LABEC AA-12BS User Manual

INSTRUCTION MANUAL
Laboratory Equipment Pty Ltd email: sales@labec.com.au Ph: 02 9560 2 811 • Fax: 02 9560 61 3 1
www.labec.com.au
2014
Series Steam Sterilizers
User Registration
On the inscription plate at back of the machine, you may find the unit's
model number, serial and reference number. Please double check the
number and fill in the form with them and quote the reference number,
when you are in contact with you distributors.
Model :
Serial number:
Reference number:
*Before operating the sterilizer, please read carefully all safetycautions and instructions of operation. This operator's manual will help you understand all functions of Autoclaves as far as possible.
*Please follow carefully instructions in this operator's manual While servicing and maintaining the units.
*Please keep this manual safe for your future references.
*Shoud error occur during operation of the machine, please get in touch with your local distributor or us for our best qualified services and assistance.
Operation manual
Important Information on Safety
Please plug the power in a correct voltage( according to nameplate) electrical socket. Don't use electrical outlets other than the correct voltage; otherwise it might cause fire or electric shock.
Don't wet your hands before plugging or unplugging the power cord.
Series Steam Sterilizers
Don't place your autoclave on an unstable table top, moving desk, tilted or shaking surface.
Don't cover or block the door, ventilation or radiation openings on the autoclave.
Don't put heavy stuff or liquid containers that might spill on the top of the machine.
Please unplug the power cord when the machine isn't used for a long while or when you travel, etc.
Operation manual
Series Steam Sterilizers
Contents
1. Description & Usage..............................................................................................................1
2. Specifications............................................................................................................................2
3. Installation.................................................................................................................................3
4. Control Panel of 12L............................................................................................ ..........5
Operation of ........
5. 12L.................................................................................................... .......7
6. Door Adjustment....................................................................................................................12
. Printer
7 User's Guide................................................................................................................13
Function setup
8. ..........................................................................................................................14
Maintenance & Maintain ......................... . .15
9. ........................................................................ ... ..
10. & Storage .
Transportation .......................................................................................... ........18
11. Warranty ..................................18
12. Accessorise ..........................
Appendix.......................................................................................................................................20
Work mode form
Error code chart...........................................................................................................................21
Cycle Chart & Schematics of 12L
................................................................................................
............................................................................................. ......19
.........................................................................................................................20
........................................................................................23
.......
Pipe diagram of 12L
Troubleshooting
Important information
...........................................................................................................................25
.................................................................................................................24
................................................................................................................26
Operation manual
Series Steam Sterilizers
1. Description & Usage
Design and mark pressurizd steam autoclaves especially to meet the needs for sterilization of instruments and devices in clinics,departments in a hospital and laboratories. Operators include doctors or his/her assistance under doctors supervisions.The autoclave is computerized, user friendly and easy to operate. Its real time display show all data during the sterilization cycle. The software program provides auto diagnose, protection against over heating or over pressurizing,and thus ensure reliability of the sterilization results. Build- in condenser water collecting system prevents discharge of waste steam and keeps the machine
clean and safe.
12L Three times pre-vacuum pressurized steam Autoclave
12L uses the advanced technology of 3 times fractionated -vacuum and Vacuum drying. It enables the hot steam to penetrate in thin tube and porous Material and laboratories to sterilize dental and surgical instruments, dressing and other 135 autoclavable stuffs.
Water reservoir
Used water tank
Printer
Door knob
Fuse
Fig 1-1
Used water tank outlet
Water reservoir outlet
Power switch
Operation manual
1
Series Steam Sterilizers
2.Specifications
Chamber inner dimensions: (Diameter x Depth) ---------------------------------200Mm x 358mm Power Supply: -------------------------------------------------------------According to the nameplate Sterilizing Temperature/Pressure --121 /(1.0-1.3) Bar (Under standard atmospheric pressure)
------------------------------------------134 /(2.0-2.4) Bar(Under standard Atmospheric pressure)
Note: Temperature/Pressure here means the chamber temperature and pressure during the phase of sterilization. The value varies under different atmospheric pressure.
Power Fuse ------------------------------------------------------------------------------------------T12A Water reservoir capacity -----------------------------------------------------------------------3.5 liters
Environment temperature---------------------------------------------------------------+5 +40
Relative Humidity Range------------------------------------------------------------------30% 75%
Figures
Tips Standard atmospheric pressure: It the atmospheric pressure at sea level. It refersto 1 Standard atmospheric pressure. The value is 0.1Mpa=1000Hpa=1.0 Bar. Atmospheric pressure decreases by higher altitude, for instance; at 3000 meter or lower incremental 1000 meter higher the altitude, the atmospheric pressure decreases
by about 0.1bar.
Package Key
Start/Stop key
Select Key
Dry Key
Test Cycle Key
Used Water Outlet
Cycle Step Key
Caution & Warning
Type B Device
Grounding
Operation manual
2
Series Steam Sterilizers
3.Installation
The Autoclave should be installed in a drafty place allowing minimum 10 cm space at all sides and 20 cm at the top. The radiation openings at the side of the machine should not be blocked. The Autoclave should be placed on a leveled counter top.(Fig 3-1)
439
369
Fig 3-1
575
Fig 3-2
Caution:
If the front of the machine is lower than the back Fault E3 (E3 means it don't dry completely)might occur. You may level up the front to correct this Fault.
Operation manual
3
Prepare the machine before use
We have installed filtration equipment in the chamber(Fig3-3).You should clean the filtration equipent when use the autoclavefor a long time The instruction how to clean please refer to P16.
Before using the device, you should make electrical connections by using the power cord that comes with the unit. Insert the hollowed end of the power cord in to inlet of the machine and plug the other end on a power outlet. Caution: The power supply can afford 1800VA load. Ok, you can turn on the autocalve now.
Series Steam Sterilizers
Fig 3-3
Caution: The power supply must afford at least 1800VA load.
Tips: Before using the Autoclave, please check on the local atmospheric pressure. If the pressure is lower than 0.095 Mpa or the Altitude is higher than 500 meters, you will need to contact your distributor who should reconfigure the controller in the device.
Fig 3-4
Operation manual
4
4. Control Panel of 12L
Control Panel of 12L (Fig 4-1).
1 Pressure Display
It displays chamber relative pressure in real time
during a cycle. The unit is per Bar.
2 Temperature
It displays chamber temperature during a cycle.
The unit is per Degrees Centigrade
Series Steam Sterilizers
1
2
3
3 State display/Error Codes
It displays all the states of autoclave. ( Look at the state form) Error Codes when alarm goes on. You may correct the error accordingly by looking it up
4
in the page of error codes and its corrections.
Caution: In case of faults, please contact your
5
distributor or the manufacturer immediately.
6
4 SELECT key
Program choose key. It is used for choosing the temperature. You can choose the right program according to the temperature that instrument can afford. You can also choose B program, it can
+
Fig 4-1
sterilize medical cotton.
5 PACKAGE key
Package choose key. It is used for choosing two types mode of package. According to fact, you can choose wrapped or unwrapped. Non-package use 3 times vacuumize, pakage use B-standard vacuumize.
8
7
10 11
9
12
6 Test key
It is used for choosing the test program. If you push the button the indication light of B&D goes on, and it means that you have chosen this testing program when you push again, the indication light of Leak goes on, it means that you have chosen this program.
Operation manual