INSTRUCTIONMANUAL
Find Quality Products Online at: sales@GlobalTestSupply.com
DIGITALLIGHTMETER
LIGHTMETERSERIES
www.GlobalTestSupply.com
KEW5204
1. SAFETY WARNINGS
This inst r uc ti on m anu al c o nt ain s war ni ng s and s af et y r ul es w hic h m ust b e
obser ved by the user to ens ure safe operat ion of the instr ument and to
retain it in safe condition.
● Read through and understand instructions contained in this manual
before starting using the instrument.
● Save and keep th e man ual h and y to en abl e qui ck r efe ren c e whe neve r
necessary.
The symbol indicated on the instrument means that the user must refer
to related p a rts of the manu a l f or safe operat i on of the instr ument. Be sure
to carefully read the instructions following each symbol in this manual.
DANGER : is reserved for conditions and actions that are likely to
WARNING: is re served for c onditions and ac tions that ca n cause
CAUTION : is reserved for conditions and actions that can cause
● Do not install substitute parts or make any modification to the
instrument. Return the instrument to Kyoritsu or your distributor for
repair or re-calibration.
● Do not expos e the i nstr ume nt to th e dire ct sun , ext reme t empe ratu res
or dew fall.
●
This instrument isn't dust & water proofed. Keep away from dust and water.
● Be sure to set t he Func t io n Se le c to r Sw it ch t o th e " OFF " p os it io n after
use. When the instrument will not be used for a long period of time,
place it in storage after removing the batteries.
● Use a damp cl oth and deterge nt for cleanin g the instrum ent. Do not
use abrasives or solvents.
● The instrument should be calibrated once a year or so for the accurate
measurement
cause serious or fatal injury.
serious or fatal injury.
minor injury or instrument damage.
This symbol is marked where the user must refer to the
instruction manual so as not to cause personal injury or
instrument damage.
This inst ru ment s at isf i es t he ma rk in g req uir eme nt def i ne d in th e
WEEE Directive. This symbol indicates separate collection for
electrical and electronic equipment.
WARNING
WARNING
CAUTION
DISTRIBUTOR
Kyoritsu reserves the rights to change specifications or designs described
in this manual without notice and without obligations.
5. Measurement
5-1 Zero adjust
Please perform zero adjustment before measurement.
Press 0 ADJ button wit h the Lens Protec t ing Cover atta ched to the Light
Sensing Probe.
"CAP-" will auto-appear on the LCD display. After "CAL-" appears, the
meter enters the automatic zero adjustment. When automatic zero
adjustment is completed, "CAL-" disappears and "0.0"appears. After
confirming the "0.0" display, remove the cap to start measuring.
IfER R1appears, check the cap to make sure it is on properly.
After t he measurement ha s been complete d, press the power bu tton
again to tur n off the powe r, and cover t he light-det ecting sur face with
the cap for protecting against stray light.
5-2 Measurement
Place the l ight sensing prob e parallel to the me asuring objec t and
measure it . At this time , please be c areful that yo ur own shadow d oes
not hit the sensing probe.
The light sensing probe can be fixed on the meter parallel or
perpendicular to th e LCD Disp lay. Or it may be detached from th e unit to
fit the measuring requirement.
REF. LEVEL

2. Features
●Wide Range 0.1〜199900 lux
●MAX/MIN function
●Automatic Zero Adjust function
●Large LCD Display with Backlight
⑩Battery cabinet cover
When battery capacity segment is less
than one, the user should replace the
battery. The battery cabinet can be
opened w ith a screw drive r and two AA
type batteries should be installed.
3. Name of Parts and Position/Operation
①Lens Protecting Cover
This is used to protect the
photoc ell f rom sc ratc hin g and
create complete dark
environment for the meter to
perform dark current
calibration.
Light Sensing Probe/Photocell
②
③LCD Display
④Power button
Find Quality Products Online at: sales@GlobalTestSupply.com
Push this button to turn on
the meter. Please push the
button again to turn it off. The
unit will automatically turn
itself off after 3 0 mi nu tes i f n o
key operation. To disable the
automatic p ower off func tion,
turn the meter off, push and
keep holding button, then push power button to turn on.
⑤Backlight button
Push this bu tt on to tu rn on a nd of f th e LCD bac kli ght. T he bac kl ight w ill
be automatically off after 30 seconds to save battery power.
⑥HOLD button
The user may hold the present and keep it on the LCD display by
pressing the button.
⑦Max/Min button
Push this button to set the meter working under MAX /MIN mode. By
pushing t h i s b utton, the user may swap the LCD readi n g among curren t ,
maximu m and minim um reading . Push and ho ld this key for 3 s econds
www.GlobalTestSupply.com
to exit MAX/MIN mode.
⑧Range selector button.
The user may u se this button to se lect the prefer red measuring o r
automatically selected by the meter. The available ranges are: AUTO
200, 2000, 20000, 200000 lx
⑨0 ADJ button (Please refer to Measurement)
4. Specifications
MODEL KEW 5204
Light detector Silicon photo diode
Measuring Range 0.0〜199,900 lx
Range 199.9/1,999/19,990/199,900 lx
Resolution 0.1 lx
Accuracy ±4%rdg±5dgt(23℃±2℃)
Angle deviation from cosine
characteristics
Temperature characteristics
Relative spectral sensitivity
characteristics
Response time
Auto power off 30 min.
Dimensions 169(L)x63(W)x37(D)mm
Weight 210g approx..
Power source R6(AA)(1.5V) x 2
Operation Temperature/
Humidity
Storage Temperature/
Humidity
Continuous measuring time Approx. 80 hours
Applicable Standards
Accessories
10° ±1.5%
30° ±3%
60° ±10%
80° ±30%
±5 % (at 23℃ reference and a range of 0 to 40℃)
Deviation from spectral luminous
efficiency:9% or less
Auto range:5s or less
Manual range:2s or less
0℃〜40℃,80%RH or less(without
condensation)
-10℃〜60℃,70%RH or less(without
condensation)
JIS C 1609-1:2006
IEC61326
Carrying case
Instruction Manual
R6(AA)(1.5V)×2
ADEQUATE LIGHT LEVELS FOR YOUR WORKING OR AT YOUR WORK AREAS.
LUXES (lx)
OFFICE
FACTORY
SCHOOL
HOSPITAL
STORE
RESTAURANT
THEATRE
CONCERT HALL
INN, HOTEL
BEAUTY SALON
BARBAR SHOP
3,000 2,000 1,500 1,000 750 500 300 200 150 100 75 50 30 20 15 10 7
Entrance hall (day-time)
Director room
Design room
Drawing room
Drawing room
Operating room
Emergency room
Important part of display,
Register, Packaging stand
Sample case Register Stair
Front desk
Counter, Office
Hair dressing, Hair dyeing
Hair set, Make-up
Most important
part of display
Extra fine visual work
Inquiry office, Dining
room
Night-duty room
Council room
Electronic computer room
Reception room
Guard station
Classroom, Faculty room,
Dining room, Gymnasium
Night-duty room
Clothing classroom,
Electronic computer room
Experimental laboratory
Reading room
Healthcare room
Consultation room
Nurse station
Pharmacist office
Pharmacist's office
General examination room
Physiological examination
room, Radioisotope room
Funeral chapel
X-ray room
Physiotherapy room,
Exercise machine room
Night-duty room
Endoscopy room, X-ray
fluoroscopy room
Lift lobby, Escalator
Negotiation room Stair
Cookroom
Dining table
Counter
Ticket office
Stall, Green room
Monitoring room
Carriage porch
Dining room
Cuisine
Guest room desk
Hair cut, Shaving
Dressing
Shampooing, Register
Stair
Library, Tearoom
Dressing room
Toilet, lavatory
Control room,
Electric room
Air conditioning machine
room, lavatory, Toilet
Auditorium, Assembly
room
Locker room
lavatory, Toilet
Nursery room
Waiting room
Visiting room
Passage for outpatients
Changing room, lavatory
Hospital charts room
Toilet
Reception room
lavatory, Toilet
Waiting room
Guest room
lavatory, Toilet
Spectator seats, Lobby
Electric room
Mechanical room
lavatory, Toilet
Banqueting hall, Salle
Lobby, lavatory, Toilet
Toilet inside the shop Passage
Warehouse, Lift, Passage
Stair
Entrance
Passage Path
Warehouse
Stair
Passage
Ward Emergency staircase
Passage, Resting room
Porch, Passage
Entrance
Resting room
Projection room
Passage
Stair
Theatre basement work
room
Stair Important parts of garden
Game room, Guest room,
Entrance, Passage,
Bathroom
Undressing room
Stair
Indoor emergency
staircase
Indoor emergency
staircase
Ophthalmic dark room
Monitoring room
(during performance)
(JIS Z9110-2010)
Projection room
(during performance)