Labconco 1300 Series A2, eq_manual_thermo_1300_series_class_ii__a2 Sistemas para archivo y comunicación de imágenes (PACS)

Page 1
Visit us online to register your warranty www.thermoscientific.com/labwarranty
User Manual
1300 Series A2
Class II, Type A2 Biological Safety Cabinet
Operating Manual 7021355 Rev. 13
Page 2
Thermo Scientific
MANUAL NUMBER 7021355
Preface
1300 Series A2 i
This manual covers the models shown below:
Dear User,
Congratulations on your purchase of a Thermo Scientific 1300 Series Class II, Type A2 biological safety cabinet! Your 1300 Series A2 biological safety
cabinet has been tested and certified in accordance to NSF/ANSI 49, and is designed to protect the user, the environment, and your research from
harmful substances and cross-contamination. This user’s manual provides instructions on how to use the 1300 Series A2 most effectively and safely. A
printed copy of this user’s manual is available from Technical Services, if required.
Containment * Comfort * Convenience™
The 1300 Series A2 offers a unique range of product features that will enhance your safety and improve overall operating efficiency. Should you have
any questions on using this product or need further explanation of any of it’s features, please contact Technical Services (Page iv), or your local
distributor.
Stainless Steel Interior Coated Interior Size Voltage
10 inch opening 8 inch opening 10 inch opening 8 inch opening
1322 1341 1320 1339 3 ft 120V
1323* 1342* 1321* 1340* 3 ft 120V
1345 1305 1355 1325 4 ft 120V
1346 1306 1356 1326 4 ft 230V
1354 4 ft 230V, China
1384* 4 ft 230V, China
1370 1352 1368 1350 5 ft 120V
1371* 1353* 1369* 1351* 5 ft 120V
1347 1307 1357 1327 6 ft 120V
1348 1308 1358 1328 6 ft 230V
1359 6 ft 230V, China
1389* 6 ft 230V China
1332 1392 1330 1390 3 ft 230V
1333* 1393* 1331* 1391* 3 ft 230V
1375* 1335* 1385* 1395* 4 ft 120V
1376* 1336* 1386* 1396* 4 ft 230V
1382 1366 1380 1372 5 ft 230V
1383* 1367* 1381* 1373* 5 ft 230V
1377* 1337* 1387* 1397* 6 ft 120V
1378* 1338* 1388* 1398* 6 ft 230V
* Includes cabinet, adjustable height stand, UV light and armrest.
Page 3
Thermo Scientificii 1300 Series A2
Preface
Important Read this instruction manual. Failure to read, understand and follow the instructions in this manual may result in damage to the unit, injury to operating personnel, and poor equipment performance.
s
Caution All internal adjustments and maintenance must be performed by qualified service personnel. s
Material in this manual is for information purposes only. The contents and the product it describes are subject to change without notice. Thermo Fisher Scientific makes no representations or warranties with respect to this manual. In no event shall Thermo Fisher Scientific be held liable for any damages, direct or incidental, arising out of or related to the use of this manual.
©2011 Thermo Fisher Scientific. All rights reserved.
CAUTION
Contains Parts and Assemblies
Susceptible to Damage by
Electrostatic Discharge (ESD)
Page 4
Thermo Scientific 1300 Series A2 iii
Preface
Important operating and/or maintenance instructions. Read the accompanying text carefully.
Potential electrical hazards. Only qualified persons should perform procedures associated with this symbol.
Equipment being maintained or serviced must be turned off and locked off to prevent possible injury.
Hot surface(s) present which may cause burns to unprotected skin, or to materials which may be damaged by elevated temperatures.
WEEE Compliance: Thermo Fisher Scientific has contracted with companies for recycling/disposal in each EU Member State. For further information, send an email to weee.recycle@thermofisher.com.
4 Always use the proper protective equipment (clothing, gloves, goggles, etc.)
4 Always dissipate extreme cold or heat and wear protective clothing.
4 Always follow good hygiene practices.
4 Each individual is responsible for his or her own safety.
Page 5
Thermo Scientificiv 1300 Series A2
Preface
Do You Need Information or Assistance on
Thermo Scientific Products?
If you do, please contact us 8:00 a.m. to 6:00 p.m. (Eastern Time) at:
1-740-373-4763 Direct 1-800-438-4851 Toll Free, U.S. and Canada 1-877-213-8051 FAX http://www.thermofisher.com Internet Worldwide Web Home Page service.led.marietta@thermofisher.com Tech Support Email Address
Certified Service Web Pagewww.unitylabservices.com
Our staff can provide information on pricing and give you quotations. We canSales Support
take your order and provide delivery information on major equipment items or make arrangements to have your local sales representative contact you. Our products are listed on the Internet and we can be contacted through our Internet home page.
Our staff can supply technical information about proper setup, operation orService Support troubleshooting of your equipment. We can fill your needs for spare or replacement parts or provide you with on-site service. We can also provide you with a quotation on our Extended Warranty for your Thermo Scientific products.
Whatever Thermo Scientific products you need or use, we will be happy to discuss your applications. If you are experiencing technical problems, working together, we will help you locate the problem and, chances are, correct it yourself...over the telephone without a service call.
When more extensive service is necessary, we will assist you with direct factory trained technicians or a qualified service organization for on-the-spot repair. If your service need is covered by the warranty, we will arrange for the unit to be repaired at our expense and to your satisfaction.
Regardless of your needs, our professional telephone technicians are available to assist you Monday through Friday from 8:00 a.m. to 6:00 p.m. Eastern Time. Please contact us by telephone or fax. If you wish to write, our mailing address is:
Thermo Fisher Scientific (Asheville) LLC 401 Millcreek Road, Box 649 Marietta, OH 45750
International customers, please contact your local Thermo Scientific distributor.
Page 6
1300 Series A2 vThermo Scientific
Table of Contents
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Safety Systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Paper Catch Grids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
HEPA Filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Use of the Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Unit Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
Chamber Lighting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-9
UV Lights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-10
Work Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-10
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Optional Exhaust Transition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Exhaust System Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Moving the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Service Valve Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Universal Piping Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
SmartPort . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Drain Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Start-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Assemble the Stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Level the Cabinet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Unlock Counterweight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Power Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
Installation Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
Locating a Certifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Unit Start-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Operating Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Loading the Chamber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
Working Recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
Unit Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
Unit Shut-down . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
Procedure Interrupt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
Section 1
Section 2
Section 3
Section 4
Page 7
vi 1300 Series A2 Thermo Scientific
Cleaning / Decontamination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
Care and Cleaning of Stainless Steel . . . . . . . . . . . . . . . . . . . . . . . . .5-1
Cleaning and Caring for Coated Surfaces . . . . . . . . . . . . . . . . . . . . .5-2
Disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
Rinsing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
UV Disinfection After Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
Change the UV Disinfection Time . . . . . . . . . . . . . . . . . . . . . . . . .5-4
Microbiological Space Decontamination . . . . . . . . . . . . . . . . . . . . . .5-4
Clean Exterior Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5
Clean Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5
Clean the Drain Pan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5
Clean the Paper Catch Grid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-6
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
Optional UV Lights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
Sample Chamber Lights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
Unit Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-3
Replacements and Repairs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-3
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-1
Certification Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-1
Classification of the Safety Cabinet . . . . . . . . . . . . . . . . . . . . . . . . . .8-1
Test Terms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-1
Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-2
Testing Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
HEPA Filter Leak Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-9
Filters that cannot be Accessed or Scanned . . . . . . . . . . . . . . . . . . .8-9
Filters that can be Accessed and Scanned . . . . . . . . . . . . . . . . . . . .8-9
Site Installation Assessment Tests . . . . . . . . . . . . . . . . . . . . . . . . .8-10
Elect. Leakage , Ground Resistance, Polarity Tests . . . . . . . . . . . .8-10
Airflow Pattern Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-10
Data Log/Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-1
Table of Contents
Section 5
Section 6
Section 8
Section 7
Section 9
Page 8
1300 Series A2 1-1Thermo Scientific
Section 1 Description
The Thermo Scientific 1300 Series A2 offers innovative SmartFlow™ technology; an automatic airflow compensation system that adjusts motor speed as filters load, without the use of a manual damper. The SmartFlow™ system ensures safe working conditions, even between annual certifications.
The plenum assembly consists of a plenum for the downflow blower and a plenum for the exhaust air blower. Each blower includes a filter. The exhaust air is discharged through an opening in the top of the cabinet.
Chamber lighting includes one fluorescent bulb for 4 foot models and two for 3, 5 & 6 foot models.
The optional, factory-installed UV light is installed in the ceiling at the front of the sample chamber.
Figure 1-1. Unit Components
RS-232
Connection
Fuseholders (2)
Exhaust opening
Power cord
Exhaust blower
Downflow blower
Plenum
access panel
Accessory
outlets
Smart Port
Supply
test hose
Armrest
Work tray
Exhaust test hose
Service valve
Electrical enclosure
Plenum
Light dome
Control panel Handle
Front window
Service valve
access ports
Stand (optional)
Page 9
1-2 1300 Series A2 Thermo Scientific
Section 1
Description
Safety Systems
Three service valve access ports are located on each side panel. When installing service valves, remove the center of the port plugs at the perforations.
The optional stand may be ordered in two configurations; a manually adjustable from 30” to 38” working height, and a fixed 34” working height with attached casters.
A stainless steel one-piece work surface is standard. Optional armrests ensure a comfortable working position for the user and minimize blockage of airflow at the front grille.
Internal outlets, provided as the power source for accessories, are provided standard in the following configuration:
1305, 1307, 1320, 1321, 1322, 1323, 1325, 1327, 1335, 1337, 1339, 1340, 1341, 1342, 1345, 1347, 1350, 1351, 1352, 1353, 1355, 1357, 1368, 1369, 1370, 1371, 1375, 1377, 1385, 1387, 1395, 1397: two duplex right and left side 1306, 1308, 1326, 1328, 1330, 1331, 1332, 1333, 1336, 1338, 1346, 1348, 1358, 1359, 1366, 1367, 1372, 1373, 1376, 1378, 1380, 1381, 1382, 1383, 1388, 1389, 1390, 1391, 1392, 1393, 1396, 1398: two single right and left side
Test hoses (aerosol challenge ports) are located under the work tray. The supply plenum hose is located on the left and the exhaust plenum hose on the right.
Note Do not remove the caps of the two test hoses to check for supply and exhaust airflow.
s
Negative pressure air system
A negative pressure air system combines with HEPA filters in the sup­ply and exhaust airflow for personnel and product protection.
Personnel protection
Inflow air along the entire working opening at a constant high velocity prevents leakage from the work opening of the chamber.
Exterior air pressure being higher than the internal air pressure creates negative pressure, ensuring containment in case of cabinet leakage.
Product protection
Steady airflow within the air system ensures constant downflow, allow­ing the HEPA filters to remove contaminants so that the samples are always surrounded by ultra-pure air. Harmful particles are not carried over the sample chamber (protection from cross-contamination).
Page 10
1300 Series A2 1-3Thermo Scientific
Section 1
Description
HEPA filters
The downflow air within the chamber and the exhaust air are cleaned by HEPA filters.
Safety lockout
To protect from UV radiation, the optional UV light will not turn on if the front window is open. While the UV light is on, the front window should remain closed. The closed window blocks out the UV rays. Opening the window will cause the UV light to turn off.
Digital Airflow Verification (DAVe)
Independent monitoring of inflow and downflow air velocities guaran­tees that product and personnel protection remain uncompromised. Airflow monitoring determines the velocity of the airflow in the sample chamber as well as the inflow velocity of air through the exhaust open­ing. As soon as airflow velocities rise above or fall below a specified safe­ty value, an audible and a visual alarm is activated.
Window position monitoring
The position sensors detect the size of the front window opening and indicate whether the window is open to the specified work position, closed (energy saving) or in an unsafe intermediate position.
SmartFlow Indicator
The SmartFlow Indicator provides assurance that the inflow is balanced properly. SmartFlow monitors the advanced motor technology adjusting airflows even as the filter resistances change due to loading or other influences. The SmartFlow Indicator visually demonstrates the cabinet’s ability to compensate to maintain personnel protection. Green indicators show the fan and filters have power and capacity in reserve to respond to additional demands. Yellow or red indicators show that while the cabinet is providing personnel and product protection, maintenance should be scheduled to restore the compensation reserve.
Safety Systems
(continued)
Page 11
The filter system consists of two HEPA filters for downflow (supply) and exhaust air.
Room air is drawn into the sample chamber through the work position window opening. The room air is drawn into the negative pressure drain pan air duct, then pulled into the air plenum in the top portion of the cabinet. The room air is mixed with recirculating HEPA filtered air from the chamber, and filtered proportionally by the downflow and exhaust filters. The filtered air is then supplied as ultra-pure air into the sample chamber of the unit, and exhausted as ultra­pure air into the room environment, or into an external exhaust system via an exhaust transition.
Sectional paper catch grids are installed beneath the work tray in the air duct between the inner and outer back wall of the unit. These grids prevent large items such as paper towels and tissues from entering the plenum where they may impair the function of the blowers or the filters. The grids can be easily removed for cleaning.
1-4 1300 Series A2 Thermo Scientific
Section 1
Description
HEPA Filters
Figure 1-2. Airflow Filtering System
Paper Catch Grids
Ultrapure air
Exhaust air filter
Exhaust blower
Room air
Supply blower
Supply filter
Ultrapure air
Downflow in chamber
Blended air
Paper catch screens
Page 12
1300 Series A2 1-5Thermo Scientific
Section 1
Description
The control panel is located on the front of the cabinet and consists of keys, status indicators and a numerical display.
The display with its 5-digit panel shows the following information, regardless of the activated operating function:
• filter life meter (reset on filter replacement or attaining 10,000 hours)
• downflow and inflow air velocities
• remaining time of the UV disinfection routine
The keys turn functions on or off.
Status indicators show:
• window work position
• airflow
• reduced blower speed
• SmartFlow Indicator
Control Panel
Figure 1-3. Control Panel
Keys
Display
ft/min
SMARTFLOW INDICATOR
Status Indicators
Page 13
The sample chamber is accessible through various positions of the front window. The working position of the window is a 8 or 10 inch opening, +0.25”, -0.75” (depending on model - check data label and first page of this manual). For loading the chamber, the window opens fully. For decontamination and energy saving, enter into Night-Set-Back mode by closing the window. When the window is closed, the blower speed is reduced, minimizing energy consumption and noise emittance without compromising particulate containment inside the work chamber.
Note The working and loading window positions are indicated on the user interface, as controlled by micro switches and displayed by status indicators on the control panel.
s
Use of the Window
1-6 1300 Series A2 Thermo Scientific
Section 1
Description
Figure 1-4. Window Openings
Window
Max opening for loading samples
1300SERIES A2
Work position
Closed position
Window
Opening height
for cleaning window
Pillar dress
panel
1300SERIES A2
Pillar dress
panel
Page 14
To place the window at the work position, move it to the area of the top ‘dimple’ (Figure 1-5) and watch for the GREEN sash position height indicator to illuminate (Figure 1-6). If the window is moved from the the work position, the sash height indicator will turn RED.
To place the window in the closed or reduced mode position, move it to the area of the lower ‘dimple’ (Figure 1-7) and watch for the BLUE reduced mode indicator to illuminate (Figure1-8). If the window is not positioned correctly, or is moved from the closed position, the BLUE indicator will be extinguished. When the BLUE reduced mode indicator is not illuminated, the unit will not operate in reduced mode, and the optional UV light will not function.
1300 Series A2 1-7Thermo Scientific
Section 1
Description
Use of the Window
(continued)
Figure 1-5. Work Position Figure 1-6. Lighted Green LED
Figure 1-7. Fully Closed Position (UV, if applicable) Figure 1-8. Lighted Blue LED
dimple
dimple
Page 15
1-8 1300 Series A2 Thermo Scientific
Section 1
Description
Our patented-pending SmartClean™ window design easily lowers below the work position for replacing the sample chamber light bulbs.
Caution If the front window jams or is sluggish to move, do not use force! Contact Technical Services immediately. Do not attempt to repair this problem.
s
The standard unit includes accessory outlets, service valve access ports, an RS-232 connection and a remote alarm connection.
The power cord to be connected to the power source is located on the top, near the front of the cabinet.
The accessory outlets (max 5A) are located on the interior chamber back wall.
On top of the cabinet are the two fuseholders for the 5A accessory outlet fuses and an RS-232 connection (signal source, not customer interface).
Service valve access ports are located on each side of the cabinet. To use these ports, remove the center perforation. Service valves are available as optional accessories.
The remote alarm accessory can be used for notification of window movement or access.
Warning If a gas burner is to be operated in the sample chamber, an appropriate shut-off valve for the gas supply must be installed. Use only laboratory safe burners in the sample chamber.
s
Use of the Window
(continued)
Unit Interface
Page 16
1300 Series A2 1-9Thermo Scientific
Section 1
Description
The Thermo Scientific 1300 Series A2 offers a bright workspace for a more comfortable working environment. Models with coated rear and side walls resist glare and make it easier for the user to work safely.
Lighting for the sample chamber includes one fluorescent bulb for 4 foot units, and two bulbs for 3, 5 & 6 foot units. The bulb(s) is installed behind the light canopy.
Figure 1-9. Interfaces (2 of 4 access ports are shown)
Chamber Lighting
Unit Interface
(continued)
L
Fuses
N
RS-232
Power cord
Accessory outlet (1 of 2 shown)
Access ports
Service valve
Page 17
1-10 1300 Series A2 Thermo Scientific
Section 1
Description
The optional, factory­installed UV light is installed at the top of the front section of the sample chamber.
The operating time of the UV light is preset for 1 hour. The UV disinfection procedure can be started by pressing the “UV” key on the control panel with the window in the closed position. If the window is opened, the procedure is immediately cancelled.
Note To protect from UV radiation, the UV light can only be turned on when the window is completely closed.
s
The single piece work tray is provided with the cabinet. It is placed onto the frame above the sample chamber drain pan. It can be removed by lifting up, using the large holes in the work tray.
The working area for optimal product protection extends over the entire width and depth of the work tray. The two optional armrests are positioned in the center of the working area 8 inches (20cm) from each other. To attach the armrests, insert into the first perforated track of the work tray.
Warning Work safety is assured only if the armrests are used correctly!
s
Work Area
UV Lights
Figure 1-10. Lights
Figure 1-11. Work tray
UV light
Chamber light
Work tray
Perforation
track
Armrest
Separation distance
Armrest
Width
Depth
Perforation track
Page 18
1300 Series A2 2-1Thermo Scientific
Section 2 Installation
The operational safety and proper function of the unit depend on the location where it is to be operated. The cabinet should be operated only in the ambient conditions listed below.
• Up to 6562 ft. (2000m) above sea level.
• Power outlet accessible to authorized personnel only. Ideally, the outlet
should be installed above the biological safety cabinet. The power cord is the mains disconnect.
• Location capable of supporting the weight of the device and non-
flammable.
• If the factory-approved stand is not purchased, the holding device must
be rated for twice the cabinet weight.
• Adequate room height. For devices that are not connected to an
exhaust system, the distance between the exhaust air opening at the device and the room ceiling must be at least 8 inches (200mm).
• Equipped with an appropriate ventilation system.
• Room temperature between 10°C and 40°C (50°F and 104°F).
• Relative humidity must not exceed 80%, up to 31°C and decreasing
linearly to 50% at 40°C.
• For indoor use only.
Room ventilation should preferably be a ventilation system that complies with the national requirements for the application.
The inlet air and exhaust air openings of the room ventilation must be located so that drafts do not impair the function of the biological safety cabinet air system.
When positioning the cabinet, make sure the counterweight on the back of the unit can move freely. The minimal distance to the wall or adjacent objects should be 3 inches (7.6cm), unless upper wall brackets are used.
Page 19
The part numbers for the optional thimble exhaust connection are 1911316 for three, four & five foot units, and 1911317 for six foot units.
The test conditions according to NSF/ANSI 49 were performed with the standard version of the safety cabinet, without exhaust system accessories.
An available accessory for the biological safety cabinet exhaust system is an exhaust thimble connection for an on-site exhaust system with blower. According to NSF/ANSI Standard 49, an exhaust alarm is required with this connection.
The exhaust system components are installed to the exhaust opening on top of the cabinet. The threaded inserts for the retaining screws are pre­installed in the unit.
Installation on 3, 4 and 5 ft. models:
1. Turn the unit off and disconnect it from the power source.
2. Thoroughly clean the surfaces around the exhaust opening and of the accessory to be installed, to ensure that they are free of debris.
3. Secure the accessory by tightening the supplied retaining screws (M5) wrench-tight.
4. Using the adjustment nuts on the sliding panel, set the gap height to an opening of 2 inches.
2-2 1300 Series A2 Thermo Scientific
Section 2
Installation
Optional Exhaust
Transition
Exhaust System
Accessories
Figure 2-1. 4 ft Thimble
Thimble
Retaining screws
Exhaust opening
Threaded inserts
Page 20
The exhaust thimble connection is installed to the top of the cabinet using an additional adapter. The threaded inserts for the retaining screws of the adapter are pre-installed to the top of the cabinet.
Installation on a 6 ft. model:
1. Turn the unit off and disconnect it from the power source.
2. Clean the contact surfaces around the exhaust opening, and the upper and lower contact surfaces of the adapter thoroughly to ensure that all grease and dust residue are removed.
3. Secure the adapter to the pre-installed threaded inserts on top of the cabinet using the supplied four retaining screws. Tighten the screws hand-tight.
4. Secure the accessory by tightening the supplied retaining screws (M 5) wrench-tight.
5. Using the adjustment nuts on the sliding panel, set the gap height to an opening of 2 inches.
1. Place the connecting sleeve onto the connecting opening of the accessory and secure it with the screws.
2. Fit the connecting pipe (200mm [8 in]) of the on­site exhaust air system to the connecting sleeve.
3. Seal the unused connecting opening with the cover. Place the cover onto the connecting opening at the housing and secure it with the screws.
1300 Series A2 2-3Thermo Scientific
Section 2
Installation
Connecting Sleeve
Exhaust System
Accessories (continued)
Figure 2-2. 6 ft Thimble and Sleeve
Thimble
Retaining screws
Exhaust opening
Threaded inserts
Adapter
Pipe
Sleeve
Cover
Sleeve
Opening
Page 21
2-4 1300 Series A2 Thermo Scientific
Section 2
Installation
Moving the Unit
To prevent tilting, always transport the cabinet separate from the stand, using equipment capable of supporting the weight, even for a move within a building.
Caution When moving the unit, lift only at the lift points shown in Figure 2-3.
s
Caution Do not allow the weight of the cabinet to rest on the drain pan! s
Warning The weight of the window is balanced by the counterweight on
the back of the cabinet. Do not move the unit unless the counterweight has been locked in place with the shipping screws (see Figure 3-2). s
Figure 2-3. Lift points
Page 22
Install the service valve(s) by removing the centers of the plug (brass) or piercing it at the pre­determined location (plastic).
Installation spacing on side of unit is indicated in Figure 2-4.
Note The plugs must only be used for the installation of service valves in accordance with the applicable national regulations.
s
At each side panel, service valves can be installed through the three service valve port plugs (P/N 249095-exterior port plug, 249096-interior port plug).
1. Move the front window to the fully open position.
2. Turn the unit off and disconnect it from the power source.
3. Score the plug circularly at the pre-determined point (inside and outside). Remove the centers.
4. Referring to Figure 2-5 on the following page, slide the bezel onto the threaded pipe of the service valve. Apply a thin bead of silicon to the inner flange of the service valve.
5. From inside the sample chamber, slide the service valve through the desired plug.
6. From the outside, slide the washer onto the service valve threaded pipe and apply a thin bead of silicon to the outer flange of the service valve.
7. Secure the service valve to the side panel using the nut.
8. Establish the connection to the supply line using a union nut.
Note If a service valve is removed, the service valve port plug should be re-installed.
s
1300 Series A2 2-5Thermo Scientific
Section 2
Installation
Service Valve
Connections
Figure 2-4. Valve Location
SmartPort
Page 23
2-6 1300 Series A2 Thermo Scientific
Section 2
Installation
Valve Connections
(continued)
Figure 2-5. Valve Assembly and Installation
If SmartPorts are installed, replace the SmartPort grommet before starting a decontamination procedure. Two replacements (P/N 114111) are shipped with each unit, or the 1910112 kit containing 4 grommets can be ordered.
SmartPort locations are indicated above (one on each side).
Replace the SmartPort
Before Decontamination
Smart Port
Service valve
Service valve
Bezel
Threaded
pipe
(3/8”MPT)
Washer
Service valve port plugs
Union
nut
Nut
Page 24
Universal piping is a factory-installed option. The piping can be installed on either, or both, sides. Access points, depending on selected installation, are available on the top, side or bottom of the cabinet. Pipe threads are 1/4” FPT. Refer to illustration below.
Note The top left access point is not available on 3 cu ft units, due to the proximity of the exhaust filter. On 3 cu ft units, the right side access points are next to the right side rope channel. The left side access points are to the left of these.
s
1300 Series A2 2-7Thermo Scientific
Section 2
Installation
Universal Piping
Connections
Figure 2-6. Possible Universal Piping Access Points
top access points
3 cu ft units only
left side
access points
bottom of cabinet access points
right side access points
bottom of cabinet access points
Page 25
The SmartPort is an optional feature on the 1300 Series A2, designed to improve work area organization inside the biological safety cabinet. It is designed to accommodate the routing of vacuum line tubing, line cords and data cables from the interior of the cabinet to the exterior working area, freeing up space inside the work area.
A 3 inch access port with an inner and outer solid rubber grommet is located on both side walls of the cabinet. Soap bubble leak test should be performed while the grommet is closed. Using a surgical scalpel, the user may cut the inner and outer grommet along the pre-cut indentations, cutting only what it needed to accommodate the tubing or cable. The negative pressure air channel within the side walls ensures that no particles escape into the lab.
At any time, new solid rubber grommets may be installed into the SmartPort to provide a complete seal in the side walls. It is recommended that new solid grommets be installed on the exterior side walls, prior to decontaminating the cabinet. Two new solid grommets arre supplied with the cabinet for this purpose, or kit # 1910112 (four grommets) can be ordered.
After the unit has been installed, either on a stand or a workstation, the drain valve should be installed. The drain valve itself is shipped in the manual bag taped to the inside of the unit.
Locate the drain piping underneath the unit on the left side. Apply thread sealant sparingly to the threads.
Install the drain valve. Ensure the valve is in the closed position.
Be aware of the valve when moving the unit from the stand or workstation.
2-8 Model 1300 Series A2 Thermo Scientific
Section 2
Installation
SmartPort
Drain Valve
Page 26
1300 Series A2 3-1Thermo Scientific
Section 3 Start-Up
Caution Correct assembly and installation is essential for proper start-up
and operation. s
For a unit without a stand, place the cabinet onto a stable structure capable of supporting two times the unit weight. Make sure the weight is not resting on the drain valve or nipple.
For a unit with a stand, first assemble the stand per the instructions included with the stand. Then fit the cabinet onto the stand. See steps below.
Caution Make sure each leveler is turned all the way in, to avoid bending an edge during installation.
s
1. Locate the four telescoping legs in the side panel packaging. Install adjustable leveler from the hardware bag into the swaged end of the telescoping legs (if not already installed). NOTE: Make sure leveler is screwed all the way in, before installation. Using a hammer, tap the leveler into place.
2. Locate the left and right side panels. Install the telescoping leg assemblies into the side panels as shown in Figure 3-1.
3. The height-adjustable stand can be set to a height range of 26.8 - 34.7 in. (680 - 880mm ). To set the stand to the desired height, pull the legs downward out of the channels to the desired height. Lock the height adjustment in place by installing two lockwashers/screws per channel to the leg.
Note All lockwashers are a wedgelock type. The wedges fit together with the smoother sides toward the unit and the screw. See Detail.
s
4. Locate the two crossmembers from the packaging. Note they are packaged separately from the side panels. With the label on the crossmember to the outside (Figure 3-1), install the crossmembers onto the retaining tabs of the side panels. Place lockwashers on screws, then install all screws by hand, before tightening to secure together.
5. To install the cabinet frame to the stand, insert four flatwashers and Allen screws loosely into the threaded holes at the bottom of the cabinet.
Assemble the Stand
Page 27
3-2 1300 Series A2 Thermo Scientific
Section 3
Start-Up
Figure 3-1. Stand Assembly
Assemble the Stand
(continued)
6. Place the safety cabinet onto the stand so that the Allen screws are routed through the holes of the keyhole tabs.
7. Slide the cabinet frame to the keyhole stop (Fig. 3-1).
8. Tighten the four Allen screws.
9. Level the cabinet (instructions follow in this section).
Detail. Order of Assembly
Channel
Leg
Lockwasher
Lockwasher
Screw
Screw
Cabinet
Side panel
Crossmembers
Crossmember
Keyhole
Stand
Retaining tab
Side
Screws
Back
Lockwashers
Telescoping leg
Leveler
Screws
Cabinet
front
Screw with
flatwasher
Keyhole
Keyhole
tab
Foot
Keyhole
stop
Page 28
1300 Series A2 3-3Thermo Scientific
Section 3
Start-Up
Unlock
Counterweight
The window counterweight is secured to the back of the cabinet to protect the window from damage during shipping.
Caution The counterweight on the cabinet back compensates for the weight of the front window. Do not try to move the front window before the counterweight has been unlocked.
s
Warning After unlocking, keep hands and fingers away from the motion range of the counterweight!
s
To unlock the counterweight, remove the shipping screws (see location arrows in Figure 3-2) and retain them for future shipping needs. A Torx wrench is shipped in the manual bag for this purpose.
Warning One person should hold the counterweight while a second person removes the screws. This facilitates screw removal and avoids the possibility of stripping the screws.
s
Note The cabinet should be leveled only after it has been located in the desired working location.
s
For cabinets without a stand, place a bubble level onto the work tray surface and adjust the cabinet until the level indicates an absolutely horizontal position in all directions.
For cabinets with an optional stand, place a bubble level onto the work tray and adjust the four levelers of the stand using an open end or adjustable wrench until the bubble level indicates an absolutely horizontal position in all directions. Check from right to left, and from back to front.
Level the Cabinet
Figure 3-2. Unlocking the Counterweight
Counterweight
Page 29
Warning Contact with electrical components can cause a lethal shock.
Warning Before connecting the unit to a power source, check plug and
power cord for possible damage. Do not use damaged components to connect the unit to the power source!
s
1. Before connecting the unit to the power source, verify that the voltage of the outlet matches the specifications on the cabinet nameplate. If the ratings given for voltage (V) and maximum current (A) do not match, do not connect the unit to the power source.
2. Connect the power cord to a properly grounded and fused outlet. The outlet must be fused separately, using a fusible link T15A or using a circuit breaker B15. If the configuration of the receptacle does not match the power plug, consult a qualified electrician to replace or install the correct receptacle.
3. Make sure that the power cord is routed away from the counterweight and cable guide. The left side of the cabinet has additional fastening points that can be used for routing the power cord.
4. Make sure that the power cord is not stretched or pinched.
To protect from accidental disconnection, the power source outlets should be located out of casual reach and be accessible only to authorized persons. Ideally, the outlet should be located above the cabinet to avoid accidental shut-off.
The power supply cord is the mains disconnect. Maintain access at all times.
If the unit includes service valves (gas, water), the valves must be installed properly to ground, to the unit’s main ground.
On connection to the power source, place the window at the working position. Press and hold the On button. A green LED in the display indicates power is being supplied to the cabinet. The cabinet is now ready for operation and can be operated using the control panel.
3-4 1300 Series A2 Thermo Scientific
Section 3
Start-Up
Power Connection
Page 30
Warning Do not operate the unit before initial operation, installation and
certification tests have been performed.
s
The installation test must be performed in accordance with the specifications of NSF/ANSI 49. The cabinet may be operated as a Class II biological safety cabinet, in accordance with NSF/ANSI 49, if the unit functions listed below were verified and if the test results are within the safety value tolerances as specified in NSF/ANSI 49, Annex F.
• Inflow velocity profile test
• Downflow velocity profile test
• HEPA filter leakage test
• Airflow smoke pattern test
• Site installation assessment tests
• A repeat test must also be performed annually, after repairs to the unit or after location changes.
• The operator should request a written test report from the authorized service technician.
Warning The operational safety of the unit, particularly the personnel and product protection, are guaranteed only if all safety functions of the unit have been tested and approved.
s
Caution Thermo will not warrant operational safety if the unit is operated without the required installation and certification test, or if these tests and repeat test are not performed by adequately trained and authorized personnel.
s
Caution The initial operation with subsequent installation test does not include any decontamination measures. The sample chamber and any accessories required must be disinfected and cleaned in accordance with the hygiene guidelines set forth for the desired application. s
Caution Alarms on this unit are not factory-set. Set the alarms to avoid product loss and ensure personnel safety. Refer to Certification section.
s
1300 Series A2 3-5Thermo Scientific
Section 3
Start-Up
Installation Tests
Page 31
Caution Service and certification must be performed by qualified
personnel. s
Biological safety cabinet certification consists of a series of tests designed to verify that the cabinet is performing within operating parameters established by the manufacturer. To assure that a biological safety cabinet is operating as intended, each cabinet should be field-tested at the time of installation and at least annually thereafter. Cabinets should be re-certified whenever HEPA filters are changed, internal maintenance is performed, or the unit is relocated.
Three industry-related organizations maintain lists of companies and individuals who are active in the certification industry. You may contact these organizations at the addresses listed below.
NSF International (NSF) and International Air Filtration Certifiers Association (IAFCA) sponsor certifier accreditation programs. Accredited certifiers have demonstrated proficiency at testing biological safety cabinets by successfully completing written and/or practical examinations.
Biohazard Cabinet Field Certifier Program
NSF International
PO Box 130140
789 N. Dixboro Rd
Ann Arbor, MI 48113-0140
Telephone (734) 769-8010 Or (800) NSF-MARK
Fax (734) 769-0109
http://www.nsf.org/Certified/Biohazard-Certifier
IAFCA
PO Box 12155
Columbus, OH 43212
Telephone (888) 679-1904
Fax (614) 486-1108
http://www.iafca.com/certifier.html
3-6 1300 Series A2 Thermo Scientific
Section 3
Start-Up
Locating a Certifier
Page 32
The Controlled Environment Testing Association (CETA) is a trade association devoted to promoting and developing quality assurance within the controlled environment testing industry. A list of active members is available by contacting the organization.
Controlled Environment Testing Association
1500 Sunday Drive
Suite 102
Raleigh, NC 27607
Telephone (919) 787-5181
Fax (919) 787-4916
http://www.cetainternational.org/members/corp_indiv.htm
Note Unless certification was expressly called for in the specification, quotes and/or purchase order, the cost for this on-site testing is to be paid for by the customer.
s
1300 Series A2 3-7Thermo Scientific
Section 2
Start-Up
Locating a Certifier
(continued)
Page 33
Page 34
1300 Series A2 4-1Thermo Scientific
Section 4 Operation
The control panel has been divided into three functions; display, keys, status indicators.
The display shows the information below during each operation.
• during normal operation, the operating hours of the unit,
• during safe work mode, the downflow and inflow*,
• with UV light (optional) timer on, the remaining disinfection time.
*Press the Hours/Velocities key to switch between the operating hours, and the downflow and inflow in safe work mode (with window in the open, safe operating work position).
Figure 4-1. Control Panel Functions
Display
Power
Indicator
Green
Hours/ Air velocities
On/Off
SMARTFLOW INDICATOR
Green
Yellow
ow
Yell
ft/min
Red
Light On/Off
Power LED
Red
Window in work
position (green)
Outlets On/Off
LED
Window
NOT in
work
position
(red)
UV Light On/Off
UV
Airflow
steady green)
Silence
Airflow
NOT
steady
(red)
Reduced blower speed/ window closed/ UV interlock indicator
Page 35
4-2 1300 Series A2 Thermo Scientific
Section 4
Operation
To enable (I) / disable (0) alarm ringback function, follow the steps below.
In OFF mode, press and hold the Hours/Air Velocities key for approximately five seconds. The display momentarily shows the operating hours, then the switching state: I = alarm acknowledge function enabled,
0 = alarm acknowledge function disabled.
Light key turns the sample chamber light on and off.
The On/Off key turns power to the unit on or off.
Outlet On/Off key turns power to the internal outlets on and off (the blue LED indicates that voltage is present).
Note Many of our units are equipped with a Ground Fault Circuit Interrupter (GFCI) on the left side of the cabinet. This device monitors
the amount of current flowing from hot to neutral. If there is an imbalance, it trips the circuit. This single GFCI protects both outlets
within the cabinet.
If the outlets do not operate when the power button is turned on, press the Reset button in the center of the GFCI outlet to ensure the GFCI is set and power is flowing through it. If this does not restore power, contact Technical Services for additional troubleshooting.
UV Light key turns the optional UV light on and off. The yellow LED indicates the UV light has been activated. If the UV light is not installed, this key has no function.
This key also activates the UV disinfection procedure. During the procedure, the display alternates between dIS and the remaining preset time (see Section 5 to change this preset).
Silence key for muting the audible alarm.
Figure 4-2. Reset Button on GFCI Outlet
Page 36
1300 Series A2 4-3Thermo Scientific
Section 4
Operation
1. To turn the unit on, press and hold the ON key until the blowers start (audible blower operation), status indicators (LEDs) illuminate, and an audible tone sounds. Note A delay of several seconds may occur between key actuation and unit response.
s
2. Move the window to the work position (8 or 10 inch opening, model dependent - see data label and first page of this manual). The correct position is reached when the green LED “front window is in work position” illuminates. The side guide rails have markings that define the lower edge of the front window. These markings can be used as an additional orientation aid.
3. Wait until the green LED “airflow is steady” illuminates.
4. The unit is ready for operation.
The following are the operating modes for this unit.
• OFF mode • Window open mode
• Working mode • Standby mode
• UV mode
Off mode means the unit is in an idle state. The unit is turned off (blowers are off). The sample chamber light is enabled. The unit is connected to a power source. The power supply for the internal outlets is enabled when the blue LED “internal power supply on” is lighted.
Window open mode is the state in which the window is open and positioned outside of the working position (8 or 10 inches, model dependent - see data label and first page of this manual). To install or remove accessories, the front window can be moved to the maximum open position. For cleaning, the front window can be lowered beyond the closed position (see Use of the Window, Section 1).
The air system blowers are switched on with the red LED “airflow not steady” lighted. The front window is not in the work position, with the red LED “front window not in work position” is lighted. The sample chamber light is enabled. The power supply for the internal outlets is enabled when the blue LED “internal power supply on” is lighted.
Working mode is active when the front window is in the working position and the air system is operating steadily. The green LEDs “front window is in work position” and “airflow steady” are lighted.
The markings on the side guide rails and the lower edge of the front window are at the same height with no audible alarm signal.
Unit Start-Up
Operating Modes
Page 37
The sample chamber light is enabled and power for the internal outlets is enabled when the blue LED “internal power supply on” is lighted.
Standby mode is the state in which the window is closed with the blower velocity reduced. The red LED “front window not in work position”, red LED “airflow not steady” and the blue LED “airflow reduced” are lighted.
The sample chamber illumination is enabled. The internal power supply in the sample chamber is enabled when the blue LED “internal power supply on” is lighted.
The blowers on the biological safety cabinet can be turned off (OFF mode) by pressing and holding the ON key for five seconds until all LEDs are off.
UV mode (lighted yellow LED) means the UV light is on with a set timer running. The window must be in the closed position to protect against UV radiation. The light will not turn on with the window open.
The yellow LED “UV disinfection routine activated” is lighted until the preset time has elapsed. Then, the UV lights turn off automatically, and the status indicator turns off.
The sample chamber illumination is disabled in this mode, as well as the internal power source in the sample chamber.
1. Move the front window to the maximum opening position. This automatically switches the blowers to full speed.
2. Install needed work materials within the sample chamber work area. Avoid blocking the front air intake grille.
3. Move the front window to the work position (the green status indicator Front Window in Work position is illuminated) and wait until the airflow has stabilized (the green status indicator Airflow steady is illuminated).
Warning Personal and product protection is ensured only if the airflow system of the device is working properly. If the alarm system issues a failure message for more than a few minutes while the front window is in work position, stop all applications that may jeopardize worker safety.
s
4. Load the work tray with samples.
5. For work breaks or for extended experimental phases without manual intervention, switch the device to standby mode by closing the window.
4-4 1300 Series A2 Thermo Scientific
Section 4
Operation
Operating Modes
(continued)
Loading the Chamber
Page 38
Failure messages are shown on the display as error codes ER 3 to ER 7. If one of these codes appears on the display, contact Technical Services immediately.
To isolate the cause of the failure, perform the following tests.
• Check to see if the exhaust opening on top of the cabinet is blocked.
• Ensure that the in-house exhaust air system is activated, if applicable .
• Close doors and windows in the laboratory to prevent drafts.
• Turn off devices in the vicinity of the biological safety cabinet that cause air turbulence, or emit excessive heat.
• Open flames in the sample chamber may impair airflow conditions.
1300 Series A2 4-5Thermo Scientific
Section 4
Operation
Error Codes
The observance of work rules ensures a minimum of operational safety when handling the biological safety cabinet.
Before starting a procedure, take off all jewelry, put on required personal protective equipment (gloves, goggles, apron), and clean and disinfect the sample chamber surfaces at regular intervals.
During operation:
• Place samples only within the defined work area of the work tray.
• Do not place unnecessary items into the sample chamber.
• Use only disinfected and cleaned accessories for the work process.
• Do not cause air turbulence by quick hand, arm or body movements in the sample chamber or in front of the work opening.
• Do not place accessories into the sample chamber that cause air turbulence or emit excessive heat.
• Do not block air circulation at the ventilation slots of the work tray.
Working
Recommendations
Page 39
• A height-adjustable working chair with an adjustable seat back should be used during extended work periods at the cabinet.
• When the forearm rests on the armrest, is should be in a nearly horizontal position.
• When the upper leg is in a horizontal position, the angle between upper and lower leg should not exceed 90°.
• To compensate between the floor and sitting height, a footrest should be used. The minimal effective size of the footrest should be 18 x 14 in (45 x 35cm). The slope should be adjustable to within a range from 5° to 15°. The adjustable height should extend to a minimum of 4 in (11cm) above the floor.
After completing a procedure,
1. Remove samples from the sample chamber and store them properly.
2. Clean and disinfect the sample chamber surfaces, including the work tray and the drain pan.
3. Clean and disinfect all accessories.
Working (continued )
4-6 1300 Series A2 Thermo Scientific
Section 4
Operation
Figure 4-3. Sitting Posture
Forearm on armrest horizontal
Upper leg horizontal
90° between upper
and lower leg
Slope 5° to 15°
Minimum 4 inches
off floor
15°
Page 40
1300 Series A2 4-7Thermo Scientific
Section 4
Operation
Procedure Interrupt
To interrupt a work process:
1. Remove all samples from the chamber and store them properly.
2. Remove any accessories from the chamber, and clean and disinfect them.
3. Clean and disinfect the chamber surfaces, work tray, and drain pan.
4. Turn the unit to standby mode by first closing the window. Then press and hold the ON key until the indicators are off (the right indicator segment of the display shows a dot to indicate that voltage is present).
Caution For safety reasons, the blowers can be switched off only when the front window is closed.
s
If the device is not to be used or is to be stored for an extended period of time, it must be completely decontaminated.
Warning To shut the device down, the sample chamber must be disinfected completely and the plenum, including the filters, must be decontaminated using the appropriate and authorized procedures.
s
1. After the device has been decontaminated, close the window completely.
2. Disconnect the unit from the power source.
All components with the exception of the HEPA filters can be discarded after having been thoroughly cleaned and decontaminated. The HEPA filters must be discarded in accordance with the applicable national, state and local regulations for special solid waste.
Warning As this unit can be used for processing and treating infectious substances, it must be decontaminated prior to disposal, in accordance with acceptable standards and procedures.
Unit Shut-down
Unit Disposal
Page 41
Page 42
1300 Series A2 5-1Thermo Scientific
Section 5 Cleaning / Decontamination
Several procedures can be used for decontaminating the biological safety cabinet. Which procedure is selected, depends on the potential risk imminent in the agents used, and/or the degree of cleanliness required by an experiment or work process.
One possible decontamination procedure is to clean, rinse and dry with a disinfection liquid or spray, and sterile distilled water.
UV disinfection can also be used. It is particularly suited as a more intense disinfection after the above procedure.
Autoclaving (sterilization with steam) can be used for treating the removable stainless steel components. The work tray(s), armrests and paper catch grids are autoclavable components.
Disinfection with formaldehyde can be performed if a sterile sample chamber is required for the work procedure. This sterilization procedure is mandatory before filters are replaced, or before the unit is discarded.
Stainless steel is a combination of many different metals including iron and chromium. Iron, the primary element in stainless steel, tends to corrode (rust) when in its natural state. The chromium content in stainless steel prevents corrosion.
Stainless steel is not corrosion or rust proof, just resistant to stains, or stains “less”. The chromium, in the presence of oxygen, forms a tough, invisible, passive layer of chromium oxide film on the steel surface. If damaged mechanically or chemically, this film is self-healing as long as it has enough oxygen. The presence of any liquid or solid that remains in contact with the stainless steel for a prolonged time can prevent oxygen contact and promote corrosion, as can prolonged contact with cleaners or disinfectants containing chlorine, ammonia, iodine or other caustic agents.
Care and Cleaning of
Stainless Steel
Page 43
5-2 1300 Series A2 Thermo Scientific
Section 5
Cleaning and Decontamination
Cleaning/Caring for
Stainless Steel (cont.)
Cleaning and Caring
for Coated Surfaces
1. To properly care for stainless steel, use cleaners and disinfectants free of caustic agents such as chlorine, iodine and ammonia.
2. Always follow the application of any cleaner or disinfectant with a minimum of two clean distilled water rinses, then a thorough drying with a clean soft cloth. By rinsing with distilled water and drying, any remaining residue is removed from contact with the stainless steel.
Caution Never use abrasive cleaners, scouring pads or steel wool when caring for stainless steel.
s
If the stainless steel does become stained, corroded or rusted, the iron
deposits left on the surface can be removed by neutralizing them with the passivation process. This process uses an acid to neutralize the “free iron” deposits left on the steel where there has been deprivation of oxygen. This stops the spread of the corrosion or rust. It will not return the stainless steel to its original finish.
All the coated surfaces of this equipment are powder coated. Powder coating is a method of applying a dry powder to electro-statically charged metal, then baked in an oven where the dry powder molecules are melted and fused together. This is by far the most durable finish available today and, if cared for properly, will last for many years. The coated surfaces should be cleaned with a neutral detergent and rinsed twice with clean distilled water, then dried thoroughly with a clean soft cloth. Cleaning or disinfecting coated surfaces without rinsing with clean distilled water and drying thoroughly will result in smearing, streaking and dulling of the coated surfaces. Never use abrasive cleaners, scouring pads or steel wool. If the coated surfaces do become dull, streaked, smeared or marred in some other way, there is no known method to restore the finish.
Page 44
1. Remove all samples from the chamber and store them properly.
2. Remove accessories from the cabinet and disinfect them using the disinfection procedure recommended by the manufacturer of the accessory.
3. The work tray(s) and stainless steel components can be removed from the chamber and disinfected separately.
4. Clean all chamber surfaces with disinfectant.
5. Do not remove the optional UV lights from the sockets. Clean thoroughly with a damp cloth. Do not saturate. Be sure to dry completely.
1. Rinse all surfaces twice with clean distilled water.
2. Discard any liquid from the drain pan. Rinse pan.
3. Dry all chamber surfaces completely.
UV disinfection can be performed by using the optional factory installed UV light.
To start the UV disinfection procedure,
1. Close the front window, the air system operates in the reduced mode (blue LED illuminated).
2. Press the UV key on the control panel. The display alternates between dIS and the remaining disinfection time.
To interrupt or cancel the UV disinfection procedure, just press the UV key (the display shows the operating hours) and slide the window up.
Rinsing
Disinfection
1300 Series A2 5-3Thermo Scientific
Section 5
Cleaning and Decontamination
UV Disinfection After
Cleaning
Page 45
Microbiological Space
Decontamination
5-4 1300 Series A2 Thermo Scientific
Section 5
Cleaning and Decontamination
The UV disinfection time is preset. However, this setting can be changed, if required.
1. Turn the unit on and move the window to the work position.
2. Press and hold the UV key until the preset time is displayed.
3. To increase the time, press the internal power supply key. Each time the key is pressed, the time increases by 30 minutes.
4. To reduce the time, press the sample chamber light key. Each time the key is pressed, reduces the time by 30 minutes.
5. To save the value, press the UV key. The display shows the operating hours again.
Note The UV disinfection time is adjustable from 0:00 (disabled) to 24:00 hours.
s
NSF/ANSI 49, Annex G recommends a microbiological space decontamination when maintenance work, filter changes, and performance tests require access to any contaminated portion of the cabinet and before the BSC is moved to another location. A microbiological space decontamination may also be required by the BSC owner or user before certification or as a part of their work procedures. NSF/ANSI 49 provides recommended procedures for microbiological space decontamination using formaldehyde and chlorine dioxide. NSF/ANSI 49 also recognizes that alternative methods including the use of vaporized hydrogen peroxide are sometimes validated and approved by BSC owners and users. The appropriate method and procedure should be reviewed and approved by interested parties including the BSC user, institution biological safety officer, and BSC certification and service personnel as appropriate.
Warning Decontamination with formaldehyde and chlorine dioxide must be performed in accordance with the specifications of NSF/ANSI 49, Annex G. As this procedure has considerable risks, it must only be performed by specially trained and authorized service personnel! Before Decontamination with chlorine dioxide, the BSC must be sealed at the exhaust opening and front opening to eliminate exposure to chlorine dioxide, of the power supplies, PCBs and other components under the front canopy as they are susceptible to damage from the process.
s
Thermo Fisher Scientific has developed validated procedures for the decontamination of the Thermo Scientific 1300 Series A2 using the Steris VHP (Vaporized Hydrogen Peroxide) system. These procedures are available on request.
Change the UV
Disinfection Time
Page 46
1300 Series A2 5-5Thermo Scientific
Section 5
Cleaning and Decontamination
Clean the exterior surfaces of the cabinet using a solution of tepid water and commercially available mild dishwashing agent. Then, dry all surfaces well, using a soft, clean cloth.
For cleaning, the window can be lowered beyond the closing position (Use of the Window, Section 1).
Creating a gap at the window’s upper edge ensures that the upper portion of the window can be cleaned and/or disinfected. Use a commercially available window cleaner to clean the window.
Clean the drain pan, using a solution of tepid water and commercially available mild dishwashing agent.
1. Remove the work tray(s) from the chamber.
2. Clean thoroughly to remove any residues and/or deposits.
3. Wipe the drain pan, using a clean cloth and plenty of clean water.
4. Discard any liquid in the drain pan. Rinse and dry thoroughly.
Note After cleaning, make sure that all cleaning product has been removed completely from the drain pan.
s
5. Re-install the work tray.
Clean Exterior
Surfaces
Clean Window
Clean the Drain Pan
Page 47
The protective paper catch grids are retained in the airduct wall by their own tension.
1. To remove a grid section, push the retaining tab down until the section can be moved out from under the inner back wall.
2. To install the grid section, first place it on the lip of the back wall, then push the retaining tab down and towards the back until the locking tabs are secured behind the inner back wall.
Caution Do not operate the unit without the paper catch grids installed!
s
5-6 1300 Series A2 Thermo Scientific
Section 5
Cleaning and Decontamination
Figure 5-1. Protective Paper Catch Grid
Clean the Paper
Catch Grid
Retaining tab
Grid section
Grid section
Inner back wall
Retaining tab
Page 48
Section 6 Maintenance
The SmartFlow Performance Indicator displays the status of the compensation reserve.
Green indicators show the fan and filters have power and capacity in reserve to respond to additional demands. Yellow or red indicators show that while the cabinet is providing personnel and product protection, maintenance should be scheduled to restore the compensation reserve.
Regardless of the display on the SmartFlow Indicator, the biological safety cabinet should be certified annually.
The annual certification consists of the following checks:
• Electrical safety in accordance with national regulations.
• Functional test of the device.
• Checking all components for possible damage.
• Checking the filter state.
• Verifying device air flows
• Checking the airflow conditions.
• Repeat test in accordance with NSF/ANSI 49.
Warning The diffuser plate on the chamber ceiling serves as protection for the downflow filter and prevents refluxing of downflow air. When scanning the filter surface for a leak test, the perforated plate must be removed.
s
1300 Series A2 6-1Thermo Scientific
Page 49
As filter replacement is an interference with the safety system of the unit, filters must only be replaced by adequately trained and authorized service personnel.
Warning For any service work in a potentially contaminated section of the biological safety cabinet, the unit must first be decontaminated.
s
The sample chamber is illuminated by one (4 ft units) or two (3, 5 & 6 ft units) fluorescent bulbs. The fluorescent bulbs are installed in the light mounting frame in the sample chamber (Fig. 6-1).
1. Switch the unit off and disconnect it from the power source.
2. Move the window to the cleaning position (below closed position) to ensure a sufficient gap between the window’s upper edge and the light dome.
3. The fluorescent bulbs are installed in rotating sockets. To remove, rotate the bulb counterclockwise to disengage the latch and remove it from the sockets.
4. To install, slide the bulb contact pins into the rotating socket grooves and rotate the bulb clockwise to latch the sockets.
The optional, factory installed UV light is installed in the chamber ceiling immediately behind the window. The UV light should be replaced after 1500 operating hours. See Figure 6-1.
1. Turn the unit off and disconnect it from the power source.
2. Move the window to the maximum open position.
3. Wear protective gloves to prevent skin oils from burning into the bulb.
The UV bulb is installed in rotating sockets. To remove, rotate the bulb counterclockwise to disengage the latch and remove it from the sockets.
4. To install, slide the bulb contact pins into the rotating socket grooves and rotate the bulb clockwise to latch the sockets.
6-2 1300 Series A2 Thermo Scientific
Section 6
Maintenance
Service
Sample Chamber Lights
Optional UV Lights
Page 50
1300 Series A2 6-3Thermo Scientific
Section 6
Maintenance
Caution Remote alarm systems can be retrofitted and integrated into the unit controls. Alarm contacts change state based on window position. Airflow changes from window position could impair personnel and product protection. Installation of these systems should only be performed by authorized service personnel.
s
The entire cabinet with the exception of the HEPA filters can be discarded after having been thoroughly cleaned and decontaminated. The HEPA filters must be discarded in accordance with the applicable national, state and local regulations for special solid waste.
Warning As this unit can be used for processing and treating infectious substances, it may become contaminated. Prior to disposal, the entire unit with filters must be decontaminated in accordance with acceptable standards and procedures.
s
Replacements and
Repairs
UV light
Figure 6-1. Light Replacement
Unit Disposal
Chamber light
Page 51
Page 52
Section 7 Specifications
1300 Series A2 7-1Thermo Scientific
Dimensions
Description
4ft Models (10” window opening work position) 1345, 1346, 1354, 1355, 1356, 1375, 1376, 1384, 1385, 1386
4ft Models (8” window opening work position)
1305, 1306, 1325, 1326, 1335, 1336, 1395, 1396
6ft Models (10” window opening work position) 1347, 1348, 1357, 1358, 1359, 1377, 1378, 1387, 1388, 1389
6ft Models (8” window opening work position)
1307, 1308, 1327, 1328, 1337, 1338, 1397, 1398
Exterior dimensions
Width mm / in 1300 / 51.2 1900 / 74.8
Depth mm / in 800 / 31.5
Height mm / in 1568 / 61.8
Interior dimensions
Width mm / in 1200 / 47.2 1800 / 70.9
Depth mm / in 630 / 24.8
Height mm / in 780 / 30.7
Front window
Working position mm / in
254 / 10 - pertains to all 4’ & 6‘ models listed
above with 10” window work position
203 / 8 - pertains to all 4’ & 6’ models listed above with 8” window work position
Opening max. mm / in 535 / 21.1
Working area height with adjustable height stand
Height Range mm / in 750 / 29.5 to 950 / 37.4
SmartPort - Diameter mm / in 76.0 / 3.0
Remaining side wall
access ports -Diameter
mm / in 23 / 0.91
SmartPort - Distance
lower edge
mm / in 365.8 / 14.4
Remaining upper ports -
Distance edge
mm / in 226.1 / 8.9
Remaining lower ports -
Distance lower edge
mm / in 353.1 / 13.9
Distance from back panel
SmartPort mm / in 182.9 / 7.2
First access port mm / in 259.0 / 10.2
Second access port mm / in 309.9 / 12.2
Third access port mm / in 360.7 / 14.2
Page 53
7-2 Model 1300 Series A2 Thermo Scientific
Section 7
Specifications
Dimensions
Description
3ft Models (10” window opening work position) -
1320, 1321, 1322, 1323, 1330, 1331, 1332, 1333
3ft Models (8” window opening work position) -
1339, 1340, 1341, 1342, 1390, 1391, 1392, 1393
5ft Models (10” window opening work position) -
1368, 1369, 1370, 1371, 1380, 1381, 1382, 1383
5ft Models (8” window opening work position) -
1350, 1351, 1352, 1353, 1366, 1367, 1372, 1373
Exterior dimensions
Width mm / in 1000 / 39.4 1600 / 63.0
Depth mm / in 800 / 31.5
Height mm / in 1568 / 61.8
Interior dimensions
Width mm / in 900 / 35.4 1500 / 59.1
Depth mm / in 630 / 24.8
Height mm / in 780 / 30.7
Front window
Working position mm / in
254 / 10 - pertains to all 3’ & 5‘ models listed
above with 10” window work position
203 / 8 - pertains to all 3’ & 5’ models listed above
with 8” window work position
Opening max. mm / in 535 / 21.1
Working area height with adjustable height stand
Height range mm / in 750 / 29.5 to 950 / 37.4
SmartPort - Diameter mm / in 76.0 / 3.0
Remaining side wall
access ports -Diameter
mm / in 23 / 0.91
SmartPort - Distance
lower edge
mm / in 365.8 / 14.4
Remaining upper ports -
Distance edge
mm / in 226.1 / 8.9
Remaining lower ports -
Distance lower edge
mm / in 353.1 / 13.9
Declaration of Conformity available from the factory, on request.
Page 54
1300 Series A2 7-3Thermo Scientific
Section 7
Specifications
Ambient conditions (indoor use only)
Temperature
Max. ambient temp during operation °C / °F 40 / 104
Min. ambient temp during operation °C / °F 10 / 50
Humidity
Max. humidity / operation % r.H. 90, non-condensing
Max. humidity / storage % r.H. 95
Heat dissipation to environment
Room temperature 20°C kJ/s / BTU/s
4ft models 10": 0.2 / 0.19
4ft. models 8": 0.18 / 0.17
6ft. models 10": 0.4 / 0.38 6ft. models 8": 0.36 / 0.34
Room temperature rise
Above room temp with window closed °K < 2 < 2
Ergonomics
Noise level* dB(A)
4ft. models 10”: 63, ±2
4ft. models 8": 62, ±2
6ft models 10”: 65, ±2
6ft. models 8": 64, ±2
*The noise level was determined in accordance with NSF/ANSI 49. The measurement uncertainty is within a range of ± 2 dB.
Volume, weights, and loads
Description
4ft Models (10” window opening work position) 1345, 1346, 1354, 1355, 1356, 1375, 1376, 1384, 1385, 1386
4ft Models (8” window opening work position) 1305, 1306, 1325, 1326, 1335, 1336, 1395, 1396
6ft Models (10” window opening work position) 1347, 1348, 1357, 1358, 1359, 1377, 1378, 1387, 1388, 1389
6ft Models (8” window opening work position) 1307, 1308, 1327, 1328, 1337, 1338, 1397, 1398
Volume
Drain pan l / gal 20 / 5.3 30 / 7.9
Weights
Cabinet kg / lb 200 / 441 280 / 617
Stand kg / lb 26.3 / 58.0 30.0 / 66.1
Loads
Maximum load on work surface kg / lb 25 / 55
Maximum load on overall working area kg / lb 50 / 110 75 / 165
Page 55
7-4 1300 Series A2 Thermo Scientific
Section 7
Specifications
Ambient conditions (indoor use only)
Temperature
Max. ambient temp during operation °C / °F 40 / 104
Min. ambient temp during operation °C / °F 10 / 50
Humidity
Max. humidity / operation % r.H. 90, non-condensing
Max. humidity / storage % r.H. 95
Heat dissipation to environment
Room temperature 20°C kJ/s / BTU/s
3ft models 10": 0.17 / 0.16
3ft. models 8": 0.15 / 0.14
5ft. models 10": 0.31 / 0.29
5ft. models 8": 0.28 / 0.27
Room temperature rise
Above room temp with window closed °K < 2 < 2
Ergonomics
Noise level* dB(A)
3ft. models 10”: 63, ±2
3ft. models 8": 62, ±2
5ft models 10”: 65, ±2
5ft. models 8": 64, ±2
*The noise level was determined in accordance with NSF/ANSI 49. The measurement uncertainty is within a range of ± 2 dB.
Volume, weights, and loads
Description
3ft Models (10” window opening work position) - 1320, 1321, 1322, 1323, 1330, 1331, 1332, 1333
3ft Models (8” window opening work position) - 1339, 1340, 1341, 1342, 1390, 1391, 1392, 1393
5ft Models (10” window opening work position) - 1368, 1369, 1370, 1371, 1380, 1381, 1382, 1383
5ft Models (8” window opening work position) - 1350, 1351, 1352, 1353, 1366, 1367, 1372, 1373
Volume
Drain pan l / gal 15 / 4.0 25 / 6.6
Weights
Cabinet kg / lb 170 / 375 230 / 507
Stand kg / lb 24.4 / 53.8 28.1 / 62.0
Loads
Max. load on single work surface module kg / lb 25 / 55
Maximum load on overall working area kg / lb 50 / 110 75 / 165
Page 56
1300 Series A2 7-5Thermo Scientific
Section 7
Specifications
Electrical data
Models (Nominal voltages ±10%)
120V Models - 4 ft / 6 ft
1305, 1307, 1325, 1327, 1335, 1337, 1345,
1355, 1375, 1385 / 1347, 1357, 1377, 1387,
1395, 1397
230V Models - 4 ft / 6 ft
1306, 1308, 1326, 1328, 1336, 1338, 1346,
1354, 1356, 1376, 1384, 1386 / 1348, 1358,
1359, 1378, 1388, 1389, 1396, 1398
120V Models - 3 ft / 5 ft
1320, 1321, 1322, 1323, 1339, 1340, 1341,
1342, 1350, 1351, 1352, 1353, 1368, 1369,
1370, 1371
230V Models - 3 ft / 5 ft
1330, 1331, 1332, 1333, 1366, 1367, 1372, 1373, 1380, 1381, 1382, 1383, 1390, 1391,
1392, 1393
Voltage
Rated voltage V 1/N/PE AC, 120V, 50/60 Hz 1/N/PE AC, 230V, 50/60 Hz
Blower voltage V 48 V DC
Current
Power consumption maximum A 8 / 10 7 / 8
Leakage current IEC 61010-1,
UL61010-1
mA < 3.5
Motherboard fusing A 2 x T 15 A
Outlet fusing A 2 x T 5 A
On-site fusing A Circuit breaker 15 A / Fuse T 15 A
Power
Power input maximum * W 1130 / 1700 1440 / 2000
Protection
Protection class I
Protection type IP 20
Overvoltage category (IEC 61010-1,
UL61010-1)
II
Contamination degree (IEC 61010-1,
UL61010-1)
2
Connecting lines
Mains connection
Power Cord (>3 m / 10 ft)
with NEMA 5-15 plug
CEE 7/7 Plug
* With additional 5A load on cabinet receptacles, blowers at 100% and cabinet lighting switched on.
Page 57
7-6 1300 Series A2 Thermo Scientific
Section 7
Specifications
Airflow system
Description
4ft Models (10” / 8” window opening work position) 1305, 1306, 1325, 1326, 1335, 1336, 1345, 1346, 1354, 1355, 1356, 1375, 1376, 1384, 1385, 1386, 1395, 1396
6ft Models (10” / 8” window opening work position) 1307, 1308, 1327, 1328, 1337, 1338,1347, 1348, 1357, 1358, 1359, 1377, 1378, 1387, 1388, 1389, 1397, 1398
Airflows / Airflow velocities
Inflow m/s 0.53
ft/min 105
Downflow m/s 0.32
Downflow ft/min 63
Air volume
Overall volume flow
m³/h
(10”/8”)
1415 / 1298 2122 / 1947
ft³/min
(10”/8”)
833 / 764 1249 / 1146
Downflow volume flow
m³/h
(10”/8”)
829 1244
ft³/min
(10”/8”)
488 732
Exhaust air volume flow
m³/h
(10”/8”)
585 / 468 878 / 702
ft³/min
(10”/8”)
344 / 276 517 / 413
Filters
Type HEPA
Material Fiberglass
Separability in MPPS % 99.995
Separability at 0,3 μm particle size % 99.999
Downflow
Width mm / in 1220 / 48 1830 / 72
Depth mm / in 457 / 18
Height mm / in 94 / 3.7
Exhaust air
Width mm / in 610 / 24 915 / 36
Depth mm / in 457 / 18 457 / 18
Height mm / in 117 / 4.6 117 / 4.6
Page 58
1300 Series A2 7-7Thermo Scientific
Section 7
Specifications
Airflow system
Description
3ft Models (10” / 8” window opening work position) - 1320, 1321, 1322, 1323, 1330, 1331, 1332, 1333, 1339, 1340, 1341, 1342, 1390, 1391, 1392, 1393
5ft Models (10” / 8” window opening work position) - 1350, 1351, 1352, 1353, 1366, 1367, 1368, 1369, 1370, 1371, 1372, 1373, 1380, 1381, 1382, 1383
Airflows / Airflow velocities
Inflow m/s 0.53
ft/min 105
Downflow m/s 0.32
Downflow ft/min 63
Air volume
Overall volume flow
m³/h
(10”/8”)
1061 / 973 1768 / 1622
ft³/min
(10”/8”)
625 / 573 1041 / 955
Downflow volume flow
m³/h
(10”/8”)
622 1037
ft³/min
(10”/8”)
366 610
Exhaust air volume flow
m³/h
(10”/8”)
439 / 351 732 / 585
ft³/min
(10”/8”)
258 / 207 431 / 344
Filters
Type HEPA
Material Fiberglass
Separability in MPPS % 99.995
Separability at 0,3 μm particle size % 99.999
Downflow
Width mm / in 915 / 36 1525 / 60
Depth mm / in 457 / 18
Height mm / in 94 / 3.7
Exhaust air
Width mm / in 610 / 24
Depth mm / in 457 / 18
Height mm / in 117 / 4.6
Page 59
Page 60
1300 Series A2 8-1Thermo Scientific
Section 8 Certification Testing
For application in the USA, the unit has been rated as a Class II safety cabinet, Type A2, in accordance with NSF/ANSI 49.
For operation as a unit of said classifications, a certification test and repeat test in accordance with NSF/ANSI 49, Annex F should be completed at the time of installation, and at least annually thereafter. As described in Annex F, recertification should be performed whenever HEPA filters are changed, maintenance repairs are made to internal parts, or the cabinet is relocated. More frequent recertification should be considered when working with safety-critical materials (three to six months).
Nominal value:
Default value as specified by Thermo.
Measured value:
Value measured at the location of the safety cabinet.
Tolerance:
Acceptable deviation from the nominal value.
Average value:
The sum of the measuring values divided by the number of tests. The average value is compared to the nominal value.
Setpoint:
Acceptable operating value for the inflow and downflow velocities.
Inflow velocity (FPM):
Velocity of the air entering the sample chamber opening.
Downflow velocity (FPM):
Velocity of the displacement flow circulating through the work chamber.
Exhaust velocity (FPM):
Velocity of the airflow discharged through the exhaust filter opening.
Exhaust airflow volume (CFM):
Amount of air discharged at the exhaust filter.
Classification of the
Safety Cabinet
Test Terms
Page 61
According to NSF/ANSI 49, tests related to the safety of the personnel, product and environment that must be conducted on-site for Class II, Type A2 biological safety cabinets, at a minimum frequency of annually, are;
• downflow velocity profile test
• inflow velocity test
• airflow smoke patterns test
• HEPA filter leak test
• site installation assessment tests
– alarm functions
– exhaust system performance (proper exhaust duct negative pressure
and canopy performance).
As described in Annex F, in addition to the above, the following tests performed at the request of the customer, or at the discretion of the certification provider.
• comfort and safety tests
– lighting intensity
– noise level
– vibration
– electrical leakage, ground circuit resistance and polarity tests
Note Service with costs: Unless certification was expressly called for in the specification, quotes and/or purchase order, the cost for this on-site testing is to be paid for by the customer.
s
8-2 1300 Series A2 Thermo Scientific
Section 8
Certification Testing
Testing
Page 62
For conducting these tests, Thermo recommends testing equipment of the manufacturers listed below.
Inflow Velocity – DIM Method (Primary)
Description: Inflow velocity using the direct inflow measurement method (DIM).
Equipment: . . .Shortridge ADM-870, or equivalent
. . . . . . . . .Flow hood Series 8400, or equivalent
Method:
1. Move the front window to the work position (25.4cm/10 in. or
20.3cm/8 in., depending on model).
2. Attach the flow hood [1] in the work opening [2] and seal the remaining open areas.
3. Operate the system blowers for approx 20 min.
4. Record at least 5 measurements of the inflow air volume.
5. Average those readings and calculate the inflow velocity (V1) as described below.
1300 Series A2 8-3Thermo Scientific
Section 8
Certification Testing
Test Equipment
Testing equipment Manufacturer Application
Smoke tubes
MSA Pittsburgh, Pennsylvania, 15230
Airflow pattern test
Thermoanemometer
TSI Shoreview, Minnesota 55126
Downflow velocity test and alternate method for inflow
velocity using constricted access
Digital Safety Inspector
Ohmic Instruments Co. Easton, Maryland 21601
Electrical leakage and ground resistance tests
GFI Circuit Tester
Leviton Manufacturing Little Neck, New York, 11362
Electrical leakage, ground resistance, and polarity tests
Photometer
Air Techniques Owing Mills, Maryland, 21117
Filter leak test
Aerosol Generator
Air Techniques Owing Mills, Maryland, 21117
Generation of aerosol mist
Vibration Meter
Quest Technologies Oconomowoc, Wisconsin, 53066
Vibration test
Flow Hood Air Data Multimeter
Shortridge Instruments. Inc. Scottsdale, Arizona, 85260
Inflow velocity test
Testing Information
Page 63
Inflow Area:
• 3ft Models (10” window opening work position) 1320, 1321, 1322, 1323, 1330, 1331, 1332, 1333: Front opening 2.46 sq ft
3ft Models (8” window opening work position) 1339, 1340, 1341, 1342, 1390, 1391, 1392, 1393: Inflow area for 8” models: 1.97 sq ft
• 5ft Models (10” window opening work position) 1368, 1369, 1370, 1371, 1380, 1381, 1382, 1383: Front opening 4.10 sq ft
5ft Models (8” window opening work position) 1350, 1351, 1352, 1353, 1366, 1367, 1372, 1373: Inflow area for 8” models: 3.28 sq ft
• 4ft Models (10” window opening work position) 1345, 1346, 1354, 1355, 1356, 1375, 1376, 1384, 1385, 1386: Front opening 3.28 sq ft
4ft Models (8” window opening work position) 1305, 1306, 1325, 1326, 1335, 1336, 1395, 1396: Inflow area for 8” models: 2.62 sq ft
• 6ft Models (10” window opening work position) 1347, 1348, 1357, 1358, 1359, 1377, 1378, 1387, 1388, 1389: Front opening 4.92 sq ft
6ft Models (8” window opening work position) 1307, 1308, 1327, 1328, 1337, 1338, 1397, 1398: Inflow are for 8” models: 3.93 sq ft
Calculation:
• Inflow Velocity (V1) = Inflow Air Volume (V2)/Actual Inflow Area (A2)
Acceptance:
• 100 – 110 FPM
Correction measures:
• Enter the service level and activate S2, change S2 until the desired veloci­ty is reached and save the new value.
• If the velocity can't be adjusted due to clogging of the filters, change the exhaust filter.
• If the filter is in order, change the exhaust blower.
Note Check of alarm limits: If S1 or S2 are changed, the alarm limits S3 and S4 must be checked and saved again.
s
8-4 1300 Series A2 Thermo Scientific
Section 8
Certification Testing
Testing Information
(continued)
Page 64
Inflow Velocity – Constricted Window Method (Secondary)
Description:
• Inflow velocity measured at several points through a constricted window.
Equipment:
• Thermal Anemometer, or equivalent (hotwire)
• Anemometer Probe Holder, part number 1911325
Method:
1. Lower the front window to a height of 3 in. (7.6cm).
2. Operate the system blowers for approximately 20 minutes.
3. Insert the thermal anemometer probe into the probe holder. Place the probe holder at the window opening work surface, ensuring it is seated properly. See Figure 8-1. Adjust the probe so the air passing through probe is centered in the window opening (1.5 in. [3.8cm]).
Note The thermoanemometer probe must be positioned at a 10° angle from front vertical and located on the same plane as the inside of the window glass.
4. Beginning 5.9 inches in from the left inside wall, record successive readings every 5.9 inches (5 readings for 3 ft, 7 readings for 4 ft, 9 readings for 5 ft, and 11 readings for 6 ft models).
5. Average those readings and correct result for true velocity.
6. Calculate the inflow velocity as described below.
1300 Series A2 8-5Thermo Scientific
Section 8
Certification Testing
Testing Information
(continued)
Figure 8-1. Constricted Window Method
5.9
5.9
3
1.5
3
Page 65
Calculation:
• Inflow velocity (V1) = average of actual inflow velocity * K factor * 0.3 for 10” window opening work position, 0.375 for 8” window opening work position
K factor = 1.0219 for 3 ft w/ 10” opening, 0.9935 for 3 ft w/ 8” opening.
K factor = 0.9918 for 4 ft w/ 10” opening, 1.047 for 4 ft w/ 8” opening.
K factor = 0.9942 for 5 ft w/ 10” opening, 1.0012 for 5 ft w/ 8” opening.
K factor = 0.9918 for 6 ft w/ 10” opening, 1.100 for 6 ft w/ 8” opening.
Acceptance:
• 100 – 110 FPM
Refer to the data plate on the lower right corner of the unit face.
s
Inflow Velocity – Calculating Exhaust Flow (Alternative)
Description:
• Exhaust velocity measurements to determine inflow velocity.
Equipment:
• Thermal Anemometer, or equivalent (hotwire)
• Freestanding Fixture, or equivalent
Method:
1. Operate the unit in the work position for approximately 20 minutes.
2. Insert the anemometer probe into the probe stand, adjusting the height to 4 inches (10cm) above the exhaust filter.
3. Record the readings on a horizontal plane 4 inches (10cm) above the effective exhaust filter area, in a grid pattern not to exceed 4 inches (10cm) x 4 inches (10cm).
4. Average the readings and calculate the inflow velocity.
Grid Pattern:
3, 4 & 5 ft Models 1305, 1306, 1320, 1321, 1322, 1323, 1325, 1326, 1330, 1331, 1332, 1333, 1335, 1336, 1339, 1340, 1341, 1342, 1345, 1346, 1350, 1351, 1352, 1353, 1354, 1355, 1356, 1366, 1367, 1369, 1370, 1371, 1372, 1373, 1375, 1376, 1380, 1381, 1382, 1383, 1384, 1385, 1386, 1390, 1391, 1392, 1393, 1395, 1396
20 points
8-6 1300 Series A2 Thermo Scientific
Section 8
Certification Testing
Testing Information
(continued)
3.94
2.98
2.98
2.98
3.94
3.94 3.74 3.74 3.74 3.74 3.94 inches
Page 66
• 6ft Models 1307, 1308, 1327, 1328, 1337, 1338,1347, 1348, 1357, 1358, 1359, 1377, 1378, 1387, 1388, 1389, 1397, 1398
32 points
Effective Exhaust Filter Area:
• 3, 4 & 5 ft Models 1305, 1306, 1320, 1321, 1322, 1323, 1325, 1326, 1330, 1331, 1332, 1333, 1335, 1336, 1339, 1340, 1341, 1342, 1345, 1346, 1350, 1351, 1352, 1353, 1354, 1355, 1356, 1366, 1367, 1369, 1370, 1371, 1372, 1373, 1375, 1376, 1380, 1381, 1382, 1383, 1384, 1385, 1386, 1390, 1391, 1392, 1393, 1395, 1396: 16.8 x 22.8”, 2.66 sq ft
• 6ft Models 1307, 1308, 1327, 1328, 1337, 1338,1347, 1348, 1357, 1358, 1359, 1377, 1378, 1387, 1388, 1389, 1397, 1398: 16.8” x 34.8”, 4.06 sq ft
Inflow Area:
• 3ft Models (10” window opening work position) 1320, 1321, 1322, 1323, 1330, 1331, 1332, 1333: Front opening 2.46 sq ft
3ft Models (8” window opening work position) 1339, 1340, 1341, 1342, 1390, 1391, 1392, 1393: Inflow area for 8” models: 1.97 sq ft
• 5ft Models (10” window opening work position) 1368, 1369, 1370, 1371, 1380, 1381, 1382, 1383: Front opening 4.10 sq ft
5ft Models (8” window opening work position) 1350, 1351, 1352, 1353, 1366, 1367, 1372, 1373: Inflow area for 8” models: 3.28 sq ft
• 4ft Models (10” window opening work position) 1345, 1346, 1354, 1355, 1356, 1375, 1376, 1384, 1385, 1386: Front opening 3.28 sq ft
4ft Models (8” window opening work position) 1305, 1306, 1325, 1326, 1335, 1336, 1395, 1396: Inflow area for 8” models: 2.62 sq ft
• 6ft Models (10” window opening work position) 1347, 1348, 1357, 1358, 1359, 1377, 1378, 1387, 1388, 1389: Front opening 4.92 sq ft
6ft Models (8” window opening work position) 1307, 1308, 1327, 1328, 1337, 1338, 1397, 1398: Inflow are for 8” models: 3.93 sq ft
Calculation:
• Average exhaust velocity (V2) = Sum of measurements / Number of readings
• Exhaust Volume (V3) = Average velocity (V2) x Effective filter area (A3)
• Inflow velocity (V1) = Exhaust volume (V3) / Work opening area (A2)
Acceptance:
• 100 – 110 FPM
1300 Series A2 8-7Thermo Scientific
Section 8
Certification Testing
3.94
2.98
2.98
2.98
3.94
3.94 3.85 3.85 3.85 3.85 3.85 3.85 3.85 3.94 inches
Testing Information
(continued)
Page 67
Downflow Velocity – Uniform Cabinet
Description:
• Measures the velocity of air flow onto the work surface through the sup­ply filter.
Equipment:
• Thermal Anemometer, or equivalent (hotwire)
• Freestanding Fixture, or equivalent
Method:
1. Operate the unit in the work position for approximately 20 minutes.
2. Insert the anemometer probe into the probe stand, adjusting for a height of 4 inches (10 cm) above the bottom edge of the window opening while in the work position.
3. Record the readings on a horizontal plane 4 inches (10 cm) above the bottom edge of the window opening while in the work position, in a grid pattern not to exceed 6 inches (15 cm) x 6 inches (15 cm).
Grid Pattern:
• 3 ft Models - 1320, 1321, 1322, 1323, 1330, 1331, 1332, 1333, 1339, 1340, 1341, 1342, 1390, 1391, 1392, 1393
• 4 ft Models 1307, 1308, 1327, 1328, 1337, 1338,1347, 1348, 1357, 1358, 1359, 1377, 1378, 1387, 1388, 1389, 1397, 1398
8-8 1300 Series A2 Thermo Scientific
Section 8
Certification Testing
Testing Information
(continued)
6
5.5
5.5
6
6 3.9 3.9 3.9 3.9 3.9 3.9 6 inches
5.7 for 8 in opening models
21 points
6
5.5
5.5
6
6 5.88 5.88 5.88 5.88 5.88 5.88 6 inches
5.7 for 8 in opening models
21 points
Page 68
Grid Pattern (continued):
• 5 ft Models - 1350, 1351, 1352, 1353, 1366, 1367, 1368, 1369, 1370, 1371, 1372, 1373, 1380, 1381, 1382, 1383
• 6 ft Models 1307, 1308, 1327, 1328, 1337, 1338,1347, 1348, 1357, 1358, 1359, 1377, 1378, 1387, 1388, 1389, 1397, 1398
Calculation:
• Average Downflow Velocity = Sum of all measurements / Number of readings
Acceptance:
• 58-68 FPM
Correction measures:
• Enter the service level and activate S1, change S1 until the desired veloc­ity is reached and save the new value.
• If the velocity can’t be adjusted due to clogging of the filters, change the supply filter.
• If the filter is in order, change the supply blower.
Note Check of alarm limits: If S1 or S2 are changed, the alarm limits S3 and S4 have to be checked and saved again.
s
1300 Series A2 8-9Thermo Scientific
Section 8
Certification Testing
Testing Information
(continued)
6
5.5
5.5
6
6 5.88 5.88 5.88 5.88 5.88 5.88 5.88 5.88 5.88 5.88 6 inches
5.7 for 8 in opening models
33 points
6
5.5
5.5
6
6 5.87 5.87 5.87 5.87 5.87 5.87 5.87 5.87 6 inches
5.7 for 8 in opening models
27 points
Page 69
8-10 Model 1300 Series A2 Thermo Scientific
Section 8
Certification Testing
Description:
• Determines the integrity of the downflow and exhaust filters.
Equipment:
• Aerosol Photometer, or equivalent
• Aerosol Generator, or equivalent
Method:
1. Allow the cabinet to operate for approximately 20 minutes.
2. Remove the work tray and protective covers, as appropriate.
3. Introduce the aerosol from the generator in the center rear of the work area, using a tee fitting to evenly distribute the aerosol.
4. If desired, and if the cabinet has not been used with hazardous materials since an effective decontamination, the upstream concentration can be measured. If desired, connect the photometer sample hose to the appropriate (supply or exhaust) test hose connection (exhaust test hose - right side / supply test hose - left side) from under the work tray.
5. Using either the appropriate upstream challenge measurement or calculated challenge value, set the photometer accordingly.
6. Scan the downstream side and perimeter of the filter as prescribed in NSF/ANSI 49, Annex F.
Acceptance:
• Filters scanned – 0.01% of upstream concentration at any point
HEPA Filter Leak Test
Filters that can be
Accessed and Scanned
Page 70
1300 Series A2 8-11Thermo Scientific
Section 8
Certification Testing
Method:
1. Set up as in Steps 1 through 5 above
2. In the duct, downstream of the filter, sample in accordance with NSF/ANSI 49.
Acceptance:
• Filters not scanned – 0.005% of upstream concentration
Description:
• Determines the internal and external behavior of the air movement.
Equipment:
• Cold smoke (titanium tetrachloride)
Method:
• Downflow:
Pass smoke across the centerline of the work surface, 4 inches (10 cm) above the work opening.
• View Screen Retention:
Pass smoke 1 inch (2.5 cm) behind the window opening, 6 inches (15 cm) above the work opening.
• Work Opening Edge Retention:
Pass smoke around the entire outside work opening, approximately 1.5 inches (3.8 cm) away from the cabinet.
• Window Seal Test:
Pass smoke up both sides and across the top of the window opening from inside the work area, approximately 2 inches (5 cm) from the edges.
Acceptance:
• No smoke shall have dead spots, reflux or escape the cabinet once drawn in.
Airflow Pattern Test
Filters that cannot be
Accessed or Scanned
Page 71
Description:
• Refer to UL 61010-1
Description:
• Verify the unit is integrated properly into the facility.
Description:
• Verify the airflow alarms will activate within 15 seconds when there is a 20% loss.
Equipment:
• Thermal Anemometer or equivalent (hotwire)
• Anemometer probe holder, part 1911325
• Shortridge ADM-870 or equivalent
• Flowhood series 8400 or equivalent
Method:
• Downflow:
Enter the service level and activate S03.
Verify that the average downflow velocity is 20% lower than the rated velocity by following the method described previously. Change and save S03 if needed and leave the service level. Enter the service level again and set S01 to the value of S03 -0.5%. The LED "AIRFLOW STA­BLE" must change from green to red. Leave the service level without saving this value by pressing the "acknowledge alarm" key.
•Inflow:
Enter the service level and activate S04.
Verify that the inflow velocity is 20% lower than the rated velocity by following one method described previously. Change and save S04 if needed and leave the service level.
Enter the service level again and set S02 to the value of S04 -0.5%.
The LED "AIRFLOW STABLE" must change from green to red. Leave the service level without saving this value by pressing the "acknowledge alarm" key.
Airflow Alarms
8-12 1300 Series A2 Thermo Scientific
Section 8
Certification Testing
Elect. Leakage , Ground
Resistance, Polarity Tests
Site Installation
Assessment Tests
Page 72
1300 Series A2 8-13Thermo Scientific
Section 8
Certification Testing
Airflow Alarms (continued)
Acceptance:
• Unit alarm activates within 15 seconds when the downflow and inflow velocity is set to the correct values and the LED „Airflow STABLE“ changes from green to red (as described above).
Description:
• Verify the window alarms are operating properly
Equipment:
• Operating manual
Method:
• Raise the window 1 inch (2.5 cm) above work position
Acceptance:
• Window alarm activates
Description:
• Verify the exhaust system is functioning properly
Canopy Connections
Equipment:
• Cold smoke (titanium tetrachloride)
Method:
• Pass smoke around the air gap, to ensure negative pressure exists
Acceptance:
• No smoke refluxes back into the room once drawn into the canopy.
Sash Alarms
Exhaust System Performance
Page 73
Page 74
1300 Series A2 9-1Thermo Scientific
Device type: Part number:
Serial number: Service number:
Location: Operator's note:
Technician / Company Notes Date Signature
Page 75
THERMO FISHER SCIENTIFIC LAMINAR FLOW EQUIPMENT WARRANTY USA
The Warranty Period starts two weeks from the date your equipment is shipped from our facility. This allows shipping time so the warranty will go into effect at approximately the same time your equipment is delivered. The warranty protection extends to any subsequent owner.
During the first thirty-six (36) months, component parts proven to be non-conforming in material or workmanship will be repaired or replaced at Thermo’s expense, including labor. Installation, calibration and certification is not covered by this warranty agreement. The Technical Services Department must be contacted for warranty determination and direction prior to performance of any repairs. Expendable items, glass, filters and gasket are covered by this warranty to the extent that Thermo Fisher Scientific determines that they were defective at the time of delivery to the carrier.
Replacement or repair of component parts or equipment under this warranty shall not extend the warranty to either the equipment or to the component part beyond the original warranty period. The Technical Services Department must give prior approval for return of any component or equipment. At Thermo’s option, all non-conforming parts must be returned to Thermo postage paid and replacement parts are shipped FOB destination.
THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, WHETHER WRITTEN, ORAL, OR IMPLIED. NO WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE SHALL APPLY. Thermo shall not be liable for any indirect or consequential damages including, without limitation, damages to lost profits or loss of products.
Your local Thermo Sales Office is ready to help with comprehensive site preparation information before your equipment arrives. Printed instruction manuals carefully detail equipment installation, operation and preventive maintenance.
If equipment service is required, please call your Technical Services Department at 1-800-438-4851 (USA and Canada) or 1-740-373-4763. We’re ready to answer your questions on equipment warranty, operation, maintenance, service, and special applications. Outside the USA, contract your local distributor for warranty information.
Rev. 0 4/12
ISO
9001
REGISTERED
THERMO FISHER SCIENTIFIC LAMINAR FLOW EQUIPMENT WARRANTY INTERNATIONAL
The Warranty Period starts two months from the date your equipment is shipped from our facility. This allows shipping time so the warranty will go into effect at approximately the same time your equipment is delivered. The warranty protec­tion extends to any subsequent owner.
During the first thirty six (36) months, component parts proven to be non-conforming in material or workmanship will be repaired or replaced at Thermo’s expense, excepting labor. Installation, calibration and certification is not covered by this warranty agreement. The Technical Services Department must be contacted for warranty determination and direction prior to performance of any repairs. Expendable items, glass, filters and gaskets are excluded from this warranty.
Replacement or repair of component parts or equipment under this warranty shall not extend the warranty to either the equipment or to the component part beyond the original warranty period. The Technical Services Department must give prior approval for return of any component or equipment. At Thermo’s option, all non-conforming parts must be returned to Thermo postage paid and replacement parts are shipped FOB destination.
THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, WHETHER WRITTEN, ORAL, OR IMPLIED. NO WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE SHALL APPLY. Thermo shall not be liable for any indirect or consequential damages including, without limitation, damages to lost profits or loss of products.
Your local Thermo Sales Office is ready to help with comprehensive site preparation information before your equipment arrives. Printed instruction manuals carefully detail equipment installation, operation and preventive maintenance.
If equipment service is required, please call your Technical Services Department at 1-800-438-4851 (USA or Canada) or 1-740-373-4763. We’re ready to answer your questions on equipment warranty, operation, maintenance, service, and special applications. Outside the USA, contract your local distributor for warranty information.
Rev. 5 4/09
ISO
9001
REGISTERED
Page 76
Thermo Fisher Scientific 401 Millcreek Road Marietta, Ohio 45750 United States
www.thermofisher.com
Loading...