Kyoritsu Electrical Instruments 8031F User manual

Incoming supply
Interior wiring
CAT.Ⅰ
Socket
CAT.
CAT.
Equipment
Distribution Board CAT.
Socket
#
WARNING
Read through and understand instructions contained in this
manual before starting using the instrument.
●Save and keep the manual handy to enable quick reference
whenever necessary.
Be sure to use the instrument only in its intended applications
and to follow measurement procedures described in the manual.
Be sure to understand and follow all safety instructions
contained in the manual. Failure to follow the above instructions cause injury instrument damage and/or damage to equipment under test.
The symbol
#
Indicated on the instrument means that the user must refer to related parts in the manual for safe operation of the instrument. Be sure to carefully read instructions following each
#
Symbol in this manual.
#
WARNING is reserved for conditions and actions that can
cause serious or fatal injury.
#
CAUTION is reserved for conditions and actions that can
cause injury or property damage.
Following symbols are used on the instrument and in the instruction manual. Attention should be paid to each symbol to ensure your safety.
Refer to this instructions in the manual.
#
This symbol is marked where the user must refer to the instruction manual so as not to cause personal injury or instrument damage.
Indicates an instrument with double or reinforced insulation.
#
WARNINGS
1. Never open the instrument when making measurements.
2. If the instrument shows the following conditions, do not try to make measurements and have the instrument checked for inspection or repair.
a) Instrument shows visual damage. b) Test leads are damaged. c) Instrument can not be operated for intended measurements. d) Instrument has been stored for a long period of time
under abnormal conditions.
e) Instrument has been subjected to severe shocks and
vibrations.
3. High voltage may be present with high energy levels.
Connect the leads to the system carefully with fingers behind
all finger guards.
4. If the all open phase lamps are not lit on, any phase may still be live-be carefull.
#
CAUTIONS
1. Make sure to never apply a voltage more than 600V ACrms. between the test leads of the instrument and earth to avoid damage to the instrument.
2. Do not measure for more than five minutes when measuring on 500V AC or more, although the instrument is designed for the use 110V through 500V AC.
Measurement categories (Over-voltage categories)
To ensure safe operation of measuring instruments, IEC61010 establishes safety standards for various electrical environments, categorized as CAT. to CAT., and called measurement categories. Higher-numbered categories correspond to electrical environments with greater energy, So a measuring
instrument designed for CAT. environments gives greater
protection than one designed for CAT.. C AT. : Equipment connected to a mains outlet socket C AT. : Fault category level at a mains socket C AT. : Fault category at distribution board level C AT. :
The circuit from the service drop to the power meter and primary over-current protection device (distribution panel).
FEATURES
Two Functions in One Unit   KEW8031F is designed to check phase sequence. Lamps
provided on the unit will also tell you if a phase is open.
Highly Reliable   Can check a wide range of 3-phase power source from 110V to
500V. Sealed against dust, the unit ensures trouble-free
performance.
INSTRUCTION MANUAL
PHASE ROTATION TESTER
KEW8031F
DESIGNATIONS
SAFETY WARNINGS
This instrument has been designed and tested according to IEC Publication 61010; Safety Requirements for Electronic Measuring Apparatus. This Instruction manual contains warnings and safetyrules which must be observed by the user to ensure safe operation of the instrument and retain it in safe condition. Therefore, read through these operating instruction before using the instrument. The voltage values used in this manual or on the meter are given in line voltage, unless otherwise specified.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
(3) Make sure that all of the three lamps for phase check are on. If so, there is no open phase. When any of the three lamps is
Not on there is open phase.
Open phase check Open phase on terminal Lamp "L1" is not on where Red alligator clip is connected.
Open phase check Open phase on terminal Lamp "L2" is not on where Whight alligator clip is connected.
Open phase check Open phase on terminal Lamp "L3" is not on where Blue alligator clip is connected.
* When the open phase check lamps are not on the rotating disc
does not turn.
(4) Check the rotating direction of the inside disc through the
phase sequence indication window.
* When the rotating disc turns counter-clockwise alternate the
connection of the two of the three alligator clips. Then, the rotating disc will turn clockwise.
* When the rotating disc turns clockwise phase sequence is
L1,L2 and L3 in order of the power source terminals where the Red, Whight and Blue alligator clips are connected.
Please Note : New European harmonized phase colours are
as follows.
Red = Brown   Protective Earth = Green yellow
Whight = Black  Neutral = Blue
Blue = Grey
Functional Design  Small, Lightweight and portable. Designed for maximum ease of
operation and ruggedness.
Safety Design   No exposed metal parts. Safety features are incorporated
including the instant push button switch operation.
SPECIFICATIONS
Standard : IEC 61010-1 CAT.Ⅲ 300V Pollution degree 2 IEC 61557-1,7 Voltage : 110V to 500V Frequency : 50Hz / 60Hz Withstand Voltage : 5,550V AC for one minute Time Limit for Continuous Use : Within 5 minutes in case voltage
is above 500V. Dimensions : 106 (L) x 75 (W) x 40 (D) mm Weight : Approx. 350g Fuse : 0.5A / 600V (F) Accessories : Instruction Manual, Carrying Case IP cord : IP30 (IEC60529)
OPERATING INSTRUCTIONS
(1) Connect colour coded alligator clips or prods to the terminals
of a 3-phase power source where a rotating electrical machine such as a motor will be connected or input to a building.
(2) Press the push switch button located on top of the unit. Keep this button pressed during phase sequence or open
phase check. When the push switch button is released it immediately goes off.