Labconco provides a warranty on all parts and factory workmanship. The warranty includes areas of
defective material and workmanship, provided such defect results from normal and proper use of the
equipment. Glassware is not warranted from breakage when dropped or mishandled.
The warranty for FreeZone® Freeze Dry Systems will expire one year from date of installation or two
years from date of shipment from Labconco, whichever is sooner. Warranty is non-transferable and
only applies to the owner (organization) of record.
This limited warranty covers parts and labor, but not transportation and insurance charges. In the
event of a warranty claim, contact Labconco Corporation or the dealer who sold you the product. If
the cause is determined to be a manufacturing fault, the dealer or Labconco Corporation will repair or
replace all defective parts to restore the unit to operation. Under no circumstances shall Labconco
Corporation be liable for indirect, consequential, or special damages of any kind. This statement may
be altered by a specific published amendment. No individual has authorization to alter the provisions
of this warranty policy or its amendments. Lamps and filters are not covered by this warranty.
Damage due to corrosion or accidental breakage is not covered.
The information contained in this manual and the accompanying products are copyrighted and all rights
reserved by Labconco Corporation. Labconco Corporation reserves the right to make periodic design
changes without obligation to notify any person or entity of such change.
Returned or Damaged Goods
Do not return goods without the prior authorization from Labconco. Unauthorized returns will not be
accepted. If your shipment was damaged in transit, you must file a claim directly with the freight carrier.
Labconco Corporation and its dealers are not responsible for shipping damages.
The United States Interstate Commerce Commission rules require that claims be filed with the delivery
carrier within fifteen (15) days of delivery.
Limitation of Liability
The disposal and/or emission of substances used in connection with this equipment may be governed by
various federal, state, or local regulations. All users of this equipment are required to become familiar with
any regulations that apply in the user’s area concerning the dumping of waste materials in or upon water,
land, or air and to comply with such regulations. Labconco Corporation is held harmless with respect to
user’s compliance with such regulations.
Contacting Labconco Corporation
If you have questions that are not addressed in this manual, or if you need technical assistance, contact
Labconco’s Customer Service Department or Labconco’s Product Service Department at 1-800-821-5525
or 1-816-333-8811, between the hours of 7:00 a.m. and 6:00 p.m., Central Standard Time.
Part #7426100, Rev. H
ECO J323
TABLE OF CONTENTS
CHAPTER 1: INTRODUCTION 1
Freeze Dry Process 1
Freeze Dry Rates 2
Freeze Dry Capacity 3
Samples Containing Volatile Substances 4
About This Manual 4
Typographical Conventions 5
Your Next Step 5
Congratulations on your purchase of a Labconco FreeZone® Freeze Dry System,
which is designed for laboratory lyophilization procedures. The refrigerant used in
the refrigeration system is CFC-free so it will not endanger the environment. The
unit is easy to install and maintain. Proper care and maintenance of this product
will result in many years of dependable service.
Freeze Dry Process
Freeze drying is an important process in sample preparation and for the preservation
and storage of biologicals, pharmaceuticals and foods. Of the various methods of
dehydration, freeze drying (lyophilization) is especially suited for substances that are
heat sensitive. Other than food processing (e.g., coffee, whole dinners), freeze drying
has been extensively used in the development of pharmaceuticals (e.g., antibiotics) and
preservation of biologicals (e.g., proteins, plasma, viruses and cell lines). The
nondestructive nature of this process has been demonstrated by the retention of
viability in freeze dried viruses and microorganisms.
Freeze drying is a process whereby water or other solvent is removed from frozen
material by converting the frozen water directly into vapor without the
intermediate formation of liquid water. The basis for this sublimation process
involves the absorption of heat by the frozen sample in order to vaporize the ice;
the use of a vacuum pump to enhance the removal of water vapor from the surface
of the sample; the transfer of water vapor to a collector; and the removal of heat by
the collector in order to condense the water vapor. In essence, the freeze dry
process is a balance between the heat absorbed by the sample to vaporize the
ice and the heat removed from the collector to convert the water vapor into ice.
Product Service 1-800-522-7658
Chapter 1: Introduction
2
SAFE TEMPERATURE AND DRYING TIMES
FOR SELECTED MATERIALS
Material
10mm Thick
Safe
Temperature
°C
Collector
Temperature
°C
Hours
(Approx.)
Milk
-5
-40
10
Urea
-7
-40
10
Blood Plasma
-10 to -25
-40
16
Serum
-25
-40
18
Vaccinia
-30 to -40
-50
22
Influenza Vaccine
-30
-50
24
Human Tissue
-30 to -40
-50
48
Vegetable Tissue
-50
-80
60
Table 1
Freeze Dry Rates
The efficiency of the Freeze Drying process is dependent upon the surface area and
the thickness of the sample, the collector temperature and vacuum obtained, the
eutectic point and solute concentration of the sample. It is important to remember
these factors when trying to obtain efficient utilization of your Freeze Dry system.
A listing of selected materials and their approximate drying times are shown in
Table 1 for your reference.
*Total sample quantities are contingent on various freeze dryer capacities.
Up to the point of overloading the system, the greater the surface area of the
sample, the faster the rate of freeze drying. By contrast, for a given surface area,
the thicker the sample the slower the rate of freeze drying. This is based on the
fact that the heat of sublimation is usually absorbed on one side of the frozen
sample and must travel through the frozen layer to vaporize water at the other
surface. In addition, as the sample is freeze dried, the water vapor must travel
through the layer of dried material. The thicker the sample, the greater the chance
that the dried layer may collapse which would cause an additional decrease in the
rate of freeze drying.
The surface area and thickness of the sample can usually be ignored when each
sample contains only a few milliliters. However, for larger volumes, the samples
should be shell frozen to maximize the surface area and minimize the thickness of
the sample. The volume of the freeze dry flask should be two to three times the
volume of the sample.
In order for lyophilization to occur, ice must be removed from the frozen sample
via sublimation. This is accomplished by the collector and the vacuum pump. The
collector, which should be at least 15 to 20°C colder than the eutectic temperature
(melting temperature) of the sample, traps vapor as ice. Since the vapor pressure
at the collector is lower than that of the sample, the flow of water vapor is from the
sample to the collector. Since this vapor diffusion process occurs very slowly
under normal atmospheric conditions, a good vacuum is essential to maintain an
Product Service 816-333-8811 or 1-800-522-7658
Chapter 1: Introduction
3
efficient rate. In many applications, the maintenance of a vacuum of 0.133 mBar
or less is recommended.
The rate of freeze drying is directly proportional to the vapor pressure and the
vapor pressure is dependent upon both eutectic temperature and solute
concentration of the sample. For example, a solution of sodium chloride and water
would freeze dry at a slower rate than pure water. The eutectic temperature of a
sodium chloride solution is about –21°C and at this temperature the vapor pressure
is about 1/16 that of water at 0°C. Although the eutectic temperature is not
dependent upon the concentration of sodium chloride, the vapor pressure of the
water would decrease as the concentration of sodium chloride increased. This is
due to the fact that as the solute concentration increases, less of the surface area of
the frozen sample is occupied by water. In general, most solutions or biological
samples will have a eutectic temperature of –10 to –25°C. However, if the sample
contains a simple sugar such as glucose or if the sample is animal or plant tissue,
the eutectic temperature may be as low as –30° to –50°C.
Freeze Dry Capacity
The volume of a sample that can be freeze dried at one time is related to factors
discussed previously and the size and design of the freeze dry system. With any
given instrument, the capacity is based on the surface area of the sample; the
eutectic temperature and concentration of the sample; and the rate and amount of
heat transferred to the frozen sample. Of these factors, the eutectic temperature is
the most important factor in determining the amount of sample that can be freeze
dried at one time, particularly when flasks are used. This is because as the eutectic
temperature decreases, the vapor pressure decreases but the rate of heat absorption
by the sample does not change. This tends to promote melting of the sample,
which leads to a marked increase in vapor pressure and ultimately overloads the
collector and vacuum pump. Samples that have eutectic temperatures of –20°C or
lower should be placed on the freeze dry system one flask at a time so that the
vacuum in the system may recover before adding another sample to the system. If
the vacuum does not recover, the capacity of the freeze dry system has been
exceeded and the sample should be removed.
If there is a problem with a particular type of sample melting when placed on the
freeze dry system, dilution of the sample with more water or providing some
insulation around the flask to decrease the rate of heat absorption by the sample
may help. If the eutectic temperature of the sample is –40 to –60°C, the freeze dry
system selected for use must be equipped with cascade type refrigeration so that
the collector temperature can be cooled to below –75°C, or a dry ice/solvent trap
may be used between the collector and the vacuum pump.
Product Service 816-333-8811 or 1-800-522-7658
Chapter 1: Introduction
4
Samples Containing Volatile Substances
In certain cases the solvent in a sample to be freeze dried may contain volatile
components such as acetonitrile, methanol, acetic acid, formic acid or pyridine. In
addition to these substances having an effect on the eutectic temperature, they may
increase the vapor pressure at the surface of the sample. Also, compared to water,
they will require the absorption of less heat for sublimation to occur. Hence,
samples that contain volatile substances will have a greater tendency to melt,
particularly when placed in flasks or exposed to room temperature. If a sample
containing a volatile substance tends to melt when placed on a freeze dry system,
dilution of the sample with more water will help keep the sample frozen. For
example, a 0.2M solution of acetic acid is much easier to freeze dry than a 0.5M
solution.
About This Manual
This manual is designed to help you learn how to install, use, and maintain your
Freeze Dryer. Instructions for performing routine maintenance and making minor
modifications to your Freeze Dryer are also included.
Chapter 1: Introduction provides a brief overview of the Freeze Dryer, explains
the organization of the manual, and defines the typographical conventions used in
the manual.
Chapter 2: Prerequisites explains what you need to do to prepare your site before
you install your Freeze Dryer. Electrical requirements are discussed.
Chapter 3: Getting Started contains the information you need to properly unpack,
inspect and install your Freeze Dryer.
Chapter 4: Using Your Freeze Dryer discusses the basic operation of your Freeze
Dryer. Information on how to attach samples and run the Freeze Dryer is included.
Chapter 5: Maintaining Your Freeze Dryer explains how to perform routine
maintenance on your Freeze Dryer.
Chapter 6: Using the RS232 Receptacle describes how to connect a computer for
monitoring the operation.
Chapter 7: Troubleshooting contains a table of problems you may encounter while
using your Freeze Dryer, including the probable causes of the problems, and
suggested corrective actions.
Chapter 8: Modifying Your Freeze Dryer Display explains how to recalibrate your
Freeze Dryer display.
Appendix A: Freeze Dryer Components contains labeled diagrams of the
components of the Freeze Dryer.
Product Service 816-333-8811 or 1-800-522-7658
Chapter 1: Introduction
5
!
Appendix B: Freeze Dryer Dimensions contains comprehensive diagrams showing
the dimensions for the Freeze Dryer.
Appendix D: Freeze Dryer Accessories lists the part numbers and descriptions of
all of the accessories available for your Freeze Dryer.
Typographical Conventions
Recognizing the following typographical conventions will help you understand and
use this manual:
Book, chapter, and section titles are shown in italic type (e.g., Chapter 3:
Getting Started).
Steps required to perform a task are presented in a numbered format.
Comments located in the margins provide suggestions, reminders, and
references.
Critical information is presented in boldface type in paragraphs that are
preceded by the exclamation icon. Failure to comply with the information
following an exclamation icon may result in injury to the user or permanent
damage to your Freeze Dryer.
Important information is presented in capitalized type in paragraphs that are
preceded by the pointer icon. It is imperative that the information contained in
these paragraphs be thoroughly read and understood by the user.
Your Next Step
If your Freeze Dryer needs to be installed, proceed to Chapter 2: Prerequisites to
ensure your installation site meets all of the requirements. Then, go to Chapter 3: Getting Started for instructions on how to install your Freeze Dryer and make all
of the necessary connections.
For information on the operational characteristics of your Freeze Dryer, go to
Chapter 4: Using Your Freeze Dryer.
If your Freeze Dryer is installed and you need to perform routine maintenance on
the Freeze Dryer, proceed to Chapter 5: Maintaining Your Freeze Dryer.
Refer to Chapter 7: Troubleshooting if you are experiencing problems with your
Freeze Dryer.
Product Service 816-333-8811 or 1-800-522-7658
6
CHAPTER2
PREREQUISITES
Before you install your Freeze Dryer, you need to prepare your site for installation.
Carefully examine the location where you intend to install your Freeze Dryer. You
must be certain that the area is level and of solid construction. An electrical source
must be located near the installation site.
Carefully read this chapter to learn:
the electrical supply requirements.
the vacuum pump requirements.
Refer to Appendix C: Freeze Dryer Specifications for complete Freeze Dryer
electrical and environmental conditions, specifications and requirements.
Electrical Requirements
The Freeze Dryer requires a dedicated electrical outlet. This outlet requires a 20
Amp circuit breaker or fuse for models rated at 115V (60 Hz). An outlet equipped
with a 15 Amp circuit breaker or fuse is required for models rated at 230V (50/60
Hz). The power cord on 115V models are equipped with a 20 Amp NEMA 5-20P
plug. Various power cord configurations are provided with 230V models. If this
does not match the available receptacle, remove this plug and replace it with an
approved plug of the suitable style.
Location Requirements
The Freeze Dryer should be located in an area that provides an unobstructed flow
of air around the cabinet. This air cools the refrigeration system. The refrigeration
system draws air through the rear panel and exhausts it through the sides. A
minimum of 3" must be allowed between the back and both sides of the Freeze
Dryer and adjacent wall surfaces. Restriction of airflow during operation could
adversely affect performance.
Product Service 1-800-522-7658
Chapter 2: Prerequisites
7
IT IS RECOMMENDED THAT THE CONSOLE MODEL
FREEZE DRYER IS SECURELY ATTACHED TO THE FLOOR
AFTER IT IS LOCATED WHERE IT WILL BE USED. THIS
WILL PREVENT ACCIDENTAL TIPPING AND POSSIBLE
INJURY. SEE INSTRUCTIONS IN CHAPTER 3 GETTING STARTED.
Refer to Appendix C: Freeze Dryer Specifications for dimensional drawings of the
Freeze Dryer. The space for benchtop models must be large enough to
accommodate the vacuum pump, which will be used with the Freeze Dryer.
Vacuum Pump Requirements
A vacuum pump must be provided by the user. A vacuum pump with a
displacement of 86 liters per minute and 0.002 mBar ultimate pressure is adequate
for most samples. The inlet fitting on the vacuum pump must be suitable for 3/4"
ID vacuum hose, which is provided. It is recommended that the vacuum pump is
equipped with an exhaust filter to minimize oil mist exhausting from the vacuum
pump. The operating vacuum level may be set on the Freeze Dryer. The higher
the pressure is set, the more likely it is that oil mist will be exhausted.
Vacuum pumps used with 115V models should be equipped with a 115V, 15 Amp
NEMA 5-15P plug. Vacuum pumps used with 230V models should be equipped
with a reverse IEC plug. This plug is included with 230V models. This will allow
the vacuum pump to be plugged into the receptacle on the back panel of the freeze
dryer. See Appendix D: Freeze Dryer Accessories for vacuum pumps available
from Labconco.
Chamber or Manifold Requirements
A freeze drying chamber or manifold is not included and must be purchased
separately. These allow samples to be attached to the Freeze Dryer. See
Appendix D: Freeze Dryer Accessories for available chambers and manifolds.
Product Service 816-333-8811 or 1-800-522-7658
8
!
CHAPTER3
GETTINGSTARTED
Now that the site for your Freeze Dryer is properly prepared, you are ready to
unpack, inspect, install and test your Freeze Dryer. Read this chapter to learn how
to:
unpack and move your Freeze Dryer.
set up your Freeze Dryer.
connect the electrical supply source to your Freeze Dryer.
properly exhaust your Freeze Dryer.
safely use solvents with your Freeze Dryer.
The FreeZone 2.5 Liter Benchtop Freeze Dryer System weighs
over 82 lbs. (37 kg). The Benchtop FreeZone Plus 2.5 Liter
Freeze Dry System weighs over 106 lbs. (48 Kg). The Console
FreeZone Plus 2.5 Liter Freeze Dry System weighs over 152 lbs.
(69 Kg). If you must lift the Freeze Dryer manually, use at least
two (2) persons and follow safe lifting guidelines.
Product Service 1-800-522-7658
Chapter 3: Getting Started
9
The United
States
Interstate
Commerce
Commission
rules require
that claims be
filed with the
delivery carrier
within fifteen
(15) days of
delivery.
Unpacking Your Freeze Dryer
Carefully unpack your Freeze Dryer and inspect it for damage that may have
occurred in transit. If your Freeze Dryer is damaged, notify the delivery carrier
immediately and retain the entire shipment intact for inspection by the carrier.
DO NOT RETURN GOODS WITHOUT THE PRIOR
AUTHORIZATION OF LABCONCO. UNAUTHORIZED
RETURNS WILL NOT BE ACCEPTED.
IF YOUR FREEZE DRYER WAS DAMAGED IN TRANSIT,
YOU MUST FILE A CLAIM DIRECTLY WITH THE FREIGHT
CARRIER. LABCONCO CORPORATION AND ITS DEALERS
ARE NOT RESPONSIBLE FOR SHIPPING DAMAGE.
DO NOT DISCARD THE CARTON OR PACKING MATERIAL
FOR YOUR FREEZE DRYER UNTIL YOU HAVE CHECKED
ALL OF THE COMPONENTS AND INSTALLED AND TESTED
THE FREEZE DRYER.
Freeze Dryer Components
Locate the model of Freeze Dryer you received in the following table. Verify that
the components listed are present and undamaged.
Product Service 816-333-8811 or 1-800-522-7658
Chapter 3: Getting Started
10
Power Cords
Catalog #
Description
Volts
Hz
PTFE
Coated
Chamber
1336400
1336500
1342100
1332601
1332701
7670520
2.5L Benchtop Freeze Dryer
115
60 X
7670521
2.5L Benchtop Freeze Dryer
115
60 X X
7670530
2.5L Benchtop Freeze Dryer
230
50 X
7670531
2.5L Benchtop Freeze Dryer
230
50 X X
7670540
2.5L Benchtop Freeze Dryer
230
60 X
7670541
2.5L Benchtop Freeze Dryer
230
60 X X
7670560
2.5L Benchtop Freeze Dryer
230
50 X
7670561
2.5L Benchtop Freeze Dryer
230
50 X X
7670570
2.5L Benchtop Freeze Dryer
230
50 X
7670571
2.5L Benchtop Freeze Dryer
230
50 X X
7670020
2.5L Benchtop Cascade Freeze Dryer
115
60 X
7670021
2.5L Benchtop Cascade Freeze Dryer
115
60 X X
7670030
2.5L Benchtop Cascade Freeze Dryer
230
50 X
7670031
2.5L Benchtop Cascade Freeze Dryer
230
50 X X
7670040
2.5L Benchtop Cascade Freeze Dryer
230
60 X
7670041
2.5L Benchtop Cascade Freeze Dryer
230
60 X X
7670060
2.5L Benchtop Cascade Freeze Dryer
230
50 X
7670061
2.5L Benchtop Cascade Freeze Dryer
230
50 X X
7670070
2.5L Benchtop Cascade Freeze Dryer
230
50 X
7670071
2.5L Benchtop Cascade Freeze Dryer
230
50 X X
7420020
2.5L Console Cascade Freeze Dryer
115
60 X
7420021
2.5L Console Cascade Freeze Dryer
115
60 X X
7420030
2.5L Console Cascade Freeze Dryer
230
50 X
7420031
2.5L Console Cascade Freeze Dryer
230
50 X X
7420040
2.5L Console Cascade Freeze Dryer
230
60 X
7420041
2.5L Console Cascade Freeze Dryer
230
60 X X
7420060
2.5L Console Cascade Freeze Dryer
230
50 X
7420061
2.5L Console Cascade Freeze Dryer
230
50 X X
7420070
2.5L Console Cascade Freeze Dryer
230
50 X
7420071
2.5L Console Cascade Freeze Dryer
230
50 X X
Part #
Component Description
7426100
User’s Manual
1291000
Plug (230V models only)
7373450
Vacuum Hose
1488800
Hose Clamps (2)
7395801
Chamber Lid
7395800
Lid Gasket
7419300
Baffle (Cascade Models Only)
7386901
Fitting
Plus the following:
Product Service 816-333-8811 or 1-800-522-7658
Chapter 3: Getting Started
11
!
If you did not receive one or more of the components listed for your Freeze Dryer,
or if any of the components are damaged, contact Labconco Corporation
immediately for further instructions.
Setting Up Your Freeze Dryer
After you verify receipt of the proper components, move your Freeze Dryer to the
location where you want to install it. Then, follow the steps listed below.
Component Orientation
The refrigeration system in the Freeze Dryer draws air in through the rear of the
cabinet and exhausts air out through the sides. A minimum of 3" should be
allowed between the back and sides of the Freeze Dryer and the adjacent wall
surface. Restriction of the airflow through the cabinet during operation could
adversely affect performance.
After the console model is positioned at the location where it will be operated, it
should be securely attached to the floor. Use (2) 1/4" x 5.0 long wood screws or
lag bolts if the floor is wood or concrete anchors if the floor is concrete. Two .312
diameter holes are located in the floor of the lower cabinet. Using these holes as a
template, mark the floor directly under the two holes. Roll the Console Freeze
Dryer aside and drill suitable pilot holes for the selected screws/anchors. Install the
vacuum pump prior to securing the Console Freeze Dryer to the floor. Roll the
Console Freeze Dryer back into position and install the screws/anchors through the
holes in the cabinet floor and into the building floor.
FAILURE TO ATTACH THE CONSOLE FREEZE DRYER TO
THE FLOOR COULD RESULT IN ACCIDENTAL TIPPING
AND POSSIBLE INJURY
Vacuum Pump Connection
A vacuum pump as described in Chapter 2: Prerequisites is required to operate
your Freeze Dry System properly. The Freeze Dryer is equipped with a 3/4" ID,
heavy wall, vacuum hose for connecting the collector chamber to the vacuum
pump.
Make certain that the oil in the vacuum pump is at the proper
level.
Place the vacuum pump near the benchtop Freeze Dryer models. For console
models, from the back of the unit, place the vacuum pump on the floor of the
cabinet.
Product Service 816-333-8811 or 1-800-522-7658
Chapter 3: Getting Started
12
TO MAXIMIZE THE STABILITY OF THE CONSOLE FREEZE
DRYER, POSITION THE VACUUM PUMP AS CLOSE TO THE
CENTER OF THE CABINET FLOOR AS POSSIBLE.
Connect the vacuum pump power cord to the receptacle on the back of the cabinet
labeled “vacuum pump.” If the vacuum pump has an off/on switch, turn the switch
on. The vacuum pump will be controlled by the Freeze Dryer. Connect the
vacuum pump inlet port to the vacuum tube on the back of the cabinet with the
vacuum hose and clamps provided. If necessary, cut the hose to proper length to
allow for gentle bends without kinks.
Models rated for use on 230V have a reverse IEC connector for attaching the
vacuum pump power cord. A plug is supplied that can be installed on the vacuum
pump power cord if necessary.
Electrical Connection
Plug the power cord into the receptacle on the back of the Freeze Dryer and plug
the other end into a suitable power receptacle.
Drying Chamber or Drying Manifold Installation
The drying chamber or drying manifold, purchased separately, may be positioned
directly above the 3-inch hole in the collector chamber lid. When vacuum is
applied to the system, the chamber will be held securely in place.
Chemical Resistance of Freeze Dryer
Components
The FreeZone Freeze Dry System is designed to be chemically resistant to most
compounds that are commonly used in freeze drying processes. However, by
necessity, the freeze dryer is comprised of a number of different materials, some of
which may be attacked and degraded by certain chemicals. The degree of
degradation is dependent on the concentration and exposure duration. Some of the
major components of the FreeZone Freeze Dry System that are susceptible to
degradation are as follows:
Product Service 816-333-8811 or 1-800-522-7658
Chapter 3: Getting Started
13
Acids
Buffers Solvents
Component
Material
Acetic Acid 20%
Formic Acid
Trifluoroacetic Acid (TFA) Calcium Chloride
Sodium Phosphate Acetone
Acetonittirle
Carbon Tetrochloride
Cyclohexane
Dioxane
Methyl t-Butyl Ether (BTBE)
Pyridine
Valve Stem
Acetal
C D D D D
Collector
Lid
Acrylic
D D D D
Hoses,
Gaskets &
Valve
Bodies
Neoprene
C D D C C D D D C
D
Flask Top
Silicon
Rubber
C D D D D D C
D
Chamber &
Fittings
Stainless
Steel
C
C – Moderate degradation; Limited use
D – Severe degradation; infrequent use recommended; immediate thorough
cleaning required.
Most common compounds used in freeze drying processes, if allowed to enter
the vacuum pump, will degrade the oil and cause damage to the vacuum pump.
Sugars and proteins typically will have minimal negative effect on any of the
materials of construction.
When using compounds in the Freeze Dryer that are hostile to the materials of
construction, it is imperative the equipment is thoroughly cleaned after use.
Rubber and plastic components that have been exposed to damaging
compounds should be removed and flushed with water.
The oil in the vacuum pump should be checked often. It must be changed if it
is cloudy, shows particles or is discolored. The useful life of vacuum pump oil
can be extended if the vacuum pump is operated for an extended period of time
after a freeze dry run. This allows contaminants to be purged from the hot oil.
This must be done with the inlet to the pump blocked off to prevent air from
free flowing through the pump. If the pump is operated at an elevated vacuum
level (> 10mBar), oil may be expelled from the pump and damage could occur.
Product Service 816-333-8811 or 1-800-522-7658
Chapter 3: Getting Started
14
!
Another way to extend the life of the vacuum pump is to install an optional
secondary trap in the line between the Freeze Dryer and the vacuum pump.
Contact Labconco for ordering information.
With prudent maintenance the FreeZone Freeze Dry System will provide years of
service. Warranty on the affected parts will be voided if maintenance has been
obviously neglected. If you have questions about using specific compounds in the
Freeze Dryer, contact Labconco Technical Service at 1-800-821-5525 or 816-3338811 or e-mail: labconco@labconco.com.
Solvent Safety Precautions
Solvents used in the Freeze Dryer may be flammable or
hazardous to your health. Use extreme caution and keep
sources of ignition away from the solvents. When using
flammable or hazardous solvents, the vacuum pump must be
vented to or operated inside a fume hood.
Hazardous materials such as strong acids or bases, radioactive
substances and volatile organics must be handled carefully and
promptly cleaned up if spilled. If a sample is spilled in the
collector chamber it must immediately be cleaned up.
WARNING: The disposal of substances used in connection with
this equipment may be governed by various Federal, State or
local regulations. All users of this equipment are urged to
become familiar with any regulations that apply in the user’s
area concerning the dumping of waste materials in or upon
water, land or air and to comply with such regulations.
Product Service 816-333-8811 or 1-800-522-7658
15
!
CHAPTER4
USING YOURFREEZEDRYER
After your Freeze Dryer has been installed as detailed in Chapter 3: Getting
Started, you are ready to begin using your Freeze Dryer. Read this chapter to learn
how to:
operate the controls.
understand the display.
connect samples.
Do not use the Freeze Dryer in a manner not specified by the
manufacturer (refer to Appendix C: Freeze Dryer Specifications). The electrical protection properties of the
Freeze Dryer may be impaired if the Freeze Dryer is used
inappropriately.
Product Service 1-800-522-7658
Chapter 4: Using Your Freeze Dryer
16
1
4
7
9
12
8 6 3
11
5 2 10
Freeze Dryer Controls
The control panel for the Freeze Dryer is shown below with a description about its
function.
1. LCD Display – Displays system operating parameters, set-up parameters and
alarm messages.
2. Menu Switch – This switch is used to change the display from operating
system parameters to set-up parameters.
3. Select Switch – Used to select set-up parameters.
4. Vacuum Switch – Used to start or stop the vacuum pump when operating in manual
start-up mode.
5. Vacuum Indicator – This green LED indicates that power is being supplied to the
vacuum pump receptacle on the back of the Freeze Dryer.
6. Manual Refrigeration Switch – Used to start only the refrigeration module.
7. Manual Operation Indicator – When lit, the green LED indicates the Freeze Dryer is
being controlled manually by the operator. Each function must be started by the
operator.
8. Auto Mode Switch – Used to start or stop the refrigeration and the Auto Mode process.
9. Auto Mode Indicator – When lit, the green LED indicates that the Freeze Dryer is in
Auto Mode. In this mode, the vacuum pump will start when the collector temperature
reaches –40°C.
10. Vacuum Graph Display – This display indicates the relative system vacuum level. The
highest LED indicates that the vacuum level is above 2.0 mBar. The indicators will
sequence down when the vacuum level reaches 2.0, 1.0, 0.8, 0.6, 0.45, 0.12 mBar. The
lower green LED flashes when the system vacuum level is 0.45 to 0.12 mBar and
illuminates steadily below 0.133 mBar.
Product Service 816-333-8811 or 1-800-522-7658
Chapter 4: Using Your Freeze Dryer
17
11. Collector Temperature Graph Display – This display indicates the temperature of the
collector. The highest LED indicates the collector temperature is warmer than 10°C.
The indicators will sequence down when the temperature reaches 10, 0, -10, -20, -30, 40°C. When the collector temperature is –40°C or lower the green indicator will light.
12. Alarm Indicator – This red LED indicates that a system alarm has occurred. Press the
Menu Switch to display the alarm message on the LCD display.
13. Main Power Switch – Turns the Freeze Dryer on or off. (Not shown, located on the
right side of the cabinet).
Operation Checklist
The following checklist should be followed prior to each use of your Freeze Dryer:
1. Wipe the interior of the collector chamber with a soft cloth or paper towel to
remove any accumulated moisture.
2. Check the collector chamber drain hose to ensure that the hose is free of
moisture and that the drain plug is securely installed. The Freeze Dryer will
not start if moisture is detected.
3. Using a soft, lint-free cloth or paper towel, wipe the collector chamber lid
gasket to remove any dirt and contaminants that could cause a vacuum leak.
Vacuum grease is not required on the lid gasket to obtain a proper vacuum seal.
4. Remove the accessory drying chamber or manifold from the connection port
and using a soft, lint-free cloth or paper towel, wipe the port gasket and sealing
surfaces of the drying chamber/manifold to remove any dirt and contaminants
that could cause a vacuum leak. Reinstall the drying chamber or manifold on
the port. Vacuum grease is not required on the port gasket to obtain a proper
vacuum seal.
5. Check that each sample valve is closed or in the “vent” position.
Vacuum Pump Ballast Setting
Most vacuum pumps are equipped with a gas ballast mechanism. The freeze dry
process requires high vacuum. Therefore, it is recommended that the gas ballast be
closed during the operation of the Freeze Dry System. If the gas ballast is left
open for extended periods of operation, the oil can be pumped out the exhaust,
causing the pump to fail.
Operating the Freeze Dryer
Set-Up
The Freeze Dryer may be configured to automatically start the vacuum pump when
the collector temperature reaches –40°C. The display units for vacuum may be
selected to be mBar, Pascal (Pa) or Torr and the temperature may be displayed as
°F or °C. The run time of the refrigeration system and the vacuum pump may be
monitored.
Product Service 816-333-8811 or 1-800-522-7658
Chapter 4: Using Your Freeze Dryer
18
VACUUM UNITS
mBar Pa Torr
TEMPERATURE UNITS
°C °F
REFRIG TOTAL HOUR: XXXX
SERVICE HOUR: XXXX
VACUUM SET POINT
XXXX YYY
To configure your Freeze Dryer, turn the main power switch ON and press MENU.
The display will show:
Press SELECT until the desired units are flashing.
Press MENU.
The display will show:
Where YYY is the units selected above.
Press SELECT until the desired vacuum operating level is displayed. See
“Setting the Operating Vacuum Level” for guidelines.
If “0” is selected, the vacuum control is disabled and the vacuum level in
the Freeze Dry System will be determined by the capabilities of the
vacuum pump.
Press MENU.
The display will show:
Press SELECT until the desired units are flashing.
Press MENU.
The display will show:
Product Service 816-333-8811 or 1-800-522-7658
Chapter 4: Using Your Freeze Dryer
19
REFRIG TOTAL HOUR: XXXX
RESET SERVICE HOUR: ?
VACUUM TOTAL HOUR: XXXX
SERVICE HOUR: XXXX
RS-232 TRANSMISSION RATE
10 SECONDS
VACUUM TOTAL HOUR: XXXX
RESET SERVICE HOUR: ?
To reset the SERVICE HOUR to 0 press SELECT. This allows you to
keep track of the time the refrigeration system operated since it was
serviced.
The display will show the refrigeration system hours.
To reset the service hours to 0, continue to hold SELECT 5 seconds until
hours are reset.
TOTAL HOUR cannot be reset.
Press SELECT.
The display will show the vacuum pump hours.
To reset the SERVICE HOUR to 0, press SELECT.
The display will show:
If you want to reset the service hours to 0, continue to hold SELECT five seconds
until hours are reset.
Press SELECT.
The display will show the RS-232 data transmission rate.
To change the rate press SELECT until the desired time interval is shown.
The time between data transmissions may be set to occur at 10, 30, 60, 300
or 600 second intervals.
Press MENU to return to display operating parameters or after a short
delay, the display will automatically switch to show operating parameters.
Product Service 816-333-8811 or 1-800-522-7658
Chapter 4: Using Your Freeze Dryer
20
Baffle
The cylindrical plastic component shipped with Cascade Freeze Dryers (Models
76700XX and 74200X) is a baffle, which is designed to evenly distribute collected
ice over the entire collector coil. Proper positioning of the baffle is required to
achieve maximum ice loading capabilities.
The baffle installs into the collection chamber with the flat flange upward. The
large boss on the flange should be positioned toward the rear of the chamber. The
arrow should point toward the front of the Freeze Dryer.
Automatic Start-Up
To run the Auto Mode, press the panel switch labeled REFRIGERATION AUTO.
The green LED above the switch will illuminate. This will start the refrigeration
system. When the collector reaches –40°C, the vacuum pump will start. The
Temperature and Vacuum Graph will indicate collector temperature and system
vacuum. The LCD display will show the actual temperature of the collector.
When the vacuum in the system is above 5 mBar the vacuum display will indicate
“HI.” At 5 mBar and below, the display will show the actual vacuum.
When the system vacuum is between 0.450 and 0.133 mBar, the lower green
vacuum graph LED will flash. When the system vacuum level is <0.133 mBar, the
green LED will be lit steadily indicating that most samples may be added.
Manual Start-Up
To manually run the freeze dry process press the REFRIGERATION MAN switch.
This will start the refrigeration system. The green LED above the switch will
illuminate. When the collector temperature reaches –40°C, the vacuum pump may
be started by pressing the VACUUM switch. The Temperature and Vacuum
Graph will indicate collector temperature and system vacuum. The LCD display
will show the actual temperature of the collector. When the vacuum in the system
is above 5 mBar the vacuum display will indicate “HI.” At 5 mBar and below, the
display will show the actual vacuum.
When the system vacuum is between 0.450 and 0.133 mBar, the lower green
vacuum graph LED will flash. When the system vacuum level is <0.133 mBar, the
green LED will be lit steadily indicating that samples may be added.
Setting the Operating Vacuum Level
The vacuum level may be set by the user to optimize the freeze dry process.
Normally, the sublimation rate will increase if there is less vacuum (a higher
pressure) in the Freeze Dryer. A good starting place is to set the vacuum so its
level is equivalent to about 10°C colder than the eutectic or collapse temperature
Product Service 816-333-8811 or 1-800-522-7658
Chapter 4: Using Your Freeze Dryer
21
Material
Solidification/Eutectic
Temperature
Pre-Freeze
Temperatures
Vacuum Set Point
Bacteria,Virus
-40°C and Colder
-50°C &
Colder
0.040 mBar &
Lower
Milk
-5 to -13
-15 to -23
1.65 to 0.77
Fungi
-40 and Colder
-50 & Colder
0.04 and lower
Vegetable Tissue
-25 to -50
-35 to -60
0.22 to 0.01
Human Tissue
-30 to –40
-40 to –50
0.12 to 0.04
Blood Plasma
-10 to –25
-20 to –35
1.03 to 0.22
Vaccine
-30 to –40
-40 to –50
0.12 to 0.04
of the sample. Adjustments to the vacuum level must be made for various freeze
drying conditions. Factors that must be considered are whether the sample is
freeze dried on heated shelves or in glassware attached to manifold valves, the
volatility of the sample itself, the size of the sample and the heat energy supplied
to the sample.
Some guidelines for setting the freeze dryer are shown below. These show prefreezing temperatures and vacuum levels. Exact protocols must be determined by
the user for the specific samples that are being freeze dried.
When the vacuum control is set to operate at less vacuum, the ice holding capacity
of the collector may be decreased. This can be improved by installing a baffle in
the collector. See Appendix D Freeze Dry Accessories for the baffle part number.
See the section titled Baffle in this chapter for the proper installation.
Pre-Freezing Samples
Appropriate containers for freeze drying include ampules, serum bottles, and wide
mouth freeze drying flasks. Shell freezing of samples is recommended for wide
mouth freeze drying flasks. Smaller samples in ampules and serum bottles may be
frozen in a freezer or in the center of the collector chamber. The sample container
size should always be at least two to three times the sample size (i.e., 40 ml
samples should be prepared in 80 ml containers or larger). The temperature
required for pre-freezing is dependent on the characteristics of the sample. Prefreezing temperature typically is at least 10° to 20°C below the eutectic or collapse
temperature of the sample.
Adding Sample
The following procedure should be followed when using sample valves in the
freeze dry process:
1. Connect a pre-frozen sample to a sample valve on the drying chamber or manifold
using an adapter. Turn the plastic valve knob to the “VACUUM” position to open
the valve, which connects the attached sample to system vacuum. The bevel on the
knob should be positioned toward the sample port to apply vacuum to the sample.
Product Service 816-333-8811 or 1-800-522-7658
Chapter 4: Using Your Freeze Dryer
22
Vent/Backfill Port
Bevel in “vent” position
Bevel in “vacuum”
position
Adapter
2. Before adding another sample, allow system vacuum to return to 0.133 mBar
or lower. Any combination of valves and sample sizes may be utilized at one
time provided that the system vacuum and collector temperature remain
sufficiently low to prevent melting of the frozen sample.
3. When all the frost has disappeared from the outer surface of the sample
container and no cold spots can be detected by handling the container, the
sample is nearly dry. To be certain of low final moisture content, dry the
sample for several hours past this point.
4. To remove a container after drying is complete, turn the plastic knob on the
valve to the “VENT” position, which closes the valve and vents the container.
Should backfilling with an inert gas be required, connect the gas supply line to
the vent port on the valve before turning the plastic knob on the valve to vent
position. The sample container may now be removed. In the vent position the
bevel on the valve knob should point away from the sample port.
5. Ampules may be flame sealed while connected to a valve by using a sealing
torch. Care must be taken not to burn the valve. An insulation material placed
between the valve and the torch is recommended.
Shut Down
At the end of a run or when a sufficient amount of condensate accumulates on the
collector coil to obstruct the flow of vapor to the collector chamber, the Freeze
Dryer should be defrosted. First, release system vacuum by turning the plastic
knob on a valve to the open position or by pulling the collector chamber drain plug
out of the drain hose. Now press the Vacuum Switch on the control panel to turn
the vacuum pump OFF. Press the Refrigeration Switch next to the illuminated
Product Service 816-333-8811 or 1-800-522-7658
Chapter 4: Using Your Freeze Dryer
23
!
LED to turn OFF the refrigeration system. Turn OFF the Main Power Switch on
the right hand side of the cabinet.
Defrosting
The following procedure should be followed when defrosting the collector coil:
1. Pull the collector chamber drain hose out from the left hand side of the Freeze
Dryer and remove the drain plug. Place the drain hose in a suitable container
to collect the condensate that will melt off the collector coil.
2. Remove the collector chamber lid and allow ambient room air to melt the ice.
Dispose of the liquid appropriately.
3. Flush the collector chamber with water and wipe chamber dry.
4. If rapid defrost is desired, pour warm water over the collector coil. Do not
allow the liquid to enter the vacuum port on the upper rear of the chamber.
5. Reinstall the drain hose plug and slide drain hose back into the side of the
cabinet. Dispose of the liquid appropriately.
Utilization of acid requires immediate cleaning and
neutralization after defrost or physical damage to the collector
chamber and collector coil will result.
Do not attempt to chip ice from the collector coil as damage
may occur to the coil.
Never attempt to start the vacuum pump when there is liquid in
the collector chamber. This could result in damage to the
vacuum pump.
Alarms
A number of unusual events may occur during a lyophilization procedure that can
adversely effect the operation of the Freeze Dryer. If an event occurs, the alarm
indicator will flash and the beeper will sound.
The beeper will automatically mute itself after one minute. The specific alarm can
be identified by pressing MENU. Pressing MENU multiple times will display
multiple alarms if they have occurred. The following “out of specification”
conditions will initiate an alarm:
Power Failure
If a power failure occurs during a Freeze Dry process, the vacuum control valve
will allow air to bleed into the Freeze Dryer System. If the failure is of a short
duration and the collector does not warm up excessively, when power is restored
the Freeze Dryer will restart and resume operation of the refrigeration and vacuum
systems. If the power failure lasts for a longer duration and the collector warms
above safe limits, when the power is restored, the Freeze Dryer will not
Product Service 816-333-8811 or 1-800-522-7658
Chapter 4: Using Your Freeze Dryer
24
POWER FAIL
HOLD SELECT TO CLEAR IT
LINE VOLTAGE ERROR
COLLECTOR TEMPERATURE
HOLD SELECT TO CLEAR IT
automatically restart. This prevents melted sample from being drawn into the
collector and prevents liquid from being sucked into the vacuum pump.
When power is restored, the alarm indicator will flash and the alarm beeper will
sound. When the MENU button is pressed the display will show:
Press and hold SELECT for five seconds until the error message is cleared from
the display.
Line Voltage Out of Range
If the voltage supplied to the Freeze Dryer varies beyond allowable limits, the
alarm will be activated. When an alarm occurs, press MENU and the display will
show:
The alarm will self-cancel when the proper voltage is restored. The high and low
alarm points are preset at the factory to correspond to the normal allowable voltage
variations based on the nominal voltage specified for the freeze dryer. Some
models may be operated outside the normal voltage limits. See Appendix C.
If necessary, the high and low alarm points may be adjusted for these models. See
Chapter 8.
Temperature Out of Range
An alarm will be activated if the collector temperature rises above –40°C. If this
occurs, the display will show:
Press and hold SELECT for five seconds until the error message is cleared from
the display.
Product Service 816-333-8811 or 1-800-522-7658
Chapter 4: Using Your Freeze Dryer
25
CHANGE VACUUM OIL
HOLD SELECT TO CLEAR IT
MOISTURE IN COLLECTOR
Service Vacuum Pump
The vacuum pump normally plugs into the vacuum pump electrical receptacle on
the back of the Freeze Dryer. When the Freeze Dryer has accumulated a total of
1000 operating hours, the alarm indicator will flash. Press MENU to display the
alarm:
Press and hold SELECT for five seconds until the error message is cleared from
the display.
This will reset the 1000 hour timer. It may be necessary to service the vacuum
pump more frequently than every 1000 hours depending on the operation of the
Freeze Dryer. The pump oil should be regularly monitored to verify that it is
clean.
Moisture in Collector
If there is liquid in the collector chamber and either a REFRIGERATION button or
the VACUUM button is pressed, neither function will start and an alarm will be
activated. When MENU is pressed, the display will show:
When the collector is drained, the alarm will self-cancel.
This feature prevents the inadvertent start of the vacuum pump when liquid is in
the collector and thereby can extend the life of the vacuum pump.
A few drops of liquid may remain the drain hose after draining. When the vacuum
pump is started these drops may be sucked into the collector chamber. This is
normal.
Product Service 816-333-8811 or 1-800-522-7658
26
CHAPTER5
MAINTAINING YOURFREEZE
DRYER
Under normal operation, the Freeze Dryer requires little maintenance. The
following maintenance schedule is recommended:
As needed:
1. Clean up all spills; remove liquids from the chamber.
2. Clean lid and gasket using soft cloth, sponge or chamois and a mild, nonabrasive soap or detergent.
3. Check oil level of the vacuum pump. It should be between MIN and MAX. If
the oil level is less than an inch (25.4 mm) above MIN, add oil to proper level.
4. If oil shows cloudiness, particles or discoloration, drain the pump and replace
with fresh oil.
5. Utilization of acids requires immediate cleaning and neutralization after a
run or physical damage to the collector chamber and collector coil will
result.
6. Check the collection chamber for condensed or frozen solvents and dispose of
appropriately. Completely empty the collector chamber before the next run or
before attempting to start the vacuum pump.
7. If sample valves leak or stems become difficult to turn, pull the stem out of the
rubber valve body and apply a thin coat of vacuum grease to the stem.
Reinsert the stem into the valve body.
Product Service 1-800-522-7658
Chapter 5: Maintaining Your Freeze Dryer
27
Monthly:
1. The rubber components on the Freeze Dryer may eventually deteriorate and
require replacement. The effective life of rubber parts depends upon both their
usage and the surrounding environment. Check all rubber hoses and gaskets
and replace any that show signs of hardening, permanent set or deterioration.
2. Using a soft cloth, sponge or chamois and a mild, non-abrasive soap or
detergent, clean the acrylic chamber lid.
3. Using a soft cloth, sponge, or chamois and a mild, non-abrasive soap or
detergent, clean the exterior surfaces of the Freeze Dryer. Liquid spray
cleaners and polishes may be used on the exterior surfaces. Do not use
solvents to remove stains from the exterior surfaces as they may damage the
finish.
Annually:
1. Every 12 months, or more often if the Freeze Dryer is operated in a dusty
environment, the refrigeration system condenser should be cleaned. Using a
vacuum cleaner with brush attachment, clean the condenser to ensure proper
airflow for peak performance.
Front Access Panel Removal (Console Models)
The front stainless steel panel can be removed by placing a flat blade screwdriver
in the rectangular hole in the center of the kick panel. Pivot the handle downward
to disengage the latch and simultaneously pull the bottom edge of the lower panel
forward. The lower panel is installed by first sliding the top edge under the lip on
the plastic control panel. Pivot the bottom inward and push until the latch is
secure.
Product Service 816-333-8811 or 1-800-522-7658
28
CHAPTER6
USINGTHE RS232
RECEPTACLE
The operation of the Freeze Dryer can be monitored using a computer when it is
connected to the RS232 receptacle on the rear panel. The computer cannot control
the operation of the Freeze Dryer. The monitored parameters are the collector
temperature in °C and vacuum in microbars.
Computer Connection for Computer
Interface
Check your computer to see which type of serial port is provided, then use a
connecting cable below:
1. Computers with a 25 pin D-sub male serial connector should use Connect
Cable, Labconco part number 7537801, to connect a computer to the
Labconco Freeze Dryer.
2. Computers with a 9 pin D-sub male serial connector should use Connect
Cable, Labconco part number 7537800, to connect a computer to the
Labconco Freeze Dryer.
The purpose of the RS232 interface is to send data to a data logging computer to
monitor the state and activity of the Freeze Dryer. This data is half duplex data.
The data properties are as follows:
1. Data Rate 2400 Baud
2. 8 Bit word length
3. 1 Start bit, 1 Stop bit
4. No parity is transmitted
5. Standard ASCII character set
Product Service 1-800-522-7658
Chapter 6: Using the RS232 Receptacle
29
B
= Base Unit
T
= Collector Temperature
-48
= Collector Temperature in °Celsius
V
= Vacuum
18
= Vacuum Level in Micro Bars
SF
= Shell Freezer Temperature (not available on 2.5L models)
9
= Feature not installed or out of range
C
= Vacuum Chamber Temperature(not available on 2.5L models)
PG
= Purge (not available on 2.5L models)
RS-232 TRANSMISSION RATE
10 SECONDS
The time between data transmissions may be varied by the user to occur at 10, 30,
60, 300 or 600 second intervals. Press the MENU button until the RS232 screen
appears on the display.
Press select until the desired time interval is shown.
The format of the transmitted message is as follows:
B<T= -48 V= 18 SF = 999 C=999 PG=9 >
Where:
There are several commercially available software packages, which can read
RS232 data and enter the data into a computer program, such as a word processor
(to create a text file) or spreadsheet (to tabulate and plot the data). Consult your
laboratory supply dealer regarding the latest software available. The following
describes how to use an IBM compatible computer with Hyper Terminal™
software (included with Windows® 95, 98, XP or 2000 operating systems) to
collect the RS232 data from your Freeze Dryer:
1. Make sure that the Freeze Dryer is properly connected to the communication
port on the computer.
2. Open Hyper Terminal™ software.
i)Windows® 95 or 98 use:
START/PROGRAMS/ACCESSORIES.
ii) Windows® 2000 use:
START/ROGRAMS/ACCESSORIES/COMMUNICATIONS
Product Service 816-333-8811 or 1-800-522-7658
Chapter 6: Using the RS232 Receptacle
30
The first time Hyper Terminal™ is opened a dialogue box requesting an area code
and phone number will appear. Enter the appropriate numbers and continue.
3. The “Connection Description” dialogue box will open. Type in a user defined
name and select an icon for the new connection. Press “OK.”
4. The “Connect To” dialogue box will open. Using the down arrow selection
button, select the communication port to which the cable has been connected.
Press “OK.”
Product Service 816-333-8811 or 1-800-522-7658
Chapter 6: Using the RS232 Receptacle
31
5. The “Com X Properties” dialogue box will open. Enter the appropriate data
properties and press “OK.”
6. When the Freeze Dryer main power switch is on, the data will be transmitted
and updated at the time intervals selected by the user.
Product Service 816-333-8811 or 1-800-522-7658
32
CHAPTER7
TROUBLESHOOTING
Refer to the following if your Freeze Dryer fails to operate properly. If the
suggested corrective actions do not solve your problem, contact Labconco for
additional assistance.
Labconco Freeze Dry Systems that are clean, dry and without samples attached
should reach a vacuum of 0.133 mBar within 10 minutes and should achieve an
ultimate vacuum of 0.040 mBar within 18 hours when the refrigeration is
operating. If the Freeze Dryer does not obtain a satisfactory vacuum, perform the
following maintenance tests.
I. Vacuum Pump
First make sure that the vacuum pump operates. If it fails to operate, check the
electrical connections of the Freeze Dryer to the power source and then check the
electrical connection of the vacuum pump to the Freeze Dryer. If the vacuum
pump has a power switch, make sure that it is turned on. If the vacuum is not
adequate when the vacuum pump is operating, proceed with the following steps:
1. Check the oil and ensure it is clear and clean. If the oil looks cloudy or has
any particulates, replace the pump oil. Sometimes it may be necessary to
flush the pump with clean oil several times. To flush the pump, run the
pump 5 to 10 minutes to allow the oil to warm up. Drain the oil and refill
with clean oil. Repeat as necessary.
2. Check the oil level in the pump. Ensure it is filled to the correct level.
3. Check vacuum hose connections from the pump to the Freeze Dryer and try
running the unit.
If vacuum problems continue, consider obtaining a second vacuum gauge capable
of reading a vacuum of 0.010 mBar. It is often useful in determining if the
vacuum pump is operating properly and the vacuum sensor reading is accurate.
Product Service 1-800-522-7658
Chapter 7: Troubleshooting
33
4. Isolate the pump by disconnecting the vacuum hose from the Freeze Dryer.
Deadhead the pump by inserting the vacuum sensor from a secondary
vacuum gauge into the end of the vacuum hose and observe the vacuum
reading obtained. Confirm that the pump is capable of achieving an
ultimate vacuum less than 0.010 mBar or approximately 10 microns. If an
inadequate vacuum reading is obtained, the pump has most likely failed
and may need to be replaced or rebuilt.
II. Gaskets, Tubing, Connections, Sample Valves
1. Check all sample valves on the drying chamber/manifold and ensure all
valves are closed or in the vent position. Check to ensure all fittings are
tight and sealed properly against the chamber. Inspect valves for cracks or
cuts.
2. Inspect gasket(s) for cracks or defects. Ensure the gasket is clean. Apply a
thin coat of vacuum grease to the gasket. Keep in mind too much vacuum
grease can actually cause leaks.
3. Check the drain line and ensure the drain plug is installed and the tube
appears to be in good condition.
4. Check all rubber vacuum tubing for signs of deterioration or cracking.
5. Check all connections and make sure they are secure and leak tight.
6. When checking the sample valves, pull as much vacuum as possible. If a
vacuum indication is displayed, wiggle or rotate the valves and watch the
gauge for any fluctuations. Fluctuations can indicate a potential vacuum
leak. If the valve seems to be in good condition, remove the valve and
apply a thin coat of vacuum grease to the stem and the outside sealing
surface of the valve body, and reinstall the valve. DO NOT APPLY
VACUUM GREASE TO THE SAMPLE VALVE GASKETS, WHICH
SEAL TO THE WALL OF THE DRYING CHAMBER. If the valve
still seems to be the source of the problem, remove the valve and stopper
the hole with a rubber stopper. Continue checking the other valves.
7. Make sure that the sample valve bulkhead nuts are tight. Either tighten by
hand or use the wrench supplied with chambers. To use the wrench, first
remove the knob and stem assembly and also the valve body. The pins on
the wrench fit into the round holes of the nut. The nut should be snug but
avoid over tightening. Replace the valve body and knob and stem
assembly.
Product Service 816-333-8811 or 1-800-522-7658
Chapter 7: Troubleshooting
34
Thrust Washer
Valve Body
Knob & Stem
Assembly
Valve Body
Trim Ring
Manifold Assembly
Knob & Stem
Assembly
Bulkhead Nut
Gasket
(No grease
required)
Bulkhead
Fitting
Valve Body
Thrust Washer
Valve Assembly (Manifold)
Valve Assembly (Stainless Steel Chamber)
Valve Body
Knob & Stem
Assembly
Valve Assembly (Clear Chamber with Valves)
The illustrations below show how the sample valve installs on a chamber and a
manifold.
Product Service 816-333-8811 or 1-800-522-7658
Chapter 7: Troubleshooting
35
Freeze Dry Flask
Isolating the freeze dry system from the accessory chamber/manifold
Gasket
III. System Components and Collection Chamber
Isolation
This test determines if the source of a leak is in the drying chamber.
1. Remove the drying chamber or manifold.
2. Leave the gasket on and turn a large freeze dry flask upside down to cover
the connection port.
3. Start the Freeze Dryer and observe the vacuum indication. If the vacuum
indication is good, then the problem is in the drying chamber or manifold
and you should return to Section II, Gaskets, Tubing, Connections, Sample Valves.
Product Service 816-333-8811 or 1-800-522-7658
Chapter 7: Troubleshooting
36
Stopper
Checking the connections to the pump and the vacuum tube
The following test determines if a vacuum leak is in the collection chamber.
1. Insert a rubber stopper in the small hole in the rear of the collector chamber
and another in the drain port in the bottom.
2. Start the Freeze Dryer and observe the vacuum indication. If the vacuum is
good, the problem is in the collection chamber. If it is bad, the problem is
in the plumbing connections from the collection chamber to the vacuum
pump.
3. If the leak is in the chamber, check the drain hose, hose connections and
the drain plug.
4. If the leak is in the plumbing connections, check all rubber parts, fittings
and clamps.
Product Service 816-333-8811 or 1-800-522-7658
Chapter 7: Troubleshooting
37
Refrigeration Module Operation
Under a no-load condition, the FreeZone Single Stage Freeze Dry System can
achieve a collector temperature of -50°C or lower within 30 minutes when the
vacuum pump is operating, depending on ambient temperature and humidity.
FreeZone Cascade Freeze Dry Systems should achieve a collector temperature of
-84°C or lower when the vacuum pump is operating. If the collector temperature
does not reach the design temperature within 30 minutes, then the refrigeration
module is not functioning properly.
Cascade Systems (those that operate @ -84°C) consist of two vapor-compression
refrigeration systems in series connected together via a heat exchanger. Each
system has a compressor, condenser, capillary tube metering device and an
evaporator. During the operation the evaporator of the “high stage” system
refrigerates the condenser of the “low stage” system. This takes place in the heat
exchanger. The low stage and high stage systems are balanced by design,
however, sometimes that balance can be disrupted during operation and the unit
will not function properly. If the unit is overloaded, or if the unit is defrosted and
restarted in a short time interval, system unbalance is possible. If system
unbalance occurs, turn the Freeze Dryer off and allow the unit to soak at ambient
temperature for 2 to 4 hours.
If any repairs are required on the refrigeration module, contact your local
laboratory supply dealer. If your laboratory supply dealer is unable to provide
assistance, please call Labconco. Repairs should only be undertaken by a
competent refrigeration technician or through an authorized Labconco service
agency.
Product Service 816-333-8811 or 1-800-522-7658
38
CHAPTER 8
MODIFYING YOUR FREEZE
DRYER DISPLAY
Your Freeze Dryer has been carefully calibrated and tested before shipping,
however under certain circumstances it may be necessary to adjust the calibration
of the temperature and/or vacuum display or to modify alarm points. Adjustments
may be necessary if the Freeze Dryer requires service or if it is operating on a line
voltage outside the voltage range at which the alarm was preset. This DOES NOT
mean that the Freeze Dryer may be operated at a voltage outside the ranges shown
in Appendix C.
Temperature and Vacuum Display
An offset may be added to the collector temperature or vacuum display. Use an
accurate traceable known reference to compare the readings.
Temperature offset may be varied ±5°C in 1°C increments. Vacuum offset may be
varied ±0.020 mBar in 0.001 mBar increments.
Line Voltage Alarm Points
The Freeze Dryer is designed to alarm if the line voltage varies more than ±10%
from nominal line voltage. Models rated to operate on 115V are preset to alarm if
the voltage drops below approximately 103V or rises above approximately 127V.
Models rated to operate on 230V will alarm if the voltage drops below
approximately 206V or rises above approximately 254V. Some models are
designed to operate below these ranges (see Appendix C). In these cases it is
permissible to lower the low voltage alarm point to prevent nuisance alarms from
occurring.
Low line limit may be lowered in 5 increments of approximately 2 to 3V
increments for 115V models or disabled.
Low line limit may be lowered in 5 increments of approximately 4 to 5V
increments for 230V models or disabled.
Product Service 1-800-522-7658
Chapter 8: Modifying Your Freeze Dryer Display
39
PRODUCTION MENU
MASTER RESET?
COLLECTOR TEMP OFFSET:
O C
High line limit may be raised in 5 increments of approximately 2 to 3V
increments for 115V models or disabled.
High line limit may be raised in 5 increments of approximately 4 to 5V
increments for 230V models or disabled.
Moisture Sensor Alarm
The moisture sensor alarm may be disabled. It is shipped enabled as the factory
default setting.
Modifying the Display
To enter an offset or to configure the alarms, perform the following steps:
Press and hold MENU button (approximately 10 seconds) until the display shows:
It is recommended that you do NOT perform a Master Reset because all offsets
and configurations that have been entered will be lost. If you must perform a
Master Reset, hold SELECT approximately 5 seconds. This will reset:
Vacuum units to mBar
Vacuum set point to 0 mBar
Temperature units to °C
Refrigeration total hours to 0
Vacuum pump total hours to 0
RS232 transmission rate to 10 seconds
Temperature offset to 0°C
Vacuum offset to 0 mBar
Low line voltage to 0 V offset
High line voltage to 0 V offset
Moisture sensor to enabled
Refrigeration to operate in Manual Mode
The display will default to the main freeze dry display information.
If you do not want to restore all factory settings, but want to adjust offsets, press
MENU briefly and the display will show:
Press SELECT until the appropriate offset is shown. Each number represents 1°C.
Product Service 816-333-8811 or 1-800-522-7658
Chapter 8: Modifying Your Freeze Dryer Display
40
SDC TEMP. OFFSET:
O C
VACUUM OFFSET:
O uBar
LOW LINE OFFSET:
0
HIGH LINE OFFSET:
0
DRYING CHAMBER:
NOT INSTALLED
Press MENU and if the drying chamber has been installed, the display will show:
Press SELECT until the appropriate offset is shown. Each number represents 1°C.
Press MENU and the display will show:
Press SELECT until the appropriate offset is shown. Each number represents
0.001mBar.
Press MENU and the display will show:
Press SELECT until the appropriate offset is shown. If the Freeze Dryer is
operating on nominal 115V, each number represents 2V. If 3 is entered into the
display, the low line alarm will alarm when the voltage drops to approximately
97V, for example 103-(2x3)=97V. If the Freeze Dryer is operating on nominal
230V each number represents 4V (if 3 is entered into the display, the low line
alarm will alarm when the voltage drops to approximately 194V for example
206-(4x3)=194V. There will be no alarm indicated if DISABLED is selected.
Press MENU and the display will show:
Press SELECT until the appropriate offset is shown. If the Freeze Dryer is
operating on nominal 115V, each number represents 2V. If 3 is entered into the
display, the high line alarm will alarm when the voltage raises to approximately
133V, for example. 127+(2x3)=133V. If the Freeze Dryer is operating on
nominal 230V each number represents 4V (if 3 is entered into the display, the low
line alarm will alarm when the voltage drops to approximately 266V, for example.
254+(4x3)=266V. There will be no alarm indicated if DISABLED is selected.
Press MENU and the display will show:
Press and hold SELECT to change the status.
Product Service 816-333-8811 or 1-800-522-7658
Chapter 8: Modifying Your Freeze Dryer Display
41
DEFROST HEATER:
NOT INSTALLED
SHELL FREEZER:
NOT INSTALLED
HOLD SELECT TO CLEAR
ALL TOTAL HOURS?
PURGE VALVE:
NOT INSTALLED
MOISTURE SENSOR:
INSTALLED
PRODUCTION MENU:
MASTER RESET?
Press MENU and the display will show:
Press and hold SELECT to change the status.
Press MENU and the display will show:
Press and hold SELECT to change the status.
Press MENU and the display will show:
Press and hold SELECT if you want to disable the moisture sensor.
Press MENU and the display will show:
Press and hold SELECT to change the status.
Press MENU and the display will show:
If you want to reset Total Hours, press and hold SELECT until an audible beep is
heard. The display will default to the main freeze dry display information.
If you do not want to set Total Hours, press MENU and the display will show:
The routine will repeat by pressing MENU or wait approximately 10 seconds for
the display to default to the main freeze dry information display information.
Product Service 816-333-8811 or 1-800-522-7658
Chapter 8: Modifying Your Freeze Dryer Display
42
PASSWORD:
In order to store offsets in memory, you Must Press and Hold MENU until display
shows:
Wait approximately 10 seconds and the display will default to the main freeze dry
display information. This can be done anytime throughout the Modifying the
Display routine.
Product Service 816-333-8811 or 1-800-522-7658
43
APPENDIX A
FREEZE DRYER COMPONENTS
The following pages list components that are available for your Freeze Dryer. The
parts shown are the most common replacement parts. If other parts are required,
contact Product Service.
Product Service 1-800-522-7658
Appendix A: Freeze Dryer Components
44
Item
Part No.
Description
1
7688500
Lid Gasket
2
7690000
Lid 3 7435800
Back Panel Assembly, 115V
7435900
Back Panel Assembly, 230V
4
7953800
Motor Fan, 115V
7953801
Motor Fan, 230V
5
7591805
Compressor, 115V, 60 Hz
7591806
Compressor, 230V, 50 Hz
7591807
Compressor, 230V, 60 Hz
6
1302300
Switch
7
7426000
Control Panel Label
8
7445500
Printed Circuit Board
9
7728000
Drain Plug
10
1643600
O-Ring
11
7645904
Drain Hose
12
7373450
Vacuum Hose (Not Shown)
13
1336400
Power Cord 115V 60Hz (Not Shown)
1336500
Power Cord 230V 50Hz EU (Not Shown)
1342100
Power Cord 230V 60 Hz (Not Shown)
1332601
Power Cord 230V 50Hz UK (Not Shown)
1332701
Power Cord 230V 50Hz China (Not Shown)
14
7445400
Vacuum Sensor
15
7515300
Temperature Sensor (Not Shown)
16
7419601
Wire Harness, Power, 115V (Not Shown)
7419701
Wire Harness, Power, 230V (Not Shown)
17
7432700
Vacuum Control/Bleed Valve
18
7429700
Moisture Sensor
19
7418400
Wire Harness, Control 115V and 230V (Not Shown)
Model 76705xx
Product Service 816-333-8811 or 1-800-522-7658
45
1
2
17
18
9-10
14
3
4
5
6
11
8
7
Model 76705xx
Appendix A: Freeze Dryer Components
Product Service 816-333-8811 or 1-800-522-7658
Appendix A: Freeze Dryer Components
46
Item
Part No.
Description
1
7395800
Lid Gasket
2
7395801
Lid
3
7435800
Back Panel Assembly, 115V
7435900
Back Panel Assembly, 230V
4
7953800
Motor Fan, 115V
7953801
Motor Fan, 230V
5
7591805
Compressor, 115V, 60 Hz Lo Stage
7591806
Compressor, 230V, 50 Hz Lo Stage
7591807
Compressor, 230V, 60 Hz Lo Stage
7591805
Compressor, 115V, 60 Hz Hi Stage
7591806
Compressor, 230V, 50 Hz Hi Stage
7591807
Compressor, 230V, 60 Hz Hi Stage
7591805
Compressor 115V/60 Hz Single Stage
7591806
Compressor 230V, 50 Hz Single Stage
7591807
Compressor 230V/60 Hz Single Stage
6
1302300
Switch
7
7426000
Control Panel Label
8
7445590
Printed Circuit Board
9
7728000
Drain Plug
10
1643600
O-Ring
11
7645906
Drain Hose
12
7373450
Vacuum Hose (Not Shown)
13
1336400
Power Cord 115V 60Hz (Not Shown)
1336500
Power Cord 230V 50Hz EU (Not Shown)
1342100
Power Cord 230V 60 Hz (Not Shown)
1332601
Power Cord 230V 50Hz UK (Not Shown)
1332701
Power Cord 230V 50Hz China (Not Shown)
14
7445400
Vacuum Sensor (Not Shown)
15
7515300
Temperature Sensor, Collector (Not Shown)
16
7427502
Wire Harness, 115V (Not Shown)
7427503
Wire Harness, 230V (Not Shown)
17
7432700
Vacuum Control/Bleed Valve
18
7429700
Moisture Sensor
19
7418400
Wire Harness, Control, 115V and 230V (Not Shown)
Models 76700xx and 74200xx
Product Service 816-333-8811 or 1-800-522-7658
Appendix A: Freeze Dryer Components
47
Models 76700xx and 74200xx
5
1
2
14
18
9-10
3 4 17 5 7
8
6
Product Service 816-333-8811 or 1-800-522-7658
48
Model 76705xx
APPENDIX B
FREEZE DRYER DIMENSIONS
Product Service 1-800-522-7658
49
Model 76700xx
Appendix B: Freeze Dryer Dimensions
Product Service 816-333-8811 or 1-800-522-7658
Appendix B: Freeze Dryer Dimensions
50
Model 74200xx
Product Service 816-333-8811 or 1-800-522-7658
51
Catalog #
Voltage
Frequency
Phase
Amperage
Nominal
Operating
Range
Nominal
w/o
Vacuum
Pump
Max. with
Vacuum
Pump
7670520
115V
103-127
60 1 6
14
100V
90-110
50 1 6
14
7670521
115V
103-127
60 1 6
14
100V
90-110
50 1 6
14
7670530
230V
198-254
50 1 2.5
7
7670531
230V
198-254
50 1 2.5
7
7670540
230V
187-253
60 1 2.5
7
7670541
230V
187-253
60 1 2.5
7
7670560
230V
198-254
50 1 2.5
7
7670561
230V
198-254
50 1 2.5
7
7670570
230V
198-254
50 1 2.5
7
7670571
230V
198-254
50 1 2.5
7
7670020
115V
103-127
60 1 10.0
16
100V
90-110
50 1 10.0
16
7670021
115V
103-127
60 1 10.0
16
100V
90-110
50 1 10.0
16
7670030
230V
198-254
50 1 5.1
10
7670031
230V
198-254
50 1 5.1
10
7670040
230V
187-253
60 1 5.1
10
7670041
230V
187-253
60 1 5.1
10
7670060
230V
198-254
50 1 5.1
10
7670061
230V
198-254
50 1 5.1
10
7670070
230V
198-254
50 1 5.1
10
7670071
230V
198-254
50 1 5.1
10
APPENDIX C
FREEZEDRYER
SPECIFICATIONS
This Appendix contains technical information about the Freeze Dryer including
electrical specifications and operating conditions.
Electrical Specifications
Product Service 816-333-8811 or 1-800-522-7658
Appendix C: Freeze Dryer Specifications
52
Catalog #
Voltage
Frequency
Phase
Amperage
Nominal
Operating
Range
Nominal
w/o
Vacuum
Pump
Max. with
Vacuum
Pump
7420020
115V
103-127
60 1 10.0
16
100V
90-110
50 1 10.0
16
7420021
115V
103-127
60 1 10.0
16
100V
90-110
50 1 10.0
16
7420030
230V
198-254
50 1 5.1
10
7420031
230V
198-254
50 1 5.1
10
7420040
230V
187-253
60 1 5.1
10
7420041
230V
187-253
60 1 5.1
10
7420060
230V
198-254
50 1 5.1
10
7420061
230V
198-254
50 1 5.1
10
7420070
230V
198-254
50 1 5.1
10
7420071
230V
198-254
50 1 5.1
10
Environmental Conditions
Indoor use only.
Maximum altitude: 6562 feet (2000 meters).
Ambient temperature range: 41° to 104°F (5° to 40°C).
Maximum relative humidity: 80% for temperatures up to 88°F (31°C),
decreasing linearly to 50% relative humidity at 104°F (40°C).
Main supply voltage fluctuations not to exceed ±10% of the nominal
voltage.
Transient overvoltages according to Installation Categories II (Overvoltage
Categories per IEC 1010). Temporary voltage spikes on the AC input line
that may be as high as 1500V for 115V models and 2500V for 230V
models are allowed.
Used in an environment of Pollution degrees 2 (i.e., where normally only
non-conductive atmospheres are present). Occasionally, however, a
temporary conductivity caused by condensation must be expected, in
accordance with IEC 664.
Product Service 816-333-8811 or 1-800-522-7658
53
PART #
DESCRIPTION
1472100
Vacuum Pump
Two stage direct drive pump, 117 liters/minute. 115 VAC,
50/60 Hz, single phase, 4.6 amps. Includes Pump Exhaust Filter
1473400. (Cat. # 1473400)
7739402
Vacuum Pump
Two stage direct drive pump, 117 liters/minute. 230 VAC,
50/60 Hz, single phase, 2.4 amps. Includes Pump Exhaust Filter
1473400. (Cat. # 1473400)
1467700
Vacuum Pump
Two stage direct drive pump, 195 liters/minute. 115 VAC,
50/60 Hz, single phase. Includes Pump Exhaust Filter.
(Cat. # 1473400)
7739403
Vacuum Pump
Two stage direct drive pump, 195 liters/minute. 230 VAC,
50/60 Hz, single phase. Includes Pump Exhaust Filter.
(Cat. # 1473400)
7584000
Vacuum Pump Chemical Resistant Hybrid
Combination two stage rotary vane pump and two stage chemistry
diaphragm pump 98/115 liters/minute. 115 VAC, 50/60 Hz, single
phase.
7584002
Vacuum Pump Chemical Resistant
Combination two stage rotary vane pump and two stage chemistry
diaphragm pump 98/115 liters/minute. 230 VAC, 50/60 Hz, single
phase.
The following accessories are available for the Freeze Dryer.
APPENDIX D
FREEZE DRYER ACCESSORIES
Product Service 816-333-8811 or 1-800-522-7658
Appendix D: Freeze Dryer Accessories
54
PART #
DESCRIPTION
1472200
Pump Inlet Filter
Disposable filter that prevents oil back streaming and protects
vacuum pump from submicron particles. Fits vacuum pumps
1472100 and 7739402.
1473400
Pump Exhaust Filter
Disposable filter that removes visible oil mist and odor from
vacuum pump exhaust. Fits vacuum pumps 1472100 and
Replacement Element, Odor, Pump Exhaust Filter,
package of 5
Fits pump exhaust filter 1473400 or vacuum pumps 1472100 and
7739402.
1988000
Vacuum Pump Oil, 1 Liter
A molecularly distilled hydrocarbon oil with low vapor pressure.
For vacuum pumps 1472100 and 7739402.
7772000
Soda Acid Trap
Secondary trap that prevents migration of corrosive chemicals into
vacuum pump.
7772100
Replacement Media for Soda Acid Trap
7772700
Vacuum Pump Oil, 1 Liter
A mineral oil with low vapor pressure.
For vacuum pumps 7438700, 7438800, 7769600, 7769800.
7670400
Replacement Filter Element
Disposable filter that removes oil mist from the vacuum pump
exhaust.
Fits vacuum pumps 7438700, 7438800, 7469600 and 7469800.
7584300
Vacuum Pump Oil, 1 Liter
For Hybrid vacuum pumps 7584000, 7584002.
Product Service 816-333-8811 or 1-800-522-7658
Appendix D: Freeze Dryer Accessories
55
PART #
DESCRIPTION
7772500
Carbon Solvent Trap
Secondary trap that prevents migration of organic solvents into
vacuum pump.
7772600
Replacement Activated Carbon Media for Carbon Solvent
Trap
7538000
Secondary Vacuum Cold Trap
Provides additional protection for the vacuum pump when
processing low eutectic samples.
9 ¾" high x 7 7/8" diameter, 304 stainless steel with ¾" vacuum
connections. For use with dry ice and solvent. Cools to
approximately –75°C.
7522800
12 Port Drying Chamber
9 ¾" high x 7 7/8" diameter, 304 stainless steel with 12 freeze dry
valves.
7509200
Product Heater
For use in chamber 7522800. Provides 3 heated shelves operating
at 43°C (110°F). 115V
7509201
Product Heater
For use in chamber 7522800. Provides 3 heated shelves operating
at 43°C (110°F). 230V
7443500
Clear Drying Chamber with 8 Valves
For bulk drying or drying in flasks.
7442100
Drying Rack- Unheated
Provides three shelves. For use in chambers 7443500 and
7522800.
7442900
Clear Drying Chamber
For bulk drying.
7522300
20 Port Manifold
10" high x 27 ½" wide x 8 5/8" deep. Type 304 stainless steel
manifold with 20 freeze dry valves.
Product Service 816-333-8811 or 1-800-522-7658
Appendix D: Freeze Dryer Accessories
56
PART #
DESCRIPTION
7868500
48 Port Ampule Manifold
28 5/8" high x 5" wide x 5" deep. Type 304 stainless steel with
48 each ¼" tube stems.
7509600
Sample Valve Kit
Includes neoprene valve body, knob and installation parts.
7522900
16 Port Drying Chamber
13" high x 138" diameter, 304 stainless steel with 16 freeze dry
valves.
7419300
Baffle
Mounts in the collector chamber to increase ice holding capacity
when operating with the vacuum control.
Product Service 816-333-8811 or 1-800-522-7658
Appendix D: Freeze Dryer Accessories
57
Product Service 816-333-8811 or 1-800-522-7658
Appendix D: Freeze Dryer Accessories
58
Product Service 816-333-8811 or 1-800-522-7658
59
Appendix D: Freeze Dryer Accessories
Product Service 816-333-8811 or 1-800-522-7658
Appendix D: Freeze Dryer Accessories
60
Ampules
Product Service 816-333-8811 or 1-800-522-7658
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.