INSTRUCTION MANUAL
Laboratory Equipment Pty Ltd
email: sales@labec.com.au
Ph: 02 9560 2 811 • Fax: 02 9560 61 3 1
www.labec.com.au
2014
OPERATION & MAINTENANCE MANUAL
FOR
SCEN-206
LED & MULTI-FUNCTION TYPE CENTRIFUGE
PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION.
INDEX
(Ⅰ) SPECIFICATION---------------------------------------------------------------- 1
(Ⅱ) ROTOR AND ACCESSORY---------------------------------------------------- 2
(Ⅲ) INSTALLATION------------------------------------------------------------------ 3
(Ⅳ) OPERATION--------------------------------------------------------------------- 4
(Ⅴ) REPAIR---------------------------------------------------------------------------- 8
(Ⅵ) PARTS LIST----------------------------------------------------------------------- 10
(Ⅶ) CIRCUIT DIAGRAM------------------------------------------------------------ 11
(Ⅰ) SPECIFICATION
Speed (rpm) 0~6,000 LED display
Max. centrifugal Force 3,980 x g
Max. Capacity 300 ml
Electronic Control
Safety Device
Automatic Lid lock, lift cover and power cut off
Timer 1 ~ 99 min. digital time setting
Weight (kg) 10
Dimension (cm) 29 x 32 x 24.5
Brake System Automatically brake system
Power 110 / 220V , 50 / 60Hz
Power Consumption 150W
Microprocessor control system
10 memory programs available
Imbalance and cut off power system
Motor at over heat protective device
Triple balance system
Glass fuse 6A
1
(Ⅱ) ROTOR AND ACCESSORY
1. Centrifugal force calculation: Use its speed and radius to calculate force as following
formula.
RCF= (rpm)
R: radius (cm)
RCF: Relative Centrifugal Force ( x g )
2. Acceptable speed and load:
Please refer to the form (Ⅱ-4) to see the max. speed each model of rotor and tube carrier
can be load. The load weight including: rotor, tube carrier, washer, samples, and so on.
3. Sterilizing and cleaning rotor:
Rotor is made of stainless steel or steel or aluminum with high density coating and highly
resistant to corrosion, but not fit to alkaline and strong acid substance.
Note: Don’t heat the rotor to a high temperature (100℃) to clean or sterilize it.
4. Rotor Specifications:
Rotor
2
x 0.00001118 x R
Capacity
(ml x pcs)
Max. RCF
(x g)
Tube size (φx l ) mm
Cat. No.
Max. Min.
RA-1508
RA-1512S
RA-5004
RA-5006
REA-1520
** Tubes are not attachable. **
15 x 8 3,940 16.5 x 122 15 x 90 2-5052-11
15 x 12 3,940 16.5 x 122 15 x 90 2-5052-21
50 x 4 3,980 29.5 x 115 27.5 x 90 2-5052-12
50 x 6 3,980 29.5 x 115 27.5 x 90 2-5052-22
1.5 x 20 3,140 11 x 48 10 x 40 2-5052-13
Before using the centrifuge, please must refer to the rotor’s max. speed setting value.
2