Labconco 3030000, 3030001, 3030002, 3030003, 3030004 User guide

...
Page 1
Fiberglass 30
Laboratory Fume Hoods
Models
3030000 3030001 3030002 3030003 3030004 3030005
Page 2
Copyright © 2021 Labconco Corporation. The information contained in this manual and the accompanying products are copyrighted and all rights reserved by Labconco Corporation. Labconco Corporation reserves the right to make periodic design changes without obligation to notify any person or entity of such change.
Returned or Damaged Goods
Do not return goods without the prior authorization from Labconco. Unauthorized returns will not be accepted. If your shipment was damaged in transit, you must file a claim directly with the freight carrier. Labconco Corporation and its dealers are not responsible for shipping damages.
The United States Interstate Commerce Commission rules require that claims be filed with the delivery carrier within fifteen (15) days of delivery.
Limitation of Liability
The disposal and/or emission of substances used in connection with this equipment may be governed by various federal, state, or local regulations. All users of this equipment are required to become familiar with any regulations that apply in the user’s area concerning the dumping of waste materials in or upon water, land, or air and to comply with such regulations. Labconco Corporation is held harmless with respect to user’s compliance with such regulations.
Contacting Labconco Corporation
If you have questions that are not addressed in this manual, or if you need technical assistance, contact Labconco’s Customer Service Department or Labconco’s Product Service Department at 1-800-821-5525 or 1-816-333-8811, between the hours of 7:30 a.m. and 5:30 p.m., Central Standard Time.
Part #2831600, Rev. I
ECO N065
Warranty
Labconco Corporation provides a warranty to the original buyer for the repair or replacement of parts and reasonable labor as a result of normal and proper use of the equipment with compatible chemicals. Broken glassware and maintenance items, such as filters, gaskets, light bulbs, finishes and lubrication are not warranted. Excluded from warranty are products with improper installation, erratic electrical or utility supply, unauthorized repair and products used with incompatible chemicals.
The warranty for Fiberglass 30 Laboratory Fume Hoods will expire one year from date of installation or two years from date of shipment from Labconco, whichever is sooner. Warranty is non-transferable and only applies to the owner (organization) of record.
Buyer is exclusively responsible for the set-up, installation, verification, decontamination or calibration of equipment. This limited warranty covers parts and labor, but not transportation and insurance charges. If the failure is determined to be covered under this warranty, the dealer or Labconco Corporation will authorize repair or replacement of all defective parts to restore the unit to operation. Repairs may be completed by 3rd party service agents approved by Labconco Corporation. Labconco Corporation reserves the rights to limit this
warranty based on a service agent’s travel, working hours, the site’s entry restrictions and unobstructed access
to serviceable components of the product.
Under no circumstances shall Labconco Corporation be liable for indirect, consequential, or special damages of any kind. This warranty is exclusive and in lieu of all other warranties whether oral, or implied.
Page 3
Original instructions
T
T
AABBLLEE
O
O
FF
C
C
OONNTTEENNTTSS
CHAPTER 1: INTRODUCTION 1 About This Manual 1 Typographical Conventions 2 Your Next Step 4
CHAPTER 2: PREREQUISITES 5 Location Requirements 6 Support Requirements 6 Exhaust Requirements 6 Electrical Requirements 7 Service Line Requirements 7 Space Requirements 7 Your Next Step 7
CHAPTER 3: GETTING STARTED 8 Unpacking Your Fume Hood 8 Removing the Shipping Skid 9 Sash Weight Release 10 Install the Hood on a Supporting Structure & Work Surface 10 Connecting to the Hood Exhaust System 12 Connecting the Electrical Supply Source to the Hood 13 Connecting the Service Lines to the Hood 15 Sealing the Fiberglass 30 Hood to the Work Surface 16 Certifying the Fiberglass 30 Fume Hood 16 Your Next Step 17
CHAPTER 4: PERFORMANCE FEATURES AND
SAFETY PRECAUTIONS 18
Performance Features 18 Safety Precautions 21 Your Next Step 24
Page 4
CHAPTER 5: USING YOUR FIBERGLASS 30 25 Operating the Vertical-Rising Sash 25 Operating the Blower 25 Operating the Lights 25 Working in Your Fiberglass 30 Fume Hood 25 Your Next Step 26
CHAPTER 6: MAINTAINING YOUR FIBERGLASS 30 27
Routine Maintenance Schedule 27 Routine Service Operations 28 Your Next Step 29
CHAPTER 7: MODIFYING YOUR FIBERGLASS 30 30 Installing Work Surfaces and Storage Cabinets 30 Installing Additional Service Fixtures 30 Installing Guardian Digital Airflow Monitor or Guardian Jr. 31 Installing an Electrical Duplex Receptacle 32 Your Next Step 32
CHAPTER 8: TROUBLESHOOTING 33
APPENDIX A: FIBERGLASS 30 HOOD COMPONENTS 35
APPENDIX B: FIBERGLASS 30 HOOD DIMENSIONS 38
APPENDIX C: FIBERGLASS 30 HOOD SPECIFICATIONS 43
APPENDIX D: REFERENCES 46
Page 5
Product Service 1-800-522-7658
1
Original instructions
C
C
HHAAPPTTEERR
1
1
I
I
NNTTRROODDUUCCTTIIOONN
Congratulations on your purchase of a Labconco Fiberglass 30 Laboratory Fume Hood. Your Fiberglass 30 Laboratory Fume Hood is designed to protect you. It
is the result of Labconco’s more than 50 years experience in manufacturing fume
hoods, and users like you suggested many of its features to us.
The Labconco Fiberglass 30 Fume Hood has been engineered to effectively contain toxic, noxious, or other harmful materials when properly installed. The Fiberglass 30 offers many unique features to enhance safety, performance, and visibility. To take full advantage of them, please acquaint yourself with this manual and keep it handy for future reference. If you are unfamiliar with how fume hoods operate, please review Chapter 4: Performance Features and Safety Precautions before you begin working in the fume hood. Even if you are an experienced fume hood user, please review Chapter 5: Using Your Fume Hood, which describes your Fiberglass 30 features so that you can use the hood efficiently.
About This Manual
This manual is designed to help you install, use, and maintain your laboratory fume hood. Instructions for installing optional equipment on your hood are also included.
Chapter 1: Introduction provides a brief overview of the laboratory fume hood, explains the organization of the manual, and defines the typographical conventions used in the manual.
Chapter 2: Prerequisites explains what you need to do to prepare your site before you install your laboratory fume hood. Electrical and service requirements are discussed.
Page 6
Chapter 1: Introduction
Product Service 1-800-522-7658
2
Original instructions
Chapter 3: Getting Started contains the information you need to properly unpack, inspect, install, and certify your laboratory fume hood.
Chapter 4: Performance Features and Safety Precautions explains how the laboratory fume hood operates and the appropriate precautions you should take when using the fume hood.
Chapter 5: Using Your Fiberglass 30 discusses the basic operation of your fume hood. Information on how to prepare, use and shut down your fume hood is included.
Chapter 6: Maintaining Your Fiberglass 30 explains how to perform routine maintenance on your fume hood.
Chapter 7: Modifying Your Fiberglass 30 explains how to modify the fume hood or add accessories.
Chapter 8: Troubleshooting contains a table of problems you may encounter while using your laboratory fume hood including the probable causes and suggested corrective actions.
Appendix A: Fiberglass 30 Hood Components contains labeled diagrams of all of the components of the fume hoods.
Appendix B: Fiberglass 30 Hood Dimensions contains comprehensive diagrams showing all of the dimensions for the laboratory fume hoods.
Appendix C: Fiberglass 30 Hood Specifications contains the electrical requirements for laboratory fume hood. Wiring diagrams are also included.
Appendix D: References lists the various resources available that deal with laboratory fume hoods.
Typographical Conventions
Recognizing the following typographical conventions will help you understand and use this manual:
Book, chapter, and section titles are shown in italic type (e.g., Chapter 3:
Getting Started).
Steps required to perform a task are presented in a numbered format. Comments located in the margins provide suggestions, reminders, and
references.
Critical information is presented in boldface type in paragraphs that are
preceded by the exclamation icon. Failure to comply with the information following an exclamation icon may result in injury to the user or permanent damage to fume hood.
!
Page 7
Chapter 1: Introduction
Product Service 1-800-522-7658
3
Original instructions
Les informations critiques sont présentées en gras dans les paragraphes qui
sont précédés par l'icône d'exclamation. Ne pas se conformer aux informations qui suivent une icône d'exclamation peut résulter à la blessure de l'utilisateur ou à des dommages irréversibles de la hotte aspirante.
Critical information is presented in boldface type in paragraphs that are
preceded by the wrench icon. These operations should only be performed by a trained certifier or contractor. Failure to comply with the information following a wrench icon may result in injury to the user or permanent damage to your hood.
Les informations critiques sont présentées en gras dans les paragraphes qui
sont précédés par l'icône de clé plate. Ces opérations devraient être seulement exécutées par un professionnel agrée. L'échec pour se conformer aux informations qui suivent une icône de clé plate peut résulter à la blessure de l'utilisateur ou à des dommages irréversibles de la hotte.
Important information is presented in capitalized type in paragraphs that are
preceded by the pointer icon. It is imperative that the information contained in these paragraphs be thoroughly read and understood by the user.
CAUTION – See Manual. When this symbol is on a fume hood it indicates a
caution that is detailed in this manual.
PRUDENCE – Consulter le Manuel. Quand ce symbole est sur une hotte
aspirante, il indique une prudence qui est détaillée dans ce manuel.
CAUTION – Hot Surface. AVERTIR – Surface Chaude
CAUTION – See Manual. This symbol on the fume hood indicates the
possibility of a pinch hazard.
PRUDENCE – Consulter le Manuel. Ce symbole sur la hotte indique la
possibilité d'un risque de pincement.
!
Page 8
Chapter 1: Introduction
Product Service 1-800-522-7658
4
Original instructions
Your Next Step
If your Fume Hood needs to be installed, proceed to Chapter 2: Prerequisites to ensure your installation site meets all of the requirements. Then, go to Chapter 3: Getting Started for instructions on how to install your laboratory fume hood and make all of the necessary connections.
If you would like to review how laboratory fume hoods operate, go to Chapter 4: Performance Features and Safety Precautions.
For information on the operational characteristics of your laboratory fume hood, go to Chapter 5: Using Your Fiberglass 30 Hood.
If your laboratory fume hood is installed and you need to perform routine maintenance on the cabinet, proceed to Chapter 6: Maintaining Your Fiberglass 30 Hood.
For information on making modifications to the configuration of your fume hood, go to Chapter 7: Modifying Your Fiberglass 30 Fume Hood.
Refer to Chapter 8: Troubleshooting if you are experiencing problems with your fume hood.
Page 9
Product Service 1-800-522-7658
5
Original instructions
C
C
HHAAPPTTEERR
2
2
P
P
RREERREEQQUUIISSIITTEESS
Before you install your laboratory fume hood, you need to prepare your site for installation. Carefully examine the location where you intend to install your hood. You must be certain that the area is level and of solid construction. In addition, a dedicated source of electrical power must be located near the installation site.
Carefully read this chapter to learn the requirements for your installation site:
The location requirements. The support requirements. The exhaust requirements. The electrical power requirements. The service line requirements. The space requirements.
Refer to Appendix B: Fiberglass 30 Hood Dimensions for complete fume hood dimensions.
Refer to Appendix C: Fiberglass 30 Hood Specifications for complete laboratory fume hood electrical and environmental conditions, specifications and requirements.
If the unit is not operated as specified in this manual it may impair the protection provided by the unit.
Si l'unité n'est pas utilisée comme spécifié dans ce manuel il peut diminuer la protection fournie par l'unité.
!
Page 10
Chapter 2: Prerequisites
Product Service 1-800-522-7658
6
Original instructions
Location Requirements
The fume hood should be located away from traffic patterns, doors, windows, fans, ventilation registers, and any other air­handling device that could disrupt its airflow patterns. All windows in the room should be closed.
La hotte aspirante devrait être localisé loin des voies de circulation, des portes, des fenêtres, des ventilateurs, des bouches de ventilation, et de tout appareil qui pourrait interrompre ses voies de flux d'air. Toutes les fenêtres dans la pièce devraient être fermées.
Support Requirements
DO NOT install the fume hood on a cart, dolly, or mobile bench. ALL Fiberglass 30 Hood installations must be permanent and stationary. The supporting structure usually consists of a base cabinet and chemically resistant work surface. See Chapter 7 Modifying Your Fiberglass 30 Hood.
NE PAS installer la hotte sur un panier, chariot, ou banc mobile. TOUS fibre de verre de 30 installations hotte doit être permanent et stationnaire. La structure de support est généralement constitué d'une armoire de base et la surface de travail résistant aux produits chimiques. Voir Chapitre 7 Modification de la fibre de verre 30 Hood.
Exhaust Requirements
The exhaust duct connection has been designed for 6" nominal duct (6.625" OD) to allow for minimum static pressure loss while operating at 100 fpm face velocities. The 6" diameter exhaust duct also allows for proper transport velocities away from the hood in the 1000 fpm to 2500 fpm range. The proper exhaust volume and static pressure loss are listed next for each hood model:
Fiberglass 30 Hood with Integral Blower, 100 fpm, 310 cfm, capable of overcoming 0.12" external static pressure.
Fiberglass 30 Hood for use with Remote Blower, 100 fpm, 310 cfm, 0.14" static pressure loss.
!
!
Page 11
Chapter 2: Prerequisites
Product Service 1-800-522-7658
7
Original instructions
Fiberglass 30 Hood with Explosion Proof Blower, 100 fpm, 310 cfm, capable of overcoming 0.12" external static pressure.
Proper blower selection can be determined from these exhaust requirements and the total system static pressure loss. Contact Labconco Customer Service for assistance in sizing a blower.
Electrical Requirements
Other than explosion proof models, the Fiberglass 30 Hood models feature internal wiring for the incandescent light assembly, light switch and blower switch. If your unit has an integral blower, then internal wiring to the blower motor and blower switch is included. Explosion Proof blowers require on site wiring. All internal wiring is terminated at the single point wiring junction box for hook-up by a qualified electrician. Refer to Chapter 3: Getting Started and Appendix C: Fiberglass 30 Specifications for the wiring diagram for proper electrical installation.
Autre que les modèles d'explosion preuve, les modèles en fibre de verre 30 capot disposent câblage interne pour l'ensemble de lumière incandescente, interrupteur de lumière et commutateur de soufflerie. Si votre appareil dispose d'une soufflerie intégrée, alors le câblage interne du moteur du ventilateur et l'interrupteur du ventilateur est inclus. Explosion Proof ventilateurs exigent le câblage du site. Tout le câblage interne est terminé à l'câblage boîte de jonction unique point pour le branchement par un électricien qualifié. Reportez-vous au Chapitre 3: Mise en route et à l'Annexe C: fibre de verre 30 Spécifications pour le schéma de câblage pour l'installation électrique appropriée.
Service Line Requirements
All service lines to the laboratory fume hood should be ¼ inch outside diameter, copper (brass for natural gas), and equipped with an easily accessible shut-off valve, should disconnection be required. Recommended operating pressure is 40 PSI, with a maximum allowable pressure of 200 PSI. Consider a pressure regulator to reduce line pressure to 40 PSI. Please check with local codes for other requirements.
Space Requirements
The dimensions for the different models are shown in Appendix B: Fiberglass 30 Hood Dimensions.
Your Next Step
After you have determined that the location you have selected accommodates the installation and operational requirements of your fume hood, you are ready to begin installation. Proceed to Chapter 3: Getting Started.
Page 12
Product Service 1-800-522-7658
8
Original instructions
C
C
HHAAPPTTEERR
3
3
G
G
EETTTTIINNGG
S
S
TTAARRTTEEDD
Now that the site for your laboratory fume hood is properly prepared, you are ready to unpack, inspect, install, and certify your unit. Read this chapter to learn how to:
Unpack and move your Fiberglass 30 Hood. Set up the fume hood with the supporting structure and work surface. Connect to an exhaust system. Connect the electrical supply source. Connect the service lines. Seal the Fiberglass 30 Hood to the work surface. Arrange certification of your Fiberglass 30 Hood.
Depending upon which model you are installing, you may need common plumbing and electrical installation tools in addition to 5/16", 3/8", 7/16", and 1/2" wrenches, ratchets, sockets, a nut driver set, a flat-blade screwdriver, a Phillips screwdriver, and a carpenter level to complete the instructions in the chapter.
Unpacking Your Laboratory Fume Hood
Carefully remove the shrink-wrap or carton on your fume hood and inspect it for damage that may have occurred in transit. If your unit is damaged, notify the delivery carrier immediately and retain the entire shipment intact for inspection by the carrier.
DO NOT RETURN GOODS WITHOUT THE PRIOR AUTHORIZATION OF LABCONCO. UNAUTHORIZED RETURNS WILL NOT BE ACCEPTED.
IF YOUR HOOD WAS DAMAGED IN TRANSIT, YOU MUST FILE A CLAIM DIRECTLY WITH THE FREIGHT CARRIER. LABCONCO CORPORATION AND ITS DEALERS ARE NOT RESPONSIBLE FOR SHIPPING DAMAGES.
The United States Interstate Commerce Commission rules require that claims be filed with the delivery carrier within fifteen (15) days of delivery.
Page 13
Chapter 3: Getting Started
Product Service 1-800-522-7658
9
Original instructions
Do not discard the shipping skid or packing material for your fume hood until you have checked all of the components and installed and tested the unit.
Do not remove the fume hood from its shipping skid until it is ready to be placed into its final location. Move the unit by placing a flat, low dolly under the shipping skid, or by using a floor jack.
THE FIBERGLASS 30 HOOD MODELS WEIGH BETWEEN 96 TO 129 LBS. (44-59 KG). THE SHIPPING SKID ALLOWS FOR LIFTING WITH A MECHANICAL LIFT TRUCK OR FLOOR JACK. IF YOU MUST LIFT THE FUME HOOD MANUALLY, FOLLOW SAFE-LIFTING GUIDELINES.
LA FIBRE DE VERRE DE 30 MODELES HOOD PESENT ENTRE 96 A 129 LBS. (44-59 KG). LA PALETTE D'EXPÉDITION PERMET LEVAGE AVEC UN CHARIOT ELEVATEUR MECANIQUE DES SOLS JACK. SI VOUS DEVEZ DECROCHER LE HOTTE MANUELLE, SUIVRE LES DIRECTIVES SAFE-LEVAGE.
NORMALLY, THE FUME HOOD CAN BE SLID OFF A HYDRAULIC LIFT TABLE AND BE PLACED INTO POSITION ON TOP OF THE WORK SURFACE. DO NOT LIFT BY THE FRONT AIR FOIL.
NORMALEMENT, LES HOTTE PEUT ETRE GLISSE OFF UN ASCENSEUR HYDRAULIQUE ET TABLE ETRE PLACE DANS POSITION SUR LE DESSUS DE LA SURFACE DE TRAVAIL. NE PAS LEVER DU FRONT AIR FOIL.
Do not move the hood by tilting it onto a hand truck.
Ne déplacez pas le capot en l'inclinant sur un camion de main.
Removing the Shipping Skid
LEAVE THE FUME HOOD ATTACHED TO ITS SHIPPING SKID UNTIL IT IS AS CLOSE TO ITS FINAL LOCATION AS POSSIBLE. MOVE THE HOOD BY USING A SUITABLE FLOOR JACK, OR BY PLACING A FURNITURE DOLLY UNDERNEATH THE SKID. DO NOT MOVE THE HOOD BY TILTING IT ONTO A HAND TRUCK.
!
!
Page 14
Chapter 3: Getting Started
Product Service 1-800-522-7658
10
Original instructions
After you verify the fume hood components, move your hood to the location where you want to install it.
Sash Weight Release
To protect the fume hood from damage in shipment, the sash weight has been secured to the back of the fume hood with four (4) screws. Simply remove the screws and make sure the sash cables are on the pulleys before operation of the sash.
NOTE: THE SASH WEIGHT ITSELF WAS INDIVIDUALLY MATCHED FOR THIS SPECIFIC HOOD AND SHOULD NOT BE EXCHANGED ON ANY OTHER UNIT.
Install the Fiberglass 30 Hood on a Supporting Structure and Work Surface
The Fiberglass 30 Hood is heavy! Use caution when lifting or moving the unit.
La fibre de verre 30 Hood est lourd! Soyez prudent lorsque vous soulever ou déplacer l'unité
When installing the Fiberglass 30 Fume Hood onto a chemically resistant work surface or benchtop, ensure that the structure can safely support the combined weight of the fume hood and any related equipment. The work surface should be at least as wide as the hood to properly support it. The work surface is aligned
with the back of the fume hood. This will provide the correct spacing under the air foil for proper bypass airflow.
WARNING: It is important to support the rear of the work surface and fume hood. The cross support provides support for the bottom of the work surface. Install the cross support after the base cabinets and work surface are leveled and before installing the hood.
AVERTISSEMENT : Il est important de soutenir l'arrière de la surface de travail et la hotte aspirante. Le support tranversal soutient le bas de la surface de travail. Installer le support transversal après que les meubles et la surface de travail soient nivelés et avant d'installer la hotte.
The following are instructions for mounting a cross support:
1. Level the base cabinets and the work surface. Work surface should be
placed flush with the back of the fume hood as shown in Figure 3-1.
2. Scribe a line on the wall or back of the base cabinet to locate the
support under the work surface.
!
!
Page 15
Chapter 3: Getting Started
Product Service 1-800-522-7658
11
Original instructions
3. Mount the support by attaching it to the wall or base cabinet.
4. Place the hood on top of the work surface and cross support.
The work surface should be smooth and durable, such as a chemically-resistant epoxy resin. The surface should be nonporous and resistant to the acids, solvents, and chemicals used.
Figure 3-1
Place Cross Support Here
Page 16
Chapter 3: Getting Started
Product Service 1-800-522-7658
12
Original instructions
Connecting to the Fiberglass 30 Hood Exhaust System
WARNING: The weight of the exhaust ductwork system must be supported independently of the hood superstructure. Do not allow this weight to be supported by the hood structure as damage to the hood may occur.
AVERTISSEMENT : Le poids du système d'aspiration de ductwork doit être soutenu d'une manière indépendante de la superstructure de la hotte. Au cas où ce poids est supporté par la structure de la hotte, des dommages à la hotte peuvent arriver.
The exhaust connection should be installed by a qualified HVAC contractor. The exhaust connection on your hood has been designed for 6" nominal pipe (6.625" OD) to allow for minimum static pressure loss with proper transport velocities away from the hood. Consult Labconco Customer Service should you require help sizing your blower for the exhaust volume and total system static pressure loss. See Chapter 2 for the hood airflow specifications.
Le raccord d'évacuation doit être installé par un entrepreneur en CVC qualifié. La connexion d'échappement sur votre hotte a été conçue pour 6 "diamètre nominal (6.625 "OD) pour permettre la perte de pression statique minimale avec des vitesses de transport appropriés à l'extérieur de la hotte. Consultez Labconco Service à la clientèle si vous avez besoin d'aide dimensionnement de votre ventilateur pour le volume de gaz d'échappement et système global de perte de pression statique. Voir le chapitre 2 pour le flux d'air de capot spécifications.
The selected exhaust duct material should match the hood procedures and chamicals used to ensure compatibility.
Le matériau de conduit d'évacuation choisi doit correspondre aux procédures de capot et chamicals utilisés pour assurer la compatibilité.
!
!
Page 17
Chapter 3: Getting Started
Product Service 1-800-522-7658
13
Original instructions
Connecting the Electrical Supply Source to the Fiberglass 30 Fume Hood
Prior to connecting any electrical wiring to the fume hood structure, refer to the hood identification plate for the proper electrical requirements of your specific model.
WARNING: The building electrical supply system for Fiberglass 30 Hoods is required to include overload protection. A switch or circuit breaker shall be in close proximity to the equipment and within easy reach of the operator. The switch or circuit breaker is to be marked as the disconnecting device for the equipment. Do not position the unit so that it is difficult to operate the main disconnect device.
AVERTISSEMENT : Le système d'alimentation électrique
pour la construction en fibre de verre 30 Hoods est nécessaire d'inclure la protection de surcharge. Un commutateur ou
disjoncteur doit être tout près de l'équipement et à portée facile de l'opérateur. Le commutateur ou le disjoncteur doit être marqué comme l'appareil débranchant pour l'équipement. Ne
pas placer l'appareil de sorte qu'il est difficile de faire fonctionner le dispositif de déconnexion principal.
The identification plate, model number, serial number, and electrical connection boxes are accessible from the front of the fume hood by removing the front panel.
The Fiberglass 30 Hood is normally wired for 115 Volt, 60 Hz, 20 Amp or 230 Volt, 50 Hz, 10 Amp electrical services. The number of circuits varies depending on the model. All of the electrical connections are terminated at the single point internal junction box for hook-up by a qualified electrician. Refer to the wiring diagram for your Fiberglass 30 Hood in Appendix C: Fiberglass 30 Fume Hood
Specifications.
!
Page 18
Chapter 3: Getting Started
Product Service 1-800-522-7658
14
Original instructions
Figure 3-2
All wiring for the fume hood SHOULD be performed by a licensed electrician and conform to all local and national electrical codes. In most cases, the hood will require the use of shielded conduit to protect the wiring into the hood. The grounding connection shall not be made to the terminal box cover.
Tout le câblage pour la hotte doit être effectuée par un électricien agréé et conforme à tous les codes électriques locaux et nationaux. Dans la plupart des cas, le capot ne nécessiter l'utilisation d'un conduit blindé pour protéger le câblage dans la hotte. La connexion de mise à la terre ne doit pas être fait à la couverture de la boîte à bornes.
The incandescent light has been mounted inside the top liner panel and is sealed from vapors inside the hood structure. To change the incandescent light bulbs in your hood, you must first remove the protective glass. Replace the defective bulbs, and reassemble.
Internal Junction Box
Page 19
Chapter 3: Getting Started
Product Service 1-800-522-7658
15
Original instructions
Connecting the Service Lines to the Fiberglass 30 Fume Hood
The hoods with service fixture kits are plumbed from the valve to the hose connector. Supply tubing shall be provided by the qualified installer. Tubing can enter the hood from above, through the back, or through the work surface to make these connections to the service fixtures.
NOTE: Inspect all fittings for leakage. Tighten the fittings slightly if needed.
NOTE: Inspectez tous les raccords pour les fuites. Serrer les raccords légèrement si nécessaire.
CAUTION: Do not use oxygen with any standard service fixture. Contact Labconco Customer Service for oxygen fixture information.
ATTENTION: Ne pas utiliser d'oxygène avec n'importe quel appareil de service standard. Contactez Labconco Service à la clientèle pour obtenir des informations de fixation de l'oxygène.
Should access to the hood plumbing fixture bodies be required, remove the service access plate on the side of the hood by loosening screws. The valve body will be fully exposed for any service work that may be necessary. The service fixture kits on your laboratory hood are designed for use with the following services:
Air
Hot Water
Vacuum
Cold Water
Natural Gas – See caution below
WARNING: Contact Labconco Customer Service directly before using any service other than those listed above in these valves to assure full compatibility.
AVERTISSEMENT: Contact Labconco directement au service client avant d'utiliser tout service autre que ceux énumérés ci­dessus dans ces vannes pour assurer une compatibilité totale.
!
!
!
Page 20
Chapter 3: Getting Started
Product Service 1-800-522-7658
16
Original instructions
CAUTION: Natural gas should be used only in the service fixture that has been pre-plumbed with brass tubing. Sulfur content of the gas could cause deterioration of standard copper supply lines.
PRUDENCE : Le gaz naturel devrait être seulement utilisé dans l'accessoire de service qui a été pré soudé avec des tuyaux de cuivre. Le contenu soufré du gaz pourrait causer la détérioration des lignes d’alimentation en cuivre standard.
Sealing the Fiberglass 30 Hood to the Work Surface
When the hood has been set in place, ducted, wired, and plumbed, it should be sealed at the work surface to prevent spilled materials from collecting under the walls of the hood. Materials such as silicone sealants are recommended to seal the hood structure.
Certifying the Fiberglass 30 Fume Hood
The combination of your laboratory hood, exhaust ductwork, and exhaust blower give you the flexibility to change the airflow at the sash opening of your hood. To determine the actual face velocity at the sash opening, airflow velocity readings will need to be taken. This should be done across the sash opening of the hood in accordance with the Industrial Ventilation Manual section on laboratory hoods. (See Appendix D – Reference) Labconco recommends an average face velocity at the sash opening of 80 to 100 feet per minute. Consult Chapter 2 for proper airflow volumes for your particular model.
The Fiberglass 30 Fume Hood has been tested per ASHRAE 110-1995. All hoods achieve an “as manufactured rating” of less than 0.10 part per million (ppm) at 4 liters per minute (lpm); AM<0.10 (consult Labconco for individual fume hood
ratings). For “field use” ASHRAE testing by a certified technician, contact
Labconco Customer Service.
NOTE: Face velocity profiles and smoke testing should be done periodically to ensure safe performance.
NOTE : Les profils de flux frontal et les tests de fumée devraient être régulièrement faits pour garantir une utilisation en toute securité.
!
!
Page 21
Chapter 3: Getting Started
Product Service 1-800-522-7658
17
Original instructions
Your Next Step
After your fume hood has been installed and certified, you are ready to proceed to Chapter 4: Performance Features and Safety Precautions.
Page 22
Product Service 1-800-522-7658
18
C
C
HHAAPPTTEERR
4
4
P
P
EERRFFOORRMMAANNCCE
E
F
F
EEAATTUURREESS
A
A
NNDD
S
S
AAFFEETTYY
P
P
RREECCAAUUTTIIOONNSS
Performance Features
The Fiberglass 30 Laboratory Hood is designed to meet the needs of the laboratory scientist. A fully featured by-pass hood with baffle and air foil can effectively contain toxic, noxious, or other harmful materials when properly installed. The by-pass airflow feature promotes full containment as the sash is moved.
The air foil sweeps the worksurface to remove heavier than air contaminants.
1. Unique sash provides maximum visibility of 28" high while conserving
energy by limiting sash travel to 20". Vertical-rising sash may be raised from a closed to 20" operating height. Exhaust volume, and blower sizing is based on the 20" height.
2. By-pass airflow design ensures containment as the sash moves.
3. Large usable interior work depth and interior height of 36" provides
ample working space.
4. Baffle (not shown) directs airflow to the rear of the interior to provide
efficient airflow. The baffle may be removed for cleaning purposes only.
5. Lift-Away™ front panel provides easy access to electrical wiring, and
sash weights.
6. Incandescent lighting provides ample illumination inside the fume hood.
No lighting is provided on explosion proof models.
7. Low mounted, factory-wired light and blower switches are ADA
compliant. No switches are provided on explosion-proof models.
Page 23
Chapter 4: Performance Features & Safety Precautions
Product Service 1-800-522-7658
19
Original instructions
8. Curved Air Foil allows air to sweep the work surface for maximum
containment.
WARNING: Do not lift hood in this area. Hood is to be lifted only by side or back.
AVERTISSEMENT: Ne pas soulever le capot dans ce domaine. Hotte doit être levée que par côté ou à l'arrière.
9. Corner posts provide maximum visibility and the flexibility to add
services after installation.
10. All hoods are factory prepared for up to 3 service fixtures. Field
installed by removing side access panel.
11. Removable side access panel for servicing valves and electrical
connections.
12. All hoods are factory prepared for one electrical duplex. A duplex can
be field mounted on the right corner post. Receptacles are field-wired to the single point junction box.
13. Shipped fully assembled and eliminates the need for costly onsite
assembly.
14. Accessory Guardian™ Digital Airflow Monitor or Guardian Jr.
Monitor continuously monitors face velocity. An audio/visual alarm alerts the user to low airflow conditions. The right corner post is factory prepared to accommodate the Guardian Monitor (sold separately).
15. Frame of epoxy-coated steel and aluminum is durable and corrosion
resistant.
16. Exhaust connection. The hood features 6" (6.625" OD pipe) exhaust
connections sized to allow for a minimum static pressure loss through the hood structure while providing a good transport velocity through the exhaust system.
17. Single point junction box provides a single connection point for electrical
wiring connections.
18. Convenient integral blower motor is readily accessible on integral blower
models.
!
Page 24
Chapter 4: Performance Features & Safety Precautions
Product Service 1-800-522-7658
20
Original instructions
Figure 4-1
16
18
17
5
15
14
11
10 9 7
12
8
2
1
Page 25
Chapter 4: Performance Features & Safety Precautions
Product Service 1-800-522-7658
21
Original instructions
Safety Precautions
Although the laboratory hood has been engineered to maintain optimum operator safety, caution should always be used while working in the hood. Prior to using the hood, check to make sure that the exhaust blower is operating and that air is entering the hood at its specified face velocity.
Bien que la hotte de laboratoire ait été réglée pour
maintenir la sécurité optimale de l’opérateur, la prudence
devrait toujours être utilisée en travaillant sous la hotte. Avant utiliser la hotte, le contrôle pour s'assurer que la soufflerie d'aspiration fonctionne et que cet air entre dans la hotte au flux spécifié.
USE GOOD HOUSEKEEPING IN THE HOOD AT ALL TIMES. CLEAN UP SPILLS IMMEDIATELY. PERIODICALLY CLEAN HOOD INTERIOR, INCLUDING LIGHT GLASS PANEL. REPLACE BURNED OUT LIGHT BULBS TO MAINTAIN MAXIMUM ILLUMINATION.
DO NOT OVERLOAD THE WORK SURFACE WITH APPARATUS OR WORK MATERIAL. THE SAFE OPERATION OF THE LABORATORY HOOD IS BASED UPON HAVING PROPER AIRFLOW THROUGH THE STRUCTURE. DO NOT PLACE LARGE, BULKY OBJECTS SUCH AS BLOCK HEATERS, DIRECTLY ON THE HOOD WORK SURFACE. INSTEAD, ELEVATE THE OBJECT 2" TO 3" ON BLOCKS TO ALLOW A FLOW OF AIR UNDER THE OBJECT AND INTO THE LOWER REAR BAFFLE EXHAUST SLOT. ENSURE BLOCKS ARE LEVEL AND SECURED IN PLACE.
Blocking the bottom of the baffle at the rear of hood will change the airflow pattern in the hood causing turbulence
and possible leakage at the face of the hood. (Don’t store
containers or supplies against the baffle, as this will affect airflow through the hood).
Le blocage de la partie inférieure de la chicane à l'arrière du capot va changer le modèle d'écoulement d'air dans la hotte provoquant une turbulence et une fuite possible à la face du capot. (Ne pas entreposer les contenants ou des fournitures contre le déflecteur, car cela aura une incidence sur le débit d'air à travers le capot).
!
!
Page 26
Chapter 4: Performance Features & Safety Precautions
Product Service 1-800-522-7658
22
Original instructions
Avoid placing your head inside hood. Keep hands out of hood as much as practical.
Evitez de placer votre tête dans le capot. Gardez les mains sur le capot autant que possible.
Always work as far back in the hood as possible. It is best to keep all chemicals and apparatus 6" inside the front of the hood.
Toujours travailler aussi loin dans le capot que possible. Il est préférable de garder tous les produits chimiques et appareils 6 "à l'intérieur de l'avant du capot.
Only explosion-proof models feature explosion-proof electrical components. Therefore, use of flammable or explosive materials in quantities above the explosive limit are not recommended.
Modèles antidéflagrants disposent composants électriques antidéflagrants. Par conséquent, l'utilisation de matériaux inflammables ou explosives dans des quantités supérieures à la limite d'explosion ne sont pas recommandés.
Do not work with chemicals in this hood without the exhaust system running. Do not store chemicals in a fume hood.
Ne pas travailler avec des produits chimiques dans la hotte sans que le système d'échappement en marche. Ne pas stocker de produits chimiques dans une hotte.
Perchloric acid use in this hood is prohibited.
Utilisation de l'acide perchlorique dans cette hotte est interdite.
High-level radioisotope materials are prohibited for use in this hood.
Matériaux de radio-isotopes de haut niveau sont interdits d'utilisation dans la hotte.
AVOID CROSS DRAFTS AND LIMIT TRAFFIC IN FRONT OF THE HOOD. AIR DISTURBANCES CREATED MAY DRAW FUMES OUT OF THE HOOD.
Page 27
Chapter 4: Performance Features & Safety Precautions
Product Service 1-800-522-7658
23
Original instructions
The use of heat-generating equipment in this hood without the exhaust system operating properly can cause damage to the hood.
L'usage d'équipement chaleur-produiant dans ce capuchon sans l'opération de système d'aspiration peut causer convenablement des dommages à la hotte.
The Fiberglass 30 Laboratory Hood should be certified by a qualified certification technician before it is initially used. The hood should be re-certified whenever it is relocated, serviced or at least annually thereafter.
La fibre de verre 30 Hotte de laboratoire doit être certifié par un technicien qualifié avant de certification, il est d'abord utilisé. La hotte doit être re-certifié à chaque fois qu'il est déplacé, l'entretien ou au moins chaque année par la suite.
Ensure that the unit is connected to electrical service in accordance with local and national electrical codes. Failure to do so may create a fire or electrical hazard. Do not remove or service any electrical components without first disconnecting the hood from electrical service.
Assurez-vous que l'appareil est connecté au service électrique conformément aux codes électriques locaux et nationaux. Ne pas le faire peut créer un risque d'incendie ou électrique. Ne pas enlever ou de réparer des composants électriques sans d'abord débrancher le capot du service électrique.
Proper operation of the fume hood depends largely upon the
hood’s location and the operator’s work habits. Consult the
Reference Manual in Appendix D.
Le bon fonctionnement de la hotte dépend largement de l'emplacement de la hotte et les habitudes de travail de l'opérateur. Consultez le manuel de référence à l'annexe D.
!
Page 28
Chapter 4: Performance Features & Safety Precautions
Product Service 1-800-522-7658
24
Original instructions
Do not touch the blower motor. The surfaces of the motor can become hot and could cause burns.
Ne pas toucher le moteur de soufflerie. Les surfaces du moteur peuvent devenir chaudes et pourrait causer des brûlures.
Your Next Step
After you understand the theory of operation and safety precautions, you are ready to proceed to Chapter 5: Using Your Fiberglass 30 Fume Hood.
!
Page 29
Product Service 1-800-522-7658
25
C
C
HHAAPPTTEERR
5
5
U
U
SSIINNG
G
Y
Y
OOUUR
R
F
F
IIBBEERRGGLLAASSS
S
3
3
0
0
Operating the Vertical-Rising Sash
Because of the Fiberglass 30 Hood counterbalanced sash mechanism, it will take only a few pounds of force to move the sash up or down, and you can operate the sash smoothly with one or two hands positioned anywhere along the handle. The vertical-rising sash may be raised to a maximum 20" operating height. The airflow requirements should be sized for the 20" operating height.
Operating the Blower
Your Fiberglass 30 Fume Hood utilizes a remote style blower or integral blower,
which can be activated by turning the blower switch to “ON.” You can validate
the hood performance by watching smoke drawn into the hood face opening. Explosion-proof models do not have a switch or wiring as this is field installed.
Operating the Lights
Your Fiberglass 30 Fume Hood is equipped with a factory-wired incandescent
light to illuminate the hood interior. Simply turn the light switch to “ON” to
operate. Explosion-proof models do not have a light.
Working in Your Fiberglass 30 Fume Hood
Planning
Thoroughly understand procedures and equipment required before
beginning work.
Arrange for minimal disruptions, such as room traffic or entry into the
room while the hood is in use.
Page 30
Chapter 5: Using Your Fiberglass 30
Product Service 1-800-522-7658
26
Original instructions
Start-up
Labconco follows the recommendations of the American National Standard Institute with regards to fume hood exhaust systems. A full list of requirements can be found in the ANSI Standard No. Z9.5 Laboratory Ventilation. If you are unsure of the proper use of the exhaust blower for your fume hood, contact your safety
officer or industrial hygienist before making any modifications to the blower’s
operational status (on/off/low, etc.). Keeping the exhaust system under constant negative pressure is an excellent way to ensure hazardous fumes in the fume hood do not enter the laboratory space inadvertently.
Turn on the incandescent light. Slowly raise the sash and verify the fume hood has proper airflow. Check the baffle air slots for obstructions. Allow the hood to operate unobstructed for two minutes. Wear a long sleeved lab coat and rubber gloves. Use protective eyewear.
Wear a protective mask if appropriate.
Loading Materials and Equipment
Only load the materials required for the procedure. Do not overload the
hood.
Do not obstruct the front air foil (sill), or rear baffle slots. Large objects should not be placed close together and spaced above the
work surface to permit airflow to sweep under the equipment.
After loading the hood, wait one minute to purge airborne contaminants
from the work area.
Work Techniques
Keep all materials at least 6 inches inside of the sash and at least two inches
from the side walls and perform all contaminated operations as far to the rear of the work area as possible.
Keep all clean and contaminated materials in the work area separate for
clear identification.
Avoid using techniques or procedures that disrupt the airflow patterns of
the hood.
Final Purging
Upon completion of the work, the hood should be allowed to operate for
two minutes undisturbed, to purge airborne contaminants from the work area before closing the sash.
Unloading Materials and Equipment
Objects in contact with contaminated material should be surface
decontaminated before removal from the hood.
All open trays or containers should be covered before being removed from
the hood.
Page 31
Chapter 5: Using Your Fiberglass 30
Product Service 1-800-522-7658
27
Original instructions
Shutdown
Turn off the incandescent light and close the sash.
Your Next Step
After you understand how to operate and work in the fume hood, you are ready to proceed to Chapter 6: Maintaining Your Fiberglass 30 Fume Hood.
Page 32
Product Service 1-800-522-7658
28
C
C
HHAAPPTTEERR
6
6
M
M
AAIINNTTAAIINNIINNG
G
Y
Y
OOUUR
R
F
F
IIBBEERRGGLLAASSS
S
3
3
0
0
Now that you have an understanding of how to work in the fume hood, we will review the suggested maintenance schedule and the common service operations necessary to maintain your fume hood for peak performance.
Only trained and experienced certification technicians should perform some of the service operations after the fume hood has been properly decontaminated. DO NOT attempt to perform these operations if you are not properly trained. The wrench icon precedes the service operations that require qualified technicians.
Techniciens de certification Seulement formés et expérimentés devraient effectuer certaines des opérations de service après la hotte a été correctement décontaminés. NE PAS tenter de réaliser ces opérations si vous n'êtes pas correctement formés. L'icône de la clé précède les opérations de service qui nécessitent des techniciens qualifiés
Routine Maintenance Schedule
Weekly
Using ordinary dish soap to clean the surface inside of the fume hood, and
the work surface.
Using an appropriate glass cleaner, clean the sash and all glass surfaces. Operate the fume hood blower, noting the airflow velocity through the
hood using a source of visible smoke.
!
Page 33
Chapter 6: Maintaining Your Fiberglass 30
Product Service 1-800-522-7658
29
Original instructions
Monthly (or more often as required)
Mensuel (ou plus souvent si nécessaire)
Determine the actual face velocity through the sash opening of the hood
where the average reading should be at the specified velocity. (Use calibrated thermal anemometer or other approved apparatus).
Déterminer la vitesse de visage réel à travers l'ouverture de la guillotine de
la hotte où la valeur moyenne doit être à la vitesse spécifiée. (Utilisez calibré anémomètre thermique ou un autre appareil approuvé)
Using a damp cloth, clean the exterior surfaces of the hood, particularly the
front of the hood, to remove any accumulated dust.
Utiliser un chiffon humide, nettoyez les surfaces extérieures de la hotte, en
particulier l'avant du capot, pour enlever toute la poussière accumulée.
Check all service valves, if so equipped, for proper operation. Vérifiez que toutes les vannes de service, si équipé, pour un bon
fonctionnement.
The hood baffles should be checked for blockages behind them to ensure
that the hood is maintaining proper airflow.
Les déflecteurs de capot doivent être vérifiés pour les blocages derrière eux
pour veiller à ce que le capot est de maintenir une ventilation adéquate.
All weekly activities. Toutes les activités hebdomadaires.
Annually
Annuellement
Have the fume hood recertified by a qualified certification technician. See
Certifying the Fiberglass 30 Fume Hood in Chapter 3
Avez la hotte recertification par un technicien de certification qualifié. Voir
la certification de fibre de verre 30 Hotte au chapitre 3.
All monthly activities. Toutes les activités mensuelles
Biannually
The sash assembly should be checked to ensure proper operation and to
make sure there are no signs of abnormal wear on the sash pulleys, cables and clamps.
Routine Service Operations
Front Panel Removal:
1. Simply lift the front panel up and then away from the hood to provide access to
the top.
Page 34
Chapter 6: Maintaining Your Fiberglass 30
Product Service 1-800-522-7658
30
Original instructions
Changing the Incandescent Lamp:
1. Turn light switch to “OFF”.
2. Raise the sash and reach inside the fume hood to unscrew the clear glass globe
that protects the incandescent bulb.
3. Install the new incandescent bulb and then reverse the removal procedure.
Do not contact blower wheel while still in motion
Your Next Step
After you understand the maintenance procedures, you are ready to proceed to Chapter 7: Modifying Your Fiberglass 30 Fume Hood.
!
Page 35
Product Service 1-800-522-7658
31
C
C
HHAAPPTTEERR
7
7
M
M
OODDIIFFYYIINNGG
Y
Y
OOUUR
R
F
F
IIBBEERRGGLLAASSS
S
3
3
0
0
There are several ways to modify the fume hood for your individual requirements. These include the addition of work surfaces, service fixtures, air monitor, and electrical duplex outlets. See Appendix A: Fiberglass 30 Hood Components for installation location.
Installing Work Surfaces and Storage Cabinets
An optional work surface is available when installing the hood on storage cabinets. Contact Labconco Customer Service for ordering information on these 30" work surfaces and storage cabinets.
4882806
Work Surface
30" x 30"
9902200
Solvent Cabinet
30"
9901200
Acid Cabinet
30"
9900200
Base Cabinet
30"
Installing Additional Service Fixtures
Additional service fixtures can be installed in the available service fixture holes in the right sidewall and corner post. The fume hood is factory set to accept up to three valves. Contact Labconco Customer Service for information. The following service fixture kits are available:
Page 36
Chapter 7: Modifying Your Fiberglass 30
Product Service 1-800-522-7658
32
Original instructions
2832600
Cold Water (CW) – Green
2832601
Hot Water (HW) – Red
2832602
Deionized Water (DW) – White
2832603
Vacuum (VAC) – Yellow
2832604
Air (AIR) – Orange
2832605
Gas (GAS) – Blue
2832606
Nitrogen (NIT) – Brown
2832607
Argon (ARG) – Gray
2832608
Steam (STM) - Black
Installing Guardian™ Digital Airflow
Monitor or Guardian™ Jr. Airflow Monitor
The Guardian Digital Airflow Monitor P/N 9743201 and 9743205 or Guardian Jr. Airflow Monitor P/N 9743202 and 9743206 allows you to continuously monitor face velocity through the fume hood opening. The fume hood right corner post is factory prepared to mount either monitor. Contact Labconco Customer Service to order.
Figure 7-3 – Hose Connector
Figure 7-1 - Knob
Figure 7-2 – Valve
Figure 7-4
Figure 7-5
Page 37
Chapter 7: Modifying Your Fiberglass 30
Product Service 1-800-522-7658
33
Original instructions
Installing an Electrical Duplex Receptacle
If you ordered your hood and want an electrical duplex outlet, then you can have one installed in the field by a qualified electrician. Contact Labconco Customer Service for ordering information. (Not acceptable on explosion-proof hoods). The following duplex kits are available:
Si vous avez commandé votre hotte et que vous voulez une sortie duplex électrique, alors vous pouvez avoir un installé sur le terrain par un électricien qualifié. Contactez Labconco Service à la clientèle pour les informations de commande. (Non acceptable sur cagoules anti-explosion). Les kits recto verso suivantes sont disponibles:
2834800 115V, 20A Duplex Kit 2834801 115V, 20A GFCI Duplex Kit
Figure 7-6
Your Next Step
After you understand the modifying procedures, you are ready to proceed to Chapter 8: Troubleshooting.
Page 38
Product Service 1-800-522-7658
34
C
C
HHAAPPTTEERR
8
8
T
T
RROOUUBBLLEESSHHOOOOTTIINNGG
Refer to the following table if your fume hood fails to operate properly. If the suggested corrective actions do not solve your problem, contact Labconco for additional assistance.
PROBLEM
CAUSE
CORRECTIVE ACTION
Remote blower and light won’t operate
Wires not connected at junction box or switches.
Check connection of switches. Check connection to control box on top of unit.
Circuit breakers tripped in building electrical supply.
Reset circuit breakers.
Remote blower
won’t operate, but
light works
Blower wiring is disconnected. Belt broken.
Inspect blower wiring and switch.
Replace belt.
Blower motor is defective.
Replace blower motor.
Fume hood blower operates but light will not operate
Lamp not installed correctly.
Inspect lamp installation.
Lamp is defective.
Replace lamp.
Circuit breaker in building is tripped.
Reset the circuit breaker.
Fume hood blower operates but light will not operate
Lamp wiring is disconnected.
Inspect lamp wiring.
Defective lamp.
Replace lamp.
Page 39
Chapter 8: Troubleshooting
Product Service 1-800-522-7658
35
Original instructions
PROBLEM
CAUSE
CORRECTIVE ACTION
Contaminants outside of fume hood
Improper user techniques for the fume hood.
See “Certifying the Hood” Chapter 3 and “Safety Precautions” Chapter 4 sections in
the manual. (Ref. Appendix D)
Restriction of the baffle air slots or blockage of the exhaust outlet.
Remove baffles to ensure that all air slots and the exhaust outlet are unobstructed.
External factors are disrupting the fume hood airflow patterns or acting as a source of contamination.
See “Location Requirements” Chapter 2, “Certifying the Hood” Chapter 3, and “Safety Precautions” Chapter 4 sections of
this manual. (Ref. Appendix D)
Fume hood has improper face velocity.
Have fume hood re-certified and check remote blower exhaust system. Hood should have average face velocity of 80­100 fpm.
Vertical-rising sash no longer operates smoothly
Cable is frayed or plastic protection is damaged.
Inspect cable and replace cable if worn or damaged immediately; otherwise injury could result.
Pulley bearing is damaged.
Replace pulley, bearing or add grease.
Cable has slipped off the pulleys.
Re-install, cable must be replaced immediately if damaged.
Weight has broken pulleys.
Replace weight pulleys.
Electrical duplex outlet no longer has power
Wires not connected or faulty duplex.
Check wire connection or replace duplex.
Circuit breakers tripped in building electrical supply.
Reset circuit breakers.
Service valves no longer operate
Faulty building supply.
Inspect building supply shut off valves and appropriate pressures below the recommended 40 PSI.
Valve no longer operates.
Replace valve and check for leaks.
Supply line or outlet line has leaks.
Inspect line for leaks and fix any leaking plumbing connections.
Page 40
Product Service 1-800-522-7658
36
A
A
PPPPEENNDDIIXX
A
A
F
F
IIBBEERRGGLLAASSS
S
3
3
0
0
H
H
OOOOD
D
C
C
OOMMPPOONNEENNTTSS
Illustration A-1 indicate the location of the following service parts:
Item
Quantity
Part No.
Description
1A 1 9823700
Valve, Labconco (Water) 1/4" Compression Fitting
1B 1 9817000
Valve, Labconco 1/4" Compression Fitting (AIR, GAS, VAC, NIT, etc.)
1C 1 9823702
Valve, Labconco Deionized 1/4" Compression Fitting
1D 1 9818000
Nut, Valve Mtg. (Labconco)
2
1
9818700-01-02-03-
04-05-06-07-08
Knobs (Gray, Green, Blue, Orange, Yellow, Red, White, Black, Brown)
3A 1 9818800
Hose Barb, Gray – Neutral or Argon – NOT SHOWN
3B 1 9818801
Hose Barb, Green – Cold Water – NOT SHOWN
3C 1 9818802
Hose Barb, Blue – Gas – NOT SHOWN
3D 1 9818803
Hose Barb, Orange – Air – NOT SHOWN
3E 1 9818804
Hose Barb, Yellow – Vacuum – NOT SHOWN
3F 1 9818805
Hose Barb, Red – Hot Water – NOT SHOWN
3G 1 9818806
Hose Barb, White – Deionized Water – NOT SHOWN
3H 1 9818807
Hose Barb, Black – Neutral or Steam – NOT SHOWN
3I 1 9818808
Hose Barb, Brown – Nitrogen – NOT SHOWN
3J 1 9819000
Nut, Hose Barb – NOT SHOWN
4 1 9825500
Label, Knob (contains all labels)
5 1 9818900
Lens, Knob
6A 1 2832600
Fixture Kits – (CW)
6B 1 2832601
Fixture Kits – (HW)
6C 1 2832602
Fixture Kits – (DI)
6D 1 2832603
Fixture Kits – (VAC)
6E 1 2832604
Fixture Kits – (AIR)
6F 1 2832605
Fixture Kits – (GAS)
6G 1 2832606
Fixture Kits – (NIT)
6H 1 2832607
Fixture Kits – (ARG)
6I 1 2832608
Fixture Kits – (STM)
7A 1 2834800
115V, 20A Duplex Kit
7B 1 2834801
115V, 20A GFCI Duplex Kit
7C 1 9818200
Cover Plate 115V Duplex
7D 1 9818100
Cover Plate, 115V GFCI
7E 1 9818300
Wallplate, Blank
8A 1 1273600
Bulb, 115V
8B 1 1273700
Bulb, 230V
8C 1 4849800
Light Fixture Assy complete with wiring, parts 8D through 8G
8D 1 4849900
Light Fixture – a) Body & Socket VFB15, b) Pendent Cover 1/2" VP1, c) Globe, 150W, 264–0006-9909
Fiberglass 30 Hood Replacement Parts
Page 41
Appendix A: Fiberglass 30 Components
Product Service 1-800-522-7658
37
Original instructions
Item
Quantity
Part No.
Description
8E 1 4850000
Wiring Harness, Light
8F 1 1279700
Gasket, Base
8G 1 1279800
Gasket, Globe
9A 1 1302300
Switch, Rocker
9B 1 1327500
Switch, Plug (Fills cutout without switch)
10A 1 2834400
Wiring Harness, Main 115V
10B 1 2834500
Wiring Harness, Main 230V
11 1 2834000
Label Set, Corner Post
12 1 2834100
Label, Front Panel
13 1 2831400
Front Panel
14A 1 2833000
Motor Plate – 115V only
14B 1 2831700
Motor Plate, 230V, 115V, EP - 230V, EP
15A 1 2807100
Motor, 115V
15B 1 1203600
Motor, 230V
15C 1 1200400
Motor, 115 VEP
15D 1 1211000
Motor, 230V EP
15E 1 1852300
Reducing Bushing – 230V only
15F 4 1662600
Rubber Grommet – 115V only
16A 1 1450000
Wheel, Blower – 115V only
16B 1 4716200
Wheel, Blower – 230V, 115V – EP, 230V – EP
17 1 2833400
Removable Side Panel – Right
18 1 2831900
Side Panel – Right
19 1 2832000
Side Panel – Left
20 1 2832500
Sash Glass Assembly
21 1 2833800
Sash Weight
22A 1 1861400
Pulley – 1-3/16" Dia.
22B 1 3627500
Thrust Washer
23 2 4949902
Cable, Sash
24 4 1920100
Clamp, Cable Replacement
25 2 1972100
S-Hook
26 1 2831500
Air Foil
27 2 1934300
Air Foil Spacer Stop
28 2 9713300
Rubber Bumper – Sash Handle
29A 1 2833500
Bracket, Cable Front – Right
29B 1 2833501
Bracket, Cable Front – Left
29C 1 1663200
Rubber Bumper – Upper Sash Stop
30 1 2831100
Corner Post, Right
31 1 2831200
Corner Post, Left
32 1 2831300
Header
30 1 9724500
Sash Stop Kit
31 1 1663200
Upper Sash Bumper
32 1 2831300
Header
Page 42
Appendix A: Fiberglass 30 Components
Product Service 1-800-522-7658
38
Original instructions
24
12
32
20
28
Figure A-1
19
23
29
22
16 Not Shown
15
14
21 Not Shown
24 Not Shown
25 Not Shown
31
27 Not shown
26 9 10 Not shown
7
1 through 6 Not Shown
7E for Air Monitor
30
17
18
8 Not Shown
11
Page 43
Product Service 1-800-522-7658
39
A
A
PPPPEENNDDIIXX
B
B
F
F
IIBBEERRGGLLAASSS
S
3
3
0
0
H
H
OOOOD
D
D
D
IIMMEENNSSIIOONNSS
Figure B-1
Page 44
Appendix B: Fiberglass 30 Hood Dimensions
Product Service 1-800-522-7658
40
Original instructions
Figure B-2
Page 45
Appendix B: Fiberglass 30 Hood Dimensions
Product Service 1-800-522-7658
41
Original instructions
Figure B-3
DIMENSION IN INCHES
Page 46
Appendix B: Fiberglass 30 Hood Dimensions
Product Service 1-800-522-7658
42
Original instructions
Figure B-4
DIMENSION IN INCHES
Page 47
Appendix B: Fiberglass 30 Hood Dimensions
Product Service 1-800-522-7658
43
Original instructions
Figure B-5
Page 48
Product Service 1-800-522-7658
44
Original instructions
A
A
PPPPEENNDDIIXX
C
C
F
F
IIBBEERRGGLLAASSS
S
3
3
0
0
H
H
OOOOD
D
S
S
PPEECCIIFFIICCAATTIIOONNSS
Environmental Conditions
Indoor use only. Maximum altitude: 10,000 feet (3,048 meters). Ambient temperature range: 41° to 104°F (5° to 40°C). Maximum relative humidity: 80% for temperatures up to 88°F (31°C),
decreasing linearly to 50% relative humidity at 104°F (40°C).
Main supply voltage fluctuations not to exceed ±10% of the nominal
voltage.
Transient over-voltages according to Installation Categories II (Over-
voltage Categories per IEC 1010). Temporary voltage spikes on the AC input line that may be as high as 1500V for 115V models and 2500V for 230V models are allowed.
Used in an environment of Pollution degrees 2 (i.e., where normally only
non-conductive atmospheres are present). Occasionally, however, a temporary conductivity caused by condensation must be expected, in accordance with IEC 664.
Model
Electrical Requirements
3030000
115 VAC – 60 Hz, 1 Phase – 5 Amps
3030001
230 VAC – 50 Hz, 1 Phase – 5 Amps
3030002
115 VAC – 60 Hz, 1 Phase – 1 Amp
3030003
230 VAC – 50 Hz, 1 Phase – 1 Amp
3030004
115 VAC – 60 Hz, 1 Phase – 1 Amp
3030005
230 VAC – 50 Hz, 1 Phase – 1 Amp
Page 49
Appendix C: Fiberglass 30 Specifications
Product Service 1-800-522-7658
45
Original instructions
Page 50
Appendix C: Fiberglass 30 Specifications
Product Service 1-800-522-7658
46
Original instructions
Page 51
Product Service 1-800-522-7658
47
Original instructions
A
A
PPPPEENNDDIIXX
D
D
R
R
EEFFEERREENNCCEES
S
Many excellent reference texts and booklets are currently available. The following is a brief listing:
Laboratory Ventilation Standards
Federal Register 29 CFR Part 1910
Non-mandatory recommendations from “Prudent Practices”.
Fume hoods should have a continuous monitoring device Face velocities should be between 60-100 linear feet per minute (lfpm) Average 2.5 linear feet of hood space per person
Occupational Health and Safety U.S. Department of Labor 200 Constitution Avenue N.W. Washington, DC 20210 (202) 523-1452
Industrial Ventilation-ACGIH
Fume hood face velocities between 60-100 lfpm Maximum of 125 lfpm for radioisotope hoods Duct velocities of 1000-2000 fpm for vapors, gasses and smoke Stack discharge height 1.3-2.0 x building height Well designed fume hood containment loss, <0.10 ppm
Industrial Ventilation, A Manual of Recommended Practice. 24th Edition, 2001 American Conference of Governmental Industrial Hygienists 1330 Kemper Meadow drive Cincinnati, OH 45240-1634 (513) 742-2020
ASHRAE 110-1995 Method of Testing Performance of Fume Hoods
Evaluates fume hood’s containment characteristics
Three part test: Smoke generation, Face velocity profile, Tracer gas release @ 4 liters
per minute
Rated As Manufactured (AM), As Installed (AI) and As Used (AU)
American Society of Heating, Refrigerating, and Air Conditioning Engineers 1791 Tullie Circle N.E. Atlanta, GA 30329 (404) 636-8400
Page 52
Appendix C: Fiberglass 30 Specifications
Product Service 1-800-522-7658
48
Original instructions
ANSI Z9.5-2011 Laboratory Standard
Covers entire laboratory ventilation system.
New and remodeled hoods shall have a monitoring device Ductless hoods should only be used with non-hazardous materials
American Industrial Hygiene Association 2700 Prosperity Avenue, Suite 250 Fairfax, VA 22031 (703) 849-8888
SEFA 1-2002
Fume hood face velocities based on toxicity levels of chemicals
Class A – 125 to 150 fpm Class B – 80 to100 fpm Class C – 75-to 80 fpm
Test method – face velocity profile and smoke generation
Scientific Equipment & Furniture Association 1028 Duchess Drive McLean, VA 22102 (703) 538-6007
NFPA 45 – 2002 Fire Protection for Laboratories Using Chemicals
Laboratory hoods should not be relied on for explosion protection Exhaust air from fume hoods should not be recirculated Services should be external to the hood Canopy hoods only for non-hazardous applications Materials of construction should have flame spread of 25 or less 80 to 120 fpm to prevent escape
NFPA 30 – 2000 Flammable and Combustible Liquids Code
Approved cabinets may be metal or wood Vent location on cabinets are required Venting of cabinets not a requirement
National Fire Protection Association 1 Batterymarch Park P.O. Box 9101 Quincy, MA 02269-9101 (800) 344-3555
General References
American Conference of Governmental Industrial Hygienists. Industrial Ventilation, A Manual of Recommended Practice, Cincinnati, OH
ASHRAE Standard Committee. ASHRAE Standard Atlanta: ASHRAE Publications Sales Department, 1995
British Standards Institution, Laboratory Fume Cupboards. Parts 1, 2 and 3, London: 1990
Department of Labor, Occupational Safety and Health Administration, 29 CFR Part 1910, Occupational Exposures to Hazardous Chemicals in Laboratories, Final Rule. Vol. 55, No. 21. Washington D.C.:1990
Page 53
Appendix C: Fiberglass 30 Specifications
Product Service 1-800-522-7658
49
Original instructions
DiBerardinis. L. et al. Guides for Laboratory Design, Health and Safety Considerations. Wiley & Sons, 1987
McDermott, Henry, Handbook of Ventilation for Contaminant Control, 2nd Edition. Butterworth Publishers, 1985.
Miller, Brinton M. et al. Laboratory Safety: Principles and Practices. American Society for Microbiology, Washington, D.C.: 1986
NIH Guidelines for the Laboratory Use of Chemical Carcinogens. NIH Publication No. 81-2385.
Rayburn, Stephen R. The Foundation of Laboratory Safety, A Guide for the Biomedical Laboratory. Springer-Verlag, New York: 1990
Sax, N. Irving and Lewis, JR., Richard J. Rapid Guide to Hazardous Chemicals in the Workplace. Van Nostrand Reinhold, 1987.
Schilt, Alfred A. Perchloric Acid and Perchlorates. The G. Frederick Smith Chemical Company, Columbus, OH: 1979.
Steere, Norman. CRC Handbook of Laboratory Safety, 2nd Edition. CRC Press, 1971.
Department of Labor, Occupational Safety and Health Administration, 29 CFR Part 1910, Occupational Exposures to Hazardous Chemicals in Laboratories, Final Rule. Vol. 55, No. 21. Washington D.C.:1990
DiBerardinis. L. et al. Guides for Laboratory Design, Health and Safety Considerations. Wiley & Sons, 1987
McDermott, Henry, Handbook of Ventilation for Contaminant Control, 2nd Edition. Butterworth Publishers, 1985.
Miller, Brinton M. et al. Laboratory Safety: Principles and Practices. American Society for Microbiology, Washington, D.C.: 1986
NIH Guidelines for the Laboratory Use of Chemical Carcinogens. NIH Publication No. 81-2385.
Rayburn, Stephen R. The Foundation of Laboratory Safety, A Guide for the Biomedical Laboratory. Springer-Verlag, New York: 1990
Sax, N. Irving and Lewis, JR., Richard J. Rapid Guide to Hazardous Chemicals in the Workplace. Van Nostrand Reinhold, 1987.
Schilt, Alfred A. Perchloric Acid and Perchlorates. The G. Frederick Smith Chemical Company, Columbus, OH: 1979.
Steere, Norman. CRC Handbook of Laboratory Safety, 2nd Edition. CRC Press, 1971.