LABEC AA-17BS User Manual

INSTRUCTION MANUAL
Laboratory Equipment Pty Ltd email: sales@labec.com.au Ph: 02 9560 2 811 • Fax: 02 9560 61 3 1
www.labec.com.au
2014
SEA Series Steam Sterilizers
User Registration
On the serial number plate at back of the machine, you may find the unit's model
number, serial number and reference number. Please double check the numbers
and fill in the form, please quote these numbers when you contact your
distributors.
Model :
Serial number:
Reference number:
*Before operating the sterilizer, please read carefully all safety cautions and instructions for operation. This operation manual will help you understand all functions of Runyes Autoclaves as far as possible.
*Please follow carefully instructions in this operator's manual while servicing and maintaining the units.
*Please keep this manual safe for your future references.
*Should error occur during operation of the machine, please get in touch with your local distributor or us for your best qualified services and assistance.
Operation manual
Important information on Safety
SEA Series Steam Sterilizers
Stability
Smother!!!
Heaviness!!!
1. Please check the voltage before plugging in the power cord. Don't use electrical outlets that have other than the correct voltage. Otherwise it might cause fire or electric shock.
2. Don't wet your hand before plugging or unplugging the power cord.
3. Never place your autoclave on an unstable, moving, tilted or shaking surface
4. Never cover or block the door or ventilation openings of the autoclave.
5. Never put heavy items or liquid containers that might spill on the top of the autoclave.
6. Please unplug the power cord when the machine will not be used for a long time.
Operation manual
SEA Series Steam Sterilizers
Contents
Section Page
1.Description and Usage ............................................................ .........1
2.Specifications................................................................... .................2
3.Installation.........................................................................................3
4.Control Panel............ .........................................................................5
5.Operation...........................................................................................6
6 .Door Adjustment..............................................................................10
7.Printer Installation and USB ( optional)............................................11
8.System Setup....................................................................................13
9.Service and Maintainance................................................................17
10.Transport and Storage.....................................................................21
11.Warranty..........................................................................................21
12.Accessories......................................................................................22
13.Appendices........... ..........................................................................23
Working mode chart........................................................................23
Error code chart...............................................................................23
working chart........................................................................................24
circuit connection chart........................................................................25
piping chart...........................................................................................26
14.Trouble shooting..............................................................................27
15.Important Information.....................................................................28
Operation manual
SEA Series Steam Sterilizers
1.Description & Usage
The autoclave is computerized, user friendly and easy to operate. Its real time display shows all data during the sterilization cycle. The software program provides auto diagnostics, protection against over heating or over pressurizing, and thus ensures reliability of the sterilization results. Built-outside used water collecting system prevents discharge of waste steam and keeps the machine clean and safe.
SEA-22L/17L-B-LCD Three Steam sterilizers times pre-vacuum pressurized
SEA-22L/17L-B-LCD uses advanced technology with 3 times pre -vacuum and vacuum drying to enable hot steam to penetrate in thin tubes and porous material. This product is suitable for dental, ophthalmic, surgical instruments, and for theatre and laboratory use to sterilize 135 autoclavable equipment.
Used water tank
Clean water tank
Printer
Door handle
USB port
Used water tank connection
Clean water tank connection
Fig 1-1
Fuse
Power switch
Operation manual
1
SEA Series Steam Sterilizers
2.Specifications
Chamber inner dimensions (Diameter X Depth)..........................................22L:249mm x 450mm
17L:249mm x 355mm
Power supply ..................................................................................According to the nameplate
Sterilizing temperature /pressure Bar(Under standard atmospheric pressure)..121 /(1.0 1.3)Bar
...........................................................................................................................134 /(2.1 2.3)Bar
Cautions
Note: Temperature/Pressure here means the chamber temperature and pressure during the phase
of sterilization. These value vary under different atmospheric pressure.
Power fuse...............................................................................................................................T12A
Water tank capacity.....................................................................................................................3.5L
Operation temperature..............................................................................................+5 +40
Relative humidity range...............................................................................................30% 75%
The maximum wight of instruments which load into the sterilizer chamber during one cycle.
Unwrapped cycle 121
22L/17L: 6.5kg
Unwrapped cycle 134
22L/17L: 6.5kg
wrapped cycle 121
22L/17L: 3.2/2.5kg
wrapped cycle 134
22L/17L: 3.2/2.5kg
Figures
Package select
Start/stop
Water outlet
Test cycle select
Last cycle select
B type device
Program select
Fast sterilizing select
Grounded correctly
Refer to the operation manual
Note of high temperature
Distilled water only
Tip:
Standard atmospheric pressure: It is the atmospheric pressure at sea level. It refers to 1 Standard atmospheric pressure. The value is 0.1Mpa=1000Hpa=1.0 Bar. Atmospheric pressure decreases by higher altitude. For instance: at or lower than 3000 meters , An increase of altitude of 1000 meters will result in a decrease of ATM PR .by 0.1Bar approximately.
Operation manual
2
SEA Series Steam Sterilizers
3.Installation
The Autoclave should be installed in a ventilated place allowing minimum of 10 cm space at all sides and 50 cm at the top. The ventilation openings at the side of the machine should not be blocked. The Autoclave should be placed on a leveled counter top as Fig 3-2. The dimension of the machine are shown in Fig 3-1:
When moving the unit DO NOT DRAG. Lift the antoclave to move.
Fig 3-1
585585
Fig 3-2
**The front of the autoclave must be higher than the rear. **Side to side the autoclave must be level. **Use level supplied to achieve correct alignment see (FIG 3-2)
Caution
If the front of the machine is lower than the back the control will display E3 (E3 means it will not dry completely). Adjust the height by adjusting foot pad of the front of Autoclave.
Operation manual
3
Prepare the machine before use
SEA Series Steam Sterilizers
We have installed the in the chamber(Fig3-3). You should clean the when use the
draining filter
draining filter regularly
autoclave for a long while. The instruction shows you how to clean it. Please refer to P16.
Before using the device, you should make electrical connections by using the power cord that comes with the unit. Insert the hollowed end of the power cord in to inlet of the machine and plug the other end on a power outlet. Caution: The power supply can afford 1800VA load.
Ok, you can turn on the autocalve now.
Fig 3-3
Fig 3-4
Caution
The power supply must be able to afford 1800VA, and grounded correctly. Tips: Before starting the Autoclave, please check the local atmospheric pressure. If the pressure is lower than
0.095 Mpa (or the Altitude is higher than 500 meters), you will need to contact your distributor who should reconfigure the controller for the device.
Operation manual
4
4.Control panel
1.LCD Screen Figure 4-1
Letter row 1 Working state: show temperature and pressure data Menu state: show menu row or page layout name Ready to work state: show sterilizer cycle name
Letter row 2 Working state: show pressure curve Menu state: show menu row Ready to work state: show selected cycle parameter
SEA Series Steam Sterilizers
1
Letter row 3 Working state: show pressure curve Menu state: show menu row Ready to work state: show selected cycle parameter
6 2
Letter row 4 Show help information
2.Menu/ C key
Press the MENU / Cancel key to display the main menu.
3.OK key
After chooseing the correct cycle on Menu state, the key means confirm, the other key means invalidation.
4.DRY key
Fast/Down key. When the "MENU" is displayed this key operates as a Down key It means quick sterilization cycle when the device ready to work. The fast key is used to choose the dry time(1 min or 5 min).This can reduce the whole time of the sterilization cycle.
5. STEP Key
Last cycle/ Up key. When the "MENU" is displayed this key operates as an "Up" key. When selecting a cycle, pressing this key will switch between Keeping Warm and Last Cycle . If the display reads Keeping Warm the chamber will keep warm after the cycle completes. This will reduce the heating time for the next cycle. If the display reads Last Cycle then the chamber will cool automatically when the cycle completes. (If no entry is made the autoclave will enter Last Cycle automatically after one hour.
Fig4-1
5
3
4
6. START / STOP Key
START / STOP key will start sterilization cycle after selecting the sterilization program. When the Menu is displayed, or the autoclave is operating, it acts as a 'Cancel' key.
Caution
In case of faults, please contact your distributor or the manufacture immediately.
Operation manual
5
5.Operation
SEA Series Steam Sterilizers
5.1 Turn on the autoclave
Please connect the power cord before you start the operation. Turn on the autoclave via the ON / OFF switch located behind the small door at the bottom right of the front of the machine. The power switch light turns on indicating that electric power is connected. 1 s later, enter to ready to work state after automatically examine. The screen will display as (Fig 5-1) show.
RUNYES
-Created by love
Please Sel. a cycle
Fig5-1
5.2 Filling Water / Refilling
Turn on the power switch, select cycle and press OK key. If the water level is lower than the minimum required, the display will show the refilling page layout (See Fig 5-2) with alarm sounding. You should add water. Open the tank cover at the top of autoclave and remove the filler filter. Fill tank with distilled water to cover the GREEN rod.
134-B-4Min-10Min
Notice
Caution
Refilling
Fig5-2
1. Distilled water will prolong the life of the autoclave.
2. Do not tilt the autoclave when there is water in the water reservoir.
The used water tank should be emptied frequently. You should empty the used water reservoir when the clean water reservoir is refilled via the used water outlet located below the ON/OFF switch
Operation manual
6