Please, read and follow strictly all the instructions in the manual.
Due to continuous improvement in the device, information contained in this
manual is subject to change without notice. Contact the company-producer
for revisions and corrections
MRC Ltd
OFFICES: HAGAVISH 3, HOLON 58817
P.O.B. 111, HOLON 58100, ISRAEL TEL: 972-3-5595252 FAX: 972-3-5594529
Ultrasonic milk analyser
SAFETY INSTRUCTIONS
1. Read this manual carefully and make sure that you understand all the
instructions.
2. For safety purposes the device is equipped with grounded power
cable. If there is no grounded electrical outlet where the device will
be used, please, install such before using the device.
3. Place the device on leveled and stable plate. In case it falls or is
severely shocked it may be damaged.
4. Connect to the electrical network in such a way that the power cable
to stay away from the side for accessing the device and not to be
stepped on.
5. Every time before cleaning the device switch it off and unplug it from
the electrical outlet. The device has to remain unplugged till the
cleaning completion.
6. Do not disassemble the unit in order to avoid possible electrical
shock. In case of malfunction contact your local dealer.
7. Handle the liquids the device works with carefully, following all the
instructions for their preparation.
8. Place the switching adaptor in such a way as to be protected from
overflow and spillage of liquids.
Ultrasonic milk analyser
PARTS AND ACCESSORIES
In the table below the standard delivery configuration of the milk analyzer is
listed:
№ Description
Item № pcs
1. Ultrasonic portable milk analyzer
LSSA001
1
1 sample measurement time
2. Operation manual
3. Plastic sample holder
4. Spare Pipes
5. 12 V DC Power Supply Cable
90 sec.
60 sec
30 sec
LSSA002
LSSA003
LSSA004
LSSA005
6. Alkaline cleaning solution Lactodaily 100 g 1
7. Acidic cleaning solution Lactoweekly 100 g 1
In the table below the milk analyzer spares and accessories, which are
delivered on customers request are listed:
№ Description
a) included in the set:
Item № pcs
b) not included in the set (may be
additionally bought):
1
2
2
1
/
8. RS232 Interface Cable - Analyser-IBM PC
9. Service Pack - CD
10. Handle
11. pH measuring system
12. pH probe with cable and holder
13. Buffer solution Ph 60 ml
LSSA006
LSSA007
LSSA008
LSSA009
LSSA010
LSSA011
(pH7.00±0.01/20˚C)
14. Buffer solution pH 60 ml
LSSA012
(pH4.00±0.01/20˚C)
15. Milk conductivity measuring system
16. Buffer solution conductivity 50 ml (5.02
LSSA013
LSSA014
(±5%) mS/cm (18±0.1˚C)
17. Real time clock
LSSA015
1
1
1
1
1
1
1
1
Ultrasonic milk analyser
18. High-fat measuring function
19. ECS POS Serial Printer
20. 12 V Serial Printer Power Supply Cable
21. RS232 Interface Cable - Milk Analyser –
Serial Printer
22. Carrying case – plastic
LSSA016
LSSA017
LSSA018
LSSA019
LSSA020
1
1
1
1
1
1 23. Plug type
1
Ultrasonic milk analyser
Ultrasonic milk analyser
2. TECHNICAL PARAMETERS
2.1. Working modes characteristics:
The program of the milk analyzer has four working modes.
2.1.1. Measurement mode milk / dairy product – first type
2.1.2. Measurement mode milk / dairy product – second type
2.1.3. Measurement mode UHT milk / dairy product – third type
These modes have been calibrated on customers’ request for 3 milk types
from the following: cow, sheep, UHT, buffalo, goat, camel milk, cream, ice
cream mixtures, whey, recovered milk, etc. before leaving the production
facilities and the text on the display will be for the corresponding types, as is
indicated on page 2 Measurement modes.
2.1.4. Cleaning
2.1.4.1. Current
2.1.4.2. Final
Ultrasonic milk analyser
2.2. Measuring range:
Fat ………………..................... …………..from 0.01% to 25% (45%*)
SNF ………………………………...…………………....from 3% to 40%
3
Density ** ………………………. ……...…….from 1000 to 1150 kg/m
Proteins …………………………. ………………………..from 2% to 7%
Lactose ………………………... …………………from 0.01 % to 20 %
Water content ………………… ……………...……..from 0 % to 70 %
o
Temperature of milk …………… from 1
C to 40oC (if measurement is
30 sec, then t° is from 15 to 40°)
o
Freezing point ****……………… …………..from – 0,400 to – 0,700
C
Salts ……………………………. …...…………...………from 0,4 to 4%
PH***……………………………...…………………..………from 0 to 14
Conductivity** ………………… ……………...…from 2 to 14 [mS/cm]
* Option, on customers’ request
** Density data are shown in an abbreviated form. For example 27.3 have to
3
be understood as 1027.3 kg/m
. To determine the milk density, write down
the result from the display and add 1000.
3
Example: result 21,20; density = 1000 + 31,20 = 1031,2 kg/m
The abbreviated form of the density is used also when entering data for
samples in working mode Recalibrate, for example:
If the measured sample density is 1034.5 kg/m3, then in the menu for
entering the samples parameters used for calibration, across the parameter
Den = , you have to enter 34.5.
*** pH and conductivity measurements are optional and are embedded in the
device on customers' request.
Accuracy is guaranteed in case of normal ambient conditions:
o
Air temperature…………………………….…from 10
C to 40oC (43 oC)
Relative humidity ……………………………………...from 30% to 80%
Power supply ………………………… …………………...220V (110V)
extent of contamination at normal environmental conditions…………..2
Accuracy values in point 2.3 are in dependence on the correctness of the
corresponding chemical method, used for component content determination.
In point 2.3. are used the following reference methods: Gerber – for fat,
gravimetric – for SNF, Kjeldahl – for protein. The boundary for maximum
variation of repeatability when the power supply voltage is from +10 to – 15%
from the nominal voltage values (220V) have to be no more than 0.8
accuracy according point 2.3. The analyser is used in conditions free of outer
electrical and magnetic fields (except the magnetic field of the Earth) and
vibrations.
2.5. Dimensions: ……………………………..240/220/100 mm, mass 3,0 kg
2.6. Continuous working time:……………………………………… non-stop
2.7 Milk sample volume per one measurement: ……….15 cm3 (= 25 ml)
Ultrasonic milk analyser
fig.1 Front panel
1. Display 6. Input pipe
2. Button “down” 7. Hanger
3. Button “Enter” 8. Sample holder
4. Button “up”
Ultrasonic milk analyser
fig. 2 Back panel
9.12 V Power supply input 15. Serial interface RS232
14. Serial printer connector 20. Switching adapter
Ultrasonic milk analyser
Fig 3. Principle working scheme
Ultrasonic milk analyser
Fig. 4 Connecting peripheral devices
Ultrasonic milk analyser
Fig. 5 Cables description
90-1801-0008
RS232 Interface Cable - Milk Analyser – Serial Printer
90-1801-0010
RS232 Interface Cable - Milk Analyser – IBM PC
2. TxD
3. RxD
5. GND
Milk
Analyser
DB 9-pin
male
PC
DB 9-pin
female
2. Receive Data
(RxD)
3. Transmit Data
(TxD)
5. Signal Ground
(GND)
90-2001-0001
Probe pH with cable
90-1801-0009
DC 12V Power Supply Milk Analyzer Cable
1. GND
2. No connection
3. No connection
4. 12V DC
Ultrasonic milk analyser
3. QUALIFICATION OF RAW MILK, THERMALLY TREATED
MILK, OTHER DAIRY PRODUCTS AND DERIVATIVES
3.1. Taking samples and preparation for analyses
In order to receive reliable results in qualification of milk, dairy products and
derivatives are needed: precise samples taking; correct samples storing (in
need to be preserved); correct preparation before making measurement. The
rules and requirements for this are described in details in Appendix Milk
sampling.
3.2. Making the measurement.
3.2.1. Preparing the analyzer for working mode
3.2.1.1. Put the analyzer on the working place, providing good ventilation and
not in the vicinity of heat providing devices or sources. The temperature in the
premises has to be in the boundaries 10-30
3.2.1.2. Check if the power switch is in "0" position and that the outlet voltage
complies with the voltage indicated on the rating plate of the analyzer.
Connect the switching adapter.
3.2.1.3. Switch on the “POWER” button, which starts the identification
procedure. For a short time the display shows the number of the software
versions, for example:
Milkanalyzer xxx
LCD vers xx
MA vers yy
MA ser. N. xxxx
where:
Milkanalyzer xxx is the time for measurement.
LCD vers XX is display control software version.
MA vers YY is the motherboard software version.
MA ser. N. xxxx is the serial number – written on the rear panel of the
analyzer.
o
C.
Ultrasonic milk analyser
If in the process of exploitation there is a need to ask a question the
company-producer, you have to send the data, written on the display during
the above described initialization procedure.
3.2.1.4.Till the analyzer is prepared for work (at about 5 minutes) the
following message is written on the display: “Getting ready”. Above pointed
time is in dependence of the environmental temperature and increases with
decreasing the temperature.
3.2.1.5. When the device is ready for work the display shows: “Ready to
start”.
The analyzer is ready to make analyses in mode 1 (normally Cow)
3.2.1.6. If you want to pass to another mode press the button Enter and hold
it pressed. The following message appears on the display:
Release button to
start menu
Release the button Enter. The display shows the possible working modes:
Milk selector
Cal1 – Cow
Cal2 – Sheep
Cal3 – UHT
-----------------------------Cleaning
Final cleaning
Using “up”▲and ”down”▼buttons, choose the working mode and press Enter
in order to start it.
3.2.2. Making analyses
To start measurement:
- pour the preliminary prepared sample in the sample holder of the
analyzer;
- put the sample holder in the recess of the analyzer;
- press the button Enter.
Ultrasonic milk analyser
The analyzer sucks the milk, makes the measurement and returns the milk in
the waste liquid reservoir. During the measurement the temperature of the
sample is shown on the display.
Ignore the results received immediately after switching on the analyzer and
after measuring distilled water. Make a second measurement with new
portion of the same sample.
3.2.3. Displaying the results
3.2.3.1. When the measurement is finished, the sample returns in the
sample-holder and the display shows the results. For example:
F= ff.ff - measured FAT in percentage;
S= ss.ss - measured SNF in percentage;
D= dd.dd - measured density in percentage;
P= pp.pp - measured protein in percentage;
L= ll.ll - measured lactose in percentage;
W= ww.ww - measured sample’s added water in percentage;
By pressing the button” Down” ▼ the display shows the second page,
containing the results:
tt.tC - sample’s temperature;
pp.pp - sample’s pH result – if there is a pH probe connected;
-0.fff - measured sample’s freezing point;
0.sss - measured salts values;
Ultrasonic milk analyser
By pressing the buttons “up”▲ and ”down”▼, the operator has the possibility
to pass from one page result to another.
If the device has an embedded option “Conductivity” and "conductivity
measurement" started, the result is shown on the display, showing the basic
results replacing lactose results in the following way:
C=xx.xx
In this case the Lactose result is shown on a new page - Page 3 Results.
xx.xx is the measured milk sample’s conductivity in [mS/cm]. If the results are
outside the limits for this type of sample (see table from the Appendix
Conductivity measurement), the cursor flashes after the letter С, reminding
that the sample is not correct. On the printout it is printed as !!!.
If the conductiity value is outside measuring range (2-14 mS/cm), the
following message appears on the display:
C=OutRg (Out of Range), and on the printout there isn’t any line with
conductivity value.
3.2.3.2. Write down the results in the form. The results remain on the display
till a new measurement is started. If the analyzer is connected to a computer
or a printer, it sends the data to the computer or prints them.
Ultrasonic milk analyser
4. CLEANING THE ANALYZER
This procedure prevents gathering milk fat residues and milk stone on the
sensor. The milk stone consists of milk solids, calcium, iron, sulphates,
magnesium, etc. All these substances form layer on the pipe and sensor’s
walls, which leads to deviations in the measurement results and blocking up
the piping.
The company-producer recommends usage of the chemicals, supplied with
the analyser – alkaline and acidic (Lactodaily and Lactoweekly). You may
order them separately or together with the analyser. Try to use only these
chemicals for cleaning the analyser.
4.1. Automatic cleaning the analyzer
Analyzer's contamination as a result of the irregular cleaning is the basic
reason for inexactness during measurement. In order to be avoid this, in the
milk analyzers with peristaltic pump is embedded automatic cleaning. For this
purpose the analyzer has to be prepared on the following way:
Fig.6 Additional pipe to the input pipe for
94 - additional pipe,
connected towards
input pipe
automatic cleaning
94
Please, pay attention to the additional pipe, connected towards the input pipe.
It has to be long enough to suck all the content in the glass, but to be bent in
order not to impede the pump function.
Ultrasonic milk analyser
The reservoir with the cleaning solution have to be connected towards the
analyser and outgoing pipe towards reservoir for collecting the measured
samples and used cleaning solution, as is shown on the scheme:
Fig.7 Connecting the reservoirs towards the milk analyzer
Pay attention to the
pipes in both reservoirs. The
pipe in the reservoir with the
cleaning solution has to be
well dipped in the cleaning
solution while the pipe in the
reservoir with the already
used samples and cleaning
solution MUST NOT be
dipped in the liquid.
Pay attention to the level of
the cleaning solution in the
reservoir. Add solution when
the level is below 2/3!
1 - cleaning solution input
2 - used samples and used
cleaning solution output
Description of the cleaning procedures:
There are two ways of cleaning the analyzer embedded in it – current and
end (final) cleaning.
4.1.1. Current cleaning.
4.1.1.1. Automatic start
Ultrasonic milk analyser
The current cleaning is made with alkaline solution (for working solution
preparation see 4.2.1.), with which the reservoir for the cleaning solution is
filled in.
It is automatically started, without operator's interference after the set time
intevals elapse
1. 55 min. after switching on the power supply of the analyser,
but idle work;
2. 15 min. after the last measurement of real milk sample.
The display shows the following message and a sound signal is emitted:
Auto clean
started!
Put empty glass
After this the cleaning is started. The display shows:
Cleaning
Please wait
If there is a glass with sample it is completely emptied and then the analyser
automatically fulfills procedure for sucking the cleaning solution and c rinsing
the analyzer's inner system. In order the input metal pipe to be cleaned out
for a short time a cleaning solution is pumped out for a short time in the
already empty glass. It is filled to the middle and then is sucked back and
poured in the reservoir with the used liquids. If there wasn't glass before the
start of the automatic cleaning and if you do not replace it or put an empty
glass then the cleaning solution will be poured in front of the device.
After the cleaning is finished the displays shows the following:
End of cleaning
After 2 seconds the display shows:
Ultrasonic milk analyser
Analyzer ready
Then the analyser is ready for normal measurement.
4.1.1.2. Manual start
The current cleaning may be completed by manual start of the menu
Cleaning. It is used before starting the menu Final Clean. It serves for
cleaning the fats from the measuring tract with alkaline cleaning solution. It is
started using the menu for choosing the working mode of the analyzer. After
the measurement is completes, by continuous pressing the button Enter, the
possible analyser operation modes are shown on the display.
Cow
Sheep
UHT
Cleaning
Final Clean
By choosing Cleaning the current cleaning is started. The display shows the
following:
Auto clean
started!
Put empty glass
In this way the current cleaning is started. The procedure and operator’s
actions are described above.
Do not switch off the device at the end of the working day before the
automatic cleaning procedure is completed. If it is not automatically cleaned
and there is not a possibility to wait starting the automatic cleaning, then start
manually the cleaning procedure with alkaline cleaning solution and if it is
necessary with acidic cleaning solution as it is described below.
Ultrasonic milk analyser
4.1.2. End (final) cleaning.
It is done with acidic cleaning detergent (for working solution preparation see
4.2.2.). It is daily. Serves for cleaning the protein deposits from the measuring
system of the analyzer, which were not removed during work with the alkaline
cleaning solution.
Do not use chemicals, which are not intended for milking equipment and
vessels in dairy industry. Pay particular attention to the acidic cleaning
solution concentration. The higher concentration may damage the measuring
sensor.
Always before the final acidic cleaning rinse the device with alkaline cleaning
solution by manual starting if it was not already automatically cleaned. It is
necessary to be done in order to remove the milk residues which could react
with the acidic cleaning solution.
It is started using the working modes menu. After the measurement is
completed, by pressing and holding pressed the button Enter, the display
shows the possible working modes of the analyser.
Cow
Sheep
UHT
Cleaning
Final Clean
After choosing the final cleaning, the following message appears on the
display:
Put filled with
Water glass
and press Enter
to Continue
Ultrasonic milk analyser
The operator has to put a glass filled with water and to press the button Enter
in order to continue the procedure. Then the following message appears on
the display:
Cleaning
Please wait
Do not miss to clean with water because the residues from the alkaline
cleaning solution could react with the acidic solution, which will lead to gas
and sediment formation.
After rinsing with water, the display shows the following:
Put filled with
Detergent glass
and press Enter
to Continue
The operator has to pour alkaline solution in the glass and to press the button
Enter, in order to continue the cleaning.
Then the following message appears on the display:
Cleaning
Please wait
Follow repeatedly sucking/pouring out the cleaning solution and heating it up.
After the cleaning is finished, the cleaning solution is emitted in the reservoir
with the waste liquids.
For maximum cleaning effect it is recommended the cleaning solution to be
preliminary heated up to 40-50 degrees centigrade.
Ultrasonic milk analyser
Follows rinsing with water:
The operator has to put a glass with water for rinsing the device.
Put filled with
Water glass
and press Enter
to Continue
After rinsing the display shows the following:
End of cleaning
After 2 seconds the following appears on the display:
Analyzer ready
After this the analyzer is ready for normal measurement or to be switched off
from the power supply.
4.2. Preparation of cleaning solution
4.2.1. Preparation of alkaline cleaning solution
Preparation of 1 % alkaline solution of Lactodaily for circulation cleaning in
the milk analyzer:
1. Take the package 100 g concentrated powder chemical Lactodaily
2. Carefully cut the upper end, paying attention not to spill it.
3. In appropriate vessel (for example bucket) pour 1 l water.
4. Add the powder and then again water up to 10 l.
Then follow the instruction for milk analyzer cleaning.
4.2.2. Preparation of acidic cleaning solution
Preparation of 1 % acidic solution of Lactoweekly for circulation cleaning in
the milk analyzer:
1. Take the package 100 g concentrated powder chemical Lactoweekly
2. Carefully cut the upper end, paying attention not to spill it.
Ultrasonic milk analyser
3. In appropriate vessel (for example bucket) pour 1 l water.
4. Add the powder and then again water up to 10 l.
Then follow the instruction for milk analyzer cleaning.
Fig. 8 Labels for acidic cleaning chemical Lactoweekly and alkaline
cleaning chemical Lactodaily
Ultrasonic milk analyser
4.3. Working with the peristaltic pumps
There are two peristaltic pumps in the analyser – 1 for the sample and 1 for
the cleaning.
Fig.9 Peristaltic pumps
Ultrasonic milk analyser
Fig. 10 Peristaltic pump SL 16
Ultrasonic milk analyser
5. POSSIBLE MALFUNCTIONS AND ERROR MESSAGES,
TROUBLESHOOTING
In the table below are described the possible malfunctions during the milk
analyzer’s exploitation and ways for their repair/remedy. If the problem
persists after all recommended measures are taken, please, connect the
nearest service center.
Error
message
MA
overheated
Accompanied
by a
continuous
sound signal
Empty
Camera
Sample
Overheat
Possible
problem /cause
Overheated milk
analyzer
Insufficient
quantity of the
milk sample
sucked in the
system or air in
the sample
Sucked
overheated
sample
Repair/remedy
Immediately switch off the analyzer.
Pay attention the analyzer to be situated away from
direct sunlight or heating devices.
Wait 5-10 minutes the device to cool down or to be
normalized the ambient temperature and switch it on
again.
The analyzer is ready to measure the next sample.
In order to avoid the future appearance of the same
error message, please, check the following:
- The sample is prepared according the
instructions and there aren’t air bubbles in it.
- There is a real suction of the sample after
starting measurement, i.e. it is obvious that the
level of the milk sample in the sample holder
decreases. In other case – there is damage in
the suction system.
- Avoid the end of the suction pipe to be above
the surface of the liquid (not dipped enough).
- Avoid curdling of the milk sample. Clean
immediately if there is a sample curdled in the
system.
- Check the pipe, connecting orifices for cleaning
solution and waste liquid (on the analyzer’s rear
panel.
- In mode Measurement, after starting the
measurement, remove the sample holder and
see if there is no milk poured back in the
sample holder.
The analyzer is ready to measure the next sample.
In order to avoid the future appearance of the same
error message, please, check the following:
-The sample is prepared according the instructions
and its temperature does not exceed the maximum
permissible sample’s temperature.
-Complete the procedure for checking the analyzer
in case of error message Empty Camera.
Loading...