Before using this product, read this entire Operator's Manual carefully. Users should
follow all of the Operational Guidelines contained in this Manual and take all necessary
safety precautions while using this product. Failure to follow these guidelines could result
in potentially irreparable bodily harm and/or property damage.
Thank you for purchasing Jeio Tech’s products.
Jeio Tech Co., Ltd. is committed to customer service both during and after the sale. If you have questions concerning
the operation of your unit or the information in this manual, contact our Sales Department. If your unit fails to operate
properly, or if you have questions concerning spare parts or Service Contracts, contact our Service Department.
ICH Chamber
Identification label
Figure A.
Please locate the identification label on the right side of the instrument. Fill in the information found on the
identification label in the spaces provided above in Figure A. Refer to this identification label information
when calling, if your unit fails to operate properly, or if you have questions concerning spare parts or
service contracts. Additionally, use this information at www.jeiotech.com when you register your
TEMPERATURE & HUMIDITY CHAMBER (TH-ICH MODEL) when you fill out the enclosed registration
card.
Manual No. Manual No. 00HAA0001227 VER.(0.0) iii
Page 4
Quality Management System
Jeio Tech Co, Ltd. is dedicated to providing world-best product quality
and customer satisfaction. To ensure we maintain this commitment we
have developed and implemented a total quality program, which
conforms to the requirements according to DIN EN ISO 9001:2000 for the
design, development, production, sales and servicing of biotechnology,
environmental chemical engineering related products, and reliable
measuring equipment for electric and electronics (ovens, incubators,
constant temperature humidity chambers, constant temperature baths,
refrigerating bath circulators, heat exchangers and shakers).
Visit our Web site at www.jeiotech.com/eng/images/ISO-Jeiotech-1.jpg to view a copy of our certificate.
Disclaimer
Jeio Tech Co., Ltd. is committed to a continuing program of product development and improvement, and
reserves the right to change information, such as specifications, appearance, and dimensions, described in
this document without notice.
Copyright
No part of this manual may be reproduced or transmitted in any form or by any means, including
photocopying, recording, or using information storage and retrieval systems, for any purpose other than the
purchaser's own use, without the express written permission of Jeio Tech Co., Ltd.
Any other product names and services identified in this manual are trademarks or registered trademarks of
their respective owners. No such use, or the use of any trade name, is intended to convey endorsement or
other affiliation with Jeio Tech Co., Ltd.
This manual is intended for individuals requiring information about the use Temperature and Humidity ICH
chamber. Use this manual as a guide and reference for installing, operating, and maintaining your Jeio
Tech Temperature and Humidity ICH chamber. The purpose is to assist you in applying efficient, proven
techniques that enhance equipment productivity
This manual covers only light corrective maintenance. No installation, service procedure or other
maintenance should be undertaken without first contacting a service technician, nor should be carried out
by someone other than a service technician with specific experience with laboratory equipment and
electricity.
1.1.2 Chapter summary
The Functional Description chapter outlines models covered, standard features, and safety features.
Additional sections within the manual provide instructions for installation, pre-operational procedures,
operation, preventive maintenance, and corrective maintenance.
The Installation chapter includes required data for receiving, unpacking, inspecting, and setup of the unit.
We can also provide the assistance of a factory-trained technician to help train your operator(s) for a
nominal charge. This section includes instructions, checks, and adjustments that should be followed before
commencing with operation of the Temperature and Humidity ICH chamber. These instructions are
intended to supplement standard laboratory procedures performed at daily and weekly intervals.
The Operation chapter includes a description of controller features along with temperature and agitation
parameter setting instructions, multi-segment program setting instructions and instructions for changing the
type of agitation and agitation amplitude.
The Accessories and Option chapter is your source for information on available accessories and option
with brief information.
The Appendix contains technical specifications, warranty and Jeio Tech technical support contact
information.
1.1.3 Model number nomenclature
This manual covers all five models of the Temperature and Humidity ICH series. The following describes
the model number nomenclature used in throughout the manual.
- 2 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
Page 11
1.2 Safety Notice
Be sure that you are completely familiar with the safe operation of this Temperature and Humidity ICH
chamber. This unit may be connected to other machinery, such as a temperature control unit. Improper use
can cause serious or fatal injury.
Installation and repair procedures require specialized skills with laboratory equipment and electricity. Any
person that installs or repairs this unit must have these specialized skills to ensure that this unit is safe to
operate. Contact Jeio Tech or their local authorized distributor for repairs or any questions you may have
about the safe installation and operation of this unit.
The precaution statements are general guidelines for the safe use and operation of this instrument. It is not
practical to list all unsafe conditions. Therefore, if you use a procedure that is not recommended in this
manual you must determine if it is safe for the operator and all personnel in the proximity to the
Temperature and Humidity ICH chamber. If there is any question of the safety of a procedure please
contact Jeio Tech before starting or stopping the Temperature and Humidity ICH chamber.
This equipment contains high voltages. Electrical shock can cause serious or fatal injury. Only qualified
personnel should attempt the startup procedure or troubleshoot this unit.
Documentation must be available to anyone that operates this equipment at all times.
Keep non-qualified personnel at a safe distance from this unit.
Only qualified personnel familiar with the safe installation, operation and maintenance of
this unit should attempt start-up or operating procedures.
Always stop the Temperature and Humidity ICH chamber before making or removing any
The following signal word panels, safety symbols and non safety symbols are used to alert you to
potential personal injury hazards or information of importance. Obey all safety messages that follow
these symbols to avoid possible personal injury or death.
1.3.1 Signal word panels
Signal word panels are a method for calling attention to a safety messages or property damage
messages and designate a degree or level of hazard seriousness. It consists of three elements: a
safety alert symbol, a signal word and a contrasting rectangular background. The following signal word
panels are in accordance with ANSI Z535.4-2007 and ISO 3864 standards.
Indicates a hazardous situation which, if not avoided, will
result in death or serious injury.
Indicates a hazardous situation which, if not avoided, could
result in death or serious injury.
Indicates a hazardous situation which, if not avoided, may
result in minor or moderate injury.
Indicates a property damage message.
- 4 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
Page 13
1.3.2 Safety symbols
Safety symbols are graphic representations—of a hazard, a hazardous situation, a precaution to avoid
a hazard, a result of not avoiding a hazard, or any combination of these messages—intended to
convey a message without the use of words. The following safety symbols are used in this manual.
1.3.3 Miscellaneous none safety symbols used in manual
The following graphic representations are intended to convey a message without words or to bring
your attention to important information about the use of the Temperature and Humidity chamber or a
feature.
European Union
electrical directive
compliance
Earth ground
Relative humidity
Altitude
Note
- 6 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
Page 15
1.4 Where to Locate Safety Labels on the Temperature and Humidity
ICH chamber.
The safety labels are attached to the Temperature and Humidity ICH chamber important information
about potential hazards and how to avoid them. All users must read this operating instruction carefully
to operate the product properly.
The following illustrations show where the safety labels should be attached to the Temperature and
Humidity ICH chamber until service of the product is discontinued. If the safety labels are damaged,
please contact your local Jeio Tech office or distributor to request new labels.
- 8 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
Page 17
1.5 Precautions for Your Temperature and Humidity ICH chamber
Our Temperature and Humidity ICH chamber is designed to provide safe and reliable operation when
installed and operated within design specifications. Make sure you read and understand all instructions
and safety precautions listed in this manual before installing or operating your unit. If you have any
questions concerning the operation of your unit or the information in this manual, contact our Sales
Department.
To avoid possible personal injury or equipment damage when installing, operating, or maintaining this
Temperature and Humidity chamber, use good judgment and follow these safe practices:
1.5.1 Warning statements
Observe all warning labels.
DO NOT remove warning labels
Check the voltage, phase and capacity of the power supply and connect properly.
DO NOT ground the Temperature and Humidity chamber to gas pipes or water pipes
DO NOT insert multiple plugs into the outlet at the same time
DO NOT operate equipment with damaged line cords
DO NOT handle or touch electrical cord and electrical parts with wet hands
DO NOT move the Temperature and Humidity chamber while it is plugged into the power source
DO NOT use or keep flammable gases near the Temperature and Humidity chamber.
DO NOT install the Temperature and Humidity chamber near environments where flammable gas may leak.
DO NOT use the machine near environments where explosion can occur due to organic evaporating gases
DO NOT put explosive and flammable chemicals (Alcohol, Benzene, and etc) into the chamber.
DO NOT let moisture, organic solvents, dust, and corrosive gas enter the control panel.
DO NOT expose the Temperature and Humidity chamber to direct sunlight.
DO NOT expose the Temperature and Humidity chamber to direct heat sources.
DO NOT use the Temperature and Humidity chamber in places where moisture is high and flooding can
occur
DO NOT install the Temperature and Humidity chamber near machinery generating high frequency noise.
DO NOT use Temperature and Humidity chamber in environments that contain industrial oil smoke and
metallic dust
DO NOT operate damaged or leaking unit.
DO NOT operate the Temperature and Humidity chamber when there is strange sound, smell and smoke
coming from the unit.
DO NOT disassemble, fix or change the Temperature and Humidity chamber other than for those items
DO NOT use doors, handles or knobs to lift or stabilize the unit
DO NOT place heavy objects on the power cord.
DO NOT put the Temperature and Humidity chamber on the power cord
DO NOT make the machine wet while cleaning
DO NOT pour water or put liquid on the Temperature and Humidity chamber when cleaning the unit.
DO NOT operate Temperature and Humidity chamber and immediately disconnect the main power supply
and request service when water may be in the unit
DO NOT sprinkle insecticide or flammable spray on the Temperature and Humidity chamber
DO NOT clean the Temperature and Humidity chamber with a strong cleanser (e.g., solvent type) and use a
soft cloth.
In addition to the safety warnings listed above, safety messages are posted throughout the manual.
These safety messages are designated by the use of a signal word panel followed by text and a safety
symbol where applicable. Read and follow these important instructions. Failure to observe these
instructions can result in permanent damage to the unit, significant property damage, personal injury or
death.
- 10 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
Page 19
1.6 Responsibility
Our Temperature and Humidity chamber are constructed for maximum operator safety when used under
standard operating conditions and when recommended instructions are followed in the maintenance and
operation of the machine.
All personnel engaged in the use of the Temperature and Humidity chamber should become familiar with
its operation as described in this manual.
Proper operation of the unit promotes safety for the operator and all workers in its vicinity.
Each individual must take responsibility for observing the prescribed safety rules as outlined. All caution,
warning and danger labels must be observed and obeyed. All actual or potential danger areas must be
reported to your immediate supervisor.
1.6.1 General responsibility
No matter who you are safety is important. Owners, operators and maintenance personnel must realize
that every day, safety is a vital part of their jobs.
If your main concern is loss of productivity, remember that production is always affected in a negative way
following an accident. The following are some of the ways that accidents can affect your production:
Loss of a skilled operator (temporarily or permanently)
Breakdown of shop morale
Costly damage to equipment and laboratory samples
Downtime
An effective safety program is responsible and economically sound.
Organize a safety committee or group, and hold regular meetings. Promote this group from the
management level. Through this group, the safety program can be continually reviewed, maintained, and
improved. Keep minutes or a record of the meetings.
Hold daily equipment inspections in addition to regular maintenance checks. You will keep your equipment
safe for production and exhibit your commitment to safety.
Please read and use this manual as a guide to equipment safety. This manual contains safety warnings
throughout, specific to each function and point of operation.
The operator’s responsibility does not end with efficient experimentation and production. The operator
usually has the most daily contact with the equipment and intimately knows its capabilities and limitations.
Plant and personnel safety is sometimes forgotten in the desire to meet incentive rates, or through a casual
attitude toward laboratory equipment formed over a period of months or years. Your employer probably has
established a set of safety rules in your workplace. Those rules, this manual, or any other safety
information will not keep you from being injured while operating your equipment.
Learn and always use safe operation. Cooperate with co-workers to promote safe practices. Immediately
report any potentially dangerous situation to your supervisor or appropriate person.
REMEMBER
NEVER place your hands or any part of your body in any dangerous location.
NEVER operate, service, or adjust the equipment without appropriate training and first
reading and understanding this manual.
Before you start the portable drying/conveying system check the following:
Remove all tools from the Temperature and Humidity chamber.
Be sure no objects, samples or chemicals are lying on the Temperature and
Humidity chamber.
If your Temperature and Humidity chamber has been inoperative or unattended, check all
settings before starting the unit.
At the beginning of your shift and after breaks, verify that the Temperature and Humidity
chamber is functioning properly.
Report the following occurrences IMMEDIATELY:
unsafe operation or condition
unusual Temperature and Humidity chamber action
leakage
improper maintenance
DO NOT wears loose clothing or jewelry, which can be caught while working on the
equipment. In addition, cover or tie back long hair.
Clean the equipment and surrounding area DAILY, and inspect the machine for loose,
missing or broken parts.
Shut off power to the Temperature and Humidity chamber when it is not in use. Turn the
power switch to the OFF position, or unplug it from the power source.
- 12 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
Page 21
1.6.3 Maintenance responsibility
Proper maintenance is essential to safety. If you are a maintenance worker, you must make safety a priority
to effectively repair and maintain equipment.
Before removing, adjusting, or replacing parts on this Temperature and Humidity chamber, remember to
turn off all electric supplies and all accessory equipment at the machine, and disconnect and lockout
electrical power. Attach warning tags where possible.
Be sure that the Temperature and Humidity chamber is correctly connected to earth grounded electrical
outlet that complies with current codes.
When you have completed the repair or maintenance procedure, check your work and remove your tools.
DO NOT restores power to the Temperature and Humidity chamber until all persons are clear of the area.
BEFORE you turn the Temperature and Humidity chamber over to the operator for production, verify the
unit is functioning properly.
1.6.4 Reporting a safety defect
If you believe that your Temperature and Humidity chamber has a defect that could cause injury, you
should immediately discontinue its use and inform Jeio Tech or local authorized distributor.
The principle factors that can result in injury are failure to follow proper operating procedures
(i.e. lockout/tag out), or failure to maintain a clean and safe working environment.
This unit is designed to test for Stability Testing of New Drug Substances and Products Q1A(R2) &
Photostability Testing of New Drug Substances and Products Q1B by the ICH Harmonised Tripartite
Guideline.
The purpose of stability testing is to provide evidence on how the quality of a drug substance or drug
product varies with time under the influence of a variety of environmental factors such as temperature,
humidity, and light, and to establish a re-test period for the drug substance or a shelf life for the drug
product and recommended storage conditions.
The following test conditions are provided in accord with drug substance and drug product by the defined in
this guideline of Q1A.
- 16 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
Page 25
The Photostability Testing should be carried out systematically including the following tests as below;
Tests on the drug substance
Tests on the exposed drug product outside of the immediate pack; and if necessary;
Tests on the drug product in the immediate pack; and if necessary;
Tests on the drug product in the marketing pack.
And also the following test conditions are provided in accord with this guideline of Q1B.
For option 2 the same sample should be exposed to both the cool white fluorescent and near ultraviolet lamp.
A cool white fluorescent lamp designed to produce an output similar to that specified in ISO 10977(1993).
A near UV fluorescent lamp having a spectral distribution from 320 nm to 400 nm with a maximum energy emission
between 350 nm and 370 nm; a significant proportion of UV should be in both bands of 320 to 360 nm and 360 to
400 nm.
%
350370320400
Wave length(nm)
The procedure is the same as the following.
samples should be exposed to light providing an overall illumination of not less than 1.2 million lux hours and an
integrated near ultraviolet energy of not less than 200 watt hours/square meter to allow direct comparisons to be
The spots in circle are indicated by the defined
test temperature/humidity points in accord with
the ICH guideline Q1A(R2).
2.2.2 Elegant design
To differentiate our brand that separates our unit from the established TH chamber, more excellent design is selected
to the unit. Using natural convection heat from lamps be disposed of to the top of the structure was designed.The
lamp is installed inside of the door and you can easily replace them.
- 18 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
Page 27
2.2.3. Uniformed distribution of temp. & humid.(Up-down flow type)
Adopting 2 Sirocco fan let it maintain
the optimal air flow inside of chamber
with an appropriate air velocity to get
the suitable temp & humid in the
chamber. Opening the door will activate
the automatic stop function of heater
and fan to prevent from contact with an
external air. Air flow is running from
inhaling inner air thru the bottom of the
chamber to the vertical flow passing
thru an evaporator and heater. So
temperature & humidity inside of
chamber will controlled by the way of air
flow.
2.2.4. Designed for enhanced user’s interface
High quality 304 stainless steel exterior and interior
Casters for easy transport and installation
Convenient maintenance - available to check total operating time
Exclusive door lock & trim
Vice lock type door catch and foamed silicone packing ensure perfect chamber seal
Door open warning alarm
Cable port for running in wires
Basic hole of 50mm diameter cable port is standard equipped on left side of chambers for wiring to the specimen.
We also provide additional 50 or 80mm diameter cable port (Optional)
Water tank for easy water supply
- Maintenance can be done easily from the front side
Door lock device (OPTION)
Periodical defrosting at below 15 degree make it for long time running with less influence of temp. & humid.
2.2.5. Safety functions
Leakage breaker for power supply
Over current protector
Overheat protector
Switch-off after alarm for over heating
Door opening alarm
Water empty
The alarm when the photostability test finished.
- 20 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
Page 29
(1) Controller
- All functions are shown on the LCD Display screen with graphic &
characters.
- Easy operation to push the control buttons.
- All the status of operation is monitored by graphic mode.
- Total Max. 100 programmable segments
Max. 30 patterns supported by each segment.
(2) Lamp control board
-NO/OFF of UV & Fluorescent lamp individually can be controlled.
Pressing the menu button investigate the current strength of the
cumulative value of research strength values
Survey, a screen switch to the desired value is set.
Current time is displayed through the use of 2 lamps, and also
Lamp Replacement time can be found intuitively.
(3) Recorder (Optional)
- 6 points, can display, record and store all temperature & humidity testing
Values and profiles.
- Each channel is sorted out & shown on the display by temp. & Humid.
Recording range: -60 to 150℃/ 0 to 100% RH
Paper wide: 100mm
- Portable type
- 6 dots (Temperature: 5 dots, Humidity: 1 dot)
- Digital display
- Free power supply (100V to 240V)
(4) Handle
- By using plastic mold product, the user can safely close the door,
designed to open the door to open and close even when the power light.
You can bring them to operate a small, sealed with the chamber and more
Designed to strengthen.
(5) Ventilation Hole
- UV Fluorescent and natural convection through the heat generated
at the door to the outside. Perforated to dissipate as the air is
discharged In this and in the perforated and can be used safely and
more prolonged.
The flow of air in the right you can see in the image.
(6) Cool Air inhalation & Hot Air ventilation Hole
-This perforated air ventilation hole to be drawn
A cool air is provided for cooling down the heat
Generated from the motor & stabilizer, etc
On the upper part, and 2 rectangular fans located
On the rear part of the unit are provided
For better exchange heat capacity and better
Performance of the unit.
(7) Access port
- Basic access port is Ø50mm, and each Ø50mm or Ø80mm Access port
can be added upon user’s request.
-Access port is used for the passing the power cable Thru chamber
for the sample test of electronic components or other Devices
(8), (9) Condensing air ventilation Hole
-Air ventilation holes (perforated plate) are located on the left side & bottom of the rear part of the unit, so hot
air will be drawn off from the refrigerating system thru this ventilation holes. And air pre-filter and air-blowing
perforate plate is installed in front of condensing unit for air ventilation.
(10) Water Tank
-In order to check the water volume directly and prompt response to problems,
Water tank (15L) is located on the front side.
For easy supplying water, Sliding rails are installed.
Auto water filling system is recommendable for the
Comfortable and more than the capacity of a water tank
(15L).
Water purifier is also recommendable when required
The cleanliness control of test sample, or products
(11) Hinge
- Its material is strong enough to stand the heavy weight
load.
-KNUCKLE JOINT made of stainless steel, excellent in strength &
durability.
- Stainless cover boasts of appearance in splendor.
(12) Electric Circuit Breaker
-In order to protect the user, sample & unit, the suitable electric circuit breaker is adopted to meet the unit’s
electrical requirements and automatically cut off the electrical power in case a short circuit, or an electric
leakage.
(13) Communication Port
- RS-232 Communication port is a default setting.
- RS-485 Communication port is an optional which can control
Up to 9 units at the same time. (optional)
(14) Ventilation Hole
-(5) this perforated plate hole is the same function as ventilation Hole.
(15) Door Limit Switch
-If door is open by chance, or accidentally, it gives the signal alarm with stopping the operation of inner blower
for user’s safety by using magnetic door limit switch. It also reduces the radical change of temperature or
humidity inside of chamber.
(16) Tempered glass
-Dual tempered glass to increase insulation and sealing effect and maintenance
of temperature and humidity on the excellent effect.
It have also been due to negligence of use of glass protect users even more
damage can be safe.
- 22 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
Page 31
(17) Shelf
- 2 shelves are basically provided for the supporting the sample, which made of stainless steel.
- Its appearance has a fine view & clean body due to electro polishing finish.
(18) Fluorescent lamp
Philips is installed by default, and the fluorescent products readily available on the market at a reasonable price, so
easy to obtain and can be replaced by the user.
- TH-ICH-300: 18W, COOLWHITE, 590mm
- TH-ICH-800: 36W, COOLWHITE, 1200mm
(19) UV Lamp
Philips is installed by default, and the fluorescent products readily available on the market at a reasonable price, so
easy to obtain and can be replaced by the user.
- TH-ICH-300: 20W, black light, 590mm
- TH-ICH-800: 40W, black light, 1200mm
(20) Caster
The unit can be placed on a level on the ground
Even if it’s not even ground by adjustment of caster foot.
(21) Water tray
- By any chance of a water leakage, or door chamber
Condensate will build up and drop into this water tray.
Just keep it clean all the time by using a dry cloth.
(22) Water Drain port
-There are 2 drain condensate ports which should be connected to a silicon rubber hose over the drain barb until is
seated against the collar by tying it with a cable tie.
2.3.1. Unit other components
Evaporator
Heater
Door lock
Water Tank
Safety device for test
sample
The established evaporators use the aluminum pins, so it has worry for
corrosion and thermal efficiency decreasing. However, our product is removed
the corrosion problem and progressed the refrigeration efficiency by using the
evaporator that uses the copper pins.
The sheath type heating element for temperature is used in the established
equipment in which takes lots of load to control the operation because it cause
a residual heat when heating and cooling, and lots of a time for the auto
tuning. However, the Nichrome(Ni-Cr) heating element newly adopted has
been improved in the response speed resulting in better performance in
controlling the temperature and reducing the time for auto tuning. And also
the heating element for humidity is made of SUS material not to cause some
problem in quality for the long time running.
Due to the characteristics of the test for long time running, the door lock
should be required to protect the test sample as an optional device.
the water tank(15L) located on the bottom of the front side is installed with
sliding tray for easy supplying water, and also when line water
supply(optional) required, the floating valve in the tank can stop the water
supply when overflowing water as the 2nd safety device.
Test sample can be protected By built-in controller preventing hightemperature, and it’s alarm signal will notify the status of operation to the
operator when the problem occurs.
Above consists of 6 florescent lamps and 3 UVA lamps.
TH-ICH-300L TH-ICH-800L
- Fluorescent lamp: 18W, cool white, ISO
10977(1993)
- UVA lamp: 20W, black light
Key features
Optical Detectors (UV sensor / VIS sensor) mounted on the inside, respectively, in the location of each light and
then detected directly and precisely control delivery, Validation City provides a clear data
Convenient lamp ON / OFF
Display fluorescent lamp and UV lamp usage time
You open the door in front of the door safety switch off automatically Lamp promote user safety
Approximately 4,000 hours of lamp life is long-term replacement costs are cheaper
- Fluorescent lamp: 36W, cool white, ISO
10977(1993)
- UVA lamp: 40W, black light
- 24 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
- 26 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
Page 35
3.1 Uncrating and inspecting the unit
3.1.1 Inspecting before removing outer container
After you have received your Jeio Tech product, inspect the shipping container carefully for any damage that
may have occurred during shipping. Report any damage to the carrier and to your local Jeio Tech office or
the distributor from which the unit was purchased. If the container and packing materials are in re-usable
condition, save them for reshipment if necessary.
3.1.2 Removing the unit from shipping container
The unit is shipped on a wooden skid with an outer and inner container. The wooden outer container is
bolted to the shipping skid and has a protective vinyl cover stapled to the container’s top. Under the outer
container the incubated shaker is covered with a cardboard box and strapped to the skid.
We have segmented the process into 3 phases. Use the following instructions to remove the incubated
Step1: Remove the staples holding the vinyl cover to the wooden crate with a straight blade screwdriver.
Step2: Lift the vinyl cover off the wooden crate.
Step3: Inspect the wooden container top for signs of moisture infiltration around the screws.
Step4: Remove the fourteen (14) #7 x 50mm Bugle head screws with a 3Pt Phillips screwdriver. See Figure 3.3..
Step5: Remove the wooden container top.
- 28 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
Page 37
Figure 3.4
Step6: Remove the seven (7) #14 x 100mm hex head lag screws holding the container front to the shipping skid and
container sides. See Figure 3.4.
Figure 3.5
Step 7: Remove the crate front from the crate supports at points A, B and C using a hammer and/or crowbar. See
Figure 3.5.
use hammer to bend over or remove, and keep hands and
Figure 3.6
Step 8: Remove the crate supports from the crate back at points D, E, and F using a hammer and/or crowbar. See
Figures 3.5 and 3.6.
.
PANEL IS HEAVY!
minor or moderate personal injury. To avoid injury have
assistance in holding and removing the crate panel.
NAILS/STAPLES ARE SHARP!
crate panel and supports contact with nails/staples could
cause minor or moder
other body parts away from nails/staples.
Dropping crate panel can cause
Be careful when removing
- 30 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
Page 39
3.1.2.2 Phase 2: Removing the unit from skid
ICH chamber is shipped with the caster feet in the locked position.
Additionally, blocking is used to lock the casters and the unit in place. Use the following procedure to remove
the unit from the skid using a forklift.
Do not use manpower in any cases because it’s heavy for men to lift.
Using forklift
Tool required
None
ICH Chamber
Figure 3.7
Step 1: Carefully slide forklift forks under the incubated shaker. Forks should be set so that outside to outside fork
dimension is 800mm. See Figure 3.7.
Step 2: Slowly lift incubated shaker above the caster blocking and carefully pull the unit out of the crate.
Step 3: Set the incubated shaker on the floor so it rests on the caster feet.
After unpacking, please check the contents to ensure you have received all of the following unit
components. Also, check the identification plate on the side of the unit to make sure you received the
model number your ordered.
If you didn’t receive one or more of the components or if the model is incorrect, contact your local Jeio Tech
office, or the distributor from which the unit was purchased.
COMPONENT QUANTITY RECEIVED
Unit
Shelf
Shelf bracket
User instruction
Controller instruction
Drain hose
1
2
4
1
1
2
Access port stopper
Plug & consent
communication software
CD(OPTIONAL)
Cable for communication
(OPTIONAL)
Cable tie
- 32 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
1
1
1
(optional)
1
(optional)
4
Page 41
3.3 Preparing the Location
3.3.1 Space requirements
ICH chamber is manufactured as the size of Figure 3.10. It is essential the unit to be situated in an area
where there is minimum space 1.2m for the front-ward due to 890mm in width of the door.
Figure 3.10
Model W(mm) D(mm) H(mm) A(mm) B(mm)
TH-ICH-300
TH-ICH-800
- 34 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
Page 43
3.3.2 Environmental setting
The unit can be operated properly under the following environmental conditions for a long time running
without any problem.
No direct sunlight on incubated shaker
Ambient temperature: 18OC to 30OC
Relative humidity not to exceed 85%
Altitude not to exceed 2000m (6,562 feet)
Keep incubated shaker away from high frequency noise produced
by equipment and/or machinery, such as electrical distribution pa
nels, welders, induction heating mechanism, and large amounts of
SCR (Silicon-controlled rectifiers). See Section 6.4—Troubleshooti
ng, Electrical—for the effects of high frequency noise.
Connect the incubated shaker to earth grounded terminals only.
The caster assembly allows for easy movement, locking and leveling of the unit. For proper agitation, it
requires the unit to be level side to side and front to back. Use the following instructions to ensure the unit
is level.
Figure 3.12
Tool required
Step 1: Place the unit in its location.
Step 2: Lower the foot of each caster
by turning the red height adjustment
counterclockwise until it hits the floor.
Turn the height adjustment a quarter turn
or more to ensure the unit’s weight is taken off
the caster. This will lock the unit in place.
Step 3: Place a level on the top of the door
and check for level side-to-side and front-to-back.
Step 4: Turn the height adjustment to raise
or lower the unit until one is level.
- 36 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
Page 45
3.5 Attaching Condensate Drain
2 Condensate drains are installed on the rear side of the unit. These condensate drains are designed,
one for eliminating any water build up from the inside of chamber heats and cools, the other for escaping
water overflowed from the inside of chamber. To connect the condensate drain to a floor drain or any other
waste the outlet push Ø6mm ID soft-walled tubing, such as: vinyl, FKM, urethane, neoprene or silicone,
over the drain barb until is seated against the collar, then, tie it with a cable tie as shown below
3.6 Pre start-up checks
Make sure all unit and wall outlet electrical connections are tight.
Make sure condense drain hose is tightly on hose barb and hose is run to a drain.
Make sure caster feet are lowered and have locked the unit in place.
Make sure the unit is level side to side and front to back.
Make sure there are no flammable or explosive liquids inside of chamber
Current mode: Current amount of light displays.
Integration mode: Integrated amount of light displays.
Setting mode: Wish integrated amount of light can be set.
4.3.2.4 How to set amount of light
1) Press MENU key 2 times to move to SETTING MODE.
2) Select either fluorescent or UV lamp by using up/down keys. Selected lamp changes dark color.
3) Set value by using up/down keys and press ENTER to save value.
4.3.2.5 In case of lamp is not on.
<Both lamps are off> <FL lamp is off> <UV lamp is off>
4.3.2.6 mark indicates lock status(KEY LOCK).
- 50 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
This is the initial display of TEMI300_Multi Monitoring program operated. It can make TEMI300
monitor maximum 9units, easily up/download the program and control the saved data.
Click left button to transfer the separate operating display.
(figure.1) initial display
- 52 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
Page 61
5.2. Separate operating display
Consist of a condition information display and controlling buttons.
Controlling button
Display current PV, SP values
Click the left button to transfer the initial display.
Graph display
(figure. 2) Separate operating display
Menu -> graph: Set the displaying for graphs for individual units and save it or not
Menu -> test mode: Set the on/off for the separate units operating.
Select the graphs saving or not and test periods on the below graph setting display.
On clicking left button, check the clicked location
data on the Date View.
Menu -> file -> open: open the saved graphic date.
Menu -> file -> print: printing current display
Menu -> view -> date view: display the Date View.
Menu -> view -> list view: display the List View.
Menu -> set -> set graph: setting the graph display condition
Menu -> zoom in/out- -> zoom in: zoom in saved graph
Menu -> zoom in/out - -> zoom out: zoom out saved graph
Menu -> convert -> excel file:converting saved graphic date to excel file.
You can set the display condition of graphs on the below graph setting display.
- 60 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
Page 69
(Figure 12) Graph setting display
You can confirm current PV, SP, Date & Time on the below Date View display
You can confirm total data of opened files on the below List View display.
(Figure 14) List View display
You can convert opened file data to Excel files on the below Excel files creation display.
The converted Excel files will be saved at the same location as the opened file.
(Figure 15) Excel file creating display
- 62 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
- 66 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
Page 75
Operating temperature 18 ~ 30 , 85%RH(non℃℃-corrosive and pure place)
capacity 15L
Water tank
quality
Interface BASICALLY RS-232C, [RS-485:OPTIONAL]
pH6.2 ~ 7.2, Electrical conductivity 20㎲/cm to below
Cable Hole
Dimension Internal 750X650X650 750X880X1160
(W×D×H,mm) External 1130X985X1520 1220X1215X1955
AC220V,1ph,
Max. Power
consumption
Weight(net) 280 ± 10kg 400 ± 10kg
1) All specifications are under ambient temperature 20 (68 ), no load ℃℉
2) Technical data (according to DIN 12880, IEC 60068)
3) It was measured in the center of front chamber with lamp vertically.
4) ICH guideline (Q1B option2)
5) Up to 98% of the set value
Control of Temp. Humidity are comply with ICH guideline Q1A (R2)※
The Temperature could go up in the condition with defrost, The humidity could be changed as well.※
Permissible ambient Condition: Temperature 18 to 30 (64.4 to 86 ), Maximum re※℃℉lative humidity 85%
→ Recommended ambient temperature: +20 (68 )℃℉
60Hz
AC230V,1ph,
50Hz
Optionally max. two holes available/ Ø 50 or 80mm (2.0 or 3.2")
If trouble occurs during product use, user can get free service for 2 year.
Limited warranty from the date of purchase.
During the warranty period, user can get free service like repairing, replacements, refund.
7.2.2 Exceptional cases even during under warranty period.
User can not be credited by warranty in case of as below.
① If trouble occurs by an act of God.
② If the equipment breaks down due to misusing of available voltage.
③ If damage occurs by dropping a product, or impact.
④ If damage occurs in an appearance by organic solvents such as thinner, benzene.
⑤ If damage occurs without following to notice in the manual.
⑥ If damage occurs by fixing the equipment by any person who is not related with Jeio tech.
⑦ If damage occurs by a mistake of a customer.
- 68 - ⓒ2009. All Rights Reserved. Jeio Tech Co., Ltd
Page 77
7.3 Technical assistance
7.3.1. Oversea
U.S.A.
Jeio Tech, Inc.
1-A Gill Street
Woburn, MA 01801 USA
TEL: +1 781-376-0700
FAX: +1 781-376-0704
E-mail: nasupport@jeiotech.com