Kyoritsu Electrical Instruments 8122, 8121, 8123 User manual

INSTRUCTION MANUAL
1. Safety warnings
This instrument has been designed, manufactured and tested
according to IEC 61010: Safety requirements for Electronic Measuring apparatus, and delivered in the best condition after passing quality control tests. This instruction manual contains warnings and safety rules which have to be observed by the user to ensure safe operation of the instrument and to maintain it in safe condition.
  Therefore, read through these operating instructions before using the
KEW 8121 KEW 8122
KEW 8123
KEW 8122WP
CLAMP SENSOR
CLAMP SENSOR Series
KEW 8121/8122/8122WP/8123
KYORITSU ELECTRICAL INSTRUMENTS
WORKS, LTD.
Read through and understand instructions contained in this manual
# WARNING
before using the instrument.
Keep the manual at hand to enable quick reference whenever
necessary.
The instrument is to be used only in its intended applications. The operating instructions described in the manual must be
observed.
Understand and follow all the safety instructions contained in the
manual. It is essential that the above instructions are adhered to. Failure to follow the above instructions may cause injury and or instrument damage.
The symbol # indicated on the instrument, means that the user must refer to the related parts in the manual for safe operation of the instrument. It is essential to read the instructions wherever the # symbol appears in the manual.
# DANGER is reserved for conditions and actions that are likely to
cause serious or fatal injury.
# WARNING is reserved for conditions and actions that can cause
serious or fatal Injury.
# CAUTION is reserved for conditions and actions that can cause
minor injury or instrument damage.
DISTRIBUTOR
Kyoritsu reserves the rights to change specifications or designs described in this manual without notice and without obligations.
5. Operating instructions
# DANGER
Never make measurement on a circuit in which the electrical
potential exceeds AC300V when using KEW 8121 and AC600V
using KEW 8122/ 8122WP and 8123 in order to avoid possible
shock hazard.
The transformer jaws are made of metal and their tips are not
completely insulated. Be especially careful about the possible
shorting where the measured object has exposed metal parts.
# CAUTION
Take sufficient care to avoid shock, vibration or excessive force
when handling the instrument. Otherwise, precisely adjusted
transformer jaws will be damaged.
When transformer jaws do not fully close, never try to close them
by force, but make them free to move and try again. If a foreign
substance is stuck in the jaw tips, remove it.
Do not force to open the Transformer jaws when they are frozen.
When making current measurements, keep the transformer jaws
fully closed.
Otherwise, accurate measurements cannot be taken. Maximum conductor size is as follows.    KEW 8121 : 24mm in diameter    KEW 8122WP : 40mm in diameter    KEW 8123 : 55mm in diameter
Hold the inserting part (except for the cable) and disconnect the
Output connector from the measuring instrument so as not to cause
a break in the cord.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
# DANGER
Never make measurement on a circuit in which the electrical
potential exceeds AC300V when using KEW 8121 and AC600V using KEW 8122/ 8122WP and 8123.
Do not make measurement when thunder rumbling. If the
instrument is in use, stop the measurement immediately and remove the instrument from the equipment under test.
Do not attempt to make measurement in the presence of flammable
gasses.
  Otherwise, the use of the instrument may cause sparking, which
can lead to an explosion.
Use insulated protective gears, such as insulated gloves, for your
safety when possible electric shock hazards are concerned.
The transformer jaws are made of metal and their tips are not
completely insulated. Be especially careful about the possible shorting where the measured object has exposed metal parts.
Never use these sensors when their surface or your hand is wet. Do not wet the output connectors because they arent dust/ water-
proof.
Do not exceed the maximum allowable input of any measuring range.
# WARNING
Never attempt to make any measurement, if any abnormal
conditions are noted, such as broken case, and exposed metal parts.
Do not install substitute parts or make any modification to the
instrument.
  Return the instrument to your local Kyoritsu distributor for repair or
re-calibration in case of suspected faulty operation.
Always keep your fingers and hands behind the barrier on the
instrument to avoid the possible shock hazard.
Measurement procedures  (1) Connect the Output connector to the Input terminal of the
measuring instrument.
 (2) Press the Trigger to open the transformer jaws and clamp onto one
conductor.
   Clamped conductor should be at the center of the closed
transformer jaws.
 (3) Ensure that the tips of transformer jaws are firmly closed.
Conductor
# CAUTION
Do not step on or pinch the cord to prevent the jacket of cord from
being damaged.
The output connector shall be removed or connected without
clamping a conductor. Otherwise, it may cause a failure.
Do not expose the instrument to direct sunlight, high temperatures,
humidity or dew.
Never give shocks, such as vibration or drop, which may damage
the instrument.
Use a damp cloth and mild detergent for cleaning the instrument.
Do not use abrasives or solvents.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Safety symbols
# 
Measurement Category: To ensure safe operation of measuring instruments, IEC 61010
establishes safety standards for various electrical environments, categorized as O to CAT IV, and called measurement categories. Higher-numbered categories correspond to electrical environments with greater momentary energy, so a measuring instrument designed for CAT III environments can endure greater momentary energy than one designed for CAT II.
O : Circuits which are not directly connected to the mains power
CAT II : Electrical circuits of equipment connected to an AC electrical
CAT III : Primary electrical circuits of the equipment connected directly
CAT IV : The circuit from the service drop to the service entrance, and
Incoming wire
Refer to the instructions in the manual.
Indicates instruments with double or reinforced insulation.
Indicates that this instrument can clamp on live bare conductors when the voltage to be tested is below Circuit - Ground-to-Earth voltage against the indicated Measurement Category.
Indicates AC
supply.
outlet by a power cord.
to the distribution panel, and feeders from the distribution panel to outlets.
to the power meter and primary over-current protection device (distribution panel).
Interior wiring
Socket
Device which is
O:
not directly connected to the mains power supply
2. Features
Clamp sensor for recording apparatus
Designed to international safety standard IEC 61010-2-032 (Pollution
Degree 2) KEW 8121 CAT.III 300V KEW 8122 /8122WP /8123 CAT.III 600V
Designed to meet IEC 60529 IP 54. Possible to perform
measurements under adverse weather.(KEW 8122WP only)
3. Instrument layout
Transform Jaws Barrier Trigger Cable Output connector
4. DIN plug pin assignment
* Lower figure shows the pin assignment seeing the Clamp sensor from output
connector part. The figure of the pin assignment of connection terminal is symmetrical to lower figure.
6. Specifications
Model KEW 8121 KEW 8122 /8122WP KEW 8123
Rated current AC100Arms(141Apeak) AC500Arms(707Apeak) AC1000Arms(1414Apeak)
Output voltage
Measuring range AC0 ~ 100A AC0 ~ 500A AC0 ~ 1000A
Accuracy (Input: sine wave)
Temperature & humidity range (Guaranteed accuracy)
Operating temperature range
Storage temperature range
Maximum permissible input
Output impedance Approx. 9.5Ω Approx. 1.9Ω Approx. 1.5Ω
Location for use Altitude up to 2000m, Indoors
Applicable standards
Withstand voltage
Insulation resistance
Conductor Size Approx.24mm in diameter (max.) Approx.40mm in diameter (max.) Approx.55mm in diameter (max.)
Dimension 97(L)×59(W)×26(D)mm 128(L)×81(W)×36(D)mm 170(L)×105(W)×48(D)mm
Cable length Approx. 2m
Output connector MINI DIN 6PIN
Weight Approx. 150g
Accessories
Option
Find Quality Products Online at: sales@GlobalTestSupply.com
AC0 ~ 500mV (AC500mV/AC100A):5mV/A
±2.0%rdg±0.3mV(50/60Hz) ±3.0%rdg±0.5mV(40 ~ 1kHz)
23±5ºC, relative humidity: 85% or less (no condensation)
0 ~ 40ºC, relative humidity: 85% or less (no condensation)
-20 ~ 60ºC, relative humidity: 85% or less (no condensation)
AC100Arms continuous(50/60Hz) AC500Arms continuous (50/60Hz) AC1000Arms continuous (50/60Hz)
IEC 61010-1, IEC 61010-2-032 Measurement CAT. III (300Vrms) Pollution degree 2 IEC 61326-1(EMC), EN 5081(RoHS)
AC3470Vrms (50/60Hz)for 5 sec. between Jaw and enclosure between enclosure and output connector between Jaw and output connector
50MΩ or greater at 1000V between Jaw and enclosure between enclosure and output connector between Jaw and output connector
www.GlobalTestSupply.com
AC0 ~ 500mV (AC500mV/AC500A):1mV/A
±2.0%rdg±0.3mV(50/60Hz) ±3.0%rdg±0.5mV(40 ~ 1kHz)
IEC 61010-1, IEC 61010-2-032 Measurement CAT. III (600Vrms) Pollution degree 2 IEC 61326-1(EMC), EN 50581(RoHS) IEC 60529 : 1999 IP54 (KEW 8122 WP only)
AC5160Vrms (50/60Hz)for 5 sec. between Jaw and enclosure between enclosure and output connector between Jaw and output connector
KEW 8122 Approx. 260g / KEW 8122WP Approx.270g
Instruction manual
MODEL 7146 (Banana ΦA4 adjuster plug)
Cable marker
MODEL 7147 (Extension Cable)
*Not available for KEW 8122WP
AC0 ~ 500mV (AC500mV/AC1000A):0.5mV/A
±2.0%rdg±0.3mV(50/60Hz) ±3.0%rdg±0.5mV(40 ~ 1kHz)
Approx. 360g