5
1
6
2
3
Small loading lid
Figure 1 System Connections
1
rot
blau
schwarz
gr-n
gr-n
schwarz
rot
blau
schwarz
rot
gr-n
nc.
blau
weiss
1111
schwarz
rot
gr-n
nc.
blau
weiss
a3a2
TSG
130-3-12-13
50%
6a6b6c7
Figure 2 Modules
8
Connection of containers for
system liquids
Sensor connection
Dist. water (green)
Sensor connection
Waste water (red)
Pipettor with steel tubing and level
sensor with tubing connected to
dilutor
Barcode scanner
RS 232 interface
Power supply ON/OFF
Main plug
Cuvette waste box
2
3
4
9
8
7
ESC
0
R
E
T
N
E
F4F3F2F1
456
2
1
5
3
910
CoaLAB 6000 – Operators Manual 1.8 Page 5
Introduction
R5
13
14
15
16
17
18
1
2
3
4
5
6
7
8
9
10
11
12
R6
R1
R2
R3
R4
Reagent positions:
10 ml Adapter (white)*
4 ml Adapter (blue)*
* test specific
Sample positions:
1 - 18 for patient plasmas
Cuvette positions:
6 rials for cuvette bars
Rinsing station:
central station to rinse and
wash the pipettor steel tubing
and the level sensor
The combined sample and reagent rotor consists of:
- 6 positions for the cuvette bars (6 cuvettes per bar).
- 6 numbered positions (R1 - R6) for the reagent vials.
Two different vials can be used:
- 4 ml plastic vial with blue adapter ring
- 10 ml plastic vial with white adapter ring
- 1 reagent stirrer position on which all 6 reagent positions can be stirred
just prior to sampling.
- 18 numbered sample positions
- rinsing station
Primary tubes
The following primary tubes can be selected in the Setup Menu under
< 1 Primary Tubes >:
- Sample cup
- 4.5 ml Beckton Dickinson Vacutainer (the short Vacutainer only)
- 3.0 ml Sarstedt Monovette
For further information refer to section 6.4.1 Define Primary Tubes.
CoaLAB 6000 – Operators Manual 1.8 Page 6
Figure 3 Sample rotor with positions
Introduction
The cuvette bar
The cuvette bar has 6 wells. Up to 6 bars (36 cuvettes) can be loaded onto
the sample rotor.
To ensure correct loading, the cuvette bar is designed so that it can only be
NOTE
inserted one way (with semi-round opening inserted first), see picture.
Remove metal foil before using the cuvette bar.
Each well of the bar contains a 1 x 3 mm stir bar. These stir bars are
necessary to obtain accurate measurements.
Ensure that only the original cuvette bars and stir bars provided by the
manufacturer are being use to operate the system otherwise the warranty
may become null and void.
Figure 4 Bar with loading direction
The keyboard
The keys are used for direct control of the system and lead you through the
operating software. The display informs you of the current operation mode
of the system.
ESC
7
8
9
456
1
2
F4F3F2F1
ENTER
0
3
Figure 5 The keyboard
Keys Meaning
Arrow Keys Function key to scroll through the menu
• The right arrow-key sets the decimal point.
• The left arrow-key deletes single characters.
ESC-key Function key to quit or confirm a certain menu point.
ENTER-key Function key to confirm and save menu information.
Number-keys Function keys to enter digits and enter or select special
menu points.
Function keys
(F1-F4)
Delete-key
(blank)
Function keys to select certain operation steps which
are displayed directly above.
Function key to delete Pat. ID or Lot-No completely or to
be used as line feed key.
The main connection fuses and power switch of the system are combined
and integrated into the right side of the instrument. There is also a socket
for a barcode scanner (optional) and for a connection to a host computer
(optional) via an RS 232 C interface. The connections are marked
accordingly.
Figure 6 Power filter with mains plug
The system is delivered without power supply cable. The cable must
NOTE
meet the local regulations. (e.g. IEC 320, VDE, CSA-C22.2, No. 21 and
No. 49).
The cable has to be designed with NYLHY. The recommended length
is 1.5 m, and the minimum cross section 3x 0.75 mm2. A cold device
plug is used at the system’s end.
Use only fuses recommended by the manufacturer. Observe the
instructions on the name plate located on the back side of the system.
The analyzer is set to 230V by the manufacturer, change fuses
(supplied in accessories) in power filter, if analyzer will be operated in
115V. See chapter 7.5 on changing fuses.
CoaLAB 6000 – Operators Manual 1.8 Page 8
Introduction
L
r
1.3 Measuring principle
The system operates according to the opto-mechanical measuring principle:
It is well suited for lipemic and icteric samples as well as reagents with
kaolin.
A light beam passes through the cuvette containing the test plasma onto a
photo-detector. Any change in intensity of the transmitted light, that is light
increase or decrease, is converted into an electric signal.
The period from adding the start reagent until clot formation is measured. It
can be converted into the appropriate units then (%, Ratio, INR, mg/dl, g/l).
Once the start reagent has been added the measuring channel is
repeatedly adjusted, that means the lamp intensity automatically adjusts to
the turbidity of the test sample. In this process the turbidity of the sample
plasma and the reagent are levelled out.
A stir bar in each cuvette provides homogeneity of the reagent-plasma
medium during the measuring process. At the same time a small whirl
emerges through the stir bar movement which assures that even the
smallest fibrin clot is formed in front of the photo-detector.
This stirring action combined with the optical measurement, constitute the
basic features of the patented „turbodensitometric measuring principle“.
Figure 7 Measuring principle
Test cuvette
amp
Stirrer - Motor
ELECTRONIC
DISPLAY
Detecto
perm. magnet
1.4 Software Overview
The software overview makes it easier to find the different functions desired.
After the instrument has been switched on, it will initialize and automatically
display the MAIN MENU.
MAIN MENU
< ROUTINE MENU >
By pressing the
• Arrow-keys right/left you can select a menu.
• ENTER-key opens a menu or selects a function.
• ESC-key steps back a level in the menu.
CoaLAB 6000 – Operators Manual 1.8 Page 9
Introduction
Initialising
MAIN MENU:
Analyzer Power ON
F4
user interrupt (ESC)
Yes NoStop STAT
Esc
< 5 Start measure >
Enter
< 5 UTILITY MENU >
Enter
< 4 SETUP MENU >
Enter
< 3 STAT MANAGEM. >
Enter
< 2 TEST MENU >
Enter
< 1 ROUTINE MENU >
UTILITY MENU:
< 1 Init Instrument >
PIN
SETUP MENU:
< 1 Primary Tube >
< 2 No of STAT pos >
STAT Measurement
EditInsPrntStart
F1F2F3F4
Enter
Select test:
PT aPTT FBG END
PIN
TEST MENU:
< 1 Modify test >
< 2 Read test >
ROUTINE MENU:
< 1 Erase joblist >
< 2 Insert sample >
< 3 Edit joblist >
< 4 Print j oblist >
< 3 Date / Time >
< 4 Printer setup >
< 5 Beeper Setup >
< 6 Modify PIN >
< 7 Language >
< 8 Print Setup >
Enter
Enter
F1F2F3F4
PARAMETER
< 1 Replication >sngl / DBLE
< 2 Measurement > < Parameter >
Enter
< 3 Print test >
< 4 Erase test >
< 5 Write test >
< 5 Start measure >
< 6 Result managem. >
< 7 Load reagents >
< 8 Maintenance >
Maintenance
Enter
< 1 Sample volume >
< 2 Diluent >
< 3 Reagent 1 >
< 7 Incub. time >
< 8 Start reagent / Vol ume
/ Lot . No. / Bottle type >
< 9 max meas time >
Enter
Enter
< 3 Normal range >
< 4 1st Conversion >Reference curve /..
< 5 2nd Conversion >Ratio / INR
< 6 Print test >
wash FLUSH remov
Result Management
send disp. PRINT
CoaLAB 6000 – Operators Manual 1.8 Page 10
Hazard and Precaution
2 Hazard and Precaution
This section contains the safety regulations that must be observed by all
users of the system, to ensure safe and economic operation.
Importance of safety regulations
All safety instructions in this instruction manual must be observed to avoid
damage to persons, property and the environment.
All local safety and environmental regulations and guidelines must also be
observed.
Disregarding of safety regulations
Disregarding these safety regulations or technical specifications may result
in accidents with personnel or equipment, or environmental damages.
Hazards and Precautions
The system has been delivered by the manufacturer in a technically
flawless condition. In order to keep this condition the user must follow the following safety regulations of this manual.
The system must be operated by skilled personnel only.
NOTE
Note introduces rules to be observed.
Warning signals
ATTENTION!
WARNING!
Two different warning signals with different meanings are used in this
section.
Potential danger may cause minor injury or equipment damage, if no
precautions are taken.
Potential danger may cause severe injury or death if no precautions are
taken.
CoaLAB 6000 – Operators Manual 1.8 Page 11
Hazard and Precaution
Electrical safety:
WARNING!
Ensure that the operating voltage is set correctly before the instrument is
connected to the main power supply.
Use only grounded sockets and extension cords when connecting the
instrument to a power supply. Never disconnect grounding contacts.
Never disconnect the grounding contacts intentionally.
There is the risk of an electrical shock if
- The protective conductor is interrupted within or outside the device
- The grounded contact has been disconnected from the line.
Never remove protective guards or secured components, since you
could expose electrically live parts. Even after the instrument has been
switched off, some parts may contain voltage due to electrical charge.
All current carrying parts are sources of danger for an electrical shock.
Never use the instrument on a moist surfaces (floor, worktable or
countertop), or place containers of liquids on top of the instrument. Liquids
spilled into the instrument may cause an electrical shock.
Switch off the instrument and disconnect it from all power sources
before performing maintenance or repair work. Perform only the
maintenance/repair work described in this manual; unauthorized work on
the instrument may cancel the warranty, and require expensive service
work.
All work which requires the instrument to be opened undervoltage must only be carried outby a technician who is familiar with the risks related
thereto.
Use only replacement fuses of the stated type and with the stated
minimum current;
- Never use ‘repaired’ fuses.
- Never short-circuit the fuse holder.
Fire and Explosion hazards:
WARNING!
Do not place the instrument near explosive mixtures of flammable
gases, such as oxygen or hydrogen; electrical sparks could cause fire or
explosions.
Mechanical Safety (system in
operation)
WARNING!
CoaLAB 6000 – Operators Manual 1.8 Page 12
Never remove housing parts while the instrument is on; moving parts
such as the fan or motor drives may cause injury.
Always close the sample rotor-loading inlet during operation, to avoid
contact with moving parts. The probe will not stop when the loading inlet is open.
Ensure that the pipettor does not stop over the sample rotor or the
measuring block when the instrument is switched off.
Hazard and Precaution
- Use the protection cap for the pipettor if the system will not be operated
for a long time, to avoid damage and contamination.
- Do not open covers during a run, to avoid damaging the pipettor.
If safe operation of the instrument is not possible switch the
instrument off and secure it against use if:
- The instrument is damaged
- The instrument does not work properly
- The instrument has been stored / transported under adverse conditions
- The instrument has experienced severe temperature fluctuations
Samples, Reagents
WARNING!
Follow the instructions on the directional inserts for the correct use of
reagents.
Avoid skin contact with samples and/or test reagents, as well as those
instrument parts that touch samples and/or test reagents. All instrument
parts are potentially infectious, and reagents might irritate mucous
membranes and the skin. All liquid and waste containers which carry
sample and/or test reagents are marked with a biohazard symbol.
If sample material is spilled onto the instrument, wipe it off immediately and
decontaminate the affected surfaces.
Ensure that no foam or air bubbles are on the surface of reagents and
sample material prior to loading the system: their presence could influence
the measured results and the level detection of the needle. Ensure reagents
are used at room temperature only.
Before using sample cups and reagent vials make sure to remove all
covers/lids beforehand. Do not use organic solvents unless explicitly
authorized by the manufacturer, as their use may result in damage to
the cuvette bars, primary tubes, and waste water tubes.
Waste Liquids:
WARNING!
Dispose of waste liquids must be in compliance with applicable
regulations.
Accuracy and Precision of
Results:
WARNING!
CoaLAB 6000 – Operators Manual 1.8 Page 13
Measure control samples and ensure the instrument is operating
properly to obtain accurate results.
Inaccurate results may lead to a false diagnosis and endanger the patient.
Hazard and Precaution
User Qualification:
WARNING!
Barcode Scanner:
ATTENTION:
CAUTION
LASER LIGHT
DO NOT STARE INTO BEAM
670nm LASER DIODE
1.0 MILLIWATT MAX
CLASS II LASER PRODUCT
VORSICHT
LASERSTRAHLUNG, NICHT IN DEN LICHTSTRAHL BLICKEN,
SICHTBARE LASERSTRAHLUNG, LASERKLASSE 2
The system should be operated by trained personnel only.
Be careful when using the barcode scanner laser light.
- Avoid direct eye contact with the laser beam.
- Follow the manufacturer’s instructions.
- Ensure the scanner was tested and authorized for use by the
manufacturer.
Acoustic Signals
A beeper confirms all key strokes on the system keyboard. Thus, the user
can recognise, for example, the end of a run, possible failures during
operation, or the access to STAT mode without looking at the display.
CoaLAB 6000 – Operators Manual 1.8 Page 14
Warranty
3 Warranty
The representatives or authorised dealers warrant that the system does not
have any material or manufacturing damages after installation if the
instruction manual has been observed.
The warranty does not apply for damages which are caused
- By accidents, negligence, or non-observation of the instruction manual,
- By using inadmissible reagents or disposables, or
- By operating the instrument in an environment contradicting the
specifications of this instruction manual.
Repair and service work must be carried out by technicians who have been
authorised.
CoaLAB 6000 – Operators Manual 1.8 Page 15
Service
4 Service
The representatives or authorised dealers guarantee to carry out repair
work during the official working hours from Monday to Friday with the
exception of public holidays.
If you need service outside of that time contact our parent company or our
authorised dealer.
CoaLAB 6000 – Operators Manual 1.8 Page 16
Installation
5 Installation
This section describes how to unpack, install and prepare the instrument for
use.
Safety regulations
We assume that you have read chapter 2 Hazard and Precaution carefully.
5.1 Unpacking and installation
of the system
5.1.1 How to prepare the
installation area
Select a stable and level area for installation.
NOTE
Ensure the power cable will reach the outlet without
cords.
Ensure there is enough light for proper operation, but no direct sunlight.
Do not put the analyzer on a shelf or on a roll-desk. Protect from
direct sun light, significant temperature changes, humidity, and
Ensure there will be a cleared space of at least 12 inches (30 cm) to the
voltage variations.
sides and rear of the system, to allow easy access and air circuation.
Ensure the space above the system will be clear of equipment, supplies and
shelves.
using any extension
5.1.2 Check Contents
NOTE
The instrument weighs about 35 pounds (17 kg), and can be carried or
moved by one person. For the safety of the instrument and yourself,
however, you should move the instrument using a rolling cart of have
someone assist you.
Do not damage any packaging material, and keep them for possible
later use, including the transport protection devices. Store them
away from heat and humidity.
• Place box on the floor
• Remove all small parts from the box first, and ensure all accessories
and disposables listed in section 8.5 have been received.
CoaLAB 6000 – Operators Manual 1.8 Page 17
Installation
5.1.3 Unpacking the system
• Lift the system from the box (with assistant), and place on the installation
area.
• Remove the left (1) and right (2) foam inserts and set aside (See figure 8).
Pay attention to the pipettor!
Be careful not to damage the pipettor, flat cable, steel tubing or level sensor;
any damaged or defective parts must be replaced by a service technician
before operating.
• Place the system in the correct position on the table.
• Remove all tapes and transport protection devices (see figure 9).
1
2
1
2
Figure 8 The system with foam inserts
Transport protection devices are:
See figure 9.
1. Box insert: protects the pipettor from moving in X-direction.
2. Plastic plate with cable ties: protects the pipettor from moving in Z-
direction.
3. Red plastic screw on the steel tubing (top): protects from contamination.
4. Plastic protection cap on the steel tubing (bottom): protects from
contamination.
5. Empty cuvette rack between sample rotor and measuring block.
CoaLAB 6000 – Operators Manual 1.8 Page 18
Installation
2
t
P
c
. Transport protection of
he pipettor:
lastic with
able ties
3
2
1
4
5
System
sectional view
ROTOR
1. Transport protection of
pipettor with carton
1
2
F1
F2
F3
ESC
F4
ENTER
7
8
4
0
9
5
1
6
2
3
Figure 9 The system with transport protection devices
5.1.4 Assembly of the system
Before connecting the instrument to the main power supply, complete all
instructions in this section.
5.1.4.1 Mount the diluter
syringe and tubes
Mount the 250 µl dilutor syringe to the dilutor of the system according, see
figure 10. When properly mounted, the syringe should not be wedged, and
must be watertight.
CoaLAB 6000 – Operators Manual 1.8 Page 19
Installation
S
• Press the syringe piston completely into the dilutor syringe.
• Open the locking screw.
• Push the syringe holder at the dilutor downward.
• Place the opening of the syringe piston onto the syringe holder of the
dilutor.
• Screw the top of the syringe onto the adapter of the dilutor valve.
• Tighten the locking screw.
• Connect all tubing’s to the dilutor valve.
teel tubing
pipettor
Dist. water to
Dilutor valve
Top of syringe
Pipettor
Dist. water
Syringe piston
Syringe holder
Locking screw
Steel tubing
point and
level sensor
Figure 10 Diluter and syringe
Attach the tubes
The liquid system requires tubing between the diluter and pipettor. Connect
the tube as follows:
• Take the tubing out of its plastic accessory bag.
• Remove the red plastic screw at the steel tubing of the pipettor.
• Screw the tubing (white screw) onto the diluter valve until hand-tight.
• Run the other end of the tubing through the tubing guides on the back of
the instrument.
• Screw the end into the top of the arm (black screw). Make sure it fits
properly and watertight.
• Review the entire tube path to ensure the tubes will be neither bent nor
squeezed during instrument operation.
CoaLAB 6000 – Operators Manual 1.8 Page 20
Installation
5.1.4.2 Connect the Liquid
System Containers
• Place the distilled water container (with the green label) and the empty
waste water container (with the red label) to the left rear of the system. Fill
the distilled water container with at least 1 Liter of distilled water.
• Place the colour-coded sensors into the appropriate containers (green for
dist. water, red for waste water). Ensure there are no air bubbles in the
tubing of the distilled water container, or in the container itself.
• Plug the green and red colour-coded couplings into their matching sockets
on the back of the system (see figure 11). A „click“ indicates the proper
seating of the connectors.
The system will only function if the distilled and waste water tubes have
been connected properly to the system.
NOTE
The system is switched OFF.
The couplings are color-coded:
Red for
waste water
Green for
dist water
Green
Red
Figure 11 Connect the level sensor tubes to the system
< NOTE >
Decoupling of the tube adapter
Press on the metal shield to
release the tube adapter.
The distilled water container must have at least 1 Liter of distilled
NOTE
water. The waste water container should be emptied every day. Be
careful not to have air bubbles in the distilled water container as well
as in the tube.
CoaLAB 6000 – Operators Manual 1.8 Page 21
Installation
Container
Figure 12 Container for distilled water. Container and tubes are
Figure 13 Container for waste water. Container and tubes are coded
dist
water
coded with green.
waste
water
with red.
5.1.4.3 Connect the Barcode
Scanner (Optional)
5.1.4.4 Connect a HOST or
Personal Computer to
the System
NOTE
• Connect the optional barcode scanner to the system. Use the
recommended type only.
• Connect the barcode scanner data cable to the upper socket (marked
barcode) on the system (see figure 6).
• Plug the power supply into the socket next to it.
The system has an additional interface port (RS 232 C).The RS 232
interface is a uni-directional interface which allows the transfer of patient ID
and results to a connected HOST or PC. Due to many different used
HOST/LIS PC systems the interface protocol may vary, please ask your
distributor or the manufacturer for further information on transfer of data’s.
• Before switching on the computer, the PC must have been connected to
the external periphery instruments. Use the data cable recommended by
manufacturer or dealer only. Then connect the PC to the power supply.
Do not switch the computer on yet!
Now connect the data cable to the lower socket (marked RS 232) on the
right side of the system and to the interface port of your computer.
Before switching on the analyzer and before each re-initialisation,
pay attention that all cuvettes and vials have been removed from the
rotor and the measuring block.
• Now connect the system to the power supply. The ON/OFF switch is on
the right side of the instrument.
CoaLAB 6000 – Operators Manual 1.8 Page 22
Installation
• After complete installation the instrument and, possibly, the computer can
be switched on and off accordingly.
See figure 6 Power ON/OFF, Power filter.
5.1.4.5 Insert and change
printer paper
After switching on the system you can now insert the printer paper (see
accessories). The printer paper can be inserted only when the system is
switched on
.
• Lift the printer cover above the keyboard and remove the previous paper
roll holder, if necessary.
• Place the new paper roll with the paper roll holder into the correct position.
The unrolling paper must run towards the inside of the printer
Paper roll axis
paper insert slot
Figure 14 Printer housing
• Guide the paper - with a clear-cut end- into the paper slot of the printer.
Once the paper has been captured it will be advanced automatically.
• Guide the paper through the slot of the printer cover and re-install the
cover.
Use manufacturer’s printer paper only. Otherwise, the warranty
NOTE
may become void. Over time the text will fade. It is recommended
that all printouts be copied for storage purposes.
CoaLAB 6000 – Operators Manual 1.8 Page 23
Installation
5.1.4.6 Mount the cuvette waste
box
The cuvette waste box must be attached before using the instrument. The
cuvette bars which are automatically ejected from the measuring unit will be
collected.
Be careful that plasma as well as reagents can flow out from the cuvette
bars.
Therefore it is necessary to clean the box in regular intervals.
Mounting the waste box
See figure 15 next page
• Take the waste box out of the accessory box.
• As shown in the figure, press the box gently.
• Guide the waste box, gently pressed and slightly tipped, into the housing
space so that the locking devices on the inside prevent the box from
slipping.
Use caution, as the cuvette bars that are collected in the waste box,
NOTE
may leak plasma and reagents.
If mounted properly the nose of the waste box will be placed under the
measuring block and will collect used cuvette bars. It is recommended to
clean the waste box weekly using a 10% bleach solution.
Demounting the waste box
• Remove the box by pressing it gently and slightly tipping it.
• Dispose of the cuvette bars in compliance with biohazard regulations. •
Mount waste box
1. Press box gently
see arrows
2. Tip box slightly while
inserting it
Thermo block 37˚C
Movement direction
of the cuvette rack --->>>
waste box fixpoints
3. Bring box into proper
position as shown in
the sectional
view.
Figure 15 mounting the cuvette waste box
CoaLAB 6000 – Operators Manual 1.8 Page 24
Installation
5.2 Instrument control Card
(ICC)
The manufacturer recommends to fill in the Instrument Control Card (see
attachment) of each system after its initial installation and to return it to the
manufacturer within four weeks. With this Instrument Control Card the
manufacturer will register your system. The customer confirms proper
receipt, functioning and completeness of the delivered system.
The Instrument Control Card is the basis for any warranty claims.
NOTE
Possible claims can only be called for if the Instrument Control Card
had been returned to the manufacturer after the initial installation of
the system.
The document is enclosed.
5.3 Language Setup
The analyzer is pre-set to the English language.
If you want to change the language follow section 6.4.7 Language.
CoaLAB 6000 – Operators Manual 1.8 Page 25
Operation
6 Operation
Training
Safety precautions
6.1 Routine Operation
<1 ROUTINE MENU>
This section describes how to routinely operate the system as well as
information on quality control of coagulation measurements. Furthermore
you will find hints to possible failures and troubleshooting.
Onsite training courses are offered for this instrument. For additional
information contact your distributor.
Thoroughly review the Hazard and Precautions section of this manual
before proceeding.
The system will be ready for coagulation analysis after initial set-up has
been completed, see section 6.4 General Setting (SETUP MENU). The
following descriptions explain the ROUTINE MENU and give an insight into
safe coagulation analysis.
In the ROUTINE MENU you can select between three different methods.
You can use nearly all well-known PT / aPTT and Fibrinogen reagents on
the system. Please note that due to the large variety of coagulation
reagents on today’s markets only a certain number of reagents have been
checked and adjusted to the system. See section 8.7 Adaptation of
coagulation reagents for further information.
Preparing routine operation
white
blue
- The instrument is connected to the power supply.
- The distilled water container is filled with water (without bubbles) and
connected to the system.
- The waste water container is empty and connected to the system.
- The waste box is cleared of cuvette bars.
- All tubes have been properly connected, and are neither squeezed nor
bent.
- Sufficient printer paper has been inserted.
- The liquid system is filled and without bubbles.
Reagents for the tests will be reconstituted according to the package insert
and will be filled into the reagent vials. The delivered reagent vials have the
following dead volumes:
10 ml reagent vial
(white adapter)
4 ml reagent vial
(blue adapter)
without stir bar
with stir bar
without stir bar
with stir bar
dead volume: 1.5 ml
dead volume: 2.0 ml
dead volume: 0.5 ml
dead volume: 0.8 ml
CoaLAB 6000 – Operators Manual 1.8 Page 26
Operation
6.1.1 Switch ON the system
The instrument must be free of cuvette bars, sample and reagent
NOTE
vials.
• Switch the instrument on by setting the power switch to the position I =
ON.
The display shows the instrument name and checks all internal modules
automatically (the pipettor moves to the washing station).
Diverse displays ...
User dialogue display
of the system
Then all system modules will be initialised.
The following data will be printed:
C2820501 V1.04 04/22/02
07.05.2002 08:55
Ser. No. / SW-Version / Release
Date / Time
The instrument is ready for operation.
• Use the arrow keys (</>) to select <1 ROUTINE MENU> in the MAIN
MENU.
MAIN MENU:
< 1 ROUTINE MENU >
• Press the ENTER-key to confirm.
The following display appears:
MAIN MENU:
< 1 Erase joblist >
6.1.2 Rinsing with distilled
water
For transport protection reasons the system has been delivered in dry
condition, without any liquids. Before measuring with the system it is
necessary to perform a rinsing process and fill all system modules with
liquid (pumps, dilutor and tubing’s). This procedure also removes any air
bubbles from the tubing system and the dilutor tubing. You can now check
whether all tube and screw connections have been installed properly.
Should you notice any leaking, tighten the connections to the point that they
are watertight.
• Use the arrow keys (</>) to select < 8 Maintenance > in the ROUTINE
MENU.
ROUTINE MENU:
< 8 Maintenance >
• Press ENTER to confirm and rinse the pipetting system of the system.
CoaLAB 6000 – Operators Manual 1.8 Page 27
Operation
The following display appears:
Maintenance
wash FLUSH remov
F1 F2 F3 F4
= accompanying function keys
• Press function key F2.
The instrument will rinse the system with distilled water. After the initial
installation this procedure should be repeated until all tubes are filled with
liquid completely.
By Pressing F4 = remov cuvette bars will be ejected into the waste box.
During rinsing the following display appears:
Maintenance
Maintenance
wash FLUSH remov
• Leave the menu after rinsing by pressing ESC.
The following display appears:
ROUTINE MENU:
< 8 Maintenance >
Pressing ESC will take you back to the MAIN MENU.
After this procedure the instrument pipetting system has been rinsed with
distilled water. The manufacturer recommends repeating this procedure
after the initial installation several times to secure that no air bubbles are in
the pipetting tube and that no liquids containing thrombin, for example, will
remain in the liquid system (if system has already been used). Check the
system for leakage’s again and re-mount, if necessary, all connections and
tubes before repeating the rinsing cycle.
6.1.3 Disinfection of waste tubing
(Terralin), change of FW-Filter
It is necessary to disinfect and clean the liquid system, tubing’s at regular
intervals, see section 8.1 Maintenance and Hygiene to avoid reagent or
sample residues, possibly causing inaccurate results or improper functioning
of the analyzer. Therefore, regular disinfection and cleaning cycles must be
run to guarantee proper functioning of the system.
Follow the instructions for disinfection of the system tubing’s:
Clean the distilled water container
and change filter
1. Remove the Distilled Water Sensor (green mark) from the Distilled
Water Container.
2. Remove the white fresh water filter
from the end of the Distilled Water
Sensor and replace it with a new fresh water filter.
CoaLAB 6000 – Operators Manual 1.8 Page 28
Operation
3. Empty the Distilled Water Container and clean the inside using a 70%
alcohol solution.
4. Rinse the Distilled Water Container thoroughly with distilled water to
remove any traces of the alcohol solution.
5. Refill the Distilled Water Container and return the Distilled Water
Sensor to thecontainer.
Disinfection of Wash Station and Waste
tubing with Cleaning Solution
6. Make a working solution of Cleaning Solution. Fill 2 Liters of distilled
water into the Cleaning Solution Container (white label) and add 20
ml of concentrated Cleaning Solution. Swirl gently.
7. Place the Waste Water Container carrying the Waste Level Sensor
(red label, sensor) beside or behind the analyzer on the same height
level. The container may not be
placed underneath the analyzer. Make
sure the waste tubing (red marked) and the electrical plug of the waste
tubing are correctly connected to the fluid panel of the back side of the
analyzer.
NOTE
Do not disconnect any of the level sensor tubing’s (red/green) and
electrical plugs at the fluid panel at the back side of the analyzer.
8. Take out 4ml of the diluted working Cleaning Solution from the
Cleaning Container using a standard commercial pipette, e.g.
Eppendorf.
9. Guide the pipette over the Wash Station, located in the center of the
reagent and sample rotor.
Make sure that neither the steel needle nor the level sensor of the
NOTE
probe will be damaged or bent during this operation. The system may
not be in operation, moving.
10. Dispense the diluted Cleaning Solution out of the pipette into the wash
station. Repeat this process until you have added 4ml into the Wash
Station. Leave the working Cleaning Solution in the Wash Station for
at least 2 min.
Make sure the Cleaning Solution inside the Wash Station remains on
NOTE
the same level. It may not disappear inside the Wash Station. In this
event place the Waste Water Container on a higher level, meaning
above the height of the wash station of the analyzer and repeat adding
the Cleaning Solution into the Wash Station. Further make certain not
to overflow the wash station to avoid any liquid flowing into the
sample/reagent rotor.
Flush after cleaning
After the disinfection/cleaning cycle with Cleaning Solution it is
absolutely necessary to rinse the liquid system with distilled water to
ensure that no Cleaning Solution may remain in the liquid system at
all.
CoaLAB 6000 – Operators Manual 1.8 Page 29
Operation
NOTE
NOTE
Follow the instructions to rinse the system with distilled water:
11. Make sure the Distilled Water Sensor is equipped with a NEW fresh
water filter and the Distilled Water Sensor is placed correctly in the
Distilled Water Container.
12. Run flush cycles with distilled water described as follows:
• Use the arrow keys (</>) to select <5 UTILITY MENU> in the MAIN
MENU.
MAIN MENU
< 5 UTILITY MENU >
• Press ENTER to confirm.
The following display appears:
UTILITY MENU
<1 Init Instrument>
• Press ENTER to confirm the message.
The following display appears:
Init Instrument
Init ... (ESC=Stop)
The motors and the diluter will be initialised. During this phase the cuvette
racks in the measuring block will be moved one position to the right and will
be ejected from the measuring block.
The following display appears:
Init Instrument
Flush... (ESC=Stop)
Then the liquid system will be initialised. Liquids are pumped into the
tubing’s. Press ESC if you want to finish initialisation.
This flush process should run a minimum 5-8 minutes or until no foam
residues of Cleaning Solution will remain in the liquid system, waste
tubing (red marked). Check if waste water runs through the waste
tubing into the Waste Water Container and no foam remains in the
waste tubing.
The manufacturer recommends repeating this procedure at least twice
a month. Once a year it is recommended to renew all liquid containing
parts of the system. Ask your local dealer
available maintenance/service kits.
or the manufacturer for
CoaLAB 6000 – Operators Manual 1.8 Page 30
Operation
6.1.4 Rinsing with Washing
Solution
In the ROUTINE MENU you can choose to clean the pipetting system of the
instrument with an extra Washing Solution. This is especially recommended
after the use of reagents containing thrombin, such as Fibrinogen. The
system disposes of an automatically wash cycle if Thrombin containing
reagents, e.g. Fibrinogen are being used. Once the reagents are
introduced, see section 6.1.5 and are requested to run in a joblist, section
6.1.7 the system automatically washes the system with washing solution
after the use of such thrombin containing reagents.
NOTE
This intensive washing cycle must be run daily after routine work in
order to avoid any residues of previous measurements.
Follow the instructions to wash the system:
• Use the arrow keys (</>) to select < 8 Maintenance > in the ROUTINE
MENU.
ROUTINE MENU
< 8 Maintenance >
• Press ENTER to confirm.
The following display appears:
Maintenance
WASH flush remov
F1 F2 F3 F4
• Press F1= WASH and follow the instructions of the display.
The following display appears:
1: washer:
LOAD 3.0ml skip
F1 F2 F3 F4
If no washing solution has been loaded the following display appears. Put
the white adapter ring into reagent position 1: = R1 extended automatically
and put the 10 ml reagent vial filled with min. 3 ml Washing Solution into the
reagent position. Confirm either with F1= Load or with ENTER.
Maintenance
To cancel this procedure press F4= skip. After having confirmed with F1 or
ENTER the analyzer will start an automatic wash cycle to clean the tubing’s
(consumption per wash cycle ca. 110 µl).
The next display appears:
Maintenance
WASH flush remov
• Press ESC to leave the menu.
CoaLAB 6000 – Operators Manual 1.8 Page 31
Operation
6.1.5 Loading and unloading
reagents
NOTE
The following display appears:
ROUTINE MENU
< 8 Maintenance >
The system has now been cleaned with Washing Solution. If necessary, this
procedure can be repeated. Press ESC to get back to the upper menu level.
Before entering a joblist it is possible to load and later unload reagents for
daily routine. Observe the following preconditions:
- All reagents must be reconstituted according to the package insert.
- The system is preset to the supplied reagent vial types. To avoid
inaccurate level detection you should use the reagent vials with the
adapter rings only. Make sure that only predefined reagent vial types (4 ml
or 10 ml) are positioned in the sample or reagent rotor. For definition of
reagent vial types see section 6.2.2 Enter Method Parameter for definition
of reagent vial types.
10 ml reagent vial
(white adapter)
4 ml reagent vial
(blue adapter)
without stir bar
with stir bar
without stir bar
with stir bar
dead volume: 1.5 ml
dead volume: 2.0 ml
dead volume: 0.5 ml
dead volume: 0.8 ml
The dead volumes mentioned above have been considered when the
volume is asked for.
Loading PT, aPTT and Fibrinogen reagent is described in the following
example.
Remember to use the provided adapter rings (4 ml vial = blue ring;
10ml vial = white ring). Always position the reagent vials with the
adapter rings into the correct reagent position in the sample rotor and
make sure to use predefined reagent vial types only to avoid
inaccurate level detection.
The system has an automatic reagent stir position in order to avoid
sedimentation’s. If reagents are stirred, Teflon stir bars must be put
into the reagent vials. To select reagent vial types, see section 6.2.2
Enter Method Parameter.
• Use the arrow keys (</>) to select <7 Load reagents> in ROUTINE
MENU.
ROUTINE MENU
< 7 Load reagents >
• Press ENTER to confirm.
The following display appears:
1: washer:
LOAD unload
F1 F4
CoaLAB 6000 – Operators Manual 1.8 Page 32
Operation
The reagent position for the Washing Solution will be extended
automatically. Should you have loaded Washing Solution already, the above
display appears.
• Press F1 to confirm loading.
By pressing F4 ‘remove’ you inform the system that you have
removed reagent and/or washing solution from the sample rotor.
NOTE
‘Remove’ will always be displayed if you have loaded the sample rotor
with reagents and/or washing solution beforehand. Once the sample
rotor has been loaded with reagents and/or washing solution ‘remove’
will be displayed to ‘unload’ the sample rotor.
If ‘skip’ is displayed instead of ‘remove’ you can press F4 to skip over
loading with reagent and/or Washing Solution. Remember however,
that the analyzer cannot run measurements or start cleaning cycles if
reagents and/or Washing Solution have not been loaded. This
message will always be displayed when no reagents and/or Washing
Solution have been loaded to the sample rotor.
The following display appears:
2: Buffer: 9999992
LOAD skip
The sample rotor position will be extended automatically. Load the sample
rotor with Buffer in position 2: = R2 and confirm with F1.
If you want to skip over this position and not load Buffer press F4.
The following display appears:
3: FBG 8888
LOAD skip
The sample rotor position will be extended automatically. Load the sample
rotor with Fibrinogen reagent in position 3: = R3 and confirm with F1.
If you want to skip over this position and not load Fibrinogen press F4.
The following display appears:
4: CaCl2 88888
LOAD skip
The sample rotor position will be extended automatically. Load the sample
rotor with Calcium Chloride in position 4: = R4 and confirm with F1.
If you want to skip over this position and not load Calcium Chloride press
F4.
The following display appears:
5: aPTT 99999995
LOAD skip
CoaLAB 6000 – Operators Manual 1.8 Page 33
Operation
The sample rotor position will be extended automatically. Load the sample
rotor with aPTT reagent in position 5: = R5 and confirm with F1.
If you want to skip over this position and not load aPTT reagent press F4.
The following display appears:
6: PT 88888
LOAD skip
The sample rotor position will be extended automatically. Load the sample
rotor with PT reagent in position 6: = R6 and confirm with F1.
If you want to skip over this position and not load PT reagent press F4.
After all reagents have been positioned on the sample rotor the following
display appears.
ROUTINE MENU
< 7 Load reagents >
• Press ESC to get back into the MAIN MENU.
6.1.6 Delete joblist, delete
samples
If you want to delete a joblist from the memory follow the instructions:
• Use the arrow keys (</>) to select <1 Erase joblist > in the ROUTINE
MENU.
The following display appears:
ROUTINE MENU
< 1 Erase joblist >
Press ENTER to confirm.
The following display appears:
Really erase joblist?
YES no
If you want to delete the joblist confirm with F1 = YES. If you do not want to
delete it, press F2 = NO. Confirm with F1.
The following display appears:
Remove samples 1-9
and close cover
• Remove primary tubes 1 to 9 from the sample positions extended
automatically and confirm with ENTER.
If more than 9 primary tubes are in the sample rotor the following message
appears:
Remove samples 10and close cover
CoaLAB 6000 – Operators Manual 1.8 Page 34
Operation
• Remove the remaining primary tubes 10 - 18 and confirm with ENTER.
The following display appears:
ROUTINE MENU
< 1 Erase joblist >
• Press ESC to get back into the MAIN MENU.
6.1.7 Enter a new joblist
Before entering a joblist you must define whether single or double
determinations shall be made, see section 6.2.2.1
The following examples show how a joblist is to be entered:
Example 1: 01 = Sample 1 S-ID: 30001 = consecutive no. PT and APTT =
test to be measured
Example 2: 02 = Sample 2 S-ID: 30002 = consecutive no PT, APTT, FBG =
test to be measured
Example 3:
03 = Sample 3
S-ID: 123 = individual no PT =
test to be measured
Example 1:
• Use the arrow keys (</>) to select < 2 Insert sample > in the ROUTINE
MENU.
ROUTINE MENU
< 2 Insert sample >
• Press ENTER to confirm.
The following display appears:
01 S-ID: 30001
pt aptt fbg all
F1 F2 F3 F4
The sample rotor will automatically advance to the position to be loaded
with sample 1 (01 S-ID:). Load the sample rotor with the first primary tube
and select the tests with keys F1-F4. In this case the tests pt=F1 and
aptt=F2 have been selected.
The following display appears:
01 S-ID: 30001
pt APTT fbg all
After the tests have been selected they will appear in capital letters.
• Press ENTER to confirm.
CoaLAB 6000 – Operators Manual 1.8 Page 35
Operation
Example 2
Example 3
NOTE
The following display appears:
02 S-ID: 30002
pt aptt fbg all
The sample rotor will automatically advance to the position to be loaded
with sample 2 (02 S-ID:). Load the sample rotor with the second primary
tube and select the tests with keys F1-F4. In this case the tests pt, aptt and
fbg have been selected, i.e. all = F4.
• Press ENTER to confirm.
The following display appears:
03 S-ID: 123
pt aptt fbg all
The sample rotor will automatically advance to the position to be loaded
with sample 3 (03 S-ID:). Load the sample rotor with the third primary tube
and select the tests with keys F1-F4. In this case the test pt=F1 has been
selected. If you want to change the patient ID press the left arrow key (<)
until the patient ID is no longer displayed. You can now enter the individual
ID, here 123.
• Press ENTER to confirm.
The following display appears:
04 S-ID: 30003
pt aptt fbg all
The sample rotor will extend automatically the position to be loaded with
sample 4 (04 S-ID:). Repeat this procedure to insert more patient ID´s.
Press ESC to leave this Menu.
The analyzer offers two possibilities to enter patient ID’s. If no ID is entered
the system will automatically assign consecutive numbers.
1.) Enter the patient ID manually.
Using the number keys (0-9), enter the
patient ID up to 10 digits.
2) Enter the patient ID by using the barcode scanner.
Press the button on
the barcode scanner and direct the scanner window to the barcode
label of the primary tube. Continue pressing the button until a beep
confirms entry of ID.
A large number of barcodes can be used. If the system does not
accept a certain barcode contact the manufacturer or enter the patient
ID manually.
Press ESC to leave the Menu.
CoaLAB 6000 – Operators Manual 1.8 Page 36
Operation
The following display appears:
ROUTINE MENU
< 2 Insert sample >
The sample rotor has a maximum of 18 sample positions of which 15 can
be used for routine work. Up to 3 positions can be used for STAT
measurements. To load the sample positions repeat the procedures as
described in this chapter.
Press ESC to get back to the MAIN MENU.
6.1.8 Handle the joblist
After entering the joblist into the memory, entries may be reviewed for
accuracy or modification.
• Use the arrow keys (</>) to select <3 Edit joblist> in the ROUTINE MENU.
ROUTINE MENU
< 3 Edit joblist >
• Press ENTER to confirm.
The following display appears:
01 S-ID: 30001
PT APTT fbg all
The desired test of each sample is displayed in capital letters. All other tests
will be displayed in lowercase. By pressing the function keys F1-F4, tests
may be selected or deselected or further tests can be added.
With the up and down arrow keys you can select the next sample.
• Press ESC to leave this menu point.
The following display appears:
ROUTINE MENU
< 3 Edit joblist >
• Press ESC to get back to the MAIN MENU.
6.1.9 Print of a joblist
The description how to print a joblist follows in this section. The printed
joblist will provide an overview of the patient samples and methods.
• Use the arrow keys (</>) to select <4 Print joblist> in the ROUTINE
MENU.
ROUTINE MENU
< 4 Print joblist >
• Press ENTER to confirm.
CoaLAB 6000 – Operators Manual 1.8 Page 37
Operation
NOTE
The following display appears:
printing joblist
The joblist will be printed on the internal printer. After printing this function
will re-appear.
No Sample-ID T1 T2 T3
1: 30001 x x
2: 30002 x x x
3: 123 x
------------------------------------------
Have the joblist printed after entry. Thus you can control whether it
contains all selected entries. See chapter 6.1.8 Handle the joblist if
you want to modify anything.
- Test name (short form)
- selected tests
6.1.10 Start measuring, load
cuvette bars
After printing the joblist start measurement. Follow the instructions in this
section to run measurements.
• Use the arrow keys (</>) to select <5 Start measure> in the ROUTINE
MENU.
ROUTINE MENU
< 5 Start measure >
• Press ENTER to confirm.
The following display appears:
Select test:
PT aPTT FBG all
CoaLAB 6000 – Operators Manual 1.8 Page 38
Operation
With the function keys F1-F3 below the display a single test will be run.
With the function key F4 = all, all three tests will be run automatically
(RANDOM ACCESS).
Example: PT = F1
The following display appears:
Prepare measure
The following display appears:
INFO
check waste box
The message ‘check waste box’ is displayed for a short moment every time
you enter a new joblist. Empty the waste box if used cuvette bars are still in.
The following message will be displayed:
Insert cuvette-bar
and close cover
The system will calculate according to the joblist how many cuvette bars are
needed. The sample rotor automatically advances to the next position for a
cuvette bar. Place the cuvette bar properly into its position.
This procedure will be repeated until all bars for the measurements have
been positioned.
• Press ENTER to confirm.
If no reagents are loaded the sample rotor will automatically advance to the
reagent position. Place the adapter ring and the predefined reagent vial of
the selected test (here PT) into it. See section, 6.2.2 Enter Method
Parameter.
The display will show the reagent volume necessary for the tests.
6: PT 123456789
LOAD 2.4ml skip
The displayed figure already considers the necessary dead volume of the
reagent vial.
Dead volumes:
10 ml reagent vial
(white adapter)
4 ml reagent vial
(blue adapter)
without stir bar
with stir bar
without stir bar
with stir bar
dead volume: 1.5 ml
dead volume: 2.0 ml
dead volume: 0.5 ml
dead volume: 0.8 ml
Having positioned the reagent vial with adapter into the sample rotor you
can confirm this in two ways:
F1 (load) or =
ENTER
The reagent has been loaded and the entry is
confirmed.
CoaLAB 6000 – Operators Manual 1.8 Page 39
Operation
F4 (skip) =
The reagent has not been loaded, the entry was not
confirmed.
• In this case press F1 or ENTER to confirm.
Remember that the supplied reagent vials consider sufficient dead
volume when calculating the reagent volume. Therefore the reagent
NOTE
volumes are larger than what is actually needed for the tests. Always
use the adapter rings and the pre-defined reagent vials as the volume
calculation of the reagent will be influenced negatively otherwise.
The instrument will check the reagents for sufficient volume. If reagents are
missing or the volume is insufficient the following display will appear:
For example:
6: PT 123456789
only x.xml of x.xml
Load the missing or insufficient reagent in position R6. Remember the
necessary dead volume.
• Press ENTER to confirm end of loading.
If you do not want to load the missing reagents press ESC. The analyzer
will measure as many samples as possible with the available reagents.
Samples not measured will be identified with a note on the printout.
By pressing ESC and thus refuse to load the missing reagents, not all
measurements will be worked off your joblist. The system will do all
NOTE
measurements possible with the reagent volume. Follow the steps in
this section to work off the remaining measurements.
The following display appears:
Pipettor M-Block
37.4 38.4
The message Pipettor / M-Block indicates the temperature of the measuring
block and the pipettor. This message will be displayed until the correct
temperature has been reached.
The following display appears:
4 PT Samples <
No Results available
The system will start measurements. The number (e.g. 4 PT samples) of
measurements for each method will be displayed. The results will
automatically be printed on the internal printer.
The display will show ‘no results available’.
The rotating symbol (<) in the upper right hand corner of the display means
that the instrument is measuring presently.
CoaLAB 6000 – Operators Manual 1.8 Page 40
Operation
At the completion of all tests a short beep will be heard and the following
message appears on the display:
STATE
Normal Termination
All cuvette bars that were not completely used will remain in the measuring
block, and be used in the next testing cycle.
Example of a printed result:
C2820501 V1.04 04/22/02
07.05.2002 08:44
Analyzer Mode : Batch
Research Mode : Off
Test PT 07.05.2002 10:30
No Sample-ID Result Cuv.
1: 123 online 01a
PT time1 15.4 sec
PT INR 1.74
PT conc 43.5 %
---------------------------------------------
Finished 07.05.2002 10:45
State: Normal Termination
6.1.11 Interrupt a current run
(User interrupt)
Follow the instructions in this section to interrupt a current run. The system
is measuring as described in section 6.1.10 Start measuring, load cuvette
bars.
The following display appears:
X PT Samples
03 123 --.--
• Press ESC to interrupt the run.
There may be a slight delay, before interruption of the current run is
confirmed on the display.
The following display appears:
User interrupt (ESC)
Yes no stop stat
F1 F2 F3 F4
Select the method of interruption by using the function keys F1-F4:
CoaLAB 6000 – Operators Manual 1.8 Page 41
Operation
Function
Instruction Meaning
key
F1 Yes The run will only be interrupted after all cuvette
bars in the measuring block have been
completed, and results are available. The
cuvette bars remaining in the sample rotor will
be used later.
F2 or No The run will not be interrupted.
F3 Stop The current run is interrupted immediately. The
cuvette bars in the incubation and measuring
position will not be used. The joblist will remain
active.
F4 Stat Places the instrument in STAT management
mode, see section 6.3 STAT Management.
Cuvette bars in the measuring block will be first
completed.
• Press F4 to get into the STAT measurement and follow the instructions as
indicated in section 6.3 STAT Management.
STAT Measurement
Edit Ins Prnt Start
• Press ENTER to run the selected tests.
6.1.12 Print results, send results
to HOST
This section describes how to print results or send to a HOST computer.
The system has an internal thermal printer to print all results, test and setup
parameters, and error messages. Moreover, results can be sent to a
connected HOST computer.
Use the RS 232C interface port to connect the HOST computer to the
NOTE
analyzer. The port is on the right side of the system below the socket
for the optional barcode scanner. The port is marked RS 232. For
further instructions, see section 5.1.4.4.
• Use the arrow keys (</>) to select <6 Result managem.> in the ROUTINE
MENU.
ROUTINE MENU
< 6 Result managem. >
• Press ENTER to confirm.
The following display appears:
Result managem.
send list PRINT
F1 F2 F3
Use F1 = send to send results to a HOST. To list the results in the display
press F2 = list. Use the arrow keys (</>) to select between the patients.
Use F3 = print to print the results on the internal printer.
CoaLAB 6000 – Operators Manual 1.8 Page 42
Operation
By pressing F1 = send the following message will appear on the display for
a short moment. The results and patient ID´s will be sent to the HOST
computer.
sending results
By pressing F2 = list the results of each patient and test will be displayed
accordingly. Using the arrow keys (</>) will display different patient results
per test. Using the arrow keys up/down will display different test results per
patient.
123 PT
1 result
After pressing F3 = print the following message will appear on the display
for a short moment. The results will be printed on the internal printer.
printing results
The following display appears:
ROUTINE MENU
< 6 Result managem. >
After sending or printing the ROUTINE MENU will appear again. Press ESC
to get back to the MAIN MENU.
6.1.13 Refill liquids, clean
containers
NOTE
It is absolutely necessary to check the level in the distilled water container
(green label) constantly and at the same time clean the waste water
container regularly. If possible, the distilled water container should have a
minimum of 1 litre. Always take care that no bubbles will be in the container.
The bubbles might get into the tubing system which would influence level
detection. Use Distilled Water (aqua dest.) only!
The containers should be checked every time before the system is switched
on, before the measurement run at the latest.
The waste water container should be cleaned in a regular cycle. See
section 8.1 Maintenance and Hygiene for further instructions.
The distilled water sensor (marked green) must be placed into the distilled
water container (green label) before using the system. This sensor monitors
the liquid level in the container and gives an error message if the level is too
low. The following error message will be displayed:
STATE
Out of system liquid
The current measurement will be interrupted if not enough liquid is in
the container during a measurement as the steel tubing cannot be
washed.
CoaLAB 6000 – Operators Manual 1.8 Page 43
Operation
The following error message will be printed:
Finished 07.05.2002 15:45
Status: no system liquid
To refill and clean the distilled water container follow the instructions below:
• Ensure that the system is not in a washing cycle or in menu
< 5 Start measure >.
• Remove the Distilled Water Sensor from the Distilled Water Container.
• Empty the Distilled Water Container completely.
• Clean the inside of the container using a 70% alcohol solution and rinse it
with distilled water. Avoid any alcohol residues in the container.
• Refill the Distilled Water Container with demineralised or distilled water.
Do not use tap water.
• Place the Distilled Water Sensor into the container and continue with
routine measurements.
• Every time the Distilled Water Container is refilled the Waste Water Container must be emptied.
Always check the liquid levels to avoid unintended interruption of runs.
6.1.14 Empty and clean the
waste water container
NOTE
It is absolutely necessary to check the level in the waste water container
(red label) constantly. The container should be checked every time before
the system is switched on, before the measurement run at the latest.
The waste water sensor (marked red) must be placed into the waste water
container (red label) before using the system. This sensor monitors the
liquid level in the container and gives an error message if the level is too
high. The following error message will be displayed:
STATE
Waste bottle full
The current measurement will be interrupted if the level is too high in
order to avoid an overflow.
The following error message will be printed:
Finished 07.05.2002 15:45
Status: Waste bottle full
To empty and clean the waste water container follow the instructions below:
• Make sure that the instrument is not in a washing cycle or in menu
< 5 Start measure >.
• Remove the Waste Water Sensor from the Waste Water Container.
• Empty the Waste Water Container completely. (See section Hazard and
Precaution)
CoaLAB 6000 – Operators Manual 1.8 Page 44
Operation
p
• Clean the inside of the Waste Water Container with 70% alcohol solution
and rinse it with distilled water. Avoid any alcohol residues in the
container.
• Return the Waste Water Sensor to the container and continue with
measurements.
NOTE
The waste water contains patient samples and test reagents and
must be treated potentially infectious. Avoid any direct skin contact.
Dispose of the waste water in compliance with biohazard and legal
rovisions.
6.2 Enter Test Parameter
Follow the instructions to change pre-defined test parameters. The
descriptions will lead you through the TEST MENU and will help you to
enter test parameters correctly.
6.2.1 Enter Reagent Specific
Parameters
In this chapter learn how to enter reagent-specific parameters.
6.2.1.1 Enter a reference curve
(PT, for example)
Follow the steps mentioned below to enter a reference curve. Here PT is
used as an example since the reference curves for all methods will be
entered in the same way. Therefore use the same steps to enter reference
curves for the other methods, if necessary.
• Use the arrow keys (</>) to select <2 TEST MENU> in the MAIN
MENU.
• Press ENTER to confirm.
NOTE
The following display appears:
• Enter the PIN code with the number keys (default=11111).
The following display appears:
MAIN MENU:
< 2 TEST MENU >
Before getting into the TEST MENU it is necessary to enter a number
code (PIN). The system number code is preset to 11111 by the
manufacturer. Change this code as described in section 6.4.6 Set PIN
Code. Remember to note the code down and keep it at a safe place. It
should be known to those users only who are authorised to operate
the instrument.
TEST MENU:
PIN
TEST MENU:
< 1 Modify test >
CoaLAB 6000 – Operators Manual 1.8 Page 45
Operation
NOTE
• To change or check test parameters press ENTER to confirm.
You can now select from the following display:
Select test:
PT aPTT FBG END
• Select method PT by pressing F1. To enter the reference curve for
Fibrinogen press F3 accordingly. Press F4 = END to leave the menu
without making any changes.
• Use the arrow keys (</>) until <4 1st Conversion> is displayed.
PT PARAMETER:
< 4 1st conversion >
• Press ENTER to confirm.
With the arrow keys (</>) choose between the following reference curve
types:
• Use the arrow keys (</>) until < 3 curve interpol > is displayed.
• Press ENTER to confirm.
The following display appears:
Time Val[%]
1 11.5 100.0
The displayed values can be overwritten directly. First enter the time. Press
the right arrow key (>) to set the decimal point. Press ENTER to confirm.
Enter the % - value with the number keys. Press ENTER to confirm. Use the
left arrow key (<) to delete or overwrite.
The system checks the reference curve points automatically whether they
are plausible, i.e. falling concentration/activity must have rising time values
and reverse. If the points are not plausible the system will beep.
The following display appears:
Time Val[%]
2 14.5 50.0
Enter the 2.reference curve point. Follow the instructions above. A minimum
of 2 points must be defined; a maximum of 9 points can be entered.
• Press ENTER to confirm the entry.
To enter any additional point after having entered the second or any
further point enter 0.0 s for the last time value. The remaining points
will not be displayed. To enter more points next time enter the actual
time value instead of 0.0 s and the remaining points will be displayed.
CoaLAB 6000 – Operators Manual 1.8 Page 46
Operation
The following display appears:
PT PARAMETER:
< 4 1st conversion >
• Press ESC to leave this menu point.
Select <1 none> in <1st Conversion> menu to enter no reference curve.
• Press ENTER to confirm <1 none> and to deactivate the reference curve.
The following display appears:
TEST MENU:
< 1 Modify test >
• Press ESC again to get back into the MAIN MENU.
6.2.1.2 Enter ISI value
Follow the instructions to enter the ISI value.
• Use the arrow keys (</>) to select <2 TEST MENU> in the MAIN
MENU.
MAIN MENU:
< 2 TEST MENU >
• Press ENTER to confirm.
The following display appears:
TEST MENU:
PIN
• Enter the PIN code with the number keys (default= 11111).
The following display appears:
TEST MENU:
< 1 Modify test >
• Press ENTER to confirm <1 Modify test> in the TEST MENU.
The following display appears:
Select test:
PT aPTT FBG END
• Press F1 to select method PT.
Press F4 = END to leave the menu without making any changes.
The following display appears:
• Use the arrow keys (</>) until <5 2nd Conversion> is displayed.
CoaLAB 6000 – Operators Manual 1.8 Page 47
Operation
PT PARAMETER:
< 5 2nd conversion >
• Press ENTER to confirm.
• Use the arrow keys (</>) until < 3 INR > is displayed.
The following display appears:
PT 2nd conversion
< 3 INR >
• Then press ENTER to confirm.
The following display appears:
INR:
100% = 11.5 s
• Enter the 100% reference value (in sec.). The 100% Value can be taken
from the tested calibration curve or from the used INR. INR is defined
through standard plasma with a given mean value of 6 determinations
definitions classification and the average value to be given in seconds.
The displayed value can be overwritten with the number keys. A maximum
of 3 digits before the decimal point can be entered. Press the right arrow
key (>) to set the decimal point.
Use the left arrow key (<) to delete or overwrite any entry.
• Press ENTER to confirm the entry.
The following display appears:
INR:
ISI: 1.100
Enter the ISI-value according to the package insert of the reagent
manufacturer. The displayed value can be overwritten with the number
keys. A maximum of 3 digits after the decimal point can be entered. Press
the right arrow key (>) to set the decimal point. Use the left arrow key (<) to
delete or overwrite any entry. After the ISI-value has been entered the
system will calculate the INR for each sample automatically.
• Press ENTER to confirm the entry.
The following display appears:
PT PARAMETER:
< 5 2nd conversion >
• Press ESC to leave this menu point.
The following display appears:
TEST MENU:
< 1 Modify test >
• Press ESC again to get back into the MAIN MENU.
CoaLAB 6000 – Operators Manual 1.8 Page 48
Operation
6.2.1.3 Enter RATIO (APTT for
example)
Follow the instructions to enter RATIO. Here aPTT is used as an example
however Ratio for all methods will be entered in the same way. Therefore
use the same steps to enter Ratio for the other methods, if necessary.
• Use the arrow keys (</>) to select <2 TEST MENU> in the MAIN
MENU.
MAIN MENU:
< 2 TEST MENU >
• Press ENTER to confirm.
The following display appears:
TEST MENU:
PIN
• Enter the PIN code with the number keys (default= 11111).
The following display appears:
TEST MENU:
< 1 Modify test >
• Press ENTER to confirm <1 Modify test> in the TEST MENU.
The following display appears:
Select test:
PT aPTT FBG END
• Press F2 to select method aPTT.
If you want to enter Ratio for one of the other methods press either F1 for PT. Press F4 = END to leave the menu without making any changes.
• Use the arrow keys (</>) until <5 2nd Conversion> is displayed.
The following display appears:
aPTT PARAMETER:
< 5 2nd conversion >
• Press ENTER to confirm.
• Use the arrow keys (</>) until <2 RATIO> is displayed.
The following display appears:
aPTT 2nd conversion
< 2 RATIO >
• Press ENTER to confirm.
The following display appears:
CoaLAB 6000 – Operators Manual 1.8 Page 49
Operation
Ratio:
100% = 20.0s
• Enter Ratio in seconds.
The displayed value can be overwritten directly. Press the right arrow key
(>) to set the decimal point. Use the left arrow key (<) to delete or overwrite
any entry.
• Press ENTER to confirm the entry.
The following display appears:
aPTT PARAMETER:
< 5 2nd conversion >
• Press ESC to leave this menu point.
The following display appears:
TEST MENU:
< 1 Modify test >
• Press ESC again to get back into the MAIN MENU.
6.2.1.4 Enter normal ranges
(for example PT)
Follow the instructions to enter the normal ranges. Here PT is used as an
example since the normal ranges for all methods will be entered in the
same way. Therefore use the same steps to enter the normal ranges for the
other methods, if necessary.
• Use the arrow keys (</>) to select <2 TEST MENU> in the MAIN
MENU.
MAIN MENU:
< 2 TEST MENU >
• Press ENTER to confirm.
The following display appears:
TEST MENU:
PIN
• Enter the PIN code with the number keys (default= 11111).
The following display appears:
TEST MENU:
< 1 Modify test >
• Press ENTER to confirm <1 Modify test> in the TEST MENU.
CoaLAB 6000 – Operators Manual 1.8 Page 50
Operation
The following display appears:
Select test:
PT aPTT FBG END
• Press F1 to select method PT.
To set the normal ranges for one of the other methods press F2 for aPTT or
F3 for Fibrinogen. Press F4 = END to leave the menu without making any
changes.
• Use the arrow keys (</>) until <3 Normal ranges> is displayed.
The following display appears:
PT PARAMETER:
< 3 Normal range >
• Then press ENTER to confirm.
The following display to the maximum %-value appears:
PT Normal range:
max: 130.0 %
• Use the number keys to enter the %-value.
Press the right arrow key (>) to set the decimal point. A maximum of 4 digits
before the decimal point can be entered. Use the left arrow key (<) to delete
or overwrite any entry.
• Press ENTER to confirm the entry.
The following display to enter the minimum %-value appears:
PT Normal range:
min: 70.0 %
• Use the number keys to enter the %-value
A maximum of 2 digits before the decimal point can be entered. Press the
right arrow key (>) to set the decimal point. Use the left arrow key (<) to
delete or overwrite any entry.
• Press ENTER to confirm the entry.
The following display appears:
PT PARAMETER:
< 3 Normal range >
• Press ESC to leave this menu point.
The following display appears:
TEST MENU:
< 1 Modify test >
• Press ESC again to get back into the MAIN MENU.
CoaLAB 6000 – Operators Manual 1.8 Page 51
Operation
6.2.2 Enter Method Parameter
This section describes how to enter the method parameters.
6.2.2.1 Enter single or double
determinations
Follow the instructions to enter single or double determinations. In the
following find two examples since the single and double determinations for
all methods will be entered in the same way. Therefore use the same
procedure accordingly.
Example 1
Set PT from double to single determination
• Use the arrow keys (</>) to select <2 TEST MENU> in the MAIN
MENU.
• Press ENTER to confirm.
The following display appears:
• Enter the PIN code with the number keys (default= 11111).
The following display appears:
• Press ENTER to confirm <1 Modify test> in the TEST MENU.
The following display appears:
• Press F1 to select method PT.
To set another method press F2 for aPTT or F3 for Fibrinogen. Use F4 = END to leave the menu without making any changes.
• Use the arrow keys (</>) until <1 Replication> is displayed.
The following display appears:
• Then press ENTER to confirm.
MAIN MENU:
< 2 TEST MENU >
TEST MENU:
PIN
TEST MENU:
< 1 Modify test >
Select test:
PT aPTT FBG END
PT PARAMETER:
< 1 Replication >
CoaLAB 6000 – Operators Manual 1.8 Page 52
Operation
The following display appears:
PT Replication:
sngl DBLE
F1 F2
DBLE = double
• Press F1 to select single (sngl) determination. The determination
selected will be displayed in capital letters.
The following display appears:
PT PARAMETER:
< 1 Replication >
• Press ENTER again to get to <1 Replication >.
Here, you can check whether single determination was accepted. SNGL will
be displayed in capital letters.
• Press ESC to leave this menu point.
Example 2
Set APTT with 10% variation coefficient from single to double
determination
<1 Modify test> in the TEST MENU has been selected.
TEST MENU:
< 1 Modify test >
• Press ENTER to confirm the selection.
The following display appears:
Select test:
PT aPTT FBG END
• Press F2 to select method aPTT.
To set another method, press either F1 for PT or F3 for Fibrinogen. Use F4 = END to leave the menu without making any changes
• Use the arrow keys (</>) until <1 Replication> is displayed.
The following display appears:
aPTT PARAMETER:
< 1 Replication >
• Press ENTER to confirm.
The following display appears:
aPTT Replication:
SNGL dble
SNGL = single
• Press F2 = dbl to select double determination. The determination selected
will be displayed in capital letters.
CoaLAB 6000 – Operators Manual 1.8 Page 53
Operation
The following display appears:
aPTT max. CV:
10%
Use the number keys to enter the variation coefficient. A maximum of 2
digits can be entered. With the left arrow key (<) you can delete or overwrite
your entry.
• Press ENTER to confirm.
The following display appears:
aPTT PARAMETER:
< 1 Replication >
• Press ESC to leave this menu point.
The following display appears:
TEST MENU:
< 1 Modify test >
• Press ESC to get back into the MAIN MENU.
6.2.2.2 Enter PT Test Parameter
In the following we have described how to enter reagent-specific data for
PT.
• Use the arrow keys (</>) to select <2 TEST MENU> in the MAIN
MENU.
MAIN MENU:
< 2 TEST MENU >
• Press ENTER to confirm.
The following display appears:
TEST MENU:
PIN
• Enter the PIN code with the number keys (default= 11111).
Enter the sample volume
The following display appears:
TEST MENU:
< 1 Modify test >
• Press ENTER to confirm <1 Modify test> in the TEST MENU.
CoaLAB 6000 – Operators Manual 1.8 Page 54
Operation
The following display appears:
Select test:
PT aPTT FBG END
• Press F1 to select method PT.
Use F4 = END to leave the menu without making any changes.
• Use the arrow keys (</>) until <2 Measurement> is displayed.
PT PARAMETER:
< 2 Measurement >
• Press ENTER to confirm.
• Use the arrow keys (</>) until <1 Sample volume> is displayed.
The following display appears:
PT Measurement:
< 1 Sample volume >
• Press ENTER to confirm.
The following display appears:
PT Sample volume:
50 µl
• Use the number keys to change the sample volume.
Remember that the total test volume must be between 125µl and 250µl. A
maximum of 3 digits can be entered. Use the left arrow key (<) to delete or
overwrite any entry.
• Press ENTER to confirm the entry.
Enter incubation time
• Use the arrow keys (</>) until <7 Incub. time> is displayed.
The following display appears:
PT Measurement:
< 7 Incub. Time >
• Press ENTER to confirm.
The following display appears:
PT Incubation:
70s
• Use the number keys to change the incubation time. A maximum of 3
digits can be entered. Use the left arrow key (<) to delete or overwrite any
entry.
• Press ENTER to confirm the entry.
CoaLAB 6000 – Operators Manual 1.8 Page 55
Operation
Enter start reagent volume and
the reagent lot number
• Use the arrow keys (</>) until <8 Start reagent> is displayed.
The following display appears:
PT Measurement:
< 8 Start reagent >
• Press ENTER to confirm.
The following display appears:
PT volume:
100 µl
• Use the number keys to change the volume of the start reagent.
Remember that the total test volume must be between 125µl and 250µl. A
maximum of 3 digits can be entered. Use the left arrow key (<) to delete or
overwrite any entry.
• Press ENTER to confirm the entry.
Enter the reagent lot number
Select reagent vial types
The following display appears:
PT Lot No.:
8888
• Use the number keys to enter or change the lot number of each method.
Take the lot number from the package insert. A maximum of 9 digits can
be entered. Use the left arrow key (<) to delete or overwrite any entry.
• Press ENTER to confirm the entry.
The system has an integrated reagent stir position to avoid sedimentation of
the coagulation reagents. The reagents will be stirred immediately before
pipetting and automatically within a defined time frame. In order to inform
the system of the reagents to be stirred it is necessary to preset the reagent
vial type and the stir function.
The manufacturer offers two reagent vial types. You should use those reagent vials only since the functionality of the system might be influenced
otherwise. A Teflon Stir Bar (supplied with the accessories) must be placed into the reagent vials. Also the adapter rings (supplied with the
accessories) must be used.
The following display to select a reagent vial type appears:
PT Bottle type:
< 6 4ml bottle >
• Use the arrow keys (</>) to determine which reagent vial type shall be
used and whether the reagent shall be stirred.
CoaLAB 6000 – Operators Manual 1.8 Page 56
Operation
Which varieties can be selected shows the following list.
10 ml reagent vial without stir bar white no
10 ml reagent vial with stir bar white yes
4 ml reagent vial without stir bar blue no
4 ml reagent vial with stir bar blue yes
Before positioning a reagent vial into the sample or reagent rotor
make sure that it has been predefined for the method. The reagent
vial types supplied by the manufacturer have different filling
NOTE
heights depending on their sizes (10 ml or 4ml). If you use the
wrong vial type the level might not be detected correctly or the
needle will touch the bottom of the wrong vial.
Enter maximum measuring time
If the vial type was selected the following display appears:
PT Measurement:
< 9 max meas time >
• Press ENTER to define the maximum measuring time.
The following display appears:
PT max. M-time:
80s
• Use the number keys to enter or change the maximum measuring time. A
maximum of 3 digits can be entered. Use the left arrow key (<) to delete or
overwrite any entry.
• Press ENTER to confirm the entry.
The following display appears:
PT Measurement:
< 1 Sample volume >
• Press ESC to leave this menu point.
The following display appears:
PT PARAMETER:
< 2 Measurement >
• Press ESC to leave this menu point.
The following display appears:
TEST MENU:
< 1 Modify test >
Press ESC to leave the menu without any further changes of methodspecific data for other methods.
CoaLAB 6000 – Operators Manual 1.8 Page 57
Operation
To change more press ENTER and follow the instructions as per section
6.2.2.3
6.2.2.3 Enter APTT Test Parameter
The following describes how to enter method-specific data for aPTT.
• Use the arrow keys (</>) to select <2 TEST MENU> in the MAIN
MENU.
MAIN MENU:
< 2 TEST MENU >
• Press ENTER to confirm.
The following display appears:
TEST MENU:
PIN
• Enter the PIN code with the number keys (default= 11111).
Enter the sample volume
The following display appears:
TEST MENU:
< 1 Modify test >
• Press ENTER to confirm <1 Modify test> in the TEST MENU.
The following display appears:
Select test:
PT aPTT FBG END
• Press F2 to select method aPTT.
Press F4 = END to leave the menu without making any changes.
• Use the arrow keys (</>) until <2 Measurement> is displayed.
aPTT PARAMETER:
< 2 Measurement >
• Press ENTER to confirm.
• Use the arrow keys (</>) until <1 Sample volume> is displayed.
The following display appears:
aPTT Measurement:
< 1 Sample volume >
• Press ENTER to confirm.
CoaLAB 6000 – Operators Manual 1.8 Page 58
Operation
The following display appears:
aPTT Sample volume:
50 µl
• Use the number keys to enter or change the sample volume.
Remember that the total test volume must be between 125µl and 250µl. A
maximum of 3 digits can be entered. Use the left arrow key (<) to delete or
overwrite any entry.
• Press ENTER to confirm the entry.
Enter 1. Reagent Volume
• Use the arrow keys (</>) until <3 Reagent 1> is displayed.
The following display appears:
aPTT Measurement:
< 3 Reagent 1 >
• Press ENTER to confirm.
The following display appears:
aPTT volume:
50 µl
• Use the number keys to change the reagent volume for APTT reagent.
Remember that the total test volume must be between 125µl and 250µl. A
maximum of 3 digits can be entered. Use the left arrow key (<) to delete or
overwrite any entry.
• Press ENTER to confirm the entry.
Enter a Lot Number
The following display appears:
aPTT Lot No.:
123456789
• Use the number keys to enter or change the lot number.
You will find the number on the reagent package insert. A maximum of 9
digits can be entered. Use the left arrow key (<) to delete or overwrite any
entry.
• Press ENTER to confirm the entry.
Select reagent vial types
The system has an integrated reagent stir position to avoid sedimentation of
the coagulation reagents. The reagents will be stirred immediately before
pipetting and automatically within a defined time frame. In order to inform
the system of the reagents to be stirred it is necessary to predefine the
reagent vial type and the stir function.
CoaLAB 6000 – Operators Manual 1.8 Page 59
Operation
NOTE
The manufacturer offers two reagent vial types. You should use those
reagent vials only since the functionality of the system might be influenced otherwise. A Teflon Stir Bar (supplied with the accessories) must be
placed into the reagent vials. Also the adapter rings (supplied with the
accessories) must be used.
The following display appears:
aPTT Bottle type:
< 6 4ml Bottle >
• Use the arrow keys (</>) to determine which reagent vial type shall be
used and whether the reagent shall be stirred.
Which varieties can be selected shows the following list.
10 ml reagent vial without stir bar white no
10 ml reagent vial with stir bar white yes
4 ml reagent vial without stir bar blue no
4 ml reagent vial with stir bar blue yes
Before positioning a reagent vial into the sample or reagent rotor
make sure that it has been predefined for the method. The reagent
vial types supplied by the manufacturer have different filling
heights depending on their sizes (10 ml or 4ml). If you use the
wrong vial type the level might not be detected correctly or the
needle will touch the bottom of the wrong vial.
Enter incubation time
• Use the arrow keys (</>) until <7 Incub. time> is displayed.
The following display appears:
aPTT Measurement:
< 7 Incub. time >
• Press ENTER to confirm.
The following display appears:
aPTT Incubation:
120 s
• Use the number keys to enter or change the incubation time.
A maximum of 3 digits can be entered. Use the left arrow key (<) to delete
or overwrite any entry.
• Press ENTER to confirm the entry.
CoaLAB 6000 – Operators Manual 1.8 Page 60
Operation
Enter Start Reagent Volume
• Use the arrow keys (</>) until <8 Start reagent> is displayed.
The following display appears:
aPTT Measurement:
< 8 Start reagent >
• Press ENTER to confirm.
The following display appears:
CaCl2 volume:
50 µl
• Use the number keys to enter or change the reagent volume for Calcium
Chloride.
Remember that the total test volume must be between 125µl and 250µl. A
maximum of 3 digits can be entered. Use the left arrow key (<) to delete or
overwrite any entry.
• Use the number keys to enter or change the lot number.
The number is on the reagent package insert. A maximum of 9 digits can be
entered. Use the left arrow key (<) to delete or overwrite any entry.
• Press ENTER to confirm the entry.
Select the reagent vial type for calcium chloride and decide whether the
reagent shall be stirred. To change the setting for the reagent vial type
follow the instructions above.
The following display appears:
CaCl2 Bottle type:
< 6 4ml Bottle >
• Use the arrow keys (</>) to determine which reagent vial type shall be
used and whether the reagent shall be stirred.
Which varieties can be selected shows the following list.
10 ml reagent vial without stir bar white no
10 ml reagent vial with stir bar white yes
4 ml reagent vial without stir bar blue no
4 ml reagent vial with stir bar blue yes
CoaLAB 6000 – Operators Manual 1.8 Page 61
Operation
NOTE
Enter Maximum Measuring Time
The following display appears:
aPTT Measurement:
< 9 max meas time >
• Press ENTER to confirm the maximum measuring time.
The following display appears:
aPTT max M-time:
120 s
• Use the number keys to enter or change the maximum measuring time.
A maximum of 3 digits can be entered. Use the left arrow key (<) to delete
or overwrite any entry.
• Press ENTER to confirm the entry.
The following display appears:
aPTT Measurement:
< 1 Sample volume >
• Press ESC to leave this menu point.
The following display appears:
aPTT PARAMETER:
< 2 Measurement >
• Press ESC to leave this menu point.
The following display appears:
TEST MENU:
< 1 Modify test >
Press ESC if no further change of method-specific data for other methods is
requested.
To change more parameters press ENTER and follow the instructions as
per section 6.2.2.4
Before positioning a reagent vial into the sample or reagent rotor
make sure that it has been predefined for the method. The reagent
vial types supplied by the manufacturer have different filling
heights depending on their sizes (10 ml or 4ml). If you use the
wrong vial type the level might not be detected correctly or the
needle will touch the bottom of the wrong vial.
CoaLAB 6000 – Operators Manual 1.8 Page 62
Operation
6.2.2.4 Enter Fibrinogen Test Parameter
Follow the instructions to enter test-specific parameters.
• Use the arrow keys (</>) to select <2 TEST MENU> in the MAIN
MENU.
MAIN MENU:
< 2 TEST MENU >
• Press ENTER to confirm.
The following display appears:
TEST MENU:
PIN
• Enter the PIN code with the number keys (default= 11111).
The following display appears:
TEST MENU:
< 1 Modify test >
• Press ENTER to confirm <1 Modify test> in the TEST MENU.
The following display appears:
Select test:
PT aPTT FBG END
• Select method FBG by pressing the function key F3. Press F4 = End to
leave this menu without making any changes.
• Use the arrow keys (</>) until <2 Measurement> is displayed.
The following display appears:
FBG PARAMETER:
< 2 Measurement >
• Press ENTER to confirm.
Enter the Sample Volume
• Use the arrow keys (</>) until <1 Sample volume> is displayed.
The following display appears:
FBG Measurement:
< 1 Sample volume >
• Press ENTER to confirm.
The following display appears:
FBG Sample volume:
100 µl
CoaLAB 6000 – Operators Manual 1.8 Page 63
Operation
• Use the number keys to enter or change the sample volume.
Remember that the total test volume must be between 125µl and 250µl. A
maximum of 3 digits can be entered. Use the left arrow key (<) to delete or
overwrite any entry.
• Press ENTER to confirm the entry.
Enter 1. Reagent Volume (Dilution)
• Use the arrow keys (</>) until <2 Diluent> is displayed.
The following display appears:
FBG Measurement:
< 2 Diluent >
• Press ENTER to confirm.
The following display appears:
FBG Diluent 1 + n:
9.0
• Use the number keys to change the reagent volume (1+n) for the
Dilution (Buffer).
A maximum of 3 digits can be entered. Press the right arrow key (>) to set
the decimal point. Use the left arrow key (<) to delete or overwrite any entry.
• Press ENTER to confirm the entry.
Enter a Lot Number (Buffer)
Enter Incubation Time
The following display appears:
FBG Diluent Lot No.:
123456789
• Use the number keys to enter or change the lot number.
The number is on the reagent package insert. A maximum of 9 digits can be
entered. Use the left arrow key (<) to delete or overwrite any entry.
• Press ENTER to confirm the entry.
• Use the arrow keys (</>) until <7 Incub. time> is displayed.
The following display appears:
FBG Measurement:
< 7 Incub. time >
• Press ENTER to confirm.
The following display appears:
FBG Incubation:
90 s
CoaLAB 6000 – Operators Manual 1.8 Page 64
Operation
• Use the number keys to enter or change the incubation time.
A maximum of 3 digits can be entered. Use the left arrow key (<) to delete
or overwrite any entry.
• Press ENTER to confirm the entry.
Enter Start Reagent Volume
• Use the arrow keys (</>) until <8 Start reagent> is displayed.
The following display appears:
FBG Measurement:
< 8 Start reagent >
• Press ENTER to confirm.
The following display appears:
FBG volume:
50 µl
• Use the number keys to enter or change the reagent volume for
Fibrinogen.
Remember that the total test volume must be between 125µl and 250µl. A
maximum of 3 digits can be entered. Use the left arrow key (<) to delete or
overwrite any entry.
• Press ENTER to confirm the entry.
Enter a Lot Number
Select Reagent Vial Types
The following display appears:
FBG Lot No.:
123456789
• Use the number keys to enter or change the lot number of the method.
You will find the number on the reagent package insert. A maximum of 9
digits can be entered. Use the left arrow key (<) to delete or overwrite any
entry.
• Press ENTER to confirm the entry.
Select the reagent vial type for Fibrinogen and decide whether the reagent
shall be stirred. To change the setting for the reagent vial type follow the
instructions above.
The following display appears:
FBG Bottle type:
< 7 4ml Bottle-R >
• Use the arrow keys (</>) to determine which reagent vial type shall be
used and whether the reagent shall be stirred.
CoaLAB 6000 – Operators Manual 1.8 Page 65
Operation
Which varieties can be selected shows the following list.
• Use the arrow keys (</>) until <9 max meas time> is displayed.
The following display appears:
FBG Measurement:
< 9 max meas time >
• Press ENTER to confirm the maximum measuring time.
The following display appears:
FBG max M-time:
100 s
• Use the number keys to enter or change the maximum measuring time.
A maximum of 3 digits can be entered. Use the left arrow key (<) to delete
or overwrite any entry.
• Press ENTER to confirm the entry.
The following display appears:
FBG Measurement:
< 1 Sample volume >
• Press ESC to leave this menu point.
The following display appears:
FBG PARAMETER:
< 2 Measurement >
• Press ESC to leave this menu point.
The following display appears:
TEST MENU:
< 1 Modify test >
• Repeat pressing ESC until you are back in the MAIN MENU.
10 ml reagent vial without stir bar white no
10 ml reagent vial with stir bar white yes
4 ml reagent vial without stir bar blue no
4 ml reagent vial with stir bar blue yes
Before positioning a reagent vial into the sample or reagent rotor
make sure that it has been predefined for the method. The reagent
vial types supplied by the manufacturer have different filling
heights depending on their sizes (10 ml or 4ml). If you use the
wrong vial type the level might not be detected correctly or the
needle will touch the bottom of the wrong vial.
CoaLAB 6000 – Operators Manual 1.8 Page 66
Operation
6.2.2.5 Print of Test Parameters
(Fibrinogen for example)
The following describes how to print all reagent and method specific data
(Print test) of a method. Here Fibrinogen is used for example however the
other methods PT and APTT will be printed in the same way. Therefore
follow the same steps to print the other methods. To print all methods
together follow the instructions in section 6.2.5 Print of Test Parameters
(Print test).
• Use the arrow keys (</>) to select <2 TEST MENU> in the MAIN
MENU.
MAIN MENU:
< 2 TEST MENU >
• Press ENTER to confirm.
The following display appears:
TEST MENU:
PIN
• Enter the PIN code with the number keys (default= 11111).
The following display appears:
TEST MENU:
< 1 Modify test >
• Press ENTER to confirm <1 Modify test> in the TEST MENU.
The following display appears:
Select test:
PT aPTT FBG END
• Select method FBG by pressing the function key F3.
If you want to print another method press either F1 for PT or F2 for APTT. If
you press F4 = End you can leave this menu without changing anything.
• Use the arrow keys (</>) until < 6 Print test > is displayed.
FBG PARAMETER:
< 6 Print test >
• Press ENTER to confirm.
‘Print test’ will be displayed for a short moment.
Print test
CoaLAB 6000 – Operators Manual 1.8 Page 67
Operation
The following display appears:
FBG PARAMETER:
< 6 Print test >
The reagent and method specific data will be printed automatically on the
internal printer.
• Press ESC to leave this menu after printing is completed.
The following display appears:
TEST MENU:
< 1 Modify test >
• Press ESC again in order to get back into the MAIN MENU.
Example of a test parameters print
C2820501 V1.04 04/22/02
07.05.2002 11:50
T3 = FBG Fibrinogen
--------------------------
Replication : Double
max. CV : 10
Sample volume : 100µl
Diluent : Bufer
999992 : 1 + 9.0
Incub. time : 90 s
Start reagent : FBG
8888 : 50 µl
4ml Bottle-R
Airgap volume : 5 µl
Waste volume : 5 µl
max meas time : 120 s
Normal range
150 -- 400 mg/dl
1st Convers. : Interpol
In the following you will find how to load method and reagent specific data
with a ChipCARD. Besides loading these data manually (see sections 6.2.1
and 6.2.2) you can use a ChipCARD. These ChipCARDs have been
programmed with the method and reagent specific data of each reagent.
Each system is preset by the manufacturer with the test settings for PT,
APTT and FIBRINOGEN. These test settings can be changed or overwritten
with a ChipCARD.
3 tests can be loaded on the system at one time, e.g. PT, APTT, and
NOTE
Fibrinogen. If you want to load a new test with changed settings
one test from the system will be overwritten, i.e. deleted.
CoaLAB 6000
Default ChipCard
PT
30.000.1196
P-ID G 15 2 0220
_____________________
To install, refer to operators manual prior to use.
Weitere Informationen zur Installation finden Sie im Benutzerhandbuch.
Reportez-vous au manuel d´utilisation pour l´installation.
Para la instalación, consultar antes el Manuel del usuario.
Per l´installazione, leggere Il manuale di instruzioni prima dell´utillizo.
_____________________
The ChipCARDs contain all reagent and test specific data such as lot no.,
reference curve, incubation or activating time, expiry date of the reagent,
ISI, etc. Once you have loaded a ChipCARD no further settings have to be
made at the system. All data which had been entered manually beforehand
will be overwritten with the data of the ChipCARD.
For further information regarding availability, prices and application of
ChipCARDs contact your distributor or manufacturer.
• Use the arrow keys (</>) to select <2 TEST MENU> in the MAIN
MENU.
LABiTec GmbH
22926 Ahrensburg, Germany
MAIN MENU:
< 2 TEST MENU >
• Press ENTER to confirm.
The following display appears:
TEST MENU:
PIN
• Enter the PIN code with the number keys (default= 11111).
• Select <2 Read test> in the TEST MENU.
TEST MENU:
< 2 Read test >
• Lift the printer cover above the keyboard.
The ChipCARD reader is below the printer module.
• Insert the ChipCARD with the chip facing down into the slot of the
reading module. Leave the ChipCARD in the reader.
• Press ENTER to confirm.
CoaLAB 6000 – Operators Manual 1.8 Page 69
Operation
The following display appears:
ChipCARD
Keypad
Figure 15 How to load a ChipCARD
ChipCARD Reader
Printer unit
Overwrite test:
PT aPTT FBG END
• Use the function keys F1 - F3 to select the test to be overwritten by the
ChipCARD. By pressing F4 = End you can leave this menu without
making any change.
The following display appears:
Read test
‘Read test’ will be displayed for a short moment. The instrument reads the
ChipCARD.
The following display appears:
Test loaded:
Name of loaded test
A message follows that the test of the system was overwritten by the
ChipCARD test.
• Press ENTER to confirm.
If the ChipCARD was inserted wrong or not at all the following error
message will appear:
E R R O R
No CARD found
• Press ENTER to confirm this message and repeat the procedure.
CoaLAB 6000 – Operators Manual 1.8 Page 70
Operation
If you inserted a ChipCARD of which the test had been loaded already the
following error message will be displayed:
E R R O R
Test already loaded
• Press ENTER to confirm this message and repeat the procedure with
another ChipCARD.
Withdraw the ChipCARD from the reader and close the printer cover again.
If you want to overwrite more tests of the system repeat the above steps for
all new tests.
The following display appears:
TEST MENU:
< 2 Read test >
• Press ESC in order to get back into the MAIN MENU.
6.2.4 Save Tests on a
ChipCARD
Find below how to write test parameters, including calibration curve, to a
ChipCARD and how to save the data. Editable ChipCARDs can be bought
from the manufacturer, see Chapter 8.5 Accessories and Consumables.
It is possible to save test parameters for a specific reagent lot so that you
can use this test again when a reagent of the same lot is used. No
parameters have to be entered manually.
Since the procedure of saving test parameters on a ChipCARD is the same
for all tests we have taken the PT test as an example.
• Use the arrow keys (</>) to select <2 TEST MENU> in the MAIN
MENU.
MAIN MENU:
< 2 TEST MENU >
• Press ENTER to confirm.
The following display appears:
TEST MENU:
PIN
• Enter the PIN code with the number keys (default= 11111).
The following display appears:
TEST MENU:
< 1 Modify test >
• Press the arrow keys (</>) until <5 Write test> is displayed.
TEST MENU:
< 5 Write test >
CoaLAB 6000 – Operators Manual 1.8 Page 71
Operation
NOTE
• Press ENTER to select the function.
• Insert the ChipCARD into the reader. Use either a blank CARD or an
already edited ChipCARD. If you select a used ChipCARD remember that
all data on the ChipCARD will be overwritten.
Be careful not to use and overwrite the Default ChipCARD (also
supplied in the accessories).
The following display appears:
Select test:
PT aPTT FBG END
F1 F2 F3 F4
• Select the test to be saved by pressing one of the function keys F1 - F3.
By pressing F4 = End you can leave this menu without making any
changes.
The following display appears:
Write test
‘Write test’ will be displayed while the data is written to the ChipCARD.
Test written:
Name of written test
An acoustic signal will be heard at the end of saving.
• Press ENTER to confirm.
• Remove the edited ChipCARD and write all test relevant information on it
for proper identification.
• Put the printer cover back into its position.
• Press ESC to get back into a higher menu level.
Error messages
If the ChipCARD has been inserted wrong or not at all the following
message will appear:
E R R O R
No CARD found
• Press ENTER to confirm and repeat the procedure after the ChipCARD
has been inserted.
If you used the Default CARD the following message will appear:
E R R O R
Write protected card
• Press ENTER to confirm and repeat the procedure with another
ChipCARD.
CoaLAB 6000 – Operators Manual 1.8 Page 72
Operation
If a defective ChipCARD was used the following message will appear:
E R R O R
No CARD found
• Press ENTER to confirm and repeat the procedure with a new ChipCARD.
6.2.5 Print of Test Parameters
(Print test)
Follow the instructions to print all reagents and test specific data (print test)
of a method. We have used Fibrinogen for example however the other
methods PT and APTT will be printed in the same way. Therefore follow the
same steps to print the other methods.
• Use the arrow keys (</>) to select <2 TEST MENU> in the MAIN
MENU.
MAIN MENU:
< 2 TEST MENU >
• Press ENTER to confirm.
The following display appears:
TEST MENU:
PIN
• Enter the PIN code with the number keys (default= 11111).
• Use the arrow keys (</>) until <3 Print test> is displayed.
TEST MENU:
< 3 Print test >
• Press ENTER to confirm.
The following display appears:
Select test:
PT aPTT FBG END
• Select method FBG by pressing the function key F3. If you want to print
another method press either F1 for PT or F2 for APTT. If you press F4 = End you can leave this menu without changing anything.
The following display appears:
Print test
The test specific data for Fibrinogen will be printed automatically on the
internal printer.
• Press ESC to get back into the MAIN MENU.
CoaLAB 6000 – Operators Manual 1.8 Page 73
Operation
6.2.6 Delete Test
In the following we have described how to delete tests which are stored in
the system. Remember that all reagent and test specific data of this method
will be deleted completely.
• Use the arrow keys (</>) to select <2 TEST MENU> in the MAIN
MENU.
MAIN MENU:
< 2 TEST MENU >
• Press ENTER to confirm.
The following display appears:
TEST MENU:
PIN
• Enter the PIN code with the number keys (default= 11111).
• Use the arrow keys (</>) until <3 Print test> is displayed.
TEST MENU:
< 4 Erase test >
• Press ENTER to confirm.
The following display appears:
Select test:
PT aPTT FBG END
• Using the function keys F1 - F4 select the test to be deleted. By pressing
F4 = End you can leave this menu without making any changes.
The following display appears:
Really erase test?
YES no
You will be asked again whether you really want to delete the selected test.
Use either F1 = YES or F2 = NO.
TEST MENU:
< 4 Erase test >
• Confirm deletion by pressing F1 = YES.
• Press ESC to get back into the MAIN MENU.
CoaLAB 6000 – Operators Manual 1.8 Page 74
Operation
6.3 STAT Management
In the following we have described the STAT management which explains
how to run emergency tests during routine operation.
Introduction
Up to 3 samples may be preset for STAT on the sample/reagent rotor, see
section 6.4.2 Define STAT positions in the sample rotor. If you want to run
STAT tests you can either:
1) Preset up to 3 samples on the rotor for STAT beforehand, see section
6.4.2. Positions 16 to 18 will be used on the rotor.
2) Or make no reservations of any special position on the rotor. The
system will position STAT tests automatically into the next available rotor
positions. If all sample rotor positions are occupied by routine
measurements and no STAT positions have been reserved you have to
clear the complete joblist first.
NOTE
We recommend to pre-set 1-3 positions for emergency measurements
to run STAT tests during routine operation.
Follow section 6.1.11 Interrupt a current run (User interrupt) run if you are in
routine operation and want to run a STAT test. All routine tests which have
not been worked off will remain in the joblist and will be measured after the
STAT test.
Follow the instructions for STAT measurements:
• Use the arrow keys (</>) to select <3 STAT MANAGEMENT> in the MAIN
MENU.
MAIN MENU:
< 3 STAT MANAGEMENT >
• Press ENTER to confirm.
The following display appears:
STAT MANAGEMENT:
Edit Ins Prnt Start
If you are in routine operation you can go directly into this method by
pressing F4 = STAT (see section 6.1.11 Interrupt a current run (User
interrupt). Before entering the STAT joblist the running measurements from
routine operation will be completed. Using the function keys F1 - F4 select
the desired action from the table:
Function key Order Meaning
F1 Edit If there is a joblist from routine operation you
can transfer patients from this list into the
STAT job list. Further you can edit or change
any existing STAT joblist.
F2 Ins To enter new emergency patients into the
STAT joblist. The routine joblist will not be
influenced, if free positions are available.
F3 Prnt To print the STAT joblist.
F4 Start To begin STAT measurements.
CoaLAB 6000 – Operators Manual 1.8 Page 75
Operation
Example F1 = Edit
• Change from ROUTINE into STAT operation:
• Press F1 for editing.
The following display appears:
01 S-ID: 30001
pt aptt fbg all
The first patient of your ROUTINE joblist will be displayed and can be
transferred into the STAT joblist. With the function keys F1-F4 you can
select the methods. The selected test will be displayed in big letters. With
the arrow keys up and down the next sample can be called for or skipped
over.
The patient samples that have been transferred from ROUTINE into the
STAT joblist will be deleted in the ROUTINE list. All patient samples which
have not been transferred into the STAT list will remain in the ROUTINE list
for later measurements.
• Press ENTER to confirm.
The following display appears:
STAT MANAGEMENT:
Edit Ins Prnt Start
If you want to print the STAT joblist press F3 = print. If you want to start
STAT measurements press F4 = Start. For further entry of new STAT
patients press F2 = Ins and follow the example below.
Example F2 = Insert
Enter new patient samples into the STAT joblist:
• Press F2 for inserting.
The following display appears:
05 S-ID: 30005
pt aptt fbg all
The system will extend the next available sample rotor position
automatically so that you can place the new primary tube. Press F1-F4 to
select the methods. If you want to change the patient ID press the left arrow
key (<) until the displayed ID is deleted. Now you can enter another ID with
the number keys.
• Press ENTER to confirm.
The following display appears:
06 S-ID: 30006
pt aptt fbg all
If you want to insert further patient samples into the STAT joblist follow the
steps as mentioned before. Press ESC to leave this menu.
CoaLAB 6000 – Operators Manual 1.8 Page 76
Operation
The following display appears:
STAT MANAGEMENT:
Edit Ins Prnt Start
Example F3 = Prnt
Print STAT joblist:
• Press F3 for printing.
The following display appears:
Printing joblist
The STAT job list will be printed automatically on the internal printer.
After the print the following message will be displayed:
STAT MANAGEMENT:
Edit Ins Prnt Start
Example of the printed STAT joblist:
C2820501 V1.04 04/22/02
07.05.2002 11:50
STAT Joblist
T1 = PT PT
T2 = APTT APTT
T3 = FBG Fibrinogen
No Sample-ID T1 T2 T3
1 30001 x x x
2 30002 x x x
3 30005 x x x
-------------------------
Example F4 = Start
Start the STAT joblist:
• Press F4 for start.
The following display appears:
Prepare measure…
The following display appears:
INFO
Check waste box
CoaLAB 6000 – Operators Manual 1.8 Page 77
Operation
‘Check waste box’ will be displayed for a short moment every time a job
list has been entered. If used cuvette racks are still in the waste box we
recommend to empty it.
The following display appears:
Insert cuvette bar
and close cover
The system calculates from the joblist automatically how many cuvette
racks will be needed. The sample rotor extends the next position
automatically to load a cuvette rack. This procedure will be repeated until all
cuvette racks necessary for the measurements have been positioned in the
sample rotor.
• Press ENTER to confirm.
The sample rotor will extend the next available reagent position
automatically. Insert the adapter ring and the reagent vial for the selected
method. Remember to use predefined reagent vial types only (see Chapter
6.2.2.2). If reagents are already loaded the analyzer will check reagent
volume and starts measurement automatically.
Test will not function without any reagents.
The necessary reagent volume will be displayed.
1: washer:
LOAD 0.5ml skip
If the reagent vial is in the sample rotor you can either select
F1 (load) or = the reagent has been loaded and confirmed.
ENTER
F4 (skip) = the reagent has not been loaded no entry confirmed
continue to next reagent.
• Press F1 or ENTER in this case to confirm.
The following display appears:
2: Buffer 9999992
LOAD 0.4ml skip
• The sample rotor position will be extended automatically. Load Buffer into
the sample rotor and confirm with F1.
If you want to skip over this position and not load Buffer press F4.
The following display appears:
3: FBG 8888
LOAD 0.3ml skip
• The sample rotor position will be extended automatically. Load
Fibrinogen into the sample rotor and confirm with F1.
If you want to skip over this position and not load Fibrinogen press F4.
CoaLAB 6000 – Operators Manual 1.8 Page 78
Operation
The following display appears:
4: CaCl2 12345678
LOAD 0.3ml skip
• The sample rotor position will be extended automatically. Load Calcium
Chloride into the sample rotor and confirm with F1.
If you want to skip over this position and not load Calcium Chloride press
F4.
The following display appears:
5: aPTT 999999995
LOAD 0.3ml skip
• The sample rotor position will be extended automatically. Load aPTT
reagent into the sample rotor and confirm with F1.
If you want to skip over this position and not load aPTT reagent press F4.
The following display appears:
6: PT 123456789
LOAD 0.3ml skip
• The sample rotor position will be extended automatically. Load PT reagent
into the sample rotor and confirm with F1.
If you want to skip over this position and not load PT reagent press F4.
The following display appears:
Prepare measure…
After having placed the last reagent into the sample rotor of the system will
check the reagents automatically. Especially the volume will be checked. If
reagents are missing or the volume is insufficient the following message will
be displayed:
6: PT 123456789
Only x.x of x.x ml
You can load the missing reagent now.
• Press ENTER to confirm.
The following display appears:
Pipettor M-Block
37.4 38.4
The message Pipettor / Tubing gives a short status on the temperature of
the measuring block and the pipettor. If the temperature is not correct, i.e.
37°C or 38°C, this message will be displayed until the temperature has
been reached.
CoaLAB 6000 – Operators Manual 1.8 Page 79
Operation
NOTE
The following display appears:
18 Samples
No Results available
The system starts measuring. The message displayed shows how many
samples still have to be worked off. The results will be printed automatically
on the internal printer.
If results are not available yet the message ‘No results available’ will be
displayed.
You can interrupt the STAT measurement by pressing the ESC-key.
Follow the instructions as per Chapter 6.1.11 Interrupt a current run
(User interrupt).
After the measurement has been completed a beep will be heard and the
following message will be displayed:
STATUS
Normal Termination
Example for a result print:
C2820501 V1.04 04/22/02
07.05.2002 11:50
Analyzer Mode : Batch
Research Mode : Off
**** STAT MANAGEMENT ****
Test PT 07.05.2002 15:10
No Sample-ID Result Cuv.
1 30001 online 01ab
PT time1 11.7 sec
PT time2 11.7 sec
PT mean 11.7 sec
PT INR 1.03
PT conc 93.75 %
Test aPTT 07.05.2002 15:17
No Sample-ID Result Cuv.
1 30001 online 01ef
aPTT time1 35.3 sec
aPTT time2 35.7 sec
aPTT mean 35.5 sec
aPTT Ratio 1.77
-------------------------**END STAT MEAUREMENT**
Finished 07.05.2002 15:45
State: Normal Termination
CoaLAB 6000 – Operators Manual 1.8 Page 80
Operation
6.4 General Setting (SETUP
MENU)
Basic settings of the system
Before working with the system in routine operation several basic settings
must be entered. The following default parameters have been preset by the
manufacturer:
Standard settings of the
system
System parameters
Name Standard setting
Primary tube Sample Cup
Reserved STAT positions 0
Analyzer mode Batch
PIN code 11111
Language English
Printer mode list
Beeper Setup long
Reagent or Method Parameters:
Name Standard setting of reagent or method parameters
Reagent volume - 50 µl 1 + 9.0 Reagent vial type - 4 ml 4 ml 10 ml
Reagent stir function - inactive - Adapter ring - blue blue White
Teflon stir bar - no no No
Incubation time 70 sec 180 sec 90 sec -
Start reagent PT CaCl2 FBG -
Reagent volume 100 µl 50 µl 50 µl Reagent vial type 4 ml 4 ml 4 ml Reagent stir function active inactive inactive Adapter ring Blue blue blue Teflon stir bar Yes no yes Airgap volume 5 µl 5 µl 5 µl Waste volume 5 µl 5 µl 5 µl Max. meas. time 80 sec 120 sec 120 sec Normal range 7.0-13.0 sec 25-45 sec 150-400 mg/dl -
1. conversion none none regression -
dep. on lot. - dep. on lot. Extrapolation range - - 45-1000 mg/dl -
2. conversion none none none -
100% dep. on lot - - ISI dep. on lot - - -
PT aPTT Fibrinogen Washing Solution
CoaLAB 6000 – Operators Manual 1.8 Page 81
Operation
NOTE
6.4.1 Define Primary Tubes
NOTE
NOTE
If you do not wish to change any of the manufacturer’s preset default
settings continue with ROUTINE operation, see section 6.1.
If you want to change any of these settings follow the steps below:
Before entering a joblist in routine operation the primary tubes can be
selected. The system has a recommended number of different types. If you
use the Sample cup which is preset by the manufacturer you can ignore the
following instructions.
To secure perfect operation of the system only recommended primary
tubes should be used.
• Use the arrow keys (</>) to select <4 SETUP MENU> in the MAIN
MENU.
MAIN MENU:
< 4 SETUP Menu >
• Press ENTER to confirm.
Before getting access to the SETUP MENU a PIN code has to be
entered with the number keys. The preset code is 11111. You can
change this code by following the instructions as per chapter 6.4.6 Set
PIN Code. Remember to write this code down and keep it at a safe
place, accessible to authorised users only.
The following display appears:
SETUP MENU:
PIN
• Enter the PIN code with the number keys (default= 11111).
• Press the arrow keys (</>) until <1 Primary Tube> is displayed in the
SETUP MENU.
The following display appears:
SETUP MENU:
< 1 Primary Tube >
• Press ENTER to confirm.
The following display appears:
Primary Tube:
< 1 Sample Cup >
The displayed primary tube is the type entered last. By pressing the arrow
keys (</>) you can select another type.
CoaLAB 6000 – Operators Manual 1.8 Page 82
Operation
• Press ENTER to confirm.
The following list will give an overview of the primary tubes which can be
used on the system.
NOTE
Never go below the recommended dead volume or results may be
affected. Only one type can be used at a time. A mixture of tubes
may damage the pipettor.
Pos No. Primary Tube Dead volume Max. Volume Manufacturer
Follow the instructions to define or reserve STAT positions in the sample
rotor. A maximum of 3 positions may be reserved for STAT samples. If no
STAT positions have been reserved, the instrument will use the next free
position available.
• Use the arrow keys (</>) to select <4 SETUP MENU> in the MAIN
MENU.
• Press ENTER to confirm.
The following display appears:
• Enter the PIN code with the number keys (default= 11111).
• Press the arrow keys (</>) until <2 No of STAT pos.> is displayed in the
The following display appears:
• Press ENTER to confirm.
MAIN MENU:
< 4 SETUP Menu >
SETUP MENU:
PIN
SETUP MENU.
SETUP MENU:
< 2 No of STAT pos >
CoaLAB 6000 – Operators Manual 1.8 Page 83
Operation
6.4.3 Date and Time
The following display appears:
No of STAT pos
No of STAT pos: 0
The displayed number of reserved STAT positions is the number entered
last. By pressing the number keys (0-3) you can enter the selected number
of STAT positions.
• Press ENTER to confirm.
The following display appears:
SETUP MENU:
< 2 No of STAT pos >
• Press ESC to get back into the MAIN MENU.
As the date and time are printed at the top of each run’s results, it is
important that they are entered accurately.
• Use the arrow keys (</>) to select <4 SETUP MENU> in the MAIN
MENU.
MAIN MENU:
< 4 SETUP Menu >
• Press ENTER to confirm.
The following display appears:
SETUP MENU:
PIN
• Enter the PIN code with the number keys (default= 11111).
• Press the arrow keys (</>) until <3 Date / Time> is displayed in the
SETUP MENU.
The following display appears:
SETUP MENU:
< 3 Date / Time >
• Press ENTER to confirm.
The following display appears:
Date / Time:
17.04.2002 16:34:58
CoaLAB 6000 – Operators Manual 1.8 Page 84
Operation
• Use the number keys to enter the date and time. With the left arrow key
(<) you can delete or overwrite.
• Press ENTER to confirm.
The following display appears:
SETUP MENU:
< 3 Date / Time >
• Press ESC to get back into the MAIN MENU.
6.4.4 Printer Setup
This section describes the printer configurations. You can set the operation
mode of the analyzer internal printer according to your requirements.
• Use the arrow keys (</>) to select <4 SETUP MENU> in the MAIN
MENU.
MAIN MENU:
< 4 SETUP Menu >
• Press ENTER to confirm.
The following display appears:
SETUP MENU:
PIN
• Enter the PIN code with the number keys (default= 11111).
• Press the arrow keys (</>) until <4 Printer setup> is displayed in the
SETUP MENU.
The following display appears:
SETUP MENU:
< 4 Printer setup >
• Press ENTER to confirm.
Printer setup:
AUTO list on off
F1 F2 F3 F4
Using the function keys F1-F4, select the desired operation mode from the
table below. The activated operation mode will be displayed in bold letters.
NOTE
Please note, by using the function key F4 = off in general no print
out will be done by the internal printer.
CoaLAB 6000 – Operators Manual 1.8 Page 85
Operation
Function
key
F1 auto All printing orders will be transmitted to the internal printer.
F2 list All printing orders will be transmitted to the internal printer.
F3 on No online print of results. To print results, activate ‘print
F4 OFF Turns the internal printed off. No printing will occur.
6.4.5 Beeper Setup
Message
Meaning
displayed
The measurement results will be printed test oriented
(online), that means according to the availability of results.
The measurement results will be printed patient oriented,
that means once all results per patient are available.
results’ see section 6.1.12 Print results, send results to
HOST. All other printing orders will be executed.
The following display appears:
SETUP MENU:
< 4 Printer setup >
• Press ESC to get back into the MAIN MENU.
This section allows the user to set the length of the signals when a status
massage or error has occurred or testing is complete. See below the 3
basic signal tones you can choose from.
1.) Long, followed by repeatedly short tones warns of future action
necessary and can be switched off by acknowledging the action and
confirming it with the ENTER key. Same tones will also sound by
pressing a key on the keyboard to acoustically confirm the entry.
2.) Short, followed by a one-time signal tone, warns that the system
waits for the next command to complete the action and remains in this
position until the operator gives the next command and confirm it with
the ENTER key. Same tones will also sound by pressing a key on the
keyboard to acoustically confirm the entry.
3.) Off, no signal tones will be carried-out.
• Use the arrow keys (</>) to select <4 SETUP MENU> in the MAIN
MENU.
MAIN MENU:
< 4 SETUP Menu >
• Press ENTER to confirm.
The following display appears:
SETUP MENU:
PIN
• Enter the PIN code with the number keys (default= 11111).
CoaLAB 6000 – Operators Manual 1.8 Page 86
Operation
• Press the arrow keys (</>) until < 5 Beeper Setup > is displayed in the
SETUP MENU.
SETUP MENU:
< 5 Beeper setup >
• Press ENTER to confirm.
The following display appears:
Beeper setup:
long short off
F1 F2 F4
With the function keys F1 + F2 you can select the beeper mode. F4 will
deactivate the beeper. The activated mode will be displayed in big letters.
The following display appears:
SETUP MENU:
< 5 Beeper setup >
• Press ESC to get back into the MAIN MENU.
6.4.6 Set PIN Code
In the following we have explained how to change the PIN code of the
system. The PIN is preset to 11111 by the manufacturer and will be active
until modified by the user. The PIN code will be asked for before entering
the TEST MENU and the SETUP MENU.
• Use the arrow keys (</>) to select <4 SETUP MENU> in the MAIN
MENU.
MAIN MENU:
< 4 SETUP Menu >
• Press ENTER to confirm.
The following display appears:
SETUP MENU:
PIN
• Enter the PIN code with the number keys (default= 11111).
• Press the arrow keys (</>) until <6 Modify PIN> is displayed in the
SETUP MENU.
The following display appears:
SETUP MENU:
< 6 Modify PIN >
• Press ENTER to confirm.
The following display appears:
CoaLAB 6000 – Operators Manual 1.8 Page 87
Operation
6.4.7 Language
Modify PIN
New PIN: 11111
The actual PIN code will be displayed. You can change it with the number
keys. A maximum of 5 numbers can be entered. The PIN code has to be
defined between 1 until 65535. Use the left arrow key (<) if you want to
delete or overwrite your entry.
• Press ENTER to confirm.
The following display appears:
SETUP MENU:
< 6 Modify PIN >
• Press ESC to get back into the MAIN MENU.
Follow the instructions if you want to change the language in the system.
You can select between English and German.
• Use the arrow keys (</>) to select <4 SETUP MENU> in the MAIN
MENU.
MAIN MENU:
< 4 SETUP Menu >
• Press ENTER to confirm.
The following display appears:
SETUP MENU:
PIN
• Enter the PIN code with the number keys (default= 11111).
• Press the arrow keys (</>) until <7 Language> is displayed in the
SETUP MENU.
The following display appears:
SETUP MENU:
< 7 Language >
• Press ENTER to confirm.
The following display appears:
Select language:
ENG deu end
F1 F2 F4
The activated language will be displayed in big letters. With the function
keys F1 = English and F2 = German you can select the desired language.
If you press F4 = end you will leave this menu without changing anything.
CoaLAB 6000 – Operators Manual 1.8 Page 88
Operation
The following display appears:
SETUP MENU:
< 7 Language >
• Press ESC to get back into the MAIN MENU.
6.4.8 Print of Setup
Parameters
Follow the instructions to print the parameters described in section 6.4
General Setting (SETUP MENU) on the internal printer. Printing the SETUP
parameters provides an overview of the general setup of the system.
• Use the arrow keys (</>) to select <4 SETUP MENU> in the MAIN
MENU.
MAIN MENU:
< 4 SETUP Menu >
• Press ENTER to confirm.
The following display appears:
SETUP MENU:
PIN
• Enter the PIN code with the number keys (default= 11111).
• Press the arrow keys (</>) until <8 Print Setup> is displayed in the
SETUP MENU.
SETUP MENU:
< 8 Print Setup >
• Press ENTER to confirm.
The following display appears:
Print Setup
The SETUP parameters will be printed on the internal printer.
The following display appears:
SETUP MENU:
< 8 Print Setup >
• Press ESC to get back into the MAIN MENU.
CoaLAB 6000 – Operators Manual 1.8 Page 89
Operation
6.5 Removing cuvette bars
from the system
In the following chapter we have described how to remove cuvette bars
from the measuring block.
• Use the arrow keys (</>) to select <5 UTILITY MENU> in the MAIN
MENU.
MAIN MENU
< 5 UTILITY MENU >
• Press ENTER to confirm.
The following display appears:
UTILITY MENU:
< 1 Init Instrument >
• Press ENTER to confirm the message.
The following display appears:
Init Instrument
Init ... (ESC=Stop)
The motors and the diluter will be initialised. During this phase the cuvette
racks in the measuring block will be moved one position to the right and will
be ejected from the measuring block.
The following display appears:
Init Instrument
Flush ... (ESC=Stop)
Then the liquid system will be initialised. Liquids are pumped into the
tubing’s.
• Press ESC if you want to finish initialisation.
The following display appears:
UTILITY MENU:
< 1 Init Instrument >
• Press ESC to get back into the MAIN MENU.
6.6 Switch OFF the system
Prior to switching off the system make sure the system is cleaned and
disinfected from all residues of reagent and sample material inside the liquid
system as well as from the outer side of the analyzer. Depending on how
long you would like to shut down the system it is necessary to follow
different wash/flush cycles to avoid any residues of reagents/samples will
remain in the liquid system.
To clean/disinfect the system follow the instructions as described in sections
6.1.2 Rinsing with distilled water, 6.1.3 Disinfection of waste tubing
(Terralin), change of FW-Filter or 6.1.4 Rinsing with Washing Solution.
CoaLAB 6000 – Operators Manual 1.8 Page 90
Operation
In case of a long term shut down it is recommended to also dry the system
completely. Follow the instructions in section 8.8 Packing of the system,
preparation of transport as described under 3. - 6. This procedure should be
done if the shut down time exceeds 4 weeks.
Remove all reagents must be removed from the sample/reagent rotor.
Follow the instructions in section 6.1.5 Loading and unloading reagents.
Remove all samples/cups from the sample/reagent rotor positions 1-18 by
erasing the joblist. Follow the instructions described in section 6.1.6 Delete
joblist, delete samples. Make sure all results of the current joblist are printed
out prior to erasing the joblist. Once the joblist has been deleted no results
will be available anymore.
Make sure that finally the instrument is free of cuvette bars, vials, cups and
adapter rings.
Now the system can be switched OFF by setting the power switch to the
position O = ON.
CoaLAB 6000 – Operators Manual 1.8 Page 91
Operation
7 Failures
Status messages:
Application errors:
Non-fatal errors:
Fatal errors:
Possible error messages, their causes and trouble shooting are described in
this section. These failures or status messages can be displayed as well as
printed. The messages will be differentiated according to the following
scheme:
System messages for the general information of the user. These are no
errors/failures.
Failures/errors to be corrected by the user.
Functional system errors to be corrected by the user.
Errors to be corrected by the technical customer service only. Contact your
authorised distributor or the manufacturer.
CoaLAB 6000 – Operators Manual 1.8 Page 92
Operation
7.1 Status Messages
Message
Module:<name> (<Adr>)
Timeout Req = <ID>
Example
Module: Control (07)
Timeout Req = 43
STATE
Out of system liquid
STATE
Waste bottle full
STATE
System liquid error
STATE
Missing reagent
STATE
User Interrupt
CauseCorrections
Message has no influence on
functions of the system.
Message only displayed and
printed out.
No influence on measurements.
The dist. water container is empty Fill water into the container see
chapter 6.1.13.
The waste water container is full. Empty the container see chapter
6.1.14.
The dist. water container is empty
and/or the waste water container
Fill dist. water container and/or
empty waste water container.
is full.
A reagent was not loaded or
positioned in the rotor
Measurement interrupted by the
Reload or position missing
reagent, see chapter 6.1.5.
See Chapter 6.1.11.
user.
STATE
A routine run was completed. Message for the user.
Normal Termination
INFO
No Space for new Sample
INFO
No samples loaded
More than 18 possible samples
were to be loaded
or
no further positions available due
to STAT positions reserved.
No joblist available, i.e. no
samples have been loaded.
Further samples must be
measured later.
Reserved STAT positions could
be reduced, see chapter 6.4.2.
Work out joblist and load
samples, see chapter 6.1.7
ERROR
Test volume too big
The permissible volume was
exceeded.
Check volume settings for sample
and reagent volume.
(Test volume 125 µl-250 µl)
ERROR
Test volume too low
The volume is too low. Check volume settings for sample
and reagent volume.
(Test volume 125 µl-250 µl).
No print out Printer is not active Switch on printer, refer to chapter
6.4.4Printer Setup
CoaLAB 6000 – Operators Manual 1.8 Page 93
Operation
7.2 Application Errors
Message
Module: Control (07)
Move Error Motor 0
Module: Control (07)
Move Error Motor 3
Module: Diluter (03)
Move Error Motor 0
ERROR
No chip card found
ERROR
Test already loaded
ERROR
Too many reagents
ERROR
Not enough sample
ERROR
Not enough reagent
ERROR
Incubation warning
ERROR
pipettor not conn.
fresh water sensor
No sensor 1 found
Waste water sensor
No sensor 2 found
ERROR
Level Sensor Error
STATE
Wash station full
STATE
Out of washer liquid
CauseCorrections
Cuvette rack or sample cup was
not positioned properly.
Cuvette rack in the rotor or in the
measuring block is not positioned
properly.
The syringe is not mounted
properly to the diluter or the space
below the syringe is too small.
A ChipCARD was not inserted
properly or not at all.
The selected test on the ChipCARD had already been loaded.
The number of reagent positions
does not correspond to the
number of loaded methods.
Not enough samples for pipetting
or a wrong primary tube used.
Not enough reagents available or
a wrong reagent vial used.
The incubation time was
exceeded.
Flat cable to pipettor is not
connected properly or cable has
been damaged.
Fresh Water Sensor is not
connected.
Waste Water Sensor is not
connected.
Liquid between Level Sensor and
needle at the pipettor
Waste Water Tube is bended or
not connected to analyzer
Too less washing solution Refill washing solution to rotor
Check correct position of racks
and/or sample cups. Does this
error appear again, see fatal
errors!
Either remove rack with Init
Instrument (Chapter 6.5) or switch
Instrument off and remove cuvette
rack manually. Does this error
appear again, see fatal errors!
Check correct mounting of
syringes (Chapter 5.1.4.1) or clear
space below
Re-insert ChipCARD, repeat
procedure (Chapter 6.2.3).
No corrections necessary.
A method must be deleted
(Chapter 6.2.6).
Refill sample or replace primary
tube (Chapter 6.4.1).
Refill reagent or replace reagent
vial.
Check method parameters
regarding incubation times,
maximum measuring times and
activation times (reagent 3).
Check correct connection of
cable. Has cable been damaged
contact customer service.
Switch Instrument off. Connect
fresh water sensor. Re-start
system.
Switch Instrument off. Connect
waste water sensor. Re-start
system.
Remove liquid carefully, press
ENTER to confirm.
Check correct connection of tube;
refer to Fatal errors if problem still
exists!
position R1.
CoaLAB 6000 – Operators Manual 1.8 Page 94
Operation
7.3 Non-Fatal Errors
Message
ERROR
Write protected card
ERROR
No valid test card
ERROR
no writing
Module: <Name> (<Adr>)
ParBlock Error <Nr>
Example:
Module: Meas1 (0A)
ParBlock Error2
break bot lim
break top lim Sample is very light and cannot
Too dark Sample is too dark and cannot be
break drift The measuring signal could not
break noise strong noise during sample
break jump signal peak during measure
No clotting found No clotting was found during
Time Range error The permissible deviation of CV’s
Conc. Overflow The extrapolation range was
Conc. Underflow The extrapolation range was not
Normal Range The result is out of the normal
CauseCorrections
A protected ChipCARD was used
to save a method.
Repeat procedure with a non-
protected ChipCARD.
The ChipCARD is invalid. Repeat procedure with a valid
ChipCARD.
ChipCARD or reader is defective.
Chip CARD was insert wrong.
Use new ChipCARD or insert
correctly. Otherwise contact
customer service.
Check sum error in the memory. The system loads all default
parameters automatically. Check
method and instrument data.
Some parameters might have to
be redefined. See Chapter 6.2.1,
6.2.2, 6.4
Sample is too turbid, too dark and
cannot be measured.
Prepare new test and repeat
measurement. If this message
appears often contact the
customer service.
Prepare new test and repeat
be measured.
measurement. If this message
appears often contact customer
service.
Prepare new test and repeat
measured.
measurement. If this message
appears frequently contact the
customer service.
Check reagent and tubes for
be adjusted or is unsteady
particles.
Check for air bubbles or particles.
adjustment.
Check sample for air bubbles.
Check the max meas time. Dilute
maximum measuring time.
sample and repeat measurement.
Repeat measurement.
was exceeded.
Check reference curve and repeat
exceeded during conversion.
measurement. This message
does not necessarily mean a
wrong result.
Check reference curve and repeat
reached.
measurement. This message
does not necessarily mean a
wrong result.
Check normal ranges unless it is
range.
sure that the patient really is
outside the normal ranges.
CoaLAB 6000 – Operators Manual 1.8 Page 95
Operation
7.4 Fatal Errors
Message
Module: Control (07)
Move Error Motor 1
Module: Control (07)
Move Error Motor 2
Module: Diluter (03)
Move Error Motor 1
Module: <Name> (<Adr>)
Range Error Par <Nr>
Example:
Module: Meas1 (0A)
Range Error Par 1
mixer error One or more stir motors in the
Module: <Name>(<Adr>)
Syntax Error Par <Nr>
Example:
Module: Control (07)
Sytax Error 0
Module: <Name>(<Adr>)
Com. Error <Nr>
Example:
Module: Control (07)
Com. Error 0
Module: <Name>(<Adr>)
Command Error <Nr>
Example:
Module: Meas1 (0A)
Command Error 2
Module: <Name> (<Adr>)
Unknown Error
Example:
Module: Control (07)
Unknown Error
STATE
Wash station full
Temperature error The temperature is not stable
CauseCorrections
X-motor is blocked and/or does
not reach its normal value. Xmotor is not connected properly or
can not reach its home-position.
Z-motor is blocked and/or does
not reach its normal value. Zmotor is not connected properly or
can not reach its home-position.
Functional failure at the diluter
valve.
Functional failure in the module
communication.
measuring cell are defective.
Functional failure in the module
communication.
Functional failure in the module
communication.
Functional failure in the module
communication.
Functional failure in the module
communication.
Waste water pump defective. Contact customer service, first
during the measurement.
Check whether the pipettor can
move freely from right to left
Otherwise contact customer
service.
Check whether the pipettor can
move freely up and down.
Otherwise contact customer
service.
Contact customer service.
Contact customer service.
Contact customer service.
Contact customer service.
Contact customer service.
Contact customer service.
Contact customer service.
refer to application errors!
Contact customer service.
CoaLAB 6000 – Operators Manual 1.8 Page 96
Operation
Message
Cause Corrections
Parameter error
Severe soft- or hardware failure. Contact customer service.
System stopped!
<Name>(<Adr>) not found
Severe soft- or hardware failure. Contact customer service.
System stopped!
Example:
Dilutor (03) not found
System stopped!
Module: Control (07)
Move Error Motor 0
Rotor motor does not reach
home-position or is not connected
Contact customer service, or see
application errors!
properly.
Module: Control (07)
Move Error Motor 3
M-Block motor does not reach
home-position or is not connected
Contact customer service, or see
application errors!
properly. Light barrier is
damaged.
7.5 Change fuses
Fuses may need to be changed if the system does not operate even though
it is properly connected.
Two fuses are required to operate the system:
230 VAC (Europe)115 VAC (USA)
4 amp/250 V time lag
according to: IEC 60127-2/III
4 amp/125 V time lag
according to: UL 248-14
Proceed as follows to change fuses:
• Switch off system
• Disconnect system from power supply and remove cable.
• Carefully lift the housing cover open, pull 1 (see figure 16).
• A red fuse carrier (= voltage selector) becomes visible.
• Place the screwdriver in the upper slot of the fuse carrier, pull 2 (see
figure 16) and drive carrier out of the housing.
• Inspect fuses to see if wires have melted and replace accordingly.
• Insert fuse carrier into housing.
• Close cover of the fuse carrier.
• Reconnect the system to power supply and check proper operation.
Contact the authorised customer service if the exchange of fuses has been
unsuccessful.
CoaLAB 6000 – Operators Manual 1.8 Page 97
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.