This instrument has been designed, manufactured and tested according to
IEC 61010: Safety requirements for Electronic measuring apparatus, and
delivered in the best condition after passed the inspection. This instruction
manual contains warnings and safety rules which must be observed by the
user to ensure safe operation of the instrument and retain it in safe
condition.
Therefore, read through these operating instructions before using the
instrument.
#WARNING
● Read through and understand instructions contained in this manual
before starting to use the instrument.
● Save and keep the manual at hand to enable quick reference whenever
necessary.
● The instrument is to be used only in its intended applications.
● Understand and follow all safety instructions contained in the manual.
Failure to follow the instructions may cause injury, instrument damage
and/or damage to equipment under test. Kyoritsu is by no means liable
for any damage resulting from the instrument in contradiction to this
cautionary note.
The symbol # indicated on the instrument means that the user must refer
to related parts in the manual for safe operation of the instrument. It is
essential to read the instructions wherever the # symbol appears in the
manual.
#DANGER is reserved for conditions and actions that are likely to
cause serious or fatal injury.
#WARNING is reserved for conditions and actions that can cause
serious or fatal injury.
#CAUTION is reserved for conditions and actions that can cause
● Never make measurement on the circuit in which electrical potential to
ground over 600V exists.
● Do not attempt to make measurement in the presence of flammable
gasses. Otherwise, the use of the instrument may cause sparking,
which can lead to an explosion.
● Never attempt to use the instrument if its surface or your hand is wet.
● Be careful not to short-circuit the power line with the metal part of the
test leads when measuring voltage. It may cause personal injury.
●
Do not exceed the maximum allowable input of any measurement range.
● Do not press the Test button with test leads connected to the
instrument.
● Never open the battery compartment cover while making a
measurement.
● The instrument should be used only in its intended applications or
conditions. Otherwise, safety functions equipped with the instrument
do not work, and instrument damage or serious personal injury may be
caused.
● Verify proper operation on a known source before use or taking
action as a result indication of the instrument.
#WARNING
● Read through and understand instructions contained in this manual
before starting using the instrument.
● Save and keep the manual handy to enable quick reference whenever
necessary.
● The instrument is to be used only in its intended applications. Understand and follow all the safety instructions contained in the
manual.
Failure to follow the instructions may cause injury, instrument damage
and/or damage to equipment under test. Kyoritsu is by no means liable
for any damage resulting from the instrument in contradiction to this
cautionary note.
● Never attempt to make any measurement, if the instrument has any
structural abnormality such as cracked case and exposed metal part.
Please refer to following explanation of the symbols used on the instrument
and in this manual.
User must refer to the explanation in the instruction manual.
#
Danger of possible electric shock
Instrument with double or reinforced insulation
Earth terminal
Protection against wrong connection is to 440V
Crossed-out wheel bin symbol (according to WEEE
Directive: 2002/96/EC) indicating that this electrical
product may not be treated as household waste, but
that it must be collected and treated separately.
Measurement categories(Over-voltage categories)
To ensure safe operation of measuring instruments, IEC 61010 establishes
safety standards for various electrical environments, categorized as o to
CAT IV, and called measurement categories.
Higher-numbered categories correspond to electrical environments with greater
momentary energy. So a measuring instrument designed for CAT III environments
can endure greater momentary energy than one designed for CAT II.
O(None, Other)
CAT II : Primary electrical circuits of equipment connected to an AC
CAT III : Primary electrical circuits of the equipment connected directly
CAT IV : The circuit from the service drop to the service entrance, and
: Circuits which are not directly connected to the mains power
supply.
electrical outlet by a power cord.
to the distribution panel, and feeders from the distribution
panel to outlets.
to the power meter and primary overcurrent protection device
(distribution panel).
The influencing variations used for calculating the operating error are
denoted as follows:
Temperature : 0Cº and 35 Cº
Supply voltage : 6.4V to 10.4V
Position : Reference position ±90°
※Prior to measurement, apply 0-Adjutment at each position.
○ AC Voltage Warning:
Warning range0-600V
Accuracy±5% of scale length
Input impedance1.2MΩ
● Typical Number of measurements (Central tendency for supply voltage
up to 6.0V)
Insulation Resistance Ranges:1MΩ at 1000V Ranges Approx. 1,400 times min.
0.5MΩ at 500V Ranges Approx. 3,500 times min.
0.25MΩ at 250V Ranges Approx. 5,500 times min.
Continuity Test (Resistance Test) Ranges:1Ω at 3Ω Range Approx. 1,500 times min.
● Applicable Standards
IEC 61010-1, -2-030 Measurement voltage CAT III 600V
1000V Range 1200V (DC+AC p-p) for 10 seconds
500V Range 600V (DC+AC p-p) for 10 seconds
250V Range 300V (DC+AC p-p) for 10 seconds
Continuity ranges:
3Ω/500Ω Range 280V (DC+AC p-p) for 10 secondsAC Voltage Warning:
1200V (DC+AC p-p) for 10 seconds
● Dimensions: 106(L) x 160(W) x 72(D) mm approx..
● Weight: 560g approx. (lncluding batteries)
● Power Source: 6 x 1.5V battery type R6P, 1.5V AA or equivalent
● Accessories: Test Probe MODEL7122B x 1 set
Pouch for test probe x 1
Shoulder strap x 1
R6P batteries x 6
Spare fuse F500mA/600V x 1
Check that the pointer lines up with the middle of the mark on the scale
correctly.
If not, adjust it by rotating the meter movement zero adjust with a
screwdriver, etc.
5-2 Battery Voltage Check
① Set the range selector switch to BATT. CHECK position.
② Press the test button.
③ Then the pointer deflects. Judge the battery status with BATT.GOOD
mark on the scale plate.
If the pointer does not move to BATT.GOOD mark, the batteries are
exhausted. Replace them with new batteries according to section 8 for
battery & fuse replacement.
5-3 Test Probe Connection
Insert the test probe fully into the terminal of the instrument.
Connect the earth clip of the test probe (black) to EARTH terminal and the
line probe (red) to LINE terminal of the connector terminal.
5-4 Test Probe Check
Set the range selector switch to 3Ω position and press and turn the test
button to lock it down.
When the test probe are connected together, the pointer should move
from the position towards the 0 position on the blue ohms scale. If not,
the probe or fuse may be faulty.
Release the test button after completion.
●
When the test button is pressed with the range selector switch in the
megaohm range position, take care not to touch the tip or the test probe
where a high voltage is present in order to avoid possible shock hazard.
● Do not keep the test button pressed or locked by turning it clockwise
④ Check the circuit under test is not energized as follows. Connect the test probe to the circuit under test and read a voltage value.
If the circuit is live, the meter indicates the voltage, the live circuit lamp
is lit, and warning buzzer sounds.
If the meter indicates 0V, the circuit is dead.
⑤ Press test button, Read the scale directly for the 500V range, multiply by
0.5 for 250V and by 2 for 1000V.
● Continuous Measurement
A lock down feature is incorporated on the test button. Pressing
and turning it clockwise, lock the test button in the continuous
operating position.
To release the lock turn the test button counterclockwise.
Never leave the test button locked down when not is use.
#DANGER
Be extremely careful not to get electric shock during insulation resistance
measurement as high voltage is present on the tip of the test probe
continuously.
⑥ With the test probe still connected to the circuit under test after testing,
release the test button to discharge capacitance in the circuit.
● Automatic Circuit Capacitance Discharge Function
This function allows the capacitance stored in the circuit under test
to be automatically discharged after testing.
Discharge can be monitored by the voltage warning range.
#DANGER
Do not touch the circuit under test immediately after testing.
Capacitance stored in the circuit may cause electric shock.
Leave the test probe connected to the circuit and never touch the circuit
until the discharge is completed.
● Never press the test button if the live circuit warning lamp is lit or the
warning buzzer sounds. This may damage the circuit.
● In case that an additional operating circuit connected in parallel to the
circuit under measurement, the measurement error might be caused
due to the effects of impedance of the circuit connected in parallel or
transient current.
① Set the range selector switch to the desired position 3 ( or 500).
② Short the line probe (red) and the earth clip of the test probe (black) and
press the test button. Adjust the ohm zero adjust to zero the pointer on
the scale.
③ Connect the test probes to the circuit under test.
④ Check the circuit under test is not energized as follows. Connect the test probe to the circuit under test a voltage value. If the circuit is live, the meter indicates the voltage, the live circuit lamp
is lit, and warning buzzer sounds, If the meter indicates 0V, the circuit is
dead.
⑤ Press test button. Read the blue ohm scale directly.
● Continuous Measurement
A lock down feature is incorporated on the test button.
Pressing and turning it clockwise, lock the test button in the
continuous operating position.
To release the lock turn the test button counterclockwise.
Never leave the test button locked down when not is use.
● Principle of Continuity Testing (Resistance Test)
Resistance value can be obtained by
applying a certain current to the
resistance under test and measuring
the voltage generated on the both
sides of the resistance under test.
◎ Cleaning the meter cover This tester is managed by our companys quality standard and is
delivered in the best condition after passed the inspection. But in the dry
time of winter static electricity sometimes builds up on the meter cover
due to the characteristic of plastic.
When the pointer deflects by touching the surface of this tester or zero
adjustment can not be made, do not try to make measurement.
When static electricity builds up on the meter cover and affects the
meter reading, use a cloth dampened with off-the shelf anti-statics agent
or detergent to wipe the meter cover surface.
If this tester should fall to operate correctly, return it to your nearest
distributors starting the exact nature of the fault.
Before returning the unit, make sure that:
a) operating instructions have been followed
b) Leads have been inspected
c) Fuse has been checked
d) Battery has been checked
e) The unit is returned with all accessory leads
Remember, the more information written about the fault, the quicker it will
be serviced.