Labcyte Echo 650, Echo 655, Echo 655T, Echo 650T Quick Start Manual

LABCYTE INC.
SITE PREP GUIDE | Echo® 650 Series Liquid Handler
Echo® 650 Series Liquid Handler (Models: 650, 655, 650T, 655T)
SITE PREP GUIDE
Server Version 3.1 | FEBRUARY 2019
This guide provides site preparation information for the Echo® 650 Series Liquid Handler. This includes models 650, 655, 650T and 655T. The topics below describe the site preparation requirements and instrument specifications for the Echo Liquid Handlers.
Pre-Installation Requirements
l Site Requirements l User Supplied Equipment Requirements l Communication Requirement s l Accommodating Your Echo Liquid Handler l Noise Ratings l Power Ratings
What You Will Be Receiving
l Shipping Dimensions l Echo Liquid Handler Components l Chiller l Junair Compressor (Optional and sold separately by Labcyte) l Vacuum Pump (Optional and sold separately by Labcyte) l Client PC (Sold separately by Labcyte)
Pre-Installation Requirements Verification Form
© 2019 LABCYTE INC. All rights reserved.
QSG-2019-1
FEB 2019
Operating environment
Temperature: 21°C ± 5°C (70°F± 9°F)
Relative humidity: 10-80%, non-condensing for the Echo Liquid Handler and 35-80% for the SMC Chiller
ACPower (2 options)
Option 1: 100-120 V, 50/60 Hz, > 10A
Option 2: 200-240 V, 50/60 Hz, > 5A
A minimum of two electrical sockets within 183 cm (6 feet) of the Echo Liquid Handler supporting a minimum of 800 W are required.
The Echo Liquid Handler and SMC chiller maximum power consumption is 600W and 500W respectively.
Up to four additional electrical sockets are required for the client PC, monitor, vacuum pump (optional) and air compressor (optional).
Compressed air source
Clean, filtered, oil-free, dry air at 552 kPa (80 PSI) minimum, 1034 kPa (150 PSI) maximum
Maximum consumption by Echo Liquid Ha ndler does not exceed 1 cfm at 150 PSI
SITE PREP GUIDE | Echo® 650 Series Liquid Handler PN | 0 01-16480

Pre-Installation Requirements

The pre-installation requirement s are described in the topics below:
l Site Requirements l User Supplied Equipment Requirements l Communication Requirement s l Accommodating Your Echo Liquid Handler l Noise Ratings l Power Ratings

Site Requirements

The following site requirements must be met before the Echo Liquid Handler can be installed:
Note: The Echo Liquid Handler power configuration is factory set and requires a Field Service visit in order to be changed on site.
Note: The Junair OF302-4B oil-less compressor can be purchased from Labcyte (Part #001-12164). For more information, see the Junair Compressor (Optional and sold separately by Labcyte) topic.
2
Vacuum source for source plate dryer (2 options)
Option 1: House supply. Minimum 200 Torr (266 mbar, 22 in Hg Vac) minimum pumping speed
8.0 m3/hr
Option 2: Standalone vacuum with surge tank (10 L). 100 Torr (133 mbar, 26 in Hg Vac), minimum pumping speed 2.0 m3/hr
Instrument mounting surface (minimal requirements)
The Echo Liquid Handler can emit vibration while in operation and is susceptible to vibration from other devices. For this reason, the Echo system should be placed on a surface that is isolated from surrounding equipment or designed to suppress the transmission of vibration from devices. The surface must be sturdy, level, and water-resistant with cross-bracing or welded joints.
The surface must be ca pable of supporting a minimum of 142 kg (313 lbs) weight with sufficient surface area to accommodate the Echo Liquid Handler in the recommended operating envelope (see Echo Liquid Handler Dimensions). If the mounting surface is a table top, the table legs should not rest on ca sters. If the table has casters, they should be locked.
If the chiller will be installed on the bench surface, it must accommodate an additional 14 kg (31 lbs).
Destination plate samples
At installation, ensure common destination plates to be used with the instrument are available for tuning the destination plate gripper.
Coupling Fluid
Use only ASTM D-1193 Type 1 water in the fluidics system
Coupling Fluid Temperature: 22.1°C ± 0.9°C (71.8°F± 1.6°F)
Hardware
Compressed air fitting compatible with 0.170 in ID / 0.250 in OD tubing between the compressed air supply a nd the Echo Liquid Handler tubing (Labcyte provides the compressed air line and fitting with the Echo Liquid Handler)
Vacuum fitting that connects 3/8 in ID, 5/8 in OD tubing to either house vacuum or a vacuum pump. The supplied vacuum line is HercoBraid, 3/8 in ID, 5/8 in OD (RyanHerco P/N 0512-072)
Client PC
Minimum Requirements: Microsoft® Windows® 7 using 32- or 64-bit OS, or Microsoft® Windows® 10 64-bit; Intel® Core™ 2 Duo processor or later; 1GB of RAM or greater; 1 GB of a vailable disk storage; display resolution of 1280 x 1024 (1920x1020 preferred); 2 x Gigabit ethernet ports
SITE PREP GUIDE | Echo® 650 Series Liquid Handler PN | 0 01-16480
Note: A standalone Vacuubrand MZ 2S-NT vacuum system can be purchased from Labcyte (Part #001-8104). For more information, see the Vacuum Pump (Optional and sold separately by Labcyte) topic.
Note: The 10 L surge tank comes standard with the Echo Liquid Handler to ensure consistent vacuum.

User Supplied Equipment Requirements

The user must provide the following items:
Note: Client PC can be purchased directly from Labcyte (Part #ECHO-Client Computer). For information on the Client PC available from Labcyte, see the Client PC (Sold separately by Labcyte) topic.
3
SITE PREP GUIDE | Echo® 650 Series Liquid Handler PN | 0 01-16480
Note: If you are integrating the Echo Liquid Handler into an automated system and need to install the Client + Server software onto a workstation, it must cont ain an Intel Core™ i5 processor or better.
Note: If you are setting up an Echo 650T or 655T Liquid Handler that will be using tubes, a capper/decapper is required before Labcyte Service and Support can install or service the instrument.
4
Server connection Echo Liquid Handling Platform (Server Version 3.1.x)
Network connection 10/100/1000 BaseT
Network protocol TCP/IP
User account
Labcyte (The client PC must have an account, with administrator privileges, that is accessible by Labcyte Service a nd Support to service the Echo Liquid Handler system.)
The Echo Liquid Handler Dimensions
53.8 cm width x 74.6 cm depth x 91.5 cm height (21.2 x 29.3 x 36.0 inches)
Echo Liquid Handler working envelope (minimal installation requirements)
Top: It is recommended to have 25.4 cm (10 in) above the instrument to remove the top cover for servicing.
Front: It is recommended to have 40.6 cm (16 in) to a llow sufficient clearance to load and unload microplates by hand or by robot.
Side: It is recommended to have 121.9 cm (48 in) minimum on left and right sides to allow for serviceability. The Echo Liquid Handler should have sufficient space to be rotated, otherwise
it will need to be removable for service.
Rear: It is recommended to have 30.5 cm (12 in) minimum behind the instrument to allow for cables, hoses, tubing, and maintenance access. The coupling fluid bottle and optional caddy fit within the rear clearance but can be relocated if necessary.
SMCChiller Dimensions
30 cm plus fittings width x 34 cm Depth x 18 cm Height (11.8 in x 13.4 in x 36.4 in). It is recommended that the chiller be installed within 121.9 cm (48 inches) of the Echo Liquid Handler.
Echo Client PC Additional space for the client PC, monitor, keyboard a nd mouse.
Other
Space must be alloca ted for:
l Any robotics in the system l Workspace for microplates
SITE PREP GUIDE | Echo® 650 Series Liquid Handler PN | 0 01-16480

Communication Requirements

The user environment must support the following communication requirements:

Accommodating Your Echo Liquid Handler

The user must set aside the space to accommodate Echo Liquid Handler installation.
Note: The instrument can be installed with a minimum of 2.5 cm (1 in) at the top if there is sufficient space to lift the top and slide it all the way forward. Talk to your Labcyte service representative for additional details.
5
Noise Rating 64 dBA
Power Rating 600W without the chiller
Heat Rating 600W without the chiller
SITE PREP GUIDE | Echo® 650 Series Liquid Handler PN | 0 01-16480
Figure 1: Instrument Layout (Front View, Top View, Side View)

Noise Ratings

The noise ratings for the Echo Liquid Handler are below:

Power Ratings

The power ratings for the Echo Liquid Handler are below:
6
Instrument shipping crate weight 278 kg (577 lbs)
Instrument shipping crate dimensions
114.3 cm width x 99.1 cm depth x 124.8 cm height (45 x 39 x 49.1 inches)
Accessory crates
Additional devices and/or consumables purchased together with the Echo Liquid Handler may ship in the crate with the Echo Liquid Handler if there is sufficient room or separately.
Vacuum line
HercoBraid, 9.5 mm (3/8 in) ID, 9.5 mm (5/8 in) OD (RyanHerco P/N 0512-072)
HFCD 17612 coupling body (Colder Products)
Vacuum Surge Tank Assembly
(Optional) Usage dependent on configuration, but included with the Echo (P/N 001-5150)
Compressed air line
4.6 meters (15 ft) of polyurethane tubing; 6.35 mm (0.25 in) OD, 4.24 mm (0.170 in) ID (SMC Pneumatics P/N TIUB07-BU)
LCD 13004 coupling body (Colder Products).
Coupling fluid bottle 1 L (P/N 001-5438)
Accessory kit Includes tubing a nd power cord for connecting the chiller (P/N 001-15272)
SMC chiller P/N 001-11149
Ethernet cable Crossover, 3 meters (10 ft)
Software Echo Client Utility installer provided on CD or USB drive
Voltage Frequency Power Fuse Part Number
100-120 V 50 Hz 8 A 250 VAC, 10 A, FAST 001-3966
200–240 V 50 Hz 5 A 250 VAC, 5 A, FAST 001-3967
Table 1: Electrical Configurations for the Echo Liquid Handler
SITE PREP GUIDE | Echo® 650 Series Liquid Handler PN | 0 01-16480

What You Will Be Receiving

Shipping Dimensions

The Echo Liquid Handler System is shipped in a crate with t he following dimensions:

Echo Liquid Handler Components

The following components are included with the Echo Liquid Handler:

AC Power, Configuration, and Fuses

The following table describes the AC Power and Fuse requirements for the Echo Liquid Handler:
Note: The power cord may need to be replaced with one conforming to local AC power receptacles. To determine which AC power configuration you have, see t he Power and Fuse Rating Label on the back panel of the Echo Liquid Handler.
7
Power Supply
The SMC chiller uses an auto switching power supply capable of running any single-phase AC voltage in the range 100-240V, 50-60 Hz
A power cord suitable for the install location is provided
Maximum power is 500W
Temperature Control
The SMC chiller controls the temperature to +/- 0.02 °C by cooling and heating as appropriate
The SMC chiller is air cooled
Ambient Operating Conditions
Temperature: 10-35 °C
Relative Humidity: 35-80%
SITE PREP GUIDE | Echo® 650 Series Liquid Handler PN | 0 01-16480

Chiller

The SMC chiller (P/N 001-11149) is included with your Echo system. The SMC chiller utilizes a stainless steel thermoelectric (Peltier) heat exchanger that regulates fluid temperature. The SMC chiller ships configured for operation with Echo Liquid Handlers:
Figure 2: SMCChiller
8
Voltage V 100 100 120 200 200 230 230 3x400 3x440
Frequency Hz 50 60 60 50 60 50 60 50 60
Motor
HP 0.60 0.60 0.60 0.60 0.60 0.60 0.60 0.60 0.60
kW 0.44 0.44 0.44 0.44 0.44 0.44 0.44 0.44 0.44
Displacement
I/min 108 138 138 108 138 108 138 108 138
CFM 3.81 4.87 4.87 3.81 4.87 3.81 4.87 3.81 4.87
FAD @ 8 bar
l/min 38 44 44 38 44 38 44 38 44
CFM 1.34 1.55 1.55 1.34 1.55 1.34 1.55 1.34 1.55
Maximum pressure
bar 8 8 8 8 8 8 8 8 8
psi 120 120 120 120 120 120 120 120 120
Maximum current A 7.5 8.7 6.6 4.1 4.6 3.4 4.1 1.4 1.4
Tank size
liters 4 4 4 4 4 4 4 4 4
gallon 1.1 1.1 1.1 1.1 1.1 1.1 1.1 1.1 1.1
Weight
kg 22 22 22 22 22 22 22 22 22
lbs 49 49 49 49 49 49 49 49 49
Noise level
dB (A)/lm
65 66 66 65 66 65 66 65 66
Thermal protection
Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes
Duty cycle 100% 100% 100% 100% 100% 100% 100% 100% 100% 100%
SITE PREP GUIDE | Echo® 650 Series Liquid Handler PN | 0 01-16480

Junair Compressor (Optional and sold separately by Labcyte)

The Echo Liquid Handler requires a compressed air source that provides clean, filtered, oil-free, dry air at 552 kPa (80 PSI) minimum and 1034 kPa (150 PSI) maximum. The Junair OF302-4B oil-less compressor can be purchased from Labcyte (Part # varies dependent on geography.)
The user must also provide a fitting between the compressed air supply and the Echo Liquid Handler. Labcyte provides the compressed air line and fitting with the Echo Liquid Handler, but the customer must provide a source compatible with 170 in ID/.250 in OD tubing.
The table below lists the specifications for the Junair OF302-4B compressor:
Note: Maximum consumption by Echo Liquid Handler for instrument pneumatics does not exceed 1 cfm at 150 PSI.
9
SITE PREP GUIDE | Echo® 650 Series Liquid Handler PN | 0 01-16480
Figure 3: Junair OF302-4B oil-less compressor
10
House supply
Minimum 200 Torr (266 mbar, 22 in Hg Vac) minimum pumping speed 8.0 m3/hr
Standalone vacuum with surge tank (10 L) 100 Torr (133 mbar, 26 in Hg Vac), minimum pumping speed 2.0 m3/hr
Permissible a mbient temperature storage/operation
14 to 140/50 - 104 F°
-10 to +60/ +10 - +40 C°
Permissible relative atmospheric moisture during operation (no condensation)
30 to 85%
No-load speed 50/60 hZ 1500/1800 rpm
Device fuse Slow blow fuse 6.3 A
Maximum pumping speed 50/60 Hz
1.3/1.4 cfm
2.2/2.4 m3/h
Ultimate vacuum (absolute)
5.2 Torr 7 mbar
Maximum permissible outlet pressure (absolute)
29 psi 2 bar
Maximum permissible different between inlet and outlet
29 psi 2 bar
Rated motor power
0.24 hp
0.18 kW
Maximum permissible range of supply voltage (+10%)
100-115 V ~50/60 Hz 120 V ~ 60 Hz 230 V ~50/60 Hz
Dual voltage motor
100-115 V ~50/60 Hz 120 V ~ 60 Hz 200- 230 V ~50/60 Hz
Maximum rated current at: 100-115 V ~50/60 Hz, 120 V ~ 60 Hz 200- 230 V ~50/60 Hz, 230 V ~50/60 Hz
3 A
1.8 A
Inlet
Hose nozzle for tubing I.D. 3/8 in Hose nozzle DN 10 mm
Outlet Silencer / G1/4 in
Dimensions L X W X H (approximate)
9.6 X 9.4 X 7.8 in 243 X 239 X 198 mm
SITE PREP GUIDE | Echo® 650 Series Liquid Handler PN | 0 01-16480

Vacuum Pump (Optional and sold separately by Labcyte)

The Echo Liquid Handler requires a vacuum source. The vacuum dries source plates before they are unloaded. This requirement can be met through a house vacuum or a vacuum pump.
The vacuum source should have the following specifications:
Customers must provide a fitting that connects the vacuum source and the Echo Liquid Handler vacuum line. The supplied vacuum line is HercoBraid, 9.5 mm (3/8 in) ID, 9.5 mm (5/8 in) OD (RyanHerco P/N 0512-072).
A standalone Vacuubrand MZ 2S-NT vacuum system can be purchased from Labcyte (Part #001-8104). This vacuum is a dual voltage/cycle compatible system.
The table below lists the specifications for the Vacuubrand MZ 2S-NT vacuum system:
11
Weight (approximate)
24.2 lbs 11 kg
Operating system Microsoft® Windows 7 32-bit OS
CPU Intel® Core™ XE2 i5-4570S 2.9 GHz Processor
Memory 4GB RAM or greater
Hard drive storage 1 TB hard drive
Video Resolution 1280 x 1024 resolution
Ethernet 2 Gigabit 10/100/1000 (Two Ethernet Cards)
SITE PREP GUIDE | Echo® 650 Series Liquid Handler PN | 0 01-16480
Figure 4: Vacuubrand MZ 2S-NT vacuum system

Client PC (Sold separately by Labcyte)

Client PC specifications (Dell OptiPlex XE2):
Included Accessories:
l Microsoft® Office 2016 l Monitor, 22 in, DVI-D/DP/VGA l Dell USB Keyboard, Black l Dell MS111 USB Optical Mouse, OptiPlex and Fixed Precision l Labcyte Mousepad
Software specifications:
Optionally, the following Echo software applications are available for purchase:
l Echo Plate Reformat l Echo Dose-Response l Echo Cherry Pick l Echo Array Maker l Echo Plate Audit l Echo Combination Screen
Note: Echo 650 Series Liquid Handlers require Echo software applications version 1.7 or higher.
12
Verified Requirement Description Yes No
Operating environment
Are the temperature and relative humidity operating environment requirements met?
Sturdy, level, water-resistant bench top
Is there a sturdy, level, water-resistant bench available for the Echo 6XX series instrument?
Operating space envelope
Does the space a vailable for the instrument meet the operating space envelope requirement?
AC power
Is AC power meeting the AC power specifications and located within 6 feet of the installation location?
Compressed air source
Is there a source of compressed air meeting the compressed air source specification?
Vacuum source Is there a vacuum source meeting the vacuum specification?
Client PC
Is there a client PC available for the installation meeting the client PC specification?
Destination plates
Are commonly used destination plates available for tuning the destination plate gripper at installation?
LABCYTE INC.
SITE PREP GUIDE | Echo® 650 Series Liquid Handler

Pre-Installation Requirements Verification Form

Please send this Pre-Installation Requirement Verification Form to the local Labcyte representative assigned to the installation. Verification of the pre-installation requirements is required before the service representative will schedule the installation.
One installation visit is provided with the purchase of an Echo 650 Series Liquid Handler instrument. If the pre-installation requirements are not met at the time of installation, it may delay the installation and require an additional visit.
If you have any questions regarding the lab environment setup or required equipment, please contact your local Labcyte Representative or email support@labcyte.com.
Echo® Liquid H andlers, Applications Software and Automation SystemsfromLabcyte Inc. are c overed by one or more of the following patents:
United States: 6,416,164; 6,548,308; 6,603,118;6,612,686;6,642,061; 6,666,541; 6,707,038; 6,710,335; 6,746,104; 6,802,593; 6,808,934; 6,809,315; 6,849,423; 6,855,925;6,869,551;
6,893,836; 6,893,115; 6,916,083; 6,932,097; 6,938,987; 6,938,995; 6,991,917; 7,070,260; 7,090,333; 7,185,969; 7,270,986; 7,354,141; 7,405,072; 7,405,395; 7,439,048; 7,454,958;
7,481,511; 7,717,544; 7,899,645; 7,900,505; 7,901,039;8,107,319, 8,389,295; 8,503,266; 8,770,69; 9,212,250; 9,586,215; 9,908,133; 10,112,212;10,118,186. EuropeanPatent EP: 1322430;
1324823;1337325; 1352112; 1366356; 1348116, 1409254, 1534526; 1585636; 1879697.Japan: 4189964; 4309131; 4434581; 4559218; 4624644;4990476; 4955901; 4990476; 5031875.
Additional patents are approvedand pendingin the UnitedStates and other countries.
All product names and brands are properties of their respective owners.
© 2019 Labcyte Inc. All rights reserved. Labcyte®, Echo®, MicroClime®, the Labcyte logo, andAccess™ are
registeredtrademarks ortrademarks ofLabcyte Inc., in the U.S. and/or other c ountries.
FOR RESEARCH USE ONLY. Not for use in diagnostic procedures.
LABCYTE INC.
170 Rose Orchard Way
San Jose, CA 95134, USA
Toll-free: + 1 877 742−6548
Fax:+ 1 408 747−2010
SUPPORT
North America
Europe
Japan
Asia
Other
+1 408 747−2000
|
+44 1543 276543
|
+81 03 5530 8964
|
+61 39018 5780
|
+1 408 747−2000
|
support@labcyte.com
|
support@labcyte.com
|
japan@labcyte.com
|
asiapacific@labcyte.com
|
support@labcyte.com
|