LABEC AA-17BS User Manual

INSTRUCTION MANUAL
Laboratory Equipment Pty Ltd email: sales@labec.com.au Ph: 02 9560 2 811 • Fax: 02 9560 61 3 1
www.labec.com.au
2014
SEA Series Steam Sterilizers
User Registration
On the serial number plate at back of the machine, you may find the unit's model
number, serial number and reference number. Please double check the numbers
and fill in the form, please quote these numbers when you contact your
distributors.
Model :
Serial number:
Reference number:
*Before operating the sterilizer, please read carefully all safety cautions and instructions for operation. This operation manual will help you understand all functions of Runyes Autoclaves as far as possible.
*Please follow carefully instructions in this operator's manual while servicing and maintaining the units.
*Please keep this manual safe for your future references.
*Should error occur during operation of the machine, please get in touch with your local distributor or us for your best qualified services and assistance.
Operation manual
Important information on Safety
SEA Series Steam Sterilizers
Stability
Smother!!!
Heaviness!!!
1. Please check the voltage before plugging in the power cord. Don't use electrical outlets that have other than the correct voltage. Otherwise it might cause fire or electric shock.
2. Don't wet your hand before plugging or unplugging the power cord.
3. Never place your autoclave on an unstable, moving, tilted or shaking surface
4. Never cover or block the door or ventilation openings of the autoclave.
5. Never put heavy items or liquid containers that might spill on the top of the autoclave.
6. Please unplug the power cord when the machine will not be used for a long time.
Operation manual
SEA Series Steam Sterilizers
Contents
Section Page
1.Description and Usage ............................................................ .........1
2.Specifications................................................................... .................2
3.Installation.........................................................................................3
4.Control Panel............ .........................................................................5
5.Operation...........................................................................................6
6 .Door Adjustment..............................................................................10
7.Printer Installation and USB ( optional)............................................11
8.System Setup....................................................................................13
9.Service and Maintainance................................................................17
10.Transport and Storage.....................................................................21
11.Warranty..........................................................................................21
12.Accessories......................................................................................22
13.Appendices........... ..........................................................................23
Working mode chart........................................................................23
Error code chart...............................................................................23
working chart........................................................................................24
circuit connection chart........................................................................25
piping chart...........................................................................................26
14.Trouble shooting..............................................................................27
15.Important Information.....................................................................28
Operation manual
SEA Series Steam Sterilizers
1.Description & Usage
The autoclave is computerized, user friendly and easy to operate. Its real time display shows all data during the sterilization cycle. The software program provides auto diagnostics, protection against over heating or over pressurizing, and thus ensures reliability of the sterilization results. Built-outside used water collecting system prevents discharge of waste steam and keeps the machine clean and safe.
SEA-22L/17L-B-LCD Three Steam sterilizers times pre-vacuum pressurized
SEA-22L/17L-B-LCD uses advanced technology with 3 times pre -vacuum and vacuum drying to enable hot steam to penetrate in thin tubes and porous material. This product is suitable for dental, ophthalmic, surgical instruments, and for theatre and laboratory use to sterilize 135 autoclavable equipment.
Used water tank
Clean water tank
Printer
Door handle
USB port
Used water tank connection
Clean water tank connection
Fig 1-1
Fuse
Power switch
Operation manual
1
SEA Series Steam Sterilizers
2.Specifications
Chamber inner dimensions (Diameter X Depth)..........................................22L:249mm x 450mm
17L:249mm x 355mm
Power supply ..................................................................................According to the nameplate
Sterilizing temperature /pressure Bar(Under standard atmospheric pressure)..121 /(1.0 1.3)Bar
...........................................................................................................................134 /(2.1 2.3)Bar
Cautions
Note: Temperature/Pressure here means the chamber temperature and pressure during the phase
of sterilization. These value vary under different atmospheric pressure.
Power fuse...............................................................................................................................T12A
Water tank capacity.....................................................................................................................3.5L
Operation temperature..............................................................................................+5 +40
Relative humidity range...............................................................................................30% 75%
The maximum wight of instruments which load into the sterilizer chamber during one cycle.
Unwrapped cycle 121
22L/17L: 6.5kg
Unwrapped cycle 134
22L/17L: 6.5kg
wrapped cycle 121
22L/17L: 3.2/2.5kg
wrapped cycle 134
22L/17L: 3.2/2.5kg
Figures
Package select
Start/stop
Water outlet
Test cycle select
Last cycle select
B type device
Program select
Fast sterilizing select
Grounded correctly
Refer to the operation manual
Note of high temperature
Distilled water only
Tip:
Standard atmospheric pressure: It is the atmospheric pressure at sea level. It refers to 1 Standard atmospheric pressure. The value is 0.1Mpa=1000Hpa=1.0 Bar. Atmospheric pressure decreases by higher altitude. For instance: at or lower than 3000 meters , An increase of altitude of 1000 meters will result in a decrease of ATM PR .by 0.1Bar approximately.
Operation manual
2
SEA Series Steam Sterilizers
3.Installation
The Autoclave should be installed in a ventilated place allowing minimum of 10 cm space at all sides and 50 cm at the top. The ventilation openings at the side of the machine should not be blocked. The Autoclave should be placed on a leveled counter top as Fig 3-2. The dimension of the machine are shown in Fig 3-1:
When moving the unit DO NOT DRAG. Lift the antoclave to move.
Fig 3-1
585585
Fig 3-2
**The front of the autoclave must be higher than the rear. **Side to side the autoclave must be level. **Use level supplied to achieve correct alignment see (FIG 3-2)
Caution
If the front of the machine is lower than the back the control will display E3 (E3 means it will not dry completely). Adjust the height by adjusting foot pad of the front of Autoclave.
Operation manual
3
Prepare the machine before use
SEA Series Steam Sterilizers
We have installed the in the chamber(Fig3-3). You should clean the when use the
draining filter
draining filter regularly
autoclave for a long while. The instruction shows you how to clean it. Please refer to P16.
Before using the device, you should make electrical connections by using the power cord that comes with the unit. Insert the hollowed end of the power cord in to inlet of the machine and plug the other end on a power outlet. Caution: The power supply can afford 1800VA load.
Ok, you can turn on the autocalve now.
Fig 3-3
Fig 3-4
Caution
The power supply must be able to afford 1800VA, and grounded correctly. Tips: Before starting the Autoclave, please check the local atmospheric pressure. If the pressure is lower than
0.095 Mpa (or the Altitude is higher than 500 meters), you will need to contact your distributor who should reconfigure the controller for the device.
Operation manual
4
4.Control panel
1.LCD Screen Figure 4-1
Letter row 1 Working state: show temperature and pressure data Menu state: show menu row or page layout name Ready to work state: show sterilizer cycle name
Letter row 2 Working state: show pressure curve Menu state: show menu row Ready to work state: show selected cycle parameter
SEA Series Steam Sterilizers
1
Letter row 3 Working state: show pressure curve Menu state: show menu row Ready to work state: show selected cycle parameter
6 2
Letter row 4 Show help information
2.Menu/ C key
Press the MENU / Cancel key to display the main menu.
3.OK key
After chooseing the correct cycle on Menu state, the key means confirm, the other key means invalidation.
4.DRY key
Fast/Down key. When the "MENU" is displayed this key operates as a Down key It means quick sterilization cycle when the device ready to work. The fast key is used to choose the dry time(1 min or 5 min).This can reduce the whole time of the sterilization cycle.
5. STEP Key
Last cycle/ Up key. When the "MENU" is displayed this key operates as an "Up" key. When selecting a cycle, pressing this key will switch between Keeping Warm and Last Cycle . If the display reads Keeping Warm the chamber will keep warm after the cycle completes. This will reduce the heating time for the next cycle. If the display reads Last Cycle then the chamber will cool automatically when the cycle completes. (If no entry is made the autoclave will enter Last Cycle automatically after one hour.
Fig4-1
5
3
4
6. START / STOP Key
START / STOP key will start sterilization cycle after selecting the sterilization program. When the Menu is displayed, or the autoclave is operating, it acts as a 'Cancel' key.
Caution
In case of faults, please contact your distributor or the manufacture immediately.
Operation manual
5
5.Operation
SEA Series Steam Sterilizers
5.1 Turn on the autoclave
Please connect the power cord before you start the operation. Turn on the autoclave via the ON / OFF switch located behind the small door at the bottom right of the front of the machine. The power switch light turns on indicating that electric power is connected. 1 s later, enter to ready to work state after automatically examine. The screen will display as (Fig 5-1) show.
RUNYES
-Created by love
Please Sel. a cycle
Fig5-1
5.2 Filling Water / Refilling
Turn on the power switch, select cycle and press OK key. If the water level is lower than the minimum required, the display will show the refilling page layout (See Fig 5-2) with alarm sounding. You should add water. Open the tank cover at the top of autoclave and remove the filler filter. Fill tank with distilled water to cover the GREEN rod.
134-B-4Min-10Min
Notice
Caution
Refilling
Fig5-2
1. Distilled water will prolong the life of the autoclave.
2. Do not tilt the autoclave when there is water in the water reservoir.
The used water tank should be emptied frequently. You should empty the used water reservoir when the clean water reservoir is refilled via the used water outlet located below the ON/OFF switch
Operation manual
6
5.3 Select cycle
Press MENU/C key to display cycle options. There are 7 cycles displayed in three page layouts (Fig 5-2). Press STEP key and QUICK key to select cycle. Press OK key to confirm selection. You need to confirm cycle choice by pressing OK otherwise "Invalidation" will be displayed.
CYCLES
Universal B cycle: 134 Three times pre-vacuum, 4 min sterilization time
18 min B cycle: 134 Three times pre-vacuum, 18 min sterilization time. Suitable for instruments that need a long sterilization time. It can achieve the best sterilization effect.
SEA Series Steam Sterilizers
UNIVERSAL B Cycle
KEEP WARMING
Please close door
Fig5-2
SELECT A CYCLE:
UNIVERSIAL B CYCLE
18 MINUTES B CYCLE
UNWRAPPED CYCLE
Unwrapped cycle: 134 one time pre-vacuum, 4 min sterilization time.
Rubber & Plastic cycle: 121 Three times pre-vacuum, 16 min sterilization time.
B & D Testing cycle: BOWIE & DICK, Leak Test.
Vacuum test cycle: It is used to test the chamber's sealing property during the vacuum test cycle.
Cleaning steam generator: Clean the steam generator and steam pipe. Refer to Page 19 for specific operation.
SELECT A CYCLE:
RUBB & PLAS.CYCLE
B&D TEST CYCLE
LEAK TEST CYCLE
SELECT A CYCLE:
CLEAN GENERATDR
Fig5-2
Operation manual
7
5.4 Start Selected Cycle
Close the door after loading the autoclave and press START key to begin the cycle. After 0.5 seconds the electrical lock will engage. If the electrical lock did not close correctly the autoclave will display a door fault. The sterilization cycle will only start after the door is locked successfully (See fig 5-4).
5.5 Stopping a Cycle
To abort a cycle during sterilization press the Start/Stop key for 3 seconds. There are two different exit modes:
1.If you press the button for 3 seconds during Pre-heating Phase, Pre-vacuum Phase or Sterilization Phase it will enter into an Exhaust Phase. After 3 minutes drying the whole cycle will end. If you need to end the cycle immediately press the Start/Stop button for 3 seconds again.
2.Press the Start/Stop button for 3 seconds during drying Phase to end the cycle immediately.
SEA Series Steam Sterilizers
Fig5-4
5.6 Cycle end
After sterilization the display will show as per fig 5-5 and an alarm will sound. The door can now be opened and the sterilized
0.05 Bar 83.5 C
The end
instruments unloaded.
Caution: If you decide to stop using the autoclave for a while, please
Open door
turn the main power switch off, or unplug the power cord should you not want to use the machine for a long time.
Fig5-5
Caution
To ensure correct operation of the autoclave:
1.When the Water Level Low alarm sounds refill the unit with distilled water.
2.When the autoclave is refilled the waste water tank should be emptied via the outlet below the On/Off switch (See section 9.2, fig-9.4).
3.Always use the tray holder to load or unload the tray with instruments in it in order to avoid burning injuries.
4.Never try to open the chamber door if and when the pressure display doesn't show 0.0 .
5.It is suggested that chemical indicator or bio-indicator be used for each and every cycle to ensure the reliability of each sterilization cycle run.
Operation manual
8
SEA Series Steam Sterilizers
5.7 The Instrument Rack
The instrument rack should be put into the chamber as the following diagrams show. The rack can hold 3 trays or be rotated clockwise by 90 to hold five trays
Three Tray Configuration
Fig1
The second way (Five Tray Configuration)
Fig3 Fig4
Notes:
Fig2
The above picture are only for reference, due to the different type or equipment renewal, instrument plate and tray will a little different, please taken the accessories as authentic without prior notice.
Operation manual
9
SEA Series Steam Sterilizers
6.Door Adjustment
Under normal circumstance the chamber door lock needs no readjustment.If steam starts leaking (the seal fails), you may use the spanner to turn the door seal tightener.
6.l Open the door first.
6.2 Insert the spanner in the gap beneath the plastic cover. Use the spanner to lock on the adjusting nut (Fig 6-2). Turn the nut counter clockwise as the figure below (Fig 6-1). This will tighten the sealing plate.
6.3 Turn the nut until the sealing plate is tight. If the door knob is too tight, you may also turn the nut clockwise to loosen it as long as the seal does not leak.
Caution
Never try to readjust the chamber door while the door is locked.when you adjust the door DO NOT touch the black sealing cover.
Tight
Loose
Fig 6-1 Fig 6-2
Operation manual
10
SEA Series Steam Sterilizers
7.Printer installation and USB( ) optional
If required a printer can be connected to the Runyes autoclave. Connect the printer to the autoclave with cables according to fig 7-1. Switch on the power and two lights on the printer panel will glow indicating the printer has been installed correctly. See section 8.1 for printer ser up.
Sterilization
Records
Date:06.04.20 C.N.:134 Wrap
-B-:18m 10m C.S.T:23:15:16 Vp1: 23:21:55
0.0 035.7 Pp1: 23:23:57
-.6 033.4 Pp2: 23:23:57
0.1 098.4 Pp2: 23:34:09
-.5 081.7 Pp3: 23:34:09
Fig7-1
7.1Print information
Date: Date Cycle: Cycle Times C.N Cycle name C.S.T Cycle start time Vp1 Vp2 Vp31 Vacuum Pp1 Pp2 Pp31 Pressure strengthen H.S.T: Sterilization start time Max Min Ave Max. Min and average pressure or temperature of chamber H.E.T: Sterilization end time D.S.T Drying cycle start time C.E.T Cycle end time Result: Result Operator: Operator, you can sign your name if it needed.
0.1 104.4 Pp3: 23:41:51
-.6 079.9 H.S.T.:23:50:24 Min:2.1 135.5 Min:2.0 134.6 Ave:2.0 135.0 H.E.T.:00:08:27 D.S.T .:00:10:44
0.2 109.2 C.E.T.:00:10:45 Result:Passed Operator:
Thanks
Fig 7-2
Caution
Runyes sterilizer only can select printer offered by Runyes at present.
Operation manual
11
SEA Series Steam Sterilizers
7.2 USB module function
7.2.1Main function: Storage of printed data of sterilization Built-in FLASH memorizer U-disk. Download
7.2.2Basic parameters: Storage Capacity: 2M bytes Quantity of printed data could be stored: 2048 Support FAT files
7.2.3Operation instruction:
1.Turn on the power, the mode of USB will start automatically. After self-test, it goes into alert with the indicator flickering-1 time.
2.After the sterilization cycle ends, the data of sterilizer is sent to the USB mode(The printer program should be turned on). Indicator will flicker-1 time after the data-transfer finish.
3.During the alert, insert U-disk. The mode indicator will be on, with the U-disk indicator flickering. And now, it is in the process of data-download. Don't turn the power off or draw out the USB-disk, which will cause the data lose.
4.After the download finish, USB mode indicator will flicker twice, reminding the user to draw out the U-disk.
5.During the sterilization cycle, avoid data download, or else, the data will lose.
6.Connect U-disk to computer. You will see files like _2000_, where the downloaded data is kept, following turns of Year, Month and Day. Easy to print it out.
7.USB mode terror(According to the flickering times of USB mode indicator) and solution:
Times
1
2
3
4
5
6
7
Caution: This function is optional to the demand of customer.
Self-test passed
Operation finish
No definition now.
U-disk needs formation
Not enough capacity
U-disk needs formation
USB mode error
Definition
Solution
Normal
Normal
No definition now.
Format U-disk in FAT files
Delete the files in U-disk
Format U-disk in FAT files
Contact us
Operation manual
12
8.System setup
Press MENU/C key for 3 seconds on ready state and the autoclave will enter into system setup (See Fig 8 -1).
Press STEP key and DRY key to select and press OK key to enter sub-menu for setup system. Press MENU/C key to cancel and return to main menu.
8.1 Date setup
With 3 sub-menus (Fig 8 - 2). Press STEP and DRY key to select. Press OK key enter this sub-menu. Press STEP and DRY key to setup the valve. Press OK key to save the setup and return to upper level of the menu. Press MENU/C key to return to upper level of menu without saving setup.
SEA Series Steam Sterilizers
DATE TIME PRINTER LANGUAGE
Fig8-1
DATE SETUP
YEAR
MONTH
DAY
Fig8-2
8.2 Time setup
With 2 sub-menus (Fig 8 -3). Press STEP and DRY to select sub-menu. Press OK key to enter this sub-menu. Press STEP and dry key to insert value. Press OK key to save the setup and return to upper level of the menu. Press MENU/C key to return to upper level of menu without saving setup.
8.3 Printer setup
Press STEP and DRY key to select ON or OFF (See Fig 8-4) Press OK key to save the setup and return to upper level of the menu. Press MENU/C key to return to upper level of menu without saving setup.
TIME SETUP
HOUR MINUTE
Fig8-3
PRINTER SETUP
ON OFF
Fig8-4
Operation manual
13
8.4 Language setup
With 2 sub-menus (fig 8-5). Press STEP and DRY key to select language. Press OK key to save the setup and return to upper level of the menu. Press MENU/C key to return to upper level of menu without saving setup. At present, we have the below language: Chinese, English, French, German, Spanish, Turkish, Malaysian, Italian, Swedish, Romanian, Polish, latvian(the others is on developing).
8.5 Time Setup
To setup or extend the sterilization time ranging from 0~3 minutes.
SEA Series Steam Sterilizers
LANGUAGE SETUP
CHINESE(S) ENGLISH
Fig8-5
time setup
00
8.6 Temperature Setup
To correct the disparity between displayed temperature and real one, ranging from -2 ~ +3. Advise to setup this value under technician's instruction.
Note
1.At the end of the cycle the autoclave displays Fig 8 - 8 and the alarm sounds. Open the door and unload the sterilized instruments
Always use the tray holder to load or unload the tray with instruments in it in order to avoid burning injuries.
Fig8-6
temperature setup
00
Fig8-7
0.05 Bar 83.5
The end
Open door
Fig8-8
C
Operation manual
14
Other Functions (Not available yet)
SEA Series Steam Sterilizers
Operation manual
15
SEA Series Steam Sterilizers
Operation manual
16
9.Service and Maintainance
SEA Series Steam Sterilizers
Caution:
*Before any kind of maintenance turn off the power to the autoclave and allow it to cool down. *Use only distilled water for sterilizing and cleaning. *The instruments being sterilized must be cleaned. Instruments with oil or impurities must be packed in a sterilization bag.
9.1 Clean sterilizer water tank and used water tank
Please clean both tanks regularly using the following process:
1. Empty the tanks (please refer to item 9.2).
2. Open the tank cover board. Use a screw driver to loosen the five screws as per Fig 9-1
3. Pull up the water tank cover board as per Fig 9-2
4. Lift up the tank cover completely, you can then disassemble the tank cover as per Fig 9-3
5. Clean the tank. Use cotton dipped in alcohol or medical disinfectant to scour the tank wall. Wash it with distilled water then empty and dry the tank.
6. Take out and clean the filter (see next page for procedure).
7. After you clean the tank and filter, reassemble the tank cover and tighten the screws.
Fig9-1
Fig9-2
9.2 Draining of water from the machine
Drain the condenser water collector and water reservoir. Connect the end of the tube without a connector to the water outlet and the other end of the tube to a drainage sink. Turn the draining knob anti-clockwise to drain the water from the machine (See Fig 9-4).
Fig 9-4
Fig9-3
Operation manual
17
SEA Series Steam Sterilizers
9.3 Clean the drain filter
Over time small impurities will deposit on the drain filter, and this moy affect vacuum and drying. In order to prolong the life of the drain filter please use the following cleaning procedure (see fig 9-5):
1.Locate the water filter.
2.Rotate off the filter screw
3.Pull off the rubber filter from the connection
4.Take off the filter from the blue rubber.
5.Clean the filter or replace with a new one.
6.After you clean or change a new filter shim, plug in the water filter back on the chamber connection.
Note: please turn the blue rubber to face down hard to touch the chamber and tighten the screw nut.
-5Fig9
Note
Please check the O ring of Drainning Filter during cleaning. If this part is broken please replace it with O ring in the accessories supplied.
9.4 Change fuse (Fig 9-6)
-6Fig9
Caution
Unplug the autoclave before changing the fuse
1.Unscrew the fuse holder with a screw driver counter clockwise.
2.Make sure to replace the faulty fuse with a correct fuse.
3.Put back the fuse holder and screw it on clockwise with a screw driver.
Operation manual
18
SEA Series Steam Sterilizers
9.5. Clean the seal plate
Over time the seal plate will have some scale build-up that may cause leaking. You should clean the seal plate regularly using a soft cloth soaked with distilled water to wipe the rubber seal and the seal plate clean (Fig 9-7 & 9-8). If there is still leaking, you may need to take the rubber seal out for cleaning. If the seal is broken you will need to replace it with a new one. Every 2000 cycles, we suggest to clean the seal plate
Fig 9-7 Fig 9-8
9.6 Change the seal
You will need a flat screw driver without a sharp tip.
1.Hold the edge of the seal with one hand and insert the screw driver beneath the seal with your other hand and then slowly pull out the seal.
2.Once a section of the seal is out, you may pull the whole seal out slowly with both hands. You can now clean the seal groove and the seal. If the seal is broken or damaged replace it with a new one.
3.After the groove and the seal are cleaned remount the seal.
Caution: The seal should be mounted evenly inside the groove with first 4 positioning points in place. Press the seal in place evenly. Attention: When inserting the seal the inner edge might protrude. You may use a screw driver to press the edge down in the groove.
Fig 9-9
Caution
Switch off machine and let it cool down before you change the seal to avoid any burning injuries.
Operation manual
19
9.7 Cleaning the Steam Generator
1. Purpose: To clean the steam generator and steam pipes, to prevent furring blocking the pipe.
2. Cycle: 20 min cleaning + 10 min drying
3. Turn on the autoclave via the ON / OFF switch. The screen will display the welcome state.
4. Cleaning Cycle Operation: Press the DRY key 6 times until CLEAN GENERATOR is displayed. Press OK.
SEA Series Steam Sterilizers
SELECT A CYCLE CLEAN GENERATOR
Fig 9-10
CLEAN GENERATOR
Please close door
5. Close the door and press START. The autoclave will now start the cleaning cycle.
Fig9-11
6. Cleaning Reminder: Every 100 cycles the B&D light and LEAK light will go on to remind you to clean the steam generator.
Note:
Cleaning function will not affect the autoclave cycle, only a remind message.
NOTICE
CLEAN REQ.
Fig9-12
9.8 Repair and change of parts must be done by qualified service engineer from the manufacturer or an authorized contractor.
9.9 If the machine needs to be maintained or the electric spare parts need to be changed please cut off the power first and ask your dealer or manufacturer to change it.
Operation manual
20
SEA Series Steam Sterilizers
10. Transportation & Storage
Power the machine off, unplug the power cord and allow the chamber to cool. The autoclave should be transported and stored in the following conditions: Temperature: -40 ~ +55 Humidity: 75% Atmospheric pressure: 500Hpa ~ 1060 Hpa
Caution
When moving the unit DO NOT DRAG. Lift the machine to move.
11. Warranty
The manufacturer provides a warranty for repair at not change to the customer for two years after the invoice date with presentation of the invoice.
Followings are the disclaimers:
1. Damage due to improper moving, installing and operating the machine;
2. Damage due to dismantling the machine by unauthorized personal;
3. Inappropriate operation on the machine without following the operator's manual
4. Damage caused by power surge, fire or other external matters.
Operation manual
21
12. Accessories
1
2
SEA Series Steam Sterilizers
3 4
5 6
1). Instrument tray rack 1 pc and Instrument tray 3 pc .
2). Power cord. 1 pc.
3). Mainboard Fuse 1 pc.
4). Water outlet tube 1 pc.
5). Tray holder 1 pc.
s
10
11
6). Filtrable net 1 pc.
7). Sealing ring 1 pc.
8). Fuse 2 pc .
9). Draining O ring 2 pc .
10). Level instrument 1 pc.
11). Door adjustment spanner 1 pc.
s
s
7 8 9
Note
1. The picture of spare parts in the attachment may be a little different to the parts supplied. Runyes keep the right of change the accessories. Please check the spare parts supplied with autoclave.
2. During disposing the obsolete device , please contact with professional recycle company Don't throw them away.
Operation manual
22
13.Appendix.
Form 1 Work mode chart
SEA Series Steam Sterilizers
B cycle
Sterilizing temperature
Sterilizing pressure
Class
Pre-heating
Vaeuuning time
Admission time
Sterilizing time
Drying time
Quick drying time
Total time
Note:
The figures in the chart are only for reference, the sterilization effect verified by user, and the user can
134
2.10 2.10 2.10 1.10 2.10
B B
0-7min 0-7min 0-7min 0-7min 0-7min
10min 10min 10min 10min
5min 5min 5min 5min 5min
4min
10min 10min 10min 10min
5min 5min 5min 5min
36(31)min 50(45)min 30(25)min 48(43)min
18min Bcycle
134 134
18min
Unwrappd
cycle
S
4min
RUB & PLAS
cycle
121
B B
16min
B&Dtest
3.5min
1min
19.5min 15min
134
Vacuum test
4min 4min
Keep pressure 10min
Clean test
Clean
Drying
30min
choose the sterilization time according to the verification results. Which cycle should be adopted is according to the type of the instruments.
Form 2 Error code chart(In case of faults, please contact your distributor or the immediately)
20 min
10 min
Item
1
2
3
4
5
6
7
8
9
Error
sensor error
super pressure >2.6Bar
super temperature >145 c
Constant temperature fails
block of steam pipe (20S<0.3Mba)
failure of the door
Steam generator over temperature (>230 C)
heating overtime (>25 min)
pressurization overtime (>60 min)
o
Solution
examine sensor
Examine inner temperature sensor or pressure sensor
examine outer temperature sensor
Return parameter autoadjus programm
Examine the steam output valve or vacuum valve
examine the door switch
Check steam generator temperature sensor
examine the electronic connection of heat ring and heat stick
examine the steam input pipe
Operation manual
23
( Bar)( Bar)
2.02.0
1.01.0
0.50.5
PressurePressure
00
SEA Series Steam Sterilizers
Working chart
134 sterilization cycle working chart
TimeTime
-1.0-1.0
11 22 33 44 55 77
1-2 Pre-heating 2-3 Pre-vacuum 3-4 Sterilizing 4-5 Exhausting 5-6 Drying 6-7 Balance period 1-7 Sterilizing cycle
66
121 seerilization cycle working chart
( Bar)( Bar)
2.02.0
1.01.0
PressurePressure
-1.0-1.0
00
11 2 2
33
44 55 77
66
TimeTime
1-2 Pre-heating 2-3 Pre-vacuum 3-4 Sterilizing 4-5 Exhausting 5-6 Drying 6-7 Balance period 1-7 Sterilizing cycle
Operation manual
24
Circuit connection chart
SEA Series Steam Sterilizers
Operation manual
25
Piping
SEA Series Steam Sterilizers
Condense pipe
Condense
Inner temperature sensor
Vacuum pump
two position three pass valve (vacuum valve)
Pressure sensor
Vent steam valve
Normally open valve
Vent water valve
Normally closed valve
Steam generator
Steam pressure lock
Safety valve
Fill air solenoid valve Normally closed valve
Chamber
Air filter
Steam valve
Normally
closed valve
Let in air pump (water pump)
manual water filling exit
Used water tank
Used water drain connection
Clean water tank
Filter net
drain connection
Operation manual
26
SEA Series Steam Sterilizers
14.Trouble shooting
1.During venting or drying if a thump sounds it means that the Condensed water collector is full. You must drain the water from the tank(see section 9.2 , Fig 9-4).
2.During a cycle if a banging sounds, steam leaks from the chamber door and the displays show an E6 code. Please verify whether the door knob is properly turned and locked. If the steam leaks at the beginning of a cycle, you will need to change the seal according to instructions at section 9.6.
3. At the end of a cycle if, there are water drops on the instruments. You can correct the problems by:
1).Checking the drain filter is not blocked (see section 9.2).
2). Reduceing the load of the instrument in the chamber.( especially the porous and dressing and other water absorbent materials.)
3). Waiting for a while at the end of a cycle Ed for better drying result. The autoclave will continue heating the chamber until you open the door.
Operation manual
27
SEA Series Steam Sterilizers
15.Important information
1.The autoclave must be installed on a leveled surface.
2.When operated at an elevation of more than 500 meters the machine must be reprogrammed. Please get in touch with your distributor for instructions.
3.Distilled water must be used for prolonging the life of the machine.
4.The ventilation openings in the machine must not be blocked or covered.
5.Instruments should be placed on tray with space between one another allowing steam to pass through.
6.Condensed water collector must be emptied from time to time.It should be emptied when refilling the water reservoir.(Draining of the collector can be followed as per item 9.2)
7
.Pls closing the door well while you use it.
8.Never try to open the chamber door if and when the pressure display doesn't show 0.0 .
9.Before any kind of maintenance turn off the power to the autoclave and allow it to cool down.
10.DO NOT drag the autoclave when it is being moved.
11.Power connection must be grounded properly.
12.Must be provided enough power (22L:2300W; 17L:1800W).
13.If the ambient temperature is low please pre-heat the autoclave without instruments for 30 minutes before sterilizing.
14.DO NOT power off the unit during sterilization,. To abort a cycle during sterilization press the Start/Stop key for 3 seconds.
15. The chart in the operation manual only for reference, which keep the actual product for real.
Operation manual
28