The information contained in this manual and the accompanying products are copyrighted and all rights
reserved by Labconco Corporation. Labconco Corporation reserves the right to make periodic design
changes without obligation to notify any person or entity of such change.
Warranty
Labconco provides a warranty on all parts and factory workmanship. The warranty includes areas
of defective material and workmanship, provided such defect results from normal and proper use of
the equipment.
The warranty for all Labconco products will expire one year from date of installation or two years
from date of shipment from Labconco, whichever is sooner, except the following;
•Purifier® Delta® Series Biological Safety Cabinets and PuriCare® Lab Animal Research
Stations carry a three-year warranty from date of installation or four years from date of
shipment from Labconco, whichever is sooner.
•SteamScrubber® & FlaskScrubber® Glassware Washers carry a two-year warranty from
date of installation or three years from date of shipment from Labconco, whichever is
sooner.
• Blood Drawing Chairs carry a ten year warranty.
• Carts carry a lifetime warranty.
• Glassware is not warranted from breakage when dropped or mi shan dl ed.
This limited warranty covers parts and labor, but not transportation and insurance charges. In the
event of a warranty claim, contact Labconco Corporation or the dealer who sold you the product. If
the cause is determined to be a manufacturing fault, the dealer or Labconco Corporation will repair
or replace all defective parts to restore the unit to operation. Under no circumstances shall
Labconco Corporation be liable for indirect, consequential, or special damages of any kind. This
statement may be altered by a specific published amendment. No individual has authorization to
alter the provisions of this warranty policy or its amendments. Lamps and filters are not covered by
this warranty. Damage due to corrosion or accidental breakage is not covered.
Returned or Damaged Goods
Do not return goods without the prior authorization from Labconco. Unauthorized return s will not be
accepted. If your shipment was damaged in transit, you must file a claim directly with the freight carrier.
Labconco Corporation and its dealers are not responsible for shipping damages.
The United States Interstate Commerce Commission rules require that claims be filed with the delivery
carrier within fifteen (15) days of delivery.
Limitation of Liability
The disposal and/or emission of substances used in connection with this equipment may be governed by
various federal, state, or local regulations. All users of this equipment are required to become familiar with
any regulations that apply in the user’s area concerning the dumping of waste materials in or upon water,
land, or air and to comply with such regulations. Labconco Corporation is held harmless with respect to
user’s compliance with such regulations.
Contacting Labconco Corporation
If you have questions that are not addressed in this manual, or if you need technical assistance, contact
Labconco’s Customer Service Department or Labconco’s Product Service Department at 1-800-821-5525
or 1-816-333-8811, between the hours of 7:00 a.m. and 6:00 p.m., Central Standard Time.
Part #3938100, Rev. C
ECO E161
T
AABBLLEE
T
CHAPTER 1: INTRODUCTION 1
ISO Class 5 Definition 2
About This Manual 2
Typographical Conventions 3
CHAPTER 2: PREREQUISITES 5
Support, Vibration & Movement Requirements 6
Location and Air Current Requirements 6
Airflow Specifications 6
Electrical Requirements 7
Space Requirements 7
CHAPTER 3: GETTING STARTED 8
Unpacking the Clean Bench 9
Installing Vertical Clean Bench on a Supporting
Structure and Work Surface 9
Verifying HEPA Filter Integrity 12
Connecting the Electrical Supply Source to the Vertical Clean
Bench 12
Setting the Downflow Velocity with Speed Control Adjustment 13
Validating the Vertical Clean Bench 13
Sealing the Vertical Clean Bench to the Work Surface 14
CHAPTER 4: PERFORMANCE FEATURES AND
Components 16
Safety Precautions 20
CHAPTER 5: APPROPRIATE APPLICATIONS FOR YOUR
Routine Daily Work Procedures 23
Congratulations on your purchase of a Labconco Purifier®
Vertical Clean Bench. Your enclosure provides product
protection. It is the result of Labconco’s more than 30 years
experience in manufacturing filtered enclosures.
These enclosures will effectively provide product protection from
airborne particulate matter. During operation, room air is drawn
through the prefilter on top, forced through the blower and then
through the HEPA filter and diffuser. The clean air in the work
area in the enclosure meets or exceeds ISO Standard Class 5
conditions defined per ISO Standard 14644-1 and Class 100
conditions as defined by Federal Standard 209E. The enclosure is
set to deliver nominal downward airflow of 40-55 feet per minute.
Each enclosure uses a single HEPA filter, which is rated at least
99.99% efficient for 0.3-micron particles. The Purifier enclosures
offer many unique features to enhance performance. To take full
advantage of them, please acquaint yourself with this manual and
keep it handy for future reference. If you are unfamiliar with how
filtered Vertical Clean Benches operate, please review Chapter 4: Performance Features and Safety Precautions before you begin
working. Even if you are an experienced user, please review
Chapter 5: Using Your Vertical Clean Bench, which describes the
enclosure features.
1
1
Product Service 1-800-522-7658
1
Chapter 1: Introduction
Table 1-1 Selected airborne particulate cleanliness classes for cleanrooms and clean zones.
classification
number (N)
ISO Class 1 10 2
ISO Class 2 100 24 10 4
ISO Class 5 Definition
Airborne particulate cleanliness inside any clean bench or
enclosure is designated by ISO Class 5, which is equivalent to
3520 particles 0.5 µm or larger per cubic meter of air per ISO
Standard 14644-1. ISO Class 5 cleanliness is illustrated in the
table to follow and is equivalent to Class 100 air conditions as
defined by Federal Standard 209E. Class 100 is equal to 100
particles 0.5 µm or larger per cubic foot of air.
ISO
Maximum concentration limits (particles/m3 of air) for particles equal to and
larger than the considered sizes shown below (concentration limits are
calculated in accordance with 3.2 of Standard 14644-1)
0.1 µm 0.2 µm 0.3 µm 0.5 µm 1 µm 5 µm
ISO Class 3 1 000 237 102 35 8
ISO Class 4 10 000 2 370 1 020 352 83
ISO Class 5 100 000 23 700 10 200 3 520 832 29
ISO Class 6 1 000 000 237 000 102 000 35 200 8 320 293
ISO Class 7 352 000 83 200 2 930
ISO Class 8 3 520 000 832 000 29 300
ISO Class 9 35 200 000 8 320 000 293 000
Table 1-1 ISO Classification Number (N)
About This Manual
This manual is designed to help you learn how to install, use, and
maintain your Clean Bench.
Chapter 1: Introduction provides a brief overview of the Clean
Bench, explains the organization of the manual, and defines the
typographical conventions used in the manual.
Chapter 2: Prerequisites explains what you need to do to prepare
your site before you install the Clean Bench. Electrical and service
requirements are discussed.
Chapter 3: Getting Started contains the information you need to
properly unpack, inspect, install, and certify the Clean Bench.
2
Product Service 1-800-522-7658
Chapter 1: Introduction
Chapter 4: Performance Features and Safety Precautions explains
how the Purifier Vertical Clean Bench operates and the appropriate
precautions you should take when using it.
Chapter 5: Appropriate Applications for your Vertical Clean
Bench discusses the basic operation of how to prepare, use and
shut down your Clean Bench.
Chapter 6: Maintaining Your Vertical Clean Bench explains how
to perform routine maintenance on the Clean Bench.
Chapter 7: Accessorizing Your Vertical Clean Bench explains
acceptable modifications to the Clean Bench or how to add
accessories.
Chapter 8: Troubleshooting contains a table of problems you may
encounter while using the Clean Bench including probable causes
and suggested corrective actions.
Appendix A: Components and Replacement Parts contains labeled
diagrams of all of the components of the Clean Benches.
showing all of the dimensions for the Clean Benches.
Appendix C: Vertical Clean Bench Specifications contains the
electrical requirements for clean benches. Wiring diagrams are
also included.
Appendix D: References lists the various resources available that
address clean rooms and Clean Benches.
Typographical Conventions
Recognizing the following typographical conventions will help
you understand and use this manual:
• Book, chapter, and section titles are shown in italic type (e.g.,
Chapter 3: Getting Started).
• Steps required to perform a task are presented in a numbered
format.
• Comments located in the margins provide suggestions,
reminders, and references.
• Critical information is presented in boldface type in paragraphs
that are preceded by the exclamation icon. Failure to comply
Product Service 1-800-522-7658
!
3
Chapter 1: Introduction
)
2'
3'
4'
with the information following an exclamation icon may result
in injury to the user or permanent damage to the enclosure.
• Critical information is presented in boldface type in paragraphs
that are preceded by the wrench icon. Only a trained certifier
or contractor should only perform these operations. Failure to
comply with the information following a wrench icon may
result in injury to the user or permanent damage to your Clean
Bench.
• Important information is presented in capitalized type in
paragraphs that are preceded by the pointer icon. It is
imperative that the information contained in these paragraphs
be thoroughly read and understood by the user.
• A number icon precedes information that is specific to a
particular model of enclosure. The 2' icon indicates the text is
specific to the 2-foot wide model. The 3' icon indicates the text
is specific to the 3-foot model, etc.
4
Product Service 1-800-522-7658
C
HHAAPPTTEERR
C
P
RREERREEQQUUIISSIITTEESS
P
Before you install the Vertical Clean Bench, you need to prepare
your site for installation. You must be certain that the area is level
and of solid construction. In addition, a dedicated source of
electrical power should be located near the installation site to
power the Clean Bench, and other apparatus. Additionally, the
enclosure should be strategically placed in the lab to provide
efficient workflow.
Carefully read this chapter to learn the requirements for your
installation site:
• The support, vibration and movement requirements.
• The location and air current requirements.
• The exhaust and blower requirements.
• The electrical power requirements.
• The space requirements.
Refer to Appendix B: Dimensions for complete Vertical Clean
Bench dimensions.
Refer to Appendix C: Vertical Clean Bench Specifications for
complete Vertical Clean Bench electrical and environmental
conditions, specifications and requirements.
2
2
Product Service 1-800-522-7658
5
Chapter 2: Prerequisites
Support, Vibration and Movement
Requirements
At a minimum, the supporting structure usually consists of a base
cabinet or base stand and chemically-resistant work surface.
Location and Air Current
Requirements
Purifier Vertical Clean Benches have been designed to provide
particulate-free clean air meeting Class 5 conditions by negating
typical cross drafts and turbulence within the opening. However,
as a precautionary safety measure and a higher level of quality
management, it is recommended that the clean bench be placed in
an area away from:
•High traffic areas where walking might cause an air
disturbance or be a nuisance.
•Overhead or wall HVAC diffusers, fans, radiators or other lab
equipment producing air currents.
•Doorways or windows that may be opened.
Airflow Specifications
Purifier Vertical Clean Benches use an integral motorized impeller
to draw room air from the top and through the enclosure. This
room air is then pushed down through the HEPA filter. The
enclosure is factory set to deliver 40-55 fpm downflow velocity
measured 6" below the upper diffuser. The Class 5 HEPA-filtered
air flows across the work area and exits out the front of the
enclosure.
Data for the outflow exhaust volume, downflow velocities and outflow face velocities are listed for each Vertical Clean Bench model
below.
Standard duplex electrical receptacles should be nearby for
connecting the Vertical Clean Bench, or other equipment. The
enclosures include iris pass-throughs to allow electrical cords
through the back of the enclosure without leaving a large hole.
Space Requirements
The dimensions for all the different models are shown in Appendix
B: Dimensions.
Product Service 1-800-522-7658
7
C
HHAAPPTTEERR
C
G
EETTTTIINNGG
G
Now that the site for your Vertical Clean Bench is properly
prepared, you are ready to unpack, inspect, install, and validate
your Clean Bench. Read this chapter to learn how to:
• Unpack and move the Clean Bench.
• Set up the enclosure with the proper supporting structure
and work surface.
• Verify HEPA filters integrity.
• Connect the electrical supply.
• Set the downflow velocity with the speed control
adjustment.
• Validate enclosure airflow.
• Seal the enclosure to the work surface.
Depending upon which model you are installing, you may need
common mechanical and electrical installation tools in addition to
wrenches, ratchets, sockets, a nut driver set, a flat-blade
screwdriver, a Phillips screwdriver, and a carpenter level to
complete the instructions in the chapter.
Each Clean Bench model weighs between 125 to
195 lbs. each (55 to 85 kg). The shipping
!
container allows for lifting with a mechanical lift
truck or floor jack. If you must lift the enclosure
manually, follow safe-lifting guidelines. Do not
lift by the lower air foil.
S
3
3
TTAARRTTEEDD
S
8
Product Service 1-800-522-7658
Chapter 3: Getting Started
Unpacking the Clean Bench
Carefully remove the shrink-wrap or carton on the Clean Bench
and inspect it for damage that may have occurred in transit. If
damaged, notify the delivery carrier immediately and retain the
entire shipment intact for inspection by the carrier.
DO NOT RETURN GOODS WITHOUT THE
PRIOR AUTHORIZATION OF LABCONCO.
)
)
Do not discard the packing material until you have checked all of
the components and tested the Clean Bench.
We recommend that you do not remove the Clean Bench from its
shipping container until it is ready to be placed into its final
location. Move the Clean Bench by placing a flat, low dolly under
the shipping skid, or by using a floor jack.
!
UNAUTHORIZED RETURNS WILL NOT BE
ACCEPTED.
IF CLEAN BENCH WAS DAMAGED IN
TRANSIT, YOU MUST FILE A CLAIM
DIRECTLY WITH THE FREIGHT CARRIER.
LABCONCO CORPORATION AND ITS
DEALERS ARE NOT RESPONSIBLE FOR
SHIPPING DAMAGES.
Do not move the Clean Bench by tilting it onto a
hand truck.
The United States
Interstate Commerce
Commission rules
require that claims be
filed with the delivery
carrier within fifteen (15)
days of delivery.
Installing the Vertical Clean
Bench on a Supporting Structure
and Work Surface
Use caution when lifting or moving the Clean Bench.
When installing the Clean Bench onto a chemical-resistant work
surface or benchtop, ensure that the structure can safely support the
combined weight of the Clean Bench and any related equipment.
Product Service 1-800-522-7658
9
Chapter 3: Getting Started
The work surface should be at least as wide and deep as the
enclosure to properly support it. The front of the Clean Bench
should be aligned within 0.30" (8 mm) of the front of the work
surface. Mounting holes are provided in the Labconco accessory
work surfaces to secure the Clean Bench.
Work Surface Specifications
The work surface should be smooth, rigid and durable, such as a
chemical-resistant epoxy resin. The surface should be non-porous
and resistant to the materials used in conjunction with the Purifier
Vertical Clean Bench. The work surface should also contain a
dished recessed area for containing primary spills.
Vertical Clean Bench Work Surface
1. Level the base cabinet or stand and the work surface. See
Figure 3-1.
2. Position the work surface in its intended location and with
the front of the work surface towards you. (Rear mounting
holes are located close to the rear edge.)
3. Secure the work surface to the base cabinet or stand with a
structural adhesive or silicone sealant.
4. Insert the supplied mounting screws in the four holes.
Allow a minimum of 1/8" clearance under the head of the
screw for positioning the Clean Bench.
5. Place the Clean Bench on the work surface and slide the
rear flange and front air foil flanges under the mounting
screw heads.
6. Tighten the four screws to complete the installation.
10
Product Service 1-800-522-7658
Chapter 3: Getting Started
Figure 3-1
Vertical Clean Bench Installation
Product Service 1-800-522-7658
11
Chapter 3: Getting Started
Verifying HEPA Filter Integrity
The HEPA filter is shipped installed with the gasket on the
downstream side. The HEPA filter is leak checked at Labconco.
A second leak check is recommended before using the enclosure
and at least annually thereafter. Consult your Safety Officer and
Chapter 6 for the HEPA Filter Leak Test. See Figure 4-2 for
HEPA filter location, HEPA filter gasket, and filter clamp bolts.
All seams downstream of the HEPA filter are jacketed by positive
pressure. This intrinsically safe design ensures Class 5 conditions
inside the Vertical Clean Bench.
HEPA Filter Appropriate Use
Filter Size
2'
3'
4'
Part No.
3707900
3707901
3707902
HEPA filters are high-efficiency particulate
air filters having a particulate removable
efficiency of 99.99% for particles with a
diameter of 0.3 micron.
Connecting the Electrical Supply
Source to the Vertical Clean
Bench
115V Models
Simply connect the 115V power cord supplied to the IEC electrical
supply plug on the back of the Clean Bench. If using at 50 Hz
operations, blower performance maximum airflow will be reduced
by 17%.
230V Models
The 230V is shipped without a plug. Install the appropriate plug
for your electrical specifications per local codes.
12
Product Service 1-800-522-7658
Chapter 3: Getting Started
Setting the Downflow Velocity
with the Speed Control
Adjustment
Adjustment of the speed control gives the correct downflow
velocity and is located behind the front panel. The downflow
velocity should be from 30-65 fpm for Class 100 (ISO class 5)
conditions. ISO Class 5 clean air conditions are maximized at a
setting within this range. The Vertical Clean Bench is factory set
at 40-55 fpm downflow velocity. (Consult your Safety Officer for
airflow recommendations for your application). Working at the
downflow velocity appropriate for the application will give the
quietest operation. Downflow velocity measurements are made
using an anemometer. An electric anemometer can be obtained
from your laboratory supply dealer. Downflow velocity
measurements should be taken 6" below the interior upper diffuser.
Using a small Phillips screwdriver, adjust the speed control to give
the required downflow velocity. See Figures 6-1 and 6-2 in
Chapter 6 to locate the speed control. The downflow velocity is
increased by turning the speed control counterclockwise and
clockwise to decrease downflow velocity.
2'
3'
Validating the Vertical Clean
Bench
To determine the actual downflow velocity, airflow velocity
readings are taken. This should be done 6" down from the top of
the interior upper diffuser. The “average downflow velocity” is
achieved by averaging the readings in the downflow test grid for
each model:
The downflow test grid consists of 9 test points, 3 points in 3 rows.
The back row of the grid is 6" from the rear of the work area. The
middle and front rows are 6.57" apart from the back to the front.
The first and last column of the downflow grid are spaced 6" in
from each sidewall with a center column spaced 5.32" away from
the first and last column.
The downflow test grid consists of 12 test points, 4 points in 3
rows. The back row of the 6" from the rear of the work area. The
middle and front rows are 6.57" apart from the back to the front.
The first and last column of the downflow grid are spaced 6" in
from each sidewall with two columns spaced 7.54" away from the
first and last column.
Product Service 1-800-522-7658
13
Chapter 3: Getting Started
4'
The downflow test grid consists of 15 test points, 5 points in 3
rows. The back row of the grid is 6" from the rear of the work area.
The middle and front rows are 6.57" apart from the back to the
front. The first and last column of the downflow grid are spaced 6"
in from each sidewall with three columns spaced 8.66" away from
the first and last column.
Refer to the exhaust specifications table in Chapter 2 for proper
airflow volumes and downflow velocities for your particular
model. A certified technician should complete validation.
The Purifier Vertical Clean Benches have been tested at Labconco
before shipment and provide ISO Class 5 clean air conditions
inside the enclosure. Labconco also performed extensive
performance testing to validate the clean benches for product
protection under ISO Class 5 conditions. For copies of these
validation reports, contact Labconco Customer Service.
Sealing the Vertical Clean Bench
to the Work Surface
When the Vertical Clean Bench has been set in place and wired, it
may be sealed at the work surface to prevent materials from
collecting under the walls. A bead of silicone sealant is
recommended to seal the Vertical Clean Bench to the work
surface.
14
Product Service 1-800-522-7658
C
HHAAPPTTEERR
C
P
EERRFFOORRMMAANNCCE
P
F
EEAATTUURREESS
F
P
RREECCAAUUTTIIOONNSS
P
The Purifier Vertical Clean Bench is designed to provide
product protection only from airborne particulate matter.
During operation, room air is drawn from the top and filtered
by the HEPA filter providing Class 5 conditions inside the
enclosure. The Purifier Vertical Clean Bench delivers Class 5
air conditions when operating at downflow velocities of 30 to
65 feet per minute. The HEPA filter is jacketed by negative
pressure preventing contaminated air inside the enclosure and
ensuring Class 5 air conditions inside the enclosure.
4
4
A
A
E
NNDD
S
AAFFEETTYY
S
The Purifier Vertical Clean Bench HEPA filter is located
behind the front panel. Users are encouraged to routinely
check airflow to satisfy Class 5 air conditions inside the
enclosure or purchase a model with an airflow monitor. For
additional surface decontamination inside the enclosure, users
are encouraged to use the model with the internal UV light and
UV sash closure.
Product Service 1-800-522-7658
15
Chapter 4: Performance Features and Safety Precautions
6
10
9
Located
Behind
4
2
1
Figure 4-1
12
5
8
Located Behind
Control Panel
16
Product Service 1-800-522-7658
19
6
Airflow
Switch with
Adjustment
Screw
15
Optional
UV light
(Not
Shown)
4
1
13
Chapter 4: Performance Features and Safety Precautions
Figure 4-2
Vertical Clean Bench Airflow Diagram
17
Figure 4-2
Vertical Clean Bench Airflow Diagram
7
11
18
Product Service 1-800-522-7658
17
Chapter 4: Performance Features and Safety Precautions
16
Figure 4-3
Optional UV Sash Closure
18
Product Service 1-800-522-7658
Chapter 4: Performance Features and Safety Precautions
1. Aerodynamic Lower Air Foil allows air to sweep the
work surface and promote ISO Class 5 air conditions inside
the enclosure. See Figure 4-1.
2. Ergonomic Slope of 10 degrees provides maximum
visibility and comfort, reduces glare, thereby minimizing
operator fatigue. See Figure 4-1.
3. Internal Depth of 25" provides necessary depth to work
inside the enclosure. See Figure 4-1.
4. Safety Glass Sash with Spring-Loaded Latch has a
wiping seal and features a spring-loaded latch to secure
sash open for loading and cleaning. The sash must be
down for normal operation. See Figure 4-1.
5. Utility Ports with Iris allows electrical cords and data
cords to pass through the back of enclosure without leaving
a large hole. The enclosure ships with solid plugs. Iris
plugs are included with the User’s Manual. See Figure 4-1
and 4-2.
monitors airflow. An audio/visual alarm alerts the user to
low airflow conditions. The Guardian™ Airflow Monitor
is an available option on all Vertical Clean Bench models.
See Figure 4-1 and 4-2.
7. Inherently Safe Impeller has a negative pressure plenum
that surrounds the positive pressure impeller so that if a
leak should occur, the unfiltered air is captured and
refiltered. See Figure 4-2.
8. Speed Control regulates the speed of the motorized
impeller and is used by the certifier to validate and adjust
the downflow velocity. Located behind the front panel.
See Figure 4-2.
9. Fluorescent Lamp is located above the work area, out of
contact with the clean air. A safety glass window beneath
the lamp distributes the light evenly across the work
surface. See Figure 4-1.
10. Filter Pressure Gauge indicates the pressure across the
HEPA filter to help predict HEPA filter loading. See
Figure 4-1.
Product Service 1-800-522-7658
19
Chapter 4: Performance Features and Safety Precautions
11. Filter Clamping Bolt evenly seals the HEPA filter to the
frame of the enclosure. See Figure 4-1 and 4-2.
12. Control Panel. The control panel, which is located above
the sash, contains the filter pressure gauge, control switches
and electronics. See Figure 4-1.
13. HEPA (High Efficiency Particulate Air) Filter is rated to
remove 99.99% of all particles 0.3 micron in size. See
Figure 4-2.
14. Lamp Ballast (Not Shown) for the fluorescent lamp is
located behind the control panel.
15. UV (Ultraviolet) Lamp (Not Shown), found only on some
Clean Bench models, allows the operator to surface
disinfect the work area of the enclosure when it is not in
use. The UV light rays will not penetrate the plane of the
sash and lower air foil.
16. UV Sash Closure(Optional) provides added safety to
prevent access inside the enclosure. See Figure 4-3.
17. Accessory Work Surface is dished and contoured to fit the
dimensions of the Clean Bench to contain spills.
18. Upper Diffuser Screen provides protection for the HEPA
Filter and promotes even distribution of downflow air. See
Figure 4-2.
19. Prefilter (roughing) enhances the life of the HEPA filter
by trapping large particles. It should be replaced quarterly.
Safety Precautions
1. Although the Clean Bench has been engineered to maintain
Class 5 air conditions, caution should always be used while
working. Prior to using the Clean Bench, check to make
sure that the blower is operating and periodically ensure
that air is flowing through the enclosure at its specified
downflow velocity of 30 to 65 fpm. The use of an airflow
monitor is recommended to alert the user to a problem with
airflow.
20
Product Service 1-800-522-7658
Chapter 4: Performance Features and Safety Precautions
2. Use good housekeeping in the Clean Bench at all times.
Clean up spills immediately. Periodically clean Clean
Bench interior.
3. Do not overload the work surface with apparatus or work
material.
4. The Purifier Vertical Clean Bench should never be used
with biohazardous materials, toxins, acids, or radionuclides.
Air from the work area is dispersed directly into the
laboratory from the Clean Bench. A qualified Safety
Officer must carefully assess any risk associated with the
use of the Purifier Vertical Clean Bench.
5. Always work with your hands as far back into the Clean
Bench as possible. Keep all materials and apparatus inside
the lower air foil of the enclosure.
6. Do not work in this Clean Bench without the blower
running.
7. Perchloric acid use in this Clean Bench is prohibited.
8. Radioactive materials are prohibited in this Clean Bench.
Consult your Safety Officer.
9. Avoid cross drafts and limit traffic in front of the Clean
Bench. Air disturbances created may disturb Class 5
conditions.
10. A qualified certification technician should test the Clean
Bench before it is initially used. The Clean Bench should
be validated annually and whenever it is relocated.
11. The sash must remain in the down position while using the
Clean Bench.
12. The HEPA filter provides Class 5 clean air conditions and
only provides product protection inside the Clean Bench.
Do not use the Clean Bench for personnel protection.
13. Never use flammable gases or solvents in the Clean Bench.
Use of an open flame must be avoided in the Clean Bench.
Open flames may disrupt the airflow patterns in the Clean
Bench, burn the HEPA filter and damage the filter’s
adhesive. Gases under high pressure should not be used in
the enclosure as they may disrupt the airflow patterns of the
Clean Bench.
14. HEPA filters are only effective for entrapment of
particulate matter.
Product Service 1-800-522-7658
21
Chapter 4: Performance Features and Safety Precautions
15. The surface of the HEPA filter is fragile and should not be
touched. Care must be taken to avoid puncturing the
HEPA filter during maintenance. If you suspect that a
HEPA filter has been damaged DO NOT use the Clean
Bench; contact a local certification agency or Labconco.
16. The HEPA filter in the enclosure will gradually accumulate
airborne particulate matter from the room. The rate of
accumulation will depend upon the cleanliness of the room
air, and the amount of time the Clean Bench is operating.
With normal usage, the HEPA filters will last two to five
years before requiring replacement.
17. Avoid direct exposure to ultraviolet (UV) radiation;
NEVER work in the enclosure when the UV light is on.
18. Ensure that the Clean Bench is connected to an electrical
service in accordance with local and national electrical
codes. Failure to do so may create a fire or electrical
hazard. Do not remove or service any electrical
components without first disconnecting the filtered
enclosure from electrical service.
19. Increase the downflow velocity by adjusting the speed
control of the motorized impeller or change the HEPA filter
when the downflow velocity falls below acceptable limits
established by your Safety Officer.
20. Ensure only trained operators use the Clean Bench. New
users should review the User’s Manual and become
familiar with the operation of the Clean Bench.
21. Proper operation of the Clean Bench depends largely upon
the enclosure location and the operator’s work habits. The
enclosure should be located away from traffic patterns,
doors, fans, ventilation registers, fume hoods, and any other
air-handling device that could disrupt its airflow patterns.
Consult Chapter 2: Prerequisites and Chapter 3: Getting Started sections of this manual for further details.
22
Product Service 1-800-522-7658
C
HHAAPPTTEERR
C
A
PPPPRROOPPRRIIAATTE
A
A
PPPPLLIICCAATTIIOONNSS FFOORR YYOOUURR
A
V
EERRTTIICCAAL
V
B
EENNCCHH
B
Now that the installation of your Vertical Clean Bench is
completed, you are ready to use it. Read this chapter to learn
about:
1. Routine Daily Work Procedures.
2. Suitable Applications.
3. Appropriate HEPA Filter Applications, Suitability and
Guidelines.
L
5
5
C
C
E
LLEEAAN
N
Routine Daily Work Procedures
Planning
• Arrange for minimal disruptions, such as room traffic or
entry into the room while the Clean Bench is in use.
Start-up
Product Service 1-800-522-7658
• Turn on blower and light. Turn off the UV light, if
equipped.
• Only raise the sash for loading and cleaning.
• Allow the Clean Bench blower to operate unobstructed for
1 minute.
23
Chapter 5: Appropriate Applications For Your Vertical Clean Bench
• Wear a long sleeved lab coat, rubber gloves and protective
eyewear. Wear a protective mask if appropriate. Consult
your Safety Officer for additional personal protective
equipment recommendations.
• Wipe down interior surfaces with a mild household
detergent or disinfectant. DO NOT use abrasive cleaners,
bleach or solvents.
Loading Materials and Equipment
• Load only the materials required for the procedure. Do not
overload the Clean Bench.
• Do not obstruct the upper air diffuser screen.
• Large objects should not be placed close together.
• After loading, wait one minute to purge airborne
contaminants from the work area.
Work Techniques
• Keep all materials inside the lower air foil, and perform all
operations as far to the rear of the work area as possible.
• Segregate all clean and contaminated materials in the work
area.
Final Purging
• Upon completion of work, the Clean Bench should be
allowed to operate for two to three minutes undisturbed, to
purge airborne contaminants from the work area before
shutting down the blower.
• If equipped, turn on the UV light to decontaminate the
work area as determined by your Safety Officer.
Unloading Materials and Equipment
• All open trays, weigh vessels or containers should be
covered before being removed from the Clean Bench.
Suitable Applications
Suitable applications for the Purifier Vertical Clean Benches are
based on guidelines established by the Institute for Environmental
Sciences. Particulates are filtered by the HEPA filter before air is
returned to the room. Biohazardous materials or other hazardous
solids cannot be used with the Purifier Vertical Clean Bench, as air
is directed over the work surface and out of the enclosure. The
Purifier Vertical Clean Bench only provides product protection
and offers no personnel protection whatsoever.
24
Product Service 1-800-522-7658
Chapter 5: Appropriate Applications For Your Vertical Clean Bench
Listed below are suitable applications for HEPA filters.
HEPA Filter Applications,
Suitability and Guidelines
•Procedures traditionally performed on an open bench where a
clean Class 5 air environment is now required.
•The HEPA filtered enclosure provides product protection from
particulate matter. Because air from the Clean Bench enters
from the room during operation, this enclosure should only be
used for operations requiring sample protection from
environmental contamination.
•HEPA filters are only effective for entrapment of particulate
matter. The surface of the HEPA filter is fragile and should
not be touched. Care must be taken to avoid puncturing the
HEPA filter during installation. If you suspect that a HEPA
filter has been damaged, DO NOT use the enclosure. See
Chapter 6 for Replacing the HEPA Filter.
•The HEPA filter will gradually accumulate airborne
particulate matter from the room. The rate of accumulation
will depend upon the cleanliness of the room air, and the
amount of time the enclosure is operating. With normal
usage, the HEPA filters will last two to five years before
requiring replacement.
Product Service 1-800-522-7658
25
C
HHAAPPTTEERR
C
M
V
B
Monitoring airflow and changing the filters are the primary
maintenance required. Certification and recertification is also
reviewed in Chapter 6.
Review this chapter on maintenance for the following:
1. Routine Maintenance.
2. Determination of when to replace the HEPA filters.
3. How to install a new HEPA filter.
4. HEPA filter leak test.
5. Speed control adjustment and setting the downflow velocity.
6. Operating and calibrating the airflow monitors.
7. Initial certification.
8. Re-certification.
9. Fluorescent light replacement.
10. UV light replacement, if applicable.
11. Motorized impeller replacement.
12. Speed control replacement.
AAIINNTTAAIINNIINNG
M
EERRTTIICCAAL
V
EENNCCH
B
H
L
6
6
C
C
Y
G
Y
LLEEAAN
OOUURR
N
26
Product Service 1-800-522-7658
Chapter 6: Maintaining Your Vertical Clean Bench
Routine Maintenance Schedule
Weekly
Monthly (or more often as required)
Annually
• Wipe down the interior surfaces of the enclosure with a
disinfectant or cleaner, depending upon the usage of the
unit and allow to dry.
• Using a damp cloth, clean the exterior surfaces of the
enclosure, particularly the front and top to remove any
accumulated dust.
• Operate the exhaust system, noting the airflow velocity out
the front of the enclosure using a source of visible smoke.
Airflow monitors are recommended for constant
monitoring.
• Determine the actual downflow velocity 6" below the upper
diffuser of the enclosure where the average reading should
be at the specified velocity. (Use calibrated thermal
anemometer or other approved apparatus). Airflow alarms
are recommended for constant monitoring.
• All weekly activities.
• Check downflow velocity. Increase speed control or
change HEPA filter when downflow velocity of the
enclosure drops below the recommended speed for your
facility or if the airflow alarm monitor alerts you. Airflow
monitors are recommended.
• Replace UV lamps if equipped.
• Have the enclosure validated by a qualified certification
technician. See Certification and Recertification to follow
in Chapter 6.
• All monthly activities.
• Change prefilter.
HEPA
When to Replace HEPA Filters
The HEPA filter in the Purifier Vertical Clean Bench gradually
accumulates airborne particulate matter from the room. The rate of
accumulation will depend upon the cleanliness of the room air, and
the amount of operational time the enclosure is operating. In
typical
Product Service 1-800-522-7658
27
Chapter 6: Maintaining Your Vertical Clean Bench
installations and usage, the HEPA filters will last two to five years
before requiring replacement. Replace HEPA filters when
downflow velocity drops below the recommended 30-65 fpm
velocity and the speed control is adjusted to full speed. Replace
HEPA filter if it fails the HEPA Filter Leak Test in Chapter 6.
How to Install a New HEPA Filter
NOTE: Only a qualified certifier should service the HEPA filter.
After the HEPA filter is replaced, the enclosure MUST be
certified. See Figure 6-1.
1. Unplug the enclosure and remove the prefilter located on
top.
2. Remove the front panel by loosening the two screws that
secure it.
3. Using a 9/16" deep socket, loosen the filter clamp bolts
located on top. Refer to Figure 6-1.
4. With the clamp bolts loosened, the HEPA filter-clamping
frame should be clear of filter. Carefully pull the filter
straight out of the enclosure and discard properly.
5. Cover the surface of the new HEPA filter gasket with a
light coating of silicone grease, if desired.
6. Install the new HEPA filter by pushing it straight into the
Vertical Clean Bench, ensuring that it is correctly oriented
with the gasket facing down on the exhaust side of the
enclosure.
CAUTION: The filter clamp bolts should only be tightened
enough to ensure a proper seal at maximum tightness. The
filter gasket should be compressed 50% or less.
7. Tighten the clamp bolts uniformly until the filter gasket is
properly compressed against the frame. Inspect the seal
thoroughly before proceeding.
8. Reinstall the front panel and prefilter.
9. Plug the Vertical Clean Bench in and have it checked for
leaks and downflow velocity.
28
Product Service 1-800-522-7658
HEPA Filter
Gasket Side
Down as
Shown
Gasket
Chapter 6: Maintaining Your Vertical Clean Bench
HEPA Filter
Sealing
Surface
Front Panel
Figure 6-1
HEPA Filter Changing Diagram & Filter Leak Test Diagram
Prefilter
Filter Clamp
Bolt
Speed Control
Adjustment Screw
Product Service 1-800-522-7658
29
Chapter 6: Maintaining Your Vertical Clean Bench
HEPA
HEPA Filter Leak Test
Purpose
After installing the new HEPA filter, the HEPA filter should be
leak checked. This test is performed to determine the integrity of
the HEPA filter, the filter housing, and the filter seal. Leak testing
must be done by a qualified technician with calibrated equipment.
Remove the upper perforated diffuser located inside the enclosure
by using a Phillips screwdriver to unfasten the screws. (See Figure
6-1). The filter passes the leak test at .01% or better. Reference
Leak Testing and Photometer scanning from the Institute of
Environmental Services (IES-RP-CC001.3)
Apparatus
1. An aerosol photometer ATI model 2D, 2E, 2G or
equivalent. Air Techniques Hamilton Associates Inc., 11403
Cron Ridge Dr., Owings Mills, MD 21117
2. One aerosol generator of the Laskin nozzle(s) type. An
aerosol of mineral oil or suitable liquid shall be created by
flowing air through it. The compressed air supplied to the
generator should be adjusted to a pressure of 10± 1 psig. during
operation. Air Techniques Inc. Model TDA-4A or equal. One
nozzle at 10 psig is (67.5 cfm x 100ug/l)/(Vol. of air), For the
2' at 55 fpm or 230 cfm, one nozzle @10 psig is 6750/230 cfm
= 30 ug/l. For the 3' at 55 fpm or 340 cfm, one nozzle @ 10
psig is 20 ug/l. For the 4' at 55 fpm or 460 cfm, one nozzle at
10 psig is 15 ug/l.
3. Mineral oil (Catalog #1491400).
4. Sampling Nozzle, Rectangular 1/2" x 3-1/4", Air Techniques,
Inc.
Procedure
For the ATI 2G Photometer
1. Turn on the photometer and allow it to operate for a minimum
of 5 minutes. Leave the valve in the “CLEAR” setting.
2. Press the “ENTER” keypad. Press the “REF” keypad.
3. The display will display “P1” for approximately 1 second, and
then display a numerical value.
30
Product Service 1-800-522-7658
Chapter 6: Maintaining Your Vertical Clean Bench
4. Using the “^” or “ν” keypads, respectively, increase or
decrease the numerical value until it equals 30 (2'), 20 (3'). 15
(4') for the Purifier Vertical Clean Bench enclosure at 55 fpm
downflow velocity.
5. Press the “ENTER” Keypad. The photometer will scan for 15
seconds, and then the “0” keypad will flash. Press the “Enter”
keypad. The unit will scan for 5 seconds, the display will read
“0000,” and the unit will sound a confirming tone.
6. Set the valve to “DOWNSTREAM.” Place the palm of your
hand over the sampling port of the pistol. There should be a
strong vacuum at this port. If the vacuum is weak, contact Air
Techniques Hamilton Associates.
7. Turn the enclosure on and let it operate for a minimum of 5
minutes.
8. If necessary, adjust the speed control of the enclosure to
maintain the following downflow airflows at 55 fpm; 2'
(230cfm), 3' (340 cfm), 4' (460 cfm.)
9. Position the aerosol generator discharge in the intake of the
prefilter on top of the enclosure.
10. Start the aerosol generator (Pressure to be 10 +/- 1 PSIG).
Ensure that one Laskin nozzle is in the “open” position.
11. Allow the generator to operate for a minimum of 15 seconds.
Scan the downstream exhaust side of the HEPA filter by
passing the sampling nozzle of the gun in slightly overlapping
strokes over the entire surface the filter, with the sampling port
not more than 1 inch from the surface of the filter media. Scan
the entire periphery of the filter and the gasket between the
filter frame and the enclosure frame. Scanning shall be done at
a traverse rate of not more than 2 inches per second.
Acceptance
Aerosol penetration shall not exceed 0.01 percent measured by the
photometer.
Product Service 1-800-522-7658
31
Chapter 6: Maintaining Your Vertical Clean Bench
Setting the Downflow Velocity
with the Speed Control
Adjustment
1. Remove the front panel by loosening the (2) Phillips screws
on top that secure the front panel.
2. The speed control is located on the electrical subassembly
located behind the switched control panel and below the
front panel. See Figure 6-1.
3. Adjust the speed control with a small Phillips screwdriver
by turning the screw counterclockwise to increase blower
speed or clockwise to decrease the blower speed. The
speed control is very sensitive, so proceed with caution.
4. Measure the downflow velocity per the averaging
technique outlined in Chapter 3 and adjust the speed
control slowly for the desired speed. Allow the speed to
stabilize and re-measure the downflow velocity to confirm.
5. Replace the front panel and tighten the screws.
Options
Calibrate and Operate the Airflow
Monitor
Optional Guardian Airflow Monitor
Refer to Figure 6-2 for operation and calibration.
The Guardian Airflow Monitor consists of a circuit board and an
airflow switch. This switch indicates airflow as safe or low. It
does not provide an actual velocity, but a small setscrew in the
back of the sensor can adjust the airflow level that it classifies as
“good/safe” or “low/alert.”
The circuit board provides power to the sensor and also contains a
“safe (green)” and “alert (red)” airflow LED indicators, as well as
a “SILENCE ALARM” button to quiet the audio alarm. When
first powered up, the Printed Circuit Board (PCB) will light both
red and green LED indicators and sound the alarm to indicate it is
working. After 5 seconds, the air monitor will indicate either good
or bad airflow based on what the connected airflow switch detects.
For low airflow, the unit will wait for 10 seconds after low
indications before it sounds both the audio alarm and the red
“alert” LED indicator. If
32
Product Service 1-800-522-7658
Chapter 6: Maintaining Your Vertical Clean Bench
the “SILENCE ALARM” button is pressed, the audio alarm will
be silenced, but the red “alert” LED will remain on. The alarm is
silenced indefinitely unless an airflow change is detected. If safe
airflow is later detected for 10 seconds, the green “safe” LED will
be lit and the “alert” (red) LED will be shut off. At any time the
airflow is safe/good, one can press the SILENCE ALARM test
button and the audio alarm and the red LED will turn on as long as
this button is held down. The PCB has also a two-pin connector
for use as an external output with isolated relay contacts that close
when the red/alert LED is lit (low airflow). These relay contacts
are not affected by the “SILENCE ALARM” button.
The PCB is mounted behind the front panel using standoffs and an
appropriate label is used to highlight the “SILENCE ALARM”
button with clear areas for the red and green LED’s.
By special order, the PCB can be prepared as a factory special with
an additional connector for the following external inputs, and
having the following possible functions:
•External Alarm allows an external signal to sound the alarm
from the building airflow system of a failed condition.
•Alarm Disable allows an external signal to prevent a “low”
airflow alarm from occurring.
•Night Setback allows an external signal to prevent a “low”
airflow alarm from occurring (not any different from Alarm
Disable above other than the terminology.)
•Contact Labconco for ordering information on this special
PCB.
Calibration
1. Ensure the flow switch and alarm circuit board are installed
and operational.
2. Allow the enclosure to operate for at least two minutes.
3. If factory installed, the monitor will alarm at 30±10 fpm
downflow velocity with the average outflow velocity set at
85±10 fpm.
4. To change the factory setting, set the downflow velocity to
the desired alarm condition using the speed control
adjustment procedure outlined in Chapter 6. Once the
alarm condition is set, use a small screwdriver to turn the
adjustment screw on the airflow switch counterclockwise
(facing the screw) until the “low” airflow red LED lights
and the audible flow alarm sounds.
5. Adjust the downflow velocity to the nominal operating
point required. The green “safe” LED should light.
Product Service 1-800-522-7658
33
Chapter 6: Maintaining Your Vertical Clean Bench
6. Over time the HEPA filter will load and eventually slow
the downflow velocity. Once the alarm condition is met,
simply increase the speed control as outlined in Chapter 6
or replace the HEPA filter if the speed control is
maximized.
7. The table below lists typical alarm conditions based on
normal operating conditions. Typical alarm conditions are
set at downflow velocities of 10 to 30 feet per minute
below the normal operating conditions due to supply air
and exhaust air fluctuations, as well as room air cross
drafts. Consult your Safety Officer or Laboratory Manager
for proper operating speeds.
Enclosure Operating
Downflow Speed
55 ± 10 fpm 25-35 fpm
45 ± 10 fpm 25-35 fpm
Alarm Condition Set Point
Speed
34
Product Service 1-800-522-7658
Light
Reflector
Chapter 6: Maintaining Your Vertical Clean Bench
Speed Control
Airflow Monitor LED
Circuit Board
Airflow Switch
with Adjustment
Control Panel
Green “Safe” LED
Silence Alarm Button
with Test/Reset
Red “Alert” LED with
Audible Alarm
Figure 6-2
Guardian Airflow Monitor with Airflow Switch
Product Service 1-800-522-7658
35
Chapter 6: Maintaining Your Vertical Clean Bench
Initial Certification
The Purifier Vertical Clean Bench has been certified at the factory
for an average downflow velocity of 40-55 fpm along with the
HEPA Filter Leak Test. The Purifier Vertical Clean Bench should
be certified for the proper downflow velocity required by your
Safety Officer. It is also a conservative recommendation to
perform the HEPA Filter Leak Test again should there be any
damage caused during transport.
Re-Certification
Under normal operating conditions, the Purifier Vertical Clean
Bench should be recertified at least annually. The certifier should
perform the minimum following tests:
• Downflow Velocity Test
• HEPA Filter Leak Test
In addition, the following tests should also be performed at the
user’s discretion:
• Measure of Line Voltage and Current
• Smoke Test to determine proper airflow patterns
• Lighting Intensity Test (when appropriate)
• Noise Level Test (when appropriate)
• Vibration Test (when appropriate)
Fluorescent Light Replacement
1. Disconnect the power.
2. Locate the small light reflector located under the control
3. Remove the light reflector by removing two Phillips screws
4. Rotate and remove the old fluorescent lamp.
5. Reinstall the new fluorescent lamp and light reflector in
6. Power the unit up and try the new fluorescent lamp.
36
panel.
on the bottom of the light reflector.
reverse order.
Product Service 1-800-522-7658
Chapter 6: Maintaining Your Vertical Clean Bench
UV Light Replacement (if
equipped)
1. Disconnect the power.
2. Locate the UV lamp located inside the enclosure.
3. Rotate and remove the old UV lamp.
4. Reinstall the new UV lamp.
5. Power the unit up and try the new UV lamp.
Motorized Impeller Replacement
The motorized impeller must be replaced as a complete unit.
When the motorized impeller is replaced, the capacitor should also
be replaced. See Appendix A for Replacement Parts Diagram. See
Figure 6-3 for an isometric view of the motorized impeller plenum
assembly. The HEPA filter rests below the motorized impeller
assembly.
1. Unplug the enclosure from the electrical outlet.
2. Remove HEPA filter per the HEPA filter removal procedure
outlined earlier in this chapter.
3. Consult the wiring diagram in Appendix C of the manual and
disconnect all the wires of the motorized impeller. Be sure to
connect wires on the new motor in the same way the old motor
was wired. The 3' and 4' enclosures have two motorized
impellers.
4. Remove four screws in each motorized impeller.
WARNING: High-speed blower. Never operate
!
5. Replace each capacitor with a new one of equal voltage and
capacity.
6. Reassemble each new motorized impeller by reversing the
assembly steps.
7. Perform a HEPA filter leak check.
impeller with housing off.
Product Service 1-800-522-7658
37
Chapter 6: Maintaining Your Vertical Clean Bench
p
p
r
p
Mounting Screws
2' Model
for Motorized
Im
elle
HEPA Filter
Motorized
Im
eller
Wire Access
Hole
Impeller
Mounting
Bracket
Hardware
Front Panel
Capacitor
Mounted here on
Electrical
Subassembly
Located behind
the Control
Panel
38
Wire Access
Holes
Motorized
Im
ellers
3' or 4' Models
Figure 6-3
Motorized Impeller Replacement
Product Service 1-800-522-7658
Chapter 6: Maintaining Your Vertical Clean Bench
Speed Control Replacement
1. Remove the bracket that the speed control is attached to. See
Figures 6-1, and 6-2.
2. Remove the two screws holding the speed control using a
Phillips screwdriver. Refer to Appendix A for Replacement
Parts Diagram.
3. Disconnect all wires leading to the speed control. Connect
wires on new speed control in the same position as the old
speed control.
4. Reassemble to the system in the same position and with the
same screws that were removed earlier.
Product Service 1-800-522-7658
39
C
HHAAPPTTEERR
C
A
CCCCEESSSSOORRIIZZIINNGG AANND
A
M
V
B
There are several ways to accessorize and modify the Vertical
Clean Bench for your individual requirements. These include the
addition of accessory work surfaces, airflow monitors, filters, and
storage cabinets or stands.
OODDIIFFYYIINNG
M
EERRTTIICCAAL
V
EENNCCH
B
1.Work Surfaces
An optional dished work surface is available to attach to the
Vertical Clean Bench.
Dished work surfaces are contoured to fit the dimensions of
the filtered enclosures to contain spills. Epoxy is chemical
resistant. See Figure 3-1 for installation.
Catalog # Description Dimensions (W x D x H)
3909900 Black, 2-foot wide 24" x 29" x 1"
3909901 Black, 3-foot wide 36" x 29" x 1"
3909902 Black, 4-foot wide 48" x 29" x 1"
3909903 Gray, 2-foot wide 24" x 29" x 1"
3909904 Gray, 3-foot wide 36" x 29" x 1"
3909905 Gray, 4-foot wide 48" x 29" x 1"
H
L
7
7
G
Y
C
C
Y
OOUUR
LLEEAAN
R
N
D
40
Product Service 1-800-522-7658
Chapter 7: Accessorizing and Modifying Your Vertical Clean Bench
2.Guardian™ Airflow Monitor
The Guardian Airflow Monitor allows you to continuously
monitor airflow through the Clean Bench. The Guardian
monitor can be placed on any Purifier Vertical Clean
Bench.
Description Order Part #’s Individual Parts Required
Guardian Airflow Monitor
115V or 230V
1 ea. 3935501 Upper Diffuser w/ Air cutout 2'
1 ea. 3935503 Upper Diffuser w/ Air cutout 3'
1 ea. 3935505 Upper Diffuser w/ Air cutout 4'
1885405
1891808
#6-32 x .31” Phillips machine screw to
mount Airflow Monitor Circuit Board
#6-32 x .50” Phillips thread cutting screw to
mount Airflow Sensor
3. Accessory Filters
HEPA Filter
HEPA Filter is 99.99% efficient on particles 0.3 micron.
Width Catalog # Size
2' 3707900 18" x 18" x 3.31"
3' 3707901 30" x 18" x 3.31"
4' 3707902 42" x 18" x 3.31"
Prefilters (Roughing)
A roughing prefilter to extend HEPA filter life.
Filter Catalog # Qty.
2' 3935700 1
3' 3935701 1
4' 3935702 1
Product Service 1-800-522-7658
41
Chapter 7: Accessorizing and Modifying Your Vertical Clean Bench
4. Storage Cabinets and Stands
Size/Description Dual Doors Right Hinge Left Hinge Dual Doors Right Hinge Left Hinge
48" 9902000 - - 9901000 - 36" 9902100 - - 9901100 - 30" 9902200 - - 9901200 - 24" - 9902300 9902400 - 9901300 9901500
18" - - - - 9901400 9901600
12" - - - - - 48" w/Self-Closing Doors 9903000 - - - - 36" w/Self-Closing Doors 9903100 - - - - 30" w/Self-Closing Doors 9903200 - - - - 24" w/Self-Closing Doors - 9903300 9903400 - - 24" ADA - 9906000 9906100 - 9905000 9905200
24" ADA w/Self-Closing
Doors
18" ADA - - - - 9905100 9905300
12" ADA - - - - - -
Size/Description Dual Doors Right Hinge Left Hinge Dual Doors Right Hinge Left Hinge
48" 9900000 - - - - 36" 9900100 - - - - 30" 9900200 - - - - 24" - 9900300 9900600 - - 18" - 9900400 9900700 - 9907000 9907100
12" - 9900500 9900800 - - 48" w/Self-Closing Doors - - - - - 36" w/Self-Closing Doors - - - - - 30" w/Self-Closing Doors - - - - - 24" w/Self-Closing Doors - - - - - 24" ADA - 9904000 9904300 - - 24" ADA w/Self-Closing
Doors
18" ADA - 9904100 9904400 - - 12" ADA - 9904200 9904500 - - -
- 9906200 9906300 - - -
- - - - - -
SOLVENT ACID
STANDARD BASE VACUUM PUMP
Base Stands
3747000 2' Telescoping Base Stand with Fixed Feet 24" w x 25.5" d x 27. 5" to 33.5" h
3746701 3' Telescoping Base Stand with Fixed Feet 36" w x 29.25" d x 27.5" to 33.5" h
3746702 4' Telescoping Base Stand with Fixed Feet 48" w x 29.25" d x 27.5" to 33.5" h
3747010 2' Telescoping Base Stand with 5" Casters 24" w x 25.5" d x 27. 5" to 33.5" h
3746711 3' Telescoping Base Stand with 5" Casters 36" w x 29.25" d x 27.5" to 33.5" h
3746712 4' Telescoping Base Stand with 5" Casters 48" w x 29.25" d x 27.5" to 33.5" h
42
Product Service 1-800-522-7658
5. IV Bar Kits
For pharmacy applications, IV Bar Kits support intravenous
bottles and bags.
3721000 2' IV Bar Kit
3721001 3' IV Bar Kit
3721002 4' IV Bar Kit
Chapter 7: Accessorizing and Modifying Your Vertical Clean Bench
Product Service 1-800-522-7658
43
C
HHAAPPTTEERR
C
T
RROOUUBBLLEESSHHOOOOTTIINNGG AANNDD
T
S
EERRVVIICCEERR
S
L
OOGG
L
Refer to the following table if your Vertical Clean Bench fails to
operate properly. If the suggested corrective actions do not solve
your problem, contact Labconco for additional assistance.
PROBLEM CAUSE CORRECTIVE ACTION
Class 5 air conditions
are not present inside
the enclosure.
External factors are
Improper user
techniques for the
enclosure.
disrupting the
enclosure airflow
patterns.
Enclosure has
improper downflow
velocity.
8
8
O
O
PPEERRAATTIINNGG
See “Certifying the Enclosure” Chapter
3 and “Safety Precautions” Chapter 4
sections in the manual. (Ref. Appendix
D).
See “Location Requirements” Chapter
2, “Certifying the Enclosure” Chapter
3, and “Safety Precautions” Chapter 4
sections of this manual. (Ref. Appendix
D).
Have enclosure certified and check
exhaust system. Check HEPA filters for
loading. Adjust enclosure speed
control. Enclosure should have an
average downflow velocity of 30-65
fpm.
44
Product Service 1-800-522-7658
Chapter 8: Troubleshooting
PROBLEM CAUSE CORRECTIVE ACTION
Blower won’t
operate.
Circuit breaker(s)
Unit not plugged
into outlet.
Plug the enclosure into appropriate
electrical service.
Reset circuit breaker.
or ground fault
Interrupter.
Blower wiring is
Inspect blower wiring.
disconnected.
Blower switch is
Replace switch.
defective.
Motorized impeller
or blower is
Replace motorized impeller or blower.
See Chapter 6.
defective.
Poor internal Class
5 air conditions
HEPA filter
Enclosure sash not
closed.
Close sash.
Replace HEPA filter or increase speed.
clogged.
Blower and lights
won’t operate.
Circuit breaker(s)
Unit not plugged
into outlet.
Plug enclosure into appropriate
electrical service.
Reset or replace circuit breaker.
tripped.
Lights do not work. Lamp not installed
Inspect lamp installation.
properly.
Lamp wiring
Inspect lamp wiring.
disconnected.
Defective lamp. Replace lamp.
Light switch is
Replace light switch.
defective.
Defective electronic
Replace ballast.
ballast.
Airflow monitor
malfunction.
No power.
No lights.
No display.
Power cable to circuit board is
disconnected and needs to be
connected. Sensor cable needs to be
connected. Verify that all airflow
monitor interface cables are connected.
Check fuses on enclosure.
No audible alarm. Alarm has been temporarily silenced
Continuous alarm. Check the downflow velocity of the
Monitor alarms; air
Audible disable
Wrong alarm set
point.
alarm.
way to airflow
monitor sensor is
blocked by insects,
dust or debris.
will not stay
operational.
Airflow monitor was not properly
adjusted. Repeat calibration steps
outlined in this manual in Chapter 6.
Check airflow and calibration of
airflow monitor. See Chapter 6.
enclosure as the airflow of the
system may have changed. If
incorrect, adjust the speed control to
increase downflow velocity. The
HEPA filter may have become
loaded. If downflow velocity is
correct, calibrate the airflow monitor
outlined in this manual in Chapter 6.
Lightly clean the airway with clean
air. Be careful not to touch sensitive
electrical components.
An alarm condition must be
continuously present before the
audible alarm can be silenced. If
flow conditions fluctuate near the
alarm set point, the airflow monitor
will automatically reset it. Action
should be taken to bring the
enclosure airflow into proper
operating parameters or adjust the
alarm set point lower.
46
Product Service 1-800-522-7658
Chapter 8: Troubleshooting
Service Operating Record Log
Customer Name Model Number
Date Installed Serial Number
Unit Location
Date Comments
Product Service 1-800-522-7658
47
A
PPPPEENNDDIIXX
A
V
EERRTTIICCAAL
V
B
EENNCCH
B
A
NND
A
P
AARRTTSS
P
The components that are available for your Vertical Clean Bench
are listed. The parts shown are the most commonly requested. If
other parts are required, please contact Product Service.
D
R
R
C
H
C
EEPPLLAACCEEMMEENNT
A
L
C
A
C
LLEEAAN
N
OOMMPPOONNEENNTTS
T
S
48
Product Service 1-800-522-7658
Appendix A: Vertical Clean Bench Components and Replacement Parts
Item Qty. Part Number Description
1 1 3911000 Glass, Side
2A 1 3911900 Glass, Sash 2'
2B 1 3911901 Glass, Sash 3'
2C 1 3911902 Glass, Sash 4'
3A 1 3938900 Sash Assembly, 2'
3B 1 3938901 Sash Assembly, 3'
3C 1 3938902 Sash Assembly, 4'
4A 1 3905600 Lower Air Foil, 2'
4B 1 3905601 Lower Air Foil, 3'
4C 1 3905602 Lower Air Foil, 4'
5A 1 3907400 Bracket, Air Foil R.H.
5B 1 3907401 Bracket, Air Foil L.H.
6 2 1554500 Stop, Sash Bumper.
7A 1 6916500 Latch, Sash
7B 1 3906700 Bracket, Latch
7C 1 1927403 Spring, Compression Latch
7D 2 1893206 Screw, 8-32 x .38" Lg. Phillips Pan S.S. Type F
9A 2 1889316 Screw, 10-24 x 1.00 PH Phil S.S.
9B 2 1912108 Washer, Plastic .194 ID. x .380 OD
9C 2 7868402 Bushing, Spacer .31 x .63
10 1.9 Ft 6913700 Wiper, Sash Seal
11A 2 1936800 Bushing, Heyco closed 1.50 Dia.
11B 2 1934601 Bushing, Heyco w/ Flex Shutter 1.50 Dia.
12A 2 3940500 Hinge, outer
12B 2 3940501 Hinge, Inner
12C 4 1591606 Rubber Bumper
12D 2 3939600 2' and 4' Sash Closure Door
12E 2 3939601 3' Sash Closure Door
12F 2 3940700 Extension Bi-Fold Door 3'
12G 2 3940701 Extension, Bi-Fold Door 4'
13A 1 3907700 Reflector, Light 2'
13B 1 3907701 Reflector, Light 3'
13C 1 3907702 Reflector, Light 4'
13D 2 1889308 Screw, Mach. 10-24 x .50 PH Phillips
14A 1 3708300 Lamp, Fluorescent 2' Enclosure (F15T8-SP35)
14B 1 9721902 Lamp, Fluorescent 3' Enclosure (F17T8)
14C 1 9721901 Lamp, Fluorescent 4' Enclosure (F25T8)
15A 1 1270100 Lamp, UV 2' or 3' Enclosure (G15T8)
15B 1 1271300 Lamp, UV 4' enclosure (G30T8)
16A 1 3911700 Bracket, Right Reflector Support
16B 1 3911701 Bracket, Left Reflector Support
16C 4 1893108 Screw, #8 x 1/2 AB Hex Washer Head
Product Service 1-800-522-7658
49
Appendix A: Vertical Clean Bench Components and Replacement Parts
See the following dimensions for the Purifier Vertical
Clean Benches.
X
B
B
S
Product Service 1-800-522-7658
53
Appendix B: Dimensions
Figure B-1
Purifier Vertical Clean Bench
54
Product Service 1-800-522-7658
A
PPPPEENNDDIIXX
A
V
EERRTTIICCAAL
V
B
EENNCCH
B
S
PPEECCIIFFIICCAATTIIOONNSS
S
This Appendix contains technical information about all
the Purifier Vertical Clean Benches including electrical
specifications, environmental operating conditions and
wiring diagrams.
•5 Amps, 115V, 60 Hz or 3 Amps, 230V, 50/60 Hz,
Purifier Vertical Clean Bench.
Environmental Conditions
• Indoor use only.
• Maximum altitude: 6562 feet (2000 meters).
• Ambient temperature range: 41° to 104°F (5° to
40°C).
• Maximum relative humidity: 80% for
temperatures up to 88°F (31°C), decreasing
linearly to 50% relative humidity at 104°F
(40°C).
• Main supply voltage fluctuations not to exceed
±10% of the nominal voltage.
• Transient over-voltages according to Installation
Categories II (Over-voltage Categories per IEC
1010). Temporary voltage spikes on the AC
input line that may be as high as 1500V for
115V models and 2500V for 230V models are
allowed.
• Used in an environment of Pollution degrees 2
(i.e., where normally only non-conductive
atmospheres are present). Occasionally,
however, a temporary conductivity caused by
condensation must be expected, in accordance
with IEC 664.
H
L
C
C
C
C
LLEEAAN
N
Product Service 1-800-522-7658
55
Appendix C: Vertical Clean Bench Specifications
C-1
56
Product Service 1-800-522-7658
Appendix C: Vertical Clean Bench Specifications
C-2
Product Service 1-800-522-7658
57
Appendix C: Vertical Clean Bench Specifications
C-3
58
Product Service 1-800-522-7658
Appendix C: Vertical Clean Bench Specifications
C-4
Product Service 1-800-522-7658
59
A
PPPPEENNDDIIXX
A
R
EEFFEERREENNCCEES
R
Many excellent reference texts and booklets are
currently available. The following is a brief listing:
Bryan, D., and R.C. Marback. 1984. Laminar-airflow
equipment certification: What the Pharmacist needs to Know. American Journal of Hospital Pharmacy. 41.1343-
1348.
General Services Administration. 1988. FED-STD-209E
Federal Standard – Clean Room and Work Station Requirements, Controlled Environment. Washington D.C.
IES Recommended Practice: IES-RP-CC-001-86.
Recommended Practice for HEPA Filters. Mt. Prospect, IL;
Institute for Environmental Sciences. 1986.
IES Recommended Practice: IES-RP-CC-001-96. Laminar Flow Clean Air Devices. Mt. Prospect, IL: Institute for
Environmental Sciences. 1986.
IES Document: IES-CC-011-85-T. A Glossary of Terms and Definitions Related to Contamination Control. Mt. Prospect,
IL: Institute for Environmental Sciences. 1985.
ISO 14644-1:1999 Cleanrooms and Associated Controlled
Environment International Organization for
Standardization.
D
D
S
60
Product Service 1-800-522-7658
DECLARATION OF CONFORMITY
Application Council Directive(s): 73/23/EEC, 89/336/EEC
Standard(s) to which conformity is declared: EN61010, EN55022,
EN50082-1
Manufacturer’s Name: Labconco Corporation
Manufacturer’s Address: 8811 Prospect Avenue
Kansas City, MO 64132 USA
Importer’s Name: See Shipping/Customs Documents*
Importer’s Address: See Shipping/Customs Documents for your
equipment
Type of Equipment: Purifier Vertical Clean Benches
3970200 Series 2' Purifier Vertical Clean Bench
3970300 Series 3' Purifier Vertical Clean Bench
3970400 Series 4' Purifier Vertical Clean Bench
Serial No.: Various – See Individual Declaration
Year of Manufacture: 2004 and subsequent
I, the undersigned, hereby declare that the equipment specified above
conforms to the above Directive(s) and Standard(s).
See individual Declaration of Conformity, which will be signed by the
importer for your country.