Labconco 34000-00,34040-00,34050-00,34000-01,34040-01,34050-01 WAM MODULE D’ACQUISITION SANS FIL MANUEL D'UTILISATION RÉF. 9515-174-50-FRE Rév. B1

Page 1
Protector Evidence Drying
Models
34000 Series 3' 34040 Series 4' 34050 Series 5'
Cabinets
Page 2
Copyright © 2018 Labconco Corporation. The information contained in this manual and the accompanying
Warranty
Labconco Corporation provides a warranty to the original buyer for the repair or replacement of parts and reasonable labor as a result of normal and proper use of the equipment with compatible chemicals. Broken glassware and maintenance items, such as filters, gaskets, light bulbs, finishes and lubrication are not warranted. Excluded from warranty are products with improper installation, erratic electrical or utility supply, unauthorized repair and products used with incompatible chemicals.
The warranty for Protector® Evidence Drying Cabinets will expire one year from date of installation or two years from date of shipment from Labconco, whichever is sooner. Warranty is non-transferable and only applies to the owner (organization) of record.
Buyer is exclusively responsible for the set-up, installation, verification, decontamination or calibration of equipment. This limited warranty covers parts and labor, but not transportation and insurance charges. If the failure is determined to be covered under this warranty, the dealer or Labconco Corporation will authorize repair or replacement of all defective parts to restore the unit to operation. Repairs may be completed by 3rd party service agents approved by Labconco Corporation. Labconco Corporation reserves the rights to limit
this warranty based on a service agent’s travel, working hours, the site’s entry restrictions and unobstructed
access to serviceable components of the product.
Under no circumstances shall Labconco Corporation be liable for indirect, consequential, or special damages of any kind. This warranty is exclusive and in lieu of all other warranties whether oral, or implied.
products are copyrighted and all rights reserved by Labconco Corporation. Labconco Corporation reserves the right to make periodic design changes without obligation to notify any person or entity of such change.
Returned or Damaged Goods
Do not return goods without the prior authorization from Labconco. Unauthorized returns will not be accepted. If your shipment was damaged in transit, you must file a claim directly with the freight carrier. Labconco Corporation and its dealers are not responsible for shipping damages.
The United States Interstate Commerce Commission rules require that claims be filed with the delivery carrier within fifteen (15) days of delivery.
Limitation of Liability
The disposal and/or emission of substances used in connection with this equipment may be governed by various federal, state, or local regulations. All users of this equipment are required to become familiar with any regulations that apply in the user’s area concerning the dumping of waste materials in or upon water, land, or air and to comply with such regulations. Labconco Corporation is held harmless with respect to user’s compliance with such regulations.
Contacting Labconco Corporation
If you have questions that are not addressed in this manual, or if you need technical assistance, contact Labconco’s Customer Service Department or Labconco’s Product Service Department at 1-800-821-5525 or 1-816-333-8811, between the hours of 7:30 a.m. and 5:30 p.m., Central Standard Time.
Part #3898400, Rev. G
ECO L479
Page 3
T
T
AABBLLEE
O
O
FF
C
C
OONNTTEENNTTSS
CHAPTER 1: INTRODUCTION 1
CHAPTER 2: PREREQUISITES 2
Space Requirements 2 Clearance 2 Location Requirements 3 Exhaust Requirements 3 Electrical Requirements 3 Drain Requirements 3 Water Requirements (Models 34000-10 thru -14, 34040-10 thru -14,
34050-10 thru -14 only) 4
CHAPTER 3: GETTING STARTED 5 Unpacking the Cabinet 5 Moving the Cabinet 6 Attaching the Cabinet to the Wall 7 Drain Options-UV Lamp Models (Models 34000-00 thru -04,
34040-00 thru -04, 34050-00 thru -04 only) 9
Drain Options-Wash Down Models (Models 34000-10 thru -14,
34040-10 thru -14, 34050-10 thru -14 only) 10
Water Supply Connection-Wash Down Models (Models 34000-10
thru -14, 34040-10 thru -14, 34050-10 thru -14 only) 10 Preparing the Evidence Drying Cabinet for Operation 11 Initial Operation 12
CHAPTER 4: PERFORMANCE FEATURES AND SAFETY
Directional Airflow 14 Optional Evidence Prefilters 15
Prefilters 15 HEPA Filters 15 Blower 16 Cabinet Air Intake and Exhaust 16 Ultraviolet (UV) Lamp (Models 34000 -00 thru -04 only) 16 Safety Precautions 17
PRECAUTIONS 13
Page 4
CHAPTER 5: USING THE CABINET 19
UV Lamp Models (Models 34000-00 thru -04, 34040-00 thru -04,
34050-00 thru -04 only) 19 Fan/UV Light Switch 19 UV Security Lock 19
Filter System Light 19 Shelf Operation 20 Drying in the Cabinet 20
Wash Down Models (Models 34000-10 thru -14, 34040-10 thru -14,
34050-10 thru -14 only) 22 Fan/Drain Pump Switch 22 Filter System Light 22 Shelf Operation 23 Drying in the Cabinet 23
CHAPTER 6: MAINTAINING THE CABINET 26 Routine Maintenance Schedule 26 Service Operations 27 Door Evidence Prefilter Replacement 27 Door Prefilter Replacement 27
Cabinet Evidence Prefilter Replacement 28 Cabinet Prefilter Replacement 28 Granular Activated Carbon and HEPA Filter Replacement 29 Changing the UV Lamp (Models 34000-00 thru -04, 34040-00
thru -04, 34050-00 thru -04) 30 Resetting a Circuit Breaker 31 Setting the Speed Control 31
CHAPTER 7: TROUBLESHOOTING 32
APPENDIX A: COMPONENTS 34
APPENDIX B: DIMENSIONS 35
APPENDIX C: SPECIFICATIONS 36
APPENDIX D: ACCESSORIES 37
Page 5
1
CChhaapptteerr 11:: IInnttrroodduuccttiioonn
Congratulations on the purchase of a Labconco Protector Evidence Drying Cabinet. The cabinet is designed to protect you, and the room environment from biohazardous aerosols and offensive odors while drying evidence. The Protector Series cabinet is the result of years of experience in manufacturing laboratory equipment, and users like you suggested many of its features to us.
The Evidence Drying cabinet models offer many unique features to enhance performance and flexibility. To take full advantage of them, please acquaint yourself with this manual and keep it handy for future reference. If you are unfamiliar with how evidence drying cabinets operate, please review Chapter 4: Performance Features and Safety Precautions before you begin using the cabinet. Even if you are an experienced cabinet user, please review Chapter 5: Using the Cabinet; it describes the cabinet’s features so that you can use it efficiently.
This manual and other technical information is available in PDF format at our website: www.labconco.com.
Product Service 1-800-522-7658
Page 6
2
CChhaapptteerr 22:: PPrreerreeqquuiissiitteess
Before you install the evidence drying cabinet, you need to plan the site for installation. Examine the location where you intend to install the cabinet. You must be certain that the area is level and of solid construction. In addition, a source of electrical power must be located near the installation site.
Carefully read this chapter to learn:
Location requirements. Electrical power requirements. Exhaust requirements. Service utility requirements. Space requirements. Drain requirements. Water Supply Requirements (Models #3400010 thru -14, 3404010 thru -14,
and 3405010 thru -14 Only)
Refer to Appendix C: Specifications, for complete cabinet electrical and environmental conditions, specifications and requirements.
Space Requirements
The dimensions for the Evidence Drying Cabinet are shown in Appendix B: Dimensions.
Clearance
A minimum clearance of at least 6 inches (150 mm) is suggested on the top of the cabinet for service. There should be 36 inches (914 mm) clearance at the front of the cabinet to allow the door to swing open without hitting any obstructions.
Product Service 1-800-522-7658
Page 7
Chapter 2: Prerequisites
3
Model #
Requirements
34000-00 & -10 34040-00 & -10 34050-00 & -10
115 VAC, 60 Hz, 6 Amps
34000-01 thru -04 & -11 thru -14 34040-01 thru -04 & -11 thru -14 34050-01 thru -04 & -11 thru -14
230 VAC, 50/60 Hz, 3 Amps
Table 2-1
Location Requirements
Note: The 3' Cabinet is equipped with locking casters to ease installation and maintenance. It should be located on a flat floor for proper cabinet alignment and door operation.
Exhaust Requirements
If you intend to connect the cabinet to the optional Exhaust Connection Kit, first examine the location to ensure that it accommodates the cabinet’s exhaust duct.
The area directly above the cabinet’s exhaust port should be clear of structural
elements, water and utility lines, or other fixed obstructions. There should be enough clearance to accommodate a 5-inch diameter duct. For further information about the cabinet’s exhaust system requirements, please refer to Appendix D: Accessories.
Electrical Requirements
The Evidence Drying Cabinet models have the following electrical requirements:
Note: An outlet with a circuit breaker rated at 10 amps for 115 volt models (5 amps for 230 volt models) should be located as close as possible to the right rear side of the cabinet. Power cords supplied with the Evidence Drying cabinet are 3 meters (10 feet) long.
Drain Requirements
If you intend to connect the optional drain kit, or if you have Model # 34000-10 thru -14, 34040-10 thru -14, or 34050-10 thru -14 you will need
to have the Evidence Drying Cabinet within 6 feet (1.8m) of a suitable floor drain.
Note: The floor drain must be located below the level of the drain hose exiting the cabinet for all models, even the drain pump-equipped models ending -10 thru -14. This will allow complete drainage of the system during wash down operations. The cabinet and floor drain should be located in an area not subject to freezing temperatures.
Product Service 1-800-522-7658
Page 8
Chapter 2: Prerequisites
4
Water Requirements (Model # 34000-10 thru -14, 34040-10
thru -14 & 34050-10 thru -14 Only)
If you intend to connect the water supply to wash down the cabinet, locate the cabinet within 6 feet (1.8 m) of a domestic water line, equipped with a water shut off valve ending in a 3/4" male garden hose connector. Water system pressure should be 10-40 PSIG (0.7-2.8 Bar).
Product Service 1-800-522-7658
Page 9
5
CChhaapptteerr 33:: GGeettttiinngg SSttaarrtteedd
Now that the installation site is properly planned, you are ready to inspect and install the Protector Evidence Drying Cabinet. This chapter covers how to:
Unpack and move the cabinet. Install the cabinet. Connect the electrical supply source.
Tools required for installation of the cabinet include two 1/2" wrenches, a flat­blade screwdriver, a #2 Phillips screwdriver, scissors, or a box knife.
Note: The 3' cabinet weighs approximately 300 lbs. (136 kg), 4' cabinet 400 lbs. (182 kg), and 5' cabinet 510 lbs. (232 kg). The shipping pallet allows for lifting with a mechanical lift truck or floor jack. If you must lift the product manually, use at least six (6) persons and follow safe-lifting guidelines. All of the cabinets
are very heavy, use a mechanical lift and safe moving practices at all times.
Unpacking the Cabinet
Carefully remove the outer carton and inspect the cabinet for damage that may have occurred in transit. If the cabinet is damaged, notify the delivery carrier immediately and retain the entire shipment intact for inspection by the carrier.
Note: United States Interstate Commerce Commission rules require that claims be filed with the delivery carrier within fifteen (15) days of delivery.
Do not return goods without the prior authorization of Labconco. Unauthorized returns will not be accepted.
If the cabinet was damaged in transit, you must file a claim directly with the freight carrier. Labconco Corporation and its dealers are not responsible for shipping damages.
Do not discard the carton or packing material for the cabinet until all of the components have been checked, installed and tested.
Product Service 1-800-522-7658
Page 10
Chapter 3: Getting Started
6
3' Cabinet Sizes
The 3' Cabinet is secured to the pallet using four transport brackets, two on each side. To remove the brackets, do the following:
1. Using the 1/2" wrenches, loosen and remove the bolts and locknuts that
secure the brackets to the pallet.
2. Using a 1/2" wrench, loosen and remove the nuts and bolts securing the
transport brackets to the bottom of the cabinet.
3. Discard the brackets, shipping nuts and bolts.
The cabinet is now ready to be removed from the pallet.
NOTE: A ramp has been provided attached to the wood pallet. Follow the instructions on the face of the cabinet for safe removal.
After removing the cabinet from the pallet, carefully remove plastic wrapping around the cabinet, and open the cabinet door. Locate the two transport brackets attached to the face of the cabinet liner. Using a #2 Phillips screwdriver, remove the screws securing the brackets, discard the brackets, and reinstall the screws.
4' & 5' Cabinet Sizes
The 4' & 5' Cabinets are secured to the pallet using four transport bolts securing the cabinet to the pallet. To remove the bolts, do the following:
1. Using the 1/2" wrench, loosen and remove the bolts that secure the cabinet
to the pallet.
2. Discard the bolts and washers.
The cabinet is now ready to be removed from the pallet. After removing the cabinet from the pallet, install the four Leveling Legs into the
bottom of the cabinet. Leveling Legs are located in the lower cabinet compartment. Then, carefully remove the plastic wrapping around the cabinet, and open the cabinet doors. Locate the two transport brackets attached to the face of the cabinet liner. Using a #2 Phillips screwdriver, remove the screws securing the brackets, discard the brackets, and reinstall the screws.
Moving the Cabinet
When moving the cabinet, DO NOT PUSH ON THE FRONT OR BACK OF THE UNIT. Move the cabinet by pushing or pulling it sideways. This will
minimize the possibility of the unit overturning when in motion.
Product Service 1-800-522-7658
Page 11
Chapter 3: Getting Started
7
Attaching the Cabinet to the Wall
WARNING: The evidence drying cabinet must be secured to a wall stud, or other suitable wall structure, using the included strap(s) and bracket(s). This will minimize the possibility of the unit overturning when in use.
1. Remove the wall bracket secured to the strap, located on the back of the
unit on top (or located in the lower cabinet compartment).
2. Locate a stud inside the wall directly behind the cabinet’s installation
location.
NOTE: The screws provided with the wall bracket are intended to be used when attaching the bracket to a wooden wall stud. If your wall is made of different materials (cinder block, concrete, etc.), you will need to attach the wall bracket to the wall using an anchor or fastener approved for that type of construction.
3. Position the bracket, centered on the stud, with the top of the strap slot 79
inches above the floor. Mark the top of each screw hole, as shown in Figure 3-1.
4. Using a 1/8" (3 mm) drill, drill pilot holes at both marked locations to a
depth of at least 1-½ inches (38 mm).
5. Using a #3 Phillips screwdriver, tighten both bracket screws until the
bracket is secured to the wall.
6. Thread the strap through both bracket slots (one strap per slot on 4' & 5'
cabinets), and then through both slots located at the top corners on the rear of the cabinet (each strap goes through only one slot on cabinet on 4' & 5' cabinets).
7. Push the cabinet back into its final location, and then tighten the strap until
there is minimal slack in the strap. Do NOT overtighten strap(s).
Product Service 1-800-522-7658
Page 12
Chapter 3: Getting Started
8
Figure 3-1
79.0" (2400 cm) from the floor to the top of the strap slot
Secure bracket to wall stud
Product Service 1-800-522-7658
Page 13
Chapter 3: Getting Started
9
Drain Options - UV Lamp Models (Model # 34000-00 thru -04,
34040-00 thru -04 & 34050-00 thru -04 only)
NOTE: These models can either drain to a bucket, already located in the lower storage area, or the bucket can be removed, and an optional Drain Kit (#3928200) can be installed. The optional drain kit allows the cabinet to drain to a nearby floor drain.
On units equipped with the drain bucket: NOTE: The drain bucket had a working capacity of approximately 1.5
gallons. Periodically monitor the liquid level in the bucket during decontamination or washdown operations.
1. Access the bucket by opening the bottom storage area door on the front of
the cabinet.
2. Before using, ensure the bucket is in its correct position under the drain.
To install the Drain Kit (#3928200):
NOTE: The drain line should be routed to a floor drain with an air gap only. NEVER directly connect the drain line to a sewer or drain system, or use a stand pipe, to prevent backflow into the cabinet.
1. Remove the bucket from the lower storage area and save for possible reuse.
2. Connect the drain hose assembly to the tail pipe of the drain (on the bottom
of the drain assembly). Push the drain assembly up onto the stand pipe until it is fully seated.
3. Hand-tighten the compression ring of the drain assembly to the tail pipe
until it is secure.
4. Route the drain hose through the appropriate hole in the bottom rear panel
of the cabinet.
5. Route the drain hose to the floor drain and secure the end of the hose to
prevent it from moving.
6. Pour water into the cabinet drain, and inspect the entire drain system for
leaks. Tighten any leaking fittings.
Product Service 1-800-522-7658
Page 14
Chapter 3: Getting Started
10
Drain Options - Wash Down Models (Model # 34000-10
thru -14, 34040-10 thru -14 & 34050-10 thru -14 only)
NOTE: This model is designed to have its drain hose routed to a nearby floor drain.
To install the drain hose assembly:
NOTE: The drain line should be routed to a floor drain with an air gap only. NEVER directly connect the drain line to a sewer or drain system, or use a stand pipe, to prevent backflow into the cabinet.
1. Remove the packing materials securing the hose to the back of the cabinet.
2. Route the drain hose to the floor drain and secure the end of the hose to
prevent it from moving.
Water Supply Connection - Wash Down Models
(Model # 34000-10 thru -14, 34040-10 thru -14 & 34050-10 thru -14 only)
NOTE: This model has a cabinet water shut off valve located behind the lower front door. The valve should be in the “off” position whenever the unit is not being washed down.
1. Connect the water supply hose (the stainless steel braided hose) to a
domestic water line, equipped with a shut off valve, and ending in a 3/4" male garden hose connector.
2. Ensure the water supply valve is in the “off” position.
3. Slowly open the water line shut off valve.
4. Check the cabinet’s water supply hose for leaks.
5. Slowly open the cabinet shut off valve, and check again for leaks.
6. Set the spray gun nozzle to “shower” and slowly pull the trigger.
7. When water flows steadily from the gun, the cabinet shut off valve can then
be turned to the “off” position.
Product Service 1-800-522-7658
Page 15
Chapter 3: Getting Started
11
Preparing the Evidence Drying Cabinet for Operation
After the evidence drying cabinet has been removed from its pallet and positioned in its final location, you must:
1. Remove the tie and/or tape securing the door, the lower storage area door,
and the filters on top of the unit.
2. Ensure that the transport brackets have been removed from the front of the
cabinet liner.
Note: The following are located in a box in the lower storage space:
User’s Manual Power Cord Product Registration Card UV Security Lock Key(s) (model #s 34000-00 thru -04,
34040-00 thru -04 & 34050-00 thru -04 only)
Drain Strainer Cup & Grid Rubber Drain Stopper
If you did not receive one or more of the components listed for the cabinet, or if any of the components are damaged, contact Labconco Corporation immediately for further instructions.
3. Using a step ladder, remove the tape and/or packaging material around the
top of the unit.
4. Plug the power cord into the back of the cabinet, and then into an
appropriate outlet. Install the Key into the UV Security lock on the upper panel on those models equipped with a UV light.
Product Service 1-800-522-7658
Page 16
Chapter 3: Getting Started
12
Initial Operation
Prior to use with evidence, you should perform the following operational tests to ensure the cabinet is operating properly:
Airflow Smoke Pattern Test
With the cabinet fan ON, introduce smoke (from a cigarette,
chemical smoke stick, etc.) into the prefilter at the bottom of the door. The smoke should flow directly into the cabinet, and then into the upper prefilter located at the top of the cabinet liner. The smoke should move from the door to the intake prefilter in approximately 5 seconds or less. If the smoke does not flow as described, see the “Adjusting the Fan Speed” section of Chapter 6.
UV Light and Interlock Tests
With the door closed, turn the “Fan/UV Light” switch to “UV” and
the “UV Security Keylock to “ON”. The UV light will then turn on. Open the door, and the UV light will turn off. Close the door, and the UV light should relight. Turn the UV Security Keylock to the
“OFF” position, and the UV light will turn off. If the UV light does
not work as described, call Labconco’s Product Service Department
at 1-800-522-7658 or 816-333-8811.
Spray Gun and Drain Pump and Tests (Model # 34000-10 thru -14,
34040-10 thru 34040-14 & 34050-10 thru -14 only)
Open the cabinet door, and ensure that the drain strainer is in place.
Open the lower door, and turn the cabinet water supply valve on. Use the spray gun in the “Shower” setting, ensuring proper operation. Turn the gun and cabinet supply valve off. Push the control switch to the “Drain Pump” position. The majority of water should be pumped out of the cabinet and to the floor drain. Some water may remain in the bottom of the cabinet due to surface tension or the floor not being level. In this case, use a small squeegee or sponge to direct the remaining water to the drain. Shut off the drain pump when finished.
Product Service 1-800-522-7658
Page 17
13
CChhaapptteerr 44:: PPeerrffoorrm
maannccee FFeeaattuurreess aanndd SSaaffeettyy
PPrreeccaauuttiioonnss
The Protector Evidence Drying Cabinet operates using the following principles:
Directional airflow. Prefiltration of large particulates by an inlet prefilter (in the door) and an
exhaust prefilter (in the top of the liner).
Filtration and retention of particulates by a High Efficiency Particulate Air
(HEPA) filter.
Adsorption of organic (odorous) chemicals by a granular activated carbon
filter.
The major components in the cabinet are:
The prefilters The blower to force air through the cabinet The HEPA filter The granular activated carbon filter
Product Service 1-800-522-7658
Page 18
Chapter 4: Performance Features and Safety Precautions
14
Figure 4-1
Room Air HEPA-filtered Air
Contaminated Air
Directional Airflow
Directional airflow plays a critical role in evidence drying cabinet performance. Air is drawn into the inlet prefilter, located on the bottom of the door. The filtered air then flows upward through the liner, passing over the evidence to be dried. Before leaving the liner, the air then passes through the exhaust prefilter, that is designed to protect the HEPA filter from large particulates, and may be removed to examine it for the collection of evidence. The air is pressurized by the blower, and then forced through the HEPA and carbon filters, to remove biohazardous particulates and odorous chemicals, respectively. This is illustrated in Figure 4-1.
Product Service 1-800-522-7658
Page 19
Chapter 4: Performance Features and Safety Precautions
15
Figure 4-2
Polyurethane seal Between media pack and filter frame
Filter frame
Filter media pack – A continuous sheet of filter media
Optional Evidence Prefilters
Optional Evidence Prefilters can be fitted on the inside (cabinet side) of the door prefilter (located on the front of the door) and cabinet prefilter (located on the inside top of the liner). These thin prefilters are designed for single-use, to be removed after the evidence has dried. After removal, the Evidence Prefilters can be examined for trace evidence, or stored with the evidence in a sealed bag for future reference. See Appendix A for reorder part numbers.
Prefilters
Both the door prefilter (located on the front of the door) and cabinet prefilter (located on the inside top of the liner) are disposable, polyester particulate filters. Both filters have a Minimum Efficiency Reporting Value (MERV) of 11, meaning they remove 65-79% of particles 1.0-3.0 microns in size, and more than 85% of particles 3.0-10 microns in size. See Appendix A for reorder part numbers.
HEPA Filters
HEPA filters are disposable, dry-type particulate filters. The filter material or media is typically made of borosilicate microfibers formed into a thin sheet, in a process similar to the production of paper. This sheet is folded, or pleated to increase its surface area. The pleats are held in place by aluminum diffusers or by beads of glue that add rigidity to the media pack. The pack is then set into a frame, and sealed as shown in Figure 4-2.
The HEPA filter manufacturer establishes the efficiency of the filter by challenging it with an aerosol of known particle size. The number of particles that penetrate the filter are quantified, and this establishes the efficiency of the filter. The filters used in the evidence drying cabinets are at least 99.99% efficient in removing particles
0.3 micron. Note: The HEPA filter media is very fragile. DO NOT touch the media. If
you think the media of a HEPA filter is damaged, DO NOT USE THE CABINET. Call Labconco’s Product Service Department at 1-800-522-7658 or
816-333-8811. Note: HEPA Filters are only effective against particulate material. Gases will pass
through the filter.
Product Service 1-800-522-7658
Page 20
Chapter 4: Performance Features and Safety Precautions
16
Blower
The blower assembly pulls room air into the intake prefilter on the front of the door, through the liner and exhaust prefilter, and then forces it through the HEPA and granular activated carbon filter. The evidence drying cabinet uses a motorized, backward curved fan for quiet, energy efficient operation.
Cabinet Air Intake and Exhaust
In order to operate properly, the air intake at the bottom of the door, the exhaust prefilter area in the top of the liner, and the top surface of the activated carbon filter must be unobstructed. Blocking or restricting any of these areas, or allowing the prefilters to become excessively loaded with dust can adversely affect the performance of the cabinet.
Ultraviolet (UV) Lamp (Model # 34000-00 thru -04, 34040-00 thru -04
& 34050-00 thru -04 only)
The optional UV lamp generates a primary wavelength of light of 254nm. A secondary emission is in the visible (blue) wavelength, resulting in the characteristic blue color while operating. UV light with a wavelength of 254 nm is biocidal, primarily by creating thymine dimers in DNA. These dimers prevent the correct transcription of the DNA into RNA, resulting in cellular death or viral inactivation. In order to be effective, the UV light must directly strike the nucleic acid, and its effectiveness can be diminished or negated by dissolved proteins or metals, or by other UV-opaque substances protecting the target nucleic acid.
Because of its limitations, UV light should be used as an adjunct to good surface disinfection practices. In order to get optimum performance from the UV light, it should be replaced after 6,000 hours of operation or less, and the exterior surface of the lamp should be kept clean and free of dust.
Note: UV irradiation is absorbed by the tempered safety glass of the door. Independent research has shown that the level of UV irradiation on the outside of the cabinet’s door is equal to background radiation levels.
Note: The UV sensitivity of a target organism varies, depending on the UV output of the lamp, the genus and species of the organism, the medium the agent is suspended in, etc. Contact the Health and Safety Officer at your facility for UV light use and recommendations.
Product Service 1-800-522-7658
Page 21
Chapter 4: Performance Features and Safety Precautions
17
Safety Precautions
Wash Down Models (Model # 34000-10 thru -14, 34040-10 thru -
14 & 34050-10 thru -14 only):
DO NOT directly connect the drain line to a sewer, drain system or place in a vertical standpipe. This type of installation will not allow the drain to function properly.
DO NOT use the wash down gun to clean up biohazardous material; Disinfect any surface or material BEFORE using the gun.
ALWAYS wear appropriate personal protective equipment, including eye and face protection when washing down the cabinet.
Ensure that the drain line is not kinked or obstructed, and flows to a floor drain with an air gap.
The Drain Pump should only be operated to remove standing liquid in the bottom of the cabinet.
Do not operate the pump dry for more than 5 minutes.
UV Lamp Models (Model # 34000-00 thru -04, 34040-00 thru -04 &
34050-00 thru -04 only):
NEVER expose ANY EVIDENCE, plastic or coated materials to ultraviolet (UV) radiation. Never bypass the UV safety interlock that only allows the UV light to work when the door is closed. UV exposure to the skin is hazardous and can result in serious eye or skin injury.
NEVER expose UV Lamp or UV Lamp connections to liquids.
For all models:
Ensure that the cabinet is connected to electrical service in accordance with local and national electrical codes. Failure to do so may create a fire or electrical hazard. Do not remove or service any electrical components without first disconnecting the cabinet from electrical service.
NEVER use this cabinet as a chemical storage cabinet. When handling evidence, or working in the cabinet, always wear appropriate
personnel protective equipment. NEVER attempt to dry evidence containing flammable gases or solvents in the
cabinet. An open flame should NOT be used in the cabinet.
Product Service 1-800-522-7658
Page 22
Chapter 4: Performance Features and Safety Precautions
18
HEPA filters only remove particulate matter. Operations generating volatile toxic chemicals are prohibited. The media of the filter is fragile and should not be touched. Avoid puncturing the HEPA filter during installation or normal operation. If you suspect that a HEPA filter has been damaged, DO NOT use the cabinet; contact Labconco at 800-821-5525 or 816-333-8811 for further information.
The HEPA filters in the cabinet will gradually accumulate airborne particulate matter from the items placed in the cabinet. The rate of accumulation will depend upon the cleanliness of the air, operating time and the nature of items being dried in the cabinet.
Do not operate the cabinet without all of the appropriate filters in place. If an odor is detected coming out of the granular activated carbon, it needs to be replaced.
The weight limit for the fold down shelves is 15 lbs (6.8 kg), and the floor of the cabinet and lower storage area is 20 lbs (9.1 kg).
When surface disinfecting the cabinet:
Always wear appropriate personnel protective equipment. NEVER use high pressure water to clean the liner; this practice could
generate biohazardous aerosols.
NEVER spray disinfectants into the blower intake at the top of the liner;
this may damage the blower.
Avoid splashing the disinfecting solution on skin or clothing. Ensure adequate room ventilation. Carefully follow the disinfectant’s safety instructions. Always dispose of disinfecting solutions in accordance with local and
national laws.
DO NOT allow bleach or disinfectants with high concentrations of free
chlorine to contact the stainless steel components of the cabinet for a long period of time. Free chlorine will corrode stainless steel after extended contact.
When handling used filters or prefilters, always wear appropriate personnel protective equipment.
Product Service 1-800-522-7658
Page 23
19
CChhaapptteerr 55:: UUssiinngg tthhee CCaabbiinneett
UV Lamp Models (Model # 34000-00 thru -04, 34040-00 thru -04
& 34050-00 thru -04 only)
Fan/UV Light Switch
The cabinet Fan/UV Light switch is located on the upper front of the cabinet, as shown in Figure 5-1. Turning the switch to the “Fan” position starts the blower,
drawing air in through the cabinet. Turning the switch to the “UV On” position
will allow the UV light to turn on when the door(s) is closed and the UV Security lock is in the “ON” position.
UV Security Lock
The UV security lock, located to the right of the Fan/UV Light switch, allows the user to lockout the UV, preventing its operation. Before evidence is loaded into the cabinet, it is recommended that the lock be turned to the “OFF” position, and the key removed. This will prevent the accidental activation of the UV light while evidence is drying. The key should be stored in a convenient, secure location, known to all of the users to prevent its loss. Recommended locations include hanging from the bar at the top of the liner, or in the lower storage area.
Filter System Light
The Filter System Light will illuminate if the differential pressure across the filters becomes excessive. If lit, first inspect or replace the two prefilters. If lit with clean prefilters, replace the HEPA filter.
NOTE: If the Filter System Light illuminates during use, finish the drying of the evidence in the cabinet, and then correct the problem before using the cabinet again.
Product Service 1-800-522-7658
Page 24
Chapter 5: Using the Cabinet
20
Figure 5-1
Shelf Operation
The three drying shelves on the left side wall can be positioned horizontally to dry small pieces of evidence, or swung down, against the left side wall of the liner, if needed. To position the shelves in the up (horizontal) position:
1. Pivot the desired shelf until it is above horizontal (the frame of the shelf is
higher than the retaining clip on the back wall of the liner.
2. Push the shelf back all the way.
3. Slowly lower the shelf to a horizontal position, ensuring that the frame of
the shelf engages the retaining clip on the back wall of the liner. The shelf is now ready to use.
To fold the shelf down (vertical position):
1. Lift up on the rear frame of the shelf, to release it from the rear retaining
clip.
2. Pull the shelf towards the front of the liner, and allow the shelf to swing
down into a vertical position.
Drying in the Cabinet
Start-up
Turn off UV light and turn the UV Security switch to “OFF”. Remove the
key, and store it in a safe place.
Ensure the drain strainer is in place. Turn the switch to “Fan”. Check the condition of the prefilters. If your protocol requires the use of
evidence prefilters, ensure a new set of both door and cabinet prefilters has been properly installed, according to the instructions in Chapter 6: Maintaining Your Cabinet. Note if the Filter System light is on.
Product Service 1-800-522-7658
Page 25
Chapter 5: Using the Cabinet
21
Wash hands and arms thoroughly with germicidal soap. Wear a long sleeved lab coat with knit cuffs and over-the-cuff rubber
gloves. Use protective eyewear. Wear a protective mask if appropriate.
Wipe-Down
Open the door to its full open position. Wipe down the interior surfaces of the cabinet with 70% ethanol, or a
suitable disinfectant, and allow it to dry.
CAUTION: Do not use bleach on the stainless steel liner. Use disinfectants compatible with stainless steel.
Loading Evidence
Load the evidence so that air flows around all surfaces of it. Do not
overload the cabinet. Do not place objects directly against the side walls.
Do not obstruct the intake prefilter (bottom of the door) or the exhaust air
discharge, on top of the cabinet.
Large objects should not be placed close together. Slowly close the door(s) and use a padlock (not supplied), or secure it with
a chain of custody seal.
Unloading Evidence
Wear a long sleeved lab coat with knit cuffs and over-the-cuff rubber
gloves. Use protective eyewear. Wear a protective mask if appropriate.
Carefully remove objects in the cabinet. Evidence that was considered
biohazardous when loaded should be handled as biohazardous. If your protocol required the use of evidence prefilters, remove them now and store them appropriately.
Wipe-Down
Wear a long sleeved lab coat with knit cuffs and over-the-cuff rubber
gloves. Use protective eyewear. Wear a protective mask if appropriate.
Wipe down the interior surfaces of the cabinet with 70% ethanol, or a
suitable disinfectant, and allow to dry.
Inspect and clean the prefilters at the bottom of the door, and at the top of
the liner.
Dispose of rubber gloves appropriately, and have lab coat laundered
properly.
Product Service 1-800-522-7658
Page 26
Chapter 5: Using the Cabinet
22
Figure 5-1a
Wash hands and arms thoroughly with germicidal soap.
Shutdown
Close the door. If UV disinfection is desired, insert the UV Security Lock
Key, and turn it to the “ON” position. Turn the Fan/UV Light switch to
“UV”. The UV light will light.
Turn the UV Light off after an appropriate disinfection time, typically 1-8
hours.
Wash Down Models (Model # 34000-10 thru -14, 34040-10 thru -14
& 34050-10 thru -14)
Fan/Drain Pump Switch
The cabinet Fan/Drain Pump switch is located on the upper front of the cabinet, as shown in Figure 5-1a. Turning the switch to the “Fan” position starts the blower, drawing air into through the cabinet. Turning the switch to the “Drain Pump” position starts the drain pump.
Do not operate the pump dry for more than 5 minutes.
Filter System Light
The Filter System Light will illuminate if the differential pressure across the filters becomes excessive. If lit, first inspect or replace the two prefilters. If lit with clean prefilters, replace the HEPA filter.
NOTE: If the Filter System Light illuminates during use, finish the drying of the evidence in the cabinet, and then correct the problem before using the cabinet again.
Product Service 1-800-522-7658
Page 27
Chapter 5: Using the Cabinet
23
Shelf Operation
The three drying shelves on the left side wall can be positioned horizontally to dry small pieces of evidence, or swung down, against the left side wall of the liner, if needed. To position the shelves in the up (horizontal) position:
1. Pivot the desired shelf until it is above horizontal (the frame of the shelf is
higher than the retaining clip on the back wall of the liner).
2. Push the shelf back all the way.
3. Slowly lower the shelf to a horizontal position, ensuring that the frame of
the shelf engages the retaining clip on the back wall of the liner. The shelf is now ready to use.
To fold the shelf down (vertical position):
3. Lift up on the rear frame of the shelf, to release it from the rear retaining
clip.
4. Pull the shelf towards the front of the liner.
5. Allow the shelf to swing down into a vertical position.
Drying in the Cabinet
Start-up
Ensure the drain strainer is in place. Turn the switch to “Fan”. Check the condition of the prefilters. If your protocol requires the use of
evidence prefilters, ensure a new set of both door and cabinet prefilters has been properly installed, according to the instructions in Chapter 6: Maintaining Your Cabinet. Note if the Filter System light is on.
Wash hands and arms thoroughly with germicidal soap. Wear a long sleeved lab coat with knit cuffs and over-the-cuff rubber
gloves. Use protective eyewear. Wear a protective mask if appropriate.
Product Service 1-800-522-7658
Page 28
Chapter 5: Using the Cabinet
24
Wipe-Down
Open the door to its full open position. Wipe down the interior surfaces of the cabinet with 70% ethanol, or a
suitable disinfectant, and allow to dry.
CAUTION: Do not use bleach on the stainless steel liner. Use disinfectants compatible with stainless steel.
Loading Evidence
Load the evidence so that air flows around all surfaces of it. Do not
overload the cabinet. Do not place objects directly against the side walls.
Do not obstruct the intake prefilter (bottom of the door) or the exhaust air
discharge, on top of the cabinet.
Large objects should not be placed close together. Slowly close the door(s) and use a padlock (not supplied), or secure it with
a chain of custody seal.
Unloading Evidence
Wear a long sleeved lab coat with knit cuffs and over-the-cuff rubber
gloves. Use protective eyewear. Wear a protective mask if appropriate.
Carefully remove objects in the cabinet. Evidence that was considered
biohazardous when loaded should be handled as biohazardous. If your protocol required the use of evidence prefilters, remove them now and store them appropriately.
Wash-Down
Wear a long sleeved lab coat with knit cuffs and over-the-cuff rubber
gloves. Use protective eyewear. Wear a protective mask if appropriate.
Wipe down the interior surfaces of the cabinet with 70% ethanol, or a
suitable disinfectant, and allow adequate contact time.
Turn the cabinet supply valve on, and ensure the spray gun is in the
Shower” position. Turn the switch to “Drain Pump”
Gently wash down the walls and back of the cabinet’s interior, avoiding
splashing whenever possible.
Do not overfill the bottom of the cabinet liner.
Product Service 1-800-522-7658
Page 29
Chapter 5: Using the Cabinet
25
Allow the pump to operate for 30 seconds to 5 minutes after the cabinet
liner has drained. Then turn off the pump. Turn the cabinet water shut off valve off.
Remove and clean out the drain strainer. Inspect and clean the prefilters at the bottom of the door, and at the top of
the liner.
Dispose of rubber gloves appropriately, and have lab coat laundered
properly.
Wash hands and arms thoroughly with germicidal soap.
Shutdown
Close the door and push the switch to the center “Off” position.
Product Service 1-800-522-7658
Page 30
26
CChhaapptteerr 66:: MMaaiinnttaaiinniinngg tthhee CCaabbiinneett
The common service operations necessary to maintain the Evidence Drying cabinet for peak performance are listed below.
Routine Maintenance Schedule
Weekly
Using 70% ethanol, or a suitable disinfectant, surface disinfect the inside of
the cabinet.
Using an appropriate glass cleaner, or LabSolutions Glass & Surface
Wipes, Labconco part # 1570000; clean the door and the surface of the UV lamp.
Operate the cabinet blower, noting if the Filter System Light turns on. Ensure the cabinet water shut off valve is in the off position.
Monthly (or more often as required)
Using a damp cloth, or LabSolutions Glass & Surface Wipes, Labconco
part # 1570000; clean the exterior surfaces of the cabinet, particularly the front and top of the cabinet, to remove any accumulated dust.
All weekly activities. Check the prefilters; replace both if needed.
Annually
All monthly activities. Replace HEPA and granular activated carbon filters.
Product Service 1-800-522-7658
Page 31
Chapter 6: Maintaining the Cabinet
27
Figure 6-1
Service Operations
Door Evidence Prefilter Replacement:
1. Pull the prefilter frame(s) straight away from the door(s), as shown in
Figure 6-1.
2. Wearing appropriate gloves, and working on the inside surface of the prefilter
frame(s), carefully pull out the Evidence prefilter(s). The Evidence prefilter(s) can now be examined for trace evidence, or stored with the evidence in a sealed bag for future reference.
3. Examine the door prefilter(s). If it is contaminated from the evidence drying
operation, or from dust, replace it with a new one, such that the orange side of the prefilter(s) faces you (away from the front of the prefilter frame).
4. Press the prefilter frame(s) back into the door, engaging the fastener strips.
Door Prefilter Replacement:
1. Follow Steps 1, 3, and 4 above.
Product Service 1-800-522-7658
Page 32
Chapter 6: Maintaining the Cabinet
28
Figure 6-2
Thumb
Screw
Prefilter
Frame
Pre-Filter
Cabinet Evidence Prefilter Replacement:
1. Open the cabinet’s glass door(s), and locate the cabinet prefilter frame on the
upper right side of the liner. See Figure 6-2.
2. To access the prefilter, loosen the captive thumbscrew until it is released. The
front of the prefilter frame will swing down.
NOTE: The prefilter frame can be removed from the cabinet for easier prefilter examination and removal. With the front thumbscrew released, and the front of the prefilter frame tilting downward, remove it by lifting the back of the frame up slightly, and then gently pull the frame straight out of the cabinet.
3. Wearing appropriate gloves, and working on the prefilter frame, carefully lift
out the prefilter, and then the evidence prefilter. The evidence prefilter can now be examined for trace evidence, or placed in a sealed plastic bag for future examination.
4. Examine the cabinet prefilter. If it is contaminated from the evidence drying
operation, or from dust, replace it with a new one, such that the orange side of the prefilter is up (away from the bottom of the prefilter frame).
5. Reinstall the prefilter frame, and hand tighten the thumbscrew.
Cabinet Prefilter Replacement:
1. Follow Steps 1, 2, 4 and 5 above.
2. replacement, such that the orange side of the prefilter faces up.
Product Service 1-800-522-7658
Page 33
Chapter 6: Maintaining the Cabinet
29
Figure 6-3
Granular Activated Carbon and HEPA Filter Replacement:
1. Use an appropriate stepladder release the two bungee cords securing the
granular activated and carbon and HEPA filters, as shown in Figure 6-3.
2. When handling used filters or prefilters, always wear appropriate personnel
protective equipment.
3. Lift the granular activated carbon and HEPA filters straight up and out of
the cabinet top.
4. Install the replacement filters, with the HEPA filter below the granular
activated carbon filter into the cabinet. The gaskets on both filters should be on the lower (bottom) surface.
5. Reinstall both bungee cords to secure the filters in place. The unit is now
ready to resume operation.
6. Filters should be disposed of properly. HEPA Filters should be treated as a
biohazard.
Product Service 1-800-522-7658
Page 34
Chapter 6: Maintaining the Cabinet
30
THE LAMP(S) IN THIS PRODUCT CONTAIN MERCURY
Manage in accordance with local disposal laws. DO NOT place lamps in trash. Dispose as a hazardous waste. For information regarding safe handling, recycling and disposal, consult www.lamprecycle.org
CETTE LAMPE DANS CE PRODUIT CONTIENT DU MERCUE
Éliminez ou recyclez conformément aux lois applicables. Pour de l‘information concernant des pratiques de manipulation sécuritaires et l’élimination sécuritaire
et le recyclage, veuillez consulter www.lamprecycle.org
Changing the UV Lamp (Model # 34000-00 thru -04, 34040-00 thru -04 & 34050-00 thru -04):
Note: For optimum performance, the UV lamp should be changed on an annual basis.
The UV lamp and the work area of the cabinet must be thoroughly decontaminated before removing the lamp.
1. Thoroughly surface decontaminate the UV lamp and the work area of the
cabinet.
2. Unplug the cabinet, and open the door(s).
3. Remove the UV lamp by rotating it 90 degrees and lifting it straight up and out
of its sockets.
4. Install new lamp by reversing the removal procedure.
Product Service 1-800-522-7658
Page 35
Chapter 6: Maintaining the Cabinet
31
Figure 6-4
Resetting a Circuit Breaker:
To reset any of the circuit breakers located on the top rear of cabinet, depress the white button until it sets.
Setting the Speed Control
Adjustment of the speed control varies the volume of air drawn through the liner while the unit is operating. Ideally, smoke released into the intake prefilter should traverse the liner, and be drawn into the exhaust prefilter in approximately 5 seconds or less.
Using a small screwdriver, adjust the speed control to give the required face velocity. See Figure 6-4 to locate the speed control. The air velocity is increased by turning the speed control counter clockwise and clockwise to decrease velocity.
Product Service 1-800-522-7658
Page 36
32
CChhaapptteerr 77:: TTrroouubblleesshhoooottiinngg
PROBLEM
CAUSE
CORRECTIVE ACTION
Cabinet blower and lights won’t turn on
Unit not plugged into outlet
Plug the cabinet into appropriate electrical service.
Check connection to the top of cabinet.
Circuit breaker(s) tripped
Reset circuit breakers.
UV light not working
Door(s) are open
Close door(s) – UV light will not work with the door(s) open.
UV Security Lock not on
The key switch for the UV Security Lock must be turned on.
Lamp is defective
Replace defective lamp.
Lamp wiring is disconnected
Inspect lamp wiring.
Defective lamp ballast
Replace lamp ballast.
UV light is dim or flickering
Lamp is defective or is at end of operating lifetime
Replace defective or worn out lamp. Defective lamp ballast
Replace lamp ballast.
Refer to the following table if the cabinet fails to operate properly. If the suggested corrective actions do not solve the problem, contact Labconco for additional assistance.
Product Service 1-800-522-7658
Page 37
Chapter 7: Troubleshooting
33
PROBLEM
CAUSE
CORRECTIVE ACTION
Filter System Light turns on
Blockage of the intake prefilters
Check the inlet prefilters; replace if needed. Ensure that there is no obstruction of the exhaust on top of the cabinet.
HEPA filter loading
Replace the HEPA filter.
Product Service 1-800-522-7658
Page 38
34
AAppppeennddiixx AA:: CCoom
mppoonneennttss
Item
Quantity
(ea per unit)
Part
No.
Cabinet
Size
Description
1 1 3872600
3'
Door Prefilter, 3FT package of 6 ea
1A
2
3872601
4'
Door Prefilter, 4FT package of 6 ea
1B
2
3872902
5'
Door Prefilter, 5FT package of 6 ea
2 1 3951802
3'
Door Evidence Prefilter, 3FT package of 10 ea
2A
2
3951822
4'
Door Evidence Prefilter, 4FT package of 10 ea
2B
2
3951832
5'
Door Evidence Prefilter, 5FT package of 10 ea
3 1 3872700
3', 4' & 5'
Cabinet (top) Prefilter, package of 6 ea
4 1 3951803
3', 4' & 5'
Cabinet (top) Evidence Prefilter, package of 10 ea
5 1 6938100
3', 4' & 5'
HEPA Filter
6 1 6938200
3', 4' & 5'
Granular Activated Carbon Filter
7 1 1270100
3', 4' & 5'
Lamp, UV 8 1
3869800
3'
Paper Liner, 3FT package of 25 ea
8A
1
3869802
4'
Paper Liner, 4FT package of 25 ea
8B
1
3869804
5'
Paper Liner, 5FT package of 25 ea
Evidence Drying Cabinet Replacement Parts
Product Service 1-800-522-7658
Page 39
35
AAppppeennddiixx BB:: DDiim
meennssiioonnss
Model Number
Size A B C D
34000-00 thru -14
3’
36.0" (1097mm)
Casters
32.3" (820mm)
25.5" (648mm)
34040-00 thru -14
4’
48.0" (1219mm)
Legs
45.3" (1151mm)
37.5" (953mm)
34050-00 thru -14
5’
60.0" (1524mm)
Legs
57.3" (1455mm)
49.5" (1257mm)
B-1
B
Interior Dimensions: 56.0" (1422mm) Height x 26.5" (673mm) Deep x “C” Lower Storage Dimensions: 12.4" (315mm) Height x 26.5" (673mm) Deep x “D”
Product Service 1-800-522-7658
Page 40
36
AAppppeennddiixx CC:: SSppeecciiffiiccaattiioonnss
Model #
Requirements
Energy Use
34000 -00 , -10
115 VAC, 60 Hz, 6 Amps
100W
34000- 01 thru -04 & -11 thru -14
230 VAC, 50/60 Hz, 3 Amps
100W
34040 -00 , -10
115 VAC, 60 Hz, 6 Amps
100W
34040- 01 thru -04 & -11 thru -14
230 VAC, 50/60 Hz, 3 Amps
100W
34050 -00 , -10
115 VAC, 60 Hz, 6 Amps
100W
34050- 01 thru -04 & -11 thru -14
230 VAC, 50/60 Hz, 3 Amps
100W
Electrical Data
Environmental Conditions
Indoor use only. Maximum altitude: 6562 feet (2000 meters). Ambient temperature range: 41° to 104°F (5° to 40°C). Maximum relative humidity: 80% for temperatures up to 88°F (31°C),
decreasing linearly to 50% relative humidity at 104°F (40°C).
Main supply voltage fluctuations not to exceed ±10% of the nominal
voltage.
Transient overvoltages according to Installation Categories II (Overvoltage
Categories per IEC 1010). Temporary voltage spikes on the AC input line that may be as high as 1500V for 115V models and 2500V for 230V models are allowed.
Used in an environment of Pollution degrees 2 (i.e., where normally only
non-conductive atmospheres are present). Occasionally, however, a temporary conductivity caused by condensation must be expected, in accordance with IEC 664.
Product Service 1-800-522-7658
Page 41
37
AAppppeennddiixx DD:: AAcccceessssoorriieess
Door Evidence Prefilters (3FT #3951802, 4FT #3951822,
5FT #3951832)
Package of 10 each Door Evidence prefilters. Evidence prefilters are single use polyester prefilters to be examined and or retained for trace evidence. Precut to fit the door prefilter frame. 1 lb. (0.5kg).
Cabinet Evidence Prefilters (#3951803)
Package of 10 each Cabinet Evidence prefilters. Evidence prefilters are single use polyester prefilters to be examined and or retained for trace evidence. Precut to fit
the cabinet prefilter frame. 1 lb. (0.5kg).
Door Prefilters (3FT #3872600, 4FT #3872601, 5FT #3872602)
Package of 6 polyester prefilters with a MERV 11 rating. Precut to fit the door of the cabinet. 1 lb. (0.5kg).
Cabinet Prefilters (#3872700)
Package of 6 polyester prefilters with a MERV 11 rating. Precut to fit the inside top of liner. 1 lb. (0.5kg).
Starter Kit (3FT #3869900, 4FT #3869910, 5FT #3869920)
Includes many of the accessories needed to use the cabinet. 15 lbs. (6.8kg).
Included:
6-door prefilters 6-cabinet prefilters 1-door evidence prefilter (qty. 2 on -10 & -20 kits) 1-cabinet evidence prefilter 10-hanging clips 10-paper liners (die cut paper to fit the bottom of the cabinet) to collect
material that falls from evidence. Paper is waxed on one side.
1-32 oz. (946 ml) bottle of germicidal cleaner with spray head.
Paper Liners (3FT #3869800, 4FT #3869802, 5FT #3869804)
A tube of 25 paper liners (die cut paper to fit the bottom of the cabinet) to collect material that falls from evidence. Paper is waxed on one side. 12 lbs. (5.5kg).
Product Service 1-800-522-7658
Page 42
38
Paper Liners (3FT #3894200, 4FT #3894201, 5FT #3894202)
A package of 100 paper liners (die cut paper to fit the bottom of the cabinet) to collect material that falls from evidence. Paper is waxed on one side. 25 lbs. (11kg).
Security Tags (#3902400)
Package of 100 each tamper-evident, labeled tags. Tags irreversibly indicate if the cabinet door has been opened during the drying cycle. 1 lb. (0.5kg).
Hanging Clips (#3869500)
Package of 10 each hanging clips, that allow small articles to be hung from the cabinet’s hanger rod or shelf. 1 lb. (0.5kg).
Exhaust System Connection Kit (#3869600)
Allows the cabinet to be connected to an exhaust system via the included 14" (356 mm) long, 5" (125 mm) diameter flexible duct. The exhaust system must be capable of exhausting 150 CFM (4.2 m3/min) at a vacuum of 0.5" of H2O (125 Pa). 10 lbs. (5kg).
Floor Drain Connection Kit (#3928200)
Allows the cabinet liner drain connector to be routed to a nearby floor drain. Includes hose barb adaptor and 10 ft. (3 meters) of 1/2" (13 mm) O.D. polyurethane tubing. 5 lbs. (2.3 kg).
Germicidal Cleaner, Individual Wipes (#3872400)
Package of 100 each individually packaged single use disinfectant wipes. 2 lbs. (1kg).
Germicidal Cleaner, Ready-to-Use Wipes (#3872500)
Package of 90 ready to use disinfectant wipes, in a convenient dispensing tub. 2 lbs. (1kg).
Germicidal Cleaner, 32 oz. (946 ml) (#3901800)
Bottle of liquid disinfectant cleaner. Includes a screw-on spray head dispenser. 2 lbs. (1kg).
Germicidal Cleaner, 1 Gal. (3785 ml) (#3901700)
Bottle of liquid disinfectant cleaner, for refilling the spray bottle. 10 lbs. (5kg).
Product Service 1-800-522-7658