Kyoto Kagaku MW2A, MW2B Instruction Manual

Page 1
Genital Unit for MW2A (male)
Instruction manual
Attention
Do not let ink from pens, newspapers, product manual or other sources contact the manikin. Ink marks on the manikin will be irremovable.
Contents
Before you start
Set includes
Preparation
Catheterization/Enema
Training session
Training session
After training
FAQs
/DOs and DON’Ts
P.1
P.6-P.11
P.12
Page 2
Page 3
Before you start
Set Includes
Before your first use, ensure that you have all components listed below.
Set includes DOs and DON’Ts
Please read carefully before use
a.
Female genitalia organ unit Valve
b.
(total 3 pcs; one is incorporated in the genitalia unit and two are for future replacement)
c.
Irrigation bottle
Lubricant
d.
Rectum unit
e.
(attached with a large metal clip 1)
Irrigation/drainage tubes
f.
(total 3 pcs; 2 are attached with one clamp each)
Instruction manual
1
3
1
1
1
3
Genitalia organ unit
genital organ frame bladder holderurinary bladder reservoirvalve
male meatus male skin
(attached with a small metal clip 1)
DOs and DON’Ts
DOs DON’Ts
Handle the manikin and the components with care.
・ ・
Storage in a dark, cool space will help prevent the skin colours from fading.
The manikin skin may be cleaned with a wet cloth,
if necessary, using mildly soapy water or diluted detergent.
Do not let ink from pens, newspapers, product
manual or other sources contact the manikin. Ink marks on the manikin will be irremovable.
Never use organic solvent like paint thinner to
clean the skin, as this will damage the simulator.
Please use the genuine lubricant included in the set.
Usage of other lubricants will cause damage to the model in a short time as in the following cases:
- Jelly-type lubricants which remain in the model will harden and cause damage to the valves.
-
Water-soluble jelly-type lubricants will be washed
out during urethral catheterization, which will
decrease lubricating ability.
Page 4
Preparation
Preparation (Catheterization/Enema)
Connection of Irrigation/drainage tubes
1. Attach the irrigation/drainage tube with clamp to the connectors at the upper end of the urinary bladder reservoir.
clamp
clamp
Genitalia unit
Connection of Irrigation/drainage tubes
connection to the irrigation bottle
The clamp is near the far end of the tube.
Irrigation tube
drainage tube
to a bucket
There are two tube connectors on the backside of the unit.
The clamp is near the bladder reservoir.
tube connectors
urinary bladder reservoir
tube connectors
2.
One of the tube connectors is for irrigation and the
clamp should be placed at the far end of the tube.
The clamp is near the far end of the tube.
3. The other tube connector is for the drainage and the clamp on the tube should be placed around 20-30cm from the bladder.
The clamp is near the bladder reservoir.
4. Connect the drain tube with no clamp to the rectum unit.
Caution
When attaching the tubes to the unit, please hold the unit with one hand while inserting the tube to the port.
Page 5
Preparation
Connect the Irrigation bottle and ll the bladder with water
1.
Connect the irrigation bottle to the irrigation tube
which has clamp at near the far end and put the
tip of drainage tubes for catheterization and enema
into the bucket.
Preparation(Catheterization/Enema)
2.
Close the clamp of irrigation tube before lling
the bottle with water. And open the clamp of the drainage tube.
Irrigation tube
Drainage tub
e
close the clamp
3. Fill the irrigation bottle with simulated urine (water). And open the clamp of the drainage tube.
Irrigation tube
open the clamp
open the clamp
Page 6
Training session
Training session
Training procedure
1. Sterilizing of meatus urethra.
For training of sterilizing of meatus urethra, use water instead of actual antiseptic.
2. Inserting the catheter
Put the catheter in the tray, drip small quantity of lubricant to its tip. Ensure that about 5cm from the tip is coated by lubricant. lubricant included in the set.
Caution
nsert the catheter through the meatus urethra
3. I holding the penis at about 90 degrees. During the insertion, the resistance thatsimulates
prostate and external sphincter muscle will be felt.
Use 14-16Fr catheter for this model. Using larger catheter causes breakage of the valve.
Please use the genuine lubricant included in the set. Usage of other lubricants will cause damage to the model in a short time as in the following cases:
- Jelly-type lubricants which remain in the model will harden and cause damage to the valves.
- Water-soluble jelly-type lubricants will be washed out during urethral catheterization, which will decrease lubricating ability.
Product code of the genuine lubricant is 11381-800.
Use the
5㎝
4. When you feel resistance, lower the penis at about 60 degrees, and insert the catheter while pulling the penis toward you.
90°
Surgical tape can be used for training of catheter
Caution
5. Removing the catheter
Discharge the water in the balloon before removing the catheter. Inated balloon may cause breakage of the valve.
Caution
Do not leave the tapes on the modelfor long time so that its surface willbe sticky with adhesive of tapes.
The bladder reservoir needs to be relled after EACH successful catheterization. In case water in the reservoir is low, water will not be ow out even when tip of the catheter reached in the bladder reservoir.
60°
retainedinthebladder
.
Page 7
Training session
Manual bladder compression
Close the clamps on the irrigation tube and drainage tube for manual bladder compression session.
The clamp of the irrigation tube The clamp of the drainage tube
Training session
Close the clamp
Training for enema
Do the training with left lateral position.
Caution
Use plenty amount of genuine lubricant for the training. Insucient lubrication may cause the diculty in the proper training. Enema tube cannot be inserted more than 7cm like real situation.
Page 8
After training After training
Drainage of water
1. Close the clamp of the irrigation tube. Drain the water from the irrigation bottle and tube.
2. Detach the rectum unit for enema training. Loosen the metal clip (large) by tucking small wings together with ngers and remove the rectum unit from the hook of the bladder reservoir. Detach from the connector. Wash the rectum unit and let it dry naturally after the training sessions.
Caution
Always keep the clamp open for storage.
Drainage of water in the bladder reservoir
1. Disassemble the genitalia unit. a) The genital organ frame and the bladder holder are xed with two notches and nail shaped
fasteners.To disengage the parts, lift the genital organ frame with ngers.
Page 9
After training
After training
Drainage of water in the bladder reservoir
b) Remove the skin with the bladder holder from the outer frame. Then peel the skin from the bladder holder.
c) Find the metal clip (small) that holds the joint between the valve and the urethral tube. Loosen the clip by tucking small wings together with ngers and pull out the urethral tube carefully.
2. Remove the bladder reservoir from the bladder holder. Detach the irrigation and drainage tubes. Then drain the water from the reservoir.
Page 10
After training
After training
Wipe of the valve and mail meatus
1. After the training, wipe the valve and the urethral tube thoroughly. Wipe the joint between the valve and the urethral tube to remove the lubricant, using mildly soapy water and let it dry naturally. Remove the lubricant which remains in the urethral tube with cotton
swabs.
Caution
Be sure to remove all lubricant thoroughly. Remaining lubricant may make slippery and make the urethral part easy to come o.
When the lubricant remains between the skin and male meatus, remove the urethral tube by pulling it with peeling the foreskin of the penis.
Then wipe its surface and inside of the foreskin.
the connection
Assembly for storage
1. After thoroughly cleaning and drying each part, insert the male meatus into the foreskin as shown
below. Pull the male meatus from the other end of the skin layer to adjust.
Caution
The urethra parts are composed of soft resin; please handle gently.
Page 11
After training
After training
Assembly for storage
2. Set the bladder reservoir into the bladder holder. Tightly press down the reservoir into the holder to leave no spaces on the bottom between the reservoir and frame.
3. Insert the projection at the bottom of the skin
in the lower hole of the bladder holder.
No space Remaining
4. Attach the urethral tube to the valve.
space
Caution
Take note of the direction of the tube and orient it so that the projection on the tube faces upward. Ensure to place the tube securely over the connector of the valve.
5. Then insert the urethral tube to the valve, and put the clip over it, holding the clip by tucking small wings together with ngers.
valve
valve
urethral tube
Page 12
After training
Assembly for storage
6. Put the skin over the bladder holder and insert two upper projections of the skin in two upper holes of the bladder holder.
After training
7.
Set the bladder holder to the genital organ frame.
a) Set the lower edge of bladder holder to the
stopper on the frame. Then, holding the frame
rmly, push the holder forward.
c) Then engage the frame and holder by pushing the corner of the genital organ frame forward (to the abdomen side), until clicking sound is heard.
b) To t the upper side of the skin in the genital
organ frame, tuck the skin under the frame with
ngers.
Or, support the fasteners from the back side by ngers and draw the genital organ frame over it toward you.
10
Page 13
After training
After training
Assembly for storage
8. Connect the rectum unit. Set the rectum unit to the tube connector of the genitalia unit. Then set it on the hook.
Caution
Set the rectum unit with correct angle as shown in the picture.
9. Secure the rectum unit with the metal clip.
11
Page 14
Trouble shooting
Quick check-up before calling the customer service, when the you have problems in using the simulator. Look in the “cases” section for a description of the problem to find a possible solution.
Catheterization training
Trouble shooting
FAQ s
Cases
Water leakage
Water does not flow out when the catheter is inserted.
Valve is easily damaged
Reason Solution
The valve is worn.
Loose connection of the valve and bladder reservoir or meatus urethra.
Water in the bladder reservoir is low.
Clamps of the irrigation and/or drainage tubes are closed.
Using a non-genuine lubricant that is not included in the set.
Using a catheter larger than the recommended size.
Replace it with the new valve
Check the connection of the valve and bladder or meatus urethra, reconnect them tightly.
Ensure that the metal clips securing the bladder or meatus urethra are firmly in place.
Fill water into the bladder reservoir by using irrigation bottle.
Open the both clamps of irrigation and drainage tubes.
Use the lubricant included in the set.
Use a catheter of the recommended size of 14-16 Fr.
Enema Training
Cases Reason Solution
Glycerin enema tube can be inserted more than 7cm.
Water leakage
Connection of the genitalia organ unit and rectum unit is not correct.
12
Check the connection of the genitalia organ unit and rectum unit. Then set them steadily.
Ensure that the metal clips securing the genitalia organ and rectum unit are firmly in place.
Page 15
Page 16
Caution
Do not let ink from pens, newspapers, product manual or other sources contact the manikin. Ink marks on the manikin will be irremovable.
For inquiries and service, please contact your distributor or KYOTO KAGAKU CO., LTD.
Consumable parts
code
11381-300
11381-410
11381-510
11381-800
11381-300
name
replacement valve for MW2A/B (a set of 6)
replacement male skin
replacement male meatus (a set of 2)
lubricant
11381-510
metal clips are
not included
11381-800
11381-410
URL http://www.kyotokagaku.com e-mail rw-kyoto@kyotokagaku.co.jp
Main Office and Factory
15 Kitanekoya-cho Fushimi-ku Kyoto 612-8388, Japan Telephone : 81-75-605-2510 Facsimile : 81-75-605-2519
LA Office ( for USA , CANADA and Mexico customers )
3109 Lomita Boulevard, Torrance, CA 90505 ,USA
Telephone : 1-310-325-8860 Facsimile : 1-310-325-8867
The contents of the instruction manual are subject to change without prior notice. No part of this instruction manual may be reproduced or transmitted in any form without permission from the manufacturer. Please contact manufacturer for extra copies of this manual which may contain important updates and revisions. Please contact manufacturer with any discrepancies in this manual or product feedback. Your cooperation is greatly appreciated.
2015.05
Loading...