ocedure below and set time on the instrument before use.
1) Install the KEW LOG SOFT 2 and the USB Driver according to the
instructions written in the Install Manual for KEW LOG Soft 2.
2) Run KEW LOG SOFT 2 once install is completed. (Refer to 4. Start
KEW LOG SOFT 2 in the Install Manual.)
3) Confirm that the LOGGER and a PC is firmly connected with a USB
Cable and click Time Synchronizing.
4) Confirm that the Logger connected with the PC is listed on the
List of detected loggers. (The USB Driver might not be installed
correctly when a message No Logger is detected. is displayed.
Refer to the backside of the Cautions for installing USB Driver
install or 6. Trouble shooting (P.8) of the Install Manual and
re-install the USB Driver.)
5) Click Time Synchronizing Button while the LOGGER is powered on
and checks are marked in the box for each connected LOGGER to
set time.
Read this manual and do your desirable settings. (Setting should be done
by using KEW LOG Soft 2. Settings for some Recording modes can be
done on the LOGGER.
Read through 1. Safety Warnings in this manual before using the
LOGGER in order to ensure safe operation of the instrument.
This in stru men t has be en des ign ed, ma nufactu red a nd te sted
according to IEC 61010: Safety requirements for Electronic Measuring
apparatus, and deli vered in the best cond ition after passed the
inspection. This instruction manual contains warnings and safety rules
which must be observed by the user to ensure safe operation of the
instrument and retain it in safe condition. Therefore, read through these
operating instructions before using the instrument.
WARNING
Read through and understand the instructions contained in this
manual before using the instrument.
K
eep the manual at hand to enable quick reference whenever
necessary.
The instrument is to be used only in its intended applications.
Understand
and follow all the safety instructions contained in the
manual.
It
is essential that the above instructions are adhered to. Failure to
follow the above instructions may cause injury, instrument damage
and / or damage to equipment under test.
The
symbol
indicated on the instrument means that the user must
refer to the related parts in the manual for safe operation of the
instrument. It is essential to read the instructions wherever the
symbol appears in the manual.
DANGER is reserved for conditions and actions that are likely
to cause serious or fatal injury.
WARNING is reserved for conditions and actions that can cause
serious or fatal injury.
CAUTION is reserved for conditions and actions that can cause
Never make measurement on a circuit in which voltage over AC300V
exists.
Do no t ma k e m e a s ure m e n t wh i l e t h u n der is r u m b ling. Stop
measurement immediately and disconnect the instrument from the
object under test.
D
o not attempt to make measurement in the presence of flammable
gasses. Otherwise, the use of the instrument may cause sparking,
which can lead to an explosion.
T
ransformer jaw tips are designed not to short the circuit under test.
If equipment under test has exposed conductive parts, however, extra
precaution should be taken to minimize the possibility of shorting.
Never attempt to use the instrument if its sur
face or your hand is wet.
Do not exceed the maximum allowable input of any measuring range.
Never open the Battery Cover during a measur
ement.
V
erify proper operation on a known source before use or taking action
as a result of the indication of the device.
WARNING
Never attempt to make measurement if any abnormal conditions, such
as broken case and exposed metal parts are found on the instrument.
Do not install substitut e part s or make a ny modific ation to the
instrument. For repair or re-calibration, return the instrument to your
local KEW distributor from where it was purchased.
D
o not try to replace the batteries if the surface of the instrument is wet.
D
isconnect all the cords and cables from the object under test and
power off the instrument before opening the Battery Cover for Battery
replacement.
CAUTION
Put the instrument on a stable place where is free from vibrations or
shocks.
Keep away Floppy Disks, Mag Cards, PCs and Displays from the
magnet, which is attached to the backside of the instrument.
Do not expose the instrument to the direct sun, high temperature and
humidity or dewfall.
Be
sure to power off the instrument after use. When the instrument will
not be in use for a long period, place it in storage after removing the
batteries.
U
se a cloth dipped in water or neutral detergent for cleaning the
Indicates a Instrument with double or reinforced insulation
Indicates that this instrument can clamp on live bare conductors
when the voltage to be tested is below Circuit – Ground-to-Earth
voltage against the indicated Measurement Category.
〜
Indicates AC
Indicates DC
Measurement categories(Over-voltage categories)
To ensure safe operation of measuring instruments, IEC 61010 establishes
safety standards for various electrical environments, categorized as
CAT.I to CAT.IV, and called measurement categories. Higher-numbered
categories correspond to electrical environments with greater momentary
energy, so a measuring instrument designed for CAT.III environments can
endure greater momentary energy than one designed for CAT.II.
CAT.I : Secondary electrical circuits connected to an AC electrical
outlet through a transformer or similar device.
CAT.II : Primary electrical circuits of equipment connected to an AC
electrical outlet by a power cord.
CAT.III : Primary electrical circuits of the equipment connected directly to
the distribution panel, and feeders from the distribution panel to
outlets.
CAT.IV : The circuit from the service drop to the service entrance, and
to the power meter and primary over-current protection device
(distribution panel).
KEW 5020 is a Data Logger capable of measuring Leakage Current,
Load Current and Voltage. (KEW5010 is for measuring Leakage Current
and Load Current).
Following sensors are used for measuring current & voltage.
* Leak/Load Current Clamp Sensor KEW 8146/8147/8148
* Load Current Clamp Sensor KEW 8121/8122/8123
* Voltage Sensor KEW 8309 (for KEW5020)
TRMS measurement & recording of AC Current (50/60Hz) & AC Voltage
(50/60Hz, only on KEW5020)
LED to indicate pre-set current/ voltage value is exceeded.
(Trigger/ Capture Recording, Power Quality Analysis modes)
Can store 60,000 data when using 1ch, and when using all 3ch, can
store 20,000 data at each channel. (Normal recording mode)
Data will not be lost at battery replacement or at low battery voltage as it
is stored in nonvolatile memory.
Long time recording is available by a use of external AC Adaptor
(opti onal part s) Installed ba tteries en ables back up at a temporal ly
electricity failure. Size AA Alkaline Battery: about 10-day measurement.
Can transfer the recorded data to PC via USB cable.
Pr
otected throughout by double (reinforced) insulation
This
instrument provides 3 Recording modes and Power Quality Analysis
mode (only KEW5020). Various insulation monitoring are available with 4
recording modes. Understand characteristic of each recording mode and
select an appropriate mode.
auto-power-off function.
Menu: Recall(CALL) P.39
Recorded qty., Max value reference, Recall
Normal recording mode
Trigger recording mode
Capture recording mode
Power Quality Analysis mode [KEW5020 only]
PC data in transit
Warning of memory clear
Error (improper Sensor is connected under
Power Quality Analysis mode/ KEW5020 only) (P.47)
All the Memory block are occupied. (P.16)
Move the data and clear the memory. (P.34)
Range Configuration
Range configuration varies depending on each connected Sensor.
Refer to "12. Specification" (P.45) in this manual.
●
Auto-ranging
Auto-ranging is avai lable only in Normal Recording mode. It isn't
available on a Sensor providing one single range.
(e.g. Voltage Sensor "KEW8309" for KEW5020)
●
Range Hold
Select the ranges with
mark to fix measurement range under
Normal Recording mode.
Measurement range in Trigger/ Capture Recording mode is decided
depending on a detection level. Switching of measurement ranges
with the button isn't possible.
●
Filter Function
Lowpass Filter works when Filter Function is ON
(
mark lights up)
and filter out the frequency in higher harmonics band.
(Cutoff Frequency: approx 160Hz, attenuation rate -24dB)
●
Range/ Filter Function for each Recording mode
Recording modeFunction
Normal recording
* Switching ranges, On/Off Filter
* Auto-ranging is available.
T
rigger/ Capture
recording
* On/Off Filter
* Range corresponds to detection level
* Auto-ranging isn't available.
1. Press the button and power off
th e ins tru men t. Co nne ct a C lam p/
Voltage Sensor to CH1 (for multiple
connection, starting from CH1) while
the instrument is being off. Connect
them firmly with careful attention to the
orientation of the connector.
2. Press the button at least 1 sec.
to power on the instrument. Release
the button when all indic ations are
displayed on the LCD.
Fir s t, ti me is d i spla y e d, an d th en
sensors being connected are displayed.
If incorrect time is displayed whenever
powering on the instrument, internal
battery for clock may be exhausted. In
this case, send back the instrument to
your local KEW distributor for repair.
3. Check the connected Sensors.
Sensor (P.7) and the full-scale value are
displayed.
(beginning from CH1)
◆
Sensor mark is displayed on the CH
No,. to which a Sensor is connected .
◆
Only CH1 is used in Capture REC mode.
◆
Voltage sensor is connected to CH1 in
PQA REC mode (only on KEW5020).
◆
Voltage sensor ca nnot be used wit h
KEW5010.
◆
Power off the instrument and repeat
steps from 1. to re-connect a Sensor.
4. Can ma ke measurement right afte r
powering on the instrument.
* Each time pressing the button,
CH1 to CH3 switches. When
(non-connect) is displayed on the
LC D, a sensor is not connec ted
to t h e selec ted ch anne l; or the
connection is incorrect.
Check the mark indicating the selected
recording mode. Refer to 5. Recording
modes and conditions in this manual
to change the record ing mode or
condition ( Recording interval / Preset
current)
Recording modeDetailsRefer to
Normal
Performs c on tinuous measurement /
recording at the pre-set intervals.
(15-kind: 1 sec. to 60min.)
P.19
Trigger
Records 8 data (0.8 sec.) in total with
time information when pre-set detection
level is exceeded.
P.22
Capture
Records 10 (50Hz) to 12 waveforms
(60Hz) when preset detection level is
exceeded.
(Graph can be viewed on PC)
P.24
(PC setting)
Power Quality Analysis
(only 5020)
Records date and time information when
preset Swell, Dip and Short interruption
value are exceeded.
but ton to c ha nge the m.
Independent setting of Range & Filter is
available on each CH.
Point
●
At Normal Re cording mod e, Range and
Filter switches each t ime pre ssing the
button.
Auto-ranging is also available.
●
At Trigger/Capture/Power Quality Analysis
mo des, o nly Fil ter on/ off is a vaila ble.
Range is selected automatically based
on the detection level (preset value for
Trigger/ Capture Recording). See Setting1
"SEt.1".
●
Se t tin g c hang e isn't a v ail a ble du ring
recording; setting should be done prior to
a recording.
mark me ans Ra n g e H o l d Func t i o n is
activated.
mark mea ns Aut o-rangi ng F unction is
activated.
5
. Wh en only t he le ftmost segment of the
Ba tte ry ma rk is f las hin g on th e LCD, it
means battery voltage is low.
Repl a c e t h e bat t e r i es wi t h new ones.
Wh en no thi ng is di spl aye d on the L CD,
the batteri es are exhaust ed. Replace the
batteries with new ones.
* Settings cannot be changed du ring recording. Ca reful ly check the
settings prior to a recording.
* Prev ious recorded d at a i s c leared and new recording starts when
changing the followings and starts recording. (In this case, "CLr" flashes on
the LCD at a start of recording.)
1) Recording mode is changed.
2) CH for Sensor is changed.
3) Sensor type is changed.
* Transfer important data to your PC first and clear Memory (P.33) since
further recording isn't activated when all the 3 blocks in Memory Block
(P.15) are full.
* Follow the procedure below to start recording.
1. Press down the
button at least 4 sec.
"
" or "CLr" mark flashes. Keep the button
pressed down until "
" mark lights up.
Recording stops when the button is released
while the mark is flashing.
2
. Measured value is displayed and "
" mark
lights up, then measurement starts.
◆
Additional three recordings (Memory Block /
see the next page) are possible.
In this case, one-time is automatically selected
and recording stops when memory becomes
full.
◆
It is recommended to clear the memory before
st arti ng to rec ord if pre viou s data i s not
necessary.
One memory block is used from the start to the end of recording.
◆
3 blocks in the Memory Block are available; it enables 3 different
recordings at different locations.
◆
Transfer important data to your PC first and clear Memory (Setting 2
or via PC) since further recording isn't activated when all the 3 blocks
in Memory Block are used.
◆
Settable Location No. at Setting 2 "SEt.2" is saved in the Memory
Block together with the measured data; and is useful to recognize the
location where the data recorded on your PC.
◆
At Trigger/ Capture/ Power Quality Analysis modes, zero is saved in
Memory Block even when no data is recorded from the start to the end
of recording.
Operation during recording
Following operations are available during recording.
◆
Displaying the measured value at each CH →
button
◆
Recording status : Displaying Max recorded valued & RECALL
Refer to 9. Confirmation of recorded data in this manual.
◆
Check of preset values at Setting 1"SEt.1" & Setting 2 "SEt.2".
Following operations aren't available during recording.
◆
Powering off the instrument
◆
Changing the measurement range.
◆
Change of preset values at Setting 1"SEt.1" & Setting 2 "SEt.2".
In case that One-time is set to "ON" at Step2: Confirmation and change of
preset value, recording is stopped automatically when Memory becomes
full.
1. Press the button at least 1 sec
to stop recording.
2. Recording stops, and the "" mark
di sap pea rs. T hen the i nst rum ent
goes back into measurement state.
Now, recording completes.
* Press the button at least 1 sec to power off the instrument.
* It is recommended to transfer the important data to a PC.
* The recorded data can be checked on the instrument.
(Refer to : 9. Confirmation of recorded data (CALL) in this manual)
* Recording mode and condition can be set on the instrument ; but PC
software "KEW LOG Soft 2" provides much easier setting.
* Only Normal and Trigger Recording modes can be set on the instrument.
Capture recording and Power Quality Analysis modes need to be set via a
PC. (Recording condition for the Capture Recording mode is changeable
on the instrument.)
●
Setting on the instrument (Menu Mode)
1. Power on the instrument, and press the
button
to enter into Menu Mode.
2
. Select and set desirable recording mode and condition at "SEt.1 on the
Menu. Setting with the instrument is mentioned on the next page.
3. Press the button at "End" on Menu Screen or press the
button and exit from the Menu Screen to return measurement status.
Max recording time is dependent on battery life (approx 10-day with size
AA Alkaline battery) Use of optional AC Adopter is recommended for long
time recording.
Using all 3 channelsUsing 2 channelsUsing only 1 channel
20,000 data30,000 data60,000 data
Recording
interval
Using all 3 channelsUsing 2 channelsUsing only 1 channel
Can be selected from; 1, 2, 5, 10, 15, 20, 30 sec.,
1, 2, 5, 10, 15, 20, 30, 60 min
◆
Press the button and proceed to the
next step when not changing the setting.
◆
To change the setting, press the
bu tton. T hen the i ndi catio n on th e LCD
flashes.
Press the or button to set
the value to the desired one. Then press the
button to confirm it.
5. Next, One-time/ Endless is indicated.
One-time: Re cordi ng s top s whe n mem ory
becomes full.
Endless : Overwrite the old data, and store
the latest data.
◆
Press the button and proceed to the
next step when not changing the setting.
◆
To change the setting, press the
bu tton. T hen the i ndi catio n on th e LCD
flashes.
Press the or button to set
the value to the desired one. Then press the
button to confirm it.
6. Now Setting 1 is complete; "End" is displayed on
the LCD. Press the button to return to
the screen on which "SEt.1" is displayed.
7. Press the button to get the instrument
ready for a measurement.
Using all 3 channelsUsing 2 channelsUsing only 1 channel
1,600 data2,400 data4,800 data
ItemRange for Setting
Default
CH1 Detection Level
0 ~ 1000 (Unit depends on Sensor)
15
CH2 Detection Level
0 ~ 1000 (Unit depends on Sensor)
15
CH3 Detection Level
0 ~ 1000 (Unit depends on Sensor)
15
One-time/ Endless
One-time/ Endless
Endless
▼
▼
Triggerrecordingmode:Currentsetvalueof15A
1. Connect a Sensor first prior to do setting.
When Sensor is not connected to the instrument
at setting, connect a Sensor, and then power
off once and on the instrument.
2. Press the button while instrument is on to
enter into Menu mode.
3. Pre ss th e but t o n wh e n "S E t .1" is
displayed on the LCD.
4. The selected recording mode is displayed.
◆
Wh en (Trig ger r eco rdi ng mod e) is
di splay ed o n the LCD , press the
button to proceed to the next setting.
◆
In case that
, or is displayed
on the LCD, press the button. Then
the indication on the LCD flashes. Press the
To change the setting, press the button.
Then the indication on the LCD flashes.
◆
Press the or button to set the
value to the desired one. Then press the
button to confirm it.
◆
When using Leakage Current Sensor, pressing
the button at 1000m A in d i c a t i o n
changes to 2A, pressing the button at
2A indication changes to 1000mA (1A)
Point
Further detection doesn't start when a value falls
to 50% or less (for voltage sensor: 1% or more)
of the detected level. Set the appropriate value
based on the test prior to recording.
6. Confirm or change the preset current value on
channel 2 and 3 as well.
7. Next, One-time/ Endless is indicated.
One-time: Re cording s top s w hen m emor y
becomes full.
Endless : Overwrite the old data, and store
the latest data.
◆
Press the button and proceed to the
next step when not changing the setting.
◆
To change the setting, press the button.
Then the indication on the LCD flashes.
Pr ess th e or b utto n to set
the value to the desired one. Then press the
Sampling period and RMS calculation
Sampling the inputs at every 1.6ms/CH and calculate the measured
value (RMS) at every 100ms. The Peak value (crest value in sampling
data) will be updated and kept.
●
Recor
ding
Average of the measured value is recorded at every recording interval.
RMS measured values, Max, Min and Peak values (sampled crest
value converted to sine RMS value) are recorded at every 10 times of
recording.
●
Display of measured value
Average of RMS value is displayed on the LCD at every 1 sec.
(10 data)
Detection and RMS value calculation
Sampling the inputs at every 1.6ms and comparing the Peak value
(sampled crest value converted to sine RMS value) and Detection level.
For current, when the Peak value exceeds the level and for voltage
when the Peak value is below the level; detection is triggered. RMS
value is calculated at every 100ms.
●
Recording
When the Detection level is exceeded, 8 data (for about 0.8 sec)
Including:
-
3 RMS values prior to the cross over point
- RMS value at the cross over point
- 4 RMS values subsequent to the cross over point
are recorded with time information. In case that events exceeding the
Detection level occur continuously, next detection doesn't start until it
drops to 50% or less of the detected value.
(for voltage, 5% or more than the detected value)
●
Display of measured value
Average of RMS value is displayed on the LCD at every 1 sec.
(10 data)
LED on the Sensor connected channel blinks when following Trigger
is detected.
<On KEW5010> Trigger is detected when the measured values
exceed the pre-set detection level under Trigger/ Capture Recording
Mode.
<On KEW5020> Trigger is detected when the measured current
values exceed the pre-set detection level under Trigger/ Capture
Recording Mode and Power Quality Analysis Mode.
For volt age value s, Trigge r is det ected whe n it is b elow the
detection level.
●
LED flashes whenever the measured values exceed the pre-set
detection level during measurements.
●
LED blin ks ever y 4 sec during recordings after one de tection
occurs. (When Auto-power off is OFF, it blinks every 2 sec.)
●
To restore the blinking LED during recordings, first press the
button once and enters into Menu mode. Then press the
button to return Measurement mode.
Detection and RMS value calculation
Sampling the inputs at every 0.55ms only at CH1 and comparing the
Peak value (sampled crest value converted to sine RMS value) and
Detection level. For current, when the Peak value exceeds the level
and for voltage when the Peak value is below the level; detection is
triggered. RMS value is calculated at every 100ms.
●
Recor
ding
When the preset current value is exceeded, instantaneous values
with corresponding time information are recorded for 200ms ( 10 or
12 waveforms) including 50ms prior to and subsequent to the cross
over point. In case that events exceeding the Detection level occur
continuously, next detection doesn't start until it drops to 50% or less
of the detected value. (for voltage, 5% or more than the detected
value)
●
Display of measur
ed value
Average of RMS value is displayed on the LCD at every 1 sec.
(10 data)
(Waveform display is only available on PC. Use the enclosed Software
and transfer the data to your PC.)
Detection and RMS value calculation
Sampling the inputs at every 0.55ms only at CH1 and comparing the
Peak value (sampled crest value converted to sine RMS value) at every
10ms and preset value to detect swell, dip and short interruption in
power supply. RMS value is calculated at every 100ms.
●
Recor
ding
When swell, dip or short interruption is detected, the detected value is
recorded as Start (S) with time and date information.
In case that a input becomes c lose to th e reference volt age by
hy steresi s value , it i s recorde d as E nd ( E) w ith tim e and dat e
information.
●
Display of measur
ed value
Average of RMS value is displayed on the LCD at every 1 sec.
(10 data)
The PC software "KEW LOG Soft 2" provides easy calculation of integral
power consumption by use of current and voltage (only on KEW5020)
recorded under Normal Recording mode.
●
On KEW 5010, measures and records current values, and then enter
any voltage value and power factor on "KEW LOG Soft 2" to calculate
simplified power consumption.
●
On KEW 5020, simplified power calculation is available by using Voltage
and Current Sensors; apply any power factor to the recorded voltage and
current values via "KEW LOG Soft 2".
Calculation with any voltage value as same as KEW5010 is also possible.
Example of Connection
◆
Single-phase 2-wire (1Φ2W)
Model NameCH1CH2CH3
KEW LOG Soft2
Fixed parameters
KEW
5010A-
-----Voltage value, Power factor
KEW 5020
A-
-----Voltage value, Power factor
VA-
--Power factor
* Can be connected with 3 systems via each channel (1 through3).
7
. Simplified Power Integration
(available via KEW LOG Soft 2)
additional recordings (up to 3 Memory Blocks),
it is recommended to set each Location No.
in advance. It is useful to identify the recorded
locations per Memory Block.
● Location No. is linked to the Location list and
allows to display the Locatio n n ame, whic h
co rre spo ndin g to the Lo cat ion No ., whe n
displaying data on PC Software.
● In ca s e of se t t i ng Loc a t i o n no . on the
instrument, it is recommended to take notes of
the Location No. and the name.
5. "Auto-power-off":
State of Auto-power-off function is displayed.
On : Enables Auto-power-off function.
OFF : Disables Auto-power-off function.
◆
Press the button and proceed to the
next step (Time) when not changing the setting.
◆
To change the setting, press the button.
Then the indication on the LCD flashes. Press
the or button to set the value to
the desired one. Then press the button
to confirm it.
Point
"
" ma rk ap pea rs on the L CD wh en th is
Function is "OFF". Power off the instrument after
use. Instrument powers off about 3 min after the
last Key operation when the Function is "ON".
LCD indication disappears during recording due
to Power Save Function while recording is being
performed.
3. Press the Button while "CALL" is displayed
on the LCD.
4
. Check of <Recorded qty.>
Percentage of recorded data agains t memory
capacity is displayed.
* Memory block with recorded data is indicated
by the Marks:
❶❷❸
. When all 3 marks are lit
up, all the Memory blocks are used. Transfer the
important data to a PC and clear the Memory to
perform next recording.
Pressing the
Button proceeds to the
next.
5
. Display of the number of the recorded data at
CH1/ the number of detected current & voltage
data in case of Trigger Recording mode.
◆
Pressing
Button moves to CH2.
◆
Press the
Butt on to ch eck max, min
and peak values on CH1 with time and date
inform at ion. See P.42 <<MAX, MIN , Instant
PEAK/ detected value>>.
◆
Pre ss the
Button un til the indi cation
becomes "5:RCL" for RECALL and proceed to
step 8.
6. Display of the number of the recorded data at
CH2/ the number of detected current & voltage
data in case of Trigger Recording mode.
Pressing the Button moves to CH3.
◆
Pr ess th e Butto n to ch eck ma x ,
mi n a n d pe ak valu e s wi th time an d d a te
information. See P.42 <<MAX, MIN, Instant
PEAK/ detected value>>.
7. Display of the number of the recorded data at
CH3/ the number of detected current & voltage
data in case of Trigger Recording Mode.
Pressing the Button moves to "RECALL".
◆
Pre s s t h e Bu t ton to check ma x ,
mi n a n d pe ak valu e s wi th time an d d a te
information. See P.42 <<MAX, MIN, Instant
PEAK/ detected value>>.
8. <RECALL> checks the latest 10 recorded data.
Press the But ton to c heck the late st
recorded data and preceding 9 data, [R01] to
[R10]. Refer to P.42 <<RECALL>>
9. That's all for CALL Function to call the recorded
da ta. Pre ssi ng t he B utt on retu rns t o
the window with "CALL" message; pressing the
Install PC software "KEW LOG Soft 2" on your PC to enable data
communication between the instrument and your PC. Refer to the
HELP contained in "KEW LOG Soft 2" which shows how to install the
software. It will be on the Desktop after installing the software, or
found in the folder of "KEW" from "Start" → "Program".
●
When connecting the logger to PC for the first time, your PC will
recognize this new hardware and install the USB driver.
Follow the instructions described in the instruction manual for "KEW
LOG Soft 2" and install it on your PC.
10-1 Connection of USB cable
(1) Connect the USB cable to the available USB port of PC.
(2) Connect the other end of USB cable to the USB terminal on the
right side of this instrument.
Note:
Remove the protective cover
of USB terminal carefully, and
connect a cable to it. When the
cover is damaged, it may cause
poor contact due to dust, etc.
(1) Power on the instrument, and get the instrument ready for a
measurement.
( No te: Data can not be tr ansfe rred whil e the ins tru ment is
performing a recording.)
(2) Start up the PC software: KEW LOG Soft 2.
10-3 Operation of PC software
Refer to the supplied instruction manual for "KEW LOG Soft 2" or
"Help" and transfer the data to your PC. The PC may not detect
the connected Logger or error message is displayed during data
transfer, even if the PC and the Logger are connected correctly
because of static electricity.
In this case, a message is displayed on the PC screen. Disconnect/
connect the USB cable once accordingly, and transfer the data
again.
10-4 Multiple connections
By using commercially available USB hub, multiple Loggers can be
connected to your PC.
◆
With PC software "KEW LOG Soft 2", the data can be transferred
to PC by selecting one Logger from the list of detected Logger.
You do not have to connect and disconnect a USB cable one by
one.
◆
Synchronization via 「KEW LOG Soft 2」 enables synchronized time
setting on multiple LOGGERs at once.
In order to avoid electrical shock, remove sensors from the instrument
when replacing batteries.
CAUTION
Do not mix new and old batteries.
In st all bat terie s in the ori entat ion as s hown i nside the bat te ry
compartment, observing correct polarity.
When only the leftmost segment of the Battery mark is flashing
on the LCD, it means battery voltage is low. Replace the batteries with
new ones. There is no influence on the measurement accuracy even
if this warning mark is flashing. Nothing even the Battery mark is
displayed on the LCD if the batteries are completely exhausted.
Replace the batteries with new ones.
(1) Loosen two Battery-cover -fixing
screws on the backside of the
instrume n t an d re m o v e the
cover.
LOGGER is automatically powered off in about 3 min after the last
key operation while Auto-power off Function is "ON" at Setting 2. This
function doesn't work during recording (LCD with
mark) although
readings aren't displayed on the LCD due to Power-saving Function.
In case of using an external power supply for making measurement,
disable the Auto-power off Function. Then long period measurement is
possible. (Be sure to power off the LOGEER after use.)
2) External power supply AC Adopter (optional)
WARNING
Use MODEL 8320 (S-8458 or TAS2903 AC Adopter made by KAGA
COMPONENTS co) only.
Use the Power cord supplied with the AC Adopter.
Confirm
the voltage of power supply and the rated voltage of the AC
Adopter, and then connect the Power Cord.
Disconnect the Power cord of the AC Adopter from the outlet when the
LOGGER isn't used for a long period.
Do not put any heating object or others on the AC Adopter or Power
Cord.
Hold the Plug part (other than cord) when disconnecting the Power
Cord from the outlet in order to prevent a break in the cord.
Specification of AC Adopter
Rated supply voltage, Frequency : AC100V / 240V,50/60Hz
Supply voltage, Range of Frequency variation
: 90~264VAC,45~66Hz
Rated output voltage of AC Adopter
: DC9.0V
Rated max Output curr
ent of AC Adopter : 1.4A
Use the optional AC Adopter for long period recordings. Installing batteries
in the Case located at the backside of the unit provides power at temporary
electricity failure. Be sure to check the battery level in advance.
Battery mark with full level will be displayed while using the AC Adaptor.
Disconnect the Adaptor to check the battery voltage level.
When defect or breakdown of the instrument is suspected, check the
following points first. If your problem isn't listed in this section, contact your
local Kyoritsu distributor.
Symptom / Check
1. Unable to measure.(
is displayed.)
Check if a sensor is connected to the pr
operly.
Insert the sensor into the connector terminal firmly.
KEW 5010 doesn't recognize the voltage Sensor
2. Unable to measure. (
is displayed.)
Sensor
other than Voltage Sensor may be connected to CH1 under
the setting for Power Quality Analysis mode for KEW 5020.
Only the Voltage Sensor can be used under Power Quality Analysis
mode.
3
. Inaccurate readings
Check
if a clamp sensor is connected properly.Insert the sensor into