On the inscription plate at back of the machine, you may find the unit's
model number, serial and reference number. Please double check the
number and fill in the form with them and quote the reference number,
when you are in contact with you distributors.
Model :
Serial number:
Reference number:
*Before operating the sterilizer, please read carefully all safetycautions and instructions of
operation. This operator's manual will help you understand all functions of Autoclaves
as far as possible.
*Please follow carefully instructions in this operator's manual While servicing and maintaining
the units.
*Please keep this manual safe for your future references.
*Shoud error occur during operation of the machine, please get in touch with your local
distributor or us for our best qualified services and assistance.
Operation manual
Important Information on Safety
Please plug the power in a correct voltage( according to
nameplate) electrical socket. Don't use electrical outlets
other than the correct voltage; otherwise it might cause
fire or electric shock.
Don't wet your hands before plugging or unplugging
the power cord.
Series Steam Sterilizers
Don't place your autoclave on an unstable table top,
moving desk, tilted or shaking surface.
Don't cover or block the door, ventilation or
radiation openings on the autoclave.
Don't put heavy stuff or liquid containers that might
spill on the top of the machine.
Please unplug the power cord when the machine isn't
used for a long while or when you travel, etc.
Design and mark pressurizd steam autoclaves especially to meet the needs for sterilization of
instruments and devices in clinics,departments in a hospital and laboratories.
Operators include doctors or his/her assistance under doctors supervisions.The autoclave is
computerized, user friendly and easy to operate. Its real time display show all data during the
sterilization cycle. The software program provides auto diagnose, protection against over
heating or over pressurizing,and thus ensure reliability of the sterilization results. Build- in
condenser water collecting system prevents discharge of waste steam and keeps the machine
clean and safe.
12L Three times pre-vacuum pressurized steam Autoclave
12L uses the advanced technology of 3 times fractionated -vacuum and Vacuum drying.
It enables the hot steam to penetrate in thin tube and porous Material and laboratories to
sterilize dental and surgical instruments, dressing and other 135 autoclavable stuffs.
Water reservoir
Used water tank
Printer
Door knob
Fuse
Fig 1-1
Used water tank outlet
Water reservoir outlet
Power switch
Operation manual
1
Series Steam Sterilizers
2.Specifications
Chamber inner dimensions: (Diameter x Depth) ---------------------------------200Mm x 358mm
Power Supply: -------------------------------------------------------------According to the nameplate
Sterilizing Temperature/Pressure --121/(1.0-1.3) Bar (Under standard atmospheric pressure)
------------------------------------------134/(2.0-2.4) Bar(Under standard Atmospheric pressure)
Note: Temperature/Pressure here means the chamber temperature and pressure during the phase
of sterilization. The value varies under different atmospheric pressure.
Power Fuse ------------------------------------------------------------------------------------------T12A
Water reservoir capacity -----------------------------------------------------------------------3.5 liters
Tips Standard atmospheric pressure: It the atmospheric pressure at sea level. It
refersto 1 Standard atmospheric pressure. The value is 0.1Mpa=1000Hpa=1.0 Bar.
Atmospheric pressure decreases by higher altitude, for instance; at 3000 meter or
lower incremental 1000 meter higher the altitude, the atmospheric pressure decreases
by about 0.1bar.
Package Key
Start/Stop key
Select Key
Dry Key
Test Cycle Key
Used Water Outlet
Cycle Step Key
Caution & Warning
Type B Device
Grounding
Operation manual
2
Series Steam Sterilizers
3.Installation
The Autoclave should be installed in a drafty place allowing minimum 10 cm space at
all sides and 20 cm at the top. The radiation openings at the side of the machine should
not be blocked. The Autoclave should be placed on a leveled counter top.(Fig 3-1)
439
369
Fig 3-1
575
Fig 3-2
Caution:
If the front of the machine is lower than the back FaultE3(E3 means it don't dry
completely)might occur. You may level up the front to correct this Fault.
Operation manual
3
Prepare the machine before use
We have installed filtration equipment in the
chamber(Fig3-3).You should clean the filtration
equipent when use the autoclavefor a long time
The instruction how to clean please refer to P16.
Before using the device, you should make electrical
connections by using the power cord that comes with
the unit. Insert the hollowed end of the power cord in
to inlet of the machine and plug the other end on a
power outlet. Caution: The power supply can afford
1800VA load. Ok, you can turn on the autocalve now.
Series Steam Sterilizers
Fig 3-3
Caution: The power supply must afford at least 1800VA load.
Tips: Before using the Autoclave, please check on the local atmospheric
pressure. If the pressure is lower than 0.095 Mpa or the Altitude is higher
than 500 meters, you will need to contact your distributor who should
reconfigure the controller in the device.
Fig 3-4
Operation manual
4
4. Control Panel of 12L
Control Panel of 12L (Fig 4-1).
1Pressure Display
It displays chamber relative pressure in real time
during a cycle. The unit is per Bar.
2Temperature
It displays chamber temperature during a cycle.
The unit is per Degrees Centigrade
Series Steam Sterilizers
1
2
3
3State display/Error Codes
It displays all the states of autoclave. ( Look at the
state form) Error Codes when alarm goes on. You
may correct the error accordingly by looking it up
4
in the page of error codes and its corrections.
Caution: In case of faults, please contact your
5
distributor or the manufacturer immediately.
6
4SELECT key
Program choose key. It is used for choosing the
temperature. You can choose the right program
according to the temperature that instrument can
afford. You can also choose B program, it can
+
Fig 4-1
sterilize medical cotton.
5PACKAGEkey
Package choose key. It is used for choosing two types mode of package. According to fact,
you can choose wrapped or unwrapped. Non-package use 3 times vacuumize, pakage use
B-standard vacuumize.
8
7
10
11
9
12
6Testkey
It is used for choosing the test program. If you push the button the indication light of B&D
goes on, and it means that you have chosen this testing program when you push again, the
indication light of Leak goes on, it means that you have chosen this program.
Operation manual
5