CANOPY HOOD
INSTRUCTION MANUAL
Models 4799000 and 7099000
Product designs are subject to change without notice
© 1994, 2007 Labconco Corporation
4809400 Revision C / ECO E427
Printed in U.S.A.

TABLE OF CONTENTS
Introduction
Components Shipped 1
General Description 2
Performance 2
Installation 3
Safety Precautions 5
Dimensions 6
Warranty & Contacting Labconco 7

PREFACE
Thank you for displaying confidence in us by selecting a Labconco Canopy Hood. Our design
engineers, assemblers and inspectors have utilized their skills and years of experience to ensure
that the new Labconco Canopy Hood meets our high standards of quality and performance.
IMPORTANT NOTICE
This manual should be read carefully by all the end users in order to become familiar with the
operation of the Canopy Hood. Recommendations are made within the manual to help you obtain
maximum performance and life from your product.
We have included sections on performance, initial set up and safety precautions to provide you
with all the tools necessary to achieve maximum performance.
If you have questions or concerns, do not hesitate to call us at 1-800-821-5525 for assistance.

INTRODUCTION
Components Shipped
Carefully check the contents of the carton for damage that might have occurred in transit. Do not
discard the carton or packing material until all components have been checked against the
following component list and the equipment has been installed and tested.
As shipped, the carton should contain the following:
Part Number Description
47990 4 Foot Canopy Hood
70990 6 Foot Canopy Hood
48094 Instruction Manual
1

INTRODUCTION
General Description
The Labconco Canopy Hood is designed for special purpose functions in which heat, steam or
odors are generated. Large or bulky apparatus such as tissue processor, steam baths or open
Kjeldahls can be used in conjunction with the Canopy Hood. The epoxy-coated superstructure has
an integral baffle to increase slot velocities and enhance the overall capture efficiency of the hood.
Performance
All component parts of the Canopy Hood are epoxy-coated except the baffle support brackets
which are stainless steel. The baffle design enhances perimeter air velocities and capture
efficiency. Static pressure loss through the hood shall not exceed .64 inches at 907 CFM and 1.25
inches at 1,360 CFM for the 4-foot unit. Static pressure loss through the 6-foot unit shall not
exceed .90 inches at 1,289 CFM and 1.65 inch at 1,934 CFM.
It is not intended that this hood be used with flammable or explosive atmospheres unless equipped
with explosion proof components and installed in accordance with all pertinent electrical codes.
2

INSTALLATION
The most effective installation is one in which the back is mounted against a wall while the front is
supported by 3/8 inch threaded rod mounted to the ceiling. Wire or cable may be substituted for
the 3/8 inch threaded rod. Please see Figure 1 for recommended installation technique.
Figure 1
Hood Efficiency
The Canopy Hood can be extremely effective when installed correctly. The effectiveness of the
Canopy Hood is dependent directly upon the proximity of the work area to the hood face. The
face velocity and static pressure expressed in the Performance section is based on a theoretical
hood face at 1 foot below the hood. If the height of the working apparatus is below the theoretical
1-foot face, then efficiencies will drop off. Conversely, the hood will become more efficient as the
height of the working apparatus enters inside the 1-foot face.
Therefore, the distance between the work surface and the hood is extremely important. Even when
used as a multipurpose installation, it is important to be within or near the 1-foot face for
maximum efficiency for all uses. Refer to Figure 2.
3

SAFETY PRECAUTIONS
Note: Canopy Hoods are not provided with explosion proof electrical components. Flammable or
explosive atmospheres should not be used in conjunction with this hood.
5

WARRANTY & CONTACTING LABCONCO
Warranty
Labconco provides a warranty on all parts and factory workmanship. The warranty includes areas of
defective material and workmanship, provided such defect results from normal and proper use of the
equipment.
The warranty for all Labconco products will expire one year from date of installation or two years from
date of shipment from Labconco, whichever is sooner, except the following;
• Purifier® Delta® Series Biological Safety Cabinets and PuriCare® Lab Animal Research
Stations carry a three-year warranty from date of installation or four years from date of shipment
from Labconco, whichever is sooner.
• SteamScrubber® & FlaskScrubber® Glassware Washers carry a two-year warranty from date of
installation or three years from date of shipment from Labconco, whichever is sooner.
• Blood Drawing Chairs carry a ten year warranty.
• Carts carry a lifetime warranty.
• Glassware is not warranted from breakage when dropped or mi shan dl ed.
This limited warranty covers parts and labor, but not transportation and insurance charges. In the event of
a warranty claim, contact Labconco Corporation or the dealer who sold you the product. If the cause is
determined to be a manufacturing fault, the dealer or Labconco Corporation will repair or replace all
defective parts to restore the unit to operation. Under no circumstances shall Labconco Corporation be
liable for indirect, consequential, or special damages of any kind. This statement may be altered by a
specific published amendment. No individual has authorization to alter the provisions of this warranty
policy or its amendments. Lamps and filters are not covered by this warranty. Damage due to corrosion or
accidental breakage is not covered.
Returned or Damaged Goods
Do not return goods without the prior authorization from Labconco. Unauthorized returns will not be accepted. If
your shipment was damaged in transit, you must file a claim directly with the freight carrier. Labconco Corporation
and its dealers are not responsible for shipping dam ages.
The United States Interstate Commerce Commission rules require that claims be filed with the delivery carrier within
fifteen (15) days of delivery.
Limitation of Liability
The disposal and/or emission of substances used in connection with this equipment may be governed by various
federal, state, or local regulations. All users of this equipment are required to become familiar with any regulations
that apply in the user’s area concerning the dumping of waste materials in or upon water, land, or air and to comply
with such regulations. Labconco Corporation is held harmless with respect to user’s compliance with such
regulations.
Contacting Labconco Corporation
If you have questions that are not addressed in this manual, or if you need technical assistance, contact Labconco’s
Customer Service Department or Labconco’s Product Service Department at 1-800-821-5525 or 1-816-333-8811,
between the hours of 7:00 a.m. and 6:00 p.m., Central Standard Time.
7