Instruction Manual
Non-Contact Safety Phase Indicator
KEW 8035
1. Safety Warnings
This instrument has been designed, manufactured and tested
acco rding to foll ow in g st an da rds an d de li vered in the best
condition after passing quality control tests.
● IEC61010-1 Measurement CAT.III 1000V/CAT.IV 600V
Pollution degree 2
● IEC61010- 031
This instruction manual contains warnings and safety rules which
have to be observed by the user to ensure safe operation of the
instrument and to maintain it in safe condition. Therefore, read
through these operating instructions before using the instrument.
WARNI NG
● Read through and understand instructions cont ained in this
manual before using the instrument.
Keep the manual at hand to enable quick reference whenever
●
necessary.
●
The instr ument is to be used only in its intended applications.
● Understand and follow all the safety instructions contained
in the manual.
It is essential that the above instructions are adhered to.
Failure to follow the above instructions may cause injury and
or instrument damage.
The symbol # indic ated on the ins tr ume nt, means that the
user must refe r to th e re la te d par ts in th e ma nu al for safe
operation of the instrument.
It is e ss ent ial to read the instr uc tions wher eve r th e symbo l
appears in the manual.
DANGER is re ser ved for conditions and act ions that are
#
likely to cause serious or fatal injury.
WARNI NG is reserved for c onditions and actions that can
#
cause serious or fatal Injur y.
CAUTION is reserved for conditions and actions that can
#
cause minor injur y or instr ument damage.
● Confir m a pr oper operation of the instrument with a we llknown power supply.
● Warning LED might not light up at live status. (earth potential
70V or less). Never touch the wire.
● Voltages may exist when the Live LED is f lashing (indicating
Earth phase). Never touc h with the wires.
● Never make mea suremen t on a circuit in w hic h the ear th
potential exceeds 100 0V to avoid electrical shocks.
● Do not make measurement when thunder is rumbling. If the
instrument is in use, stop the measurement immediately and
remove the instrument from the measured object.
● Do not attempt to make measu rem ent in the presence of
flammable gasses. Otherwise, the use of the instrument
may cause sparking, which can lead to an explosion.
● Keep your fingers and hands behind the barrier on the
instrument to avoid the possible shock ha zard.
#
DANGER
#
● Put insul ate d protecti ve gears when there is a da nger of
electrical shock hazard.
● The ti p of cl ip is ma de of metal an d it is not c om pl etely
insulated. Be especially careful about the possible shorting
where the measured object has exposed metal parts.
● Never attempt to use th e instrument if it's surface or your
hand is wet.
Other wise, electrical shock accident may occur.
● Never open the Battery compartment cover and the
instrument case when making a measurement.
● The instrument is to be used only in its intended applications
or conditions. Otherwise, safety functions equipped with the
instrument doesn’t work, and instrument damage or serious
personal injury may be caused.
● Only the qual if ied p er so n can use the inst rume nt at the
secondary side of high voltage power receiving equipments.
WARNI NG
● Never attempt to make any measurement, if any abnormal
con dit ion s are noted, such as bro ken case, and exp ose d
metal par ts.
●
Do not install substi tute par ts or make any modifi cat ion to the
instrument. Return the instrument to your local Kyoritsu distributor
for repair or re-calibration in case of suspected fault y operation.
● Always keep your fingers and hands behind the barrier on
the instr ument to avoid the possible shoc k hazard.
● Do n ot try to repl ac e batteri es if the sur face o f the
instrument is wet.
Disco nnect the clips from the measu red condu ctors fi rst
and power of f th e instr ument before op ening the Ba tter y
compar tment c over for a battery replacement.
● Do not apply shocks, vibrations or excessive forces onto the
Measurement clips.
● Never force to open the Measurement clips when they are
frozen.
● This instrument can be operated with safe at temperatures
between -10ºC and 50 ºC and altitude up to 2000m.
● Keep away from dust and water.
● Precise measurement cannot be made near a charged body
or equipment generating electromagnetic waves.
● Measurable conductor size is between dia. 2.4mm and
30mm. Accurate measurem ents for conductors out of this
range cannot be made.
● Measured results are inf luenced by voltage wires on which
twice or more of the measured voltages exist near the point
to be clipped.
The clip point should be far from such voltage wires.
●
This instrument cannot identify wiring status correctly when an
earth line is connected bet ween phases via delta c onnection.
Check the connection specific ation of the measured object.
● Incapable of measuring bus bars or shielded wires. Clip onto
a covered conductor and make a measurement.
●
All the clips should be clipped onto the covered wires and
make measurements. Otherwise, it may cause a malfunction.
#
CAUTION
#
● Do not touch the Clips during measurements to get accurate
results.
● This instrument cannot find the missing line of the earth line.
Do not pull the cable when removing the Measurement clips
●
from the measured conductors. It may cause a break in cable.
● Power off the instr ument af ter use. Remove the bat ter ies
if the instrument is to be stored and will not be in use for a
long period.
● Do not expose the instrument to direct sunlight, hightemperature and humidity or dew.
● Dry and store the instrument if it is wet.
● Do not step on or pinch the cord to prevent the jacket of
cable from being damaged.
● Bending or pulling the cord may cause a break in a cord.
● Neve r give shocks, su ch as v ibr ati on or d rop, which may
damage the instrument.
● Use a damp cloth and detergent for cleaning the instrument.
Do not use abrasives or solvents.
Safety Symbols
Refe r to the instr uct ions in the manual to protec t th e
#
user and instrument.
Indicates instrument with double or reinforced insulation
AC
This instrument satisf ies the marking requirement defined
in th e WEEE Directive. This symbol indicates separa te
collection for electrical and electronic equipment.
○ Measurement Category
To ensure safe operation of measuring instruments, IEC 61010
establishes safety standards for various electrical environments, categorized as O to CAT.IV, and called measurement
categories. Higher-numbered categories correspond to electrical
environments with greater momentary energy, so a measuring
instrument designed for CAT.III environments can endure greater
momentary energy than one designed for CAT.II.
O : Circuits which are not directly connected to the mains
power supply.
CAT.II : Electrical circuits of equipment connected to an AC
electrical outlet by a power cord.
CAT.III : Primary electrical circuits of the equipment connected
directly to the distribution panel, and feeders from the
distribution panel to outlets.
CAT.IV : The circuit from the service drop to the service
entrance, and to the power meter and primary
overcurrent protection device (distribution panel).
Incoming wire
Socket
Interior wiring
Device which is
O:
not directly
connected to the
mains power supply
2. Features
● This is a Phase indic ator and c an indicate a presence of live
line and phase sequence with the equipped LED and Buzzer
while clipping the 3-phase line over the jacket of a conductor.
● Brightness switch is equipped to make the indication visible
in dimly lit areas.
● A ma gn et on the back si de of the inst rument can fix th e
instrument onto the distribution board and enables safety and
easy measurements.
● Designed to International safety standard IEC 61010-1 (CAT.
III 1000V/ CAT.IV, 600V, Pollution degree 2)
3. Instrument layout
Barrie
Measurement clip
Put the Label for clip,
if necessary.
(*5) Labels for clip aren't supplied
with KEW8035(EU).
Test leads
Live LED (L1, Phase-U)
Live LED (L2, Phase-V)
Live LED (L3, Phase-W)
Phase sequence LED (positive)
Phase sequence LED (reversed)
Rotation LED (positive)
Rotation LED (reversed)
Power switch
Power LED
Brightness switch
DISTRIBUTOR
Kyoritsu reserves the rights to change specifications or designs
described in this manual without notice and without obligations.
6-16 92-2009C
4. Specification
Model KEW8035
Measurement
principle
Voltage range
Frequency range
Operating
temperature &
humidit y range
Storage
temperature &
humidit y range
Location for use
Applicable
standards
Dust-proof IP40 (IEC60529
Withstand
voltage
Insulation
resistance
Power source
Auto-power-of f 10 min after powering on the instr ument
Low battery
warning
Current
consumption
Continuous use
Conductor size
Cable length approx 70cm
Dimension 112(L) × 61(W) × 36(D)mm
Weight approx 38 0g (
Accessories
(*1) without batteries
(*2) powers off automatically at 3±0.2V or less
(*3) stand-by state (will be doubled (max) at measurement)
(*4) stand-by state (will be 0.5 times at measurement)
(*5) Labels for clip aren't supplied with KEW8035(EU).
(voltage to earth, c ontinuous sine wave)
-10 to 50ºC, relative humidity 80% or less
-20 to 60ºC, relative humidity 80% or less
Altitude 2000m or less, Indoor use
Measurement CAT.III 1000V/CAT.IV 600V
between the tip of Measurement clip and enclosure
between the tip of Measurement clip and enclosure
(size AA alkaline bat tery LR6 or equivalent
Power LED flashes at 4.0 ±0.2V or less (*2)
Exter nal diameter of the covered conductor
Instruction manual, battery, Soft c ase, Label for Clip(*5)
Static induction
3-phase AC70 to 1000V
45 to 66Hz
(no condensation)
(no condensation) (*1)
IEC 61010 -1
Pollution degree 2
IEC61010 - 031
IEC 61326-1,2-2 (EMC standard)
IEC 61557-1,7
EN505 81 (RoHS)
AC688 0V (rms 50/60Hz) for 5 sec
10MΩ or more/ 1000V
DC6V
1.5V AA × 4pcs)
15mA (*3)
approx 20 0 hours (*4)
Dia.2.4 to 30mm
batteries included
)
)
5. Checks and Indications
5.1. Preliminary c hecks
5.1.1. Press the Power switch and power on the instr ument.
Then all the LEDs flash in order for about 1 sec. Confirm
all the LEDs ligh t up and f lash. Only the Power LED
keeps lighted up in about 1 sec later.
5.1.2.
Apex of “▼” mark on the Measurement clip shall indicate
the center of the measured conducto r. Connect each
Measurement clip to 3-phase line as follows: Red to L1,
Phase- U, White to L 2, Phase -V, Blue to L3, Phase -W.
Lines co nne cting the apexes of
“▼” mar ks sh ould pass thr ough
the center of the conductor.
Blue
or
Grey
White
or
Black
Red
or
5.1.3. Mea su re a cover ed condu ctor AC70V or more first
5.1.4. Pre sence of live wir es and phase seq ue nc e are
Brown
to confi rm eac h live LE D lights up . Do no t use the
instrument when any of the LED doesn’t light up.
informed by LED indication and Buz zer sound as soon
as connecting the clips.
Note:
The label of KEW8035(EU) shows the harmonised and also
the pre-harmonised cable colours (UK):
KEW8035(EU)
KEW8035(EU) shows also other alphanumeric identifications on
the label as:
A B C, R S T and U V W.
5.2. Live wire chec k
DANGER
● LEDs don’t light up when voltage to earth is 70V or less.
● Voltages may exist at Ear th phase.
● It is impossible to detect the missing phase of the earth line.
Earth line and phase sequence are indicated if the earth line
has a missing phase.
State Indication
Live
Missing line
Earth line
Earth line
(Delta
connection
Positive
phase
Reversed
phase
5.3. Use the Brightness sw itch to mak e the LED indi ca ti on
br ig ht er.
Brightness of all the LEDs (except for the Power LED) is
increased while pressing down the switch.
Phase with flashing LED is
live state.
LED doesn’t light up for
or
missing line or ear th line.
Phase with flashing LED is
an earth phase.
)
When the green Rotation
LED flashes in the order of
the direc tion indicated by
the arrow mark (clockwise),
the circuit under test has a
positive phase.
The buzzer sounds
intermittently. (pi-pi-pi
When the red Rotation
LED flashes in the order
of the direction indicated
by the arrow mark
(counterclockwise), the
circuit under test has a
reversed phase. The buzzer
sounds continuously. (pi---)
#
CAUTION
#
)
6. Battery Replacement
CAUTION
● Power off the instrument and remove the Measurement clips
from the measured object when replacing batteries to avoid
electrical shocks.
● Do not mix old and new batteries.
● Inst all batterie s in corre ct polarity as indicated inside the
case.
Use the same model of batteries from the same manufacturer.
●
Whe n th e Power LED on the fron t si de of t he instr ument is
flashing, batte ry voltage is low. Repla ce batteri es with new
ones to continue further measurements.
Low battery voltage may not affect measurement accuracies.
The instrument is powered off automatically when bat teries are
exhausted.
1) Loosen the screw fixing the Battery compartment cover.
2) Slide the Battery compartment cover downwards to remove
it.
3) Re place th e batteries with new ones . Fo ur size AA LR6
alkaline or equivalent 1.5V AA type batteries should be
used.
4) Install the Batter y compartment cover and tighten the screw.
Batter y compartment
cover
#
Screw
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com