This instrument has been designed, manufactured and tested according to IEC
61010: Safety requirements for Electronic measuring apparatus, and delivered in
the best condition after passed the inspection. This instruction manual contains
warnings and safety rules which must be observed by the user to ensure safe
operation of the instrument and retain it in safe condition. Therefore, read
through these operating instructions before using the instrument.
WARNING
#
● Read through and understand the instructions contained in this manual
before using the instrument.
● Keep the manual at hand to enable quick reference whenever necessary.
● The instrument is to be used only in its intended applications.
● Understand and follow all the safety instructions contained in the manual.
It is essential that the above instructions are adhered to. Failure to follow
the above instructions may cause injury, instrument damage and/or damage
to equipment under test. Kyoritsu is by no means liable for any damage
resulting from the instrument in contradiction to these cautionary notes.
○ The symbol
to the related parts in the manual for safe operation of the instrument. It is
essential to read the instructions wherever the symbol appears in the manual.
DANGER :is reserved for conditions and actions that are likely to
#
WARNING :is reserved for conditions and actions that can cause
#
CAUTION :is reserved for conditions and actions that can cause injury
#
indicated on the instrument, means that the user must refer
Please refer to following explanation of the symbols used on the instrument and
in this manual.
User must refer to the explanations in the instruction manual.
#
Caution, risk of electric shock.
Instrument with double or reinforced.
Protection against wrong connection is up to 440V.
EARTH GROUND
This instrument satisfies the marking requirement defined in the WEEE
Directive 2002/96/EC. This symbol indicates separate collection for
electrical and electronic equipment.
# DANGER
●
Confirm a proper operation of the instrument with a well-known power supply.
● Do not use this instrument on energized (LIVE) circuits.
●
Do not attempt to make measurement in the presence of flammable gasses.
Otherwise, the use of the instrument may cause sparking, which can lead
to an explosion.
● Do not make measurement when thunder is rumbling. If the instrument is
in use, stop the measurement immediately and remove the instrument
from the measured object.
● Never attempt to use the instrument if its surface or your hand is wet.
● Never open the battery compartment cover and the instrument case when
making a measurement.
● Do not exceed the maximum allowable input of measuring ranges.
●
The instrument is to be used, only in its intended applications or conditions.
Otherwise, safety functions equipped with the instrument doesn't work,
and instrument damage or serious personal injury may be caused.
● Always Keep your fingers and hands behind the protective fingerguard on
Measurement categories (Over-voltage categories)
To ensure safe operation of measuring instruments, IEC 61010 establishes
safety standards for various electrical environments, categorized as O to CAT IV,
and called measurement categories.
Higher-numbered categories correspond to electrical environments with greater
momentary energy, so a measuring instrument designed for CAT III environments
can endure greater momentary energy than one designed for CAT II.
O : Circuits which are not directly connected to the mains power supply.
CAT II : Primary electrical circuits of equipment connected to an AC
CAT III : Primary electrical circuits of the equipment connected directly to
CAT IV : The circuit from the service drop to the service entrance, and to the
electrical outlet by a power cord.
the distribution panel, and feeders from the distribution panel to
outlets.
power meter and primary overcurrent protection device (distribution
panel).
O: Device which is
not directly
connected to the
mains power supply
MODEL-3131A is an analogue insulation tester with five ranges for insulation
resistance measurement and continuity testing (resistance tests) of low voltage
installations.
● Designed to safety standards: IEC61010-1 IEC61010-031 IEC 61557-1,2,4,10
● Dust and drip proof constrution to IP54
● Three insulation test ranges : 250V/100MΩ, 500V/200MΩ, 1000V/400MΩ
● Two continuity test ranges : 2Ω, 20Ω
● Back light function to facilitate work at night or dimly lit locations
● Easy for battery check
● Power-on indication LED lighting during battery check and measurement
● Live circuit warning indication with LED lighting and the buzzer sound
● Fuse protected.
● Color coded scales and range switch position for easy reading
● Uses only 6× 1.5V battery type R6, AA or equivalent
Measuring range to keep operating instrumetal uncertainty
0.1MΩ - 10MΩ
0.4MΩ - 40MΩ
Maximum percentage operating instrumetal uncertainty
±30%0.2MΩ - 20MΩ
Continuity Test (Resistance Test) Ranges(IEC 61557-4)
Ranges
x1Ω
x10Ω
The influencing variations used for calculating the operating instrumental uncertainty are
denoted as follows:
Temperature : 0℃ and 35℃
Supply voltage : 6.5V to 10.4V
Position : Reference position ±90°
※Prior to measurement, apply Zero Adjustment at each position
Measuring range to keep operating instrumetal uncertainty
0.2Ω - 2Ω
2Ω - 20Ω
Maximum percentage operating instrumetal uncertainty
Typical Number of Measurements (central tendency for supply voltage up to
Strap belt x 1
Test probe pouch x 1
R6P AA batteries x 6
Instruction manual x 1
6.5V)
Insulation Resistance Ranges:
1000V/400MΩ Range Approx. 500 times min.
500V/200MΩRange Approx. 1300 times min.
250V/100MΩ Range Approx. 1800 times min.
Continuity Test (Resistance Test) Ranges:
x1Ω Range/ x 10Ω Range Approx. 1000 times min.
Check that the pointer lines up with the middle of the mark on the scale
correctly. If not, adjust it by rotating the meter movement zero adjust with
a screwdriver, etc.
5-3 Battery Voltage Check
① Set the range selector switch to BATT. CHECK position. ② Press the test button. ③ Then the pointer deflects. Judge the battery status with BATT.GOOD
mark on the scale plate.
If the pointer does not move to BATT.GOOD mark, the batteries are
exhausted. Replace them with new batteries according to section 8 for
battery & fuse replacement.
5-4 Test Probe Connection
Insert the test probe fully into the terminal of the instrument. Connect the earth clip of the test probe (black) to EARTH terminal and
the line probe (red) to LINE terminal of the connector terminal.
5-5 Test Probe Check
Set the range selector switch to x1Ω position and press and turn the Test
button to lock it down. When the test probe are connected together, the
pointer should move from the ∞ position towards the 0position on the
blue continuity scale. If not, the probe or fuse may be faulty. Release the
Test button after completion.
# CAUTION
Do not keep the test button pressed or locked by turning it clockwise during
battery check.
6-1 Disconnection and check of power source of circuit under test
DANGER
#
● To avoid possible electrical shock, do not perform measurements on
energized (LIVE) circuits.
● Never make measurements with the battery compartment cover removed.
● Always Keep your fingers and hands behind the protective fingerguard on
test probe to avoid the possible shock hazard.
CAUTION
#
● Never press the test button if the live circuit warning LED is lit or the
warning buzzer sounds. This may damage the circuit.
Voltage check can be made with the range selector switch at any position.
Be sure to turn off the breaker for the circuit under test.
① Connect the earth clip of the test probe (black) to the earth side and the line
probe (red) to the line side of the circuit under test.
② Ensure that the live circuit warning LED is not lit and the audible warning
is not present. If the LED is lit and the buzzer sounds, never press the test
button. Voltage is generated in the circuit under test. Recheck that the
breaker for the circuit under test is "OFF".
● Always test the circuit or equipment to ensure it is surely de-energized
before measurement according to the instruction of 6-1.
To avoid electrical shock, measurements must be performed on de-
●
energized circuits only.
When the test button is pressed with the range selector switch in the
●
insulation position, take care not to touch the tip of the test probe and the
circuit under test where a high voltage is present in order to avoid possible
shock hazard.
●
Never make measurement with the battery compartment cover removed.
● Always Keep your fingers and hands behind the protective fingerguard on
test probe to avoid the possible shock hazard.
# CAUTION
● Never press the test button if the live circuit warning LED is lit or the
warning buzzer sounds. This may damage the circuit.
Conduct the voltage warning check before measurement to ensure that
the circuit under test is de-energized.
① Check the voltage which can be applied to the circuit under test and set
the range selector switch to the desired nominal output voltage range.
② Connect the earth clip of the test probe (black) to the earth terminal of the
circuit under test.
③ Put the tip of the line probe (red) to the circuit under test and press the
test button.
④ Read the scale directly for the 500V range, multiply by 0.5 for 250V and
Do not touch the circuit under test immediately after testing. Capacitance
stored in the circuit may cause electric shock.
Leave the test probe connected to the circuit and never touch the circuit until
the discharge is completed.
Principle of Insulation Resistance Measurement
Resistance value can be obtained by applying a certain high voltage to the
resistance (insulation resistance) and measuring the flowing current.
Resistance Value = Voltage/Current RX = V/I
14
RESISTANC
VALUE:RX
CURRENT:I
—
VOLTAGE:V
ARTH(+)
6-3 Continuity Testing (Resistance Tests)
DANGER
#
● Always test the circuit or equipment to ensure it is surely de-energized
before measurement according to the instruction of 6-1.
● To avoid electrical shock, measurements must be performed on de-
energized circuits only.
Never make measurement with the battery compartment cover removed.
●
● Always Keep your fingers and hands behind the protective fingerguard on
test probe to avoid the possible shock hazard.
# CAUTION
● Never press the test button if the live circuit warning LED is lit or the
warning buzzer sounds. This may damage the circuit.
Conduct the voltage warning check before measurement to ensure that
the circuit under test is de-energized.
● In case that an additional operating circuit is connected in parallel to the
circuit under measurement, the measurement error might be caused due
to the effects of impedance of the circuit conneted in parallel or transient
current.
① Set the range selector switch to the desired position x 1Ωor x 10Ω.
② Short the line probe (red) and the earth clip of the test probe (black) and
press the test button. Adjust the ohm zero adjust to zero the pointer on
the scale.
③ Connect the test probes to the circuit under test and press the test button.
④ Read the scale directly for x 1Ωrange, multiply by 10 for x 10Ωrange.
Fig.6
Principle of Continuity Testing (Resistance Test)
Resistance value can be obtained by applying a certain current to the
resistance under test and measuring the voltage generated on the both
sides of the resistance under test.
Resistance Value = Voltage/Current RX = V/I
6-4 Continuous Measurement
A lock down feature is incorporated on the test button. Pressing and
turning it clockwise, lock the test button in the continuous operating
position.
To release the lock turn the test button counterclockwise.
# DANGER
Be extremely careful not to get electric shock during insulation resistance
measurement as high voltage is present on the tip of the test probe
continuously.
To facilitate working at night or dimly lit situations, a back light function is
provided which illuminates the display.
To operate this function, the back light button must be pressed and released
while pressing the test button. The back light continues illuminating for
approx. 40 seconds and then turn off automatically.
When the test button is released, the back light will turn off even within the
lighting time.
This instrument is managed by our company's quality standard and is
delivered in the best condition after passed the inspection. But in the dry time
of winter static electricity sometimes builds up on the meter cover due to the
characteristic of plastic.
When the pointer deflects by touching the surface of this instrument or zero
adjustment can not be made, do not try to make measurement.
When static electricity builds up on the meter cover and affects the meter
reading, use a cloth dampened with off-the-shelf anti-static agent or detergent
to wipe the meter cover surface.
11. Service
If this tester should fail to operate correctly, return it to your nearest
distributors stating the exact nature of the fault.
Before returning the unit, make sure that:
a) probes have been checked.b) fuse has been checked.c) battery has been checked.
Remember, the more information written about the fault, the quicker it will be
serviced.